Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 1 of 49
General information for DR 5000 and DR 6000 when using Spectroquant® Test kits from Merck. Wavelength specification of the instrument: 190 to 1100 nm Photometric range: ± 3.0 A Standard cells which can be used: 16 mm (only US version), 13 mm (only European version), check your instrument before ordering the cell tests 10, 50 mm, 1-inch round and 1-inch rectangular cells (use the sample cell adapters coming with the instrument) Important: All Merck Spectroquant® cell tests have a diameter of 16 mm and can only be used in the DR 5000 and DR 6000 US-version. Our Spectroquant® round
tubes 16 mm are not applicable for the 13 mm diameter tubes from the DR 5000 and DR 6000 European version because of the 13 mm shaft. All our Spectroquant® reagent test kits using the 10, 20 and 50 mm cells can be measured with all DR 5000 instruments. The programming data can be found in the following list.
Information about Spectroquant® test kits: In our Internet homepage www.analytical-test-kits.com users can find all information about our test kits. All package inserts as well as the manuals from e.g. our Pharo spectrophotometers can be found there. This information helps to get a quick overview about the easy handling of our test kits. The package inserts include detailed descriptions about all handling steps, colour reaction, sample preparation if required, interference with other substances, analytical quality assurance data and more. Zeroing the instrument: Before measuring the sample users must enter the “Zero” value for each range. In the below enclosed programming tables it is mentioned for each parameter and measuring range, what “Zero” value is necessary. You can choose between the following options:
1) Own blank (= distilled water and reagents) 2) Cell with blank (white screw cap) 3) Distilled water 4) Clear sample without reagents
The advantages of Spectroquant® Tests kits are, that in some cases users can run as “Zero” distilled water or use the cell with blank (white screw cap) an therefore need not to waste always on determination. This makes live easier and reduce costs. The zero should be entered when the instruments request “Zero”. Message “Under range!” after zeroing the instrument: The instrument must be zeroed for each test or series of tests to establish a zero reference for the measurement. This is done by placing a zero solution (distilled water and reagents) in the cell holder and pressing ZERO. While entering a value of e.g. “25” for the low limit - and not “0” (like Hach) - the message “Under range!” appears in the display. This message doesn’t influence the results. Just go forward with the measurement. If you implement “0” for the low limit this message will not appear (identical as if you use an “original” Hach program). Please contact our local Merck companies or Merck agents for any questions or use the contact form in our internet homepage. No endorsement by Hach Company is implied or intended.
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 2 of 49
Merck order no. 1.01758* 1.01758* 1.01762 1.00594.0001* 1.14825.0001 1.14825.0001 1.14825.0001
replaced by 1.01758
Program name Acid capacity01758
Acid capacity01758
Aluminium00594 Aluminium14825 Aluminium14825 Aluminium14825
Unit mmol/l mg/l mg/l mg/l mg/l mg/l
Wavelength 605 nm 605 nm 550 nm 550 nm 550 nm 550 nm
Resolution 0.01 1 0.01 0.01 0.001 0.001
Chemical Form 1 OH CaCO3 Al Al Al Al
Calibration “a” -10,074 -503,7 -0.253 -0.165 -0.0649 -0.033
Calibration “b” 30,24 1512 1.023 0.9585 0.37735 0.1917
Calibration “c” -23,081 -1154,1 -0.599 -0.3233 -0.12728 -0.06466
Calibration “d” 10,358 517,9 0.187 0.0719 0.02829 0.01437
Upper limit 8.00 400 0.50 1.20 0.472 0.240
Lower limit 0.40 20 0.02 0.10 0.039 0.020
Timer 1 (mm:ss) ----- ----- 05:00 02:00 02:00 02:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- -----
Additional Merck information
3) 3) 3) 3) 3) 3)
Cell size used 16 mm round 16 mm round 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 3 of 49
Merck order no. 1.14739.0001* 1.14558.0001* 1.14544.0001* 1.14559.0001* 1.14739.0001* 1.14558.0001* 1.14544.0001* 1.14559.0001*
Program name Ammonium14739 Ammonium14558 Ammonium14544 Ammonium14559 Ammonium14739 Ammonium14558 Ammonium14544 Ammonium14559
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 0.001 0.01 0.1 0.1 0.001 0.01 0.1 0.1
Chemical Form 1 NH4-N NH4-N NH4-N NH4-N NH4+ NH4
+ NH4+ NH4
+
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 0.877 4.26 7.84 36.4 1.13 5.49 10.1 46.8
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 2.000 8.00 16.0 80.0 2.576 10.30 20.6 103.0
Lower limit 0.010 0.20 0.5 4.0 0.013 0.26 0.6 5.2
Timer 1 (mm:ss) 15:00 15:00 15:00 15:00 15:00 15:00 15:00 15:00
Timer 2 (mm:ss) ----- ----- ----- ----- ------ ----- ----- -----
Additional Merck information
2) 2) 2) 2) 2) 2) 2) 2)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 4 of 49
Merck order no. 1.14752 1.14752 1.14752 1.00683.0001 1.00683.0001
Program name Ammonium14752 Ammonium14752 Ammonium14752 Ammonium00683 Ammonium00683
Unit mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 0.01 0.001 0.001 0.1 0.1
Chemical Form 1 NH4-N NH4-N NH4-N NH4-N NH4-N
Calibration “a” 0 0 0 0 0
Calibration “b” 1.23 0.483 0.245 27.8 55.6
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 3.00 1.18 0.500 75.0 150.0
Lower limit 0.05 0.02 0.010 2.0 5.0
Timer 1 (mm:ss) 05:00 05:00 05:00 15:00 15:00
Timer 2 (mm:ss) 05:00 05:00 05:00 ----- -----
Additional Merck information
1) 1) 1) 1) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 5 of 49
Merck order no. 1.14752 1.14752 1.14752 1.00683.0001 1.00683.0001 1.00675.0001*
Program name Ammonium14752 Ammonium14752 Ammonium14752 Ammonium00683 Ammonium00683 AOX00675
Unit mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm 445 nm
Resolution 0.01 0.001 0.001 0.1 0.1 0.01
Chemical Form 1 NH4+ NH4
+ NH4+ NH4
+ NH4+ AOX
Calibration “a” 0 0 0 0 0 0
Calibration “b” 1.58 0.622 0.316 35.8 71.5 4.00
Calibration “c” 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0
Upper limit 3.86 1.520 0.644 96.6 193.2 2.50
Lower limit 0.06 0.024 0.013 2.6 6.4 0.05
Timer 1 (mm:ss) 05:00 05:00 05:00 15:00 15:00 15:00
Timer 2 (mm:ss) 05:00 05:00 05:00 ----- ----- -----
Additional Merck information
1) 1) 1) 1) 1) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 10 mm 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 6 of 49
Merck order no. 1.01747.0001 1.01747.0001 1.00687.0001* 1.00826.0001* 1.14839.0001 1.00605.0001 1.00605.0001 1.00605.0001
Program name Arsenic01747 Arsenic01747 BOD00687 Boron00826 Boron14839 Bromine00605 Bromine00605 Bromine00605
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 525 nm 525 nm 500 nm 410 nm 565 nm 550 nm 550 nm 550 nm
Resolution 0.001 0.001 0.1 0.01 0.001 0.01 0.001 0.001
Chemical Form 1 As As O2 B B Br2 Br2 Br2
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 0.114 0.0447 13.5 1.38 0.328 7.519 2.960 1.504
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 1.000 0.039 12.0 2.00 0.800 10.00 3.937 2.000
Lower limit 0.005 0.002 0.5 0.05 0.050 0.10 0.039 0.020
Timer 1 (mm:ss) 120:00 120:00 ----- 60:00 12:00 01:00 01:00 01:00
Timer 2 (mm:ss) ----- ----- ----- ----- 02:00 ----- ----- -----
Additional Merck information
3) 3) 3) 1) 1) 3) 3) 3)
Cell size used 10 mm 1 inch 16 mm round 16 mm round 10 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 7 of 49
Merck order no. 1.14834.0001* 1.01745.0001 1.01745.0001 1.01745.0001 1.00858.0001* 1.14815.0001 1.14815.0001
see appendix page 47
Program name Cadmium 14834
Cadmium01745 Cadmium01745 Cadmium01745 Calcium00858 Calcium14815 Calcium14815s
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 550 nm 525 nm 525 nm 525 nm 565 nm 550 nm 550 nm
Resolution 0.001 0.001 0.001 0.001 1 1 0.1
Chemical Form 1 Cd Cd Cd Cd Ca Ca Ca
Calibration “a” 0 0 0 0 -29.293 0 0
Calibration “b” 0.595 0.680 0.268 0.136 416.85 137 29.4
Calibration “c” 0 0 0 0 -252.27 0 0
Calibration “d” 0 0 0 0 99.158 0 0
Upper limit 1.000 0.010 0.005 0.002 250 160 15.0
Lower limit 0.025 0.500 0.197 0.100 10 10 1.0
Timer 1 (mm:ss) 02:00 02:00 02:00 02:00 03:00 08:00 08:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 3) 1) 1)
Cell size used 16 mm round 10 mm 1 inch 50 mm 16 mm round 10 mm 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 8 of 49
Merck order no. 1.00858.0001* 1.14815.0001 1.14815.0001 1.00858.0001* 1.14815.0001 1.14815.0001 1.19251.0001
see appendix page 47 see appendix page 47 180 *)
Program name Calcium00858 Calcium14815 Calcium14815s Calcium00858 Calcium14815 Calcium14815s Carbohyrazide see Oxygen Scavengers
Unit mg/l mg/l mg/l mg/l mg/l mg/l µg/l
Wavelength 565 nm 550 nm 550 nm 565 nm 550 nm 550 nm 562 nm
Resolution 1 1 0.1 1 1 0.1 1
Chemical Form 1 CaCO3 CaCO3 CaCO3 CaO CaO CaO Carbohy
Calibration “a” -73.418 0 0 -41.011 0 0
Calibration “b” 1045 342 73.5 583.59 192 41.2
Calibration “c” -634.44 0 0 -353.18 0 0
Calibration “d” 249.09 0 0 138.82 0 0
Upper limit 625 400 37.5 350 224 21.0 600
Lower limit 25 25 2.5 14 14 1.4 0
Timer 1 (mm:ss) 03:00 08:00 08:00 03:00 08:00 08:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----
Additional Merck information
3) 1) 1) 3) 1) 1) 1)
Cell size used 16 mm round 10 mm 10 mm 16 mm round 10 mm 10 mm 1 inch
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 9 of 49
Merck order no. 1.14730.0001* 1.14897 1.14897 1.01807.0001 1.01804.0001* 1.00595.0001* 1.00597.0001* 1.00598
Program name Chloride14730 Chloride14897 Chloride14897 Chloride01807 Chloride01804 Chlorine00595 Chlorine00597 Chlorine00598
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 525 nm 500 nm 445 nm 500 nm 445 nm 550 nm 550 nm 550 nm
Resolution 1 1 0.1 0.01 0.1 0.01 0.01 0.01
Chemical Form 1 Cl- Cl- Cl- Cl- Cl- Cl2 Cl2 Cl2
Calibration “a” 0 0 0 0,01337 -1,537 0 0 0
Calibration “b” 114 107 28.2 4,873 13,35 2.469 2.469 3.333
Calibration “c” 0 0 0 -1,329 -2,082 0 0 0
Calibration “d” 0 0 0 1,018 3,146 0 0 0
Upper limit 125 250 25.0 5.00 15.0 6.00 6.00 6.00
Lower limit 5 10 2.5 0.10 0.5 0.03 0.03 0.05
Timer 1 (mm:ss) ----- 01:00 01:00 10:00 10:00 01:00 01:00 01:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 3) 3) 3) 3)
Cell size used 16 mm round 10 mm 10 mm 50 mm 16 mm round 16 mm round 16 mm round 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 10 of 49
Merck order no. 1.00598 1.00598 1.00602 1.00602 1.00602 1.00599.0001 1.00599.0001 1.00599.0001
Program name Chlorine00598 Chlorine00598 Chlorine00602 Chlorine00602 Chlorine00602 Chlorine00599 Chlorine00599 Chlorine00599
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 550 nm 550 nm 550 nm 550 nm 550 nm 550 nm 550 nm 550 nm
Resolution 0.01 0.001 0.01 0.01 0.001 0.01 0.01 0.001
Chemical Form 1 Cl2 Cl2 Cl2 Cl2 Cl2 Cl2 Cl2 Cl2
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 1.312 0.667 3.333 1.312 0.667 3.333 1.312 0.667
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 3.00 1.000 6.00 2.36 1.000 6.00 2.36 1.000
Lower limit 0.02 0.010 0.05 0.02 0.010 0.05 0.02 0.010
Timer 1 (mm:ss) 01:00 01:00 01:00 01:00 01:00 01:00 01:00 01:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
3) 3) 3) 3) 3) 3) 3) 3)
Cell size used 1 inch 50 mm 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 11 of 49
Merck order no. 1.00086/1.00087
1.00088* 1.00086/1.00087
1.00088 1.00608.0001 1.00608.0001 1.00608.0001
Program name Chlorine00086/87/88
Chlorine00086/87/88
Chlorine dioxide00608
Chlorine dioxide00608
Chlorine dioxide00608
Unit mg/L mg/L mg/l mg/l mg/l
Wavelength 550 nm 550 nm 550 nm 550 nm 550 nm
Resolution 0.01 0.001 0.01 0.01 0.001
Chemical Form 1 Cl2 Cl2 ClO2 ClO2 ClO2
Calibration “a” 0 0 0 0 0
Calibration “b” 2.47 0.67 6.33 2.49 1.27
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 6.00 1.000 10.00 3.94 2.000
Lower limit 0.03 0.010 0.10 0.04 0.020
Timer 1 (mm:ss) 01:00 01:00 02:00 02:00 02:00
Timer 2 (mm:ss) ----- ----- 01:00 01:00 01:00
Additional Merck information
3) 3) 3) 3) 3)
Cell size used 16 mm round 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 12 of 49
Merck order no. 1.14552.0001* 1.14758.0001 1.14758.0001 1.14758.0001 1.14552.0001* 1.14758.0001 1.14758.0001 1.14758.0001
Program name Chromate14552 Chromate14758 Chromate14758 Chromate14758 Chromate14552 Chromate14758 Chromate14758 Chromate14758
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 550 nm 550 nm 550 nm 550 nm 550 nm 550 nm 550 nm 550 nm
Resolution 0.01 0.01 0.001 0.001 0.01 0.01 0.001 0.001
Chemical Form 1 Cr Cr Cr Cr CrO4 CrO4 CrO4 CrO4
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 0.971 1.30 0.511 0.260 2.17 2.90 1.14 0.579
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 2.00 3.00 1.181 0.600 4.46 6.69 2.635 1.338
Lower limit 0.05 0.05 0.020 0.010 0.11 0.11 0.044 0.022
Timer 1 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
2) 3) 3) 3) 2) 3) 3) 3)
Cell size used 16 mm round 10 mm 1 inch 50 mm 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 13 of 49
Merck order no. 1.18750.0001* 1.18751.0001* 1.18752.0001* 1.18753.0001* 1.14560.0001* 1.01796.0001* 1.14540.0001*
Hach Program 431 *) 430 *) 435 *) 435 *)
Program name COD COD COD COD COD14560 COD01796 COD14540
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 350 nm 420 nm 620 nm 620 nm 340 nm 340 nm 445 nm
Resolution 0.1 0.1 1
Chemical Form 1 COD COD COD COD COD COD COD
Calibration “a” 0 0 0
Calibration “b” -41.7 -40.7 -205,3
Calibration “c” 0 0 0
Calibration “d” 0 0 0
Upper limit 40.0 150 1500 15000 40.0 80.0 150
Lower limit 0.0 0 0 0 4.0 5.0 10
Timer 1 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 1) 1) 1)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round
*) The above-mentioned Merck COD cell tests can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach COD tubes. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents). The blank for COD is stabile for e.g. 3 month when storing dark. * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 14 of 49
Merck order no. 1.14895.0001* 1.14690.0001* 1.14541.0001* 1.14691.0001* 1.14555.0001* 1.01797.0001* 1.09772.0001*
Program name COD14895 COD14690 COD14541 COD14691 COD14555 COD01797 COD (Hg free)09772
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 445 nm 445 nm 605 nm 605 nm 605 nm 605 nm 445 nm
Resolution 1 1 1 1 10 10 1
Chemical Form 1 COD COD COD COD COD COD COD
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” -222 -397 1667 3226 4608 42553 -238
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 300 500 1500 3500 10000 90000 150
Lower limit 15 50 25 300 500 5000 10
Timer 1 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 1) 1) 1)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents). The blank for COD is stabile for e.g. 3 month when storing dark. * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 15 of 49
Merck order no. 1.09773.0001* 1.17058.0001* 1.17059.0001* 1.14553.0001* 1.14767.0001 1.14767.0001 1.14767.0001
Program name COD (Hg free)09773
COD seawater 17058
COD seawater 17059
Copper14553 Copper14767 Copper14767 Copper14767
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 605 nm 340 nm 605 nm 605 nm 605 nm 605 nm 605 nm
Resolution 1 0.1 1 0.01 0.01 0.01 0.001
Chemical Form 1 COD COD COD Cu Cu Cu Cu
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 1923 -57.5 3650 3.60 4.44 1.75 0.889
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 1500 60.0 3000 8.00 6.00 2.36 1.200
Lower limit 100 5.0 50 0.05 0.10 0.04 0.020
Timer 1 (mm:ss) ----- ----- ----- 05:00 05:00 05:00 05:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ------ ------
Additional Merck information
1) 1) 1) 2) 3) 3) 3)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 16 of 49
Merck order no. 1.14561.0001* 1.09701.0001 1.09701.0001 1.09701.0001 1.14800.0001 1.14800.0001 1.14800.0001
Program name Cyanide14561 Cyanide09701 Cyanide09701 Cyanide09701 Cyanide14800 Cyanide14800 Cyanide14800
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 605 nm 605 nm 605 nm 605 nm 585 nm 585 nm 585 nm
Resolution 0.001 0.001 0.001 0.001 0.001 0.001 0.001
Chemical Form 1 CN- CN- CN- CN- CN- CN- CN-
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 0.161 0.221 0.0871 0.0442 0.208 0.0819 0.0416
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 0.500 0.500 0.197 0.100 0.500 0.197 0.100
Lower limit 0.010 0.010 0.004 0.002 0.010 0.004 0.002
Timer 1 (mm:ss) 10:00 10:00 10:00 10:00 05:00 05:00 05:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
3) 3) 3) 3) 3) 3) 3)
Cell size used 16 mm round 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 17 of 49
Merck order no. 1.19250.0001 1.19253.0001 1.19251.0001 1.14557.0001* 1.14557.0001 1.00809.0001* 1.00809.0001
*) *) 181 *) see appendix page 47
see appendix page 47
Program name Cyanuric Acid 19250
Cyanuric Acid 19253
DEHA see Oxygen Scavengers
Fluoride14557 Fluoride14557s Fluoride00809 Fluoride00809s
Unit mg/l mg/l µg/l mg/l mg/l mg/l mg/l
Wavelength 525 nm 525 nm 562 nm 620 nm 620 nm 620 nm 620 nm
Resolution 1 1 1 0.01 0.001 0.01 0.001
Chemical Form 1 Cyanuric Acid Cyanuric Acid DEHA F- F- F- F-
Calibration “a” 0 0 0 0
Calibration “b” 1.33 0.323 1.43 0.312
Calibration “c” 0 0 0 0
Calibration “d” 0 0 0 0
Upper limit 160 160 450 1.50 0.500 1.80 0.500
Lower limit 2 2 0 0.10 0.025 0.10 0.025
Timer 1 (mm:ss) 03:00 03:00 10:00 05:00 05:00 15:00 15:00
Timer 2 (mm:ss) ---- ---- ---- ----- ----- ----- -----
Additional Merck information
4) 4) 1) 1) 1) 1) 1)
Cell size used 1 inch 1 inch 1 inch 16 mm round 50 mm 16 mm round 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box.As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 18 of 49
Merck order no. 1.14598 1.14598 1.00822.0250 1.14500.0001* 1.14678.0001 1.14678.0001 1.14678.0001
190 *)
Program name Fluoride14598 Fluoride14598 Fluoride Formaldehyde 14500
Formaldehyde 14678
Formaldehyde 14678
Formaldehyde 14678
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 620 nm 620 nm 580 nm 565 nm 565 nm 565 nm 565 nm
Resolution 0.01 0.1 0.1 0.01 0.01 0.01 0.001
Chemical Form 1 F- F- F- HCHO HCHO HCHO HCHO
Calibration “a” 0 0 0 0 0 0
Calibration “b” 2.02 21.5 3.23 4.22 1.66 0.844
Calibration “c” 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0
Upper limit 2.00 20.0 2.00 8.00 8.00 3.15 1.500
Lower limit 0.10 1.0 0.02 0.10 0.10 0.04 0.020
Timer 1 (mm:ss) 05:00 05:00 01:00 05:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 3) 3) 3) 3)
Cell size used 10 mm 10 mm 1 inch 16 mm round 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 19 of 49
Merck order no. 1.14821.0001 1.09711.0001 1.09711.0001 1.09711.0001 1.14731.0001* 1.14731.0001
see appendix page 47
Program name Gold14821 Hydrazine09711 Hydrazine09711 Hydrazine09711 Hydrogen peroxide14731
Hydrogen peroxide14731s
Unit mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 550 nm 445 nm 445 nm 445 nm 410 nm 410 nm
Resolution 0.1 0.01 0.001 0.001 0.1 0.01
Chemical Form 1 Au N2H4 N2H4 N2H4 H2O2 H2O2
Calibration “a” 0 0 0 0 0 0
Calibration “b” 6.25 0.870 0.342 0.174 38.5 10.5
Calibration “c” 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0
Upper limit 12.0 2.00 0.787 0.400 20.0 5.00
Lower limit 0.5 0.02 0.008 0.005 2.0 0.25
Timer 1 (mm:ss) ----- 05:00 05:00 05:00 02: 00 02: 00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- -----
Additional Merck information
1) 3) 3) 3) 2) 3)
Cell size used 10 mm 10 mm 1 inch 50 mm 16 mm round 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 20 of 49
Merck order no. 1.18789.0001 1.18789.0001 1.19251.0001 1.00606.0001 1.00606.0001 1.00606.0001
182 *)
Program name Hydrogen peroxide18789
Hydrogen peroxide18789
Hydroquinone see Oxygen Scavengers
Iodine00606 Iodine00606 Iodine00606
Unit mg/l mg/l µg/l mg/l mg/l mg/l
Wavelength 445 nm 445 nm 562 nm 550 nm 550 nm 550 nm
Resolution 0.01 0.01 1 0.01 0.01 0.001
Chemical Form 1 H2O2 H2O2 Hydro I2 I2 I2
Calibration “a” 0 0 0 0 0
Calibration “b” 2.65 1.04 11.9 4.69 2.38
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 6.00 2.36 1000 10.00 3.94 2.000
Lower limit 0.03 0.01 0 0.20 0.08 0.050
Timer 1 (mm:ss) 10: 00 10: 00 2:00 01:00 01:00 01:00
Timer 2 (mm:ss) ----- ----- ---- ---- ----- -----
Additional Merck information
1) 1) 1) 3) 3) 3)
Cell size used 10 mm 1 inch 1 inch 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 21 of 49
Merck order no. 1.14549.0001* 1.14896.0001* 1.14761 1.14761 1.14761 1.100796.0001 1.100796.0001 1.100796.0001
Program name Iron14549 Iron14896 Iron14761 Iron14761 Iron14761 Iron00796 Iron00796 Iron00796
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 565 nm 525 nm 565 nm 565 nm 565 nm 500 nm 500 nm 500 nm
Resolution 0.01 0.1 0.01 0.001 0.001 0.01 0.01 0.001
Chemical Form 1 Fe Fe Fe Fe Fe Fe Fe Fe
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 1.64 28.3 2.08 0.820 0.417 5.56 2.19 1.11
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 4.00 50.0 5.00 1.97 1.000 5.00 1.97 1.000
Lower limit 0.05 1.0 0.05 0.02 0.005 0.10 0.04 0.010
Timer 1 (mm:ss) 03:00 05:00 03:00 03:00 03:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
2) 2) 3) 3) 3) 3) 3) 3)
Cell size used 16 mm round 16 mm round 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 22 of 49
Merck order no. 1.19251.0001 1.14833.0001* 1.09717.0001 1.09717.0001 1.09717.0001 1.00815.0001*
183 *)
Program name Isoascorbic acid see Oxygen Scavengers
Lead14833 Lead09717 Lead09717 Lead09717 Magnesium 00815
Unit µg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 562 nm 525 nm 525 nm 525 nm 525 nm 565 nm
Resolution 1 0.01 0.01 0.001 0.001 0.1
Chemical Form 1 ISA Pb Pb Pb Pb Mg
Calibration “a” 0 0 0 0 -20.79
Calibration “b” 4.39 6.25 2.46 1.25 337.41
Calibration “c” 0 0 0 0 -513.33
Calibration “d” 0 0 0 0 458.21
Upper limit 1500 5.00 5.00 1.970 1.000 75.0
Lower limit 0 0.10 0.10 0.039 0.010 5.0
Timer 1 (mm:ss) 10:00 ----- ----- ----- ----- 03:00
Timer 2 (mm:ss) ---- ----- ----- ----- ----- -----
Additional Merck information
1) 2) 1) 1) 1) 3)
Cell size used 1 inch 16 mm round 10 mm 1 inch 50 mm 16 mm round
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to DR 5000 US version. DR 5000 European versions (with 13 mm cell holder) do not allow tests with 16 mm round cells!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 23 of 49
Merck order no. 1.00816.0001* 1.01846.0001 1.01846.0001 1.01846.0001
Program name Manganese 00816
Manganese 01846
Manganese 01846
Manganese 01846
Unit mg/l mg/l mg/l mg/l
Wavelength 445 nm 565 nm 565 nm 565 nm
Resolution 0.01 0.01 0.001 0.001
Chemical Form 1 Mn Mn Mn Mn
Calibration “a” 0 0 0 0
Calibration “b” 4.13 1.67 0.657 0.334
Calibration “c” 0 0 0 0
Calibration “d” 0 0 0 0
Upper limit 5.00 2.00 0.790 0.400
Lower limit 0.10 0.05 0.020 0.005
Timer 1 (mm:ss) 02:00 10:00 10:00 10:00
Timer 2 (mm:ss) 05:00 ----- ----- -----
Additional Merck information
2) 1) 1) 1)
Cell size used 16 mm round 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 24 of 49
Merck order no. 1.01739.0001 1.01739.0001 1.01739.0001 1.14770 1.14770 1.14770 1.19251.0001
184 *)
Program name Manganese 01739
Manganese 01739
Manganese 01739
Manganese 14770
Manganese 14770
Manganese 14770
Methylethylketoxime see Oxygen Scavengers
Unit mg/l mg/l mg/l mg/l mg/l mg/l µg/l
Wavelength 565 nm 565 nm 565 nm 445 nm 445 nm 445 nm 562 nm
Resolution 0.01 0.001 0.001 0.01 0.01 0.001 1
Chemical Form 1 Mn Mn Mn Mn Mn Mn MEKO
Calibration “a” 0 0 0 0 0 0
Calibration “b” 1.72 0.679 0.345 5.62 2.21 1.12
Calibration “c” 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0
Upper limit 2.00 0.790 0.400 10.00 3.94 2.000 1000
Lower limit 0.05 0.020 0.005 0.50 0.20 0.010 0
Timer 1 (mm:ss) 10:00 10:00 10:00 02:00 02:00 02:00 10:00
Timer 2 (mm:ss) ----- ----- ----- 02:00 02:00 02:00 ----
Additional Merck information
1) 1) 1) 3) 3) 3) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm 1 inch
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 25 of 49
Merck order no. 1.00860.0001* 1.19252.0001 1.01632.0001 1.01632.0001 1.01632.0001 1.01632.0001 1.01632.0001 1.01632.0001
Hach Program 320 *)
Program name Molybdenum 00860
Molybdenum Monochloramine01632
Monochloramine01632
Monochloramine01632
Monochloramine 01632
Monochloramine 01632
Monochloramine 01632
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 620 nm 420 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 0.01 0.01 0.01 0.001 0.01 0.01 0.001
Chemical Form 1 Mo Mo Cl2 Cl2 Cl2 NH2Cl NH2Cl NH2Cl
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 1.10 5.00 1.97 1.00 3.67 1.44 0.733
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 1.00 40.0 10.00 3.94 2.000 7.25 2.85 1.450
Lower limit 0.02 0.2 0.25 0.10 0.050 0.18 0.09 0.036
Timer 1 (mm:ss) 02:00 05:00 05:00 05:00 05:00 05:00 05:00 05:00
Timer 2 (mm:ss) ----- ---- 10:00 10:00 10:00 10:00 10:00 10:00
Additional Merck information
1) 4) 1) 1) 1) 1) 1) 1)
Cell size used 16 mm round 1 inch 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 26 of 49
Merck order no. 1.01632.0001 1.01632.0001 1.01632.0001 1.14554.0001* 1.14785.0001 1.14785.0001 1.14785.0001
Program name Monochloramine 01632
Monochloramine 01632
Monochloramine 01632
Nickel14554 Nickel14785 Nickel14785 Nickel14785
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 445 nm 445 nm 445 nm 445 nm
Resolution 0.01 0.01 0.001 0.01 0.01 0.01 0.001
Chemical Form 1 NH2Cl-N NH2Cl-N NH2Cl-N Ni Ni Ni Ni
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 0.988 0.389 0.198 3.82 4.76 1.87 0.952
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 1.96 0.77 0.392 6.00 5.00 1.97 1.000
Lower limit 0.05 0.02 0.010 0.10 0.10 0.04 0.020
Timer 1 (mm:ss) 05:00 05:00 05:00 01:00 01:00 01:00 01:00
Timer 2 (mm:ss) 10:00 10:00 10:00 02:00 02:00 02:00 02:00
Additional Merck information
1) 1) 1) 2) 3) 3) 3)
Cell size used 10 mm 1 inch 50 mm 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 27 of 49
Merck order no. 1.14542.0001* 1.14563.0001* 1.14764.0001* 1.00614.0001* 1.14542.0001* 1.14563.0001* 1.14764.0001*
Program name Nitrate14542 Nitrate14563 Nitrate14764 Nitrate00614 Nitrate14542 Nitrate14563 Nitrate14764
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 525 nm 340 nm 340 nm 340 nm 525 nm 340 nm 340 nm
Resolution 0.1 0.1 0.1 1 0.1 0.1 0.1
Chemical Form 1 NO3-N NO3-N NO3-N NO3-N NO3- NO3
- NO3-
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 7.14 14.7 27.9 128 31.6 65.3 124
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 18.0 25.0 50.0 225 79.7 110.7 221.3
Lower limit 0.5 0.5 1.0 23 2.2 2.2 4.4
Timer 1 (mm:ss) 10:00 10:00 10:00 05:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
2) 2) 2) 2) 2) 2) 2)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 28 of 49
Merck order no. 1.00614.0001* 1.14773.0001 1.14773.0001 1.14773.0001 1.14773.0001
Program name Nitrate00614 Nitrate14773 Nitrate14773 Nitrate14773 Nitrate14773
Unit mg/l mg/l mg/l mg/l mg/l
Wavelength 340 nm 525 nm 525 nm 525 nm 525 nm
Resolution 1 0.1 0.1 0.1 0.1
Chemical Form 1 NO3- NO3-N NO3-N NO3
- NO3-
Calibration “a” 0 0 0 0 0
Calibration “b” 568 9.62 3.79 42.6 16.8
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 996 20.0 7.9 88.5 34.8
Lower limit 102 0.5 0.2 2.2 0.9
Timer 1 (mm:ss) 05:00 10:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- -----
Additional Merck information
2) 1) 1) 1) 1)
Cell size used 16 mm round 10 mm 1 inch 10 mm 1 inch As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 29 of 49
Merck order no. 1.09713 1.09713 1.09713 1.09713 1.09713 1.09713
Program name Nitrate09713 Nitrate09713 Nitrate09713 Nitrate09713 Nitrate09713 Nitrate09713
Unit mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 340 nm 340 nm 340 nm 340 nm 340 nm 340 nm
Resolution 0.1 0.1 0.01 0.1 0.1 0.1
Chemical Form 1 NO3-N NO3-N NO3-N NO3- NO3
- NO3-
Calibration “a” 0 0 0 0 0 0
Calibration “b” 19.6 7.72 3.92 86.8 34.2 17.4
Calibration “c” 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0
Upper limit 25.0 9.8 5.00 110.7 43.6 22.1
Lower limit 1.0 0.4 0.10 4.4 1.7 0.4
Timer 1 (mm:ss) 10:00 10:00 10:00 10:00 10:00 10:00
Timer 2 (mm:ss)
Additional Merck information
1) 1) 1) 1) 1) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 30 of 49
Merck order no. 1.14556.0001* 1.14942.0001 1.01842.0001 1.01842.0001 1.14556.0001* 1.14942.0001 1.00609.0001*
Program name Nitrate seawater14556
Nitrate seawater14942 Nitrate01842 Nitrate01842 Nitrate
seawater14556 Nitrate
seawater14942 Nitrite00609
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 500 nm 500 nm 500 nm 500 nm 500 nm 500 nm 605 nm
Resolution 0.01 0.01 0.1 0.1 0.01 0.1 0.1
Chemical Form 1 NO3-N NO3-N NO3-N NO3- NO3
- NO3- NO2
-
Calibration “a” -0.047 0.421 -0,5041 -2.2315 -0.0442 -1.865 0
Calibration “b” 1.718 8.251 32,2843 142.913 7.182 36.524 82.64
Calibration “c” -0.438 -0.5679 -21,08326 -93.329 -1.6478 -2.515 0
Calibration “d” 0.0821 0.335 7,1650 31.717 0.3041 149 0
Upper limit 3.00 17.00 30.0 132.9 13.28 75.3 90.0
Lower limit 0.10 0.20 0.3 1.3 0.44 0.9 1.0
Timer 1 (mm:ss) 30:00 15:00 10:00 10:00 30:00 15:00 20:00
Timer 2 (mm:ss) ----- 60:00 ----- ----- ----- 60:00 -----
Additional Merck information
3) 3) 1) 1) 3) 3) 1)
Cell size used 16 mm round 10 mm 50 mm 50 mm 16 mm round 10 mm 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 31 of 49
Merck order no. 1.14547.0001* 1.14776 1.14776 1.14776 1.14547.0001* 1.14776 1.14776
Program name Nitrite14547 Nitrite14776 Nitrite14776 Nitrite14776 Nitrite14547 Nitrite14776 Nitrite14776
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 525 nm 525 nm 525 nm 525 nm 525 nm 525 nm 525 nm
Resolution 0.001 0.001 0.001 0.001 0.001 0.001 0.001
Chemical Form 1 NO2-N NO2-N NO2-N NO2-N NO2- NO2
- NO2-
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 0.274 0.376 0.148 0.0752 0.900 1.22 0.479
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 0.700 1.000 0.394 0.200 2.300 3.285 1.293
Lower limit 0.010 0.020 0.008 0.002 0.033 0.066 0.026
Timer 1 (mm:ss) 10:00 10:00 10:00 10:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
2) 3) 3) 3) 2) 3) 3)
Cell size used 16 mm round 10 mm 1 inch 50 mm 16 mm round 10 mm 1 inch As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 32 of 49
Merck order no. 1.14776 1.00613.0001* 1.14537.0001* 1.14763.0001* 1.19251.0001 1.19251.0001 1.19251.0001
Hach Program 180 *) 181 *) 182 *)
Program name Nitrite14776 Nitrogen00613 Nitrogen14537 Nitrogen14763 Oxygen Scavengers Carbohyrazide
Oxygen Scavengers DEHA
Oxygen Scavengers Hydroquinone
Unit mg/l mg/l mg/l mg/l µg/l µg/l µg/l
Wavelength 525 nm 365 nm 525 nm 365 nm 562 nm 562 nm 562 nm
Resolution 0.001 0.1 0.1 1 1 1 1
Chemical Form 1 NO2- N N N Carbohy DEHA Hydro
Calibration “a” 0 0 0 0
Calibration “b” 0.243 18.4 7.81 185
Calibration “c” 0 0 0 0
Calibration “d” 0 0 0 0
Upper limit 0.657 15.0 15.0 150 600 450 1000
Lower limit 0.007 0.5 0.5 10 0 0 0
Timer 1 (mm:ss) 10:00 10:00 10:00 10:00 10:00 10:00 2:00
Timer 2 (mm:ss) ----- ----- ----- ----- ---- ---- ----
Additional Merck information
3) 2) 1) 2) 1) 1) 1)
Cell size used 50 mm 16 mm round 16 mm round 16 mm round 1 inch 1 inch 1 inch
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 33 of 49
Merck order no. 1.19251.0001 1.19251.0001 1.14694.0001* 1.00607 1.00607 1.00607 1.01744.0001*
Hach Program 183 *) 184 *)
Program name Oxygen Scavengers Isoascorbic acid
Oxygen Scavengers Methylethylketoxime
Oxygen14694 Ozone00607 Ozone00607 Ozone00607 pH01744
Unit µg/l µg/l mg/l mg/l mg/l mg/l pH
Wavelength 562 nm 562 nm 500 nm 550 nm 550 nm 550 nm 550 nm
Resolution 1 1 0.1 0.01 0.001 0.001 0.1
Chemical Form 1 ISA MEKO O2 O3 O3 O3
Calibration “a” 0 0 0 0 6.144
Calibration “b” 13.5 2.25 0.887 0.450 3.218
Calibration “c” 0 0 0 0 -2.318
Calibration “d” 0 0 0 0 0.714
Upper limit 1500 1000 12.0 4.00 1.575 0.800 6.4
Lower limit 0 0 0.5 0.05 0.020 0.010 8.8
Timer 1 (mm:ss) 10:00 10:00 ----- 01:00 01:00 01:00 -----
Timer 2 (mm:ss) ---- ---- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 3) 3) 3) 3) 3)
Cell size used 1 inch 1 inch 16 mm round 10 mm 1 inch 50 mm 16 mm round
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 34 of 49
Merck order no. 1.14551.0001* 1.00856.0001 1.00856.0001 1.00856.0001 1.00856.0001 1.00474.0001* 1.14543.0001*
Program name Phenol14551 Phenol00856 Phenol00856 Phenol00856 Phenol00856 Phosphate00474 Phosphate14543
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 500 nm 445 nm 495 nm 495 nm 495 nm 690 nm 690 nm
Resolution 0.01 0.001 0.01 0.01 0.001 0.01 0.01
Chemical Form 1 Phenol Phenol Phenol Phenol Phenol PO4-P PO4-P
Calibration “a” 0 0 0 0 0 0 0
Calibration “b” 3.39 0.0895 8.33 3.28 1.67 1.61 1.61
Calibration “c” 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0
Upper limit 2.50 0.100 5.00 1.97 1.000 5.00 5.00
Lower limit 0.10 0.002 0.10 0.07 0.025 0.05 0.05
Timer 1 (mm:ss) 01:00 30:00 10:00 10:00 10:00 05:00 05:00
Timer 2 (mm:ss) ----- 10:00 ----- ----- ----- ----- -----
Additional Merck information
2) 1) 1) 1) 1) 2) 2)
Cell size used 16 mm round 1 inch 10 mm 1 inch 50 mm 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 35 of 49
Merck order no. 1.00475.0001* 1.14729.0001* 1.00616.0001* 1.00673.0001* 1.00474.0001* 1.14543.0001* 1.00475.0001* 1.14729.0001*
Program name Phosphate00475 Phosphate14729 Phosphate00616 Phosphate00673 Phosphate00474 Phosphate14543 Phosphate00475 Phosphate14729
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 0.1 0.1 0.1 0.1 0.01 0.01 0.1 0.1
Chemical Form 1 PO4-P PO4-P PO4-P PO4-P PO43-
PO43-
PO43-
PO43-
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 8.06 8.06 39.2 39.2 4.95 4.95 24.7 24.7
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 25.0 25.0 100.0 100.0 15.33 15.33 76.7 76.7
Lower limit 0.5 0.5 3.0 3.0 0.15 0.15 1.5 1.5
Timer 1 (mm:ss) 05:00 05:00 05:00 05:00 05:00 05:00 05:00 05:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
2) 2) 2) 2) 2) 2) 2) 2)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to DR 5000 US version. DR 5000 European versions (with 13 mm cell holder) do not allow tests with 16 mm round cells!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 36 of 49
Merck order no. 1.00616.0001* 1.00673.0001* 1.00474.0001* 1.14543.0001* 1.00475.0001* 1.14729.0001* 1.00616.0001* 1.00673.0001*
Program name Phosphate00616 Phosphate00673 Phosphate00474 Phosphate14543 Phosphate00475 Phosphate14729 Phosphate00616 Phosphate00673
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 1 1 0.01 0.01 0.1 0.1 1 1
Chemical Form 1 PO43-
PO43-
P2O5 P2O5 P2O5 P2O5 P2O5 P2O5
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 120 120 3.70 3.70 18.5 18.5 89.9 89.9
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 307 307 11.46 11.46 57.3 57.3 229 229
Lower limit 9 9 0.11 0.11 1.1 1.1 7 7
Timer 1 (mm:ss) 05:00 05:00 05:00 05:00 05:00 05:00 05:00 05:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
2) 2) 2) 2) 2) 2) 2) 2)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 37 of 49
Merck order no. 1.14848 1.14848 1.14848 1.00798.0001 1.00798.0001 1.14848
Program name Phosphate14848 Phosphate14848 Phosphate14848 Phosphate00798 Phosphate00798 Phosphate14848
Unit mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 0.01 0.01 0.001 0.1 0.1 0.1
Chemical Form 1 PO4-P PO4-P PO4-P PO4-P PO4-P PO43-
Calibration “a” 0 0 0 0 0 0
Calibration “b” 2.08 0.820 0.417 35.1 13.8 6.38
Calibration “c” 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0
Upper limit 5.00 1.97 1.000 100.0 39.4 15.3
Lower limit 0.05 0.02 0.010 1.0 0.4 0.2
Timer 1 (mm:ss) 05:00 05:00 05:00 05:00 05:00 05:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 1) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 38 of 49
Merck order no. 1.14848 1.14848 1.00798.0001 1.00798.0001 1.14848 1.14848 1.14848 1.14546.0001*
Program name Phosphate14848 Phosphate14848 Phosphate00798 Phosphate00798 Phosphate14848 Phosphate14848 Phosphate14848 Phosphate14546
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 690 nm 410 nm
Resolution 0.01 0.01 1 0.1 0.01 0.01 0.001 0.1
Chemical Form 1 PO43- PO4
3- PO43- PO4
3- P2O5 P2O5 P2O5 PO4-P
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 2.51 1.28 108 42.4 4.76 1.88 0.954 13.4
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 6.04 3.0 307 120.9 11.46 4.51 2.291 25.0
Lower limit 0.06 0.03 3 1.2 0.11 0.05 0.023 0.5
Timer 1 (mm:ss) 05:00 05:00 05:00 05:00 05:00 05:00 05:00 -----
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 1) 1) 1) 2)
Cell size used 1 inch 50 mm 10 mm 1 inch 10 mm 1 inch 50 mm 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 39 of 49
Merck order no. 1.14842.0001 1.14842.0001 1.14546.0001* 1.14842.0001 1.14842.0001 1.14546.0001* 1.14842.0001 1.14842.0001
Program name Phosphate14842 Phosphate14842 Phosphate14546 Phosphate14842 Phosphate14842 Phosphate14546 Phosphate14842 Phosphate14842
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 410 nm 410 nm 410 nm 410 nm 410 nm 410 nm 410 nm 410 nm
Resolution 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1
Chemical Form 1 PO4-P PO4-P PO43-
PO43- PO4
3- P2O5 P2O5 P2O5
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 18.0 7.09 41.2 55.2 21.7 30.8 41.3 16.2
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 30.0 11.8 76.7 92.0 36.2 57.3 68.7 27.0
Lower limit 1.0 0.4 1.5 3.1 1.2 1.1 2.3 0.9
Timer 1 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 2) 1) 1) 2) 1) 1)
Cell size used 10 mm 1 inch 16 mm round 10 mm 1 inch 16 mm round 10 mm 1 inch As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 40 of 49
Merck order no. 1.14562.0001* 1.00615.0001* 1.14683.0001* 1.14683.0001* 1.14683.0001* 1.14683.0001* 1.14683.0001* 1.14683.0001*
*) *)
Program name Potassium14562 Potassium00615 Residual Hardness14683
Residual Hardness14683
Residual Hardness14683
Residual Hardness14683
Residual Hardness14683
Residual Hardness14683
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 565 nm 565 nm 565 nm 565 nm 565 nm 565 nm
Resolution 0.1 1 0.01 0.01 0.01 0.001 0.001 0.001
Chemical Form 1 K K Ca CaCO3 CaO °d °e °f
Calibration “a” -0.46153 -1.1538 -0.64615 -0.06461 -0.08077 -0.11538
Calibration “b” 4.62293 11.557 6.4721 0.64721 0.80901 1.15573
Calibration “c” -1.14084 -2.8521 -1.59717 -0.15972 -0.19965 -0.28521
Calibration “d” 0.32358 0.8089 0.45301 0.0453 0.05663 0.08089
Upper limit 50.0 300 5.00 12.50 7.00 0.700 0.875 1.250
Lower limit 5.0 30 0.50 1.25 0.70 0.070 0.088 0.125
Timer 1 (mm:ss) 05:00 05:00 10:00 10:00 10:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- -----
Additional Merck information
3) 3) 3) 3) 3) 3) 3) 3)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 41 of 49
Merck order no. 1.14794.0001 1.14794.0001 1.14794.0001 1.00857.0001 1.00857.0001 1.00857.0001 1.00857.0001 1.01813.0001
Program name Silicate14794 Silicate14794 Silicate14794 Silicate00857 Silicate00857 Silicate00857 Silicate00857 Silicate01813
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 665 nm 665 nm 820 nm 410 nm 410 nm 410 nm 410 nm 820 nm
Resolution 0.01 0.01 0.001 0.1 0.1 1 0.1 0.0001
Chemical Form 1 Si Si Si Si Si Si Si Si
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 3.74 1.47 0.286 28.0 11.0 282 111 0.785
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 5.00 1.97 0.750 50.0 19.7 500 196.9 0.2337
Lower limit 0.10 0.04 0.005 0.5 0.2 5 2.0 0.0002
Timer 1 (mm:ss) 03:00 03:00 03:00 02:00 02:00 02:00 02:00 05:00
Timer 2 (mm:ss) 10:00 10:00 10:00 02:00 02:00 02:00 02:00 05:00
Additional Merck information
3) 3) 3) 3) 3) 3) 3) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 42 of 49
Merck order no. 1.14794.0001 1.14794.0001 1.14794.0001 1.00857.0001 1.00857.0001 1.00857.0001 1.00857.0001 1.01813.0001
Program name Silicate14794 Silicate14794 Silicate14794 Silicate00857 Silicate00857 Silicate00857 Silicate00857 Silicate01813
Unit mg/l mg/l mg/l Mg/l mg/l mg/l mg/l mg/l
Wavelength 665 nm 665 nm 820 nm 410 nm 410 nm 410 nm 410 nm 820 nm
Resolution 0.01 0.01 0.001 0.1 0.1 1 1 0.0001
Chemical Form 1 SiO2 SiO2 SiO2 SiO2 SiO2 SiO2 SiO2 SiO2
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 8.00 3.15 0.611 59.8 23.5 603 237 1.68
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 10.70 4.21 1.600 107.0 42.1 1070 421 0.5000
Lower limit 0.21 0.08 0.010 1.1 0.4 11 4 0.0005
Timer 1 (mm:ss) 03:00 03:00 03:00 02:00 02:00 02:00 02:00 05:00
Timer 2 (mm:ss) 10:00 10:00 10:00 02:00 02:00 02:00 02:00 05:00
Additional Merck information
3) 3) 3) 3) 3) 3) 3) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 43 of 49
Merck order no. 1.14831.0001 1.14831.0001 1.00885.0001* 1.14548.0001* 1.00617.0001* 1.14564.0001* 1.14791.0001
*) *) *)
Program name Silver14831 Silver14831 Sodium00885 Sulfate14548 Sulfate00617 Sulfate14564 Sulfate14791
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 550 nm 550 nm 550 nm 525 nm 525 nm 820 nm 525 nm
Resolution 0.01 0.01 1 1
Chemical Form 1 Ag Ag Na SO42- SO4
2- SO4
2- SO4
2-
Calibration “a” 0 0 0 0
Calibration “b” 2.22 0.874 263 256
Calibration “c” 0 0 0 0
Calibration “d” 0 0 0 0
Upper limit 3.00 1.18 300 250 500 1000 300
Lower limit 0.50 0.20 10 5 50 100 25
Timer 1 (mm:ss) 05:00 05:00 01:00 02:00 02:00 02:00 05:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- 07:00
Additional Merck information
1) 1) 1) 4) 4) 4) 1)
Cell size used 10 mm 1 inch 16 mm round 16 mm round 16 mm round 16 mm round 10 mm
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 44 of 49
Merck order no. 1.01812.0001 1.01812.0001 1.01812.0001 1.14779.0001 1.14779.0001 1.14394.0001* 1.14394.0001*
*) *) *)
Program name Sulfate01812 Sulfate01812 Sulfate01812 Sulfide14779 Sulfide14779 Sulfite14394 Sulfite14394
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 445 nm 445 nm 445 nm 665 nm 665 nm 410 nm 410 nm
Resolution 0.1 0.1 0.01 0.01 0.001 0.1 0.1
Chemical Form 1 SO42- SO4
2- SO4
2- S2- S2- SO3
2- SO2
Calibration “a” 0.0052 0 0 0
Calibration “b” 1.1079 0.244 8.77 7.02
Calibration “c” -0.0055 0 0 0
Calibration “d” 0.1062 0 0 0
Upper limit 50.0 19.7 10.00 1.50 0.500 20.0 16.0
Lower limit 2.5 1.0 0.50 0.10 0.020 1.0 0.8
Timer 1 (mm:ss) 02:00 02:00 02:00 01:00 01:00 02:00 02:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- -----
Additional Merck information
4) 4) 4) 3) 3) 2) 2)
Cell size used 10 mm 1 inch 50 mm 10 mm 50 mm 16 mm round 16 mm round
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 45 of 49
Merck order no. 1.14394.0001 1.14394.0001 1.01746.0001 1.01746.0001 1.14697.0001* 1.01764.0001* 1.01787.0001* 1.14622.0001*
see appendix page 48 see appendix page 48
Program name Sulfite14394s Sulfite14394s Sulfite01746 Sulfite01746 Surfact.anion. 14697
Surfact.cat. 01764
Surfact.nonion. 01787
Tin14622
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 410 nm 410 nm 410 nm 410 nm 650 nm 635 nm 605 nm 665 nm
Resolution 0.01 0.01 0.1 0.1 0.01 0.01 0.01 0.01
Chemical Form 1 SO32-
SO2 SO32-
SO2 Surfact. CTAB Triton X-100 Sn
Calibration “a” 0 0 0 0 0 0 0 0.03
Calibration “b” 1.69 1.35 29.4 23.5 1.80 1.92 6.06 1.5544
Calibration “c” 0 0 0 0 0 0 0 -0.5016
Calibration “d” 0 0 0 0 0 0 0 0.147
Upper limit 3.00 2.40 60.0 48.0 2.00 1.50 7.50 2.50
Lower limit 0.05 0.04 1.0 0.8 0.05 0.05 0.10 0.10
Timer 1 (mm:ss) 02:00 02:00 02:00 02:00 10:00 05:00 02:00 15:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 1) 1) 2) 2) 2) 3)
Cell size used 50 mm 50 mm 10 mm 10 mm 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 46 of 49
Merck order no. 1.14878.0001* 1.14879.0001* 1.00961.0001* 1.00961.0001* 1.00961.0001* 1.00961.0001* 1.00961.0001* 1.00961.0001*
Program name TOC14878 TOC14879 Total hardness 00961
Total hardness 00961
Total hardness 00961
Total hardness 00961
Total hardness 00961
Total hardness 00961
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 605 nm 605 nm 565 nm 565 nm 565 nm 565 nm 565 nm 565 nm
Resolution 0.1 1 1 1 1 0.1 0.1 0.1
Chemical Form 1 TOC TOC Ca CaCO3 CaO °d °e °f
Calibration “a” 0 0 -22.354 -55.594 -31.296 -3.1296 -3.8748 -5.5594
Calibration “b” -64.5 -645 339.98 849.94 475.97 47.597 59.261 84.994
Calibration “c” 0 0 -195.26 -489.08 -273.36 -27.336 -33.641 -48.908
Calibration “d” 0 0 69.671 175.17 97.54 9.754 11.864 17.517
Upper limit 80.0 800 215 538 301 30.1 37.6 53.8
Lower limit 5.0 50 5 13 7 0.7 0.9 1.3
Timer 1 (mm:ss) ----- ----- 03:00 03:00 03:00 03:00 03:00 03:00
Timer 2 (mm:ss) ----- ----- ----- ----- ----- ----- ----- -----
Additional Merck information
1) 1) 3) 3) 3) 3) 3) 3)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 47 of 49
Merck order no. 1.01763.0001* 1.01749.0001* 1.01749.0001* 1.01809.0001* 1.01809.0001* 1.00861.0001* 1.14566.0001* 1.14832.0001
Program name Volatile Org.Acids01763
Volatile Org.Acids01749
Volatile Org.Acids01749
Volatile Org.Acids01809
Volatile Org.Acids01809
Zinc00861 Zinc14566 Zinc14832
Unit mg/l mg/l mg/l mg/l mg/l mg/l mg/l mg/l
Wavelength 495 nm 500 nm 500 nm 500 nm 500 nm 500 nm 500 nm 565 nm
Resolution 1 1 1 1 1 0.001 0.01 0.01
Chemical Form 1 HOAc HOAc Butyric acid HOAc Butyric acid Zn Zn Zn
Calibration “a” 0 0 0 0 0 0 0 0
Calibration “b” 1430 1841 2486 1841 2486 1.36 4.88 1.08
Calibration “c” 0 0 0 0 0 0 0 0
Calibration “d” 0 0 0 0 0 0 0 0
Upper limit 3000 3000 4401 3000 4401 1.000 5.00 2.50
Lower limit 50 50 73 50 73 0.025 0.20 0.05
Timer 1 (mm:ss) 03:00 01:00 01:00 01:00 01:00 15:00 15:00 03:00
Timer 2 (mm:ss) 10:00 ----- ----- ----- ----- ----- ----- 02:00
Additional Merck information
1) 1) 1) 1) 1) 1) 2) 3)
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!’’
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 48 of 49
Appendix: Measuring range with appendix Each test kit contains a package insert with detailed information regarding the standard procedure of the measurement. In some cases, however, special points have to be considered which are not mentioned in the package inserts. In this appendix you will find specific applications and modified methods for analyses for sensitive measurements indicated with “see appendix page 47”. Calcium: with Spectroquant® Calcium Test, Cat. No. 1.14815.0001 Measuring range: 1.0 – 15.0 mg/L Ca 2.5 – 37.5 mg/L CaCO3 1.4 – 21.0 mg/L CaO Procedure: Add 10 ml of reagent Ca-1A and 8 drops each of reagent Ca-2A and Ca-3A to 1 ml of the sample solution and mix. Leave to stand for 8 min for colour development. After filling into 10 mm cells measure in the photometer at 550 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 1 ml of distilled water instead of the sample solution. Fluoride: with Spectroquant® Fluoride Cell Test, Cat. No. 1.14557.0001 Measuring range: 0.025 – 0.500 mg/l F- Procedure: Pipette 10 ml of sample solution into the reaction cell, mix and add 1 blue microspoon of reagent F-1K. Close the cell tightly and dissolve the solid by shaking. Leave to stand for 5 min for colour development. After filling into 50 mm cells measure in the photometer at 620 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 10 ml of distilled water instead of the sample solution. Fluoride: with Spectroquant® Fluoride Cell Test, Cat. No. 1.00809.0001 Measuring range: 0.025 – 0.500 mg/l F- Procedure: Pipette 10 ml of sample solution into the reaction cell, mix and add 1 blue microspoon of reagent F-1K. Close the cell tightly and dissolve the solid by shaking. Leave to stand for 15 min for colour development. After filling into 50 mm cells measure in the photometer at 620 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 10 ml of distilled water instead of the sample solution Hydrogen peroxide: with Spectroquant® Hydrogen peroxide Cell Test, Cat. No. 1.14731.0001 Measuring range: 0.25 – 5.00 mg/l H2O2 Procedure: The test method of the package insert can be used without any alteration. After filling into 50 mm cells measure in the photometer at 410 nm
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 49 of 49
Sulfite: with Spectroquant® Sulfite Cell Test, Cat. No. 1.14394.0001 Measuring range: 0.005 – 3.00 mg/l SO3
2- Procedure: Place 1 microspoon of reagent SO3-1K in the cell, close tightly and shake until the complete dissolution of the solid. Then pipette 7 ml of the sample solution into the cell and mix. After the reaction time of 2 min fill into 50 cells and measure in the photometer at 410 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 7 ml of distilled water instead of the sample solution. Please contact our local Merck companies or Merck agents for any questions. Alternatively contact: [email protected]. No endorsement by Hach Company is implied or intended.