Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
COLLEGE OF ENGINEERING Chemical, Biological & Environmental Engineering
PROJECT OVERVIEW
MASS FLUX CHAMBER EXPERIMENTAL SETUP
Flux Chamber
Design
Acquire FluxChamberMaterials
Completed
In Progress
Upcoming
Assemble Flux
Chamber
Conduct Pressure &
PID Test
Analyze Results
Conduct Blank Test
Reevaluate Flux Chamber
Assembly
GC/MSAnalysis
DevelopSampling
Device
Figure 2.The work flow diagram illustrates the sequence of
tasks performed to accomplish objectives.
Objectives(1) Utilize existing test system (mass flux chamber) to
characterize vapor intrusion
(2) Prepare concrete samples to assemble with new flux
chamber components
(3) Conduct experiments to determine VOC flux through
concrete
OpportunityVolatile organic compounds (VOCs) in groundwater diffuse
through soil and concrete foundations to contaminate
indoor air. Measuring VOC concentration is the first step to
mitigating exposure risks to people. The ultimate goal is to
develop a passive sampling device for VOC vapor intrusion.
Figure 1. A typical site contaminated with a VOC such as
trichloroethylene (TCE) is shown with the contaminant diffusing into the
indoor air of a building, causing significant hazards to human health.
Figure 4. Experimental setup for PID testing. A) pump air inlet
B) Peristaltic pump C) Tedlar® bag (impermeable to VOC) D)
Mass flux chamber E) Top chamber air inlet
Figure 5. Tools used in experimentation. F) 5 and 100
mL gas-tight syringes G) Tedlar® bag H) PID.
How It Works
(1) VOC is loaded into the bottom chamber through the septum with a 5 mL
liquid syringe.
(2) VOC volatizes into a gas and diffuses through the concrete which is sealed
by heat shrink tubing to prevent gas from circumventing the sample.
(3) A constant flowrate of air is pumped through the top chamber inlet to
clear the heptane gas, providing a constant mass transfer driving force.
(4) Outlet gas is collected in a Tedlar® bag and VOC concentration is tested
using a Photoionization detector (PID). VOC concentration in the bottom
chamber is tested using gas-tight syringes through the septum.
Volatile Organic Compounds
Trichloroethylene (TCE)
Primary Uses
• Industrial Solvent
• Degreaser
• Found in adhesives, paint removers,
carpet cleaner
Regulations
• OSHA Standard is 100 ppm for 8-hour
exposure
Heptane (Experimental Substitute)
• Non-toxic VOC
• Similar molecular weight and vapor
pressure to TCE
• Easily detected using PID and GC/MS
Decane (Experimental Substitute)
• Non-toxic VOC
• Low Vapor Pressure Low
Concentration
• Easily detected using PID and GC/MS
Concrete
Figure 7. Different modes of diffusion in
concrete.
Project Work Flow
Figure 6. Concrete and aluminum blank that
are 1.5”–3.0” tall with a 2.75” diameter.
Table 1. Physical property comparison of VOC.
Acknowledgements• Mike Niemet, Ben Thompson, Mike Novak, Jake
Donally, Katie Rabe, Blake Wimer, CH2M Sponsors
• Columbia Concrete Sawing Co. - Concrete samples
• King Machine - Fabricated metal components
• Dr. Pommerenck - COMSOL Model support
• Dr. Semprini– Concrete diffusion and adsorption
• Dr. Gibson – Mathematical model
• Dr. Harding - Project guidance and support
References[1] Musielak, M. “Indoor Air Pollution: Study of
Trichloroethylene Transport through Soil and Basement.”
Université de Toulouse. 2010.
[2] Welty, JR. et al. “Fundamentals of Momentum, Heat,
and Mass Transfer.” Wiley and Sons, Inc. 5th ed. 2008.
[3] Green, Bradley A., Shea, D., and Ashton, A . "Use of
VOC Mass Flux to Estimate Vapor Intrusion Impacts."
Sanborn. Web. 2012.
Future Work• Develop passive sampling device for vapor intrusion
• Test flux chamber using different VOC’s and concrete
subslabs
• Field test of passive sampling device
Experimental Results• Pressurizing the chamber verified that a sufficient
seal was provided and no air could circumvent the
concrete.
• Replacing the concrete with an impermeable
aluminum cylinder during PID testing confirmed that
vapor was not circumventing the subslab.
• Steady state conditions for the heptane chamber
were observed in the system after 450 hours at a
value of 37 (±3) g/m2/day illustrated in Figure 9.
• The decane chamber data confirms the time to
reach steady state is longer and the steady state
flux value will be lower than heptane as a result of
lower vapor pressure also seen in Figure 9.
• PID reading accuracy of ±5% was verified using
GC/MS.
Figure 9. Steady state was observed at 450 hours for heptane
with a value of 37 g/m2/day. Steady state has not been
achieved with the decane chamber, but the value is expected
to be much less than the Heptane steady state value .
Mass Transfer Influenced by:
• Water/Cement Ratio
• Aggregate Size
• Cracking: Micro/Macro
“vapor Intrusion.” Navaf. <www.navfac.navy.mil>
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0
5
10
15
20
25
30
35
40
0 200 400 600 800 1000
Mass
Flu
x (
g/m
²/day)
Mass
Flu
x (
g/m
²/day)
Time (Hours)
Mass Flux of Heptane and Decane
Heptane Flux
Decane Flux
Vapor Intrusion Assessments for Concrete SubslabsJacob Lum, John Collin Hall, Amro Al-Habsi
Figure 3. Experimental setup for PID testing. A) Air flow inlet B) Air flow outlet C) Heat
shrink tubing D) Concrete subslab E) Syringe septums
COMSOL Multiphysics
Welty. Fundamentals of Momentum, Heat, and Mass Transfer.
Figure 9. A steady-state COMSOL concentration
simulation developed at 20ºC using a laminar flow
and species transport in porous media package.
• Porosity
• Diffusivity
VOC Contamination
Mass Flux Model:
𝑁𝐴 =𝐷𝐴,𝐵𝐶𝐴,𝑠𝐿−2𝐷𝐴,𝐵𝜋
𝑛=1
∞𝐶𝐴,𝑠𝜋
𝐿𝑐𝑜𝑠𝑛𝜋𝑧
𝐿𝑒𝑥𝑝(−𝐷𝐴,𝐵𝑛
2𝜋2𝐿−2𝑡)
Mathematical Model
Fick’s 2nd Law: 𝜕𝐶𝐴
𝜕𝑡= 𝐷𝐴,𝐵
𝜕2𝐶𝐴
𝜕𝑧2
IC : CA(x,0) = 0 for all x
BC(s): CA(0,t) = CAs for t > 0
CA(L,t) = 0 for t > 0
Heptane Flux
ChamberDecane Flux
Chamber
0.38
VOC Flux
2.0𝑥10−4
0.05
0.18
0.09
1.9
0.9
0.3
(𝒊𝒏𝒄𝒉𝒆𝒔)
(𝒊𝒏𝒄𝒉𝒆𝒔)
(𝒎𝒐𝒍/𝒎𝟑)