BA058D/06/en/10.03 50098512 Valid as of software version: V 1.06.XX (amplifier) V 1.03.XX (communication) PROline promass 80 Coriolis Mass Flow Mesuring System Description of Device Functions MASS FLOW VOLUME FLOW DENSITY TEMPERATURE UNIT VOLUME UNIT MASS FLOW UNIT MASS COMMISSIONING LANGUAGE ASSIGN LINE 1 SUM ASSIGN TOTALIZER ASSIGN CURRENT OPERATION MODE ASSIGN STATUS OUTPUT ASSIGN STATUS INPUT TAG NAME ASSIGN LF-CUTOFF INSTALLATION DIRECTION K-FACTOR ACTUAL SYSTEM CONDITION FAILSAFE MODE SERIAL NUMBER SW-REV. AMP. ACCESS CODE ASSIGN LINE 2 OVERFLOW CURRENT SPAN ASSIGN FREQUENCY ON-VALUE AKTIVE LEVEL TAG DESCRIPTION ON-VALUE MEASURING MODE ZERO POINT PREVIOUS SYST. CONDITION SIMULATION MEASURAND SW-REV.-NO. SENSOR SENSOR TYPE I/O-MODUL TYPE SW-REV. I/O-MODUL UNIT VOLUME FLOW DEFINE PRIVATE CODE 100% VALUE LINE 1 100% VALUE LINE 2 UNIT SUM VALUE 0_4 mA VALUE 20 mA END VALUE FREQUENCY OFF-VALUE MIN. PULSE WIDTH BUS ADDRESSE OFF-VALUE POSITVE ZERO RETURN NOMINAL DIAMETER ASSIGN SYSTEM ERROR VALUE SIMULATION STATUS ACCESS VALUE F LOW TIME CONSTANT SIMULATION STATUS HART PROTOCOL EPD SYSTEM DAMPING SENSOR TYPE ERROR CATEGORY MEASURING VALUES SYSTEM UNITS QUICK SETUP OPERATION USER INTERFACE TOTALIZER CURRENT OUTPUT PULSE/FREQ.- OUTPUT STATUS OUTPUT STATUS INPUT COMMUNICATION PROCESS PARAMETER SYSTEM PARAMETER SENSOR DATA SUPERVISION SIMULATION SYST. SENSOR VERSION AMP.VERSION MASS FLOW VOLUME FLOW DENSITY TEMPERATURE UNIT VOLUME UNIT MASS FLOW UNIT MASS COMMISSIONING LANGUAGE ASSIGN LINE 1 SUM ASSIGN TOTALIZER ASSIGN CURRENT OPERATION MODE ASSIGN STATUS OUTPUT ASSIGN STATUS INPUT TAG NAME ASSIGN LF-CUTOFF INSTALLATION DIRECTION K-FACTOR ACTUAL SYSTEM CONDITION FAILSAFE MODE SERIAL NUMBER SW-REV. AMP. ACCESS CODE ASSIGN LINE 2 OVERFLOW CURRENT SPAN ASSIGN FREQUENCY ON-VALUE AKTIVE LEVEL TAG DESCRIPTION ON-VALUE MEASURING MODE ZERO POINT PREVIOUS SYST. CONDITION SIMULATION MEASURAND SW-REV.-NO. SENSOR SENSOR TYPE I/O-MODUL TYPE SW-REV. I/O-MODUL UNIT VOLUME FLOW DEFINE PRIVATE CODE 100% VALUE LINE 1 100% VALUE LINE 2 UNIT SUM VALUE 0_4 mA VALUE 20 mA END VALUE FREQUENCY OFF-VALUE MIN. PULSE WIDTH BUS ADDRESSE OFF-VALUE POSITVE ZERO RETURN NOMINAL DIAMETER ASSIGN SYSTEM ERROR VALUE SIMULATION STATUS ACCESS VALUE F LOW TIME CONSTANT SIMULATION STATUS HART PROTOCOL EPD SYSTEM DAMPING SENSOR TYPE ERROR CATEGORY MEASURING VALUES SYSTEM UNITS QUICK SETUP OPERATION USER INTERFACE TOTALIZER CURRENT OUTPUT PULSE/FREQ.- OUTPUT STATUS OUTPUT STATUS INPUT COMMUNICATION PROCESS PARAMETER SYSTEM PARAMETER SENSOR DATA SUPERVISION SIMULATION SYST. SENSOR VERSION AMP.VERSION COMMISSIONING LANGUAGE ASSIGN LINE 1 SUM ASSIGN TOTALIZER ASSIGN CURRENT OPERATION MODE ASSIGN STATUS OUTPUT ASSIGN STATUS INPUT TAG NAME ASSIGN LF-CUTOFF INSTALLATION DIRECTION K-FACTOR ACTUAL SYSTEM CONDITION FAILSAFE MODE SERIAL NUMBER SW-REV. AMP. ACCESS CODE ASSIGN LINE 2 CURRENT SPAN ASSIGN FREQUENCY ON-VALUE AKTIVE LEVEL TAG DESCRIPTION ON-VALUE MEASURING MODE ZERO POINT PREVIOUS SYST. CONDITION SIMULATION MEASURAND SENSOR TYPE I/O-MODUL TYPE DE V F MIN BU P AS QUICK SETUP OPERATION USER INTERFACE TOTALIZER CURRENT OUTPUT PULSE/FREQ.- OUTPUT STATUS OUTPUT STATUS INPUT COMMUNICATION PROCESS PARAMETER SYSTEM PARAMETER SENSOR DATA SUPERVISION SIMULATION SYST. SENSOR VERSION AMP.VERSION

promass80

Embed Size (px)

DESCRIPTION

prmas 80

Citation preview

  • BA058D/06/en/10.0350098512

    Valid as of software version:V 1.06.XX (amplifier)V 1.03.XX (communication)

    PROline promass 80Coriolis Mass Flow Mesuring System

    Description of Device Functions

    MASS FLOW VOLUME FLOW DENSITY TEMPERATURE

    UNIT VOLUMEUNIT MASS FLOW UNIT MASS

    COMMISSIONING

    LANGUAGE

    ASSIGN LINE 1

    SUMASSIGNTOTALIZER

    ASSIGN CURRENT

    OPERATIONMODE

    ASSIGN STATUSOUTPUT

    ASSIGN STATUSINPUTTAG

    NAMEASSIGN

    LF-CUTOFFINSTALLATION

    DIRECTION

    K-FACTOR

    ACTUAL SYSTEMCONDITION

    FAILSAFE MODE

    SERIALNUMBERSW-REV.

    AMP.

    ACCESS CODE

    ASSIGN LINE 2

    OVERFLOW

    CURRENT SPAN

    ASSIGNFREQUENCY

    ON-VALUE

    AKTIVE LEVEL

    TAGDESCRIPTION

    ON-VALUE

    MEASURINGMODEZEROPOINT

    PREVIOUS SYST.CONDITIONSIMULATIONMEASURAND

    SW-REV.-NO.SENSOR

    SENSORTYPE

    I/O-MODULTYPE

    SW-REV.I/O-MODUL

    UNITVOLUME FLOW

    DEFINE PRIVATECODE

    100% VALUELINE 1

    100% VALUELINE 2

    UNIT SUM

    VALUE 0_4 mA VALUE 20 mA

    END VALUEFREQUENCY

    OFF-VALUE

    MIN. PULSE WIDTH

    BUS ADDRESSE

    OFF-VALUE

    POSITVE ZERORETURNNOMINAL

    DIAMETERASSIGN SYSTEM

    ERRORVALUE

    SIMULATION

    STATUS ACCESS

    VALUE F LOW

    TIME CONSTANT

    SIMULATIONSTATUS

    HART PROTOCOL

    EPD

    SYSTEMDAMPINGSENSOR

    TYPEERROR

    CATEGORY

    MEASURINGVALUES

    SYSTEM UNITS

    QUICK SETUP

    OPERATION

    USER INTERFACE

    TOTALIZER

    CURRENTOUTPUT

    PULSE/FREQ.-OUTPUTSTATUSOUTPUTSTATUSINPUT

    COMMUNICATION

    PROCESSPARAMETER

    SYSTEMPARAMETER

    SENSOR DATA

    SUPERVISION

    SIMULATION SYST.

    SENSOR VERSION

    AMP. VERSION

    MASS FLOW VOLUME FLOW DENSITY TEMPERATURE

    UNIT VOLUMEUNIT MASS FLOW UNIT MASS

    COMMISSIONING

    LANGUAGE

    ASSIGN LINE 1

    SUMASSIGNTOTALIZER

    ASSIGN CURRENT

    OPERATIONMODE

    ASSIGN STATUSOUTPUT

    ASSIGN STATUSINPUTTAG

    NAMEASSIGN

    LF-CUTOFFINSTALLATION

    DIRECTION

    K-FACTOR

    ACTUAL SYSTEMCONDITION

    FAILSAFE MODE

    SERIALNUMBERSW-REV.

    AMP.

    ACCESS CODE

    ASSIGN LINE 2

    OVERFLOW

    CURRENT SPAN

    ASSIGNFREQUENCY

    ON-VALUE

    AKTIVE LEVEL

    TAGDESCRIPTION

    ON-VALUE

    MEASURINGMODEZEROPOINT

    PREVIOUS SYST.CONDITIONSIMULATIONMEASURAND

    SW-REV.-NO.SENSOR

    SENSORTYPE

    I/O-MODULTYPE

    SW-REV.I/O-MODUL

    UNITVOLUME FLOW

    DEFINE PRIVATECODE

    100% VALUELINE 1

    100% VALUELINE 2

    UNIT SUM

    VALUE 0_4 mA VALUE 20 mA

    END VALUEFREQUENCY

    OFF-VALUE

    MIN. PULSE WIDTH

    BUS ADDRESSE

    OFF-VALUE

    POSITVE ZERORETURNNOMINAL

    DIAMETERASSIGN SYSTEM

    ERRORVALUE

    SIMULATION

    STATUS ACCESS

    VALUE F LOW

    TIME CONSTANT

    SIMULATIONSTATUS

    HART PROTOCOL

    EPD

    SYSTEMDAMPINGSENSOR

    TYPEERROR

    CATEGORY

    MEASURINGVALUES

    SYSTEM UNITS

    QUICK SETUP

    OPERATION

    USER INTERFACE

    TOTALIZER

    CURRENTOUTPUT

    PULSE/FREQ.-OUTPUTSTATUSOUTPUTSTATUSINPUT

    COMMUNICATION

    PROCESSPARAMETER

    SYSTEMPARAMETER

    SENSOR DATA

    SUPERVISION

    SIMULATION SYST.

    SENSOR VERSION

    AMP. VERSION

    COMMISSIONING

    LANGUAGE

    ASSIGN LINE 1

    SUMASSIGNTOTALIZER

    ASSIGN CURRENT

    OPERATIONMODE

    ASSIGN STATUSOUTPUT

    ASSIGN STATUSINPUTTAG

    NAMEASSIGN

    LF-CUTOFFINSTALLATION

    DIRECTION

    K-FACTOR

    ACTUAL SYSTEMCONDITION

    FAILSAFE MODE

    SERIALNUMBERSW-REV.

    AMP.

    ACCESS CODE

    ASSIGN LINE 2

    O

    CURRENT SPAN

    ASSIGNFREQUENCY

    ON-VALUE

    AKTIVE LEVEL

    TAGDESCRIPTION

    ON-VALUE

    MEASURINGMODEZEROPOINT

    PREVIOUS SYST.CONDITIONSIMULATIONMEASURAND

    SSENSORTYPE

    I/O-MODULTYPE I

    DEF

    1

    VA

    EF

    O

    MIN.

    BUS

    O

    PO

    DASS

    S

    QUICK SETUP

    OPERATION

    USER INTERFACE

    TOTALIZER

    CURRENTOUTPUT

    PULSE/FREQ.-OUTPUTSTATUSOUTPUTSTATUSINPUT

    COMMUNICATION

    PROCESSPARAMETER

    SYSTEMPARAMETER

    SENSOR DATA

    SUPERVISION

    SIMULATION SYST.

    SENSOR VERSION

    AMP. VERSION

  • Device Functions PROline Promass 80

    2 Endress+Hauser

  • Device Functions PROline Promass 80 Table of Contents

    Endress+Hauser 3

    Table of Contents

    1 Function matrix PROline Promass 80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 The function matrix: layout and use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.2 Graphical illustration of the function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    2 Group MEASURING VALUES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    3 Group SYSTEM UNITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    4 Group QUICK SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

    5 Group OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    6 Group USER INTERFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    7 Group TOTALIZER 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

    8 Group HANDLING TOTALIZER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    9 Group CURRENT OUTPUT 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

    10 Group PULSE/FREQUENCY OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

    11 Group STATUS OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3811.1 Information on the response of the status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4011.2 Switching action of the status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    12 Group STATUS INPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

    13 Group COMMUNICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

    14 Group PROCESS PARAMETER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

    15 Group SYSTEM PARAMETER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

    16 Group SENSOR DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

    17 Group SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

    18 Group SIMULATION SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

    19 Group SENSOR VERSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

    20 Group AMPLIFIER VERSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

    21 Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5821.1 SI units (not for USA and Canada) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5821.2 US units (only for USA and Canada) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

    22 Index of key words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

  • Table of Contents Device Functions PROline Promass 80

    4 Endress+Hauser

    Registered trademarks

    HART

    Registered trademark of HART Communication Foundation, Austin, USA

    S-DAT, T-DAT, F-CHIPRegistered trademarks of Endress+Hauser Flowtec AG, Reinach, CH

  • Device Functions PROline Promass 80 1 Function matrix PROline Promass 80

    Endress+Hauser 5

    1 Function matrix PROline Promass 80

    1.1 The function matrix: layout and use

    The function matrix is a two-level construct: the groups form one level, functions the other.The groups are the highest-level grouping of the control options for the measuring device.Each group comprises a number of functions.You select a group in order to access the individual functions for controlling or parameterizing the measuring device.

    You will find an overview of the groups in the table of contents on Page 3 and in the graphical representation of the function matrix on Page 6.You will also find an overview of the functions on Page 6, complete with the page references of the detailed function descriptions.The descriptions of the individual functions start on Page 7.

    Example of how to parameterize a function (in this case changing the language for the user interface): Entry into the function matrix (F-key). Select the OPERATION group. Select the LANGUAGE function, change the setting from ENGLISH to DEUTSCH

    with OS and save F (all texts displayed on the user interface appear in German). Exit the function matrix (ESC > 3 seconds).

    F-x0

    xxxx

    xx-19

    -xx

    -xx-

    xx-0

    00

    >3s

    - + EEsc

    E

    E

    E

    E

    E E E E E

    +

    +

    Esc

    +

    Esc

    +

    Esc

    E

  • 1 Function matrix PROline Promass 80 Device Functions PROline Promass 80

    6 Endress+Hauser

    1.2 Graphical illustration of the function matrixUN

    IT L

    ENG

    TH

    (p. 12

    )

    ACTU

    AL FR

    EQ.

    (p. 33

    )

    DEN

    SITY

    SE

    T VA

    LUE

    (p. 47

    )

    DEN

    SITY

    CO

    EF. C

    1(p.

    53

    )

    UNIT

    TEM

    PER

    ATUR

    E(p.

    11

    )

    TEST

    D

    ISPL

    AY(p.

    19

    )

    VALU

    E SI

    M. C

    URR.

    (p.

    28

    )

    FAIL

    SAFE

    VA

    LUE

    (p. 33

    )

    VALU

    E SI

    MUL

    ATIO

    N

    PULS

    E (p.

    37

    )

    ZERO

    PO

    INT

    ADJU

    ST

    (p. 47

    )

    DEN

    SITY

    CO

    EF. C

    0(p.

    53

    )

    OPE

    RAT

    ION

    HO

    URS

    (p. 55

    )

    UNIT

    R

    EF. D

    ENSI

    TY

    (p. 11

    )

    BACK

    LIG

    HT(p.

    19

    )

    SIM

    ULAT

    ION

    CU

    RR.

    (p. 27

    )

    FAIL

    SAFE

    M

    OD

    E(p.

    33

    )

    SIM

    ULAT

    ION

    PULS

    E(p.

    37

    )

    FIX.

    R

    EF. D

    ENSI

    TY(p.

    47

    )

    CAL.

    CO

    EF. KD

    2

    (p. 52

    )

    MAX

    . TE

    MP.

    CA

    RR

    .(p.

    53

    )

    SYST

    EM R

    ESET

    (p. 55

    )

    UNIT

    DEN

    SITY

    (p. 11

    )

    CON

    TRAS

    T LC

    D(p.

    19

    )

    ACTU

    AL CU

    RREN

    T(p.

    27

    )

    TIM

    E CO

    NST

    ANT

    (p. 32

    )

    FAIL

    SAFE

    M

    OD

    E(p.

    36

    )

    VAL.

    SI

    M. S

    WIT

    . PN

    T(p.

    39

    )

    EPD

    R

    ESPO

    NSE

    TIM

    E (p.

    46

    )

    CAL.

    CO

    EF. KD

    1

    (p. 52

    )

    MIN

    . TEM

    P. CA

    RR

    .(p.

    53

    )

    ALAR

    M D

    ELAY

    (p. 55

    )

    TEM

    PER

    ATUR

    E(p.

    7)

    UNIT

    CO

    RR

    . VO

    LUM

    E (p.

    10

    )

    DIS

    PLAY

    DA

    MPI

    NG

    (p. 19

    )

    RES

    ET T

    OTAL

    .

    (p. 21

    )

    FAIL

    SAFE

    M

    OD

    E(p.

    27

    )

    OUT

    PUT

    SIG

    NAL

    (p. 32

    )

    OUT

    PUT

    SIG

    NAL

    (p. 36

    )

    SIM

    . SW

    ITCH

    PO

    INT

    (p. 39

    )

    DEV

    ICE

    ID

    (p. 44

    )

    EPD

    VAL

    UE H

    IGH

    (p. 46

    )

    TEM

    P. CO

    EF. KT

    (p. 52

    )

    MAX

    . TE

    MP.

    M

    EAS.

    (p. 53

    )

    ERRO

    R CA

    TEG

    ORY

    (p. 55

    )

    REF

    EREN

    CE D

    ENSI

    TY

    (p. 7)

    UN

    IT CO

    RR

    . VO

    L.

    FLOW

    (p.

    10

    )

    ACCE

    SS CO

    DE

    COUN

    TER

    (p. 16

    )

    FOR

    MAT

    (p. 18

    )

    TOTA

    LIZE

    R M

    OD

    E(p.

    21

    )

    TIM

    E CO

    NST

    ANT

    (p. 27

    )

    VALU

    E F

    HIG

    H

    (p. 30

    )

    PULS

    E W

    IDTH

    (p. 35

    )

    ACTU

    AL ST

    ATUS

    (p. 39

    )

    VAL.

    SI

    M. ST

    AT. IN

    (p. 43

    )

    MAN

    UFAC

    TURE

    R ID

    (p. 44

    )

    EPD

    VA

    LUE

    LOW

    (p. 46

    )

    PRES

    SUR

    E(p.

    48

    )

    FLO

    W DA

    MPI

    NG

    (p.

    51

    )

    TEM

    P. CO

    EF. KM

    2

    (p. 52

    )

    MIN

    . TEM

    P. M

    EAS.

    (p. 53

    )

    ASSI

    GN

    PR

    OC.

    ER

    R.

    (p. 54

    )

    DEN

    SITY

    (p. 7)

    UNI

    T VO

    LUM

    E(p.

    9)

    STAT

    US AC

    CESS

    (p. 16

    )

    100%

    VA

    LUE

    (p. 18

    )

    UNIT

    TOT

    ALIZ

    ER(p.

    21

    )

    VALU

    E 20

    m

    A(p.

    24

    )

    VALU

    E F

    LOW

    (p. 30

    )

    PULS

    E VA

    LUE

    (p. 35

    )

    TIM

    E CO

    NST

    ANT

    (p. 39

    )

    SIM

    . ST

    ATUS

    IN

    (p. 42

    )

    HAR

    T PR

    OTO

    COL

    (p. 44

    )

    EMPT

    Y PI

    PE D

    ET.

    (p. 46

    )

    PRES

    SURE

    M

    OD

    E(p.

    48

    )

    DEN

    SITY

    DA

    MPI

    NG

    (p. 51

    )

    TEM

    P. CO

    EF. KM

    (p. 52

    )

    DEN

    SITY

    CO

    EF. C

    5(p.

    53

    )

    ERR

    OR

    CAT

    EGO

    RY

    (p. 54

    )

    SW-R

    EV. I/O

    (p. 57

    )

    COR

    R.V

    OL.

    -FL

    OW

    (p. 7)

    UNIT

    VO

    LUM

    E FL

    OW(p.

    9)

    DEF

    . PR

    IVAT

    E CO

    DE

    (p. 16

    )

    100%

    VA

    LUE

    (p. 18

    )

    OVER

    FLOW

    (p. 20

    )

    VALU

    E 0_

    4 m

    A(p.

    24

    )

    END

    VAL

    UE FR

    EQ.

    (p. 29

    )

    ASSI

    GN

    PU

    LSE

    (p. 34

    )

    OFF

    -VAL

    UE

    (p. 38

    )

    MIN

    . PUL

    SE W

    IDTH

    (p. 42

    )

    BUS

    ADD

    RES

    S(p.

    44

    )

    OFF

    -VAL

    . LF

    -CU

    T O

    FF(p.

    45

    )

    RES

    TORE

    O

    RIG

    INAL

    (p. 48

    )

    POS.

    ZE

    RO R

    ETUR

    N (p.

    50

    )

    NO

    MIN

    AL D

    IAM

    ETER

    (p. 52

    )

    DEN

    SITY

    CO

    EF. C

    4(p.

    53

    )

    ASSI

    GN

    SY

    S. ER

    ROR

    (p. 54

    )

    VALU

    E SI

    M. M

    EAS.

    (p. 56

    )

    SW-R

    EV. S-

    DAT

    (p. 57

    )

    I/O M

    OD

    UL TY

    PE(p.

    57

    )

    VOLU

    ME

    FLOW

    (p. 7)

    UNIT

    MAS

    S(p.

    8)

    ACCE

    SS CO

    DE

    (p. 15

    )

    ASSI

    GN

    LIN

    E 2

    (p. 17

    )

    SUM

    (p. 20

    )

    FAIL

    SAFE

    MO

    DE

    (p. 22

    )

    CURR

    ENT

    SPAN

    (p. 23

    )

    ASSI

    GN

    FR

    EQUE

    NCY

    (p. 29

    )

    VALU

    E SI

    M. FR

    EQ.

    (p. 34

    )

    ON-

    VALU

    E (p.

    38

    )

    ACTI

    VE LE

    VEL

    (p. 42

    )

    TAG

    D

    ESCR

    IPTI

    ON

    (p. 44

    )

    ON

    -VA

    L. LF

    -CU

    T O

    FF(p.

    45

    )

    DEN

    SITY

    AD

    JUST

    (p. 48

    )

    MEA

    SURI

    NG M

    OD

    E (p.

    49

    )

    ZERO

    PO

    INT

    (p. 52

    )

    DEN

    SITY

    CO

    EF. C

    3(p.

    53

    )

    PREV

    . SY

    S. CO

    ND

    (p. 54

    )

    SIM

    . MEA

    SURA

    ND

    (p.

    56

    )

    SEN

    SOR

    TYPE

    (p. 57

    )

    LAN

    GUA

    GE

    GRO

    UP(p.

    57

    )

    MAS

    S FL

    OW

    (p. 7)

    UNIT

    MAS

    S FL

    OW(p.

    8)

    UN

    IT PR

    ESSU

    RE

    (p. 12

    )

    QS CO

    MM

    ISSI

    ON

    (p.

    13

    )

    LAN

    GUA

    GE

    (p. 15

    )

    ASSI

    GN

    LI

    NE

    1(p.

    17

    )

    ASSI

    GN

    TOTA

    LIZE

    R(p.

    20

    )

    RES

    ET AL

    L TO

    TALI

    Z-ER

    S (p.

    22

    )

    ASSI

    GN

    CU

    RREN

    T (p.

    23

    )

    OPE

    RAT

    ION

    MO

    DE

    (p. 29

    )

    SIM

    ULAT

    ION

    FR

    EQ.

    (p. 34

    )

    ASSI

    GN

    ST

    ATUS

    (p. 38

    )

    ASSI

    GN

    ST

    ATUS

    IN

    (p. 42

    )

    TAG

    N

    AME

    (p. 44

    )

    ASSI

    GN

    LF

    -CU

    T O

    FF

    (p. 45

    )

    MEA

    SURE

    FL

    UID

    (p. 47

    )

    INST

    L. D

    IR. S

    ENSO

    R

    (p. 49

    )

    K-FA

    CTO

    R(p.

    52

    )

    DEN

    SITY

    CO

    EF. C

    2(p.

    53

    )

    ACTU

    AL SY

    S. CO

    ND.

    (p. 54

    )

    SIM

    . FA

    ILSA

    FE M

    OD

    E (p.

    56

    )

    SER

    IAL

    NUM

    BER

    (p.

    57

    )

    SW-R

    EV. AM

    P.

    (p. 57

    )

    MEA

    SUR

    ING

    VA

    LUES

    (p. 7)

    SYST

    EM UN

    ITS

    (p. 8)

    QUIC

    K SE

    TUP

    (p. 13

    )

    OPE

    RAT

    ION

    (p. 15

    )

    USER

    IN

    TER

    FACE

    (p. 17

    )

    TOTA

    LIZE

    R 1

    /2(p

    . 20

    )

    HA

    ND

    LIN

    G TO

    TALI

    ZER

    (p. 22

    )

    CURR

    ENT

    OUT

    PUT

    1/2

    (p. 23

    )

    PULS

    E/FR

    EQ. O

    UT.

    (p. 29

    )

    STAT

    US O

    UTPU

    T (p

    . 38

    )

    STAT

    US IN

    PUT

    (p. 42

    )

    COM

    MUN

    ICAT

    ION

    (p. 44

    )

    PRO

    CESS

    PA

    RA

    M.

    (p. 45

    )

    SYST

    EM PA

    RA

    MET

    ER

    (p. 49

    )

    SEN

    SOR

    DATA

    (p. 52

    )

    SUPE

    RVIS

    ION

    (p. 54

    )

    SIM

    ULAT

    . SY

    STEM

    (p. 56

    )

    SEN

    SOR

    VER

    SIO

    N(p

    . 57

    )

    AM

    P. H

    W V

    ERSI

    ON

    (p. 57

    )

  • Device Functions PROline Promass 80 2 Group MEASURING VALUES

    Endress+Hauser 7

    2 Group MEASURING VALUES

    Function description MEASURING VALUES

    ! Note! The engineering unit of the measured variable shown here can be set in the SYSTEM UNITS group

    (see Page 8). If the fluid in the pipe flows backwards, a negative sign prefixes the flow reading on the display.

    MASS FLOW In this function, the currently measured mass flow appears on the display.

    Display shows:5-digit floating-point number, including unit and sign(e.g. 462.87 kg/h; 731.63 lb/min; etc.)

    VOLUME FLOW In this function, the currently measured volumetric flow appears on the display. The volumetric flow is derived from the measured mass flowrate and the den-sity of the fluid measured.

    Display shows:5-digit floating-point number, including unit and sign(e.g. 5.5445 dm3/min; 1.4359 m3/h; 731.63 gal/d; etc.)

    CORRECTED VOLUME FLOW

    The calculated corrected volume flow appears on the display. The calculated corrected volume flow is derived from the measured mass flow and the refer-ence density of the fluid (density at reference temperature, measured or fixed entry).

    User Interface:5-digit floating-point number, including unit and sign (e.g. 1.3549 Nm3/h; 7.9846 scm/day; etc.)

    DENSITY In this function, the currently measured density or the specific gravity appears on the display.

    Display shows:5-digit fixed-point number, including unit(e.g. 1.2345 kg/dm3; 993.5 kg/dm3; 1.0015 SG_20 C; etc.)

    REFERENCE DENSITY The density of the fluid, at reference temperature, appears on the display. The reference density can be measured or also specified via the function FIXED REFERENCE DENSITY.

    User Interface:5-digit floating-point number, including unit, corresponding to 0.1000...6.0000 kg/dm3

    (e.g. 1.2345 kg/dm3; 993.5 kg/m3; 1.0015 SG_20 C; etc.)

    TEMPERATURE In this function, the currently measured temperature appears on the display.

    Display shows:5-digit floating-point number, including unit and sign(e.g. 23.4 C; 160.0 F; 295.4 K; etc.)

  • 3 Group SYSTEM UNITS Device Functions PROline Promass 80

    8 Endress+Hauser

    3 Group SYSTEM UNITS

    Function description SYSTEM UNITS

    You can select the unit for the measured variable in this function group.

    UNIT MASS FLOW Use this function to select the unit for displaying the mass flow (mass/time).

    The unit you select here is also valid for: Current output Frequency output Switching points (limit value for mass flow, flow direction) Low flow cut off

    Options:Metric:gram g/s; g/min; g/h; g/daykilogram kg/s; kg/min; kg/h; kg/dayMetric ton t/s; t/min; t/h; t/day

    US:ounce oz/s; oz/min; oz/h; oz/daypound lb/s; lb/min; lb/h; lb/dayton ton/s; ton/min; ton/h; ton/day

    Factory setting:Country dependent (kg/h or US lb/day)

    UNIT MASS Use this function to select the unit for displaying the mass.

    The unit you select here is also valid for: Pulse value (e.g. kg/p)

    Options:Metric g; kg; t

    US oz; lb; ton

    Factory setting:Country dependent (kg or US lb)

    ! Note!The unit for the totalizer is independent of your choice here, it is selected separately in the TOTALIZER group (see Page 20).

  • Device Functions PROline Promass 80 3 Group SYSTEM UNITS

    Endress+Hauser 9

    UNIT VOLUME FLOW Use this function to select the unit for the volume flow.

    The unit you select here is also valid for: Current output Frequency output Switching points (limit value for volume flow, flow direction) Low flow cut off

    Options:Metric:Cubic centimeter cm3/s; cm3/min; cm3/h; cm3/dayCubic decimeter dm3/s; dm3/min; dm3/h; dm3/dayCubic meter m3/s; m3/min; m3/h; m3/dayMilliliter ml/s; ml/min; ml/h; ml/dayLiter l/s; l/min; l/h; l/dayHectoliter hl/s; hl/min; hl/h; hl/dayMegaliter Ml/s; Ml/min; Ml/h; Ml/day

    US:Cubic centimeter cc/s; cc/min; cc/h; cc/dayAcre foot af/s; af/min; af/h; af/dayCubic foot ft3/s; ft3/min; ft3/h; ft3/dayFluid ounce oz f/s; oz f/min; oz f/h; oz f/dayGallon gal/s; gal/min; gal/h; gal/dayMillion gallon Mgal/s; Mgal/min; Mgal/h; Mgal/dayBarrel (normal fluids: 31.5 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/dayBarrel (beer: 31.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/dayBarrel (petrochemicals: 42.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/dayBarrel (filling tanks: 55.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day

    Imperial:Gallon gal/s; gal/min; gal/h; gal/dayMega gallon Mgal/s; Mgal/min; Mgal/h; Mgal/dayBarrel (beer: 36.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/dayBarrel (petrochemicals: 42.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day

    Factory setting:Country dependent (m3/h or US Mgal/day)

    UNIT VOLUME Use this function to select the unit for the volume.

    The unit you select here is also valid for: Pulse value (e.g. m3/p)

    Options:Metric cm3; dm3; m3; ml; l; hl; Ml

    US cc; af; ft3; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petrochemicals); bbl (filling tanks)

    Imperial gal; Mgal; bbl (beer); bbl (petrochemicals)

    Factory setting:Country dependent (m3 or US Mgal)

    ! Note!The unit for the totalizer is independent of your choice here, it is selected separately in the TOTALIZER 1/2 group (see Page 20).

    Function description SYSTEM UNITS

  • 3 Group SYSTEM UNITS Device Functions PROline Promass 80

    10 Endress+Hauser

    UNIT CORRECTED VOL-UME FLOW

    Use this function to select the unit for displaying the corrected volume flow (corrected volume/time).

    The unit you select here is also valid for: Current outputs Frequency outputs Relay switch points (limit value for corrected volume flow, flow direction) Low flow cut off

    Options:Metric:Nl/sNl/minNl/hNl/dayNm3/sNm3/minNm3/hNm3/day

    US:Sm3/s; Sm3/min; Sm3/h; Sm3/dayScf/s; Scf/min; Scf/h; Scf/day

    Factory setting:Nm3/h

    UNIT CORRECTED VOL-UME

    Use this function to select the unit for displaying the corrected volume.

    The unit you select here is also valid for: Pulse value (e.g. Nm3/p)

    Options:Metric:Nm3

    Nl

    US:Sm3

    Scf

    Factory setting:Nm3

    ! Note!The unit of the totalizers is independent of your choice here. The unit for each totalizer is selected separately for the totalizer in question.

    Function description SYSTEM UNITS

  • Device Functions PROline Promass 80 3 Group SYSTEM UNITS

    Endress+Hauser 11

    UNIT DENSITY Use this function to select the unit for displaying the fluid density.

    The unit you select here is also valid for: Current output Frequency output Switching points Density adjustment value Density responce value for EPD

    Options:Metric g/cm3; g/cc; kg/dm3; kg/l; kg/m3; SD 4 C, SD 15 C, SD 20 C; SG 4 C, SG 15 C, SG 20 C

    US lb/ft3; lb/gal; lb/bbl (normal fluids); lb/bbl (beer); lb/bbl (petrochemi-cals); lb/bbl (filling tanks)

    Imperial lb/gal; lb/bbl (beer); lb/bbl (petrochemicals)

    Factory setting:kg/l

    SD = Specific Density, SG = Specific GravityThe specific density is the ratio of fluid density to water (at water temperature = 4, 15, 20 C)

    UNIT REFERENCE DEN-SITY

    Use this function to select the unit for displaying the reference density.

    The unit you select here is also valid for: Current outputs Frequency outputs Relay switch points (limit value for density) Fixed reference density (for calculation of corrected volume flow)

    Options:Metric:kg/Nm3

    kg/Nl

    US:g/Scckg/Sm3

    lb/Scf

    Factory setting:kg/Nl

    UNIT TEMPERATURE Use this function to select the unit for displaying the temperature.

    The unit you select here is also valid for: Current output Frequency output Switching points

    Options:C (CELSIUS)

    K (KELVIN)F (FAHRENHEIT)R (RANKINE)

    Factory setting:C (CELSIUS)

    Function description SYSTEM UNITS

  • 3 Group SYSTEM UNITS Device Functions PROline Promass 80

    12 Endress+Hauser

    UNIT LENGTH Use this function to select the unit for the unit of length for nominal diameter.

    The unit you select here is also valid for: nominal diameter of the sensor (see the NOMINAL DIAMETER function on

    Page 52).

    Options:MILLIMETERINCH

    Factory setting:Country dependent (MILLIMETER or INCH)

    UNIT PRESSURE Use this function to select the unit for pressure.

    The unit you select here is also valid for: Specified pressure (see function DRUCK auf Page 48)

    Option:BAR GPSI GBAR APSI A

    Factory setting:BAR G

    Function description SYSTEM UNITS

  • Device Functions PROline Promass 80 4 Group QUICK SETUP

    Endress+Hauser 13

    4 Group QUICK SETUP

    ! Note! The display returns to the QUICK SETUP COMMISSION cell if you press the ESC key

    combination during programming of a parameter.

    Only units not yet configured in the current Quick Setup are offered for selectionin each cycle. The unit for mass and volume is derived from the corresponding flow unit.

    The YES option remains visible until all the units have been parameterised.NO is the only option displayed when no further units are available.

    Only the outputs not yet configured in the current Quick Setup are offered for selection in each cycle.

    The YES option remains visible until all the outputs have been parameterised.NO is the only option displayed when no further outputs are available.

    The automatic parameterization of the display option contains the following basicsettings/factory settings:YES: Main line = Mass flow; Additional line = Totalizer 1.NO: The existing (selected) settings remain.

    : see next page

    Function description QUICK SETUP

    QUICK SETUP COMMISSION

    Use this function to start the Setup menu for commissioning.

    Options:NOYES

    Factory setting:NO

  • 4 Group QUICK SETUP Device Functions PROline Promass 80

    14 Endress+Hauser

    ++ +E EEsc

    E- +

    Yes

    Yes

    Yes

    No

    No

    No

    Mass flow

    Current output 1

    Volume flow Density

    Frequency/Pulse output

    Selection System Units

    Selection output type

    Configuration another unit ?

    Configuration another output ?

    Automatically configuration display ?

    Temperature Quit

    Quit

    UnitMass flow

    AssignCurrent

    Current span

    Value0/4 mAValue20 mA

    Time constant

    Time constantFailsafe mode

    Failsafe mode

    Failsafe mode

    AssignFrequencyEnd valuefrequency

    ValueF Low

    ValueF High

    Output signal

    AssignPulse

    Pulse value

    Pulse width

    Output signal

    UnitflowVolume

    UnitDensity

    Operation mode

    UnitTemperature

    MeasuringMode

    UnitTotalizer

    UnitTotalizer

    Frequency Pulse

    Automaticallyparameterization

    of the display

    Language

    Pre-setting

    Quick Setup

    HOME-POSITION

    Quick SetupCommission

    AssignCurrent

    Current span

    Value0/4 mAValue20 mA

    Time constant

    Failsafe mode

    Current output 2

    Corr. Vol. flow

    UnitCorr. Vol. flow

    UnitRef. Density

    FixRef. Density

    F06-80xxxxx-19-xx-xx-en-000

  • Device Functions PROline Promass 80 5 Group OPERATION

    Endress+Hauser 15

    5 Group OPERATION

    Function description OPERATION

    LANGUAGE Use this function to select the language for all texts, parameters and mes-sages shown on the local display.

    ! Note!The displayed options depend on the available language group shown in the LANGUAGE GROUP (8226) function.

    Options: Language group WEST EU / USA:ENGLISHDEUTSCHFRANCAISESPANOLITALIANONEDERLANDSPORTUGUESE

    Language group EAST EU / SCAND:ENGLISHNORSKSVENSKASUOMIPOLISHRUSSIANCZECH

    Language group ASIA:ENGLISHBAHASA INDONESIAJAPANESE (syllabary)

    Factory setting:Country-dependent (see Page 58 ff.)

    ! Note! If you press the OS keys at startup, the language defaults to ENGLISH.

    ACCESS CODE All data of the measuring system are protected against inadvertent change. Programming is disabled and the settings cannot be changed until a code is entered in this function.If you press the OS key in any function the measuring system automatically goes to this function and the prompt to enter the code appears on the display (programming disabled).

    You can enable programming by entering the personal code (Factory setting = 80, see the DEFINE PRIVATE CODE function)

    User input:max. 4-digit number: 0...9999

    ! Note! The programming levels are disabled if you do not press a key within 60

    seconds following return to the HOME position. You can also disable programming in this function by entering any number

    (other than the defined private code). The Endress+Hauser service organization can be of assistance if you lose

    your personal code.

  • 5 Group OPERATION Device Functions PROline Promass 80

    16 Endress+Hauser

    DEFINE PRIVATE CODE Use this function to define a personal code number for enabling programming.

    User input:max. 4-digit number: 0...9999

    Factory setting:80

    ! Note! If you define the personal code number = 0, the programming is always

    enabled. Programming has to be enabled before this code can be changed.

    When programming is disabled the function cant be changed, this precaution prevents others from changing your personal code without your knowledge and consent.

    STATUS ACCESS Use this function to check the access status for the function matrix.

    Display shows:ACCESS CUSTOMER (Parameterization enabled)LOCKED (Parameterization disabled)

    ACCESS CODE COUNTER

    Displays how often the customer code, service code or the digit 0 (code-free) has been entered to gain access to the function matrix.

    Display:max. 7-digit number: 0...9999999

    Factory setting:0

    Function description OPERATION

  • Device Functions PROline Promass 80 6 Group USER INTERFACE

    Endress+Hauser 17

    6 Group USER INTERFACE

    Function description USER INTERFACE

    ASSIGN LINE 1 Use this function to define the display value assigned to the main line (the upper line of the local display) for display during normal measuring operation.

    Options:OFFMASS FLOWMASS FLOW IN %VOLUME FLOWVOLUME FLOW IN %DENSITYTEMPERATURETOTALIZER 1TOTALIZER 2CORRECTED VOLUME FLOWCORRECTED VOLUME FLOW IN %REFERENCE DENSITY

    Factory setting:MASS FLOW

    ASSIGN LINE 2 Use this function to define the display value assigned to the additional line (the bottom line of the local display) for display during normal measuring operation.

    Options:OFFMASS FLOWMASS FLOW IN %VOLUME FLOWVOLUME FLOW IN %DENSITYTEMPERATURETOTALIZER 1TAG NAMEOPERATION/SYSTEM CONDITIONDISPLAY FLOW DIRECTIONMASS FLOW BARGRAPH IN %VOLUME FLOW BARGRAPH IN %TOTALIZER 2CORRECTED VOLUME FLOWCORRECTED VOLUME FLOW IN %CORRECTED VOLUME FLOW BARGRAPH IN %REFERENCE DENSITY

    Factory setting:TOTALIZER

  • 6 Group USER INTERFACE Device Functions PROline Promass 80

    18 Endress+Hauser

    100% VALUE(Line 1) ! Note!

    This function is not available unless one of the following was selected in the ASSIGN LINE 1 function: MASS FLOW IN % VOLUME FLOW IN % CORRECTED VOLUME FLOW IN %

    Use this function to define the flow value to be shown on the display as the 100% value of the variable assigned to line 1.

    User input:5-digit floating-point number

    Factory setting: 10 kg/s (if MASS FLOW IN % or MASS FLOW BARGRAPH IN % is selected)

    10 l/s (if VOLUME FLOW IN % or VOLUME FLOW BARGRAPH IN % is selected)

    100% VALUE(Line 2) ! Note!

    This function is not available unless one of the following was selected in the ASSIGN LINE 2 function: MASS FLOW IN % VOLUME FLOW IN % MASS FLOW BARGRAPH IN % VOLUME FLOW BARGRAPH IN % CORRECTED VOLUME FLOW BARGRAPH IN %

    Use this function to define the flow value to be shown on the display as the 100% value of the variable assigned to line 2.

    User input:5-digit floating-point number

    Factory setting: 10 kg/s (if MASS FLOW IN % or MASS FLOW BARGRAPH IN % is selected)

    10 l/s (if VOLUME FLOW IN % or VOLUME FLOW BARGRAPH IN % is selected)

    FORMAT Use this function to define the maximum number of places after the decimal point displayed for the reading in the main line.

    Options:XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX

    Factory setting: X.XXXX

    ! Note! Note that this setting only affects the reading as it appears on the display, it

    has no influence on the accuracy of the system's calculations. The places after the decimal point as computed by the measuring device

    cannot always be displayed, depending on this setting and the engineering unit. In these instances an arrow appears on the display between the meas-ured value and the engineering unit (e.g. 1.2 kg/h), indicating that the measuring system is computing with more decimal places than can be shown on the display.

    Function description USER INTERFACE

  • Device Functions PROline Promass 80 6 Group USER INTERFACE

    Endress+Hauser 19

    DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to severely fluctuating flow variables, either very quickly (enter a low time con-stant) or with damping (enter a high time constant).

    User input:0...100 s

    Factory setting:1 s

    ! Note!Setting the time constant to zero seconds switches off damping.

    CONTRAST LCD Use this function to optimize display contrast to suit local operating conditions.

    User input:10...100%

    Factory setting:50%

    BACKLIGHT Use this function to optimize the backlight to suit local operating conditions.

    User input:0...100%

    ! Note!Entering the value 0 means that the backlight is switched off. The display then no longer emits any light, i.e. the display texts can no longer be read in the dark.

    Factory setting:50%

    TEST DISPLAY Use this function to test the operability of the local display and its pixels.

    Options:OFFON

    Factory setting:OFF

    Test sequence:1. Start the test by selecting ON.

    2. All pixels of the main line and additional line are darkened for minimum 0.75 second.

    3. Main line and additional line show an 8 in each field for minimum 0.75 second.

    4. Main line and additional line show a 0 in each field for minimum 0.75 second.

    5. Main line and additional line show nothing (blank display) for minimum 0.75 second.

    6. When the test completes the local display returns to its initial state and the setting changes to OFF.

    Function description USER INTERFACE

  • 7 Group TOTALIZER 1/2 Device Functions PROline Promass 80

    20 Endress+Hauser

    7 Group TOTALIZER 1/2

    Function description TOTALIZER 1/2

    ASSIGN TOTALIZER Use this function to assign a measured variable to the totalizer.

    Options:MASS FLOWVOLUME FLOWCORRECTED VOLUME FLOW

    Factory setting:MASS FLOW

    ! Note!The totalizer is reset to 0 as soon as the selection is changed.

    SUM Use this function to view the total for the totalizer measured variable aggregated since measuring commenced. The value can be positive or negative.

    Display shows:max. 7-digit floating-point number, including sign and unit (e.g. 15467.04 kg)

    ! Note!The totalizer response to faults is defined in the FAILSAFE MODE function (see Page 22).

    OVERFLOW Use this function to view the overflow for the totalizer aggregated since measuring commenced.

    Total flow quantity is represented by a floating decimal point number consisting of max. 7 digits. You can use this function to view higher

    numerical values (>9,999,999) as overflows. The effective quantity is thus the total of OVERFLOW plus the value returned by the SUM function.

    Example:Reading for 2 overflows: 2 E7 kg (= 20,000,000 kg)The value returned by the SUM function = 196,845.7 kgEffective total quantity = 20,196,845.7 kg

    Display shows: Integer with exponent, including sign and unit, e.g. 2 E7 kg

  • Device Functions PROline Promass 80 7 Group TOTALIZER 1/2

    Endress+Hauser 21

    UNIT TOTALIZER Use this function to define the unit for the totalizer measured variable, as selected beforehand.

    Options (for the MASS FLOW assignment):Metric g; kg; tUS oz; lb; ton

    Factory setting:Depends on nominal diameter and country,[value] / [g ...kg or US oz...US ton]corresponding to the totalizer unit factory setting (see Page 58 ff.)

    Options (for the VOLUME FLOW assignment):Metric cm3; dm3; m3; ml; l; hl; Ml

    US cc; af; ft3; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petro-chemicals); bbl (filling tanks)

    Imperial gal; Mgal; bbl (beer); bbl (petrochemicals)

    Options (for the CORRECTED VOLUME FLOW assignment):Metric Nl; Nm3

    US Sm3; Scf

    Factory setting:Depends on nominal diameter and country,[value] / [dm3 ...m3 or US gal...US Mgal]corresponding to the totalizer unit factory setting (see Page 58 ff.)

    TOTALIZER MODE Use this function to define how the flow components are to be totalized.

    Options:BALANCEPositive and negative flow components. The positive and negative flow com-ponents are balanced. In other words, net flow in the flow direction is regis-tered.

    FORWARD Only positive flow components are totalized.

    REVERSE Only negative flow components are totalized.

    Factory setting:Totalizer 1 = BALANCETotalizer 2 = FORWARD

    RESET TOTALIZER Use this function to reset the sum and the overflow of the totalizer to zero (= RESET).

    Options:NOYES

    Factory setting:NO

    ! Note!If the device has a status input and is appropriately configured, a reset for the totalizer can also be triggered by a pulse.

    Function description TOTALIZER 1/2

  • 8 Group HANDLING TOTALIZER Device Functions PROline Promass 80

    22 Endress+Hauser

    8 Group HANDLING TOTALIZER

    Function description HANDLING TOTALIZER

    RESET ALL TOTALIZ-ERS

    Use this function to reset the totals (including all overflows) of the totalizers (1...2) to zero (= RESET).

    Options:NOYES

    Factory setting:NO

    ! Note!If the device has a status input and if it is appropriately configured, a reset for the totalizer (1...2) can also be triggered by a pulse (see the ASSIGN STATUS INPUT function ).

    FAILSAFE MODE Use this function to define the response of all totalizers (1...2) to a fault.

    Options:STOPThe totalizer is paused until the fault is rectified. The totalizer stops at the last value prior to the occurrence of the error.

    ACTUAL VALUEThe totalizer continues to count is based on the current flow measured value. The fault is ignored.

    HOLD VALUEThe totalizer continues to count the flow is based on the last valid flow value (before the fault occurred).

    Factory setting:STOP

  • Device Functions PROline Promass 80 9 Group CURRENT OUTPUT 1/2

    Endress+Hauser 23

    9 Group CURRENT OUTPUT 1/2

    Function description CURRENT OUTPUT 1/2

    ASSIGN CURRENT Use this function to assign a measured variable to the current output.

    Options:OFF MASS FLOWVOLUME FLOWCORRECTED VOLUME FLOWDENSITYREFERENCE DENSITYTEMPERATURE

    Factory setting:MASS FLOW

    ! Note!If you select OFF, the only function shown in this group is these (ASSIGN CURRENT) function.

    CURRENT SPAN Use this function to define the current span. The selection specifies the opera-tional range and the lower and upper signal on alarm. For the current output 1 the option HART can be defined additionally.

    Options:020 mA420 mA420 mA HART (only current output1)420 mA NAMUR420 mA HART NAMUR (only current output1)420 mA US420 mA HART US (only current output1)020 mA (25 mA)420 mA (25 mA)420 mA (25 mA) HART (only current output1)

    Factory setting:420 mA HART NAMUR (fr current output 1)420 mA NAMUR (fr current output 2)

    ! Note! The option HART is only supported by the current output designated as cur-

    rent output 1 in the device software, (terminals 26 and 27). When switching the hardware from an active (factory setting) to a passive

    output signal select a current span of 420 mA (please refer to the Operat-ing Instructions PROline promass 80, BA 057D/06/en)

  • 9 Group CURRENT OUTPUT 1/2 Device Functions PROline Promass 80

    24 Endress+Hauser

    CURRENT SPAN(contd)

    Current span, operational range and signal on alarm level

    A = Current span1 = Operational range (measuring information)2 = Lower signal on alarm level3 = Upper signal on alarm level

    ! Note! If the measured value exceeds the measuring range (as defined in the func-

    tions VALUE 0_4 mA and VALUE 20 mA a notice message is generated (#351...352, current span).

    In case of a fault the behaviour of the current output is according to the selected option in the function FAILSAFE MODE. Change the error category in the function ASSIGN SYSTEM ERROR to generate a fault message instead of a notice message.

    VALUE 0_4 mA! Note!This function is not available unless the DENSITY, REFERENCE DENSITY or TEMPERATURE option was selected in the ASSIGN CURRENT function.

    Use this function to assign a value to the 0/4 mA current, (see Setting the span by means of the 0_4 mA and 20 mA value on Page 25).

    Option:5-digit floating-point number (with sign for the TEMPERATURE measured variable)

    Factory setting:0.5 [kg/l] or 50 [C]

    VALUE 20 mA Use this function to assign a value to the 20 mA current, (see Setting the span by means of the 0_4 mA and 20 mA value on Page 25).

    Option:5-digit floating-point number (with sign for the MASS FLOW, VOLUME FLOW, CORRECTED VOLUME FLOW and TEMPERATURE measured variables)

    Factory setting:Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [C]

    Function description CURRENT OUTPUT 1/2

    12 3

    I[mA]

    A 1 2 30-20 mA

    4-20 mA

    4-20 mA NAMUR

    4-20 mA US

    0-20 mA (25 mA)

    4-20 mA (25 mA)

    0 - 20.5 mA 0 22

    4 - 20.5 mA 2 22

    3.8 - 20.5 mA 3.5 22.6

    3.9 - 20.8 mA 3.75 22.6

    0 - 24 mA 0 25

    4 - 24 mA 2 25

    4-20 mA HART 4 - 20.5 mA 2 22

    4-20 mA HART NAMUR 3.8 - 20.5 mA 3.5 22.6

    4-20 mA HART US 3.9 - 20.8 mA 3.75 22.6

    4-20 mA (25 mA) HART 4 - 24 mA 2 25 F06-

    x3xx

    xxxx

    -05-

    xx-x

    x-xx

    -017

  • Device Functions PROline Promass 80 9 Group CURRENT OUTPUT 1/2

    Endress+Hauser 25

    Setting the span by means of the 0_4 mA and 20 mA value

    The span for the measured variable selected in the ASSIGN CURRENT function is specified via the VALUE 0_4 mA and VALUE 20 mA functions.

    The span is defined differently, depending on the measured variable selected:

    MASS FLOW, VOLUME FLOW and CORRECTED VOLUME FLOW The VALUE 0_4 mA function is not available; the value for the zero flow (0

    kg/h or 0m3/h) is assigned to the 0/4 mA current. The flow value for the 20 mA current is defined in the VALUE 20 mA function,

    (input range 99999 to +99999). The appropriate unit is taken from the UNIT MASS FLOW, UNIT VOLUME FLOW or UNIT CORR. VOL. FLOW function.

    Example (for standard measuring mode):

    = Initial value (0...20 mA) = Lower signal on alarm level:: depends on the setting in the function CURRENT SPAN = Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN = Maximum current value: depends on the setting in the function CURRENT SPAN = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT SPAN and FAILSAFE MODE,

    A = Measuring range (the minimum measuring range has to exceed the value that corre-lates with a flow velocity of 0.3 m/s)

    (Continued on next page)

    Function description CURRENT OUTPUT 1/2

    0

    3

    45

    6I [mA]

    2

    A

    1Q

    F06-

    80xx

    xxx-

    05-xx

    -xx-

    xx-00

    8

  • 9 Group CURRENT OUTPUT 1/2 Device Functions PROline Promass 80

    26 Endress+Hauser

    Setting the span by means of the 0_4 mA and 20 mA value(contd)

    DENSITY and REFERENCE DENSITY The density value for the 0/4 mA current is defined in the

    VALUE 0_4 mA function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY function.

    The density value for the 20 mA current is defined in the VALUE 20 mA function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY and UNIT REF. DENSITY function.

    Example (for standard measuring mode):

    = Initial value (0...20 mA) = Lower signal on alarm level:: depends on the setting in the function CURRENT SPAN = Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN = Maximum current value: depends on the setting in the function CURRENT SPAN = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT SPAN and FAILSAFE MODE

    A = Measuring range (the minimum measuring range has to exceed the value that corre-lates with a flow velocity of 0.3 m/s)

    TEMPERATURE The temperature value for the 0/4 mA current is defined in the

    VALUE 0_4 mA function, (input range 99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

    The temperature value for the 20 mA current is defined in the VALUE 20 mA function, (input range 99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

    ! Note!Values with different signs cannot be entered for the 0_4 mA and 20 mA val-ues if SYMMETRY is the option selected in the MEASURING MODE function, (see Page 49), The message INPUT RANGE EXCEEDED appears on the dis-play.

    Example (for standard measuring mode)

    = Initial value (0...20 mA) = Lower signal on alarm level:: depends on the setting in the function CURRENT SPAN = Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN = Maximum current value: depends on the setting in the function CURRENT SPAN = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT SPAN and FAILSAFE MODE

    A = Measuring range (the minimum measuring range has to exceed the value that corre-lates with a flow velocity of 0.3 m/s)

    Function description CURRENT OUTPUT 1/2

    0

    3

    45

    6I [mA]

    2

    A

    1Q

    F06-

    80xx

    xxx-

    05-x

    x-xx

    -xx

    -00

    9

    0

    3

    45

    6I [mA]

    2

    A

    1Q

    F06-

    80xx

    xxx-

    05-xx

    -xx

    -xx-

    010

  • Device Functions PROline Promass 80 9 Group CURRENT OUTPUT 1/2

    Endress+Hauser 27

    TIME CONSTANT Use this function to enter a time constant defining how the current output signal reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).

    User input:Fixed-point number 0.01...100.00 s

    Factory setting:1.00 s

    FAILSAFE MODE For safety reasons it is advisable to ensure that the current output assumes a predefined state in the event of a fault. The setting you select here affects only the current output. It has no effect on other outputs and the display (e.g. total-izers).

    Options: MIN. CURRENTThe current output adopts the value of the lower signal on alarm level (as defined in the function CURRENT SPAN.

    MAX. CURRENTThe current output adopts the value of the upper signal on alarm level (as defined in the function CURRENT SPAN.

    HOLD VALUE (not recommended)Measuring value output is based on the last measuring value saved before the error occurred .

    ACTUAL VALUE Measured value output is based on the current flow measurement.The fault is ignored .

    Factory setting:MIN. CURRENT

    ACTUAL CURRENT Use this function to view the computed value of the output current.

    Display shows:0.00...25.00 mA

    SIMULATION CURRENT Use this function to activate simulation of the current output.

    Options:OFFON

    Factory setting:OFF

    ! Note! The SIMULATION CURRENT OUTPUT message indicates that simulation

    is active. The measuring device continues to measure while simulation is in progress,

    i.e. the actual measured values are output correctly via the other outputs.

    " Caution!The setting is not saved if the power supply fails.

    Function description CURRENT OUTPUT 1/2

  • 9 Group CURRENT OUTPUT 1/2 Device Functions PROline Promass 80

    28 Endress+Hauser

    VALUE SIMULATION CURRENT ! Note!

    This function is not available unless the SIMULATION CURRENT function is active (= ON).Use this function to define a selectable value (e.g. 12 mA) to be output at the current output. This value is used to test downstream devices and the flow-meter itself.

    User input:floating-point number: 0.00...25.00 mA

    Factory setting:0.00 mA

    " Caution!The setting is not saved if the power supply fails.

    Function description CURRENT OUTPUT 1/2

  • Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

    Endress+Hauser 29

    10 Group PULSE/FREQUENCY OUTPUT

    Function description PULSE/FREQUENCY OUTPUT

    OPERATION MODE Use this function to configure the output as a pulse or frequency output.The functions available in this function group vary, depending on which option you select here.

    Options:PULSE FREQUENCY

    Factory setting:PULSE

    ASSIGN FREQUENCY! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    Use this function to assign a measured variable to the frequency output.

    Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOWDENSITYREFERENCE DENSITYTEMPERATURE

    Factory setting:MASS FLOW

    ! Note!If you select OFF, the only functions shown in this function group are ASSIGN FREQUENCY and OPERATION MODE.

    END VALUE FREQUENCY ! Note!

    This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    Use this function to define a full scale frequency for the frequency output.You define the associated measured value of the measuring range in the VALUE F HIGH function described on Page 30.

    User input:4-digit fixed-point number: 2...1000 Hz

    Factory setting:1000 Hz

    Example: VALUE F HIGH = 1000 kg/h, full scale frequency = 1000 Hz: i.e. a frequency

    of 1000 Hz is output at a flow of 1000 kg/h. VALUE F HIGH = 3600 kg/h, full scale frequency = 1000 Hz: i.e. a frequency

    of 1000 Hz is output at a flow of 3600 kg/h.

    ! Note!In the FREQUENCY operating mode the output signal is symmetrical (on/off ratio = 1:1). At low frequencies the pulse duration is limited to a maximum of 2 seconds, i.e. the on/off ratio is no longer symmetrical.

  • 10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

    30 Endress+Hauser

    VALUE F LOW! Note!This function is not available unless the DENSITY, REFERENCE DENSITY or TEMPERATURE option was selected in the ASSIGN FREQUENCY function.

    Use this function to assign a value to the start value frequency (0 Hz), (see Setting the span by means of the f-min. and f-max. value on Page 30).

    Option:5-digit floating-point number (with sign for the TEMPERATURE measured variable)

    Factory setting:0.5 [kg/l] or 50 [C]

    VALUE F HIGH Use this function to assign a value to the END VALUE FREQUENCY, (see Setting the span by means of the f-min. and f-max. value on Page 30).

    Option:5-digit floating-point number (with sign for the MASS FLOW, VOLUME FLOW, CORRECTED VOLUME FLOW and TEMPERATURE measured variables)

    Factory setting:Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [C]

    Setting the span by means of the f-min. and f-max. value

    The span for the measured variable selected in the ASSIGN FREQUENCY function is specified via the VALUE F LOW and VALUE F HIGH functions.

    The span is defined differently, depending on the measured variable selected:

    MASS FLOW, VOLUME FLOW and CORRECTED VOLUME FLOW The VALUE F LOW function is not available; the value for the zero flow

    (0 kg/h or 0m3/h) is assigned to the start value frequency. The flow value for the end value frequency is defined in the VALUE F HIGH

    function, (input range 99999 to +99999). The appropriate unit is taken from the UNIT MASS FLOW, UNIT VOLUME FLOW or UNIT CORRECTED VOL-UME FLOW function.

    Example (for standard measuring mode):

    = Flow value at which a frequency of 0 Hz should be output (preset, cannot be edited).

    = Flow value at which the, in the function END VALUE FREQUENCY defined, frequency should be output (entry in VALUE F HIGH function).

    (Continued on next page)

    Function description PULSE/FREQUENCY OUTPUT

    125

    100

    Measuring range Measured variable (amount)

    Value f minValue f max

    Frequency [%]

    0 F06-

    80xx

    xxxx

    -05-

    xx-x

    x-e

    n-01

    1

  • Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

    Endress+Hauser 31

    Setting the span by means of the f-min. and f-max. value(contd)

    DENSITY and REFERENCE DENSITY The density value for the start value frequency is defined in the

    VALUE F LOW function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY and UNIT REF. DENSITY function.

    The density value for the end value frequency is defined in the VALUE F HIGH function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY and UNIT REF. DENSITY function.

    Example (for standard measuring mode):

    = Density value and reference density at which a frequency of 0 Hz should be output (entry in VALUE F LOW function).

    = Density value and reference density at which the, in the function END VALUE FREQUENCY defined, frequency should be output (entry in VALUE F HIGH function).

    TEMPERATURE The temperature value for the 0/4 mA current is defined in the

    VALUE F LOW function, (input range 99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

    The temperature value for the 20 mA current is defined in the VALUE F HIGH function, (input range 99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

    ! Note!Values with different signs cannot be entered for the VALUE F LOW and VALUE F HIGH if SYMMETRY is the option selected in the MEASURING MODE function (see Page 49). The message INPUT RANGE EXCEEDED appears on the display.

    Example (for standard measuring mode):

    = Temperature value at which a frequency of 0 Hz should be output (entry in VALUE F LOW function).

    = Temperature value at which the, in the function END VALUE FREQUENCY defined, frequency should be output (entry in VALUE F HIGH function).

    Function description PULSE/FREQUENCY OUTPUT

    125

    100

    Measuring range Measured variable (amount)

    Value f minValue f max

    Frequency [%]

    0

    F06-

    80xx

    xxxx

    -05

    -xx-

    xx-e

    n-0

    12

    125

    100

    Measuring range Measured variable (amount)

    Value f minValue f max

    Frequency [%]

    0 F06-

    80xx

    xxxx

    -05-

    xx-xx

    -en

    -01

    3

  • 10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

    32 Endress+Hauser

    OUTPUT SIGNAL! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    Use this function to select the polarity of the frequency signal.

    Options:PASSIVE - POSITIVEPASSIVE - NEGATIVE

    Factory setting:PASSIVE - POSITIVE

    PASSIVE:

    ! Note!For continuous currents up to 25 mA (Imax = 250 mA / 20 ms)

    TIME CONSTANT! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    Use this function to enter a time constant defining how the frequency output signal reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).

    User input:floating-point number: 0.00...100.00 s

    Factory setting:0.00 s

    Function description PULSE/FREQUENCY OUTPUT

    1 5 8 37+

    -

    =

    OpenCollector

    U = 30 V DCExternal power supply

    max

    F-xx

    xxxx

    x-04

    -xx

    -xx

    -en

    -00

    0

    PASSIVE-NEGATIVE

    conducting conducting

    nonconducting

    nonconducting

    t t

    transistor transistor

    PASSIVE-POSITIVE

    F-xx

    xxxx

    x-05

    -xx

    -xx-

    en

    -00

    2

  • Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

    Endress+Hauser 33

    FAILSAFE MODE! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    For reasons of safety it is advisable to ensure that the frequency output assumes a predefined state in the event of a fault. Use this function to define this state. The setting you select here affects only the frequency output. It has no effect on other outputs and the display (e.g. totalizer).

    Options:FALLBACK VALUEOutput is 0 Hz.

    FAILSAFE LEVELOutput is the frequency specified in the FAILSAFE VALUE function.

    HOLD VALUEMeasured value output is based on the last measured value saved before the fault occurred.

    ACTUAL VALUEMeasured value output is based on the current flow measurement. The fault is ignored.

    Factory setting:FALLBACK VALUE

    FAILSAFE VALUE! Note!This function is not available unless FREQUENCY was selected in the OPERATION MODE function and FAILSAFE LEVEL was selected in the FAILSAFE MODE function.

    Use this function to define frequency that the measuring device outputs in the event of an fault.

    User input:max. 4-digit number: 0...1250 Hz

    Factory setting:1250 Hz

    ACTUAL FREQUENCY! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    Use this function to view the computed value of the output frequency.

    Display shows:0...1250 Hz

    Function description PULSE/FREQUENCY OUTPUT

  • 10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

    34 Endress+Hauser

    SIMULATION FREQUENCY ! Note!

    This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

    Use this function to activate simulation of the frequency output.

    Options:OFFON

    Factory setting:OFF

    ! Note! The SIMULATION FREQUENCY OUTPUT message indicates that

    simulation is active. The measuring device continues to measure while simulation is in progress,

    i.e. the currently measured values are output correctly via the other outputs.

    " Caution!The setting is not saved if the power supply fails.

    VALUE SIMULATION FREQUENCY ! Note!

    This function is not available unless FREQUENCY was selected in the OPERATION MODE function and the VALUE SIMULATION FREQUENCY function is active (= ON).

    Use this function to define a selectable frequency value (e.g. 500 Hz) to be output at the frequency output. This value is used to test downstream devices and the flowmeter itself.

    User input:0...1250 Hz

    Factory setting:0 Hz

    " Caution!The setting is not saved if the power supply fails.

    ASSIGN PULSE! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

    Use this function to assign a measured variable to the pulse output.

    Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOW

    Factory setting:MASS FLOW

    ! Note!If you select OFF, the only functions shown in this function group are ASSIGN PULSE and OPERATION MODE.

    Function description PULSE/FREQUENCY OUTPUT

  • Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

    Endress+Hauser 35

    PULSE VALUE! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

    Use this function to define the flow at which a pulse is triggered. These pulses can be totaled by an external totalizer, and the total flow quantity since measuring started can be registered in this way.

    User input:5-digit floating-point number

    Factory setting:Depends on nominal diameter and country,[value] [dm3 ...m3 or US gal...US Mgal] / pulsescorresponding to the factory setting for the pulses value (see Page 58 ff.)

    ! Note!The appropriate unit is taken from the corresponding UNIT function in the group SYSTEM UNITS (see Page 8).

    PULSE WIDTH! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

    Use this function to enter the maximum pulse width of the output pulses.

    User input:0.5...2000 ms

    Factory setting:100 ms

    Pulse output is always with the pulse width (B) entered in this function.The intervals (P) between the individual pulses are automatically configured. However, they must at least correspond to the pulse width (B = P).

    B = Pulse width entered (the illustration applies to positive pulses)P= Intervals between the individual pulses

    ! Note!When entering the pulse width, select a value that can still be processed by an external totalizer (e.g. mechanical totalizer, PLC, etc.).

    " Caution!If the pulse number or frequency resulting from the pulse value entered, (see function PULSE VALUE on Page 35), and from the current flowis too large to maintain the pulse width selected (interval P is smaller than the pulse width B entered), a system error message (pulse memory) is generated after buffering/balancing time.

    Function description PULSE/FREQUENCY OUTPUT

    B=P

    B

    PP

    B

    B< P

    F06-xxxxxxxx-05-xx-xx-xx-012

  • 10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

    36 Endress+Hauser

    OUTPUT SIGNAL! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

    Use this function to configure the output in such a way that it for example matches an external totalizer. According to the application, the direction of the pulses can be selected here.

    Options:PASSIVE - POSITIVEPASSIVE - NEGATIVE

    Factory setting:PASSIVE - POSITIVE

    PASSIVE:

    ! Note!For continuous currents up to 25 mA (Imax = 250 mA / 20 ms)

    FAILSAFE MODE! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

    For reasons of safety it is advisable to ensure that the pulse output assumes a predefined state in the event of a fault. Use this function to define this state. The setting you select here affects only the pulse output. It has no effect on other outputs and the display (e.g. totalizer).

    Options:FALLBACK VALUEOutput is 0 pulse.

    HOLD VALUEMeasured value output is based on the last measured value saved before the fault occurred.

    ACTUAL VALUEMeasured value output is based on the current flow measurement. The fault is ignored.

    Factory setting:FALLBACK VALUE

    Function description PULSE/FREQUENCY OUTPUT

    1 5 8 37+

    -

    =

    OpenCollector

    U = 30 V DCExternal power supply

    max

    F-xx

    xxxx

    x-04

    -xx

    -xx

    -en

    -00

    0

    B B

    PASSIVE-POSITIVEpulses

    conducting conducting

    nonconducting

    nonconducting

    t t

    transistor transistor

    PASSIVE-NEGATIVEpulses

    B = Pulse width

    F-xx

    xxxx

    x-05

    -xx

    -xx-

    en

    -00

    1

  • Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

    Endress+Hauser 37

    SIMULATION PULSE ! Note!This function is not available unless the PULSE option was selected in the OPERATING MODE function.

    Use this function to activate simulation of the pulse output.

    Options: OFFCOUNTDOWNThe pulses specified in the VALUE SIMULATION PULSE function are output.

    CONTINUOUSLYPulses are continuously output with the pulse width specified in the PULSE WIDTH function. Simulation is started once the CONTINUOUSLY option is confirmed with the F key.

    ! Note!Simulation is started by confirming the CONTINUOUSLY option with the F key.The simulation can be switched off again via the SIMULATION PULSE function.

    Factory setting:OFF

    ! Note! The notice message #631 SIM. PULSE indicates that simulation is active. The on/off ratio is 1:1 for both types of simulation. The measuring device continues to measure while simulation is in progress,

    i.e. the current measured values are output correctly via the other outputs.

    " Caution!The setting is not saved if the power supply fails.

    VALUE SIMULATION PULSE ! Note!

    This function is not available unless the COUNTDOWN option was selected in the SIMULATION PULSE function.

    Use this function to specify the number of pulses (e.g. 50) which are output during the simulation. This value is used to test downstream devices and the measuring device itself. The pulses are output with the pulse width specified in the PULSE WIDTH function. The on/off ratio is 1:1.

    Simulation is started once the specified value is confirmed with the F key. The display remains at 0 if the specified pulses have been output.

    User input:0...10 000

    Factory setting:0

    ! Note!Simulation is started by confirming the simulation value with the F key.The simulation can be switched off again via the SIMULATION PULSE function.

    " Caution!The setting is not saved if the power supply fails.

    Function description PULSE/FREQUENCY OUTPUT

  • 11 Group STATUS OUTPUT Device Functions PROline Promass 80

    38 Endress+Hauser

    11 Group STATUS OUTPUT

    Function description STATUS OUTPUT

    This group is not available unless the measuring device is fitted with a status output.

    ASSIGN STATUS Use this function to assign a switching function to the status output.

    Options:OFFON (operation)FAULT MESSAGENOTICE MESSAGEFAULT MESSAGE or NOTICE MESSAGEEMPTY PIPE DETECTION (only if function is active)FLOW DIRECTIONLIMIT MASS FLOWLIMIT VOLUME FLOWLIMIT CORRECTED VOLUME FLOWLIMIT DENSITYLIMIT REFERENCE DENSITYLIMIT TEMPERATURE

    Factory setting:FAULT MESSAGE

    ! Note! The behaviour of the status output is of the quiescent-current type, in other

    words the output is closed (transistor conductive) when normal, error-free measuring is in progress.

    Please read and comply with the information on the switching characteris-tics of the status output (see Page 40, 41).

    If you select OFF, the only function shown in this function group is this func-tion, in other words ASSIGN STATUS.

    ON-VALUE! Note!This function is not available unless LIMIT MASS FLOW, LIMIT VOLUME FLOW, LIMIT CORRECTED VOLUME FLOW, LIMIT TEMPERATURE, LIMIT DENSITY, LIMIT REFERENCE DENSITY or FLOW DIRECTION was selected in the ASSIGN STATUS function.

    Use this function to assign a value to the switch-on point (status output conductive). The value can be greater or less than the switch-off point.Positive and negative values are permissible.

    User input:5-digit floating-point number

    Factory setting:0 [kg/h] or 0 [m3/h] or 2 [kg/l] or 200 [C]

    OFF-VALUE! Note!This function is not available unless LIMIT MASS FLOW, LIMIT VOLUME FLOW, LIMIT CORRECTED VOLUME FLOW, LIMIT TEMPERATURE, LIMIT DENSITY, LIMIT REFERENCE DENSITY was selected in the ASSIGN STATUS function.

    Use this function to assign a value to the switch-off point (status output not conductive). The value can be greater or less than the switch-on point.Positive and negative values are permissible.

    User input:5-digit floating-point number

    Factory setting:0 [kg/h] or 0 [m3/h] or 2 [kg/l] or 200 [C]

  • Device Functions PROline Promass 80 11 Group STATUS OUTPUT

    Endress+Hauser 39

    TIME CONSTANT! Note!This function is not available unless one of the following was selected in the ASSIGN STATUS function:FLOW DIRECTIONLIMIT MASS FLOWLIMIT VOLUME FLOWLIMIT CORRECTED VOLUME FLOWLIMIT DENSITYLIMIT REFERENCE DENSITYLIMIT TEMPERATURE

    Use this function to enter a time constant defining how the status output reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant). The purpose of damping, therefore, is to prevent the status output changing state continuously in response to fluctuations in flow.

    User input:5-digit floating-point number: 0.00...100.00 s

    Factory setting:0.00 s

    ACTUAL STATUS Use this function to check the current status of the status output.

    Display shows:NOT CONDUCTIVECONDUCTIVE

    SIMULATIONSWITCH POINT

    Use this function to activate simulation of the status output.

    Options:OFFON

    Factory setting:OFF

    ! Note! The SIMULATION STATUS OUTPUT message indicates that simulation is

    active. The measuring device continues to measure while simulation is in progress,

    i.e. the current measured values are output correctly via the other outputs.

    " Caution!The setting is not saved if the power supply fails.

    VALUE SIMULATION SWITCH POINT ! Note!

    This function is not available unless the ON setting was selected in the SIMU-LATION SWITCH POINT function.

    Use this function to define the switching response of the status output during the simulation. This value is used to test downstream devices and the flow-meter itself.

    User input:NOT CONDUCTIVECONDUCTIVE

    Factory setting:NOT CONDUCTIVE

    " Caution!The setting is not saved if the power supply fails.

    Function description STATUS OUTPUT

  • 11 Group STATUS OUTPUT Device Functions PROline Promass 80

    40 Endress+Hauser

    11.1 Information on the response of the status output

    GeneralIf you have configured the status output for LIMIT or FLOW DIRECTION, you can define the requisite switching points in the ON-VALUE and OFF-VALUE functions. When the measured variable in question reaches one of these predefined values, the status output signal switches as shown in the illustrations below.

    Status output configured for direction of flowThe value you entered in the ON-VALUE function defines the switching points for the positive and negative directions of flow. If, for example, the switching point you define is = 1 kg/h, the status output is not conductive at 1 kg/h and is conductive at +1 kg/h. Set the switching point to 0 if your process calls for direct switchover (no switching hysteresis). If low flow cut off is used, it is advisable to set hysteresis to a value greater than or equal to the low flow cut off rate.

    Switch-off / switch-on point

    a = Status output conductiveb = Status output not conductive

    Status output configured for limit valueThe status output signal switches as soon as the measured variable falls below or exceeds a defined switching point. Application: Monitoring flow or process-related boundary conditions.

    Measured variable

    = ON OFF-VALUE (maximum safety) = ON > OFF-VALUE (minimum safety) = Status output off (not conductive)

    F-xx

    xxxx

    x-05

    -xx

    -xx-

    xx-0

    04

    t

    F-xx

    xxxx

    x-05

    -xx

    -xx-

    xx-00

    5

  • Device Functions PROline Promass 80 11 Group STATUS OUTPUT

    Endress+Hauser 41

    11.2 Switching action of the status output

    Function State Open Collector (Transistor)ON (operation) System in measuring mode

    conductive

    System not in measuring mode (power supply failure) not

    conductive

    Fault message System OK

    conductive

    (System or process error)Fault Error response of outputs, inputs and totalizer

    not conductive

    Notice message System OK

    conductive

    (System or process error)Fault Continuation of measuring

    not conductive

    Fault messageorNotice message

    System OK

    conductive

    (System or process error)Fault Response to error or Info Continuation of measuring

    not conductive

    Empty pipe detection (EPD)

    Fluid density above response level, e.g. full measuring tube

    conductive

    Fluid density below response level, e.g. empty measuring tube

    not conductive

    Flow direction forward

    conductive

    reversenot

    conductive

    Limit value mass flow volume flow corrected vol-

    ume flow density corrected density temperature

    Limit value not overshot or undershot

    conductive

    Limit value overshot or undershot not

    conductive

  • 12 Group STATUS INPUT Device Functions PROline Promass 80

    42 Endress+Hauser

    12 Group STATUS INPUT

    Function description STATUS INPUT

    This group is not available unless the measuring device is fitted with a status input I/O module.

    ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.

    Options:OFFRESET TOTALIZER 1POSITIVE ZERO RETURNZEROPOINT ADJUSTMENTRESET TOTALIZER 2RESET ALL TOTALIZERS

    Factory setting:OFF

    ! Note!Positive zero return is active as long as the active level is available at the status input (continuous signal). All other assignments react to a change in level (pulse) at the status input.

    ACTIVE LEVEL Use this function to define whether the assigned function (see ASSIGN STATUS INPUT function) is released when the signal level is present (HIGH) or not present (LOW).

    Options:HIGHLOW

    Factory setting:HIGH

    MINIMUM PULSE WIDTH

    Use this function to define a minimum width which the input pulse must achieve in order to trigger the defined switching function.

    User input:20...100 ms

    Factory setting:50 ms

    SIMULATION STATUS INPUT

    Use this function to activate simulation of the status input, in other words to trigger the function assigned to the status input (see the ASSIGN STATUS INPUT function on Page 38).

    Options: OFFON

    Factory setting:OFF

    ! Note! The SIMULATION STATUS INPUT message indicates that simulation is

    active. The measuring device continues to measure while simulation is in progress,

    i.e. the current measured values are output correctly via the outputs.

    " Caution!The setting is not saved if the power supply fails.

  • Device Functions PROline Promass 80 12 Group STATUS INPUT

    Endress+Hauser 43

    VALUE SIMULATION STATUS INPUT ! Note!

    This function is not available unless the ON setting was selected in the SIMULATION STATUS INPUT function.

    Use this function to select the level to be assumed at the status input during the simulation.

    Options:HIGHLOW

    Factory setting:LOW

    " Caution!The setting is not saved if the power supply fails.

    Function description STATUS INPUT

  • 13 Group COMMUNICATION Device Functions PROline Promass 80

    44 Endress+Hauser

    13 Group COMMUNICATION

    Function description COMMUNICATION

    TAG NAME Use this function to enter a tag name for the measuring device. You can edit and read this tag name at the local display or via the HART protocol.

    User input:max. 8-character text, permissible: A-Z, 0-9, +,, punctuation marks

    Factory setting:_ _ _ _ _ _ _ _ (without text)

    TAG DESCRIPTION Use this function to enter a tag description for the measuring device. You can edit and read this tag description at the local display or via the HART protocol.

    User input:max. 16-character text, permissible: A-Z, 0-9, +,, punctuation marks

    Factory setting:_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ (without text)

    BUS ADDRESS Use this function to define the address for the exchange of data with the HART protocol.

    User input:0...15

    Factory setting:0

    ! Note!Addresses 1...15: a constant 4 mA current is applied.

    HART PROTOCOL Use this function to display if the HART protocol is active.

    Anzeige:OFF = HART protocol not aktiveON = HART protocol aktive

    ! Note!The HART protocol can be activated with the selection 4-20 mA HART resp. 4-20 mA (25 mA) HART in the CURRENT SPAN function (see Page 23).

    MANUFACTURER ID Use this function to view the manufacturer ID in decimal numerical format.

    Display shows:Endress+Hauser17 = ( 11 hex) for Endress+Hauser

    DEVICE ID Use this function to view the device ID in hexadecimal numerical format.

    Display shows:50 = ( 80 dez) for Promass 80

  • Device Functions PROline Promass 80 14 Group PROCESS PARAMETER

    Endress+Hauser 45

    14 Group PROCESS PARAMETER

    Function description PROCESS PARAMETER

    ASSIGN LOW FLOW CUT OFF

    Use this function to assign the switching point for the low flow cut off.

    Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOW

    Factory setting:MASS FLOW

    ON VALUE LOW FLOW CUT OFF

    Use this function to assign the on value for the low flow cut off. Low flow cut off is active if the setting is a value not equal to 0. The sign of the flow value is highlighted on the display to indicate that low flow cut off is active.

    User input:5-digit floating-point number

    Factory setting:depends on nominal diameter

    ! Note!The appropriate unit is taken from the corresponding UNIT function in the group SYSTEM UNITS (see Page 8).

    OFF VALUE LOW FLOW CUT OFF

    Use this function to enter the switch-off point for low flow cut off.Enter the switch-off point as a positive hysteresis value from the switch-on point.

    User input:Integer 0...100%

    Factory setting:50%

    Example:

    Q = Flow [volume/time]t = TimeH = Hysteresisa = ON VALUE LOW FLOW CUT OFF = 200 g/hb = OFF VALUE LOW FLOW CUT OFF = 10%c = Low flow cut off active1 = Low flow cut off is switched on at 200 g/h2 = Low flow cut off is switched off at 220 g/h

    1

    c c

    Q

    t

    1

    2b

    a

    2

    H

    F06-

    80xx

    xxxx

    -05-

    xx-x

    x-xx

    -007

  • 14 Group PROCESS PARAMETER Device Functions PROline Promass 80

    46 Endress+Hauser

    EMPTY PIPE DETECTION (EPD)

    Use this function to activate the empty pipe detection (EPD). With empty measuring tubes the density of the fluid falls below a specified value (see EPD VALUE LOW function).

    Options: OFFON

    Factory setting:OFF

    " Caution! Select a correspondingly low EPD VALUE LOW so that the difference to the

    effective density of the fluid is sufficiently large enough.This ensures that totally empty measuring tubes and not partially filled ones are detected.

    For gas measurement we strongly recommend to switch off empty pipe detection.

    EPD VALUE LOW! Note!This function is not available unless the ON selection was selected in the EMPTY PIPE DETECTION (EPD) function.

    Use this function to set an lower threshold for the measured density value, in order to detect possible problems in the process indicated by too low density.

    User input:5-digit floating-point number

    Factory setting:0.2000 g/cc

    EPD VALUE HIGH! Note!This function is not available unless the ON selection was selected in the EMPTY PIPE DETECTION (EPD) function.

    Use this function to set an upper threshold for the measured density value.

    User input:5-digit floating-point number

    Factory setting:6.0000 g/cc

    EPD RESPONSE TIME Use this function to enter the time span for which the criteria for an empty pipe have to be satisfied without interruption before a notice message or fault mes-sage is generated.

    User input:fixed-point number: 1.0...60.0 s

    Factory setting:1.0 s

    Function description PROCESS PARAMETER

  • Device Functions PROline Promass 80 14 Group PROCESS PARAMETER

    Endress+Hauser 47

    FIXED REFERENCE DENSITY

    In this function, a fixed value for the reference density can be entered, with which the corrected volume flow or corrected volume is calculated.

    User input:5-digit floating-point number

    Factory setting:1 kg/Nl

    ZERO POINT ADJUST This function enables a zero point adjustment to be automatically carried out. The new zero point determined by the measuring system is adopted by the ZERO POINT function (see Page 52).

    Options:CANCELSTART

    Factory setting:CANCEL

    " Caution!Before carrying out the calibration, please refer to BA 057D/06/en Promass 80 Operating Instructions where a detailed description of the zero point adjust-ment is given.

    ! Note! Programming is locked during zero point adjustment and the display shows:

    ZERO ADJUST RUNNING. If the zero point adjustment is not possible, e.g. with a flow velocity

    > 0.1 m/s, or has been cancelled, then the alarm message ZERO ADJUST NOT POSSIBLE is shown on the display.

    If the Promass 80 measuring electronics are fitted with a status input, then the zero point can also be activated by using this input.

    After Zero point adjustment is completed, the new zero point can be called up with theF key. If the F key is pressed again, you return to the ZERO POINT ADJUST function.

    DENSITY SET VALUE In this function, enter the density set value of the particular fluid for which you want to carry out a field density adjustment.

    User input:5 digit floating-point number, incl. units (corresponding to 0.1...5.9999 kg/l)

    ! Note! The preset density entered here should not vary from the actual fluid density

    by a more than 10%. The appropriate unit is taken from the corresponding UNIT function in the

    group SYSTEM UNITS (see Page 8).

    MEASURE FLUID In this function the actual density of the fluid is measured for the density adjustment.

    Options:CANCELSTART

    Function description PROCESS PARAMETER

  • 14 Group PROCESS PARAMETER Device Functions PROline Promass 80

    48 Endress+Hauser

    DENSITY ADJUST With this function a density adjustment can be carried out on site. The density set value will thus be recalculated and stored. This ensures that the values dependent on density calculations (e.g. volume flow) are as accurate as possi-ble.

    " Caution!Before carrying out a density adjustment, please refer to BA 057D/06/en Promass 80 Operating Instructions where a detailed description of the density adjustment is given.

    ! Note!The density adjustment can be executed if: The sensor does not accurately measure the density which the operator

    expects based on laboratory trials. The characteristics of the fluid are outside the measuring points set at the

    factory or reference conditions under which the flowmeter has been calibrated.

    The plant is used solely for measuring a fluid whose density is to be determined very accurately under constant conditions.

    User input:CANCELDENSITY ADJUST

    Factory setting:CANCEL

    RESTORE ORIGINAL With this function the original density coefficient determined at the factory are restored.

    Options:NOYES

    Factory setting:NO

    PRESSURE MODE Use this function to configure an automatic pressure correction. In this way, the effect of a pressure deviation between the calibration and process pressures on the measured error for mass flow is compensated for, (see also Operating Instructions PROline promass 80, BA 057D/06/en, Accuracy Chapter).

    Options:OFFFIX (A fixed process pressure for pressure correction is specified)

    Factory setting:OFF

    PRESSURE! Note!This function is not available unless FIX was selected in the PRESSURE MODE function.

    Use this function to enter the value for the process pressure which should be used during pressure correction.

    User input:7-digit floating-point number

    Factory setting:0 bar g

    ! Note!The appropriate unit is taken from the function group SYSTEM UNITS (see Page 8).

    Function description PROCESS PARAMETER

  • Device Functions PROline Promass 80 15 Group SYSTEM PARAMETER

    Endress+Hauser 49

    15 Group S