Radiant Technologies, Inc.Pt/PNZT/Pt Data using Precision Premier
Piezoelectric Displacement Results of Measuring Pt/PNZT/Pt Capacitor on Seiko Instruments
SPA400 AFM. Radiant Technologies, Inc.
August 7, 2001
Radiant Technologies, Inc.Pt/PNZT/Pt Data using Precision Premier
Sample Description
• Capacitor Area: 12µ x 12µ
• Construction: Pt/PNZT/Pt
• Composition: 4/20/80 PNbZT
• PNZT Thickness: 1µm
Radiant Technologies, Inc.Pt/PNZT/Pt Data using Precision Premier
Displacement at 40V
40V Displacement Loop
-15
-10
-5
0
5
10
15
20
25
30
-60 -40 -20 0 20 40 60
Volts
An
gst
rom
s 40V
1u 4/20/80 PNZT w/ Pt electrodes
measured w/SII
Displacement measured at 40Vwith a measurement period of 1 second.
Radiant Technologies, Inc.Pt/PNZT/Pt Data using Precision Premier
Displacement at 40V
Displacement in Angstroms Corrected for Chuck Drift
-15
-10
-5
0
5
10
15
20
25
30
-60 -40 -20 0 20 40 60
Vdrive
Ang
stro
ms
Displacement
1u 4/20/80 PNZT w/ Pt electrodes measured w/SII SPA400 AFM.
Average of 10 loops.
Displacement loop averaged 10 times.
Radiant Technologies, Inc.Pt/PNZT/Pt Data using Precision Premier
Nested Displacement Loops
Nested Displacement Loops
-15
-10
-5
0
5
10
15
20
25
30
-60 -40 -20 0 20 40 60
Volts
An
gs
tro
ms
40V
30V
20V
10V
1u 4/20/80 PNZT w/ Pt electrodes
measured w/SII
Displacement measured at 40V, 30V, 20V, and 10V. Measurement period is 1 second.
Radiant Technologies, Inc.Pt/PNZT/Pt Data using Precision Premier
Conclusion
• Precision Premier integrates with the Seiko Instruments AFM family to collect displacement measurements of thin ferroelectric films.
• The 1µm thick Radiant 4/20/80 PNbZT film displaced 23Å at 40V giving the film a d33 value of 57.5.
• The measured value is the not the true d33 value of the film due to strong clamping by the substrate.
• Piezoelectric fatigue, frequency response, and imprint can also be measured using the Vision software.