Klinik und Poliklinik für Augenheilkunde Pupil centre shift between bright light and dim light conditions C. Winkler von Mohrenfels 1 , J Salgado 1, R. Khoramnia 1 , Christian Wüllner 2 , Peter Riedel 2 , C Donitzky 2 1Augenklinik, Klinikum rechts der Isar, Technische Universität München, Munich, Germany, Financial interest: none 2 WaveLight AG Erlangen, financial interest: employee

Pupil centre shift between bright light and dim light conditions

  • Upload
    hu-ball

  • View
    25

  • Download
    0

Embed Size (px)

DESCRIPTION

Pupil centre shift between bright light and dim light conditions. C. Winkler von Mohrenfels 1 , J Salgado 1, R. Khoramnia 1 , Christian Wüllner 2 , Peter Riedel 2 , C Donitzky 2 1Augenklinik, Klinikum rechts der Isar, Technische Universität München, Munich, Germany, Financial interest: none - PowerPoint PPT Presentation

Citation preview

Page 1: Pupil centre shift  between bright light and dim light conditions

Klinik und Poliklinik für Augenheilkunde

Pupil centre shift between bright

light and dim light conditionsC. Winkler von Mohrenfels1, J Salgado1, R. Khoramnia1, Christian

Wüllner2, Peter Riedel2, C Donitzky2

1Augenklinik, Klinikum rechts der Isar, Technische Universität München, Munich, Germany, Financial interest: none

2 WaveLight AG Erlangen, financial interest: employee

Page 2: Pupil centre shift  between bright light and dim light conditions

Klinik und Poliklinik für Augenheilkunde

ASCRS 2010 2

Problem

- different pupil size between dim and bright light condition

- centration of laser ablation: pupil center, apex, line of sight

- eye tracking: pupil, iris

- pupil shift between the different light conditions could lead to

miss alignment of the laser ablation.

Method

- 47 healthy subjects both eyes (12♂, 35 ♀),

- mean age 36 years (range 23 to 60)

- WaveLight Allegro Topolyzer: Pupil software (dim [20 lux] and bright

light [455 lux] conditions)

Page 3: Pupil centre shift  between bright light and dim light conditions

Klinik und Poliklinik für Augenheilkunde

ASCRS 2010 3

Pupil difference and age

y = -0.0252x + 3.8457R2 = 0.2163

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

10 20 30 40 50 60 70

age in [years]

Pu

pil

dev

itio

n b

etw

een

dim

an

d b

ritg

h l

igh

t co

nd

itio

ns

in [

mm

]

Page 4: Pupil centre shift  between bright light and dim light conditions

Klinik und Poliklinik für Augenheilkunde

ASCRS 2010 4

Pupil center shift right eye

MW 105µm (SD 65; min 18µm; max 330µm)

-0.5

-0.4

-0.3

-0.2

-0.1

0.0

0.1

0.2

0.3

0.4

0.5

-0.5 -0.4 -0.3 -0.2 -0.1 0.0 0.1 0.2 0.3 0.4 0.5

x axis

• y-a

xis

nasaltemporal

Page 5: Pupil centre shift  between bright light and dim light conditions

Klinik und Poliklinik für Augenheilkunde

ASCRS 2010 5

Pupil center shift left eye

MW 107µm (SD 64; min 11µm; max 348µm) -0.5

-0.4

-0.3

-0.2

-0.1

0.0

0.1

0.2

0.3

0.4

0.5

-0.5 -0.4 -0.3 -0.2 -0.1 0.0 0.1 0.2 0.3 0.4 0.5

x-axis

y- a

xis

temporalnasal

Page 6: Pupil centre shift  between bright light and dim light conditions

Klinik und Poliklinik für Augenheilkunde

ASCRS 2010 6

Conclusions

- the average pupil shift is 0.11mm

- most pupils shift to temporal side

- huge variation 0.01mm to 0.348mm

- the larger pupils shifts could lead to a miss alignment

between tracking and laser ablation.

→ risk of decentration and poor visual outcome