Quality and dose control for CR/DR

John GibbsGlobal Product Manager Modality Workstations

+32-3-444-7560

Page 2: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

2Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

From analogue to digital

What is the impact of going digital on my daily work?

• Radiographers’ questions• How do I know my work quality is OK?

• Radiology Supervisor’s questions• How can I monitor team quality?

• Physicist’s questions• How can I monitor the system quality?

Page 3: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

3Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

NX Tools

• Radiographer• E.I - Exposure Index• Dose Bar-graph

• Radiology supervisor• Repeat Reject Report• Extended Dose Monitoring New• IHE REM Profile New

• Physicist• Auto AC2• Auto QC Mammo

Page 4: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

4Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

NX Tools for everyone

• Radiographer• E.I - Exposure Index• Dose Bar-graph

• Radiology supervisor• Repeat Reject Report• Extended Dose Monitoring New• IHE REM Profile New

• Physicist• Auto AC2• Auto QC Mammo

Page 5: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

5Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

* For Brightness -16 Contrast - 10

Why Exposure Index ?

200023532.862408920

100020522.5612017710

50017512.26603555

25014501.96307102.5

IEC Exposure Index

Carestream

EI***

Agfa LGM**

Canon Rex*

Fuji S number

Exposure in µGy

**At 400 Speed Class

***Kodak specific EI

Four Systems four ways of measuring exposure at the pla te

Page 6: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

6Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

• The Exposure Index (EI) is related to receptor exposure• The Exposure Index provides a relative exposure measurement

within an exam type

• It is not (and should not be used as) a calibrated patient dose meter!!!

• The dose in µGy to the plate can only be determined when a a flat field exposure is made under calibrated (RQA-5) X-Ray beam conditions

• Three Parameters• EI Exposure Index• TEI Target Exposure Index• DI Deviation Index

What is Exposure Index ? ( IEC 62494-1)

Page 7: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

7Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Target Exposure Index and Deviation Index

• Target Exposure Index: TEI

• The reference exposure index for a particular exposure* • Extremities - 1000*, Chests – 500*, Abdomens – 250*

• Can be determined by statistical averaging (50 exposures)→ preferred scenario

• Can be preset (fixed) by the user

• What is it used for • To calculate the Deviation Index

* Example values - actual values should be based on customer statistics

Page 8: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

8Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Target Exposure Index and Deviation Index

• Deviation Index :

• Expresses how far the exposure is away from a reference value

• Provides a relative indication for under/over exposure

• In a perfect situation the DI would be zero (rarely the case)

• 1 deviation unit equals ~ 25% (+1 or -1) on a AEC

• 3 deviation units equals 2x exposure or ½ exposure (+3 or -3)

DI =10 * LogTEI

[ EI ]

Page 9: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

9Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Dose bar-graph

• Has been introduced to improve interpretation of Lgm or E.I./D.I (Exposure Index/Deviation Index) by the radiographer.

• This feature allows radiographers to personally monitor their own exposure technique quality.

=> Are they working inline with the agreed average dose for each exposure type?

Page 10: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

10Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Dose bar-graph

• Colour improves interpretation of Lgm / D.I.• -0.3 < Lgm < +0.3 Green bar • -0.6 < Lgm < -0.3 or +0.3 < Lgm < +0.6 Yellow bar• Lgm < -0.6 or +0.6 < Lgm < Red bar• -3 < D.I. < +3 Green bar • -6 < D.I. < -3 or + 3 < D.I. < +6 Yellow bar• D.I. < -6 or +6 < D.I < Red bar

• D.I.= Deviation Index is displayed on an Exposure Index acquisition system (= all new CR30, DX-G and DX-D systems)

Page 11: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

11Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Dose between average and double the average dose.

Dose between average and double the average dose.

Page 12: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

12Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Dose higher than double, but lower than 4x the average.

Dose higher than double, but lower than 4x the average.

Page 13: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

13Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Dose higher than 4x average dose

Dose higher than 4x average dose

Page 14: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

14Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Low Dose Extremities

Historic Film Speed 600 400 200 100

Exposure Index ~200 ~300 ~600 ~1000

Deviation Index "+ or - 3 units maximum"

General Work

*These are guidelines only!!!! The customer has to decide on exposures settings using the ALARAprinciple. The dose level used corresponds with the noise level the customer accepts.

Exposure Index Guidelines*

Page 15: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

15Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Who?

• Radiographer• E.I - Exposure Index• Dose Bar-graph

• Radiology supervisor• Repeat Reject Report• Extended Dose Monitoring New• IHE REM Profile New

• Physicist• Auto AC2• Auto QC Mammo

Page 16: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

16Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Repeat Reject

• Reject workflow: rejecting an image automatically generates a new exposure for the repeat, user can assign a reject reason.

Page 17: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

17Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Repeat Reject report

• Formatted report available for Repeat/Reject

• When exporting statistics now two file are created, the xml file(as before) and a html file that contains a readable/printable summary of the exported data.

HTML Document

Page 18: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

18Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Extended Dose Reporting New NX8600

• Allows searching the dose database of 60.000 exposures to create reports

• Search for a certain time period, digitizer, x-ray room, exam type or operator

• Create report for statistics of all exposures or only look at the extremes (DI outliers)

Page 19: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

19Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Extended Dose Reporting

Page 20: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

20Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Extended Dose Reporting

Page 21: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

21Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Extended Dose Reporting

Page 22: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

22Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Select a single line to see graphical representation of the results

For each Protocol code, statistics are shown for the received DAP values

A Diagnostic Reference Level (DRL) can be entered for each code

Page 23: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

23Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Page 24: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

24Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Page 25: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

25Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Select a single Protocol code to see statistics of the exposures that are part of the protocol code.

Page 26: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

26Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Limit that determines outliers can now be configured.

Page 27: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

27Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Data for individual exposures can be exported for further analysis

Page 28: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

28Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Forced entry of operator ID per exam

Operator selection box will pop up at first ID of each open exam.

Select ‘force operator IDCreate operator list in configtool

Select or type in name

Page 29: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

29Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Extended Dose Reporting

• Limitations:

• CR DR only• Agfa modalities only

• Hospital wide dose reporting � IHE REM

Page 30: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

30Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

IHE REM Profile

• Radiation Exposure Monitoring

• We will send DICOM Structured Reports of radiation events (Dose objects) to Radiation Reporting Systems.• Some of the key details include:

• All modalities – kVP, mA, collimation, filters, etc. – Patient/Order/Study details

• Projection X-Ray – DAP, Dose@RP, Imaging geometry, Fluoro Dose, Fluoro Time

• Mammography – AGD, Entrance Exposure@RP, Compression, Half Value Layer

• What is in report depends on what is provided by the modality!• OR manually entered.

Page 31: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

31Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

IHE REM Profile New

Page 32: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

32Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

IHE REM Profile New

NX

Page 33: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

33Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Agfa Dose Info Reporter

• Under development

• CT + Agfa CR• US + Canada• Vendor Neutral• Data accessible by AGFA PACS /RIS

Page 34: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

34Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Who?

• Radiographer• E.I - Exposure Index• Dose Bar-graph

• Radiology supervisor• Repeat Reject Report• Extended Dose Monitoring New• IHE REM Profile New

• Physicist• Auto AC2• Auto QC Mammo

Page 35: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

35Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Auto QC 2

• Hardware • One test phantom• Filters, filter module,

positioning template

• Delivered in one suitcase

Page 36: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

36Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Auto QC 2 - Product information

• Software• Automatic calculation on raw images• Artifact inspection (visual check)• Automated report• Search & history functions• Based on NX GUI, with 4 main panes • Follows the NX licensing (protected by dongle & ELMS)

Page 37: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

37Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Auto QC 2 - Overview of tests

• Periodic quality control & Acceptance test procedure

• Plate inventory • Monitor check

•Scan Line Drop Outs•Overshoots•Signal-to-Noise Ratio (SNR)•Sensitivity •Dynamic Range •Signal Uniformity (banding) •Spatial Resolution [MTF]•Laser Jitter•System Linearity•Distance Accuracy

•Sweep Error•Pixel Aspect Ratio (V/H)•Pixel Size Horizontal & vertical•Skew•Image Width •Image Length•Radiation Quality (KVP, tube-wear)•Dark Noise•Erasure Efficiency

Page 38: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

38Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

Auto QC 2 software UI

Page 39: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

39Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

AutoQC Mammo - Phantom + PMMA

Page 40: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

40Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

AutoQC Mammo - Software

• Software package installable on NX or on a separate (physicists) PC

Page 41: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

41Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

QC for DR - Overview of tests

• Periodic quality control & Acceptance test procedure•Overshoots/Ringing •Signal to Noise Ratio (SNR)•System Sensitivity •System Linearity (Dose response) •Dynamic Range determination •Slow Scan Banding•Scan Line Uniformity•Spatial Resolution (MTF) •Laser Beam Jitter •Laser beam positioning error (Sweep Error)•Pixel Aspect Ratio

•Pixel Size / Resolution•Image Skew •Image Width and Height•(Radiation Quality) Only relevant for automated Auto QC calculations

•Artifact Analysis

•AEC test•Bad pixels•Low contrast

•Dark Noise(ATP)•Erasure Efficiency(ATP)

Ghosting is not relevant due to Dark Images correction

Page 42: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

42Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012

QC for DR - Overview of tests

• DR Periodic quality control & Acceptance test procedure

• Signal to Noise Ratio (SNR) (covered by low contrast test)

• System Sensitivity **• System Linearity (Dose response) *• Dynamic Range determination *• Spatial Resolution (MTF) *• Artifact Analysis **• AEC test• Bad pixels ***• Low contrast *

• Current Procedure* Tested with Normi 13 phantom** Tested with Flatfield exposure*** Panel type dependant

Page 43: Quality and Dose Control for CR / DR - Agfa HealthCare and Dose... · Quality and dose control for CR/DR ... Quality and Dose Control CR DR, update March 2012 * For Brightness -16

43Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012Quality and Dose Control CR DR, update March 2012