System Installation

Note:

does not have a steel (iron) top

Retain all packing materials.

1.

a. provides a draft free, dust free, dry environment. The temperature of the room must be between 50 °F and 85 °F(10 °C - 29 °C). Temperature should not vary more than 10 °F (5 °C) per day. Relative humidity range should be0-85%.

b. is free from vibration of large equipment and/or excessive walk-through traffic.

c. is capable of supporting the combined weight of the SMART Turbo and the SMART Trac II magnet {165 lbs. (75kg)} and, if applicable,

d. . Materials such as wood, plastic, aluminum, or non-magnetic stainless steelare suitable surfaces. Ensure that no steel bar, supporting the work surface, runs directly beneath the magnet.

e. provides adequate bench space — 44 in. (112 cm) in length and 30 in. (76 cm) in depth minimum - SMARTTurbo and SMART Trac II Magnet58 in. (148 cm) in length and 30 in. (76 cm) in depth minimum - SMARTTurbo, SMART Trac II Magnet and SMART Trac II Processor

f. allows the instrument to be connected to a dedicated, grounded outlet as specified in the instrumentspecifications and capable of accepting the plug of the power conditioner illustrated in this manual.

2.

3. Cut the bands securing the cardboard on the magnet.

Following the instructions provided in the SMART Turbo shipping carton, carefully remove the instrument and placeit on the vibration-free workbench or laboratory table in a location that

the SMART Trac II processor module {combined weight of 202 lbs. (95 kg)}.

Carefully remove the SMART Trac II processor module from its shipping carton and, based on available space, placeit on the workbench, on a shelf above the workbench or on the floor under the workbench. The processormodule must be within 9 ft. (3 m) of the magnet.

4. Lift the cardboard from the magnet.

Note:

2

Page 3: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

5. Remove the foam from the top of the magnet.

6. With two people positioned on the sides of the magnet, one on each side, use the straps and carefully lift the magnetfrom the skid and place it on the workbench beside the SMART Turbo. The magnet weighs approximately 110pounds (50 kg).

7. Inspect each component of the system for visual damage.

WARNING

If damage is noted, do not attempt instrument operation.

AVERTISSEMENT

Ne pas mettre en marche si l’instrument est endommagé.

8. If any instrument component has been damaged in shipping, contact the freight carrier to report the damage and tofile a damage report. Contact CEM Service or the nearest subsidiary or distributor to request service information.

9. Verify that all accessories illustrated and listed on page 6 are included.

3

Page 4: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Heater & Heater Block

Cable Connections

4

Power Conditioner

Magnet

SMART Turbo

SMART Trac II Processor Module

Page 5: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

TracTubes

TracFilm

HeaterBlock

ExternalUSBDrive

NylonSplitLoom

NMR/SMART TracCable

Accessory Kit containing:• Trac Tube (160505) 5• Trac Film (159875)• Standards Kit (568450)

• Heater Block (159885)

• Frequency Oil Standard (160065)• Normalization Standard (160570)

• External USB Drive (277200)• NMR/SMART Trac Cable (160270)• 3/8” Nylon Split Loom (BR198859)• SMART Trac Operation Manual (600xxx) (Not illustrated)• Maintenance Guide (SM0044)

Cable Kit (160660) containing:

BNC Cable

Ground Cable

Shim Cable

Short SMB – SMB Cable

Long SMB – SMB Cable

Cross-over Ethernet Cable

BNC CableTx out to Tx in)

Ground Cable

SMB – SMB

Short Cable

(Tx in to Tx out)

37-way

Shim Cable

Cross-over

Ethernet Cable

SMB – SMB

Long Cable

(Rx out to Rx in)

USB toSerial Cablefor CEMService Use(Do Not Discard)

NormalizationStandard

FrequencyOil Standard

5

Page 6: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

10. Unpack the power conditioner and place it on the work bench or the floor as far away from the SMART Trac IIMagnet as possible.

Ensure that the power conditioner is not placed in close proximity with the SMART Trac II magnet to preventdemagnitization of the magnet.

11. Carefully connect each cable provided in the SMART Trac II cable kit and the cable in the accesssory kit. Ensure thateach cable is connected properly. (Refer to the illustration on pages 4 and 5. Use caution to prevent bending ordamaging the pins of the connectors.

12. Unpack the heater and heater block. Plug the heater power cord into one of the receptacles of the power conditioneras illustrated on page 4. Place the heater block inside the heater. Press the “Power” button first, then press the“On/Off” button as illustrated below. Follow the manufacturer’s instructions included with the heater block to set thetemperature of the heater to 40 °C. Place the Trac Standards (p/n 160065 and 160570) in the heater block andpermit the standards to heat for at least four (4) hours. Note: The label on the bottom of the heater indicates theactual voltage

CAUTION

13. Plug the power cords from the SMART Turbo and the processor module into the power conditioner.

14. Plug the power cord of the power conditioner into the appropriate electrical outlet. Turn the power conditioner on.

Note: The above power cord is applicable to the U.S. only. Power cords for instruments installed in other countries will besupplied based on electrical specifications for the applicable country.

15. Position the power switch of the SMART Turbo and the SMART Trac II processor module in the “on” position.

Note: CEM Corporation recommends that the SMART Trac II System power cords be connected to electrical outlets andthe power switches remain in the “on” position at all times. After 15 minutes of idle time, the SMART Turbo assumes a“sleep” mode.

ON

OFF

6

Page 7: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

P.O. Box 200, Matthews • 3100 Smith Farm Road • Matthews, NC 28104 • (704) 821-7015 • Fax: (704) 821-7894E-mail: [email protected] Web: http:\\www.cem.com

U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail: [email protected]. S. Service Support (800) 726-5551 Fax: (704) 821-4368 E-mail: [email protected]

16. Permit the system to warm up.

CAUTION

The SMART Trac II System requires a warm up period of at least 4 - 8 hours prior to operation. If theinstrument is switched off for an extended period of time (weekend, holidays, etc.), ensure that warmuptime is permitted prior to use.

17. When the warm-up time is complete, the blue indicator light on the front of the magnet will remain constant.indicating that the heater is functioning properly.

18. Perform the QuickCheck MONITOR tests as outlined in the instructions supplied in the MONITOR package.

20. The SMART Trac II is ready for operation. Contact the local sales representative to arrange initial training or refer tothe SMART Trac II Operation Manual to begin analysis.

19. Perform the magnet temperature and homogeneity tests.

a. Temperature - From the SMART Turbo Main Menu, select “7 - SMART Trac.” Select “Next Page,” “1 - Diag-nostics,” and “1-System Check.” Variable number 17 is the magnet temperature. The temperature should bebetween 39 °C and 41 °C.

b. Homogeneity - Press any key on the keyboard. Select “4 - Magnet Homogeneity.” Follow the on-screeninstructions. The displayed homogeneity should be <20.0.

7

Page 8: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

 

System Setup

Page 9: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

SMART Trac™ II

Setup Guide

Page 10: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

1. Insert the Oil Standard and Normalization Standard into the heater block. Turn the SMART Trac II instrument on and perform steps 2 through 6 immediately.

Note: Standards must be thermally conditioned at 40C for at least 4 hours.

Oil Standard

Normalization Standard

Page 11: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

2. The SMARTNMR and CEM Communicator activate automatically upon instrument startup

SMARTNMR System Tray Icon

CEM Communicator

Page 12: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

3. Right-click SMARTNMR and select Tools System Setup

Page 13: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

4. Select on System Setup.

Page 14: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

5. Once the System Setup dialog window is open, select Auto Run.

“Auto Run” will automatically run the following procedures, stopping between each procedure to prompt for user input:

• Magnet Temperature Offset• SHIM Setup• Cal-lock• Frequency Optimization• Pulse Width Optimization

Correct standard to use for each procedure:

SHIM Setup Oil Standard

Cal-lock Normalization Standard

Frequency Optimization Oil Standard

Pulse Width Optimization Oil Standard

Page 15: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

6. When prompted for the Magnet Temp. Offset, lift the blue cover on the magnet, and enter the temperature offset number shown on the label. Select OK.

Page 16: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

SHIM Setup prompt for Oil Standard

Cal-lock prompt for Normalization Standard

Page 17: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Frequency Optimization Prompt for Oil Standard

Pulse Width Optimization Prompt for Oil Standard

Page 18: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

7. When the Status window indicates Complete, select OK and proceed with Unit Normalization.

Record the data shown in Results:Auto SHIM CompleteOffsetFrequencyP90P180

Frequency Range: 23.3800 – 23.4200P90 Range: 3.0 – 6.5P180 Range: 6.0 – 15.0

Page 19: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Unit Normalization

1. Main Menu

2. Press 7 – SMART Trac

3. Press 7 – Unit Normalization

Note: Methods can not be created or imported until unit

normalization is performed.

4. Select SMART Trac II. Press 1 to toggle between SMART Trac I & II.

5. Press Ready

6. Remove sample or oil standard if present on the magnet.

7. Press Ready.Note: Process will take about 5mins.

8. Collecting data at gain 10, 25 and 50 will be shown on the screen.

Page 20: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Unit Normalization

9. Insert Normalization Standard into SMART Trac II

10.Press ReadyNote: Process will take about 1 hour.

11.SUS Temperature Stabilization will be shown on the screen.

12.Collecting data at gain 10, 25 and 50 will be shown on the screen.12.Collecting data at gain 10, 25 and 50 will be shown on the screen.

13.When done, screen will go back to the SMART Trac menu.

After unit normalization, you can now begin developing/importing

method/s on the SMART Trac II.

Page 21: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Backup SMART Trac™ II Methods(for customer)

• Main Menu

• Press 7 – SMART Trac

• Press 5 – Export Method

• Select Method• Select Method

• Insert USB Stick

• Press Ready

• Leave USB stick with customer

Page 22: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Complete Backup SMART Trac™ II(for installer)

• Insert USB stick

• Right click the SMART NMR icon

• Highlight “Tools → Export Diagnostic Data”

• Select “Export Diagnostic Data”• Select “Export Diagnostic Data”

• Select “Yes”

Page 23: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

 

Method Importation

Page 24: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Importing Methods for SMART Trac II

1. From the system Main Menu, press “7” on the instrument keypad for “7.SMART Trac”

2. From the SMART Trac menu, press “4” to select “4.IMPORT METHOD.”

3. Ensure that a USB stick containing the method(s) to import is installed in the SMART Trac II USB port.

4. Press “Ready.”

5. Select the method(s” to import from the USB stick to the SMART Trac II system.

Page 25: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

 

Method Creation

Page 26: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Create a Fat NMR Method on the SMART Trac II

From the Main Menu Select #2. Edit/Create Method Select #1 New Method Name Method: (spell out name of method); Select (“Ready”) Edit Method

#1. Moisture/Fat NMR (“toggle”) Caution: Do Not Choose Moisture/Fat this selection is for the FAS-9001

#2. Constant Weight (Default) #3. Turbo ON (Default)

Moisture/Fat NMR Type #1. Standard

Enter Parameters for Moisture: Power: 100% Delta Weight: 0.2mg Delta Time: 10sec Max Time: 10min (Default) Bias: +0 (Default) Min Result: 0% (Default) Max Result: 100% (Default) (“Next Page”)

Max Temp: 100oC Based on AOAC oven method Min Wt Range: 2.0g (2.0 is Default) Max Wt Range: 2.5g Wt Compensation: ON (“Next Page”)

Fat Range <80%, >80% or <1% (<80 Default, <1% Condensed Skim) Fat Bias: 0% 80% Butter, Mayonnaise Warm up delay: 0 Used with Fat Only Option, powders or butter. ● Run Time 8 sec. (Default) (“Next Page”)

Page 27: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Edit References: Need High, Middle, Low reference values to create Linear Fat Equation. Example: 12.20%, 10.56%, 5.30% Fat

Press “ADD” for each reference value to be entered: 1. 12.20 2. 10.56 3. 5.30

Collect Standards Data: Choose item number “1” to collect data for 12.20. EDIT Standard - 12.20 (Start with highest fat sample)

Press “ADD Standard” to collect standards data for 12.20 to generate signal/mass.

Using the12.20%F sample, repeat “Add Standard” step a minimum of 3 times (up to 8).

Edit Standard – 12.20% %Fat S/M Mass ------------ ----- ------ *1. ------- 1539.6340 2.2315 g *2. ------- 1534.0634 2.3152 g *3. ------- 1537.0634 2.5315 g

Note: Fat calculation will not be displayed until standards data has been collected for at least two reference points (12.20 and 5.30).

After enough data points have been collected for 12.20, Press “Prev Page” and choose next item number “3” to collect data for the lowest fat sample (5.30).

EDIT Standard – 5.30 (lowest fat sample) Press “ADD Standard” and repeat at least 3 times to collect signal/mass. Now that you have the second point, the fat values will be calculated and shown.

Select the points (press line number to toggle on/off) that are within specifications of reference to create your Linear Fat Equation.

Note: Each time you select a point it will recalculate the line and therefore recalculate the other standards data (even those under 12.20).

Edit Standard – 5.30% %Fat S/M Mass ------------ ----- ---------- *1. 5.26 1036.0111 2.1222 g *2. 5.35 1040.1713 2.2231 g *3. 5.30 1038.0225 2.1315 g

After enough data points have been collected, Press “Prev Page” and choose next item number “2” to collect data for the middle point (10.60).

Edit Standard – 10.56 (middle point to validate line) Press “ADD Standard” and repeat at least 3 times to collect signal/mass. Select points (press line number to toggle on/off) that are within specifications.

Edit Standard – 10.56% %Fat S/M Mass ------------ ----- ---------- *1. 10.56 1414.2478 2.4311 g *2. 10.48 1409.0634 2.3315 g *3. 10.61 1417.2236 2.5314 g

Page 28: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Data In Review

Edit References:

1. 12.20 2. 10.56 3. 5.30

Edit Standard – 12.20%

%Fat S/M Mass ------------ ----- ---------- *1. 12.24 1539.6340 2.2315 g *2. 12.15 1534.0634 2.3152 g *3. 12.19 1537.0634 2.5315 g

Edit Standard – 10.56%

%Fat S/M Mass ------------ ----- ---------- *1. 10.56 1414.2478 2.4311 g *2. 10.48 1409.0634 2.3315 g *3. 10.61 1417.2236 2.5314 g

Edit Standard – 5.30%

%Fat S/M Mass ------------ ----- ---------- *1. 5.26 1036.0111 2.1222 g *2. 5.35 1040.1713 2.2231 g *3. 5.30 1038.0225 2.1315 g

Once appropriate points are selected and there are at least 3 points for each reference

within specification, exit to Main Menu (saves method). Then, EXPORT METHODS to a USB stick (refer to REFERENCE PROCEDURES). Please, EMAIL METHODS to your CEM Representative for back up.

Page 29: Quick Start Guide - cem.comcem.com/media/contenttype/media/literature/515_Manual_TracII_Quick... · U. S. Sales and Application Support (800) 726-3331 Fax: (704) 821-5185 E-mail:

Reference Procedures EXPORT METHODS - To Export Methods to USB Stick:

Exit to the Main Menu Select #7. SMART Trac Select #5. Export Method

EMAIL METHODS - To email methods:

After exporting methods to USB stick, place stick in computer’s USB port. Create email for your CEM Representative Under the cem folder, attach the corresponding .cal, .mth, and .pse files

For example: If emailing cream method attach: cream.cal, cream.mth, cream.pse To Edit an Existing Method:

To Edit FAT References: Exit to the Main Menu Select #7. SMART Trac Select #3. Edit Method Standards

Note: EXPORT edited method to a new USB Stick

To Edit Moisture Parameters: Exit to the Main Menu Select #2. Edit/Create Method Select Method Navigate Through Screens Make Changes Exit to the Main Menu

*Note: Do NOT go past the “fat range” screen. This will delete your method.