Instructions

QuickStart8DPO Demonstration Boardfor TDS7104 and TDS7054Oscilloscopes

071-0715-00

www.tektronix.com

Page 2: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Copyright Tektronix, Inc. All rights reserved.

Tektronix products are covered by U.S. and foreign patents, issued andpending. Information in this publication supercedes that in all previouslypublished material. Specifications and price change privileges reserved.

Tektronix, Inc., 14200 SW Karl Braun Drive, Beaverton, OR 97077

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Page 3: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

WARRANTY

Tektronix warrants that this product will be free from defects in materials andworkmanship for a period of three (3) months from the date of shipment. If any suchproduct proves defective during the three-month period, Tektronix, at its option, either willrepair the defective product without charge for parts and labor, or will provide areplacement in exchange for the defective product.

In order to obtain service under this warranty, Customer must notify Tektronix of thedefect before the expiration of the respective warranty period and make suitablearrangements for the performance of service. Tektronix will provide such service atCustomer’s site without charge during the warranty period, if the service is performedwithin the normal on-site service area. Tektronix will provide on-site service outside thenormal on-site service area only upon prior agreement and subject to payment of all travelexpenses by Customer. When or where on-site service is not available, Customer shall beresponsible for packaging and shipping the defective product to the service centerdesignated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the returnof the product to Customer if the shipment is to a location within the country in which theTektronix service center is located. Customer shall be responsible for paying all shippingcharges, duties, taxes, and any other charges for products returned to any other locations.

This warranty shall not apply to any defect, failure or damage caused by improper use orimproper or inadequate maintenance and care. Tektronix shall not be obligated to furnishservice under this warranty a) to repair damage resulting from attempts by personnel otherthan Tektronix representatives to install, repair or service the product; b) to repair damageresulting from improper use or connection to incompatible equipment; c) to repair anydamage or malfunction caused by the use of non-Tektronix supplies; or d) to service aproduct that has been modified or integrated with other products when the effect of suchmodification or integration increases the time or difficulty of servicing the product.

THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THELISTED PRODUCTS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS ORIMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIEDWARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULARPURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACEDEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDYPROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANYINDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGESIRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HASADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Page 4: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board i

Table of Contents

General Safety Summary v. . . . . . . . . . . . . . . . . . . . . . . . . . . .

Preface vii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Contacting Tektronix viii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Getting StartedRequired Materials 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Prerequisites 1–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up Your QuickStart8 Board 1–2. . . . . . . . . . . . . . . . . . . . . Configuring Your QuickStart8 Board 1–3. . . . . . . . . . . . . . . . . . . . Applying and Removing Power 1–4. . . . . . . . . . . . . . . . . . . . . . . . . Overview of Signals 1–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Operating BasicsUsing the QuickStart8 board 2–1. . . . . . . . . . . . . . . . . . . . . . . . . Test Points 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Detailed Signal Descriptions 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . .

Switching Power Supply Signals: V SWITCH, I SWITCH 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Quadrature Amplitude Modulation Baseband Signals 2–6. . . . A 250 mV Fast Rise Signal (Tr < 200 ps) 2–8. . . . . . . . . . . . . . Dual Random Anomalies Signals 2–9. . . . . . . . . . . . . . . . . . . . RESET ANOM 2–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Phase Locked VCO Output Signals 2–11. . . . . . . . . . . . . . . . . . . RF Modulation 2–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pseudo Random Anomalies Signals 2–14. . . . . . . . . . . . . . . . . . Universal Serial Bus 2–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Simulated Ground Bounce Signal 2–16. . . . . . . . . . . . . . . . . . . .

Appendix A: DiagramsAppendix B: Instrument CareIndex

Page 5: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Table of Contents

ii QuickStart8 DPO Demonstration Board

FiguresFigure 1–1: Connecting the QuickStart8 board 1–5. . . . . . . . . .

Figure 2–1: Signal locations on the QuickStart8 board 2–4. . . .

Figure 2–2: V switch, I switch 2–5. . . . . . . . . . . . . . . . . . . . . . . . .

Figure 2–3: QAM I, QAM Q, and QAM CLK signals 2–6. . . . .

Figure 2–4: XY display 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Figure 2–5: Y-T FastAcq display 2–7. . . . . . . . . . . . . . . . . . . . . .

Figure 2–6: XYZ FastAcq display 2–7. . . . . . . . . . . . . . . . . . . . .

Figure 2–7: Fast Rise Signal 2–8. . . . . . . . . . . . . . . . . . . . . . . . . .

Figure 2–8: Random Anomaly with cheat signal 2–9. . . . . . . . .

Figure 2–9: Dual Channel Random Anomalies 2–9. . . . . . . . . .

Figure 2–10: Low resolution display 2–11. . . . . . . . . . . . . . . . . . .

Figure 2–11: High resolution display 2–11. . . . . . . . . . . . . . . . . . .

Figure 2–12: Modulated RF display 2–13. . . . . . . . . . . . . . . . . . .

Figure 2–13: Pseudo-Random digital signals 2–14. . . . . . . . . . . .

Figure 2–14: USB signals 2–15. . . . . . . . . . . . . . . . . . . . . . . . . . . .

Figure 2–15: Simulated Ground Bounce signal 2–16. . . . . . . . . .

Page 6: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Table of Contents

QuickStart8 DPO Demonstration Board iii

TablesTable 1–1: VCO MODE SELECT (S2810) 1–3. . . . . . . . . . . . . .

Table 2–1: Oscilloscope signals 2–1. . . . . . . . . . . . . . . . . . . . . . .

Table 2–2: USB Data and switch functions 2–3. . . . . . . . . . . . .

Page 7: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Table of Contents

iv QuickStart8 DPO Demonstration Board

Page 8: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board v

General Safety Summary

Review the following safety precautions to avoid injury and preventdamage to this product or any products connected to it. To avoidpotential hazards, use this product only as specified.

Only qualified personnel should perform service procedures.

While using this product, you may need to access other parts of thesystem. Read the General Safety Summary in other system manualsfor warnings and cautions related to operating the system.

To Avoid Fire or Personal InjuryGround the Product. This product is indirectly grounded through thegrounding conductor of the mainframe power cord. To avoid electricshock, the grounding conductor must be connected to earth ground.Before making connections to the input or output terminals of theproduct, ensure that the product is properly grounded.

Observe All Terminal Ratings. To avoid fire or shock hazard, observe allratings and markings on the product. Consult the product manual forfurther ratings information before making connections to the product.

Avoid Exposed Circuitry. Do not touch exposed connections andcomponents when power is present.

Do Not Operate With Suspected Failures. If you suspect there is damageto this product, have it inspected by qualified service personnel.

Do Not Operate in an Explosive Atmosphere.

Do Not Operate in Wet/Damp Conditions.

Keep Product Surfaces Clean and Dry.

Page 9: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

General Safety Summary

vi QuickStart8 DPO Demonstration Board

Safety Terms and SymbolsTerms in This Manual. These terms may appear in this manual:

CAUTION. Caution statements identify conditions or practices thatcould result in damage to this product or other property.

Terms on the Product. These terms may appear on the product:

DANGER indicates an injury hazard immediately accessible as youread the marking.

WARNING indicates an injury hazard not immediately accessible asyou read the marking.

CAUTION indicates a hazard to property including the product.

Symbols on the Product. These symbols may appear on the product:

CAUTIONRefer to Manual

Protective Ground(Earth) Terminal

Page 10: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board vii

Preface

This instruction manual contains specific information about theQuickStart8 board for use with the TDS7104 and TDS7054Oscilloscopes.

Getting Started covers how to set up and configure the Quick-Start8 board and an overview of the output signals; plus anyrequired materials you may need to use the board.

Basic Operations provides a quick reference testpoint table and adetailed description of the output signals.

The Appendix includes circuit diagrams for reference only, andinstructions on how to maintain and ship the QuickStart8 board.

Page 11: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Preface

viii QuickStart8 DPO Demonstration Board

Contacting Tektronix

Phone 1-800-833-9200*

Address Tektronix, Inc.Department or name (if known)14200 SW Karl Braun DriveP.O. Box 500Beaverton, OR 97077USA

Web site www.tektronix.com

Sales support

1-800-833-9200, select option 1*

Service support

1-800-833-9200, select option 2*

Technicalsupport

Email: [email protected]

1-800-833-9200, select option 3*1-503-627-2400

6:00 a.m. – 5:00 p.m. Pacific time

* This phone number is toll free in North America. After officehours, please leave a voice mail message.Outside North America, contact a Tektronix sales office ordistributor; see the Tektronix web site for a list of offices.

Page 12: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Getting Started

Page 13: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board 1–1

Getting Started

The QuickStart8 board produces signals specifically designed todemonstrate the timing and voltage measurement capabilities ofTektronix test and measurement oscilloscopes.

The board is especially useful for demonstration and trainingpurposes, since its uses output signals that incorporate imperfectionsfound in a testing environment.

This section includes the following information:

Required materials

Instructions for setting up and configuring the board

Overview of the output signals

Required MaterialsThe QuickStart8 board comes with the following standard accessory:

Instruction manual

You will need to supply the following items:

A Tektronix general-purpose oscilloscope

Choose one: P6139A 10X Passive probe, TCP 202 Current probe,or Differential probe (P6246 or P6247)

50 Ω BNC cable with a BNC/SMA adapter

Page 14: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Getting Started

1–2 QuickStart8 DPO Demonstration Board

PrerequisitesThe examples in this manual are based on the following prerequisitesand assumptions:

You are familiar with the basics of taking timing and voltagemeasurements

That the oscilloscope and any accessories are properly set up (forexample, it is assumed that any oscilloscope probe that youintend to use is properly calibrated)

Setting Up Your QuickStart8 Board

CAUTION. Static discharge can damage the board. To prevent staticdamage, handle components only in a static-free environment.

Always wear a grounding wrist strap, heel strap, or similar devicewhile handling the board.

1. To discharge any static electricity, touch the ground connectorlocated on the board. Then, before you remove the board fromthe protective bag, touch the bag to discharge stored staticelectricity.

2. Remove the QuickStart8 board from the bag and place the boardparts-side up on any flat surface.

3. Turn the board so that the rubber feet are placed on the worksurface.

4. Set the power switch, S2780, to off (see Figure 2–1 on page 2–4).

Page 15: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Getting Started

QuickStart8 DPO Demonstration Board 1–3

Configuring Your QuickStart8 Board Configure the QuickStart8 board using the VCO Mode Select switch,S2810. See Figure 2–1 on page 2–4 for the switch location. Theconfiguration you choose will determine the appearance of theoutput signals at test pins J2410, J2520 and J1.

VCO MODE SELECT. Switch S2810 consists of four DIP switches,labeled 1 through 4. DIP switches 1 and 2 are not used. DIP switches3 and 4 of S2810 allow selection of several modes of operation.Table 1–1 lists the VCO switch configurations and outputs.

Table 1–1: VCO MODE SELECT (S2810)

VCO Mode Switch

1 2 3 4PLL VCO OUTPUT

X X C C Output varies between FL and FH every 800 ms

X X C O Output varies between FL and FH every 3 seconds

X X O C FL*

X X O O FH *

* FL is approximately 68 MHz, and FH is approximately 85 MHz.

Page 16: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Getting Started

1–4 QuickStart8 DPO Demonstration Board

Applying and Removing Power

Power Source. The QuickStart8 board can be powered directly fromthe TDS 7000 Series oscilloscope using the USB (Universal SerialBus) port or from an optional transformer.

CAUTION. Connecting the QuickStart8 board incorrectly can damageyour keyboard (see Figure 1–1)

To power on the board using the USB port:

1. Connect the USB cable into the USB IN port on the board (seeFigure 1–1 and 2–1).

2. Plug the USB cable into the USB connector on back of theTDS 7000 Series instrument.

3. Plug the keyboard cable into the USB OUT port on the board (seeFigure 1–1 and 2–1).

To power on the board using an optional wall transformer:

CAUTION. To prevent damage to the oscilloscope when using a walltransformer, do not use the USB connector to attach between theQuickStart8 board and oscilloscope.

1. Connect the power cord to the power jack on the board (seeFigure 2–1).

2. Plug the wall transformer into a wall socket.

NOTE. Use a wall transformer that is 9 volts or between 6 – 12 voltsat 5 watts. The outside diameter of the power-jack connector is5.5 mm.

To remove power follow these steps:

1. Set the power switch (see Figure 2–1) to off.

2. Disconnect the power cord or the USB IN cable.

Page 17: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Getting Started

QuickStart8 DPO Demonstration Board 1–5

YES

NO

TDS7000

QuickStart8

QuickStart8

MouseKeyboard

MouseKeyboard

TDS7000

USB OUTport

USB INport

Figure 1–1: Connecting the QuickStart8 board

Page 18: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Getting Started

1–6 QuickStart8 DPO Demonstration Board

Overview of Signals The following is an overview of the signals you can use on theQuickStart8 board to demonstrate a variety of timing and voltagemeasurements.

Signals

The following signals are available on the QuickStart8 board:

Switching Power Supply Signals (J2880); switched voltage andcurrent (see page 2–5)

Quadrature amplitude modulated baseband signals (see page 2–6)

Fast rise voltage step, J600 (see page 2–8)

Dual random anomalies signals (see page 2–9)

Voltage controlled oscillator outputs (see page 2–11)

Phase locked loop control signal, J2410 (see page 2–11)

QAM modulated RF outputs, J1 (see page 2–13)

Pseudo-random digital signals with clock, J13 (see page 2–14)

Universal serial bus (USB) signal outputs, J2 (see page 2–15)

Simulated ground bounce signal, J14 (see page 2–16)

Page 19: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

Page 20: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board 2–1

Using the QuickStart8 board

This section includes a quick reference for testpoints (see Table 2–1)and a detailed description of oscilloscope signals beginning onpage 2–5.

Test PointsMost of the testpoints are arranged along the edge of the board.Table 2–1 lists the oscilloscope signals and a numerical reference inthe left-hand column for testpoint locations on Figure 2–1.

Table 2–1: Oscilloscope signals

See Figure2–1

Testpoint Description Function

1 J1 MODULATED RF OUT(see page )

Demonstrates QAM modu-lated RF

2 J13 Pseudo-random DigitalSignals (see page 2–14)

Demonstrates general digitaltriggering and display

3 J14 Ground BounceSignal (see page 2–8)

Demonstrates Intermittentbehavior capture

4 J600 FAST RISE STEPVOLTAGE (see page2–8)SMA connection

Demonstrates step responseand time delay measurements

5 J2410 PLL VCO OUTPUT(see page 2–11)

Selects VCO mode

6 J2520 VCO TUNINGVOLTAGE (see page2–11

Demonstrates voltage thatprograms the VCO frequency

7 J2880 V SWITCH: Switchingpower supply voltagesignal (see page 2–5)

Demonstrates general oscillo-scope measurements

Page 21: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–2 QuickStart8 DPO Demonstration Board

Table 2–1: Oscilloscope signals (Cont.)

See Figure2–1 FunctionDescription

Testpoint

8 — I SWITCH current loop:Switching power supplycurrent signal (see page2–5)

Demonstrates general oscillo-scope measurements

9 TP6 RAND ANOM I: A dualrandom anomaly sig-nals (see page 2–9)

Demonstrates FastAcq

10 TP11 CHEAT II: Signal thatcoincides with theanomaly signal (seepage 2–9)

Demonstrates FastAcq

11 TP13/TP14

QAM I and Q: Base-band quadrature ampli-tude modulated signals(see page 2–6)

Demonstrates: X-Y XYZ FastAcq

12 TP16 RAND ANOM II: A dualrandom anomaly signal(see page 2–9)

Demonstrates FastAcq

13 TP19 CHEAT I: Signal thatcoincides with theanomaly signal (seepage 2–9)

Demonstrates FastAcq

14 TP21 QAM CLK: Signal thatcoincides with theanomaly signal (seepage 2–6)

Demonstrates: FastAcq X-Y XYZ

Page 22: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–3

The following Table 2–2 lists the USB DATA testpoint and switcheson the QuickStart8 board. See Figure 2–1 for testpoint and switchlocations.

Table 2–2: USB Data and switch functions

Switches Function Switch position Mode

J2 USB DATA: Makes avail-able signals from thehost computer

USB cable must beused to power theboard

S2780 Controls power to theboard

On/Off —

S2810 VCO MODE SELECT 1 OPEN/SHUT2 OPEN/SHUT3 OPEN/SHUT4 OPEN/SHUT

Not usedNot used(See Table1–1 on page1–3)

S2830 RESET ANOM: Causesrandom changes to theoutput frequency, ampli-tude, and anomaly fre-quency

— Reset ran-dom ANOM

Page 23: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–4 QuickStart8 DPO Demonstration Board

J600

VCO Switch Reset switch

J13

Power switch

Power jackUSB ports

USB data

1

2

3

4

5

6

7

8

9

10

11

12

13

14

SMA

Figure 2–1: Signal locations on the QuickStart8 board

Page 24: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–5

Detailed Signal Descriptions This section covers output signals on Quickstart 8 board thatdemonstrate the features of the TDS7000 series of oscilloscopes.

Switching Power Supply Signals: V SWITCH, I SWITCH

The V SWITCH (J2880) and I SWITCH signals are derived from theswitching power supply on the board. This power supply switches atabout 260 kHz. These signals were included to demonstrate thecurrent, voltage and power measurement capabilities on theoscilloscope (see Figure 2–2).

Figure 2–2: V switch, I switch

Page 25: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–6 QuickStart8 DPO Demonstration Board

Quadrature Amplitude Modulation Baseband Signals

QAM I, QAM Q, and QAM CLK. These signals emulate a basebandquadrature amplitude modulated digital signal. The clock rate of thesignal is about 1 MHz. This signal is designed to demonstrate anumber of features including FastAcq, XY and XYZ as shown inFigures 2–3 through 2–6.

Figure 2–3: QAM I, QAM Q, and QAM CLK signals

Figure 2–4: XY display

Page 26: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–7

Figure 2–5: Y-T FastAcq display

XYZ Mode

Using the XYZ mode in place of the XY mode, permits theinstrument to sample signals at precise times when the incomingsignal is expected to be at the correct logic level. The Z parametercomes from connecting an additional probe from Channel 3 to theQAM CLK signal. The resulting QAM pattern will have data pointsat precise locations without the noise shown in the XY QAM pattern.

Figure 2–6: XYZ FastAcq display

Page 27: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–8 QuickStart8 DPO Demonstration Board

A 250 mV Fast Rise Signal (Tr < 200 ps)

FAST RISE SIGNAL. This signal (J600) has an edge that is faster thanthe rise time of most oscilloscope, but Tektronix oscilloscope canmeasure this signal. There is a long (relative to rise time) 50 runon the circuit board that is exposed and plated to allow for probing.This is intended to demonstrate the ability of a probe and oscillo-scope system to display small time differences or skew. This signalcan also be used as a deskew signal source (see Figure 2–7).

Figure 2–7: Fast Rise Signal

Page 28: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–9

Dual Random Anomalies Signals

RAND ANOM I, CHEAT I, RAND ANOM II, CHEAT II. The RandomAnomalies signals are square waves that have occasional glitches,short pulses and other signal changes inserted at random intervals.Two channels are provided for demonstrating two channel anomalydetection (see Figures 2–8 and 2–9).

Figure 2–8: Random Anomaly with cheat signal

Figure 2–9: Dual Channel Random Anomalies

Page 29: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–10 QuickStart8 DPO Demonstration Board

RESET ANOM

RESET ANOM. This switch (S2830) changes the signal outputfrequency and amplitude in a random fashion. The number ofglitches per second is also varied by the RESET ANOM switch. Theamplitude will be a value between 2.7 and 3.2 volts and thefrequency will vary between 180 kHz and 320 kHz. The dual randomanomaly outputs can be synchronized or unsynchronized bycontinuing to depress the RESET ANOM switch.

Page 30: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–11

Phase Locked VCO Output Signals

PLL VCO OUTPUT. This voltage controlled oscillator output signal(J2410) provides a 2.5 volt (peak to peak) signal that is either85 MHz or 68 MHz. For example, the peak-to-peak voltage isapproximately 3.1 V when oscilloscope channel has 1M ofresistance or 1.8 V peak-to-peak when the oscilloscope channel has aresistance of 50

VCO TUNING VOLTAGE. This voltage signal drives the VCO. Phaselocked loop parameters determine how fast the VCO frequency willchange to a new programmed value.

The PLL VCO and VCO TUNING (J2520) signals are useful fordemonstrating the effect of different resolutions when comparing ahigh frequency envelope of a signal to a slowly varying controlsignal. This is demonstrated in Figures 2–10 and 2–11.

Figure 2–10: Low resolution display

Page 31: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–12 QuickStart8 DPO Demonstration Board

Figure 2–11: High resolution display

VCO MODE SELECT. This signal modifies the VCO output accordingto Table 1–1 on page 1–3. Switches 3 and 4 of S2810 (a DIP switch)allow selection of several modes of operation.

Page 32: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–13

RF Modulation

You can obtain a modulated RF signal by connecting a probe to J1located near the SMA connector (see page 2–8). The waveform hasan amplitude that modulates between 40 and 80 mV when viewingthe individual waveforms while the frequency modulation can beseen with a more compressed time scale.

Figure 2–12: Modulated RF display

Page 33: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–14 QuickStart8 DPO Demonstration Board

Pseudo Random Anomalies Signals

A five pin connector is included in the center of the board (J13) thatgenerates four random digital signals and a clock. These signals areuseful for demonstrating boolean trigger, state trigger, and otherkinds of triggers (see Figure 2–13).

Figure 2–13: Pseudo-Random digital signals

Page 34: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

QuickStart8 DPO Demonstration Board 2–15

Universal Serial Bus

USB DATA. USB signals from the host computer are available from J2(see Figure 2–14). This testpoint will generate signals only when theUSB cable is used to power the board. It is also necessary to connecta device such as a mouse to the USB output connector on theQuickStart8 board.

Figure 2–14: USB signals

Page 35: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Operating Basics

2–16 QuickStart8 DPO Demonstration Board

Simulated Ground Bounce Signal

An unexpected positive transition on channel 1 (J14, pin 1) isgenerated whenever the signal on channel two (J14, pin 2) transitionsfrom positive to negative. This positive transition occurs at intervalsof 0.1 seconds to several seconds (see Figure 2–15).

Figure 2–15: Simulated Ground Bounce signal

Page 36: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendices

Page 37: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board A–1

Appendix A: Diagrams

This section contains the schematic diagrams for the QuickStart8 board.

SymbolsGraphic symbols and class designation letters are based on ANSIStandard Y32.2-1975. Abbreviations are based on ANSI Y1.1-1972.

Logic symbology is based on ANSI/IEEE Standard 91-1984 in terms ofpositive logic. Logic symbols depict the logic function performed andcan differ from the manufacturer’s data.

The pound (*) after a signal name indicates that the signal performs itsintended function when in the low state.

Other standards used in the preparation of diagrams by Tektronix, Inc.,include the following:

Tektronix Standard 062-2476 Symbols and Practices for SchematicDrafting

ANSI Y14.159-1971 Interconnection Diagrams

ANSI Y32.16-1975 Reference Designations for Electronic Equip-ment

MIL-HDBK-63038-1A Military Standard Technical Manual WritingHandbook

Page 38: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–2 QuickStart8 DPO Demonstration Board

Component ValuesElectrical components shown on the diagrams are in the following unitsunless noted otherwise:

Capacitors: Values one or greater are in picofarads (pF).Values less than one are in microfarads (F).

Resistors: Values are in ohms ().

Graphic Items and Special Symbols Used in ThisManualEach assembly in the instrument is assigned an assembly number (forexample, A5). The assembly number appears in the title on the diagramand the circuit board illustration.

DigitalGround

AssemblyNumber

InternalScrew

AdjustmentFunction

Block TitleOnboardJumper

Refer toAssembly& DiagramNumber

OffboardConnector

Active LowSignal

Signal FromAnother Diagram,

Same Board

PowerTermination

Strap

PanelControl

FemaleCoaxial

Connector

Heat Sink

DecoupledVoltage

DiagramNumber

DiagramName

Component onback of board

*

Page 39: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–3

A1 QuickStart8 board component locator

Page 40: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–4 QuickStart8 DPO Demonstration Board

Page 41: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–5

Page 42: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–6 QuickStart8 DPO Demonstration Board

Page 43: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–7

Page 44: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–8 QuickStart8 DPO Demonstration Board

Page 45: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–9

Page 46: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–10 QuickStart8 DPO Demonstration Board

Page 47: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–11

Page 48: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–12 QuickStart8 DPO Demonstration Board

Page 49: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–13

Page 50: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–14 QuickStart8 DPO Demonstration Board

Page 51: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–15

Page 52: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–16 QuickStart8 DPO Demonstration Board

Page 53: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–17

Page 54: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–18 QuickStart8 DPO Demonstration Board

Page 55: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

QuickStart8 DPO Demonstration Board A–19

Page 56: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix A: Diagrams

A–20 QuickStart8 DPO Demonstration Board

Page 57: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board B–1

Appendix B: Instrument Care

This section includes:

Care and Maintenance

Shipping the QuickStart8 board

Care and MaintenanceThe QuickStart8 board does not require scheduled or periodicmaintenance. However, to keep good electrical contact and efficientheat dissipation, keep the board free of dirt, dust, and contaminants.When not in use, store the QuickStart8 board in the original shippingbag.

Cleaning. Clean dirt and dust with a soft bristle brush. For moreextensive cleaning, use only a damp cloth moistened with deionizedwater; do not use any other chemical cleaning agents.

Preventing Electrostatic Discharge. When handling the QuickStart8board, adhere to the following precautions to avoid damagingelectronic components.

CAUTION. Static discharge can damage semiconductor componentson the QuickStart8 board.

1. Minimize handling of the board by touching only the edges.

2. Transport and store the board in the static-protected container.

3. Discharge the static voltage from your body by wearing agrounded antistatic wrist strap while handling the training board.

4. Do not place anything capable of generating or holding a staticcharge on the work station surface.

Page 58: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Appendix B: Instrucment Care

B–2 QuickStart8 DPO Demonstration Board

Shipping the QuickStart8 board.

To commercially transport the QuickStart board or transformer,package the board or transformer as follows:

1. Obtain a corrugated card board shipping carton with insidedimensions at least six inches greater than the board ortransformer dimensions and with a carton test strength of at least889.6 Newton Lbs/200 pounds.

2. If you are shipping the package to a Tektronix service center forWarranty service, attach a tag to the board or transformershowing the following:

Owner’s name and address

Name of a person who can be contacted

Board or transformer type and serial number

Description of the problem

3. Place the board in an antistatic bag to protect static-sensitivecomponents.

4. Tightly pack dunnage or urethane foam between the carton andthe board or transformer (allowing 7.62 cm/three inches on eachside) to cushion the board or transformer on all sides.

Page 59: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Index

Page 60: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

QuickStart8 DPO Demonstration Board Index–1

Index

AAbout this manual, viiAddress, Tektronix, viii

CCare and maintenance, B–1CHEAT I, 2–9CHEAT II, 2–9Component locations

power jack and USB ports, 2–4switches, 2–4testpoints, 2–4

Configurationconnecting the board, 1–5VCO MODE SELECT switch,

1–3Contacting Tektronix, viii

FFAST RISE SIGNAL, 2–8

II SWITCH, 2–5

MManual, how to use, vii

PPhone number, Tektronix, viiiPLL VCO output (J2410), 2–11Power

adapter, 1–4applying, 1–4removing, 1–4transformer, 1–4USB port, 1–4

Prerequisites, 1–2Product support, contact informa-

tion, viii

QQuick reference

switches, 2–3testpoints, 2–1USB data, 2–3

QuickStart8 board, setting up, 1–2

RRAND ANOM I, 2–9RAND ANOM II, 2–9Random anomalies signals (J13),

2–14Required materials, 1–1

BNC/SMA, 1–1probes, 1–1

RESET ANOM (J2830), 2–10

Page 61: QuickStart8 DPO Demonstration Board for TDS7104 …€¦ · Instructions QuickStart8 DPO Demonstration Board for TDS7104 and TDS7054 Oscilloscopes 071-0715-00

Index

Index–2 QuickStart8 DPO Demonstration Board

SService support, contact informa-

tion, viiiShipping, B–2Signal descriptions, 2–5

250 mV fast rise (Tr < 200 ps),2–8

phase locked VCO output, 2–11quadrature amplitude modulation

baseband, 2–6random anomalies signal, 2–9simulated ground bounce signal,

2–16switching power supply, 2–5

Signal descriptions, pseudo-randomdigital, 2–14

Signals, overview, 1–6Switches, RESET ANOM (J2830),

2–10

TTechnical support, contact informa-

tion, viiiTektronix, contacting, viiiTestpoints

FAST RISE SIGNAL (J600), 2–8PLL VCO output (J2410), 2–11QAM I, QAM Q, and QAM

CLK, 2–6quick reference, 2–1RAND ANOM I, CHEAT I,

RAND ANOM II, CHEAT II,2–9

random anomalies signals (J13),2–14

simulated ground bounce signal(J14), 2–16

USB DATA (J2), 2–15V Switch, I Switch, 2–5VCO MODE SELECT, 2–12VCO TUNING VOLTAGE

(J2520), 2–11

UURL, Tektronix, viiiUSB DATA, 2–15

VV SWITCH, 2–5VCO MODE SELECT, 2–12VCO TUNING VOLTAGE

(J2520), 2–11

WWeb site address, Tektronix, viii

XXYZ mode, 2–7

YY-T mode, 2–7