R Wyllie, R Wakai, T G Walker University of Wisconsin, Madison Spin-Exchange Relaxation Free Heart signal frequency spectrum from DC-100Hz Adult heart signals have a peak amplitude ~100pT. A fetus signal has a peak amplitude of ~1pT. To detect fetal MCG thus requires a sensitivity of better than 100 fT/Hz 1/2 over the 100Hz bandwidth. The signals below are adult MCG taken in a 3-layer mu-metal room with our atomic magnetometer operating at about 40 fT/Hz 1/2 sensitivity. Biomagnetic Signals We have developed a portable atomic magnetometer (AM) suited for biomagnetic measurements. Using this magnetometer, we have been able to collect adult magnetocardiograms (MCG) and present a sensitivity which should enable the detection of fetal MCG. An Alkali vapor cell is optically pumped, creating an atomic alignment. This alignment then precesses in the presence of an orthogonal magnetic field. This precession causes a Faraday rotation of a mutually orthogonal probe beam, which is detected using a balanced polarimeter. For small fields, the signal is proportional to the rotation angle of the probe polarization, x ~ P x , the net atomic polarization. We can solve the rate equation to get P x : Magnetometer Design and Results Z-Mode Theory Pump optics Probe collimation optics Rb cell in ceramic and Teflon oven Polyamide Thermal Insulation Probe detection optics A Rb-87 cell with 400 Torr N 2 buffer gas is heated to 180C. A circularly polarized optical pump tuned to the D1 line provides a net electronic polarization along the pump axis. Another beam detuned from the D2 line is linearly polarized, and a rotation in probe polarization is detected on the balanced polarimeter. Atomic magnetometers work by detecting magnetically induced Larmor precession of electronic spins. The minimum detectable field goes as t time Measuremen T time coherence Spin spin sample of Number 1 N T N g B B s Spin exchange collisions cause a decoherence of the individual spins, and would limit usable densities for detecting individual spins. But generally, AMs are sensitive to net spin, which doesn’t decohere under spin-exchange, so densities can be turned up until other collisional processes become important. Three regimes of precession to collision rate, from Happer and Tang, PRA 16 1977 SE 1 T SE 1 T regime (SERF 1 SE T ) 1 2 ( : levels F different for different is Frequency Larmor 2 1 I B g B J I F The lasers are both couples to the optics tubes through optical fibers, and the entire apparatus is inside tri-axial helmholtz coils and is operated either in a small 4- layer mu-metal sheild or a 3-layer mu-metal room. Apply a large oscillating field along the pump direction and use lock-in detection. As shown below, the signal has two frequency components proportional to orthogonal field components. We also retain the same bandwidth as the original magnetometer. To the left, we demonstrate a noise reduction while using Z-Mode modulation. z y x x t t B z cos 0 x x B ~ z y x y t B z y y B ~ ) ( ) ( 2 ' ) ( 1 0 u J u J S S x z z x ) 2 sin( ) ( ) ( 2 ' ) 2 ( 2 0 t u J u J S S z x z z x q u 0 A Portable SERF Magnetometer for Biomagnetic Measurements This work is supported by the National Institutes of Health B g sR P P P B s z z z z x y z x , R ' , ' ' ' ' ' pr SD 2 2 2 2 2 2 Where R is the optical pumping rate, are spin-decoherence rates, and are Larmor frequencies. If DC residual fields are cancelled, then the polarization and frequency response are 2 2 P 1 P 2 6 ) P , 2 / 3 ( , P I, q S ) q( F y z x P P ' 2 2 2 1 1 ' q P P y z x Where q is the “slowing down factor,” which is the ratio of the free electron to net sample precession rates (see box on spin- exchange collisions). If a spin temperature distribution is assumed in the atomic gas, then we can write a closed form for Rb-87 with a polarization P: Thus, the probe beam polarization will, in this case, be sensitive to oscillating fields in the y-direction, and will have a low-pass filter type frequency response with a characteristic bandwidth determined by atomic species and experimental parameters.

R Wyllie, R Wakai, T G Walker University of Wisconsin, Madison Spin-Exchange Relaxation Free Heart signal frequency spectrum from DC-100Hz Adult heart

Embed Size (px)

Citation preview

Page 1: R Wyllie, R Wakai, T G Walker University of Wisconsin, Madison Spin-Exchange Relaxation Free Heart signal frequency spectrum from DC-100Hz Adult heart

R Wyllie, R Wakai, T G WalkerUniversity of Wisconsin, Madison

Spin-Exchange Relaxation Free

Heart signal frequency spectrum from DC-100Hz

Adult heart signals have a peak amplitude ~100pT.A fetus signal has a peak amplitude of ~1pT. To detect fetal MCG thus requires a sensitivity of better than 100 fT/Hz1/2 over the 100Hz bandwidth.

The signals below are adult MCG taken in a 3-layer mu-metal room with our atomic magnetometer operating at about 40 fT/Hz1/2 sensitivity.

Biomagnetic Signals

We have developed a portable atomic magnetometer (AM) suited for biomagnetic measurements. Using this magnetometer, we have been able to collect adult magnetocardiograms (MCG) and present a sensitivity which should enable the detection of fetal MCG.

An Alkali vapor cell is optically pumped, creating an atomic alignment. This alignment then precesses in the presence of an orthogonal magnetic field. This precession causes a Faraday rotation of a mutually orthogonal probe beam, which is detected using a balanced polarimeter. For small fields, the signal is proportional to the rotation angle of the probe polarization, x~ Px , the net atomic polarization. We can solve the rate equation to get Px:

Magnetometer Design and Results Z-ModeTheoryPump optics

Probe collimation optics

Rb cell in ceramic and Teflon oven

Polyamide Thermal Insulation

Probe detection optics

A Rb-87 cell with 400 Torr N2 buffer gas is heated to 180C. A circularly polarized optical pump tuned to the D1 line provides a net electronic polarization along the pump axis. Another beam detuned from the D2 line is linearly polarized, and a rotation in probe polarization is detected on the balanced polarimeter.

Atomic magnetometers work by detecting magnetically induced Larmor precession of electronic spins. The minimum detectable field goes as

t timeMeasuremenT timecoherenceSpin

spins sample ofNumber

1

NTNg

BBs

Spin exchange collisions cause a decoherence of the individual spins, and would limit usable densities for detecting individual spins. But generally, AMs are sensitive to net spin, which doesn’t decohere under spin-exchange, so densities can be turned up until other collisional processes become important.

Three regimes of precession to collision rate, from Happer and Tang, PRA 16 1977

SE1T

SE

1T

regime) (SERF 1

SET

)12(

:levels Fdifferent for different isFrequency Larmor 21

I

Bg BJIF

The lasers are both couples to the optics tubes through optical fibers, and the entire apparatus is inside tri-axial helmholtz coils and is operated either in a small 4-layer mu-metal sheild or a 3-layer mu-metal room.

Apply a large oscillating field along the pump direction and use lock-in detection. As shown below, the signal has two frequency components proportional to orthogonal field components. We also retain the same bandwidth as the original magnetometer.To the left, we demonstrate a noise reduction while using Z-Mode modulation.

z

y

x

x

ttBz cos0

xx B~

z

y

x

y

tBz

yy B~

)()(2'

)( 10 uJuJSS xzzx

)2sin()()(2'

)2( 20 tuJuJSS zx

zzx

qu 0

A Portable SERF Magnetometer for Biomagnetic Measurements

This work is supported by the National Institutes of Health

BgsRP

PP

Bsz

z

z

zxyzx

,R' ,''

'

'

'

prSD22

22

22

Where R is the optical pumping rate, are spin-decoherence rates, and are Larmor frequencies.If DC residual fields are cancelled, then the polarization and frequency response are

2

2

P1P26)P,2/3(,PI,

qS)q(F

yz

xP

P

' 222

11

' qPP yzx

Where q is the “slowing down factor,” which is the ratio of the free electron to net sample precession rates (see box on spin-exchange collisions). If a spin temperature distribution is assumed in the atomic gas, then we can write a closed form for Rb-87 with a polarization P:

Thus, the probe beam polarization will, in this case, be sensitive to oscillating fields in the y-direction, and will have a low-pass filter type frequency response with a characteristic bandwidth determined by atomic species and experimental parameters.