Radiation Exposure Monitoring(REM) Profile
Kevin O’DonnellToshiba Medical Research Inst. - USACo-chair, IHE Radiology Planning Cmte
IHE Radiology
2
MotivationMotivation• X-Ray based imaging can provide tremendously
useful information
• Patient Dose is an important consideration
• Potential benefit > potential risk … but risk should be managed responsibly
3
A New IHE ProfileA New IHE Profile• IHE Radiation Exposure Monitoring Profile – Integration of systems reporting dose and systems
which receive, store, or process those reports– Modalities, PACS, RIS, Workstations, Registries
– Facilitate compliance with Euratom 97/43,ACR Guidelines, etc.
– Directly based on DICOM Dose Reports
4
IHE Radiation Exposure Monitoring ProfileIHE Radiation Exposure Monitoring Profile
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
National Registry
ArchiveDose Information
Reporter
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
Outlier: #Performing Phys.Over Target: 12.2%
Outlier: #Performing Phys.Over Target: 12.2%
Outlier: #Performing Phys.Over Target: 12.2%
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
some text: #Numerical Details 12.2 14.5 11.8 7.6 9.5 10.9
Numer 12.2 14.5 9.5 10.9
5
REM Profile: Actors & TransactionsREM Profile: Actors & Transactions
6
DICOM Dose ReportsDICOM Dose Reports• “SR Objects” – DICOM Structured Reports– Easily ingested (and regurgitated) by PACS
• Granularity : “Irradiation Event”– & Accumulated Dose over Study, Series
• Templates : – CT, Projection X-Ray (Mammo, Fluoro)– DR/CR
• Not yet addressed: NM, RT
7
Key MeasurementsKey Measurements• CT Dose– DLP, CTDIvol, kVP, mA, sec– Effective Dose [ Optional ]
(Reference estimation method)
• Projection X-Ray Dose– DAP, Dose@RP, kVP, mA, sec– Fluoro Dose, Fluoro Time
• Mammography Dose– AGD, Entrance Exposure@RP, kVP, mA, sec– Compression, Half Value Layer
ftp://medical.nema.org/medical/dicom/2009/09_16pu.pdf
8
Other Details in Dose SROther Details in Dose SR• Full Patient / Order / Study Details• Unique ID for each Irradiation Event• Equipment ID, Ordering Doc, Performing Tech
• Patient Position, Anatomy imaged• Imaging Geometry (projection)
• Collimation• X-Ray Filters, Anode Target Material• Calibration, Phantom, Dosimeter
DICOM SR TemplateDICOM SR Template
9
10
Creating SR Dose ReportsCreating SR Dose Reports• Modalities– CT, XA/XRF, X-Ray, Mammography
• “Readers”– CR, DX, Film Digitizers– Note challenge if tube and detector don’t talk…
• 3rd Party Workstations
• RIS, PACS
11
Storing SR Dose ReportsStoring SR Dose Reports
• Usually stored in the Study folder– Archive, Backup, Reconciliation– Query / Retrieve
• Can be sent to other destinations
12
Using SR Dose ReportsUsing SR Dose ReportsPossible applications:• Radiation QA
– Periodically Query / Retrieve Reports from Archive– Set policies/standards and flag deviations– Set goals for improvement and track; – Implement protocol changes and compare difference in dose
• Patient Impact Evaluation– e.g. if Patient identified as pregnant post-facto
• Dose Mapping– Store data in realtime from Modality to Mapping Workstation
13
Using SR Dose ReportsUsing SR Dose Reports• National Registries– Anonymize and submit Dose Reports to Register– Compile Population Risk Estimations– Derive Dose Reference Levels– Provide Site-Site Comparisons
• Individual Dose Record– Collect Dose Reports over time
• Clinical Trials– Collect Dose together with Images– Demonstrate both improved detection & reduced dose
IHE REM Changes• Final Text CPs
– CP195 - Internationalize Series Description (RAD-62)– CP197 - Replace Register with Registry– CP198 - Remove weakening of DICOMDIR (RAD-63)
• Almost Final Text CPs– CP196 - REM De-identification (RAD-63)– CP199 - Patient Longitudinal Record Use Case– CP200 - Alternative Submission Transport Mechanisms (RAD-63)– CP203 - Require Irradiation Event UID in Images– CP204 - Record Patient Characteristics in Dose Reports
( See www.ihe.net/Technical_Framework/ )
14
DICOM SR ChangesDICOM SR Changes• Incorporated CPs
– Use DICOM 2009 PS3.16 – incorporates CP-874 and more
• Final Text CPs– CP-957 – Add CT Exposure Modulation Type– CP-963 – Add DateTime Started to Irradiation Event
• WIP CPs– CP-1063 – Clean-up Dose Reports– CP-1065 – Improve Device Identification– CP-1068 – CT Scan Location– CP-1070 – Additional Codes for Dose Information Source– CP-1075 – CT Dose for Constant Angle Acquisitions– CP-1076 – Fix participant relationship in TID1020– CP-1077 – CR Dose Report
( See dclunie.com/dicom-status/status.html )15
Profile Overviews:Profile Overviews:wiki.ihe.net/index.php?title=Profiles
Profile Specs.:Profile Specs.:www.ihe.net/Technical_Framework
17
Why not use Image Headers?Why not use Image Headers?
• Mostly Exposure details (for interpretation)
• No image; no dose– deleted due to quality / patient motion
• More images; more dose– extra reconstructions; post-processing
• Lacks complete dose details
18
Why not use MPPS?Why not use MPPS?
• Modality Performed Procedure Steps– Status messages back to RIS / PACS
• Transient Information– designed for workflow; not persistent archiving
• Lacks complete dose details