Scanditronix Usermeeting

June 2017 Kopenhagen

Re-engineering on an Ion Source Base

Page 2: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Heidelberg Ion Source MC32NI

Page 3: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Issue: Anode Base

• Finding the right drawings • Creation of an 3D-Model • 3D printing • Proof of plausibility

Page 4: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Issue: Anode Base

• Is the outer ring really out of oxygen free copper or is Elmedur X possible?

• Mixing of materials with different thermal expansion rate could lead into bending or cracks

Page 5: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Material: ElmedurX, Cowodur, Oxygenfree Copper?

• Comparing our „museum“ of Ion Sources and other stuff

• Polishing the parts and comparing with material samples from workshop

Page 6: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Secret solder reservoir and gap precission

Capillary Effect Design

Page 7: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Atmospheric condition soldering trys

Page 8: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Choosing temperature gradient

950°C Surface disintegration (real datasheet value 893°C from company) ->ElmedurX is not an eutectic system. It is warm hardened

Page 9: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Surface Preperation - Cleaning

Cleaning test with 1mol HCL for 100 hours -> No weight or size changes within tolerances

ElmedurX sample with nailpaint covers (thanks to our secretary Sandra) to measure thickness of plating

Page 10: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Surface Preperation - Electroplating

Page 11: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Choosing the right solder

• Success with „honey like“ solder

• Liquid solder flows to much during long heating and cooling time

Page 12: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Soldering

• Vacuum: 1x10-8mbar • Duration: Whole night • Temperature drift ±10°𝐶

(after reflectance / emittance calibration)

• Final temperature 810°C

MPI Kernphysik Heidelberg

Martin Beckmann

Page 13: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Sucessful Soldering !!!

Gold is totally absorbed into the metallic crystal structure

Page 14: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Pressure Testing Device

Page 15: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Trial at Copenhagen by HRS

Page 16: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Holgers Try

Page 17: Re-engineering on an Ion Source Base 2017 Sebastian Stricker.pdf · emittance calibration) • Final temperature 810°C MPI Kernphysik Heidelberg Martin Beckmann . Scanditronix Usermeeting

Scanditronix Usermeeting

June 2017 Kopenhagen

Before Automation

Special Thanks to: Precission Engineering Workshop DKFZ Heidelberg