References 1. Kohnen T. How far we have come: from Ridley’s first intraocular lens to modern IOL technology. 1. Kohnen T. How far we have come: from Ridley’s first intraocular lens to modern IOL technology. J Cataract Refract Surg 2009;35:2039. J Cataract Refract Surg 2009;35:2039. 2. Dick HB, Krummenauer F, Schwenn O, et al. 2. Dick HB, Krummenauer F, Schwenn O, et al. Objective and subjective evaluation of photic Objective and subjective evaluation of photic phenomena after monofocal and multifocal intraocular lens implantation. Ophthalmology phenomena after monofocal and multifocal intraocular lens implantation. Ophthalmology 1999;106:1878-1886. 1999;106:1878-1886. 3. Neuhann IM, Kleinmann G, Apple DJ. 3. Neuhann IM, Kleinmann G, Apple DJ. „Special“ intraocular lens design: why do they sometimes „Special“ intraocular lens design: why do they sometimes fail? Compr Ophthalmol Update 2006; 7:53-60. fail? Compr Ophthalmol Update 2006; 7:53-60. 4. Abela-Formanek C, Amon M, Schauersberger J, et al. 4. Abela-Formanek C, Amon M, Schauersberger J, et al. Results of hydrophilic acrylic, hydrophobic Results of hydrophilic acrylic, hydrophobic acrylic, and silicon intraocular lenses in uveitic eyes with cataract. Comparison to a control acrylic, and silicon intraocular lenses in uveitic eyes with cataract. Comparison to a control group. J Cataract Refract Surg 2002;28:1141-1152. group. J Cataract Refract Surg 2002;28:1141-1152. Visual and refractive outcomes after implantation Visual and refractive outcomes after implantation of a new intraocular lens: 6-month follow-up. of a new intraocular lens: 6-month follow-up. Giuseppe Migliorati, Maria L. Salvetat, Marco Zeppieri and Paolo Brusini Giuseppe Migliorati, Maria L. Salvetat, Marco Zeppieri and Paolo Brusini Department of Ophthalmology, S. Department of Ophthalmology, S. Maria della Misericordia Hospital, Udine, Italy. Maria della Misericordia Hospital, Udine, Italy. Introduction This preliminary study shows that the new “Apeiron” 42F IOL seems to be an effective and safe surgical alternative to traditional IOLs. It provided favorable clinical performances, which included ease of implantation, minimal postoperative uveitis, and good optic centration. BCVA significantly improved after surgery in all cases without significantly induced astigmatism. The slight postoperative hyperopic refractive error induce by the IOL is probably due to the posterior optical axial displacement. 9 A modified proposed constant may reduce the surgically induced refractive error. No eyes showed postoperative visually significant anterior capsule contraction or PCO. This IOL is advantageous in that it completely distends the capsular bag, which theoretically reduces IOL dislocation and PCO, thus providing an excellent barrier between the anterior and posterior eye segment. Moreover, the posterior position of the IOL, which is provided by the large angle between the optic disc and the loops, should limit the posterior vitreous from moving forward, thus reducing the risk of postoperative retinal detachment. Further long-term multicenter studies with this new IOL are needed to determine the rate of postoperative complications, especially PCO and retinal detachment. Financial disclosure: GM is the inventor of the new IOL. The authors have no financial interest in any devices used in the study. Since Ridley’s original idea in 1949, several artificial intraocular lens (IOL) types have been designed to optimize visual outcomes after cataract surgery. 1 Modern research in this field is centred on IOL design and material that have optical properties similar to the natural, transparent, young human lens. New IOLs are currently being designed to reduce postoperative complications, such as optical aberrations, decentration, tilting, posterior capsular opacification (PCO), and risk of retinal detachment. 2,3 The “Apeiron” 42F is a new IOL designed by GM, produced by Morcher GmbH, and commercially available since 2009 (Fig.1). It is an acrylic hydrophilic foldable IOL with a 6 mm optics and an external circular loop diameter of 11 mm; the loops are angled at 22° (Fig.2). It can be folded and inserted through a 2.2 mm clear corneal incision (CCI) (Fig.3). Methods Conclusions ARVO Meeting Fort Lauderdale, Florida, USA May 2-6, 2010 Poster: D939 Purpose : To evaluate visual acuity (VA), refractive outcomes, optic IOL position and postoperative complications after implantation of the new foldable Apeiron 42F Morcher IOL. Fig.1. “Apeiron” 42F Morcher IOL This prospective surgical study included 25 eyes of 25 consecutive subjects (mean age of 71±12 years, range 55-86) undergoing cataract extraction with phacoemulsification and implantation of an IOL in the capsular bag. Biometry was performed with the IOLMaster optical biometer (Carl Zeiss Meditec AG). IOL power was chosen to obtain an emmetropic postoperative refraction. The same experienced surgeon (PB) performed all surgeries using a standardized technique and peribulbar anesthesia. The technique included a temporal 2.5 mm CCI, continuous curvilinear capsulorhexis of 5.5 mm, phacoemulsification, and IOL “Apeiron” 42F (Morcher GmbH, Stuttgart, Germany) implantation in the capsular bag with a 2.2 mm cartridge system (Microglide injector, Medicel AG, Switzerland) (Fig.3). Best corrected VA (BCVA), manifest refractive spherical-equivalent (MRSE), refractive astigmatism (RA), and corneal endothelial cell density (ECD) were assessed 6 months postoperatively and compared with the preoperative values using the paired Student t-test. The MRSE and RA were evaluated with the Automatic Refractor/Keratometer 599 (Zeiss Humphrey, Dublin, CA). ECD was assessed using the non-contact specular microscope Tomey EM-3000 (Tomey, Japan). Postoperative IOL position was assessed with the ultrasound biomicroscopy (UBM) (Vumax 2, Sonomed Inc., NY,USA) using a 35 MHz transducer -The 11-mm diameter external circular loop may provide a better centration inside the capsular bag, which is important in postoperative visual outcome. IOL decentration usually gives rise to coma-like aberrations. 2 -The 22° angle between the IOL optic disc and the external circular loop is similar to that found between the equator and the posterior pole in the human crystalline lens, which may limit the anterior displacement of the vitreous body, thus reducing the risk of vitreo-retinal tractions or retinal ruptures; 7 - The foldable IOL insertion through a small CCI tends to minimize surgically induced astigmatism and optical aberrations, resulting in better postoperative visual outcomes. 8 All surgeries were completed within a time period of 10 minutes. No intraoperative complications were observed. 5. Tetz M, Wildeck A. Evaluating and defining the sharpness of intraocular lenses. Part 1: influence 5. Tetz M, Wildeck A. Evaluating and defining the sharpness of intraocular lenses. Part 1: influence of optic design on the growth of the lens epithelial cells in vitro. J Cataract Refract Surg of optic design on the growth of the lens epithelial cells in vitro. J Cataract Refract Surg 2005;31:2172-2179. 2005;31:2172-2179. 6. Vock L, Menapace R, Stifter E, et al. Posterior capsule opacification and neodymium:YAG laser 6. Vock L, Menapace R, Stifter E, et al. Posterior capsule opacification and neodymium:YAG laser capsulotomy rates with a round-edged silicone and a sharp-edged hydrophobic acrylic intraocular lens capsulotomy rates with a round-edged silicone and a sharp-edged hydrophobic acrylic intraocular lens 10 years after surgery. J Cataract Refract Surg 2009;35:459-465. 10 years after surgery. J Cataract Refract Surg 2009;35:459-465. 7. Fechner PU, Kania J, Kienzle S. The value of a zero power intra-ocular lens. J Cataract Refract 7. Fechner PU, Kania J, Kienzle S. The value of a zero power intra-ocular lens. J Cataract Refract Surg 1988;14:436-440. Surg 1988;14:436-440. 8. Elkady B, Aliò JL, Ortiz D, et al. 8. Elkady B, Aliò JL, Ortiz D, et al. Corneal aberrations after microincision cataract surgery. J Corneal aberrations after microincision cataract surgery. J Cataract Refract Surg 2008;34:40-45. Cataract Refract Surg 2008;34:40-45. 9. Erikson P. Effects of intraocular lens position errors on post-operative refractive error. J 9. Erikson P. Effects of intraocular lens position errors on post-operative refractive error. J Cataract Refract Surg 1990;16:305-311. Cataract Refract Surg 1990;16:305-311. Results Fig.5 The RA did not change significantly 6 months after surgery (p=0.47). Fig.7 The mean ECD loss 6 months after surgery was 3.3 ± 3.4% (range 0.5 – 11.6%) compared to pre-op data (p=0.02). Fig.6 A tendency toward a slight hyperopic refractive error was observed 6 months after surgery. Fig.4 The mean BCVA significantly improved 6 months postoperatively compared to the preoperative data (p=0.0001). Fig.8A. UBM image of a cataractous crystalline lens before surgery. Fig.8B. A traditional IOL posizionated inside the capsular bag in the fellow eye. Fig.8C. The “Apeiron” 42F Morcher IOL positioned inside the capsular bag at 6 months after surgery. All UBM images taken 6 months after surgery showed good IOL centration in the capsular bag and posterior optic displacement. Fig.2. Technical characteristics of the 42F IOL. The new “Apeiron” 42F Morcher IOL theoretically provides several advantages: - The hydrophilic acrylic material has shown to provide high uveal biocompatibility. 4 - The square-edge optic design has shown to decrease the development of PCO due to the reduced space between the IOL and the posterior capsule. This may limit lens epithelial cell proliferation and metaplasia; 5 the issue, however, remains controversial. 6 The “Apeiron” 42F IOL is folded and inserted in the capsular bag through a CCI of 2.5 mm. Fig.3 Fig.2 B Fig.8 A C R efractive astigm atism D iopters 1.2 1.0 0.8 0.6 0.4 0.2 0.0 pre-op post-op 0.5 ± 0.5 0.6 ± 0.4 diopters B estcorrected visualacuity Decim a 1.0 0.8 0.6 0.4 0.2 0.0 0.37 ± 0.2 0.9 ± 0.1 pre-op post-op Snellen lines M anifestrefractive sphericalequivalent D iopters 3 2 1 0 -1 -2 -3 -4 -5 pre-op post-op -0.9 ± 2.0 -0.75 ± 0.4 diopters The mean distance between the endothelium and the IOL optic posterior surface 6 months post-op was 6.1 ± 0.5 mm (range 5.2 – 6.5). C ornealendothelialcelldensity cells/mm 2 1600 1800 2000 2200 2400 2600 2800 pre-op post-op 2256 ± 303 2183 ± 321 cells/mm 2

References 1. Kohnen T. How far we have come: from Ridley’s first intraocular lens to modern IOL technology. J Cataract Refract Surg 2009;35:2039. 2. Dick

Embed Size (px)

Citation preview

Page 1: References 1. Kohnen T. How far we have come: from Ridley’s first intraocular lens to modern IOL technology. J Cataract Refract Surg 2009;35:2039. 2. Dick

References1. Kohnen T. How far we have come: from Ridley’s first intraocular lens to modern IOL technology. J Cataract Refract Surg 1. Kohnen T. How far we have come: from Ridley’s first intraocular lens to modern IOL technology. J Cataract Refract Surg 2009;35:2039.2009;35:2039.2. Dick HB, Krummenauer F, Schwenn O, et al. 2. Dick HB, Krummenauer F, Schwenn O, et al. Objective and subjective evaluation of photic phenomena after monofocal Objective and subjective evaluation of photic phenomena after monofocal and multifocal intraocular lens implantation. Ophthalmology 1999;106:1878-1886.and multifocal intraocular lens implantation. Ophthalmology 1999;106:1878-1886.3. Neuhann IM, Kleinmann G, Apple DJ. 3. Neuhann IM, Kleinmann G, Apple DJ. „Special“ intraocular lens design: why do they sometimes fail? Compr Ophthalmol „Special“ intraocular lens design: why do they sometimes fail? Compr Ophthalmol Update 2006; 7:53-60.Update 2006; 7:53-60.4. Abela-Formanek C, Amon M, Schauersberger J, et al. 4. Abela-Formanek C, Amon M, Schauersberger J, et al. Results of hydrophilic acrylic, hydrophobic acrylic, and silicon Results of hydrophilic acrylic, hydrophobic acrylic, and silicon intraocular lenses in uveitic eyes with cataract. Comparison to a control group. J Cataract Refract Surg 2002;28:1141-1152.intraocular lenses in uveitic eyes with cataract. Comparison to a control group. J Cataract Refract Surg 2002;28:1141-1152.

Visual and refractive outcomes after implantation Visual and refractive outcomes after implantation of a new intraocular lens: 6-month follow-up.of a new intraocular lens: 6-month follow-up.

Giuseppe Migliorati, Maria L. Salvetat, Marco Zeppieri and Paolo BrusiniGiuseppe Migliorati, Maria L. Salvetat, Marco Zeppieri and Paolo Brusini Department of Ophthalmology, S.Department of Ophthalmology, S. Maria della Misericordia Hospital, Udine, Italy. Maria della Misericordia Hospital, Udine, Italy.

IntroductionIntroduction

This preliminary study shows that the new “Apeiron” 42F IOL seems to be an effective and safe surgical alternative to traditional IOLs. It provided favorable clinical performances, which included ease of implantation, minimal postoperative uveitis, and good optic centration. BCVA significantly improved after surgery in all cases without significantly induced astigmatism. The slight postoperative hyperopic refractive error induce by the IOL is probably due to the posterior optical axial displacement.9 A modified proposed constant may reduce the surgically induced refractive error. No eyes showed postoperative visually significant anterior capsule contraction or PCO. This IOL is advantageous in that it completely distends the capsular bag, which theoretically reduces IOL dislocation and PCO, thus providing an excellent barrier between the anterior and posterior eye segment. Moreover, the posterior position of the IOL, which is provided by the large angle between the optic disc and the loops, should limit the posterior vitreous from moving forward, thus reducing the risk of postoperative retinal detachment. Further long-term multicenter studies with this new IOL are needed to determine the rate of postoperative complications, especially PCO and retinal detachment.

This preliminary study shows that the new “Apeiron” 42F IOL seems to be an effective and safe surgical alternative to traditional IOLs. It provided favorable clinical performances, which included ease of implantation, minimal postoperative uveitis, and good optic centration. BCVA significantly improved after surgery in all cases without significantly induced astigmatism. The slight postoperative hyperopic refractive error induce by the IOL is probably due to the posterior optical axial displacement.9 A modified proposed constant may reduce the surgically induced refractive error. No eyes showed postoperative visually significant anterior capsule contraction or PCO. This IOL is advantageous in that it completely distends the capsular bag, which theoretically reduces IOL dislocation and PCO, thus providing an excellent barrier between the anterior and posterior eye segment. Moreover, the posterior position of the IOL, which is provided by the large angle between the optic disc and the loops, should limit the posterior vitreous from moving forward, thus reducing the risk of postoperative retinal detachment. Further long-term multicenter studies with this new IOL are needed to determine the rate of postoperative complications, especially PCO and retinal detachment.

Financial disclosure: GM is the inventor of the new IOL. The authors have no financial interest in any devices used in the study. Financial disclosure: GM is the inventor of the new IOL. The authors have no financial interest in any devices used in the study.

Since Ridley’s original idea in 1949, several artificial intraocular lens (IOL) types have been designed to optimize visual outcomes after cataract surgery.1 Modern research in this field is centred on IOL design and material that have optical properties similar to the natural, transparent, young human lens. New IOLs are currently being designed to reduce postoperative complications, such as optical aberrations, decentration, tilting, posterior capsular opacification (PCO), and risk of retinal detachment.2,3 The “Apeiron” 42F is a new IOL designed by GM, produced by Morcher GmbH, and commercially available since 2009 (Fig.1). It is an acrylic hydrophilic foldable IOL with a 6 mm optics and an external circular loop diameter of 11 mm; the loops are angled at 22° (Fig.2). It can be folded and inserted through a 2.2 mm clear corneal incision (CCI) (Fig.3).

Since Ridley’s original idea in 1949, several artificial intraocular lens (IOL) types have been designed to optimize visual outcomes after cataract surgery.1 Modern research in this field is centred on IOL design and material that have optical properties similar to the natural, transparent, young human lens. New IOLs are currently being designed to reduce postoperative complications, such as optical aberrations, decentration, tilting, posterior capsular opacification (PCO), and risk of retinal detachment.2,3 The “Apeiron” 42F is a new IOL designed by GM, produced by Morcher GmbH, and commercially available since 2009 (Fig.1). It is an acrylic hydrophilic foldable IOL with a 6 mm optics and an external circular loop diameter of 11 mm; the loops are angled at 22° (Fig.2). It can be folded and inserted through a 2.2 mm clear corneal incision (CCI) (Fig.3).

MethodsMethods

ConclusionsConclusions

ARVO Meeting Fort Lauderdale,

Florida, USA May 2-6, 2010Poster: D939

ARVO Meeting Fort Lauderdale,

Florida, USA May 2-6, 2010Poster: D939

Purpose:Purpose: To evaluate visual acuity (VA), refractive outcomes, optic IOL position and postoperative complications after implantation of the new foldable Apeiron 42F Morcher IOL. To evaluate visual acuity (VA), refractive outcomes, optic IOL position and postoperative complications after implantation of the new foldable Apeiron 42F Morcher IOL.

Fig.1. “Apeiron” 42F Morcher IOL Fig.1. “Apeiron” 42F Morcher IOL

This prospective surgical study included 25 eyes of 25 consecutive subjects (mean age of 71±12 years, range 55-86) undergoing cataract extraction with phacoemulsification and implantation of an IOL in the capsular bag. Biometry was performed with the IOLMaster optical biometer (Carl Zeiss Meditec AG). IOL power was chosen to obtain an emmetropic postoperative refraction. The same experienced surgeon (PB) performed all surgeries using a standardized technique and peribulbar anesthesia. The technique included a temporal 2.5 mm CCI, continuous curvilinear capsulorhexis of 5.5 mm, phacoemulsification, and IOL “Apeiron” 42F (Morcher GmbH, Stuttgart, Germany) implantation in the capsular bag with a 2.2 mm cartridge system (Microglide injector, Medicel AG, Switzerland) (Fig.3). Best corrected VA (BCVA), manifest refractive spherical-equivalent (MRSE), refractive astigmatism (RA), and corneal endothelial cell density (ECD) were assessed 6 months postoperatively and compared with the preoperative values using the paired Student t-test. The MRSE and RA were evaluated with the Automatic Refractor/Keratometer 599 (Zeiss Humphrey, Dublin, CA). ECD was assessed using the non-contact specular microscope Tomey EM-3000 (Tomey, Japan). Postoperative IOL position was assessed with the ultrasound biomicroscopy (UBM) (Vumax 2, Sonomed Inc., NY,USA) using a 35 MHz transducer probe.

This prospective surgical study included 25 eyes of 25 consecutive subjects (mean age of 71±12 years, range 55-86) undergoing cataract extraction with phacoemulsification and implantation of an IOL in the capsular bag. Biometry was performed with the IOLMaster optical biometer (Carl Zeiss Meditec AG). IOL power was chosen to obtain an emmetropic postoperative refraction. The same experienced surgeon (PB) performed all surgeries using a standardized technique and peribulbar anesthesia. The technique included a temporal 2.5 mm CCI, continuous curvilinear capsulorhexis of 5.5 mm, phacoemulsification, and IOL “Apeiron” 42F (Morcher GmbH, Stuttgart, Germany) implantation in the capsular bag with a 2.2 mm cartridge system (Microglide injector, Medicel AG, Switzerland) (Fig.3). Best corrected VA (BCVA), manifest refractive spherical-equivalent (MRSE), refractive astigmatism (RA), and corneal endothelial cell density (ECD) were assessed 6 months postoperatively and compared with the preoperative values using the paired Student t-test. The MRSE and RA were evaluated with the Automatic Refractor/Keratometer 599 (Zeiss Humphrey, Dublin, CA). ECD was assessed using the non-contact specular microscope Tomey EM-3000 (Tomey, Japan). Postoperative IOL position was assessed with the ultrasound biomicroscopy (UBM) (Vumax 2, Sonomed Inc., NY,USA) using a 35 MHz transducer probe.

-The 11-mm diameter external circular loop may provide a better centration inside the capsular bag, which is important in postoperative visual outcome. IOL decentration usually gives rise to coma-like aberrations.2

-The 22° angle between the IOL optic disc and the external circular loop is similar to that found between the equator and the posterior pole in the human crystalline lens, which may limit the anterior displacement of the vitreous body, thus reducing the risk of vitreo-retinal tractions or retinal ruptures;7

- The foldable IOL insertion through a small CCI tends to minimize surgically induced astigmatism and optical aberrations, resulting in better postoperative visual outcomes.8

-The 11-mm diameter external circular loop may provide a better centration inside the capsular bag, which is important in postoperative visual outcome. IOL decentration usually gives rise to coma-like aberrations.2

-The 22° angle between the IOL optic disc and the external circular loop is similar to that found between the equator and the posterior pole in the human crystalline lens, which may limit the anterior displacement of the vitreous body, thus reducing the risk of vitreo-retinal tractions or retinal ruptures;7

- The foldable IOL insertion through a small CCI tends to minimize surgically induced astigmatism and optical aberrations, resulting in better postoperative visual outcomes.8

All surgeries were completed within a time period of 10 minutes. No intraoperative complications were observed. All surgeries were completed within a time period of 10 minutes. No intraoperative complications were observed.

5. Tetz M, Wildeck A. Evaluating and defining the sharpness of intraocular lenses. Part 1: influence of optic design on the growth 5. Tetz M, Wildeck A. Evaluating and defining the sharpness of intraocular lenses. Part 1: influence of optic design on the growth of the lens epithelial cells in vitro. J Cataract Refract Surg 2005;31:2172-2179.of the lens epithelial cells in vitro. J Cataract Refract Surg 2005;31:2172-2179.6. Vock L, Menapace R, Stifter E, et al. Posterior capsule opacification and neodymium:YAG laser capsulotomy rates with a round-6. Vock L, Menapace R, Stifter E, et al. Posterior capsule opacification and neodymium:YAG laser capsulotomy rates with a round-edged silicone and a sharp-edged hydrophobic acrylic intraocular lens 10 years after surgery. J Cataract Refract Surg edged silicone and a sharp-edged hydrophobic acrylic intraocular lens 10 years after surgery. J Cataract Refract Surg 2009;35:459-465.2009;35:459-465.7. Fechner PU, Kania J, Kienzle S. The value of a zero power intra-ocular lens. J Cataract Refract Surg 1988;14:436-440.7. Fechner PU, Kania J, Kienzle S. The value of a zero power intra-ocular lens. J Cataract Refract Surg 1988;14:436-440.8. Elkady B, Aliò JL, Ortiz D, et al. 8. Elkady B, Aliò JL, Ortiz D, et al. Corneal aberrations after microincision cataract surgery. J Cataract Refract Surg 2008;34:40-45.Corneal aberrations after microincision cataract surgery. J Cataract Refract Surg 2008;34:40-45.9. Erikson P. Effects of intraocular lens position errors on post-operative refractive error. J Cataract Refract Surg 1990;16:305-311.9. Erikson P. Effects of intraocular lens position errors on post-operative refractive error. J Cataract Refract Surg 1990;16:305-311.

Results Results

Fig.5 The RA did not change significantly 6 months after surgery (p=0.47).

Fig.5 The RA did not change significantly 6 months after surgery (p=0.47).

Fig.7 The mean ECD loss 6 months after surgery was 3.3 ± 3.4% (range 0.5 – 11.6%) compared to pre-op data (p=0.02).

Fig.7 The mean ECD loss 6 months after surgery was 3.3 ± 3.4% (range 0.5 – 11.6%) compared to pre-op data (p=0.02).

Fig.6 A tendency toward a slight hyperopic refractive error was observed 6 months after surgery.

Fig.6 A tendency toward a slight hyperopic refractive error was observed 6 months after surgery.

Fig.4 The mean BCVA significantly improved 6 months postoperatively compared to the preoperative data (p=0.0001).

Fig.4 The mean BCVA significantly improved 6 months postoperatively compared to the preoperative data (p=0.0001).

Fig.8A. UBM image of a cataractous crystalline lens before surgery. Fig.8B. A traditional IOL posizionated inside the capsular bag in the fellow eye. Fig.8C. The “Apeiron” 42F Morcher IOL positioned inside the capsular bag at 6 months after surgery. All UBM images taken 6 months after surgery showed good IOL centration in the capsular bag and posterior optic displacement.

Fig.8A. UBM image of a cataractous crystalline lens before surgery. Fig.8B. A traditional IOL posizionated inside the capsular bag in the fellow eye. Fig.8C. The “Apeiron” 42F Morcher IOL positioned inside the capsular bag at 6 months after surgery. All UBM images taken 6 months after surgery showed good IOL centration in the capsular bag and posterior optic displacement.

Fig.2. Technical characteristics of the 42F IOL. Fig.2. Technical characteristics of the 42F IOL.

The new “Apeiron” 42F Morcher IOL theoretically provides several advantages:- The hydrophilic acrylic material has shown to provide high uveal biocompatibility.4

- The square-edge optic design has shown to decrease the development of PCO due to the reduced space between the IOL and the posterior capsule. This may limit lens epithelial cell proliferation and metaplasia;5 the issue, however, remains controversial.6

The new “Apeiron” 42F Morcher IOL theoretically provides several advantages:- The hydrophilic acrylic material has shown to provide high uveal biocompatibility.4

- The square-edge optic design has shown to decrease the development of PCO due to the reduced space between the IOL and the posterior capsule. This may limit lens epithelial cell proliferation and metaplasia;5 the issue, however, remains controversial.6

The “Apeiron” 42F IOL is folded and inserted in the capsular bag through a CCI of 2.5 mm. The “Apeiron” 42F IOL is folded and inserted in the capsular bag through a CCI of 2.5 mm. Fig.3Fig.3

Fig.2

BBFig.8 AFig.8 A CC

Refractive astigmatism

Dio

pters 0,0

0,2

0,4

0,6

0,8

1,0

1,2

1,4

1.2

1.0

0.8

0.6

0.4

0.2

0.0pre-op post-op

0.5 ± 0.5 0.6 ± 0.4

dio

pte

rs

Best corrected visual acuity

Decim

al no

tation

0,0

0,2

0,4

0,6

0,8

1,0

1,2

1.0

0.8

0.6

0.4

0.2

0.0

0.37 ± 0.2

0.9 ± 0.1

pre-op post-op

Sn

elle

nli

nes

Manifest refractive spherical equivalent

Dio

pters

-5

-4

-3

-2

-1

0

1

2

33

2

1

0

-1

-2

-3

-4

-5pre-op post-op

-0.9 ± 2.0 -0.75 ± 0.4

dio

pte

rs

The mean distance between the endothelium and the IOL optic posterior surface 6 months post-op was 6.1 ± 0.5 mm (range 5.2 – 6.5).The mean distance between the endothelium and the IOL optic posterior surface 6 months post-op was 6.1 ± 0.5 mm (range 5.2 – 6.5).

Corneal endothelial cell density

cells/ mm

2

1600

1800

2000

2200

2400

2600

2800

pre-op post-op

2256 ± 303 2183 ± 321

ce

lls/m

m2