Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
Research Department P,EPORT No. L.016 10th April, 1940
Serial No. 1940/5
'Work can-:ied out by -H.D. Ellis
Drawings Nos. L.016.1 to L.016.3
Summary:
General·
TOTAL HARiVrONIC ANALYSER
In Control Rooms and elsewhere where audio frequency equipment is used it is hig.b..ly desrrable to provide apparatus for checking the distortion characteristics from time to time Q For this purpose a han-nonic analyser which measures the total harmonic is adequate, but it rmtst be stable, simple and quick to operate, and, should require no extra equipment beyond a standard. tony source. The instruInent described in this report ope:mtes on the bal~mced bridge principle and satisfies the above requirements" besides being portable and mains operated. .
The design of apparatus for routine di.stortion measurements
provides a somewhat different problem from that of a wave analyser. In
the first place: it is necessary to indicate only the total swn of the • I
harmoniccomponpnts generated in an amplifier to deterraine whether or
not the apparatus has cieteriorated in service. Secondly, it is
essential that the apparatus slmll be simple in use, though not
necessarily in conception, so that it may be operated by relatively
unskilled engineers. Third j it must be capable of rapid manipulation
if it is to be of.any real value in maintenance work. Finally, it
BBC R & 0
111111111111111111111111111111111111111111111111111111111I11 300008367 R
'------- --~;
- 2 -
, shou~d not depend for its accuracy on a very .stable and distortion-free
tone source, or upon the use of clli~bersome filters. Briefly in the
wave analyser- the essentia1 features are accuracy rmcl individual
selection of harrnonic components, whereas in the routine dis
- 3 -
it is particularly adaptable to the measurement of individual harmonic
components and does not lend itself to the measurement of the total sum.
Finally the balance mcthod was chosen. This method consists
in balancing out the fundamental tone and measuring what is left, and it
has several distinct ad.vrcntagcs. First, variations of frequency of the
tone source are quite unimportant since they affect both input and
output similarly cnd therefore do not disturb the h:Llcmce. Secondly
the test fr8quency may be chosen at will since there c:re no resonant or
filter circuits to restrict the frequencies which may be used. Thirdly,
distortion in the s'.~i.)ply tone only introduces a seconc, order error in
the result, as will be explained at a later stage.
The Balance Methoq,
The balance method consists in comparing the input to the
. apparatus under test with the output from it by means of a bridge
circuit. 'The difference between these two quantities is thEm a
measure of the dis.tortio:r: introduced by the apparatus. The test
circuit is. shown diagrarnma tic211yin Fig. 1. The signal levels at A
and B are arranged to be approximately equal for convenience; by means
of the attenuator shown, since this simplifies the design of the bridge
network.
The bridge netvlork itself must provide means for balanCing··
. out the fu..Yldamentc.l accurately, both inampli tude and also in phase, so
- 4 -
that the output of fundamental at the detector point C may be reduced
virtually to zero. Tho output at C is fed via a calibratedattenuator
and an amplifier to the indicating meter, which must have a square law
character~istic if the true r.m. s. total of all the harmonic components
is to be recorded.
Suppose the source voltage at A is E and the voltage
appearing at B is E + E2 -+ E3 +....... representing thefundarnental together with the various hannonics generated. If the bridge is
properly balanced the output at C will be aE2 + aE3 -1- •••• , a being the
loss ill the bridge, and the calibr.q.ted attenuator may be adjusted until
a suitable indication is shovrn on the meter. This indication,
therefore, gives B.. measure of the r.m.s. voltage at C due to the
If the input to the
bridge from B is removed, the output at C will be aE and the
attenuator may be re-set to obtain the same indication on the meter as
before. The two settings of the. attenuator, therefore, give
iID~ediately a figure for the ratio Le. the E
percentage total distortion.
The Bridge Circuit
The balanCing network must provide both amplitude and phase
adjustment for balancing out. the fundamental. It is highly desirable
- 5 -
that these two controls shall not be inter-dependent, Le. adjustment
of the c-w'1lplitude control shall not affect the phase s}lift Gnd vice-verso ..
, Unless this is achieved the process of balancing becomes laborious or
even exceedingly difficult, according to the degree of inter-dependence 0
By i::-ltroducing sufficient ettenuation the phase shift may be rendered
almost independent of amplitude cdjustmont. The BIllpli tude balance cells
for no comment save that it covers a mngs of 3 db. and there:fore links
up ,with the 2 db, steps assumed for the attenuator T in Fig. 1. It is
exceedingly difficult, however, to alter pho.se without altering amplit1J,de
and efter mnny circuits haQ beon investigated, given prolonged trial anci
finally rejected, the circuit about to be described W1C.S evolv89- which
satisfied all demands.
- Besidos the condition of non-interdependence of controls
mentioned above, it is essential that sufficient phase shift nhould be
available -to match that introduced by the amplifier or chain of
apparatus lmder test. At low frequency this may be considerable and in
certllin special Cllses may have any value between 00 and 360°, The
circuit finally adopted allows any value to be matched betwoen OOand
1800 ; ve,lues between 1800 and 3600 are ce.tered for merely by reverse.l
of one input lirw to the bridge.
The basic phe.se' shifting circuit used is illustrated in
Fig. 2. Referring to Fig. 2(e.), tone is applied e.t AB and C is the
- 6 -
electrical centre point beb'men A and B. In Fig. 2(b), vrhich is a
vol tage vector diagrrun, the voltage across AB is flhovm by the vector
AB with C at the centre point •. Provided ha load is applied between C and
D the eurrent I which flows through RI also flows through X, so that
the voltage vectors representing the potential differences across
AD and DB will be at Tight angles ~1Sshovm. in addition their vector
SUlIl is equal to the vector AB. If the resistance RI is reduced the
relative magnituc1es of the two vectors AD and DB are varied accordingly,
while still maintaining their right-angled relationship as represented.
by ADI and DIB .. ,By simple geometry it follows that the locus of D is
a semicircle with centre at C so that the vector CD,which represents the
voltage betvleen C and D, remains constant in21nplitude but varies in
phase relative to AB. By varJing either RI or X, therefore, the
potential between C and D varies in phase relative to that across AB
over the whole range 0 0 to 1800 while maintc1'"ining co'nstant amplitude,
thereby satisfying all the demands for the circuit.
For use at audio frequency the capacity becomes too large to
be conveniently variable so that· the resistance is made the adjustable
arm. This can be reduced to zero but cannot be increased above a
fini te maximum, so that the full 180° cannot be cov~red on this one
control. A similar circuit is therefore included in the other input
line, whieh covers only a limited range and serves both to supplement
.l
- 7 -
the restricted :range and to provide a fine control. Since the amplitude
remains independent6f the impedances X 2"ld HI this circuit may be used
in the line;carrying the. hc-'lrmonics, but in the case of phase shifting
circuits where the amplitude is Et fUllction of the circuit impedances,
and thereforeoftbi'3 frequency of the h.D.nnonic, the~le circuits would
have to be confined to the pure tone line only.
,Although the output should not in theory be loaded at 8.11,
it is found. that if fLn impedance not lE-SS than lOB. (Fig. 2(0)) be
connected across CD, the behaviour of the circuit is not seriously
upset. . '1'he whole. ctr~ui t arrangement 'will be seen in the diagr&m Fig. 3·
To maintain a reasonable lmv of phase shift versUEl rotation oftho
variable resistances in the· tvv'O phase shifting circuits, the capacities
are varied in four steps selected by a ganged switch, each position
covering a frequency band. The appropriate position is then chosen
according to the test frequency selected. In order to maintain complete
separation, of, the uT():ne" 'and "Test" circuits (A fmd D, Fig.l) the two
circuits are, foqt9twP separate shieldedwindings of a transformer, a
third windingpickirtg off the difference potential and feeding it to.,
the detector circ:c!.i,to .
Calibrated AttenucJ.tor and Amplifier.
These tYro i toms call for no special comment 0 The attenuator
is, in fact, ,apotentiometer giving 2 dbo por stop oxcept near tho
- 8 -
bottom, and is engraved direct in figures for percentage :b..armonic dovm
to 9.1%. The amplifier is str!l.ightforviaro. with two stages giving [-1
mHximwn gain of 66 db. from firot grid. to seconc~ anoc!.e. A gain control
is provided to standardise the·gain 2.ccoro.ingtb the level at which the
measurement is being taken.
Detector
In orde r that tJ:1e r.m. s. total harmonic shall be indicated
the eIetector must l)e truly square law. Theia - v characteristic of g
a normal valve gives a ver;/ close approximation to & square Lw! near
its bottoJ! bend, but it suffers from the disadvantage that the' maximum
signal which can be applied gives only a small rise of current compared
with the standing D.e. The latter can be backed off, but· all backing
off circuits are critical and apt to get ovt of adjustment~ and are
therefore to be avoided if possible. By using two valves with their .'
grids in push-pull andanod.es in' parallel, the oper&ting point may be
much closer to' the bottom bond and a true parabolic characteristic
obtained. The standing D.C. may thon:;fore be much reduced ana. the
, .
increase of current with applied signa,l bccoJT.l8s much "larger by comparison,
see Fig. 4 (a) and (b). Two AC/SP3valves 'are used for uniformity here,
as in the remainder of the apparatus'. lm adjustment is provided to . set
the standing D~C. in this stage to ensure'that the v&1VGS are operating
on the correct part of their cho.racteristics, b.nd· a pre-set bal~mcing
- 9 -
control is also providedl:
The s~nsitivity,of;.- the detector is sl1ch that the standing
current (0.5rfLAj' is doubled when a signal level of --16 d1:). ,is appiied to
the primarybf the'inputtransformero
Operation
The sequence of ,events in the operation of the analyser is as
follows.
(1) Having pwitched on the SUP1)lies and allowed time to heat up
(say 1 to 2 minutes) set the calibrated attenuator to CD, tum the gain
control down to minimumJ and set the standing feed of the detector to
the liSet" mark by means of the zero controL
(2) Plug l1p tM.tone source, attenu[,tor, amplifier under test
and dummy, load and conne ct the t! Tone" ci rcui t ,of the, analyse I' ,to ,the
point A, Fig. 1 and the "Test" circuit to point B.
, (3) Set the, output level of the tone Source to the output : ': ,"., .
'level at which the amplifier under test is to be measured and set the 'o' ,"; I". ' ,
calibrated attenuatorto the stud marked IIFII • Break the IITesMline " '" - .
to the analyser by pressing the key marked IITest offlland bring:up the /' > • • • -.., ;
'gain till the meter needle reaches the '0' mark on the scale •. :Break
the "Tone" line and adjust T, Fig .1, until the meter reads as nearly
10' as pos8ible.
(4). Adjust the coarse phCCse shift knob when the meter should
- 10 -
pass through a minimLm. If it does not, 'reverse one of the lines to
the analyse r. Continue to balance to a minimwnbysuccesslve
adjustments on the fine amplitude and phase controls~ increasing the
sensitivity as required by increasing the ce.librated nttenuator only.
(5) HaviY'.g obtained the best balance possible, retu.rn the
calibrated attenuD,tor to "FfI fild check the level of fundamental by
pressing the "Test off" key and adjust the gain control if necessary
to bring the needle to '0' . l1elease the,key and turn up the calibrated ~~.,,.,(:~,,!"Iit
potentiometer until the ',same reading is again obtained on the meter,
when the percentage distortion may be read off the scale. The meter is
calibrated in db. above and below the '0' mark, so tha.t if the correct
setting of the calibratedattenuatorlies between two studs the ,
percenta.ge distortion may be estimated with fair accuracy.
Accuracy
A jack isp:tovided so -thi3.t phones may be plugged to the output
of the detector amplifier and the fundamental balanced out by ear.
This facility is particularly-valuable vlhen distortion greater than 10% \
is being measured, when it will be found that the minimwnis difficult
to determine on the 'meter and maybe much more accurately determined
with the aid of phones.
If there is a negligible amount of phe.se shift in the apparatus
under test, d,istortion of the supply tone is unimportant since it is
'.
- 11 -
automc,tically balanced out? In P..ny c~se the error so introduced is .!,- . \ ..
only of the second or6er.. Some idea of the magnitude. of the error may
be seen from tlfefollOi~ingexample. If a perfectly linear ampl:i.fier is
under test which,)101uever, shifts the phas.e of the fundamentGl relative
to the secondh8.:rn1onic by 30°, Le. shifts the funclamental 50° and the
harmonic 20 0 , for example, then the hannonic indicated will be one half
that present irithe tone source.
The frequency characteristic of the detector amplifier is
sensibly fiat from 50 to 10.1000 cycJ,es/sec. and fcills off slowly above
this,. which ensures tho.t all harmonics of importance are recorded.
Application
.. The input' impedance to each side of the bridge network is
approximately 2.,000 or.ms up.der normal conditions " At the extreme limit
of the phase shift·coIltrol' the ,imped",,-nce of the tone branch may fgll to
about 750 obms but this will very rarely occur.
The apparatus will operate over e range of inputs from ..,.10 to
+30 db. whichirvill c:over most requirements in service and, as mentioned
earlier, Dny value of phase shift may be covered. Any test frequency ;,' ," ." .:--
maybe employedbetwe'en' 25 and 8,000 . cycles/sec. provided the ganged
frequency selEietionsYfi ten is set to the appropriatepcisition. There
'are four positionscQvering the bq.nds25. - 100, 100 - 450 J 450 - 2,000, ,
and 2,000-8,000 cyclos/sec.
1
- 12 -
Each input line is completely separete from the other and is
approximately balanced about earth, so that the apparatus may boused
on balanced, unbalanced 'or ea:rthed lines at will.
Finally~ in cases where it is necessary to t&ke a meEcsurement ~ . . .. . , . : '
y,hen the pure output from the tone soure') is not av&ilab1e; the analyser
may still be used in conjunction with Cl. 10wp';:Lss filter. The necess[lr"J
pure tone is .obtained by tappi:'.1g off from the output of the apparatus
under test and passing it. through this filter. This method is
particularly valuable, for example, in order to meas~ire the distortion
pret1ent in recording systems, in fact, this analyse I' is the only
instrument on which any reliable figures 11flve been obtained. The
usual difficulty is the variation in frequenc:r of the fundamental tone,
but since this apparatus is unaffected by such viHiations of frequency
it is ideal for this purpose. ,
Existing Test Apparatus
Routine distortion tests are ~)t present tnade at transmitters
on the General Radio or Marconi-Ekco Noise and Distortion Measuring
apparatus. This apparatus uses the elimination method of remOving the .' .
fundamental and incorporates a high-pass filter. rts 'use is therefore
confined to tests at 400 cycles/sec. arid it requires El. very pure source
of tone. A further disadvantage is trl8.t the measuring circuit is not
square law, so that if two or more harmonics of approxin1atelysimi1ar
- 13 -
amplitude are present the indication is not the true r.m.f'. sum.
This error is probably not. very serious.
As well as being suitable for all Control Room measurements)
the harmonic analyser described in this report J1'l[;,Y be used to replnce
the General Raclio Noise and .Dis-bortion Meter at tn,nsmi tters) for
measuring distortion at any audio frequency Rt will ccnd for mecsuring
hum and background noise. The indications give t~:'U3 r.m.s. velues J
E!.nd t~le operation is equally slmple. It vrould be usec~ in conjunction
with the Modu18tion~vloni to I' 0
\4- ";
C ---
TONE
~_~QCE
I I L ___ , _____ J
/ /
POSITION OF ATTENUt>.IOQ IF APPARATU5 UNDER TEST HAS GAIN LESS THAN UNITY .
FIG.L
,----- -APPAQA"TU5
UNDea TEST
BRIDGE NElWORK
l
c
1 CA.LI8C2ATED J
ATTENUATOQ
~P J -r ,,
- -_.-•
(5 - 5 - 40 .
A
B
(a.)
SING'.E VALVe:
FIG.4 (a)
D
Io.
FIG.Z.
I Ia.
1215E DUE TO ::SIGNAL
Ys-rANOING DC
A
( b)
I2ESOL..IAN-r C.Hbd:ll>.CTERISTIC .
I
lIa.
R ISE D U E 1 0 T 51~NAL
,/
I ~ I
:--r------+-l SIAND ING
J_ OC
,....- - --MA,)( . SIGNAL -- - . I
TWO VAL.Ves.
F=IG.4(b)
B.B.C. RESEARCH DEPT.
TOTA..L HARMONIC ANALYSEr
>-." ."
~
£;: ~
r
I l> 70 ~ o z -(1 p Z po
~ (j) m 1;]
0 ,., ~
~
• 1J O m .- \J
~ ~ . - ~ W
TO E
~ 0'
n
~ m ~ m > ?D () :r 0 m ~ ~
O·f F SW f.
°2j5 9 I G A NGED TO
5W.2
TEST OF F
0-2.,.,.,A / /')-!()?Y1~
r----+--------r-+~/ 0--5 v 0-500v
12
20000
AC SP3 \
~ o o 52
"-IEON fOO".
+300,
HT
r-----~----------_o4y
~_+-4~----------------+_~----------------------------------_i~----+---~VVVV---------U6v
~ D:::NO~ES MI'--L A""~.~E'TER 5f-H)",T. ;:;> lE ~ :r,::~ /r)L'f.A"'''''E'R I2ESIS-ANCE
F IG . 3 .
-------------------------------------------------------~-