98
RESL CUSTOMER EXPORT CONTROL AGREEMENT It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to conduct business in accordance with all applicable U.S. export control laws and regulations. It is also RESL’s policy that its Customers comply with U.S. export control laws and regulations. Therefore, Customer agrees to the following: 1. Because products, technical data, and technical assistance (i.e., services) provided to Customer by RESL may be subject to U.S. export control laws and regulations, (i) transactions with certain persons and companies and (ii) the export or reexport of certain types and levels of products, technical data, and services are prohibited or restricted. 2. Customer acknowledges that it is responsible for its own compliance with U.S. export control laws and regulations. Customer further agrees that it assumes the responsibility to obtain all necessary U.S. export licenses or other U.S. governmental authorizations, as well as all liability for the failure to do so. 3. Customer acknowledges that export control requirements may change and that the export or reexport of RESL products, technical data, and services without an export license or other appropriate governmental authorization may result in criminal and/or civil liability. 4. The obligations and requirements described herein shall survive the expiration or termination of any agreement or contract between RESL and Customer.

RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

Page 1: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

RESL CUSTOMER EXPORT CONTROL AGREEMENT It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to conduct business in accordance with all applicable U.S. export control laws and regulations. It is also RESL’s policy that its Customers comply with U.S. export control laws and regulations. Therefore, Customer agrees to the following: 1. Because products, technical data, and technical assistance (i.e., services) provided to

Customer by RESL may be subject to U.S. export control laws and regulations, (i) transactions with certain persons and companies and (ii) the export or reexport of certain types and levels of products, technical data, and services are prohibited or restricted.

2. Customer acknowledges that it is responsible for its own compliance with U.S. export control laws and regulations. Customer further agrees that it assumes the responsibility to obtain all necessary U.S. export licenses or other U.S. governmental authorizations, as well as all liability for the failure to do so.

3. Customer acknowledges that export control requirements may change and that the export or

reexport of RESL products, technical data, and services without an export license or other appropriate governmental authorization may result in criminal and/or civil liability.

4. The obligations and requirements described herein shall survive the expiration or termination

of any agreement or contract between RESL and Customer.

Page 2: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

GrF33 Participating LaboratoriesLab Code Lab Name Matrix Code

ADEM01 Alabama Department of Environmental Management GrF

ADFC99 Abu Dhabi Quality and Conformity Council‐Central Testing Lab GrF

AFOH01 USAFSAM/OEA GrF

ANTE01 ALS Environmental ‐ Fort Collins, CO GrF

ARGO01 Idaho National Laboratory GrF

ARPL01 Analytical Support Operations ‐ Radiochemical Processing Lab GrF

ARSL01 ARS International, LLC GrF

ASUK99 NAS Radiochemistry Group GrF

AY1201 Consolidated Nuclear Security, LLC, Analytical Chemistry Organization Laboratory GrF

CDHS01 California Department of Public Health GrF

CESL01 Lawrence Livermore National Laboratory ‐ EMRL GrF

CHMH01 222‐S Laboratory GrF

CMRC01 Carlsbad Environmental Monitoring and Research Center GrF

CRCE99 Centre for Radiation, Chemical and Environmental Hazards GrF

DEHS01 Department of Environmental Health & Safety GrF

DINL99 Departamento Ingeniería Nuclear y Mecánica de Fluidos GrF

DPHE01 Colorado Dept. of Public Health & Env. / Laboratory Services Div. GrF

ELCS01 Energy Labs ‐ College Station GrF

ELIW01 Energy Laboratories, Inc. GrF

ERCL01 Washington State Public Health Laboratories GrF

EULC01 EnergySolutions, LLC GrF

FDHE01 Florida Dept of Health Environmental Laboratory GrF

FDOH01 Florida Dept. of Health, Mobile Environmental Radiological Lab GrF

GENE01 GEL Laboratories, LLC GrF

GPCL01 Georgia Power Company Environmental Laboratory GrF

HECR01 SC Dept. Health and Environmental Control Radiological Laboratory GrF

HERR01 Washington Closure Hanford GrF

HPAL01 Los Alamos National Laboratory GrF

ISUP01 ISU ‐ Department of Physics/Health Physics/EAL GrF

JAEC99 Research Laboratories and Information Directorate GrF

KSCF01 Kennedy Space Center, HP Laboratory GrF

LAEC99 Lebanese Atomic Energy Commission ‐ Environmental Radiation GrF

LDRA99 Laboratori de Radiologia Ambiental‐Universitat de Barcelona GrF

LERL99 LARUEX Environmental Radioactivity Laboratory GrF

LGCL99 LGC Ltd GrF

LOCK03 Advanced Test Reactor (ATR) Complex Radioanalytical Laboratory GrF

MAHP59 Materials Assessment and Health Physics Group GrF

MART03 Radioactive Material Analysis Laboratory GrF

NARL02 USEPA ‐ NAREL ‐ MERL GrF

NESI01 B&W Technical Services‐Radioisotope & Analytical Chemistry Laboratory GrF

NJDH01 New Jersey Dept. of Health, PHEL, ECLS GrF

NOCS99 National Oceanography Centre, Southampton GrF

NRLL99 Environmental Radioactivity ‐ National Centre for Radiation Science GrF

ODHL01 Ohio Department of Health Laboratory GrF

ORIS01 ORISE/IEAV GrF

OTLI01 Outreach Laboratory, a Division of ESC Lab Sciences GrF

QUAN01 TestAmerica St. Louis GrF

Page 3: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

QUAN03 TestAmerica Laboratories ‐ Richland GrF

RAVR99 Radiactividad Ambiental y Vigilancia Radiologica GrF

RPSC01 Radiation Protection Service GrF

SANC99 RadioAnalysis, South Africa Nuclear Energy Corp. GrF

SEML01 SRS Environmental Monitoring Laboratory GrF

SOUT01 Southwest Research Institute GrF

SRPD01 Sandia National Laboratories, Radiation Protection Sample Diagnostics GrF

TDHL01 Texas Department of State Health Services Laboratory GrF

TELE01 TELEDYNE BROWN ENGINEERING ‐ ENVIRONMENTAL SERVICES GrF

TELE02 ATI Environmental, Inc., Midwest Lab GrF

TMAO01 EBERLINE SERVICES OAK RIDGE LABORATORY GrF

TNUT01 St. Louis USACE FUSRAP Laboratory GrF

UPVL99 Lab Radiactividad Ambiental.  Univ. Politecnica Valencia GrF

URSL01 UNLV Radioanalytical Services Laboratory GrF

USED99 National Center for Nuclear Energy, Sciences and Techniques GrF

WEST04 PACE ANALYTICAL SERVICES, PITTSBURGH GrF

WIPP01 WIPP Laboratories GrF

WSHL01 Wisconsin State Laboratory of Hygiene GrF

WSTP99 Cavendish Nuclear Limited GrF

WVDP01 WVDP Environmental Laboratory GrF

WVPL01 WVDP Radiation Protection Lab GrF

ZIOS99 ZVD Institute of Occupational Safety GrF

Page 4: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

Laboratories Not Reporting GrF33

Lab Code Lab Name Matrix Code

DRMG01 CNS Pantex ‐ D&RMG GrF

FMEC99 Foods and Water Laboratories Center GrF

HCAL01 Lawrence Livermore National Laboratory GrF

KDHE01 Kansas Dept. of Health & Environment GrF

MDPH01 MDPH‐Radiation Control Program GrF

NARL01 National Analytical Radiation Environmental Laboratory GrF

NCNS99 Mansour OUESLATI ‐ Radioanalytical Laboratory GrF

RSAL01 RSA Laboratories, Inc. GrF

Page 5: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

Study Reference ValuesMAPEP-15-GrF33 Radiological Reference Date: 08/01/2015

Analyte RefValue

RefUnc

Radiological Units:(Bq/sample)

Gross alpha 0.90 0.03Gross beta 1.56 0.05

Printed 1/06/2016

Page 6: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

Sample Statistical SummaryMAPEP-15-GrF33 Radiological Reference Date: 08/01/2015

Analyte T(1) A(2) Grand(3)Mean

StdDev

RefValue

RefUnc

AcceptanceRange

Radiological Units: (Bq/sample)Gross alpha 68 56 0.70 0.26 0.90 0.03 0.27 - 1.53 Gross beta 69 67 1.52 0.19 1.56 0.05 0.78 - 2.34

Note: (1) T = Total number of laboratories reporting analyte.(2) A = Number of laboratories with 'Acceptable' performance.(3) Mean excludes values derived as total metals and values indicated as "Not Acceptable".

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations(Result/Uncertainty > 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, does not include zero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standarddeviations (Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty attwo standard deviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations(Result/Uncertainty > 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, does not include zero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standarddeviations (Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty attwo standard deviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/06/2016

Page 7: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

Flag Summary ReportMAPEP-15-GrF33

Radiological

Analyte A W RW NGross alpha 56 12Gross beta 67 2

Printed 1/06/2016

Page 8: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ADEM01) Alabama Department of Environmental ManagementMontgomery LaboratoryMontgomery, AL 36110

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0001N -0.0922Gross alpha -89.80.90 0.27 1.53 N0.0001N -0.4218Gross beta -73.01.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 9: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ADFC99) Abu Dhabi Quality and Conformity Council-Central Testing LabRadiation LabAbu Dhabi, - -

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.3A -0.84Gross alpha -6.70.90 0.27 1.53 N0.32A -1.25Gross beta -19.91.56 0.78 2.34 W

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 10: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(AFOH01) USAFSAM/OEA2510 Fifth Street, Area BWright-Patterson AFB, OH 45433-7913

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.11A -0.71Gross alpha -21.10.90 0.27 1.53 W0.10A -1.74Gross beta 11.51.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 11: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ANTE01) ALS Environmental - Fort Collins, CO225 Commerce DriveFort Collins, CO 80524-1416

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.054A -0.602Gross alpha -33.10.90 0.27 1.53 A0.10A -1.26Gross beta -19.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 12: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ARGO01) Idaho National LaboratoryINL, Materials and Fuels ComplexIdaho Falls, ID 83415

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.17A -0.89Gross alpha -1.10.90 0.27 1.53 W0.27A -1.37Gross beta -12.21.56 0.78 2.34 W

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 13: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ARPL01) Analytical Support Operations - Radiochemical Processing LabPO Box 999Richland, WA 99354

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.022N -0.215Gross alpha -76.10.90 0.27 1.53 A0.045A -1.34Gross beta -14.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 14: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ARSL01) ARS International, LLC2609 North River RoadPort Allen, LA 70767

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.039A -0.300Gross alpha -66.70.90 0.27 1.53 A0.190A -1.54Gross beta -1.31.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 15: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ASUK99) NAS Radiochemistry GroupA38.1 AWE,Reading, Berkshire RG7 4PR

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.038A -1.091Gross alpha 21.20.90 0.27 1.53 A0.022A -1.919Gross beta 23.01.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 16: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(AY1201) Consolidated Nuclear Security, LLC, Analytical Chemistry Organization LaboratoryY12, NSC, Bldg. 9995, Rm 142Oak Ridge, TN 37831-8189

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.065A -0.938Gross alpha 4.20.90 0.27 1.53 A0.078A -1.77Gross beta 13.51.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 17: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(CDHS01) California Department of Public HealthDrinking Water & Radiation Lab.Richmond, CA 94804-6403

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0143A -0.307Gross alpha -65.90.90 0.27 1.53 A0.0183A -1.57Gross beta 0.61.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 18: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(CESL01) Lawrence Livermore National Laboratory - EMRL7000 East AvenueLivermore, CA 94551

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

2.69E-03N -2.49E-01Gross alpha -72.30.90 0.27 1.53 N6.08E-03A -1.82E+00Gross beta 16.71.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 19: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(CHMH01) 222-S Laboratory2355 Stevens DriveRichland, WA 99354

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

7.56E-02A -7.287E-01Gross alpha -19.00.90 0.27 1.53 A6.70E-02A -1.178E+00Gross beta -24.51.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 20: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(CMRC01) Carlsbad Environmental Monitoring and Research Center1400 University Dr.Carlsbad, NM 88220

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.018A -0.751Gross alpha -16.60.90 0.27 1.53 A0.15A -1.64Gross beta 5.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 21: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(CRCE99) Centre for Radiation, Chemical and Environmental HazardsPublic Health EnglandDidcot, OXON OX11 0RQ

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.02A -0.41Gross alpha -54.40.90 0.27 1.53 A0.11A -2.05Gross beta 31.41.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 22: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(DEHS01) Department of Environmental Health & SafetyNorth Carolina State Univ.Raleigh, NC 27695-8007

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.04A -0.84Gross alpha -6.70.90 0.27 1.53 A0.06A -1.40Gross beta -10.31.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 23: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(DINL99) Departamento Ingeniería Nuclear y Mecánica de FluidosAlameda de Urquijo s/nBilbao, Vizcaya 48013

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.044A -1.11Gross alpha 23.30.90 0.27 1.53 A0.041A -1.56Gross beta 0.01.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 24: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(DPHE01) Colorado Dept. of Public Health & Env. / Laboratory Services Div.8100 Lowry BlvdDenver, CO 80230

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.04A -0.44Gross alpha -51.10.90 0.27 1.53 A0.04A -1.45Gross beta -7.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 25: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ELCS01) Energy Labs - College Station415 Graham RDCollege Station, Tx 77845

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.18A -1.13Gross alpha 25.60.90 0.27 1.53 W0.06A -1.58Gross beta 1.31.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 26: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ELIW01) Energy Laboratories, Inc.2393 Salt Creek HWYCasper, Wy 82601

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

.18A -0.71Gross alpha -21.10.90 0.27 1.53 W

.18A -1.2Gross beta -23.11.56 0.78 2.34 ARadiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 27: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ERCL01) Washington State Public Health Laboratories1610 N.E. 150th StreetShoreline, WA 98155-9701

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

.023N -.234Gross alpha -74.00.90 0.27 1.53 A.03A -1.52Gross beta -2.61.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 28: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(EULC01) EnergySolutions, LLCI-80, Exit 49Clive, UT 84029

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

.04N -.25925Gross alpha -71.20.90 0.27 1.53 W.3A -1.6338Gross beta 4.71.56 0.78 2.34 W

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 29: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(FDHE01) Florida Dept of Health Environmental LaboratoryPO Box 680069Orlando, FL 32868-0069

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.05A -0.83Gross alpha -7.80.90 0.27 1.53 A0.07A -1.54Gross beta -1.31.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 30: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(FDOH01) Florida Dept. of Health, Mobile Environmental Radiological LabPO Box 680069Orlando, FL 32868-0069

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.08A -0.87Gross alpha -3.30.90 0.27 1.53 A0.05A -1.51Gross beta -3.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 31: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(GENE01) GEL Laboratories, LLC2040 Savage RoadCharleston, SC 29407

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.130A -0.999Gross alpha 11.00.90 0.27 1.53 A0.0591A -1.57Gross beta 0.61.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 32: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(GPCL01) Georgia Power Company Environmental Laboratory2480 Maner RoadAtlanta, GA 30339

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0346N -0.0890Gross alpha -90.10.90 0.27 1.53 N0.0283A -1.61Gross beta 3.21.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 33: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(HECR01) SC Dept. Health and Environmental Control Radiological Laboratory8231 Parklane RoadColumbia, SC 29223

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.039N -0.247Gross alpha -72.60.90 0.27 1.53 W0.07A -1.51Gross beta -3.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 34: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(HERR01) Washington Closure HanfordRadiological Counting FacilityRichland, WA 99352

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

.03N -.11Gross alpha -87.80.90 0.27 1.53 W

.08A -1.57Gross beta 0.61.56 0.78 2.34 ARadiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 35: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(HPAL01) Los Alamos National LaboratoryMilan Gadd, MS G761Los Alamos, NM 87545-1663

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

.09A -.97Gross alpha 7.80.90 0.27 1.53 A

.02A -1.6Gross beta 2.61.56 0.78 2.34 NRadiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 36: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ISUP01) ISU - Department of Physics/Health Physics/EAL785 S. 8th Ave, Rm 120Pocatello, ID 83209-8106

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.01A -0.43Gross alpha -52.20.90 0.27 1.53 A0.01A -1.27Gross beta -18.61.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 37: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(JAEC99) Research Laboratories and Information DirectorateJordan Atomic Energy CommissionAmman, Jordan 70 (11934)

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.054A -0.534Gross alpha -40.70.90 0.27 1.53 A0.099A -1.873Gross beta 20.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 38: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(KSCF01) Kennedy Space Center, HP LaboratoryIHA-023Kennedy Space Center, FL 32899

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0098N -0.2313Gross alpha -74.30.90 0.27 1.53 A0.0513A -1.533Gross beta -1.71.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 39: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(LAEC99) Lebanese Atomic Energy Commission - Environmental RadiationNational Council for Scientific ResearchBeirut, Beirut 1107 2260

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.06A -0.74Gross alpha -17.80.90 0.27 1.53 A0.06A -1.41Gross beta -9.61.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 40: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(LDRA99) Laboratori de Radiologia Ambiental-Universitat de BarcelonaFacultat de QuimicaBarcelona, Catalunya 08028

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

1.9N -26.4Gross alpha 2833.30.90 0.27 1.53 A1.6N -45.1Gross beta 2791.01.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 41: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(LERL99) LARUEX Environmental Radioactivity LaboratoryDpt. Applied Physics, Faculty of Vet. SciencesCaceres, Caceres 10003

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.056A -0.430Gross alpha -52.20.90 0.27 1.53 A0.12A -1.26Gross beta -19.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 42: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(LGCL99) LGC LtdQueens RoadTeddington, Middlesex TW11 0LY

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.11A -0.75Gross alpha -16.70.90 0.27 1.53 A0.09A -1.15Gross beta -26.31.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 43: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(LOCK03) Advanced Test Reactor (ATR) Complex Radioanalytical LaboratoryINL/Battelle Energy Alliance, LLCIdaho Falls, ID 83415-7111

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.022A -0.304Gross alpha -66.20.90 0.27 1.53 A0.08A -1.70Gross beta 9.01.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 44: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(MAHP59) Materials Assessment and Health Physics GroupImperial College Reactor CentreAscot, Berkshire SL5 7TE

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.10A -0.95Gross alpha 5.60.90 0.27 1.53 A0.12A -1.45Gross beta -7.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 45: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(MART03) Radioactive Material Analysis LaboratoryORNLOak Ridge, TN 37830

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.023N -0.182Gross alpha -79.80.90 0.27 1.53 A0.045A -1.47Gross beta -5.81.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 46: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(NARL02) USEPA - NAREL - MERL540 S Morris AveMontgomery, AL 36115

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.011N -0.246Gross alpha -72.70.90 0.27 1.53 A0.022A -1.39Gross beta -10.91.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 47: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(NESI01) B&W Technical Services-Radioisotope & Analytical Chemistry LaboratoryLynchburg Technology CenterLynchburg, VA 24504-5447

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.032A -0.600Gross alpha -33.30.90 0.27 1.53 A0.038A -1.61Gross beta 3.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 48: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(NJDH01) New Jersey Dept. of Health, PHEL, ECLS3 Schwarzkopf DriveWest Trenton, NJ 08628

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.014A -0.8954Gross alpha -0.50.90 0.27 1.53 N0.016A -1.5466Gross beta -0.91.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 49: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(NOCS99) National Oceanography Centre, SouthamptonGAU-RadioanalyticalSouthampton, Hampshire SO14 3ZH

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.02A -0.47Gross alpha -47.80.90 0.27 1.53 A0.03A -1.84Gross beta 17.91.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 50: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(NRLL99) Environmental Radioactivity - National Centre for Radiation SciencePO Box 25099Christchurch, Christchurch 8144

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.022A -0.850Gross alpha -5.60.90 0.27 1.53 A0.029A -1.413Gross beta -9.41.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 51: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ODHL01) Ohio Department of Health Laboratory8995 E Main StreetReynoldsburg, OH 43068

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.041A -0.734Gross alpha -18.40.90 0.27 1.53 A0.021A -1.439Gross beta -7.81.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 52: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ORIS01) ORISE/IEAVPO Box 117Oak Ridge, TN 37831-0117

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.019A -0.586Gross alpha -34.90.90 0.27 1.53 A0.018A -1.569Gross beta 0.61.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 53: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(OTLI01) Outreach Laboratory, a Division of ESC Lab Sciences311 N. AspenBroken Arrow, OK 74012-

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.013A -0.272Gross alpha -69.80.90 0.27 1.53 A0.024A -1.61Gross beta 3.21.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 54: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(QUAN01) TestAmerica St. Louis13715 Rider Trail NorthEarth City, MO 63045-1205

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0855A -1.23Gross alpha 36.70.90 0.27 1.53 A0.0734A -1.41Gross beta -9.61.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 55: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(QUAN03) TestAmerica Laboratories - Richland2800 GEORGE WASHINGTON WAYRICHLAND, WA 99354-

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.103A -0.779Gross alpha -13.40.90 0.27 1.53 A0.112A -1.57Gross beta 0.61.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 56: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(RAVR99) Radiactividad Ambiental y Vigilancia RadiologicaCIEMAT (Ed 70 P2 D11)Madrid, Madrid 28040

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.014A -0.624Gross alpha -30.70.90 0.27 1.53 A0.070A -1.483Gross beta -4.91.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 57: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(RPSC01) Radiation Protection ServiceOntario Ministry of LabourWeston, Ontario M9P 3T1

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

-NRGross alpha 0.90 0.27 1.530.19A -1.94Gross beta 24.41.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 58: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(SANC99) RadioAnalysis, South Africa Nuclear Energy Corp.Sample Receipt Gate 1Pretoria, Gauteng 0001

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.023A -0.283Gross alpha -68.60.90 0.27 1.53 A0.05A -1.45Gross beta -7.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 59: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(SEML01) SRS Environmental Monitoring LaboratoryBldg 735-BAiken, SC 29808

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.079A -0.78Gross alpha -13.30.90 0.27 1.53 A0.057A -1.61Gross beta 3.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 60: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(SOUT01) Southwest Research Institute6220 Culebra Rd.San Antonio, TX 78228-0510

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0752A -1.07Gross alpha 18.90.90 0.27 1.53 A0.0952A -1.57Gross beta 0.61.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 61: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(SRPD01) Sandia National Laboratories, Radiation Protection Sample DiagnosticsPO Box 5800, MS1103Albuquerque, NM 87185-1103

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.015A -1.109Gross alpha 23.20.90 0.27 1.53 N0.033A -1.417Gross beta -9.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 62: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(TDHL01) Texas Department of State Health Services Laboratory1100 W 49th StreetAustin, TX 78756

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.020A -0.498Gross alpha -44.70.90 0.27 1.53 A0.025A -1.469Gross beta -5.81.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 63: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(TELE01) TELEDYNE BROWN ENGINEERING - ENVIRONMENTAL SERVICES2508 Quality LaneKnoxville, TN 37931-6819

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.039A -0.497Gross alpha -44.80.90 0.27 1.53 A0.068A -1.34Gross beta -14.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 64: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(TELE02) ATI Environmental, Inc., Midwest Lab700 Landwehr RoadNorthbrook, IL 60062-

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.04A -0.30Gross alpha -66.70.90 0.27 1.53 A0.09A -1.85Gross beta 18.61.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 65: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(TMAO01) EBERLINE SERVICES OAK RIDGE LABORATORY601 SCARBORO RDOAK RIDGE, TN 37830-

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.025A -0.43Gross alpha -52.20.90 0.27 1.53 A0.044A -1.525Gross beta -2.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 66: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(TNUT01) St. Louis USACE FUSRAP Laboratory8945 LATTY AVEBERKELEY, MO 63134-

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

.0245A -.3409Gross alpha -62.10.90 0.27 1.53 A

.0432A -1.5390Gross beta -1.31.56 0.78 2.34 ARadiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 67: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(UPVL99) Lab Radiactividad Ambiental. Univ. Politecnica ValenciaCAMINO DE VERA, 14VALENCIA, VALENCIA 46022

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.04A -0.59Gross alpha -34.40.90 0.27 1.53 A0.02A -1.56Gross beta 0.01.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 68: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(URSL01) UNLV Radioanalytical Services LaboratoryCampus Mail Stop 453037Las Vegas, NV 89154-3037

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.04A -0.81Gross alpha -10.00.90 0.27 1.53 A0.07A -1.37Gross beta -12.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 69: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(USED99) National Center for Nuclear Energy, Sciences and TechniquesCNESTEN BP 1382RABAT, RABAT RP 10001

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0444A -1.27Gross alpha 41.10.90 0.27 1.53 A0.0463A -1.11Gross beta -28.81.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 70: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(WEST04) PACE ANALYTICAL SERVICES, PITTSBURGH1638 Roseytown RoadGreensburg, PA 15601

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0974A -0.518Gross alpha -42.40.90 0.27 1.53 W0.304A -1.69Gross beta 8.31.56 0.78 2.34 W

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 71: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(WIPP01) WIPP Laboratories1400 University DriveCarlsbad, NM 88220

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

6.50e-02A -6.97e-01Gross alpha -22.60.90 0.27 1.53 A4.48e-02A -1.31e+00Gross beta -16.01.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 72: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(WSHL01) Wisconsin State Laboratory of Hygiene2601 Agriculture DriveMadison, WI 53718

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.053A -0.551Gross alpha -38.80.90 0.27 1.53 A0.034A -1.324Gross beta -15.11.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 73: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(WSTP99) Cavendish Nuclear LimitedGreeson CourtCumbria, UK CA24 3HZ

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.08A -0.88Gross alpha -2.20.90 0.27 1.53 A0.27A -1.72Gross beta 10.31.56 0.78 2.34 W

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 74: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(WVDP01) WVDP Environmental Laboratory10282 Rock Springs RoadWest Valley, NY 14171

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0473A -0.594Gross alpha -34.00.90 0.27 1.53 A0.0194A -1.48Gross beta -5.11.56 0.78 2.34 N

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 75: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(WVPL01) WVDP Radiation Protection Lab10282 Rock Springs RdWest Valley, NY 14171

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.0562A -0.889Gross alpha -1.20.90 0.27 1.53 A0.0323A -1.5253Gross beta -2.21.56 0.78 2.34 A

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 76: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

MAPEP-15-GrF33 Combined Report

Department of Energy RESL - 1955 Fremont Ave, MS4149 - Idaho Falls, ID 83415

Laboratory Results For MAPEP-15-GrF33(ZIOS99) ZVD Institute of Occupational SafetyChengdujska Cesta 25Ljubljana-Polje, Slovenia SI-1260

AnalyteUnc

ValueUncFlagResult

RefValue Flag

Bias(%)

AcceptanceRangeNotes

Units: (Bq/sample)Radiological

0.13A -0.57Gross alpha -36.70.90 0.27 1.53 W0.32A -1.54Gross beta -1.31.56 0.78 2.34 W

Radiological Reference Date: August 1, 2015

Gross Alpha Flags:

A = Result acceptable, Bias <= +/- 70% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 70% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Gross Beta Flags:

A = Result acceptable, Bias <= +/- 50% with a statistically positive result at two standard deviations (Result/Uncertainty >2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standard deviations, does not includezero).

N = Result not acceptable, Bias > +/- 50% or the reported result is not statistically positive at two standard deviations(Result/Uncertainty <= 2, i.e., the range encompassing the result, plus or minus the total uncertainty at two standarddeviations, includes zero).

Uncertainty Flags:

1) NOT ACCEPTABLE...........................RP<2%

2) ACCEPTABLE..................................2%<=RP<=15%

3) ACCEPTABLE WITH WARNING...........15%<RP<=30%

4) NOT ACCEPTABLE............................RP>30%

RP = Relative Precision

Printed 1/6/2016Issued 12/2/2015

Page 77: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

Gross alpha

MAPEP-15-GrF33

Notes:The chart shows only data points with values between -0.61 and 2.01 (± 5 Standard Deviations).

Page 78: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

Gross beta

MAPEP-15-GrF33

Notes:The chart shows only data points with values between 0.56 and 2.48 (± 5 Standard Deviations).

Page 79: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

The intent of the distribution graphs contained within this report is to graphically demonstrate to users how % Bias data within the current MAPEP Series appears when examined by matrix, by analyte, by method of sample preparation or by method of detection. The data points (small 'dots') plot the bias (or z-score for organic constituents) of each of the analytes in each of the matrices. The BLACK small vertical line inside the small rectangle at the top of the data distribution graph is the mean of the population of the bias shown for that analyte in the matrix.

The box plots summarize the distribution of points for each analyte. The ends of the box are the 25th and 75th quantiles. The difference between the quartiles is the interquartile range. The line across the middle of the box identifies the median bias value. Each box has lines, sometimes called whiskers, which extend from each end. The whiskers extend from the ends of the small box to the outermost data point that falls within the distances computed. Any potential outliers are represented as small black dots.

The box plot of the bias data points and the mean visually illustrate the breadth of the distribution and where any potential outliers in the distribution might lie. The moments for the distribution plot are provided below the Bias plot. In some cases, N becomes very small and thus developed statistics may not accurately reflect estimates of the population if N were a significantly larger value.

Possible Outliers Mean Confidence Diamond

25th Percentile 75th Percentile Interquartile Range

Page 80: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

���������������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

��������������������������������������������

��

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB'HWHFWLRQ *URVV�DOSKD%HWD�&RXQWLQJ�����SL�JDV�IORZ�SURSRUWLRQDO�FRXQWHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 81: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

���

���

��

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

���������������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

������������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB'HWHFWLRQ *URVV�DOSKD%HWD�&RXQWLQJ���OLTXLG�VFLQWLOODWLRQ�FRXQWHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 82: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

����

���

���

���

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������������������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������������

��

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB'HWHFWLRQ *URVV�DOSKD��*URVV$OSKD�%HWD�����SL�JDV�IORZ�SURSRUWLRQDO�FRXQWHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 83: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

����

����

����

����

����

����

����

����

����

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

�������������

���������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV$QDO\WHB'HWHFWLRQ *URVV�DOSKD��2WKHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 84: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

������

������������

��������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������������

��

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB'HWHFWLRQ *URVV�EHWD��%HWD&RXQWLQJ�����SL�JDV�IORZ�SURSRUWLRQDO�FRXQWHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 85: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

��

��

��

��

��

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

����������������

��������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB'HWHFWLRQ *URVV�EHWD%HWD�&RXQWLQJ���OLTXLG�VFLQWLOODWLRQ�FRXQWHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 86: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������

������������������

�����������������

������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������������

��

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB'HWHFWLRQ *URVV�EHWD��*URVV$OSKD�%HWD�����SL�JDV�IORZ�SURSRUWLRQDO�FRXQWHU

*U)�'LVWULEXWLRQ�E\�'HWHFWLRQ�0HWKRG

Page 87: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

���

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

�����������������������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

�������������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�DOSKD��$FLGGLVVROXWLRQ�E\�VWURQJ�$TXD�5HTLD��K\GURIOXRULF�DFLG��HWF�

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 88: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

���

��

��

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

�������������������������

��������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

����������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�DOSKD$FLG�OHDFKLQJ�ZLWKRXW�K\GURIOXRULF�DFLG

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 89: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

���

���

���

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

������������������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

�������������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�DOSKD��(3$�����5DGLRDFWLYLW\��*URVV�$OSKD�%HWD�6FUHHQLQJ��������������

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 90: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

����

���

���

���

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������

����������������

������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������������

��

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�DOSKD1R�SUHSDUDWLRQ���DQDO\]HG�DV�UHFHLYHG

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 91: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

���

���

���

���

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

���������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

��������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV$QDO\WHB0HWKRG *URVV�DOSKD��2WKHU

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 92: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

���

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

�������������������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

����������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�DOSKD��:HW�DVK��$FLG�GLJHVWLRQ���WKH�XVH�RI�R[LGL]HUV�WR�GHVWUR\�RUJDQLFV

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 93: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

��

��

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

�����������������������

�������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�EHWD��$FLGGLVVROXWLRQ�E\�VWURQJ�$TXD�5HTLD��K\GURIOXRULF�DFLG��HWF�

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 94: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

��

��

��

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

�������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�EHWD$FLG�OHDFKLQJ�ZLWKRXW�K\GURIOXRULF�DFLG

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 95: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

������������������

��������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������

������������������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�EHWD��(3$�����5DGLRDFWLYLW\��*URVV�$OSKD�%HWD�6FUHHQLQJ��������������

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 96: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

���

���

���

���

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������

������������������

������������

�������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������������������������������������������

��

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�EHWD1R�SUHSDUDWLRQ���DQDO\]HG�DV�UHFHLYHG

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 97: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

��

��

��

��

��

��

��

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV$QDO\WHB0HWKRG *URVV�EHWD��2WKHU

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG

Page 98: RESL CUSTOMER EXPORT CONTROL AGREEMENT · 2016-01-25 · RESL CUSTOMER EXPORT CONTROL AGREEMENT . It is the Radiological and Environmental Sciences Laboratory’s (RESL) policy to

����

����

����

����

����

����

����

����

�����������������������������������������������������

PD[LPXP

TXDUWLOHPHGLDQTXDUWLOH

PLQLPXP

��������������������������������������������

4XDQWLOHV

0HDQ6WG�'HY6WG�(UU�0HDQ8SSHU�����0HDQ/RZHU�����0HDQ1

���������

6XPPDU\�6WDWLVWLFV

%LDV

'LVWULEXWLRQV�$QDO\WHB0HWKRG *URVV�EHWD��:HW�DVK���$FLGGLJHVWLRQ���WKH�XVH�RI�R[LGL]HUV�WR�GHVWUR\�RUJDQLFV

*U)�'LVWULEXWLRQ�E\�3UHS�0HWKRG