Upload
brian-dellamora
View
53
Download
0
Embed Size (px)
Citation preview
Analysis of Material Choice and Geometry of Coronary StentsSara Beck, Josh Becker, Emily Burgess, Brian Della Mora, William Turri
MSE 365 Tuesday LabDepartment of Materials Science and Engineering, University of Michigan
Introduction
Metal Analysis
• Used Scanning Electron Microscopy (SEM) to visualize stents on amicroscopic scale
• Sample was mounted in Bakelite powder, ground, and polished to seemicrostructure
• Homogeneous microstructure observed• Etching needed for microstructure but not possible due to lack of supplies
• Composition of the BMS and DESanalyzed using EDAX and WD-XRF
• EDAX and WD-XRF were consistent• BMS and DES showed virtually the
same composition• WD-XRF data displayed was used to
identify the alloy choice as MP35N1
Component BMS (wt%) DES (wt%)Cr 19.9 19.6Co 31.0 30.8Ni 36.5 35.4
Mo 10.9 12.1
• Alloy has high strength, toughness, and corrosion resistance• Stents are wire drawn, plastically deformed into a sinusoidal shape, and
laser fused at certain junctions• Surface is then electropolished to ensure microscopic uniformity2
1 “Data Sheet: MP35N Alloy.” Specialty Steel Supply. 2008. Web.2 Loya, Mariana. Design and fabrication of advanced surface microstructures: surface modification of cardiovascular stent wires via RF plasma processing. Dissertation, University of California San Diego.Web.3 “Plastic Deformation During the Expansion of a Biomedical Stent.” Application Gallery. COMSOL Multiphysics. Web. 4 Ahmad, Hasan, et al. "Synthesis of Biocompatible Sterically-Stabilized Poly(2-(methacryloyloxy)ethyl Phosphorylcholine) Latexes via Dispersion Polymerization in Alcohol/Water Mixtures." Langmuir (2009): 11442-1449. Print.5 Udipi, K. “Development of a Novel Biocompatible Polymer system for Extended Drug Release in A Next-generation Drug-eluting Stent.” Journal of Biomedical Materials Research(2007): 1064-071. Web.6 “Indications, Safety, and Warnings.” Coronary Stent. Web.
Polymer AnalysisComputational Modeling
• Intended to run DSC and FT-IR to gather percent crystallinity, glasstransition, and functional groups
• Polymer coating was only 4.3 µm thick (not enough to run successfultests)
• A literature review was conducted to better understand properties of thepolymers and why they were chosen
• The two DES are the Endeavor and Resolute Integrity
• The coating needs to be both hydrophilic and hydrophobic to: Control drug release Maintain an interface with the blood vessel wall
• The release rate of Zotarolimus was optimized5
Cell Membrane
Hydrophilic Polymer
Hydrophobic Polymer & DrugStent
Summary
• The metal was identified as MP35N alloy processed by wire drawing• Diameter expansion comparison with theoretical values was consistent• Stent application increases circulation• BMS geometry can be simplified without losing functionality• COMSOL modeling shows importance of laser fusing in radial expansion• Gained an understanding of the polymer-drug system mechanism
Acknowledgements & References
Stent Polymer CoatingEndeavor Phosphorylcholine4
Resolute C19/C10/PVP5
• Polymers must be: Robust (rupture resistant) Biocompatible Hydrophilic
• Zotarolimus is a hydrophobic drugused to treat thrombosis andstenosis6
3
3.2
3.4
3.6
3.8
4
5 6 7 8 9 10 11 12 13 14 15 16
Inte
rnal
Dia
met
er (m
m)
Pressure (atm)
ExperimentalTheoretical
400
300
200
100
0
Von Mises Stress Distribution (MPa)
Future Recommendations• For testing of metal alloy and functional polymers, obtain bulk samples to
use for in situ material testing• Apply etchants to determine crystal structure and grain size• Model fluid flow more accurately using 3-D geometry• Determine radial expansion for BMS geometry
1.40
1.35
1.30
1.25
1.20
1.45
1.15
Flow Velocity (m/s)
• Heart stents are used to treat blood clots• Clots affect the flow of blood in vessels, leading to strokes or heart attacks• Blood flow modeled in COMSOL to compare clotted vessel to open vessel• When the blood clot is present, blood flow is significantly slower
Demonstrates decreased circulation• After stent implantation, circulation should return to normal
• Stents radially expand by applying an internalpressure to keep vessel open and alleviate clot
• The radial expansion of a stent was modeled3
Must plastically deform without failure Functionality of modeled geometry comparable
to BMS• Laser fusion:
Joins stent sections Reduces stress Decreases chance of stent failure
BMS Geometry
Special thanks to Michael Palazzolo at Medtronic for providing us with the stents and to Justin Scanlon for helping us set up experiments.
Images:"Coronary Stenting - Dr I.B.A. Menown MD FRCP Consultant Cardiologist." Coronary Stenting - Dr I.B.A. Menown MD FRCP Consultant Cardiologist. Web. Huang, Kenneth. "Mr. Heart." Future Science Leaders. 17 Jan. 2015. Web.BioLinx Polymer System. Biovisioning and Medtronic. Video.
Coronary stents are cylindrical instrumentsimplanted in blood vessels. These devices werefirst used in 1986 by Jacques Puel in France andhave since been optimized for the treatment ofthrombosis and stenosis. An in-depth study ofthe effects of material choice, geometry, andprocessing of bare metal stents (BMS) and twodifferent polymer coated drug-eluting stents (DES) made by Medtronic wascompleted.