RICH-MAR THERASOUND 3 SERIESULTRASOUND OPERATIONHANDBOOK AND MANUAL
Part # MN 2430Rev. I
Batch 001
CAUTION
This device is not designed to be connected with any electrical equipment unless manufactured and ap-proved by Rich-Mar.
NOTE: This includes whirpools and carbon electrodes NOT manufactured by Rich-Mar.
CAUTION: When using carbon electrodes with any Rich-Mar stimulator, a moistened interface (cloth orsponge) MUST be utilized between these electrodes and the patient to avoid skin irritation and/or electricalburns.
TABLE OF CONTENTS
Therasound 3 Series Warranty..........................................6Ultrasound Indications for Treatment...............................7Ultrasound Contraindications & Warnings........................7Introduction....................................................................8Therasound 3 Series Operation........................................8
Therasound 3.1/3.2, 3.4, and 3.5 Front Panel Illustrations.........9Combining Therasound 3 Units with Stimulation........................10
Ultrasound Calibration & Tuning Procedure.....................11Trouble Shooting............................................................ 13Therasound 3 Series Specifications..................................13Changing the Voltage of the Therasound 3 Series..............14
Appendix AUltrasound Technical Information
Appendix BTherasound 3 Series Parts List
Appendix CTherasound 3 Series Schematics
LIMITED WARRANTY
This equipment is sold under an exclusive two-year warranty from date of sale, which warrants it to be freefrom defects in material and workmanship. We agree to repair or replace at the point of manufacture,without charge, all parts showing such defects, provided the unit is delivered to us, prepaid to our factory,intact for our examination, within two years from date of sale, and provided such examination discloses inour final judgement that it is defective.
This warranty does not apply if the equipment has been subject to misuse, neglect, accidents, incorrectwiring (not our own), improper installation, or put to use in violation of instructions furnished by us, has beendamaged by excess voltage or has been repaired or altered outside our factory or if the equipment has hadits serial number altered or removed.
Changes: Rich-Mar reserves the right to modify or change the equipment in whole or in part, at any timeprior to delivery, in order to include refinements deemed appropriate by the Company but without incurringany liability to modify or change equipment previously delivered, or to supply new equipment in accordancewith earlier specifications. This warranty will be honored only if the enclosed card is filled out and returnedto the factory. This warranty is valid only to original purchaser.
This warranty is expressly in lieu of all other warranties expressed or implied including the warranties ofmerchantability and fitness for use and all other obligations on our part, and we neither assume, nor autho-rize any other person to assume for us, any other liability in connection with the sale or use of this equip-ment. In no event shall we be liable for consequential or special damages. We make no warranty whatso-ever in respect to accessories or parts not supplied by us.
6
7
Ultrasound Indications for Treatment(Therapeutic Ultrasound)
Rich-Mar Ultrasound devices are indicated to produce therapeutic deep heat for the following condi-tions:1) Relief of pain.2) Muscle spasms.3) Joint contractures.
But not for the treatment of malignancies.
WARNING - Federal law restricts this device to sale by or on the order of a physician or any other practitio-ner licensed by the law of the state in which said person practices.
Ultrasound Contraindications
ContraindicationsUltrasound should not be used in the following areas:1) Near or over the heart.2) Near or over the eyes.3) On the head.4) Near or over reproductive organs.5) On the lower back during pregnancy or over the pregnant uterus.6) Directly over the spinal column.7) Over growing bone in children.8) Where the skin suffers from any sensory impairment.9) Over areas of malignancies.10) In the area of visceral plexus and large autonomous ganglion.11) Over the thoracic area if the patient is using a cardiac pacemaker.12) Over a healing fracture.13) Over ischemic tissues in individuals with vascular disease where the blood supply would be unable tofollow the increase in metabolic demand and tissue necrosis might result.
PrecautionsPrecautions should be taken when used:1) Over anesthetized areas.2) On patients with hemorrhagic diastheses.3) Ultrasound treatment should not be performed over an area of the spinal cord following laminectomy (i.e.-when major covering tissues have been removed).
Caution1) Excessive doses of ultrasound may cause damage to tissue. Periosteal pain is an indication of excessintensity and if it occurs, the power should be reduced; the transducer should be moved more rapidly over thearea being treated; or a lower pulsed duty cycle should be used.2) If the soundhead has been operated unloaded for an extended period of time, the transducer will get hot. Ifthe soundhead is applied to the patient while the transducer is hot, a burn may result.
WarningDo not operate the soundhead in an unloaded condition. It is possible that unrepairable damage may occur tothe transducer in an unloaded state.
To pause treatment, press the “Pause” button. Timewill stop flashing and the intensity will begin to flash.Notice that the ultrasound active indicator is no longerilluminated, indiacting that the treatment has beenpaused. To resume treatment, press the “Resume”button and the time will flash and the ultrasound activeindicator will illuminate. When the treatment time ends,a loud beep will sound and time will stop flashing andrevert to the last treatment time entered. The intensitywill revert to zero and the ultrasound active light willextinguish.
To stop a treatment, press the “Stop” button. Time willstop flashing and revert to the last treatment timeentered. Intensity will revert to zero and the ultrasoundactive light will extinguish.
All most recently used parameters, with the exceptionof intensity, will remain saved as the default param-eters after a treatment ends. To use the same treat-ment again, enter the desired intensity and press the“Start” button to begin.
Therasound 3.4 Operation (see Fig. 2, page 9)The Therasound 3.4 device is a dual-frequency (1MHzand 3MHz) ultrasound utilizing a 5cm2 transducer.
The unit is operated in the same manner as theTherasound 3.1, except that the user is able to choosebetween the 1MHz and 3Mhz frequencies. Afterchoosing the desired duty cycle, select the frequencyby pressing the corresponding button.
Therasound 3.5 Operation (see Fig. 3, page 9)The Therasound 3.5 device is a dual-frequency (1MHzand 3MHz) ultrasound utilizing the patented “TherapyHammer” 2cm2/5cm2 transducer. Continue with theoperation instructions for the Therasound 3.1.
The unit is operated in the same manner as theTherasound 3.1, except that the user is able to choosebetween the 1MHz and 3MHz frequencies and the2cm2 and 5cm2 transducers. After choosing thedesired duty cycle, select the frequency by pressingthe corresponding button. Then select the desiredtransducer (2cm2 or 5cm2) by pressing the “Trans-ducer” button. Continue with the operation instructionsfor the Therasound 3.1.
When finished making changes, return to the openingscreen by pressing stop/clear or start treatment bypressing the start button.
8
IntroductionThe Rich-Mar Therasound 3 series was designed toprovide therapy professionals with the best technologyat affordable prices. The Therasound 3 series offersthe most flexible treatment possibilities in a convenient,easy-to-use package.
This manual is meant to familiarize the user with thecontrols, operations, and ultrasound therapies availablein the Therasound 3 units. The simple control of theunits allow the user to master the units’ vast capabili-ties quickly and easily.
Therasound 3.1/3.2 Operation (see Fig. 1, page 9)The Therasound 3.1 ultrasound is a 1MHz unit utilizinga 5cm2 transducer and the Therasound 3.2 ultrasoundis also a 1MHz unit utilizing a 10cm2 transducer. Tooperate the unit, make sure that the power cord isplugged into the power receptacle on the back of theunit. The power receptacle is located below the On/Off switch. Turn the switch to the “1,” or on position.The device should then activate.
The Therasound unit will then go through a quickdiagnostic check where some letters and numbersappear on the display. This is normal. After the diag-nostic check green digits will appear in the time andintensity windows.
To set up a treatment, select the parameters desired bypressing their corresponding buttons. A beep will soundevery time a button is pressed on the device. Select theduty cycle (100%, 50%, 20%, or 10%) by pressing theDuty Cycle Select button. Then select the desiredtreatment time by pressing the “+” or “-” buttons in the“Time Set” area. Select how the intensity will bedisplayed, either in W/cm2 or Watts, by pressing theSelect button in the “Intensity” area. Set the intensityby using the “+” or “-” buttons in the “Intensity” area.
Once all treatment parameters have been set asdesired, apply ultrasound-coupling lotion or gel to thepatient and place the transducer on the patient. Pressthe “Start” button and the “Time” indicator will start toflash. This indicates that the treatment is running andtime is counting down. Notice that the green “Ultra-sonic Active” indicator light next to the soundheadcable on the front of the unit has activated. Thisindicates that the device is outputting ultrasound.
Please note that any treatment parameter may bemodified or changed while the device is operatingsimply by pressing the button corresponding to theparameter to be changed.
9
Fig. 1 Therasound 3.1/3.2 front panel
Fig. 2 Therasound 3.4 front panel
Fig. 3 Therasound 3.5 front panel
NOTE: Ultrasonic Active LED is located next to the soundhead cable exiting below the front panel.
100%
10%20%50%
Select
Duty Cycle Pause/ResumeStart/Stop
Time - Minutes
Select
Intensity
Watts
W/cm2
Frequency
Select
Time - Minutes
Duty Cycle
100%
10%20%50%
3 MHz
1 MHzSelect
Start/Stop
Intensity
Select
WattsW/cm 2
Pause/Resume
Duty Cycle
Time - Minutes
50%20%10%
100%
Select
Start/Stop
Intensity
Select
W/cmWat t s
2
Pause/ResumeFrequency
3 M H z
1 M H zSelect
Transducer
Select5cm
2cm 2
2
Combining Ultrasound withRich-Mar Stimulation
The Rich-Mar Therasound 3 Series ultrasounds aredesigned to be connected to any Rich-Mar musclestimulator, thus enabling the user to provide combina-tion therapy to patients.
To connect the Therasound unit to a stimulator, simplyplug an electrode lead from the stimulator into the jackon the lower right rear side of the Therasound ultra-sound unit.
Using the indifferent electrode of the stimulator tocomplete the circuit with the soundhead of theTherasound, the user will be able to provide electricalstimulation as well as ultrasound through the trans-ducer.
CAUTION: When using a combination treatment, notethat both faces of the “Therapy Hammer” transducerwill output stimulation.
10
Disinfecting RecommendationsTo disinfect the soundhead between therapy treat-ments, Rich-Mar recommends using a disinfectantcleaner for ultrasound. OSHA addresses the need forprudent infection control (OSHA Instruction CPL 2-2.33C) to include decontamination of equipmentbetween patients.
11
Ultrasound Calibration andTuning Procedure
Ultrasound Service InformationRich-Mar Corporation recommends that all Rich-Marultrasonic therapy products be returned to the factoryor to a servicing Rich-Mar distributor for service orcalibration. It is recommended that the device becalibrated annually or when any major component ischanged.
CautionCalibration and peaking adjustments must not beattempted unless the person performing these adjust-ments has the proper test equipment, which mustinclude an acceptable ultrasonic wattmeter, such as theOhmic UPM-30 or equivalent. Degassed water mustbe used to obtain accurate readings (4 parts per millionof oxygen).
WarningUse of controls or adjustments or performance ofprocedures other than those specified herein mayresult in hazardous exposure to ultrasonic energy.
Calibration and Tuning Procedure
Tuning Frequency and Setting Voltage5cm @ 1MHz (applies to Therasound 3.1, 3.4, 3.5)1) Install the 5cm soundhead into the wattmeter
making sure that only the face of the transducer issubmerged. Zero the wattmeter.
2) Hold down the Start/Stop switch and turn on theunit. The software version should appear on thescreen.
3) Let off the Start/Stop switch and the frequency willappear on the screen for 5cm@1MHz. Find thefrequency peak and add 5kHz.
Note: Normal range of operation is 930 to 960.
4) Press Start/Stop switch and the voltage will appear.Set maximum output to 10 watts.
* If Therasound 3.1 see “Setting Thermistor Value” onnext page to complete calibration.
5cm @ 3MHz (applies to Therasound 3.4 and 3.5)1) ) Press the Start/Stop switch.
2) Find the frequency peak for 5cm @3MHz and add7kHz.
Note: Normal range of operation is 3050 to 3150.
3) Press the Start/Stop switch and the voltage willappear. Set maximum output to 10 watts.
* If Therasound 3.4 see “Setting Thermistor Value” onnext page to complete calibration.
2cm @ 1MHz (applies to Therasound 3.5 only)1) Press the Start/Stop switch.
2) Find the frequency peak for 2cm @ 1MHz and add5kHz.
Note: Normal range of operation is 950 to 980.
3) Press the Start/Stop switch and the voltage willappear. Set maximum output to 4 watts.
2cm @ 3MHz (applies to Therasound 3.5 only)1) Press the Start/Stop switch.
2) Find the frequency peak for 2cm @ 3MHz and add5kHz.
Note: Normal range of operation is 3050 to 3150.
3) Press the Start/Stop switch and the voltage willappear. Set maximum output to 4 watts.
* If Therasound 3.5 see “Setting Thermistor Value” onnext page to complete calibration.
Note:Therasound 3.1 - you need to complete only the
5cm @ 1MHz portion.Therasound 3.2 - you need to complete only the
10cm @ 1 MHz portion.Therasound 3.4 - you need to complete both the
5cm @ 1 MHz and the 5cm@3 MHzportions.
Therasound 3.5 - you need to complete the 5cm @1MHz, 5cm @ 3MHz, 2cm @ 1MHz, and2cm @ 3MHz portions.
12
Cable DiagnosticTherasound units are equipped with a diagnosticfeature that will detect a broken cable. The diagnosticcheck is run when the device is powered on. If thedevice detects a variation in cable connection it willdisplay 7E S7 (test) on the display for 2 seconds.If this occurs, conduct the water test outlined in theOutput Trouble section on page 9. If the water doesnot cavitate, or bubble, the device may have a brokencable or it may need to be tuned and calibrated. If thedevice is outputting ultrasound, there may have been aslight variation in the diagnostic. Rich-Mar recom-mends that the device be calibrated at least once everyyear.
10cm @ 1MHz (applies to Therasound 3.2 only)1) Install the 10cm soundhead into the wattmeter
making sure that only the face of the transducer issubmerged. Zero the wattmeter.
2) Hold down the Start/Stop switch and turn on theunit. The software version should appear on thescreen.
3) Let off the Start/Stop switch and the frequency willappear on the screen for 10cm@1MHz. Find thefrequency peak and add 5kHz.
Note: Normal range of operation is 900 to 950.
4) Press Start/Stop switch and the voltage will appear.Set maximum output to 20 watts.
Complete calibration by Setting the Thermistor Value.
Setting the Thermistor Value1) After the last frequency and voltage calibration has
been made, press Start/Stop. Verify that thesoundhead temperature thermistor value is set at“H0 73.”If so, press Start/Stop to enter this number.
If not, set at H0 73, using the “+” or “-” switches toobtain the correct reading. Once setting is correctpress Start/Stop twice.
13
Therasound 3 Series Specifications
Therasound 3.1/3.2Dimensions: 7.5"W x 7.5"D x 4"H
Weight: 5.5 lbs.
Power Input: 115 VAC, 1.0 A, 60Hz or230 VAC, 0.5A, 50Hz
Output: 0-2 w/cm2
SoundheadERA: 5cm2 - 3.1
10cm2 - 3.2
Frequency: 1MHz
BNR: 5.5:1 maximum
Therasound 3.4Dimensions: 7.5"W x 7.5"D x 4"H
Weight: 5.5 lbs.
Power Input: 115 VAC, 1.0 A, 60Hz or230 VAC, 0.5 A, 50Hz
Output: 0-2 w/cm2
SoundheadERA: 5cm2
Frequency: 1MHz & 3MHz
BNR: 5.5:1 maximum
Therasound 3.5Dimensions: 7.5"W x 7.5"D x 4"H
Weight: 5.5 lbs.
Power Input: 115 VAC, 1.0 A, 60Hz or230 VAC, 0.5 A, 50Hz
Output: 0-2 w/cm2
SoundheadERA: 2cm2 & 5cm2
Frequency: 1MHz & 3MHz
BNR: 5.5:1 maximum
Trouble-ShootingRich-Mar Corporation takes pride in its TechnicalSupport Hotline: 1-918-543-2222. We have an out-standing staff ready to take your calls and help withdiagnosing and troubleshooting problems. Listed beloware several options for troubleshooting the Therasound3 Series.
1.) The device fails to turn on.Check to verify that the power cord is fully pluggedinto the corcom power entry module below the ACswitch on the rear of the device. Also verify that thepower cord is fully plugged into the AC wall outlet.
Check the fuse to verify that it has not blown.CAUTION: To ensure personal saftey, unplug the unitbefore checking the fuse.
2.) Ultrasonic Active LED fails to illuminateVerify that the soundhead connections are pluggedfirmly into the main board.
3.) Display reads “HEAd.”The “HEAd” warning is a safety feature wherein thesoundhead temperature shuts off to avoid overheating.This warning occurs when the soundhead overheats,usually due to an un-coupled or poor coupling situation.Turn the device off for 3-4 minutes to allow it to resetand turn it back on and begin treatment. If the“HEAd” warning is still displayed, please call thefactory.
When using the Therasound 3.5, check that thesoundhead in use is active. Also, be sure to use anadequate amount of coupling gel or lotion specificallydesigned for ultrasound transmission.
Output TroubleshootingRich-Mar suggests that ultrasound output be testedonce a week. To do this, turn on the device and start atreatment at 100% output. The outputting head shouldbe facing up towards the user. Pour some water on thesoundhead and increase the intensity. If the device isoutputting ultrasound, the water should cavitate, orbubble, on the soundhead.CAUTION: Do not operate the ultrasound in an“unloaded” condition for more than one minute atintensities greater than 1.0 W/cm2. It is possible thatthe transducer may overheat and damage the unit.
14
Changing the Voltage from 115V to 230VWARNING: RISK OF ELECTRICAL SHOCK.
Unplug unit prior to changing the voltage from 115 to 230.
1. After unplugging the unit, openthe outlet cover with a flatheadscrewdriver by inserting it into thenotch on the top.
2. Once the outlet cover is open,insert the flathead screwdriver intothe notch on the top of the redvoltage converter.
6. Finally, close the outlet cover, ensuring that thedesired voltage appears through the window in theoutlet cover. Also be sure to use an adapter forthe AC plug to suit local wall outlets in use.
3. Holding the regulator so that115V is right-side up, remove the1amp slow blow fuse from the rightside.
4. Rotate the regulator until 230Vis right-side up. Remove the clipfrom the right-hand side and inserttwo .5amp slow-blow fuses, oneon each side.
5. SAVE THE CLIP AS IT IS NEEDED TO SWITCH BACK TO115V. After correct fuses are inserted into each side, insert the regulatorinto the slot with 230V reading right-side up.
115V to 230V Quick Check, Take out 1amp slow blow fuse and remove clip on back of 230V side., SAVE CLIP, Insert two .5 amp slow blow fuses., Insert regulator with 230V appearing through the window in outlet cover.
Changing the Voltage from 230V to 115VWARNING: RISK OF ELECTRICAL SHOCK.
Unplug unit prior to changing the voltage from 230 to 115.
1. After unplugging the unit, openthe outlet cover with a flatheadscrewdriver by inserting it into thenotch on the top.
2. Once the outlet cover is open,insert the flathead screwdriver intothe notch on the top of the redvoltage converter.
3. Remove the two .5 amp slowblow fuses from the voltageregulator.
4. Hold the regulator so that230V is right-side up. Insert theclip into the right-hand side of theregulator.
5. Rotate the regulator so that 115V is right-side up and insert one 1amp slow blow fuse into the right side. Insert the regulator into the slotin the unit with 115V reading right-side up.
6. Close the outlet cover, ensuring that 115V appears through thewindow in the outlet cover. Also be sure to use an adapter for theAC plug to suit local wall outlets in use.
230V to 115V Quick Check, Take out the two .5 amp slow blow fuses., INSERT THE CLIP INTO THE RIGHT HAND SIDE OF THE 230V REGULATOR, Insert one 1 amp slow blow fuse., Insert regulator with 115V appearing through the window in outlet cover.
15
APPENDIX AULTRASOUND TECHNICAL INFORMATION
Near Field Distribution
Beyond this point, the beam has a more uniform intensityand is called the “far field”. Below is shown the far fielddistribution at 16cm from the transducer face.
Far Field DistributionThe preceding descriptions apply for radiation emittedinto the equivalent of an infinite medium of distilled,degassed water at 30°C and with line voltage variationsin the range of +/-10% of the rated value
Transducer Parameters and Tolerances:The Rich-Mar ultrasound units operate at frequencies ofeither 1MHz or 3MHz +/- 10%. The effective radiatingareas (ERA) of the transducers are ten, five,
Ultrasound TechnicalInformation
Applicator Type:The ultrasonic radiation fields produced by Rich-Martherapeutic ultrasound transducers are of the planewave type and are essentially cylindrical in shape. Thistype of applicator is referred to as a collimatingapplicator.
Applicator Label:Each Rich-Mar applicator is labeled to provide the userwith information on its applicable parameters. Thefollowing abbreviations are used on the label.
Gen: The Rich-Mar ultrasonic generatorfor which the applicator is intended.
f: The operating frequency in MHz forthe applicator.
Area: The effective radiating area of theapplicator in square centimeters.
BNR: The Beam Nonuniformity Ratio.
Type: Coll-means collimating applicator.
Near Field/ Far Field:If measurements are made of the sound intensity alongthe central axis of the beam produced by the applica-tor, the intensity distribution shows maxima and minimanear the applicator and then a gradual decline beyondthe last maximum intensity.The “interference” or “near field” is the area in theultrasound beam extending from the applicator surfaceto the location of the most distant intensity maximum.In this area, maxima and minima of intensity arelocated close to each other. This is the area in whichmost therapeutic application occurs. This is shown inthe following figure measured 0.5cm from the trans-ducer face.
18
19
or two square centimeters, depending upon the size ofthe transducer being used. The tolerance for the ERAis +/-25% on the 2 and 5 square centimeter transduc-ers. The tolerance for the 10 square centimetertransducers is +0. -25%. The Beam-Nonuniformity-Ratio (BNR) of any Rich-Mar transducer is 5.5:1 orless.
100% ModeWhen operated in the 100% mode, the generatorproduces a non-interrupted sinusoidal waveform of oneor three MHz. The peak power and average powerare therefore the same.The error in indication of radiated power in intensityfor the continuous mode does not exceed +/- 14%allowing for a 6% error in the wattmeter, which equals+/- 20%.
Pulsed ModeWhen operated in the pulsed mode, the generatorproduces a square-wave burst of sinusoidal waveformof 1MHz or 3MHz of 2.5 milliseconds in duration.Depending upon the Rich-Mar model of therapeuticultrasound in use, the duty cycle can be chosen be-tween 5% and 95% duty. This then implies the repeti-tion rate is selectable between 20 and 380 pulses persecond. (This is computed by taking the inverse of theduty cycle 1/380 = .95, 1/20 = .05). The tolerance forthe pulsed mode is +/- 20%.See the following chart for second comparison on%Duty cycle to pulses.
% Duty Cycle Pulses/Second (Indicated on front panel of device)
5 2010 4015 6020 8025 10030 12035 14040 16045 18050 20055 22060 24065 26070 28075 30080 32085 34090 36095 380
The error in indication of radiated power in intensityfor the pulsed mode does not exceed +/-14% allowingfor an allowable 6% error in the wattmeter, whichequals +/-20%.
Timer AccuracyThe Food and Drug Administration requires that thetreatment timer accuracy is to within 0.5 minutes forthe preset duration of emission for settings less thanfive minutes, to within 10% of the preset duration ofemission for settings from five to ten minutes, and towithin one minute of the preset duration of emission forsettings greater than ten minutes.
Ratio of Temporal Peak to Temporal Average(Rtpa):The ratios of temporal peak to temporal averageintensities (Rtpa) will vary with the pulse rate of thedevice. Depending upon the Rich-Mar model oftherapeutic ultrasound in use, the duty cycle can bechosen between 5% and 95% duty.The Rtpa is calculated in the following manner:Rtpa = (1/Duty):1Example 5% duty = .05 (min. duty, max. Rtpa)Rtpa = (1/.05):1Rtpa = 20:1Example 95% duty = .95 (max. pulsed duty, min. Rtpa)Rtpa = (1/.95):1Rtpa = 1.05:1See the following chart for %Duty cycle to Rtpacomparison.
% Duty Cycle Rtpa (Indicated on front panel of device)
5 20:110 10:115 8.33:120 5:125 4:130 3.33:135 2.86:140 2.5:145 2.22:150 2:155 1.82:160 1.66:165 1.54:170 1.43:175 1.33:180 1.25:185 1.18:190 1.11:195 1.05:1
The Rtpa tolerance does not exceed +/- 20%.The temporal maximum intensity for each duty cycleas well as the 100% modulation is whatever is indi-cated on the meter.
The temporal average intensity for each duty cycle willbe the meter indication multiplied by the percentageduty cycle.
Temporal Average = (Duty) x (Meter Indication)Example, 5 Watts, 35% DutyTemporal Average = .35 x 5 Watts = 1.75 Watts
The Spatial Average Intensities for each of thesesetting will be divided by the transducer’s EffectiveRadiating Area (ERA)
Spatial Average = (Temporal Average)/(ERA)Example, 5 Watts, 35% Duty, 5cm2 Transducer
Spatial Average = (1.75 Watts)/(5cm2) = 0.35 Watts/cm2
The pulse width (On time) of all Rich-Mar therapeuticultrasound devices is 2.5 milliseconds (mS). The timebetween pulses (Off time) in milliseconds is calculatedas follows:
Pulse width (On time) = 2.5mSOff time = [2.5-2.5(%Duty cycle)]/(%Duty cycle)Where %Duty cycle is represented as a decimal.
Please see the following example for computing theOff time for a 10% Duty cycle:
Off time=[2.5-2.5(0.10)]/(0.10)=22.5 milliseconds.
Additional Technical Notes:The peak power is the same in the pulsed modes as inthe 100% modulated mode.Unless otherwise stated, all technical parameters areaccurate within +/- 20%.When in the pulse modes the unit is still generatingtherapeutic heat, although it is an amount reduced by afactor directly related to the duty cycle. The pulserates are used to allow the practitioner to treat areas ofbony prominences without creating periosteal pain.The line leakage is tested in both the forward andreverse polarities to be less than 50 microamperesexceeding all standards for medical devices in thisclass.The device is designed to meet or exceed UL Stan-dards 544 for medical devices and the CanadianStandards Association (CSA No. 125).
20
APPENDIX BPARTS LIST
The
raso
und
3 Se
ries
Par
ts L
ist
The
raso
und
3.1,
3.4
, and
3.5
Ric
h-M
ar P
art N
o.Pa
rt N
ame
Des
crip
tion
Qty
/Boa
rd06
00#4
SO
LDER
LU
GG
RO
UN
D S
TRA
P2
0704
#4 x
1/4
” SP
AC
ERD
ISPL
AY
MO
UN
TIN
G4
0712
6-32
x 1
/2”
MA
IN B
OA
RD
(GR
OU
ND
) MO
UN
TIN
G S
PAC
ER1
0730
PLA
STIC
SPA
CER
MA
IN B
OA
RD
MO
UN
TIN
G4
3269
10FN
NEZ
10 S
ELF
LOC
KIN
G2
3510
SMIT
H 2
182
FEET
437
45N
EW S
TYLE
RIG
HT
BA
ILH
AM
MER
137
97TH
ERA
SOU
ND
3 S
ERIE
S1
4161
AM
P 64
0440
-2FA
N C
ON
NEC
TOR
(.1”
2PI
N F
EMA
LE)
241
8364
0433
-3TR
AN
SFO
RM
ER C
ON
NEC
TOR
(.15
6” 3
PIN
FEM
ALE
)1
4200
6404
40-6
LEA
D C
ON
NEC
TOR
(.1”
6 P
IN F
EMA
LE)
142
5126
PIN
HEA
DER
RIB
BO
N C
AB
LE C
ON
NEC
TOR
248
04TS
2.5,
TM
3C/3
P FA
N1
4806
FAN
GU
AR
D1
5008
SLO
W B
LOW
1 A
MP
155
1660
97 P
IN JA
CK
(KEY
STO
NE)
CO
MB
INA
TIO
N1
5905
GRE
EN L
EDU
A A
CTI
VE
159
11M
E-35
2-00
02 L
ED H
OLD
ER1
7403
5cm
CR
YST
AL
& D
IAPH
RA
GM
178
40PS
OSX
SOO
O (U
S500
, 750
, 100
0)C
OR
CO
M P
OW
ER E
NTR
Y M
OD
ULE
(AC
SW
ITC
H &
FU
SE)
183
41FD
7-36
MA
GN
ATE
KTR
AN
SFO
RM
ER1
8694
THER
MIS
TOR
191
12FE
MA
LE S
PAD
E C
ON
NEC
TOR
491
16N
P-07
-GY
(NU
T FO
R M
S 91
15)
TRA
NSD
UC
ER S
TRA
IN R
ELIE
F M
OU
NTI
NG
191
50A
T-40
SPE
AK
ER1
9161
KEY
STO
NE
7623
3/1
6” C
LIP
STR
AIN
REL
IEF
FOR
TR
AN
SFO
RM
ER W
IRES
191
7177
21-3
PPS
#4 W
ASH
ER4
9182
THER
MA
LLO
Y 5
3-77
-4 S
IL P
AD
491
86TH
ERM
ALL
OY
772
1-7P
PS4
9843
AC
CO
RD
/US
500/
750/
1000
1
The
raso
und
3.1
only
Ric
h-M
ar P
art N
o.Pa
rt N
ame
Des
crip
tion
Qty
/Boa
rd26
79U
S 50
0 M
AIN
BO
AR
D1
2680
US
750/
1000
FR
ON
T PA
NEL
BO
AR
D1
6352
US
500
FRO
NT
PAN
EL1
The
raso
und
3 Se
ries
Par
ts L
ist
The
raso
und
3.1
only
Ric
h-M
ar P
art N
o.Pa
rt N
ame
Des
crip
tion
Qty
/Boa
rd63
61TH
ERA
SOU
ND
3.1
PA
NEL
LA
BEL
174
36U
S500
3.1
5cm
191
39R
B14
-250
F D
ISC
ON
NEC
T1
The
raso
und
3.4
only
Ric
h-M
ar P
art N
o.Pa
rt N
ame
Des
crip
tion
Qty
/Boa
rd26
80U
S750
/100
0 FR
ON
T PA
NEL
BD
.1
2684
US1
000/
750
MA
IN B
OA
RD
163
59U
S 75
01
6363
THER
ASO
UN
D 3
.4 P
AN
EL L
AB
EL1
The
raso
und
3.5
only
Ric
h-M
ar P
art N
o.Pa
rt N
ame
Des
crip
tion
Qty
/Boa
rd26
80U
S 75
0/10
00 F
RO
NT
PAN
EL B
D.
126
84U
S100
0/75
0 M
AIN
BO
AR
D1
6364
THER
ASO
UN
D 3
.5 P
AN
EL L
AB
EL1
7402
2cm
CR
YST
AL
& D
IAPH
RA
GM
174
40U
S 10
00 3
.5 H
H1
Ref
eren
ce D
esig
nato
r(s)
Par
t Num
ber
Des
crip
tion
Q
ty/B
oard
.(U
1)(U
2)(Q
1)(q
2
4
880S
HEA
TSIN
K, H
AR
DW
AR
E K
IT, T
O22
0
4
(U1-
U2)
(Q1-
Q2)
H
S522
1H
EATS
INK
, BA
R, R
’M C
UST
OM
, TO
220
2(U
14)
8215
74-1
SKT,
44-
PIN
PLC
C, T
IN0
(U5)
2-
6404
63-1
SKT,
8-P
IN D
IP, T
IN, .
100
1(U
5)
58
010B
0000
0H
EATS
INK
, IC
CLI
PON
, DIP
81
C1
C2
C3
C5
C6
C
1812
C10
4M1R
AC
CA
P, C
ER, 0
.1uF
,100
V, 2
0%, X
7R, 1
812
5C
10
ECE-
A1J
U22
1C
AP,
ALU
M, 2
20uF
, 63V
, 20%
, RA
DIA
L1
C11
E
CE-
A1H
U33
1C
AP,
ALU
M, 3
30uF
, 50V
, 20%
, RA
DIA
L1
C12
C14
C15
C16
C19
C22
C
0805
C10
4M5R
AC
CA
P, C
ER, 0
.1uF
, 50V
, 20%
, X7R
, 080
516
C23
C26
C29
C31
C33
C37
C40
C42
C7
C8
C13
C08
05C
102M
5RA
C
CA
P, C
ER, 0
.001
uF, 5
0V, 2
0%, X
7R, 0
805
1C
17
593
D47
6X00
20E2
TC
AP,
TA
NT,
47u
F,20
V, 2
0%, E
-SIZ
E1
C18
EC
H-S
1272
JZC
AP,
PPS
, 270
0pF,
100V
, 5%
, RA
DIA
L1
C20
5
95D
686X
0025
R2T
CA
P, T
AN
T, 6
8uF,
25V
, 20%
, R-S
IZE
1
The
raso
und
3 Se
ries
Par
ts L
ist
Ref
eren
ce D
esig
nato
r(s)
Par
t Num
ber
Des
crip
tion
Q
ty/B
oard
.C
21 C
24EC
A-1
HFQ
330
CA
P, A
LUM
, 33u
F, 5
0V, 2
0%, R
AD
IAL
2C
25T3
22A
275K
010A
SC
AP,
TA
NT,
2.7
uF, 1
0V, 1
0%, A
XIA
L1
C27
ECH
-S15
61JZ
CA
P, P
PS, 5
60pF
, 100
V, 5
%, R
AD
IAL
1C
28C
0805
C56
1MR
5AC
CA
P, C
ER, 5
60pF
, 50V
, 20%
, X7R
, 080
51
C30
595D
107X
9016
C2T
CA
P, T
AN
T, 1
00uF
, 16V
, 10%
, C-S
IZE
1C
3059
5D10
7X00
16C
2TC
AP,
TA
NT,
100
uF,1
6V, 2
0%, C
-SIZ
E1
C32
C34
ECE-
A1C
U22
0C
AP,
ALU
M, 2
2uF,
16V
, 20%
, RA
DIA
L2
C35
T491
B10
6K00
6AS
CA
P, T
AN
T, 1
0uF,
6V
, 10%
, B-S
IZE
1C
3659
3D22
7X00
10E2
TC
AP,
TA
NT,
220
uF,1
0V, 2
0%, E
-SIZ
E1
C38
C39
C06
03C
220M
5RA
CC
AP,
CER
, 22p
F, 5
0V, 2
0%, X
7R, 0
603
1C
4EC
E-A
1JU
471
CA
P, A
LUM
, 470
uF, 6
3V, 2
0%, R
AD
IAL
1C
41T4
91A
474M
020A
SC
AP,
TA
NT,
0.4
7uF,
20V
, 20%
, A-S
IZE
1C
9EC
E-A
2AU
471
CA
P, A
LUM
, 470
uF, 1
00V
, 20%
, RA
DIA
L1
D1
D2
D3
D6
S1JB
DIO
, GLA
SS P
ASS
, 600
V, 1
A, D
0214
4D
10 D
11 D
4 D
8I1
0BQ
040
DIO
, SC
HO
TTK
Y, 4
0V, 1
A, S
MB
4D
12 D
13 D
9LL
4148
DIO
, 1N
4148
, 75V
, .15
A, S
0D-8
03
D14
BZX
84C
5V1Z
DIO
, ZEN
ER, 5
.1V
, 250
mA
, 5%
, SO
T23
1D
5K
BU
4JR
ECTI
FIER
, FU
LL-W
AV
E B
RID
GE
1D
7LN
1251
CLE
D, R
ED, J
-TY
PE, S
MT
1F1
60R
110
FUSE
, PTC
RES
ETTA
BLE
1F2
60R
075
FUSE
, PTC
RES
ETTA
BLE
1J1
7191
2-12
6H
DR
, 26-
PIN
DU
AL
RO
W,
EJEC
T/LA
TCH
ING
, .10
01
K1
K2
DS2
E-S-
DC
12V
REL
AY
, DPD
T, 1
2V, 2
A, P
CM
OU
NT
2L1
CTM
2299
7IN
DU
CTO
R, W
ITH
TA
P C
OIL
, 13.
0uH
1L2
PE-5
3827
SIN
DU
CTO
R, P
OW
ER1
L3PE
-538
19S
IND
UC
TOR
, PO
WER
1L4
CC
TM15
399
IND
UC
TOR
, WIT
H T
AP
CO
IL, 4
.84u
H,
1P1
P2
P764
0456
-2H
DR
, 2-P
IN, V
ERTI
CA
L FR
ICTI
ON
LO
CK
, .10
03
P364
0445
-3H
DR
, 3-P
IN, V
ERTI
CA
L FR
ICTI
ON
LO
CK
.156
1P5
P6
RL3
59H
DR
, RC
A JA
CK
, 3-P
IN2
P81-
6404
56-0
HD
R, 1
0-PI
N, V
ERTI
CA
L FR
ICTI
ON
LO
CK
, .10
01
PCB
SET0
2DPC
B, R
M U
S100
0 SE
RIE
S M
AIN
BO
AR
D1
Q1
Q2
IRL5
20N
FET,
N-C
H, P
OW
ER, T
O-2
20-U
P2
Q3
ZVP3
306F
TAFE
T, P
-CH
, SO
T23
1
The
raso
und
3 Se
ries
Par
ts L
ist
Ref
eren
ce D
esig
nato
r(s)
Par
t Num
ber
Des
crip
tion
Q
ty/B
oard
.Q
4 Q
5FE
T, N
-CH
. EN
HA
NC
EMEN
T M
OD
E, S
OT2
3-G
SD1
R1
ERJ-
14R
SJR
10U
RES
, 0.1
, 1/4
W, 5
%, 1
210
1R
10 R
12C
RC
W08
05-1
004F
RT1
RES
, 1M
, 1%
, 1/1
0W, 1
00V
, 080
52
R11
R9
CR
CW
0805
-002
02
R13
CR
CW
0805
-110
3FR
T1R
ES, 1
10K
, 1%
, 1/1
0W, 1
00V
, 080
51
R14
R17
R18
R19
R20
R22
CR
CW
0805
-100
2FR
T1R
ES, 1
0K, 1
%, 1
/10W
, 100
V, 0
805
13R
29 R
30 R
31 R
37 R
39 R
5 R
8
R15
R40
CR
CW
0805
-100
1FR
T1R
ES, 1
K, 1
%, 1
/10W
, 100
V, 0
805
2R
16C
RC
W08
05-4
991F
RT1
RES
, 4.9
9K, 1
%, 1
/10W
, 100
V, 0
805
1R
2C
RC
W08
05-8
871F
RT1
RES
, 8.8
7K, 1
%, 1
/10W
, 100
V, 0
805
1R
21 R
32C
RC
W08
05-2
002F
RT1
RES
, 20K
, 1%
, 1/1
0W, 1
00V
, 080
52
R23
R6
R7
CR
CW
0805
-100
3FR
T1R
ES, 1
00K
, 1%
, 1/1
0W, 1
00V
, 080
53
R24
CR
CW
0805
-301
3FR
T1R
ES, 3
01K
, 1%
, 1/1
0W, 1
00V
, 080
51
R25
CR
CW
0805
-374
1FR
T1R
ES, 3
.74K
, 1%
, 1/1
0W, 1
00V
, 080
51
R26
CR
CW
0805
-221
1FR
T1R
ES, 2
.21K
, 1%
, 1/1
0W, 1
00V
, 080
51
R27
R36
CR
CW
0805
-000
0JR
T1R
ES, 0
OH
M, 5
%, 1
/10W
, 100
V, 0
805
2R
28 R
35TB
DR
3C
RC
W08
05-1
960F
RT1
RES
, 196
, 1%
, 1/1
0W, 1
00V
, 080
51
R33
CR
CW
0805
-100
5FR
T1R
ES, 1
0M, 1
%, 1
/10W
, 100
V, 0
805
1R
34C
RC
W08
05-5
761F
RT1
RES
, 5.7
6K, 1
%, 1
/10W
, 100
V, 0
805
1R
38C
RC
W08
05-4
700F
RT1
RES
, 470
, 1%
, 1/1
0W, 1
00V
, 080
51
R4
CR
CW
0805
-100
0FR
T1R
ES, 1
00, 1
%, 1
/10W
, 100
V, 0
805
1R
P1 R
P3 R
P5B
CN
164A
102J
7R
ES N
ET, I
SOLA
TED
, 1K
, 8-P
INS
3R
P2 R
P4 R
P6 R
P7B
CN
164A
103J
7R
ES N
ET, I
SOLA
TED
, 10K
, 8-P
INS
4R
P8B
CN
164A
471J
7R
ES N
ET, I
SOLA
TED
, 470
, 8-P
INS
1R
T1 R
T2ER
T-D
2FG
L202
STH
ERM
ISTO
R, N
TC, 2
.0K
, RA
DIA
L
2
T1
PE-5
762
TRA
NSF
OR
MER
, GA
TE D
RIV
E1
U1
U2
LM31
7HV
TIC
, REG
, VO
LTA
GE
- AD
JUST
AB
LE, T
O-2
20-U
P2
U10
LT14
91C
SIC
, OP
AM
P, S
O14
1U
11LM
2594
M-1
2IC
, VO
LTA
GE
REG
, SW
ITCH
ING
, 12V
, SO
81
U12
LM25
94M
-5.0
IC, V
OLT
AG
E RE
G, S
WIT
CHIN
G, 5
V, S
O8
1U
1374
HC
T904
6AD
IC, P
LL W
ITH
BA
ND
GA
P C
ON
TRO
LLED
VC
O,
SO16
1U
14A
T89C
52-1
2JC
IC, M
CU
, 8-B
IT W
ITH
8K
FLA
SH, 1
28 R
AM
, 12M
HZ,
PLC
C44
U15
AT2
5010
-10S
CIC
, EEP
RO
M, 1
K, S
PI S
ERIA
L C
MO
S, S
O8
1
The
raso
und
3 Se
ries
Par
ts L
ist
Ref
eren
ce D
esig
nato
r(s)
Par
t Num
ber
Des
crip
tion
Q
ty/B
oard
.U
16IS
PLSI
1016
-60L
TIC
, CPL
D, I
SP, 6
4K G
ATE
S, T
QFP
441 U
17IC
L766
0CB
AIC
, VO
LTA
GE
CO
NV
ERTE
R, D
C-D
C, S
O8
1U
18TD
A70
52A
TDIC
, AM
P, 1
W A
UD
IO W
ITH
DC
VO
LUM
E C
ON
TRO
L, S
O8
U3
U4
OPA
445A
UIC
, OP
AM
P, H
IGH
VO
LTA
GE,
SO
82
U5
EL72
22C
NIC
, FET
DR
IVER
, DIF
FER
ENTI
AL,
DIP
8
1
U6
TLC
1541
CD
WIC
, AD
C, 1
1-C
H, 1
0-B
IT, S
O20
W1
U7
TLC
5620
CD
IC, D
AC
, QU
AD
8-B
IT, S
ERIA
L IN
TER
FAC
E, S
O14
1U
8TC
74H
C27
3AFN
IC, O
CTA
L D
-TY
PE F
F C
R,
SO20
W1
U9
AD
8402
AR
100
IC, D
IGIT
AL
POT,
DU
AL,
100
K, S
O14
1V
1EZ
J-S2
VB
223Z
VA
RIS
TOR
, 6V
CD
, 25A
, W/E
MI,
080
51
Y1
EC
S-11
0.5-
20-5
P
C
RY
STA
L, 1
1.05
9MH
z, 2
0pF,
SM
T1