RS-600 Operations Manual Ver. 005-10/03

RS-600 Syntron Multifunction Test SystemOperations Manual

Radian Research, Inc.3852 Fortune DriveLafayette, IN 47905

Ph: 765-447-0535Fax: 765-448-4614

Page 2: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 3: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Preface

Using This Manual

The contents of this manual provide the core procedures, software functionality, andapplication descriptions required to effectively operate the RS-600 SyntronMultifunction Test System. While it is our intent to be as detailed as possible, theflexibility of the system makes it difficult to cover every variation of functionality orapplication. Contents of this manual are subject to change without notice. If there areany questions that are not clearly or adequately addressed in this manual, please contactRadian Research for individual technical support.

This Operations Manual is divided into four sections:

Syntron System DetailsProvides a thorough description of the various RS-600 components.

Software OverviewA detailed tour of the Control Program.

Specific OperationsStep-by-step procedures for specific operations.

Test Connection DiagramsExplains how to properly make connections to the given device under test.Illustrations as well as procedures are provided as a reference.

WARNING!

Any work with the RS-600 Multifunction Test System, RS-600 accessories, energizedstandards and energized meters can present the danger of electrical shock. The RS-600and its accessories should be operated by qualified personnel. The informationprovided in this manual is intended to serve as a guide for properly qualified electricutility personnel. This manual is not intended to replace existing electric utility safetyprocedures and those listed in the Handbook for Electricity Metering.

Operation of the RS-600 should not be conducted if the work area is wet or damp.Operation should also not be conducted if flammable gases or fumes are present in thework area.

Radian Research, Inc.3852 Fortune Drive

Lafayette, IN 47905Ph: 765-447-0535Fax: 765-448-4614

Microsoft Windows® is a registered trademark of the Microsoft Corp.IB-10® is a registered trademark of the General Electric Company.

SC-10® is a registered trademark of Scientific Columbus.Microjoule® is a registered trademark of Scientific Columbus.

1234

When using the RS-600 never make voltage and currentconnections/disconnections when the system is live.

For service or repairs to the RS-600 contact RadianResearch, Inc. Do not attempt to service or makemodifications to the RS-600 due to the risk of electricalshock.

Radian Research, Inc. assumes no liability for failure tocomply with existing applicable safety precautions aswell as those listed in this warning statement.

This RS-600 Operations Manual is proprietaryinformation and is the property of Radian Research, Inc.Any unauthorized reproduction or distribution of thismanual without the permission of Radian Research isstrictly prohibited.

Page 4: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 5: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Syntron System Details

SECTION 1: Syntron System Details

RS-600 System Specifications ..............................................................................1-3

Testing Capability ..................................................................................................1-3

Normal Operating Conditions ................................................................................1-3

Physical Description ..............................................................................................1-3

RS-600 Base System Parts List .............................................................................1-4

RS-600 Overview ...................................................................................................1-5

RS-686 System Computer .....................................................................................1-5

RS-610 System Power Supply ..............................................................................1-5

RS-611 Syntron Signal Source .............................................................................1-6

RS-640 System Connection Module .....................................................................1-6

RS-770 Meter Test Socket ....................................................................................1-6

RS-771 Meter Adapter ...........................................................................................1-6

RS-735 Current Tap Changer ................................................................................1-7

RS-ACF Advanced Calibration Function Software Module ...................................1-7

RS-AWG Advanced Waveform Generation Software Module ..............................1-7

RS-600 Warranty Information ..............................................................................1-7

RS-600 System Module Diagram ..........................................................................1-9

RS-600 Connection Diagram .............................................................................. 1-11

RS-600 Syntron Multifunction Test System ...................................................... 1-13

RS-600 Desk Dimensions ................................................................................... 1-15

Page 6: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 7: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

RS-600 System SpecificationsAccuracy: Worst case ±0.02%Stability: Included within accuracy specificationPower Requirements: 120 VAC, 15 A, singlephaseSupply Frequency: 48-62 HzRecalibration Interval 365 daysMeasurement Functions Watthours, VARhours, QhoursTest Voltage: 60-600 V (1 volt increments)Test Current: 0.2-50 A (0.25 amp increments)Test Frequency: 47-63 Hz (1 Hz increments)Test Phase Angle: 0-360 degrees (1 degree increments)

The RS-600 System has been calibrated at an ambient temperature of 23°C ±5° usingstandards and instruments which are traceable to the National Institute of Standards andTechnology (NIST). The referenced set (three RM-11 Primary Standards and an RM-109Digital Comparator) which support this calibration system are calibrated on a schedulewhich is adjusted to maintain traceability at the required accuracy level.

Testing CapabilityReference Standards: Two (2) Radian Standards, four (4) with RS-640/4 option

Two (2) SC-10 Standards, four (4) with RS-640/4 optionTwo (2) IB-10 Standards, four (4) with RS-640/4 option

Watthour Meters: One (1) with the RS-771 OptionInstruments: One (1) Voltmeter with RS-ACF Option

One (1) Ammeter with RS-ACF OptionOne (1) Watt Meter with RS-ACF OptionOne (1) Phase Angle Meter with RS-ACF Option

Normal Operating ConditionsAmbient Temperature: 15°C to 35°C (Optimum = 22° to 25° C)Relative Humidity: 0 to 95%Warm-up Time: 30 minutes

Physical DescriptionSize: Desk- 81.3 cm (32 in.) Height

115.6 cm (42.5 in.) Width81.3 cm (32 in.) Depth

Base System Weight: 148 kg (325 lbs.)Shipping Dimensions: Same as overall Desk dimensions (ground shipment)

Syntron System Details

10

1-3

Page 8: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Syntron System Details

1-4

RS-600 Base System Parts List

1 Desk Work Station1 Tower

1 RS-640 System Connection Module(Voltage/Current Amplifier Panel and Data Collection Module)

1 RS-611 Syntron Signal Source1 RS-610 System Power Supply1 RS-686 System Computer

1 Pentium Computer1 101 Enhanced Keyboard1 512 Mbyte Hard Disk (Internal of Computer)1 16 Mbyte Ram (Internal of Computer)1 3.5", 1.44 MB Floppy (Internal of Computer)1 15" High Resolution Super VGA Monitor (ViewSonic)1 Mouse (Microsoft Mouse)1 Super VGA Video Card (Internal of Computer1 3.5' Computer Power Cable1 3.5' Monitor Power Cable1 Syntron/Data Collection Fiber Optic Communication Card (Internal of Computer)1 Radian Mouse Pad

1 Radian RS-600 Control Program1 Microsoft DOS Version 6.221 Microsoft Windows for Workgroups Version 3.111 RS-600 Operations Manual

2 8' Voltage Cables (Triax to Spade)2 6' Current Cables4 6' Power Supply (Internal: Triax to Triax for Power to Syntron and Data Col.)1 Power Splitter (Internal: for fans and LEDs)1 6' Main Power Cord

2,4 6' BNC to BNC Cables1,3 1' Current Jumpers3,6 2' Potential and Auxiliary Power Jumpers

1 set 16.5' Syntron Synthesizer Fiber Optic Cables1 set 16.5' Syntron Current Fiber Optic Cables1 set 16.5' Syntron Voltage Fiber Optic Cables1 set 16.5' Data Collection Fiber Optic Cables

1 Desk Work Station2 Computer / Power Supply Keys2 Cabinet Keys

Cabinets

System Modules

Computer Peripherals and Cables

Software and Documentation

System Cables and Accessories

System Cables and Accessories

Page 9: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

10Syntron System Details

1-5

RS-600 Overview

The RS-600 Syntron Multifunction Test System is an automated laboratory calibrationsystem. The core application of the RS-600 is the testing of field and primary standards.The RS-600 will test these devices with worst case accuracies of 0.02% includingstability. The RS-600 will replace existing standards, references and loading devicesso as to provide a turnkey primary reference system of exceptional accuracy andoptimum efficiency. The RS-600 supports watthour, VARhour and Qhour functionswhich enables testing of multifunction standards and solid state meters.

A major benefit of the RS-600 is its extremely flexible functionality. Althoughdesigned to serve as a primary reference system for testing field and lab standards,various upgrades and accessories allow for the testing of watthour billing meters,voltmeters, ammeters, wattmeters and phase angle meters as well.

The RS-600 can also be upgraded to allow creation of arbitrary waveforms withharmonic noise components. With the RS-600 Control Program, the operator canspecify the harmonic (1 thru 11), the axis (voltage and/or current), and the magnitude(% distortion up to 30% of fundamental). After the waveform is specified, the softwarestores it and the synthesizer will generate that particular arbitrary waveform for testingpurposes. This arbitrary waveform capability allows electric utilities to effectivelyevaluate the performance of solid state meter designs with a full range of testingcriteria.

RS-686 System Computer

The RS-686 controls the entire system and collects all test data. The RS-600 ControlProgram located on the system computer will automate almost all aspects of the testingprocess. Powerful commercial software packages will run on the computer andtransferring data to them is straightforward and quick. The RS-686 can readily beinterfaced to other computers including mainframes and networks. Permanent harddisk storage will be available if test records or other data is to be collected andmaintained within the RS-600. Self-diagnostic tests will isolate most system problemsshould they occur.

The RS-686 contains not only its software, but also the software for all the remainingmodules. Upon power-up and initiation of the control program, the RS-686 transferssoftware to all the other system modules. Updates are handled by updating only thesoftware in the RS-686.

RS-610 System Power Supply

The RS-610 supplies power to all modules except the RS-686. The RS-610 assists inproviding high accuracy in two ways. First, the RS-610 AC couples the power comingin so there is no common ground between modules. This makes it much easier to sendhigh accuracy signals to the various modules. Second, the RS-610 operates at a highfrequency (1khz) making individual power supplies much more compact and easilyshielded both of which are essential for high accuracy.

Page 10: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Syntron System Details

1-6

RS-611 Syntron Signal Source

The RS-611 Syntron Signal Source generates low level voltage and current signals ofextreme accuracy and extremely low distortion. The test frequency range is 47 to 63Hz in 1 Hz increments. The voltage amplifier can source any voltage from 60 to 600volts with a resolution of one volt increments. The current amplifier can source anycurrent from 0.25 amps to 50 amps with a resolution of 0.25 amps. The test phase anglebetween voltage and current has a range of 0 to 360 degrees in one degree increments.

RS-640 System Connection Module

The RS-640 serves two primary purposes. First, as a data collection module itinterfaces to the pulse outputs of the standards under test and transfers the data to theRS-686 computer. A card from the RS-640 connects directly to the RS-686 computerfor direct memory access. This permits rapid data collection from multiple ofstandards. The RS-640 will interface directly to the Radian RM-1H optical pickupwhich is used to sense the calibration LED on solid state meters. Also, the RS-640 willinterface with the RM-DS Meter Disk Sensor which is used to sense disk rotation whentesting induction meters. The RS-600 base system has two data collection inputchannels. The RS-640/4 upgrade increases the number of channels to four.

The second purpose of the RS-640 is as a Voltage/Current Amplifier connection panel.The voltage and current leads used to connect devices under test stem from this module.In connection diagrams that refer to the RS-640 Voltage/Current Amplifier, theconnection is to this module even though the actual voltage and current amplifiersreside in the RS-611 Syntron Signal Source module. Also any reference to a datacollection module refers to the RS-640.

RS-770 Meter Test Socket (Optional)

The RS-770 permits testing of induction and solid state meters. It features a manual testsocket compatible with all socket based billing meters. The RS-770 has provisions forvoltage and current jumpers that internally route the voltage and current applied to themeter. It comes with a jumper guide to enable quick setup and testing of billing meters.

RS-735 Current Tap Changer (Optional)

The RS-735 automatically changes the taps of manual current ranging standards such asthe IB-10 and the SC-10. It connects to the RS-686 and is controlled by the RS-600Control Program automatically during calibration. The RS-735 has no measurableeffect on error but the burden must be considered when sizing loads and currentamplifiers. The RS-735 has four outputs and a common. The four outputs typicallyconnect to the 1,5,12.5 and 50 ampere taps of the unit under test. One RS-735 is requiredfor each manual ranging standard to be tested. The RS-735 modules may be seriallyconnected together to test multiple non-autoranging standards. The RS-735 is rated for-25 to +75 degrees Celsius so that it may be used in an environmental chamber.

Page 11: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

10Syntron System Details

1-7

RS-ACF Advanced Calibration Function Software Module `(Optional Upgrade)

For kW, VAh, kVA, kVAR, mVH, mAH, V2H, A2H, volts, amps and phase anglefunctions. Enables testing of voltmeters, ammeters, watt meters and phase anglemeters. Also enables testing of Radian RM-15 Multifunction Standards and multifunctionsolid state meters.

RS-AWG Arbitrary Waveform Generation Software Module(Optional Upgrade)

For creation of arbitrary waveforms with 2nd to 11th harmonic components. Multipleharmonic components can be used to create a waveform that can be applied to either thevoltage or current axis. This capability enables testing and analysis for the effects ofharmonics on measurement accuracy, an invaluable tool in the evaluation of solid statemeter designs.

Warranty Information

Radian Research warrants the RS-600 Syntron Multifunction Test System to be freefrom defects in material and workmanship. Radian’s obligation under this warranty isto repair or replace any instrument or component therein which, within two years aftershipment, proves to be defective upon examination.

The RS-600 is warranted to be substantially stable in calibration over time. If within oneyear after factory calibration the RS-600 does not meet its specifications, RadianResearch will repair and recalibrate the unit at no charge.

This warranty is made void by disassembly or any of the individual modules from thesystem. If service is required, contact your local Radian Research representative or theRadian Research headquarters in Lafayette, Indiana.

Page 12: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Syntron System Details

1-8

Page 13: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 1-9

RS-600 System Module DiagramSyntron Multifunction Test System

10

Page 14: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 15: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 1-11

RS-6

00 C

onne

ctio

n Di

agra

mSy

ntro

n M

ultif

unct

ion

Test

Sys

tem

Page 16: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 17: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 1-15

RS-600 Syntron Multifunction Test System10

Page 18: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 19: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 1-17

RS-600 Syntron Multifunction Test SystemDesk Dimensions

10

Page 20: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 21: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Section 2: Software Overview

Introduction to the RS-600 Control Program ......................................................... 2-3

Useful Microsoft Windows Terminology ................................................................. 2-4

Core Concepts in the RS-600 Control Program ...................................................... 2-4

Configuration View .................................................................................................. 2-6

Channel Table View ............................................................................................... 2-10

Device View ........................................................................................................... 2-11

Test View ............................................................................................................... 2-14

Test Group View .................................................................................................... 2-16

Report View ........................................................................................................... 2-17

Autonull View ........................................................................................................ 2-18

File Open / File Save as / File Delete Browsers ....................................................... 2-18

Control Program Menu Structure .......................................................................... 2-20

Diagnostics View............................................................................................................. 2.20

Menu Item Overview ............................................................................................. 2-21

Toolbar Icon Descriptions .................................................................................... 2-23

Page 22: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 23: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

Introduction to the RS-600 Control Program

The RS-600 Syntron Multifunction Test System provides a multitude of testing capabilities.The Control Program is the part of the system that makes the RS-600’s advanced capabilitiesbecome a reality. The Control Program puts the power of custom test program creation in thehands of the operator. The Microsoft Windows-based Control Program provides multitaskingcapability for increased flexibility and efficiency. In addition, the Control Program containsenhanced graphically oriented operations allowing for increased user-friendliness and simplifiedtest program creation.

If the operator of the RS-600 is new to computers then he/she will find this manual easy to followand the Control Program quick to master. If the operator is an experienced computer user thenhe/she will appreciate the logical design of the software and the power of its multitaskingcapabilities. Nevertheless, regardless of the operator’s computer experience, he/she will findthe RS-600 to be an effective as well as powerful laboratory instrument. By design, the powerof this laboratory instrument is at the discretion of the operator via the Control Program.

Microsoft Windows®

Microsoft Windows® is a software package that provides a graphical environmentfor the operation of the computer. Windows provides multitasking capabilitieswhereby more than one operation or program can be operating at the same time. WithMicrosoft Windows®, whatever work is in progress on the computer screen isdisplayed in a “window.” These “windows” may be maneuvered around on thecomputer screen and they may be maximized (enlarged) or minimized (shrunk) by theoperator. The RS-600 Control Program operates under the Windows environment forefficient use of the computer’s capabilities as well as to enhance the simplicity of thesoftware. It is assumed that the user of the RS-600 Control Program has a rudimentaryunderstanding of the Microsoft Windows® operating environment. While great carehas been taken to give as detailed instruction as possible, a detailed overview ofWindows conventions and functionality will not be given. For in-depth instructionon Microsoft Windows®, please consult the operations manual that came with youroriginal Windows diskettes or seek other avenues of Windows training.

The RS-600 Control Program is based on the Windows 95/98 operating systems. Effortsare in process to implement a RS-600 Control Program that will work with Windows 2000,Windows XP, etc. As development advances in this area, all applicable Radian RS-600customers will be notified.

2-3

20

Page 24: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

Useful Microsoft Windows Terminology

WindowThe primary convention for user interface screens within a Microsoft Windowsbased software program. Windows can be scaled and moved around the screen. Themultitasking environment supported by Windows allows multiple windows to beopen at one time.

DialogA type of window that either solicits some information from the user or providesa message of some sort. Dialogs are also characterized by their inability to bescaled in size.

ViewA generic term that refers to either a specific window or dialog within the ControlProgram.

File BrowserA browser is a special type of Windows dialog which enables the user to “browse”the file and directory structure of the computer’s disk drive. Browsers are used toOpen, Save, and Delete various file types and directories within the ControlProgram.

Toolbar IconsGraphical buttons arranged across the top of a program’s Main View that give one-step access to some of the core menu options. Every function performed by aToolbar Icon can also be performed by a menu selection.

ClickRefers to a single push of the left mouse button coinciding with specific placementof the cursor on a button or area of the screen.

Double ClickRefers to pushing the left mouse button two times in rapid succession whilepointing to a specific area of the screen with the cursor. Core Concepts in the RS-600 Control Program

Core Concepts in the RS-600 Control Program

Each of the following entities has a primary “view” associated with it. From these viewsvarious parameters of the entity may be entered, modified, saved, or acted upon withfunctionality specific to that type of entity.

DevicesDevice parameter files that define a type of electrical device that will be testedwith the RS-600. Examples of Devices are standards, revenue meters, andvoltmeters. Creation of new and editing of existing Devices is accomplishedrespectively via the File New/File Open Device menu selections. The DeviceView can also be accessed by clicking on the New Device or Open Device toolbaricon.

2-4

Page 25: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

TestsTest parameter files with “spreadsheet like” test grid for defining the combinationof voltage and current points that the test will encompass. Creation of new andediting of existing Tests is accomplished respectively via the File New/File OpenTest menu selections. The Test View can also be accessed by clicking on theNew Test or Open Test toolbar icons.

Test GroupsFiles that contain logically grouped individual Tests that can be run together.Creation of new and editing of existing Test Groups is accomplished respectivelyvia the File New/File Open Test Group menu selections. The Test Group View canalso be accessed by clicking on the New Test Group or Open Test Group toolbaricons.

Variations of the above entities are stored as files by the RS-600 Control Program.To distinguish between a file and a physical item, all file references will becapitalized in this manual.

ConfigurationShows and allows setting of all global parameters and preferences for the RS-600Control Program. Accessed via the Options Configure menu selection.

ChannelRepresentation of an input channel on the RS-640 Data Collection Module. Aftera device that is to be tested is physically connected to the system cables, it mustalso be “connected” via the Channel Table View in the RS-600 Control Program.The Channel Table View is accessed via a View Channel Table menu selection orby clicking the Channel Table toolbar icon. The Channel Table View is also theview open when the Control Program is first launched, reminding the user thatdevices to be tested with the RS-600 must first be connected to a channel. By“connecting” or specifying a particular device to a particular channel in theChannel Table, the RS-600 Control Program will know the characteristics of thatgiven device. The characteristics include the pulse constant, min/max currentsand min/max voltages.

ReportA view of past test result data stored according to the serial number of the deviceand further filtered by the exact times and dates the tests occurred. Reports arealso printable directly from the Report View that is accessed via a View Reportmenu selection or by clicking the Report View toolbar icon.

AutonullAutonull is a self-maintenance procedure that the RS-600 system performs on aregular basis to clear any traces of transformer core saturation that may haveoccurred through exposure to DC currents.

DiagnosticsSystem diagnostics is a tool that allows for troubleshooting and diagnostics ofthe various RS-600 system modules. If error codes are received during testing,the Diagnostics View may be instrumental in determining the cause of said error.

2-5

20

Page 26: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

2-6

Configuration View

The configuration view shows and allows for the setting of all global parameters and preferencesfor the RS-600 Control Program. The configuration window appears as below and is accessedthrough the Options - Configure menu selections of the RS-600 Control Program.

Result Tab

Result Display

Defines how test result information will be displayed in the Test View. This selectionalso defines exactly in what form the test results will be saved. Once data is savedusing a certain data display format, there is no way to easily change the format of thatdata.

• Percent Error will show the percent of error with 0.0% being no error.• Percent Registration will show percent of registration with 100.0% being no

error• Correction Factor will show the correction factors that need to be applied for

the device under test to reach zero errors.• Measured Value will show the actual number of pulses counted or the reading

that was manually entered from a device under test.• Calibration Value

Defines the number of decimal digits to display the results in.

• 2 will display as 0.00• 3 will display as 0.000

Result Digits

Page 27: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 2-7

20Print results at end of each test

As a test is completed, the results will be printed to the default printer. The defaultprinter is set from the File-Printer setup menu.

Save results to floppy at end of each test

As a test is completed, the results will be saved to Drive A:\ under the file name <ser/num>.dat where ser/num is the serial number of the device under test. This saving process isin addition to the default saving of test results to the RS-600 system computer’s hard disk.

Point Order

Defines the order of how the test points are traversed on the screen.

• Col moves to the test points in a down, then up, serpentine fashion. Col isselected to reduce the number of voltage changes.

• Row moves to the test points in a horizontal fashion. Row is selected toreduce the number of current changes.

Time Out Multiplier

This is a multiplier selected and used to increase the test time beyond that as specifiedin the test sheet. This feature is automatically activated if an inadequate number ofpulses have been counted to provide a reliable accuracy measurement for the devicebeing tested.

Test Tab

By clicking on the Test tab, the configuration view changes as shown below.

Software Overview

Page 28: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Hardware Tab

By clicking on the Hardware tab, the configuration view changes as shown below.

Optional Equipment

Informs the RS-600 Control Program of optional equipment being used.

• Selecting Tap Changers informs the RS-600 system that RS-735 Current TapChangers are being used. Tap Changers are used for automatically re-directing current to the proper taps on devices that do not supportautomatic current ranging. This selection must be checked if Tap Changerswill be used with the RS-600 system.

• Selecting Meter Adapter informs the RS-600 system the RS-770 MeterAdapter is being used. This selection must be checked if the RS-770 MeterAdapter is used with the RS-600.

Phase View

Defines how to display the phase angle difference between the voltage and currentwaveforms on the Test View.

• Lead/lag will display the phase difference in terms of Unity, Leading, orLagging values.

• Degrees will display the phase difference in terms of degrees.• Radians will display the phase difference in terms of radians.

Pause at the end of each test

At the completion of each test, the system will pause and display a dialog to inform you thetest has completed. If a Test Group was run, the next test will not start until you acknowledgethe pause.

Software Overview

2-8

Page 29: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Comm Tab

By clicking on the Comm tab, the configuration view changes as shown below.

Comm Enable

Selecting Comm Enable allows communications between the RS-600 and the standardunder test.

Communication cable types

Informs the RS-600 of the type of communication cables being used.

• Selecting RM-PCA Rev 1 cables allows for the use of the RM-PCA Revision 1cables that have a gray band on the connector end.

• Selecting RM-PCA Rev 2 cables allows for the use of the RM-PCA Revision 2cables that have a blue band on the connector end.

• Selecting RD-xx direct allows for the use of RS-232 connection between the RDstandards and the RS-600 system.

Communication Loop Testing

This is an in-house Radian Research testing tool that the customer should not haveaccess to. If access is allowed, the customer should not select it because it will slowdown testing due to a tremendous amount of data tying up the communication serialport.

Select Comm Ports

Allows the user to select which comm port on the computer they wish to use for communicatingto the device under test.

Software Overview

20

2-9

Page 30: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Channel Table View

The Channel Table is used to inform the RS-600 system which devices are connected tothe RS-640 System Connection Module channels. It is always visible along the bottomof the screen at initialization of the RS-600 Control Software. If desired, the ChannelTable may be repositioned by clicking on it then dragging it to a new location.

Software Overview

2-10

Configuring the Channel Table

1. Click on the numbered button of the channel you want to connect a deviceto. This will bring up a Select Device browser. Select the Device thatmatches the physical device you wish to test and click the OK button.

2. Type in the Serial number of the device.

3. If you are using RS-735 Current Tap Changers select the Tap Changercheck box.

Clicking the Edit button will open the Device View window for the particulardevice selected and allow for the changing of the device properties for theselected device. Clicking on the Disconnect button will disconnect theselected device from the RS-600 Control Program. Clicking on the Help buttonwill launch the help program.

The Comm Connections window will show the devices connected to the RS-600 systemand clicking on the Refresh button will update the communication connections betweenthe RS-600 and the devices connected to it.

Page 31: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

2-11

Device ViewThe creation of new devices is accomplished via the File-New-Device menu selections.The editing of existing devices may be accomplished by the File-Open menu selections.Clicking on the Open Device toolbar icon will also provide access to the Device View.The Device window for creating a new device appears as below.

20

Setup tab

Device Type

Tells the RS-600 program what type of device this is.

Phases

Defines the number of phases this device uses.

Minimum Pulse Count

Specifies the minimum number of pulses needed for this device to have a valid test.

Testing Method

Defines whether the device will be connected to the RS-640 Data Collection Modulefor automatic reading of the pulse count or if the reading will be manually entered bythe operator when prompted.

Pulse Output Pullup

Specifies the resistor size to use for the RS-640 Data Collection Module connectionto ensure a proper pulse reading. All Radian products, including pulse pickupaccessories and watthour standards, utilize a pullup resistor value of 150 ohms. Non-Radian products should have a pullup resistor value set at 1,000 ohms.

Page 32: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/032-12

Software Overview

Standard OptionsSpecial settings for devices with limitations not present in the Radian Research lineof electronic watthour standards

• Continuous voltage is for devices that need the voltage to stay applied continuouslyor stabilization. Normally, between test points the voltage and current are rampeddown, as the new point is being setup. With Continuous Voltage checked, the voltagewill stay applied unless the voltage or power factor is changing for the next point.

• High DC Sensitivity is for devices that will report erroneous results if DC is present.This tells the system to clear any DC before running the test point.

• Run Autonull before test is a maintenance procedure that should be performed on aregular basis to clear any traces of transformer core saturation that may have occurredthrough exposure to DC currents. The Autonull procedure takes approximately 5minutes to run.

Ramp Rates

Specifies the amount of time for ramping voltage and current levels.

Ranges tab

By clicking on the Ranges tab, the Device View window will change as shown below. The Rangeswindow is where the voltage and current test parameters are set for a device.

Page 33: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 2-13

Software Overview

Current Ranging

Defines if the device can support all currents automatically or if the device requirestap settings for different current ranges.

• Auto should be selected if the device is autoranging and only the maximumcurrent the device can support needs to be specified.

• Manual should be selected if the device requires tap settings. Nominal andMaximum currents for each tap setting need to be specified. Nominal settingsshould be the exact tap values shown on the physical device.

Maximum settings are the maximum current values you want to allow for thatparticular tap while testing the device. This distinction allows you to test a devicewith a higher current than is indicated on a particular tap.

Voltage Ranging

Defines if the device can support all voltages automatically or if the device requiresmanual tap settings for various voltage ranges.

• Auto is selected if the device is autoranging and only the maximum voltage thedevice can support needs to be specified.

• Auto With Taps is selected if the device is autoswitching and therefore uses tapsinternally.

• Manual should be selected if the device requires external tap settings. If thedevice requires tap settings (manual or auto), then you need to specify theNominal and Maximum voltages for each tap setting. Nominal settingsshould be the exact tap values shown on the physical device.

Maximum settings are the maximum voltage values you want to allow for that particular tapwhile testing the device.

20

Page 34: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

2-14

Functions tab

By clicking on the Functions tab, the Device View window will change as shown below. Onlyfunctions supported by the RS-600 system purchased will appear in the function window.Additional measurement functionality may be procured as a system upgrade.The Function window is where the parameter to be tested along with its pulse factor isselected.

Test View

The Test View window is used for defining and running tests on the devices connected tothe Channel Table. The Test allows for current to vary on one axis and for the Power Factorand Voltage to vary on the other axis. Creation of new and editing of existing Tests isaccomplished respectively via the File New/File Open Test menu selections. The Test Viewmay also be accessed by clicking on the Open Test toolbar icon.

As the test is running, the results and status are updated on the display. If multiple devicesare being tested simultaneously, different channels can be viewed via the View ChannelsChannel # menu choice. Only one channel can be viewed at a time even though multipledevices can be tested simultaneously. Following is an example and overview of theparameters on the Test View.

Page 35: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

Power FactorClicking on any of the Power Factor value buttons will cause a window to appear allowingthe operator to specify a desired value for the phase difference between the voltage andcurrent waveforms. The value can be entered directly either in degrees or radians. Thetwo common settings of Unity and 0.5 lagging Power Factor can also be selected byclicking on the appropriate choice.

VoltageClicking on any of the Voltage value buttons will cause a drop down checklist to appearallowing the operator to specify a desired value. The most common choices are availablefor checking. However, any voltage within the RS-600 operating range and resolutioncan be entered by typing the value in on the keyboard and pressing Enter.

CurrentClicking on any of the current value buttons will cause a drop down checklist to appearallowing the operator to specify a desired value. The value can be selected by clickingon it or by directly typing it in and pressing Enter.

TapClicking on any of the Tap buttons will cause a drop down checklist to appear allowingthe operator to specify the desired Current Tap to be used for that particular test point.This is applicable when testing devices with Manual Current Ranging settings in theirDevice View.

FunctionsClicking on this button causes a drop down scroll list to appear allowing the operatorto specify the measurement function being tested for this particular Test. Scrolling tothe desired function then clicking selects that function. Note the distinction betweenthis Function selection and the one encountered in the Device View. Here only onefunction can be selected per Test. In the Device View, all functions supported by thedevice are selected.

Pulse ConstantIf a value is specified by the operator in this field then it will override the value specifiedin the Device View parameters. By leaving the value in this field set a the default 0.0000setting the pulse constant will be automatically read from the Device parameters of thedevice under test. As a general rule to eliminate confusion, the Pulse Constant on theTest View should never be changed from this default setting.

Stabilization TimeIn this field the operator can specify the amount of time that the device is allowed tostabilize before actually running the test. The following are recommended values forvarious standards: RM-10 = 15 seconds typical; RM-11 = 15 seconds typical; SC-10 =60 seconds minimum; Rotating standards = 15 seconds typical.

Test TimeIn this field the operator can specify the amount of time that pulses will be counted fromthe device under test. The following are recommended values for various standards: RM-10 = 30 seconds; RM-11 = 30 seconds; SC-10 = 60 seconds; Rotating standards = 60seconds.

2-15

20

Page 36: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/032-16

Software Overview

Run/StopRun begins execution of the Test shown in the active Test View. The Stop button replacesthe Run button when a Test is actually running. Clicking on Stop ends the test and promptsthe operator to save the results data.

CloseWill close the Test View window.

HelpWill bring up the RS-600 Help screen.

Test Group View

Test Groups give the ability to create logical groups of tests then run them together. This isbeneficial when a single test sheet is not large enough for all your desired test points or fortesting different functions on the same device. Currently, only one function can be tested pertest sheet, requiring multiple test sheets to fully test a multi-function device. Creation of newand editing of existing Test Groups is accomplished respectively via the File New/File OpenTest Group menu selections. The Test Group View can also be accessed by clicking on theOpen Test Group toolbar icon. The Test Group window appears as follows.

Adding a TestClick on the Available Tests list box in the center. Traverse directories with the directoriesbrowser on the left of the Test Group View dialog.

Removing a TestClick on the test in the Selected Tests list box on the right.

Saving a Test GroupSelect the File Save menu item or click on the Save toolbar icon. This will bring up a SaveBrowser allowing you to name and save the Test Group.

Running the Test GroupClick on the Run button at the bottom of the Test Group View dialog. This dialog will remainopen in the background as the Test View showing the first Test of the group appears andtesting begins.

Running Autonull with Test GroupClick on the Autonull button.

Page 37: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03 2-17

Software Overview

Report View

This feature is used to view and print saved test results. Access this view from the menuitem View Report or via the View Test Results toolbar icon. The Report View appears asfollows.

20

Using the Report View

1. Select the Serial Number of the device you wish to view/print by clicking in the SerialNumber box and choosing from the drop down list that appears.

2. Select the Date of the test that you wish to view/print by clicking on the Date boxand choosing from the available dates in the drop down list that appears.

3. To print test results, simply click on the Print button at the bottom.

4. To see a graphical representation of the data, click on the Graph button.

5. To close the window, click on the close button.

6. To access the help files, click on the Help button

Page 38: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Autonull View

Autonull is a self-maintenance procedure that the RS-600 system performs on a regularbasis to clear any traces of transformer core saturation that may have occurred throughexposure to DC currents. The Autonull procedure takes approximately 5 minutes to run.The last Autonull date is recorded and shown at the top of the Autonull. Autonull can beuser initiated at any time via the Autonull View dialog accessed by the Options/Autonullmenu choice.

Software Overview

2-18

Diagnostics View

System diagnostics is a tool that allows for troubleshooting and diagnostics of the variousRS-600 system modules. The Select All button may be used to perform diagnostics on allmodules. By clicking in the small gray box following each module’s name, individualmodules may be selected. The De-select All button will de-select modules selected fordiagnostics. Run will start the diagnostics process for those modules selected.

Page 39: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

File Save as allows an open Device, Test or Test Group to be saved to a new name. Thecurrent Device, Test or Test Group isnot changed. dialog allows you tocreate a new directory to save theobject. Select this button and type inthe new directory name. The newdirectory will be created under thecurrent directory.

File Delete allows a Device, Test orTest Group to be deleted.

Caution: deletions can not be undone.

The Dir... button in the File Delete dialog allows you to delete a directory. Highlight thedirectory you wish to delete and press the Dir... button.

Caution: Everything in the directory will be deleted also.

20Software Overview

2-19

File Open /File Save as /File Delete Browsers

A browser is a special type of Windows dialog which enables the user to “browse” the file and directorystructure of the computer’s disk drive. Browsers are used to Open, Save, and Delete various file types anddirectories within the Control Program. Directories are traversed by double clicking on names in the Directoriesbox on the right side of the browser. The selected directory appears above the box and all files within that

directory are then displayed in the Filesbox on the left side of the browser.Below the Files box on the left side is a“List Files of Type” drop down listwhich allows the selection of eitherDevice, Test, or Group files. Clickingon a file name selects it.

File Open allows the selection of anexisting Device, Test or Test Group.Choosing this selection will open anexisting View associated with this typeof item.

Page 40: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Control Program Menu Structure

Software Overview

2-20

Page 41: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Menu Item Overview

File NewCreates a new window of the type selected; Device, Test, or Test Group. After editing,you will be prompted to save changes to a new file from the File Save As browser.

File OpenBrings up the File Open browser to allow selection of an existing Device, Test, or TestGroup to open.

File CloseCloses the current window. If any changes had been made that were not saved, a promptwill appear asking whether to save the changes that have been made.

File SaveWill save changes to the current window.

File Save AsAllows saving of the current window Device, Test, or Test Group to a new name. Theexisting name will not change nor be deleted, but a copy of the object is created and anew name is assigned. The File Save As browser is invoked to allow selection of thenew name and directory. This feature can be used to easily create new Device, Test, orTest Group files that are slight variations of existing ones.

File PrintPrints the current report to the printer. This is only available for Reports of testresults. First select the menu item “View Reports” to display the test results.Specify the Serial number and test date of the results you want printed. Then selectPrint either from the menu, toolbar, or the print button on the Report View dialog.

2-21

Software Overview

20

File DeleteBrings up the File Delete dialog for selection of the file or directory to be deleted. Beforedeleting, you will be prompted to confirm the deletion.

File Printer SetupSelects and sets up a printer to use for printing of test result Reports.

File ExitWill close all open windows and exit the application. If any data has not been saved,you will be prompted to save or discard the data first.

View ReportDisplays the Report View where past test results can be easily viewed and/or printed.

View ChannelAllows the selection of which channel to be displayed on the Test View while a test isrunning or completed. Only channels that have devices connected to them via theChannel Table will be available for viewing. If an average was specified in the ChannelTable it can be viewed as well.

Page 42: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Software Overview

2-22

Test Run/StopBegins execution of the Test shown in the active Test View. It can also be triggered byclicking on the Run button located at the bottom of the Test View window. This menuselection is only available when a Test View is the active window. When the Test isrunning, the menu item becomes Test Stop and the Run button becomes Stop. Selectingeither of these options will cause the Test to stop and the operator will be prompted tosave the test results.

Test Continuous RunWill start the execution of the Test in the active Test View and will continue running thetest indefinitely until Test Stop is selected from the menu or the Stop button in the TestView window is clicked. This menu selection is only available when a Test View is theactive window.

Test ClearSelecting Test Clear from the menu will clear the results of the active Test. If the test isrunning it will be stopped and the data cleared. You will have the option of saving theexisting data before it is cleared from the active Test View. This menu selection is onlyavailable when a Test View is the active window.

Test Pause/ContinueIf a test is running, selecting Test Pause will pause the execution of the active runningtest and ramp the voltage and currents down to zero. This menu item will then changefrom Pause to Continue. Selection of Test Continue menu item will resume the test at thelast test point prior to the initial Pause command. Can also be triggered by clicking on thePause/Continue button located at the bottom of the Test View window. This menuselection is only available when a Test View is the active window.

Test Lock/UnlockIf a test is running, selecting Test Lock will freeze the test at the present test point andleave the voltage and current at the ramped level. This menu item will then change to TestUnlock. Selection of Test Unlock will then resume the test at the last test point prior tothe initial Lock command. Can also be triggered by clicking on the Lock/Unlock buttonlocated at the bottom of the Test View window. This menu selection is only available whena Test View is the active window.

Caution: while a test is locked the voltage and current will be ramped up.

Test Skip PointWhen a test is running, selecting Test Skip Point from the menu will skip the test pointcurrently running and begin testing the next point. Can also be triggered by clicking onthe Skip button located at the bottom of the Test View window. This menu selection isonly available when a Test View is the active window.

Options ConfigureSelecting the Options Configure menu item opens the Configuration View window. Thisview allows the configuration of global preferences and parameters for your RS-600system.

Page 43: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Options AutonullSelecting the Options Autonull menu item opens the Autonull View dialog. This view allowsyou to initiate the Autonull procedure and displays the date when the last Autonull wasperformed.

Options Re-InitializeThis selection repeats the system initialization procedure which normally occurs each timethe RS-600 Control Program is launched.

Options DiagnosticsThis selection is used to perform various diagnostics on RS-600 system modules.

Options CalibrateThis provides the system calibration functionality for your system and is not available foryour use.

Window CascadeRearranges all open windows into a cascaded tier as is standard in most Windowsapplications.

Help IndexBrings up the help system to the main help level.

Help Getting StartedBrings up the help system to the Getting Started topic.

Help Using HelpBrings up help on how to use the help system.

Help AboutBrings up a dialog that describes this software and release version.

Toolbar Icon Descriptions

The Toolbar of the Control Program provides one button access to many of the commonmenu commands. Every toolbar icon corresponds to a specific menu choice.

Going from left to right, the menu functions for the toolbar icons are as follows.

1. Open a Device 2. Open a Test 3. Open a Test Group 4. View Test Results 5. Save Changes 6. Print

Software Overview

20

2-23

Page 44: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 45: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

SECTION 3: Specific Operations

Create a Device ........................................................................................................ 3-3

Create a Test ........................................................................................................... 3-3

Create a Test Group ................................................................................................. 3-4

Run a Test ............................................................................................................... 3-4

View and Print Saved Test Results .................................................................................. 3.6

Page 46: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 47: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Create a Device

1. To create a new device, select the File New Device menu item. This will bring up a newDevice View window with the parameters that define a device.

2. By clicking on the Setup, Ranges, and Functions tabs, fill in the parameters for the Device.See Device View in the Software Overview section for details.

3. Select File Save menu item. This will prompt you with a Save browser to name andsave the Device accordingly.

Specific Operations

3-3

Create a Test

1. To create a new test, select the File New Test menu item. This will bring up a new TestView window with the parameters that define a test.

2. The new Test View will open with parameters already filled in. To change theparameters, simply click on the parameter and enter a new value. See Test View inSoftware Overview section for details.

3. Select File Save menu item. This will prompt you for the test name to Save As.

30

Page 48: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Specific Operations

Create a Test GroupA Test Group is a way to create logical groups of test sheets that can be saved then runtogether.

1. To create a new test group, select the File New Test Group menu item. This will bringup a new Test Group View window.

2. Select the Tests you want to include in the Test Group by clicking on the Test namesin the Available Tests box on the left of this dialog. Tests can be chosen from anydirectory by traversing the directory browser in the middle section of this dialog. AsTests are selected they are immediately added to the Selected Test box on the rightside of the dialog.

3. Select File Save menu item. This will prompt you for the Test Group name to save as.

3-4

Run a Test

1. Connect device(s) to the RS-640 System Connection Module. For more specific instructionssee Test Connection Diagrams in section 4 of this manual.

2. Configure the Channel Table with the appropriate Devices corresponding to the physicaldevice connection made in above in step #1.

Page 49: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Specific Operations

3-5

3. Select a Test to run from the File Open.

4. Click the Run button on the Test View screen. The results and status of the test will beupdated on the display.

30

Page 50: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Specific Operations

3-6

View and Print Saved Test Results

1. Select View Report from the menu or click on the View Test Results toolbar icon to bringup the Report View window.

2. Select the Serial Number of the device you which to view or print by clicking in the SerialNumber box and choosing from the drop down list that appears.

3. Select the Date of the test that you wish to view/print by clicking on the Date box andchoosing from the available dates in the drop down list that appears.

4. To print, click on the Print button at the bottom.

Page 51: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

SECTION 4: Test Connection Diagrams

Testing a Radian RM-10 or RM-11 Standard .......................................................... 4-3

Testing Multiple Radian RM-10 or RM-11 Standards ............................................. 4-4

Testing a Multifunction Radian Standard ............................................................... 4-5

Testing a Rotating Standard ................................................................................... 4-6

Testing Several Multifunction Radian RM Standards .............................................. 4-7

Testing a Non-autoranging Electronic Standard ..................................................... 4-8

Testing Several Non-autoranging Electronic Standards ......................................... 4-9

RS-771 Meter Adapter Operation .......................................................................... 4-10

Single-Phase Meter Adapter Current Cable Chart ................................................... 4-11

Testing an Induction Meter ................................................................................... 4-12

Testing Solid State Meters from Infrared Calibration LED .................................... 4-13

Testing a Solid State Meter from its KYZ Output ................................................... 4-14

Testing a Microjoule ............................................................................................. 4-15

Page 52: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Page 53: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

Testing a Radian RM-10 or RM-11 Standard

1. Connect the BNC cable from the Data Collection Module channel to the output of thestandard. In this manner the Data Collection Module will receive pulses from thestandard during test.

2. Connect the current leads from the Current Amplifier to the current input of thestandard.

3. Connect the potential leads from the Voltage Amplifier to the potential input of thestandard. Note to observe polarity when making this connection.

4. Connect the Auxiliary Power input of the standard to a 120 volt wall outlet.

4-3

40

Page 54: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Testing Multiple Radian RM-10 or RM-11 Standards

1. Connect the BNC cable from the first channel of the Data Collection Module to the outputof the first standard. Connect the BNC cable from the next channel of the Data CollectionModule to the output of the second standard. In this manner the two channels of theData Collection Module will receive pulses from the standards during the test.

2. Connect the current leads from the Current Amplifier to the standards in series.

3. Connect the potential leads from the Voltage Amplifier to the potential input of the firststandard. Jumper from the potential input of the first standard to the next standard. Noteto observe polarity when making this connection.

4. Connect the Auxiliary Power input of the standard to a 120 volt wall outlet. Jumper fromthe Auxiliary Power input of the first standard to the next standard. Note the polarity whenmaking this connection.

Test Connection Diagrams

4-4

Page 55: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

Testing a Multifunction Radian Standard

1. Connect the BNC cable from the Data Collection Module channel to the output of thestandard. In this manner the Data Collection Module will receive pulses from thestandard during the test.

2. Connect the current leads from the Current Amplifier to the potential input of thestandard. Note to observe polarity when making this connection.

3. Connect the Auxiliary Power input of the standard to a 120 volt wall outlet.

4. The Select button on the standard is used to choose the measurement mode.

4-5

40

Page 56: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Testing a Rotating Standard

1. Connect the Current Tap Changer to the current inputs of the standard making sure thatthe communications cable is also connected to the Data Collection Module.

2. Connect the negative current lead from the Current Amplifier to the first current tapof the standard.

3. Connect the positive current lead from the Current Amplifier to the Current TapChanger.

4. Connect the potential leads from the Voltage Amplifier to the potential input of thestandard. Note to observe polarity when making this connection.

5. Connect the BNC of the RM-DS to the Data Collection Module.

6. Connect the AC adapter into a 120 volt wall outlet.

7. Affix the suction cup mount of the RM-DS to the face of the standard. Align the RM-DS so that it will trigger (red and green LEDs should flash) as the dial hand rotates.Note that there is one red and one green LED on the RM-DS. When both the red andgreen flash it signifies that the RM-DS is operating at the optimum level. The screwon the RM-DS can be used to adjust the signal strength. It is recommended that thesupplied screwdriver be used in order to not damage the adjustment screw.

Test Connection Diagrams

4-6

Page 57: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

Testing Several Multifunction Radian RM Standards

1. Connect the BNC cable from the first channel of the Data Collection Module to the outputof the first standard. Connect the BNC cable from the next channel of the Data CollectionModule to the output of the second standard. In this manner the two channels of theData Collection Module will receive pulses from the standards during the test.

2. Connect the current leads from the Current Amplifier to the standards in series.

3. Connect the potential leads from the Voltage Amplifier to the potential input of the firststandard. Jumper from the potential input of the first standard to the next standard. Noteto observe polarity when making this connection.

4. Connect the Auxiliary Power input of the standard to a 120 volt wall outlet. Jumper fromthe Auxiliary Power input of the first standard to the next standard. Note to observe thepolarity when making this connection.

5. The Select button on the standards is used to choose the measurement mode they arein. When testing multiple standards in this manner make sure that all standards are inthe same measurement mode.

4-7

40

Page 58: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Testing a Non-autoranging Electronic Standard

1. Connect the Current Tap Changer to the current inputs of the standard.

2. Connect the negative current lead from the Current Amplifier to the first current tap ofthe standard.

3. Connect the positive current lead from the Current Amplifier to the Current Tap Changer.

4. Connect the BNC cable from the Data Collection Module channel to the Current TapChanger. Also connect the communications cable to the same Data Collectionchannel.

5. Connect the potential leads from Voltage Amplifier to the potential input of thestandard. Note to observe polarity when making this connection.

6. Jumper from the potential input on the standard to the Auxiliary Power input on thestandard.

Test Connection Diagrams

4-8

Page 59: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

Testing Several Non-autoranging Electronic Standards

1. Connect the first Current Tap Changer to the current inputs of the first standard. Connectthe second Current Tap Changer to the current inputs of the second standard.

2. Connect the negative current lead from the Current Amplifier to the first current tap ofthe first standard.

3. Connect the Current Tap Changer of the first standard to the first current input of thesecond standard.

4. Connect the positive current lead from the Current Amplifier to the Current TapChanger on the second standard.

5. Connect the BNC cable from the first channel of the Data Collection Module to thefirst Current Tap Changer. Do the same with the communications cable.

6. Connect the BNC cable from the second channel of the Data Collection Module to thesecond Current Tap Changer. Do the same with the communications cable.

7. Connect the potential leads from Voltage Amplifier to the potential input of the firststandard. Jumper from the potential input of the first standard to the second standard.Note to observe polarity when making this connection.

8. Jumper from the potential input on the first standard to the Auxiliary Power input onthe first standard. Jumper from the Auxiliary Power input of the first standard to thesecond standard. Note to observe polarity when making this connection.

4-9

40

Page 60: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

4-10

RS-770 Meter Test Socket Operation

1. Connect the potential leads from the RS-640 Voltage Amplifier to thepotential input of the meter test socket. Make voltage jumper connectionson the test socket according to the chart for the proper meter form.

2. Connect the current leads from the RS-640 Current Amplifier to thecurrent input of the meter test socket. Make current jumper connectionson the test socket according to the chart for the proper meter form.

3. The meter form should be specified in the Device View of the ControlProgram.

Page 61: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Form Link Position Red (+) Black (-)1S - D C2S OPEN R C3S - D L*3S (Alt) - D M4S - B M5S - W/X Y/Z6S - L/M B/Z8S - L/R M/Z9S - L/M/R Z10S - L/M/R X/B/Z12S OPEN L C/R*12S (Alt) OPEN M C/R13S - W/X Y/Z14S OPEN L/R Z15S OPEN L/R M/Z16S OPEN L/M/R Z26S - L/M B/Z

Form Red (+) Black (-) Jumper Jumper1S E C - -2S D C E/F -3S C E - -*3S (Alt) C E - -4S D C E/F -5S C F E/D -6S C F E/Y W/D8S C F E/W Y/D9S C F E/W Y/D10S C F E/W Y/D12S F C E/D -*12S (Alt) F C E/D -13S C F E/D -14S C F E/Y W/D15S C F E/Y W/D16S C F E/W Y/D26S C F E/D -

* (Alt): Alternate positions of movable voltage terminal. The * indicates the voltage terminal in the 9 o’clock position.

Test Connection Diagrams

4-11

40Single-Phase Meter Test Socket Current Cable Chart

PotentialPotentialPotentialPotentialPotential Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing

CurrentCurrentCurrentCurrentCurrent Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing Jumper Connections for Single-Phase and Series Single-Phase Testing

Page 62: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

4-12

Testing an Induction Meter

1. Connect the Voltage Amplifier leads to the appropriate inputs of the meter test socket.

2. Connect the Current Amplifier leads to the appropriate inputs of the meter test socket.

3. Jumper additional current and voltage inputs appropriately on the meter test socketaccording to the Meter Test Socket Connection Diagram.

4. Connect the BNC of the RM-DS to the Data Collection Module.

5. Connect the AC adapter to the RM-DS and plug the adapter into a 120 volt wall outlet.

6. Align the RM-DS so that it is triggered from the edge of the disk. Note that there is onered and one green LED on the RM-DS. When both the red and green flash it signifiesthat the RM-DS is operating correctly. The screw on the RM-DS can be used to adjustthe signal strength. It is recommended that the supplied screwdriver be used in order tonot damage the adjustment screw.

TO

RS-640 DATA

COLLECTION

MODULE

Page 63: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

4-13

40Testing a Solid State Meter from the Calibration LED

1. Connect the Voltage Amplifier leads to the appropriate inputs on the meter test socket.

2. Connect the Current Amplifier leads to the appropriate inputs on the meter test socket.

3. Jumper additional current and voltage inputs appropriately on the meter test socketaccording to the Meter Socket Connection Diagram.

4. Connect the BNC of the RM-1H to the Data Collection Module.

TO

RS-640 DATA

COLLECTION

MODULE

Page 64: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Test Connection Diagrams

4-14

Testing a Solid State Meter from its KYZ Output

1. Connect the Voltage Amplifier leads to the appropriate inputs on the meter test socket.

2. Connect the Current Amplifier leads to the appropriate input on the meter test socket.

3. Jumper additional current and voltage inputs appropriately on the meter test socketaccording to the Meter Test Socket Connection Diagram.

4. Connect the RM-KYZ terminations to the KYZ output of the meter under test.

5. Connect the BNC of the RM-KYZ to the Data Collection Module.

TO

RS-640 DATA

COLLECTION

MODULE

Page 65: RS-600 Syntron Multifunction Test System - PowerRS-600 Operations Manual Ver. 005-10/03 10 Syntron System Details 1-5 RS-600 Overview The RS-600 Syntron Multifunction Test System is

RS-600 Operations Manual Ver. 005-10/03

Testing a Microjoule

1. Connect the Current Amplifier leads to the common and to the appropriate tap for the firsttest point.

2. Connect the Voltage Amplifier leads to the voltage input and jumper the Auxiliary Powerinput.

3. Connect the Hi Pulse Output to the Data Collection Module.

Test Connection Diagrams

4-15