Software● Real–Time cell growth logging● 3D graphical representati on of OD or growth rate over ti me over unit● Pause opti on● Save/Load opti on● Report opti on: PDF and Excel● Connect up to 12 units (recommended) simultaneously to 1 computer● Remote monitoring opti on (requires internet connecti on)● Cycling/Profi ling opti ons● User manual calibrati on possibility for most cells
RTS-8 is a personal bioreactor that uti lizes patented Re-verseSpin® technology that applies non-invasive, mecha-nically driven, low energy consumpti on, innovati ve type of agitati on where cell suspension is mixed by the single-use falcon bioreactor tube rotati on around its axis with a change of directi on of rotati on moti on resulti ng in highly effi cient mixing and oxygenati on for aerobic culti vati on. Combined with a nearinfrared opti cal system it is possible to register cell growth kineti cs non-invasively in real ti me.
Features● Parallel culti vati on enables to save ti me and resources for bioprocess opti mizati on● Individually controlled bioreactor accelerates opti mizati on process● Possibility to culti vate microaerophilic and obligate anaerobic microorganisms (not strict anaerobic conditi ons)● Reverse–Spin® mixing principle enables non-invasive biomass measurement in real ti me● Near-infrared opti cal system makes it possible to register cell growth kineti cs ● Free of charge soft ware for storage, demonstrati on and analysis of data in real ti me● Compact design with low profi le and small footprint for personal applicati on ● Temperature control for bioprocess applicati ons● Acti ve cooling for rapid temperature control, e.g. for temperature fl uctuati on experiments● Task profi ling for process automati zati on● Cloud data storage to remotely monitor the process of culti vati on while at home or using a mobile phones
RTS-8 Multi-channel BioreactorWith real time biomass, pH and pO2 measurement
Benefi tsRTS mixing principle has a number of advantages, e.g. automati c balancing, defoaming single-use, etc., but one of the more important points is that, due to the individual rotati on principle and thermostati ng of each bioreactor tube, diff erent conditi ons for supporti ng the temperature (from 4 to 70°C) and aerati on conditi ons can be applied (from anaerobic to aerobic).
Applications● Fermentati on real ti me growth kineti cs● Clone candidate screening● Protein expression● Temperature stress and fl uctuati on experiments● Media screening and opti mizati on● Growth characterizati on● Inhibiti on and toxicity tests● Strain quality control
LTF Labortechnik GmbH & Co. KG • Hattnauer Straße 18 • D-88142 Wasserburg/Bodensee • Tel. +49 (0) 83 82 / 98 52-0 • Fax +49 (0) 83 82 / 98 [email protected] • www.labortechnik.com
www.labortechnik.com
Page 2/2
Measurement range 0–100 OD600
Measurement wavelength (λ) 850 ± 15 nm
Light source Laser
Measurement periodicity per hour 1-60
Temperature setting range +4°C – +60°C
Temperature control range 15°C below ambient up to +60°C
Heating/coolling rate 0,5°C/min
Sample volume 3 – 50 ml
Speed control range 50 – 2700 rpm
Max. number of units connectedwith the software 12 (recommended)
Display LCD
Minimum PC requirements Intel/AMD Processor, 1 GB RAM, Windows XP* /Vista/7/8/8.1, 2.0 USB port
Optimal PC requirementsIntel/AMD Processor, 3 GB RAM Windows XP*/Vista/7/8/8.1/10,
USB 2.0 port *not guaranteed because OS not supported by producer
Dimensions (W×D×H) 330 x 580 x 320 mm
Weight 15 kg
Internal power supply Input AC 240 V 50/60 Hz; Output DC 12 V
O2 sensor
Range 0 – 100%
Accuracy ± 0.01% O2 at 0.21%, ± 0.1% O2 at 20.9%
Drift < 0.015% O2 per Day
Temperature range up to 40°C
pH sensor
Range pH 4.0 – pH 8.5
Accuracy ± 0.05 pH at pH 7 with onepoint calibration, ± 0.10 pH at pH 7 with precalibration
Drift < 0.005 pH per Day
Temperature range pH up to 40°C