Samples are loaded and AUC has reached temperature equilibration. Open a new file.

Samples are loaded and AUC has reached temperature

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Samples are loaded and AUC has reached temperature

Samples are loaded and AUC has reached temperatureequilibration.

Open a new file.

Page 2: Samples are loaded and AUC has reached temperature

Check velocity andinterference for all cells, choose rotor, set speed, time (hold)and temperature.

Page 3: Samples are loaded and AUC has reached temperature

set scan options set scan method. Here is 1 min between scans

Start the AUC and bring up to the desired speed.

save scan file

0 hrs 1 min 0 sec

Page 4: Samples are loaded and AUC has reached temperature

Details window for cell# 4.

Page 5: Samples are loaded and AUC has reached temperature

Laser set up window forcell # 4.

Set zoom to 50% and move sliding bar to right side to visualize the entire fringe pattern.

Page 6: Samples are loaded and AUC has reached temperature

Next, we have to adjust the Laser Delay.

By re-setting zoom, we can now see the top, meniscus and bottom of the cell

Top (inside) Meniscus Bottom (outside)

Page 7: Samples are loaded and AUC has reached temperature

Setting the laser delay aligns the laser beam so that it falls exactly down through the center of the center piece sector.

First, on the laser delay bar, left click on the right arrow box to movethe laser.

As the laser moves out of alignment, the right end of the fringe pattern begins to darken.When it looks like the fringe pattern example below, stop and then, using the left arrow box, count each left click of the mouse (1 click =.1degree) as you move the laser back through the middle and out of alignment in the other direction. Again, the right fringe pattern darkens as below. Then, to bring the laser to mid-line, left click on the right arrow box and move the laser back one-half the counted distance.

Page 8: Samples are loaded and AUC has reached temperature

Laser out of alignment at72.6 degrees.

Page 9: Samples are loaded and AUC has reached temperature

Laser out of alignment at71.2degrees.

Page 10: Samples are loaded and AUC has reached temperature

Laser is centered at 71.9.

Now, duration, brightness and contrast need to be adjusted

Page 11: Samples are loaded and AUC has reached temperature

For a starting point, set Laser duration to .2,bring down the brightness to 22 andset the contrast softer to 107

Page 12: Samples are loaded and AUC has reached temperature

To finish, bring down the contrast to100 and increase brightness to 26.

Page 13: Samples are loaded and AUC has reached temperature

Keep fringes and spaces so they are not too dark, and not too white so that you don’tsaturate the bright and dark range of the camera.

Too bright. Too much white space between fringes

Not bright enough and too much contrast. The black areas are too dark.

Too bright and not enough contrast. The black areas are not dark enough.

Page 14: Samples are loaded and AUC has reached temperature

To set Inside radius, place pointer just before the meniscus and double left clickwith mouse.

Page 15: Samples are loaded and AUC has reached temperature

To set Outside radius, placepointer at right edge anddouble left click with mouse.

Page 16: Samples are loaded and AUC has reached temperature

To begin radial calibration for IF scans, from the Interference top menu, open RadialCalaibration.

This window opens. Hit the LaserSetup button.

Page 17: Samples are loaded and AUC has reached temperature

This window appears. Make sureCell # 8 is checked. We calibrate the laser using the counterbalance. Note Fringe pattern above.

Page 18: Samples are loaded and AUC has reached temperature

Fringe pattern of a counter balance.

hole in counter balance solid area opposite holeair vs air no transmitted light air vs air

A gray area appears as the border between the counter balance hole and the solidpart.

Page 19: Samples are loaded and AUC has reached temperature

Adjust Laser Delay as before, by movinglaser out of alignment.

Page 20: Samples are loaded and AUC has reached temperature

Laser out of alignment again.

Page 21: Samples are loaded and AUC has reached temperature

Fringes are adjusted for laser delay, duration, brightness and contrast.

Page 22: Samples are loaded and AUC has reached temperature

To set inside radius, left click on theborder between air and gray space.

Page 23: Samples are loaded and AUC has reached temperature

To set outside radius, left click on theborder between air and gray space.

Page 24: Samples are loaded and AUC has reached temperature