Upload
hoangliem
View
226
Download
7
Embed Size (px)
Citation preview
SES Instruments Pvt. Ltd.SES Instruments Pvt. Ltd.
24. Four Probe Set-up: DFP-RM (Research Model) 21
25. Four Probe Set-up (Resistivity Mapping): FP-01 23
26. Measurement of Magnetoresistance: MRX-01 23
27. Two Probe Method for Resistivity Measurement TPX-01 25
28. Electron Spin Resonance Spectrometer: ESR-104 26
29. Study of Thermoluminescence of F-centres TLX-02 27
30. Hall Effect Experiment: HEX-21 28
31. Hall Effect in Metals: HEM-01 29
32. Dependence of Hall Coefficienton Temperature: HEX-22 31
33. Quinck's Tube Method (QTX-01) 32
34. Gouy's Method: GMX-01 33
35. Magnetic Hysteresis Loop Tracer: HLT-111 34
36. Study of P-N Junctions(Bandgap & Diffusion Potential): PN-1 35
37. Semiconductors Diodes Characteristics(Advance Model): SDC-02 36
Physics and Electronics Experimental Setups
38. Study of a Transistor Amplifier (RC Coupled)Cum-Feed Back Amplifier: RC-01 37
39. Study of Multivibrators: MV-01 37
40. Study of Characteristics of SemiconductorsDiodes (Basic Model): D-1 38
41. Study of Hybrid Parametersof a Transistor: HP-01 38
42. Study of a Solid State Power Supply: SSPS-1 39
43. Study of Modulation & Demodulation withBuilt-in Carrier Frequency (Solid State): MD-1 39
44. Study of a Basic Operational Amplifier: 741-01 39
45. Study of Op. Amp. 74l Applications: 741-02 40
46. Study of Astable & Monostable Multivibrators using Timer IC: 555 40
47. Study of an Integrated Circuit Regulator: 723 40
Contents P a g e
General Laboratory Instruments
1. Digital D.C. Microvoltmeter: DMV-001 3
2. Digital Nanoammeter: DNM-121 3
3. Digital Picoammeter: DPM-111 4
4. High Voltage Power Supply: EHT-11 4
5. True RMS A.C. Millivoltmeter: ACM-102/ ACM-103 5
6. Electromagnet & Power Supply: EMU-75 & DPS-175 5
7. Electromagnet & Power Supply: EMU-50 & DPS-50 6
8. Digital Gaussmeter: DGM-102 6
9. Digital Gaussmeter: DGM-202 7
10. Digital Gaussmeter: DGM-103 7
11. Digital Gaussmeter: DGM-204 8
12. PID Controlled Oven: PID-200 CT 8
13. Travelling Microscope: TVM-02 9
14. Travelling Microscope: TVM-03 9
15. Regulated Power Supply: PS-12 10
16. Function Generator: FG-01 10
Physics and Material Science Lab Experiments
17. Magnetic Field Measurement Apparatus: MFM-01 11
18. Study of Dielectric Constant: DEC-01 12
19. Planck's Constant Experiment(By Photoelectric Effect): PC-101 14
20. Study of Planck's Constant (Using LED's): PCA-01 16
21. Franck Hertz Experiment: FH-3001 18
22. Four Probe Experiment : DFP-02 (Basic Model) 19
23. Four Probe Experiment : DFP-03 (Advance Model) 20
Application : For the measurement of low currents down to 100pA
Current Range : 100nA to 100mA full scale in 4 decade ranges with 100%overranging (100pA minimum)
Accuracy : ±0.2% for all ranges
Display : 3½ digit, 7 segment LED with autopolarity and decimal indication
Input : Through BNC connector
Very low temperature drift
Low dc input bias current-10pA
Measures voltage down to 1mV
Recorder facility (optional)
Measures current downto 100pA
All solid state and IC design
Accepts either polarityof input current
DI AL N AMMEGIT ANO TERM e D - 121od l NM
I NT FI E I MESC E I C QU P NT
ROORKEE
I A M V EDIG T L ICRO OLTM TERMode DNM - 1l 00
SCIENT F C EQUIPMENTI IR ORKEEO
Digital Nanoammeter D N M - 1 2 12
Digital Microvoltmeter D M V - 0 0 11
Application : For very low d.c. voltage measurement at very high impedance
Range : 1mV, 10mV, 100mV, 1V & 10V with 100% over-ranging.
Resolution : 1mV
Accuracy : ±0.2%
Stability : Within ±1 digit
Input Impedance : >1000MW (10MW on 10V range)
Display : 3½ digit, 7 segment LED with autopolarity and decimal indication
3
Measures current down to 1pA
All solid state and IC design
Very low offset current
4
3 Digital Picoammeter
High Voltage Power Supply
D P M - 1 1 1
E H T - 1 1
Continuously variable
Electronically regulated
Fully solid state circuit
Low power consumption
Application
Current Range
Display
Accuracy
For the measurement of very low currents down to 1pA
1nA to 100mA full scale in 6 decade ranges with 100%over-ranging (1pA minimum)
3½ digit 7 segment LED display
±0.2% for all ranges
:
:
:
:
Application
Polarity
Range
Max. Current
Stability
Regulation
Display
Protection
EHT source for photomultiplier tubes and other radiation detectors
+ve or -ve, as required
0-200V & 0-1500V continuously adjustable
1mA
±0.05% for ±10% mains variation
±0.05% for 0 to 1mA load
3½ digit, 7 segment LED
Fully protected against overload and short circuit by currentlimiting technique
:
:
:
:
:
:
:
:
4
5 True RMS A.C. Millivoltmeter
Measures True RMS Voltage
Accuracy 1%
High Input Impedance
High Stablility
Excellent Linearity
A C M - 1 0 2 / 1 0 3
Application
Voltage Range
Freq. Range
Display
Input Impedance
Accuracy
ACM-103 (Additional Features)
Oscillator Output
Frequency
General purpose, indispensable to any electronics laboratory
20mV, 200mV, 2V and 20V
10Hz to 200KHz
3½ digit, 7 segment LED with autopolarity and decimal indication
1 Mohm shunted by 25pf on all ranges
1% (10Hz to 100KHz); 2% (100KHz to 200KHz)
:
:
:
:
:
:
0-300mV continuously variable
1KHz (sine) fixed
:
:
5
Constant Current Power Supply, DPS-175
Application : IC regulated constant current sourcessuitable for electromagnet EMU-75
Current : 0-3A per coil, smoothly adjustable
Line Regulation : ±0.1% for 10% mains variation
Load Regulation : ±0.1% for load variation from 0 to max.
Display : 3½ digit, 7 segment LED display
Protection : Protected against overload/short circuit.
6 Electromagnet & Power Supply E M U - 7 5 , D P S - 1 7 5
Electromagnet, EMU-75
Application : General purpose laboratory electromagnet
Pole Pieces : f75mm flat in EMU-75: f75mm tappered to 25mm in EMU-75T
Field : EMU-75: 10.75KG at 10mm airgap: EMU-75T: 17.5KG at 10mm airgap
Energising Coils :
Two, each having a resistance of approx. 13W
Power Requirement :
0-90Vdc, 3A, if coils are connected in series
0-45Vdc, 6A, if coils are connected in parallel
T F C I M NSCIEN I I EQU P E T
ORO RKEE
DIGITAL GAUSSMETERod M 1M el DG - 02
8
7 Electromagnet & Power Supply
Electromagnet,EMU-50
Application: Low costelectromagnet isspecially designed forHall Effect andMagnetic Susceptibilityby Quinck’s methodexperiments
Pole Pieces : f50mm diameter flat in EMU-50: f50mm tapered to 20 mm in EMU-50T
Field : EMU-50: 7.5KG, at 10mm airgap: EMU-50T: 9.5KG at 10mm airgap
Energising Coils: Two,each with a resistanceof about 3W
Power Requirement:0-30Vdc, 4A, if coilsare connected in series
Digital Gaussmeter
EM U-5 0 , DPS-50
D G M - 1 0 2
Mag. Field Measurement
Excellent Linearity
IC Controlled Circuit
Excellent Stability
Application
Range
Accuracy
Display
Transducer
Special Feature
Suitable for the measurement of magnetic field from 1gauss to 20KG
0-2KG, 0-20KG
±0.5%
3½ digit, 7 segment LED display
Hall Probe (InAs)
Indicate the direction of the magnetic field
:
:
:
:
:
:
6
Constant Current Power Supply, DPS-50
Application
Current
Line Regulation
Load Regulation
Display
Protection
IC regulated constant current sourcesspecially suitable for EMU-50
0-4A (smoothly adjustable)
±0.1% for ±10% mains changes
±0.1% for no to full load
3½ digit, 7 segment LED display
Protected against overload/short circuit.
:
:
:
:
:
:
10
9 Digital Gaussmeter
Wide Range (1G to 20KG)
Excellent Linearity
Excellent Stability
Interchangeable Hall Probes
Digital Gaussmeter D G M - 1 0 3
D G M - 2 0 2
Range
Accuracy
Display
Transducer
Special Feature
0-2KG, 0-20KG
±0.5%
3½ digit, 7 segment LED display with auto polarity andoverflow indication
Hall Probe (InAs)
Indicates the direction of the magnetic field
:
:
:
:
:
Differential mode facility
Wide Range (1G to 40KG)
Excellent Linearity
Excellent Stability
Easy replacement of Hall Probe
Application
Range
Accuracy
Display
Transducer
Special Feature
Suitable for the measurement of magnetic field from 1 gaussto 40KG with differential mode facility
0-2KG, 0-20KG & 0-40KG
±0.5%
3½ digit, 7 segment LED display
Hall Probe (InAs)
Indicates the direction of the magnetic field
:
:
:
:
:
:
7
Application :
Wide application in industry, where accurate measurements of magnetic field is required. Easy inter-changeability of probe enables both the axial and transverse field measurement with same unit
12
11 Digital Gaussmeter
Measures down to 0.1G
Excellent Linearity
Excellent Stability
Interchangeable Hall Probes
PID Controlled Oven
Introduction
This is high quality temperature controlled oven suitable for testing of electronic components & study of temperature transducers etc. The oven has been designed for fast heating and cooling rates which enhances the effectiveness of the controller. While the basic design of the controller is around the PID configuration for its obvious advantages, wastage of power is avoided by using a Pulse Width Modulated (PMW) switch. This combination has the advantages of both on-off controller and linear PID controller. The result is a good stable and accurate temperature control.
D G M - 2 0 4
P I D - 2 0 0 C T
Application : Suitable for magnetic field measurement from 0.1 gauss to 40KG. Easy interchangeability of probe enables both axial and transverse field measurement with same unit
Range
Accuracy
Resolution
Display
Transducer
Special Feature
0-200G, 0-2KG, 0-20KG and 0-40KG
±0.5%
0.1 gauss at 200G range
3½ digit, 7 segment LED with auto polarity and overflow
Hall Probe (InAs)
Indicate the direction of the magnetic field
:
:
:
:
:
:
Specifications of the Oven
Temperature Range
Stability
Measurement Accuracy
Oven Dimensions
Sensor
Display
Ambient to 200°C
Short Range: ±0.2°C Long Range: ±0.5°C
±0.5°C (typical)
Heating Chamber Inner Diameter: 25mm; Height: 100mm
RTD (A class)
3½ digit, 7 segment LED (12.5 mm)
:
:
:
:
:
:
8
14
13 Travelling Microscope
Travelling Microscope T V M - 0 3
T V M - 0 2
Introduction
The bed is of a heavy casting, thoroughly aged and machined, is fitted with leveling screws. On the dovetail guide ways slides the horizontal carriage which can be clamped at any position by means of a thumbscrew. A second sliding carriage slides along a gunmetal vertical pillar fitted on the horizontal carriage. The slow motion guide bars are made of sturdy material and the motion is very smooth.
(i) True achromatic objective with 7.5cm focussing distance from object
(ii) 10X Ramsden Eyepiece with fine cross wire
Horizontal scale
Vertical scale
Least Count
180mm with a screw gauge type motion
150mm with a screwgauge type motion
0.01mm on both scales
:
:
:
Scale
Optics
Introduction
The bed is of a heavy casting, thoroughly aged and machined, is fitted with leveling screws. On the dovetailed guide ways slide the carriage which can be clamped at any position by means of a thumbscrew. A sliding carriage slides along a gunmetal Vertical pillar fitted on the horizontal carriage. The slow motion guide bars are made of sturdy material and the motion is very smooth. Microscope tube is inclinable in any angle. True vertical and horizontal positions marked focussing. Scales and verniers are made of lifetime Stainless Steel.
9
Optics
True achromatic objective with 7.5cm focussing distance from object; 10XRamsden Eyepiece with fine cross wire is provided.
Horizontal scale
Vertical scale
Lateral Scale
Least Count
180mm with a screw gauge type motion
150mm with a vernier scale
60mm with a vernier scale
0.01mm on all scales
:
:
:
:
Scale
16
15 Regulated Power Supply
Application: General purposelow cost power source ideallysuited for analog and digital IC’s
Output: Four fixed voltages viz.+5V, -5V, +12V, and -12V
Current: 300mA max for each
Regulation: ±0.2% for noload to full load
Protection: Internally protected byover-current and thermal shutdown
Function Generator
4-digit counter basedfrequency display –Maximum error ONE LSB
Square, sine and triangularoutput upto 2MHz
IC based circuit – highstability, low distortion
PS-12
FG-01
This is an economical priced signal source. The IC used is a high frequency functiongenerator that produce low distortion sine, square and triangular waveform from lessthan 1Hz to 20MHz
Applications
Waveforms
Frequency
Amplitude
Output Impedance
Frequency Readout
General purpose instrument, indespensable to anylaboratory
Square/ Sine/ Triangular (switch selectable)
10Hz – 2MHz in six ranges
0-3V (p-p)
50W
4-digit counter
:
:
:
:
:
:
10
SCIENTIFI EQUIPMENTC
ROORKEE
I L C REGU ATED POWER SUPPLY
Model PS 12
F MUNCTION GENERATOR odel : FG-01
SCIENT FIC EQUIPMENTI
ROORKEE
Magnetic Field Measurement Apparatus M F M - 0 117
11
The experiment consists of two coils, Constant Current Power Supply and Gaussmeter. The Gaussmeter probe is mounted on a rail with a scale. It can move smoothly and precisely for measurement of magnetic field along the axis of the coils.
The following studies can be carried out with the set-up:
1. Study of magnetic field due to one coil and calculation of its diameter.
2. Study of Principal of super-imposition of magnetic field due to 2 coils by keeping the distance between the coils at a, >a and <a, where a is the radius of the coil.
Legend:
Line 1 - Magnetic Profile when the distance between the coils is >a
Line 2 - Magnetic Profile when the distance coils is =a
Line 3 - Magnetic Profile when the distance between coils is <a- Superimposition overlaps completely
Apparatus consists of the following:-
1. Digital GaussmeterRange : 0-200Resolution : 0.1GAccuracy : ± 0.5%Display : 3½ digit 7 segment LED with autopolarity.
2. Two CoilDiameter : 200 mmNumber of turn : 1000
3. Constant Current Power SupplyCurrent : 0-0.5A Smoothly adjustableLine Regulation : ± 0.2% for 10% mains variation.Load Regulation : ± 0.2 % for 0 to full loadDisplay : 3½ digit 7 segment LED Display.Protection : Against overload/short circuit.
The 2 coils are mounted on platform one coil is fixed and other coil moves smoothly on a rail along the axis of the coils.
1
2
3
Ma
gn
etic F
ield
Distance (mm)
12
18 Study of Dielectric Constant
Study of Dielectric Constant and Curie Temperature of Ferroelectric Ceramics
Dielectric or electrical insulating materials are understood as materials in which electrostic field can persist for long times. Layers of such substances are commonly inserted into capacitors to improve their perfomance, and the term dielectric refers specifically to this application.
An electric field polarizes the molecules of dielectric, producing concentrations of charge on its surface that create an electric field opposed (antiparallel) to that of capacitor. This reduces the electric potential. Considered in reverse, this means that, with a dielectric, a given electric potential causes the capacitor to accumulate a larger charge.
All ferroelectric materials have a transition point called the Curie point (T ). At T>T , c c
the crystal does not exhibit ferroelectricity, while for T<T it is ferroelectric. If there are c
more than one ferroelectric phase, the temperature at which the crystal transforms from one phase to another is called transition temperature. Near the Curie temperature point or transition temperatures, the thermodynamic properties including dielectric, elastic, optical and thermal constants show an anomalous behaviour.
The graph on the next page shows the variation of dielectric constant with temperature for Lanthanum doped Lead Zirconate Titanate (PLZT) ceramic, which is cooled from its paraelectric cubic phase to ferroelectric rhombohedral phase.
1. Probes Arrangement, DEA-01
It has two individually spring loaded probes. The probes arrangement is mounted in a suitable stand, which also holds the sample plate. To ensure the correct measurement of sample temperature, the RTD is embedded in the sample plate
Description of the Experimental Set-up
DEC-01
Perovskite ABO structure with the A3
and B cations on the corner and bodycentre positions, respectively. Three oxygenanions per unit cell occupy the faces andform octahedra surrounding the B-site.
A
B
Temp. Range
Display
Resolution
Accuracy
Stability
Power
Ambient to 200°C
3½ digit, 7 segment LED with autopolarity & decimal indication
0.1°C
±0.5°C (typical)
±0.2°C
150W
:
:
:
:
:
:
Range
Resolution
Display
50-6000 pf
1 pf
3½ digit, 7 segment LED
:
:
:
(ii) Digital Capacitance Meter
This is a compact direct reading Instrument for the measurement of capacitance of the sample.
6400
6200
6000
5800
5600
5400
5200
5000
4800
4600
4400
4200
4000
3800
3600
3400
3200
3000
2800
2600
2400
2200
2000
1800
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230
Die
lec
tric
Co
ns
tan
t (e
)
DIELECTRIC CONSTANTSAMPLE : Barium Titanate
T (Curie Point)C
Temperature (K)
13
just below the sample. This stand also serves as the lid of temperature controlled oven. Proper leads are provided for connection to Capacitance Meter and Temperature Controller.
2. DEC-Sample
Barium Titanate (BaTiO )3
3. Oven, DEO-01
This is a high quality temperature controlled oven. The oven has been designed for fast heating and cooling rates, which enhances the effectiveness of the controller.
4. Main Unit
The Set-up consists of two units housed in the same cabinet.
(i) Oven Controller
Platinum RTD (A class) has been used for sensing the temperature. A Wheatstone bridge and an instrumentation amplifier are used for signal conditioning. Feedback circuit ensures offset and linearity trimming and a fast accurate control of the oven temperature.
19 Planck's Constant Experiment
1. Determination of Planck's Constant and Work Function of Materials by Photoelectric Effect
It was observed as early as 1905 that most metals under influence of radiation, emit electrons. This phenomenon was termed as photoelectric emission.The experimental facts provided the strong evidence that the electromagnetic field is quantified and the field consists of quanta of energy E= hn where n is the frequency of the radiation and h is the Planck's constant. These quanta are called photons.Further it is assumed that electrons are bound inside the metal surface with an energy ef, where f is called work function. It then follows that if the frequency of the light is such that
hn > ef
it will be possible to eject photoelectrons, while if hn<ef, it would be impossible. In the former case, the excess energy of quantum appears as kinetic energy of the electron, so that
2 hn= mv + ef (1)
which is the famous photoelectrons equation formulated by Einstein in 1905.The energy of emitted photoelectrons can be measured by simple retarding potential techniques as is done in this experiment. When a retarding potential V is used to 0
measure kinetic energy of electrons E , we have, e
2E = mv = eV or V = n - fe 0 0
So when we plot a graph of V as a function of n, the slope of the straight line yields h and 0
the intercept of extrapolated point n=0 can give work function f.
2. To verify inverse square law of radiation using a photoelectric cell
If L is the luminous intensity of an electric lamp and E is the illuminanscence (intensity of illumination) at point r form it, then according to inverse square law.
E =
If this light is allowed to fall on the cathode of a photo-electric cell, then the photo-electric current (I) would be proportional to E.
E = =K.I
Hence a graph between and I is a straight line, which verify the inverse square law of radiation.
PC-101
L2r
L2r
12r
14
15
THE APPARATUS CONSIST OF THE FOLLOWING :
1. Photo Sensitive Device : Vacuum photo tube.2. Light source : Halogen tungsten lamp 12V/35W.3. Colour Filters : 635nm, 570nm, 540nm, 500nm & 460nm.4. Accelerating Voltage : Regulated Voltage Power Supply
5. Current Detecting Unit : Digital Nanoammeter It is high stability low current measuring instrument
6. Power Requirement : 220V ± 10%, 50Hz. or 110V ± 10%, 60Hz. as required.
7. Optical Bench : The light source can be moved along it to adjust distance between the light source and the phototube, scale length is 400mm. A tube is provided to install colour filter and a focus lens is fixed in the back end.
The set-up is complete in all respect, no additional accessory required.
Output
Display
Accuracy
±15V continuously variable through multi-turn pot
3 ½ digit 7-segment LED
±0.2%
:
:
:
Range
Resolution
Display
Accuracy
1000 mA, 100 mA, 10 mA & 1mA with 100 % over ranging facility
1nA at 1mA range
3 ½ digit 7-segment LED±0.2%
:
:
:
:
2
1.5
1
0.5
0
0.5
1
1.5
2
1 2 3 4 5 6 7 80
Frequency (n)-1 14
(sec X 10 )
DV = 0.810
14Dn = 1.9 X 10
h =
0.81141.9x10
-19x1.60x10
-346.82x10 Joule.sec
1.6V
eDV0
Dn
f
=
=
=
Ac
ce
lea
rtin
g P
ote
nti
al
Re
tard
ing
Po
ten
tia
l (-
V)
0
460
500
540
570
635
Fig.: Typical graph of V (stopping potential)0
as a function of n0
Fig.: Verification of Inverse Square Law2 Graph : 1/r vs I
2.4
2.2
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0.0
Cu
rre
nt I (m
A)
0.00 1.00 2.00 3.00 4.00 5.00
20 Study of Planck’s Constant
Study of Planck’s Constant by means of LED
Several proposals to measure the Planck’s Constant for didactical purposes, using the current – voltage (I-V) characteristics of a light emitting diode (LED) have been made quite regularly in the last few years. The physical interpretation however is not completely clear and this has raised many discussions, which have been published almost with same regularity as the proposals themselves.
The present experiment is based on diode current for V<V , using the diode law.0
l = l exp [-e(V – V)/ hkT]0 0
where, e is electronic charge, k is Boltzmann constant, T is absolute temperature and h is material constant which depends on the type of diode, the location of recombination region, etc.
The correct method to determine the real height of the potential energy barrier V is 0
to directly measure the dependence of the current on temperature keeping the applied voltage V slightly below V . The idea is that the disturbance to V is as little 0 0
as possible. The slope of lnI vs. 1/T curve gives e(V – V)hk (Fig. 1). The constant h 0
is determined from I-V characteristics of the diode (Fig. 2) at room temperature from the relation
h = (e/kt) (DV/Dlnl)
Compare with previous methods, this determination of V is more precise and more 0
accurate and at the same time the physical interpretation is more transparent. The Planck’s constant is then obtained by the relation
h = e V l /c0
The wavelength (l) of the light emitted by the diode can be measured by a transmission grating spectrometer normally available in the lab.
The value of Plank’s constant obtained from this method is within 5% of accepted -34value (6.62 x 10 Joules.sec).
16
C TI E I MES IEN FIC QU P NT
ORO RKEE
PCA-01
AN S ONSTA A PARAT SPL CK' C NT P UMo l C 0de - P A- 1
V
CV - 1 - T 1E TXP TEXP
FOF
NCURRE T
S T EMPE TON
mA
mA
ON
LED
The set-up is a self contained unit. All the necessary facilities and measuring devices are built in a single unit, as a result only minimum of external connections need to be made.
[I] Dependence of current (I) on temperature (T) at constant applied voltage (V)
The following facilities are built in for this
Current Meter
A highly stable current source with 3½ digit display.Range: 0-2mA with resolution of 1mA
Oven, PCO-01
It is a small temperature controlled oven with built-in RTD sensor.Temperature adjustable from ambient to 338KDigital display of temperatureHigh stability 1K
Variable Voltage Source (0-2V)
A high stability voltage source with 3½ digit display.
[II] Material Constant h
To draw I-V characteristics of LED for determination of h, a variable voltage source and a current meter are provided with 3½ digit display.
[III] The wavelength l of light emitted by LED
These are taken from LED datasheet or measured by a transmission grating,and are provided with the set of LED’s.
Fig. 1 Fig. 2
17
T-I Characteristics I-V Characteristics
1.2
1.1
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
8
7
6
5
4
3
2
1
Sample: LED YellowAmbient Temp: T = 300KSample: LED Yellow
V = 1.8 V
DInl
DV
1.4 1.6
Junction Voltage (V)
1.8 22.8 2.9
In I (
I in
mA
)
In I (
I in
mA
)
3 3.1 3.2
DInl
-1DT
-1 -3T x 10
3.3 3.4 3.5
18
21 Franck-Hertz Experiment
In 1914 Franck & Hertz experimentally proved that the energy transferred from electrons to atoms always has discreet values and the values obtained for energy levels are in agreement with spectroscopic results. Thus experimentally proving the Bhor's Model. It is a landmark experiment and can be performed in any college lab.
The experiment consists of the following:
Argon filled tetrode
Filament Power Supply
3.6 - 4.3 continously variable
Power Supply for VG K1
1.3 - 5V continously variable
Power Supply for VG A2
1.3 - 15V continously variable
Power Supply for VG K2
0 - 95V continuously variable
Saw tooth waveform for CRO display
Scanning Voltage : 0-80V
Scanning Frequency : 115±20 Hz
Multirange Digital Ammeter
Display : 3½ digit, LED-7 -8 -9 Range : 10 , 10 & 10 A
The instrument can not only lead to a plot of the amplitude spectrum curve by means of point by point measurement, but also directly display the amplitude spectrum curve on the oscilloscope screen. This instrument can thus be used as classroom experiment as well as for demonstration to a group of students.
F H - 3 0 0 1
Nanoammeter
Fig. 1
V
V
V
G K1
G K2
G A2
A
G2
G1
K
Plot of Beam Current Vs. Accelarting Voltage
Accelarating Voltage V (V)G2K
-8E
lec
tro
n B
ea
m C
urr
en
t (A
x1
0)
1400
1200
1000
800
600
400
200
0
0 10 20 30 40 50 60 70 80 90 100
in Franck Hertz Experiment, FH-3001
47V
47V
Oscilloscope display of Franck-Hertz experiment
Ele
ctr
on
Be
am
Cu
rre
nt
Accelerating Voltage
Fig.The resistivity of a Germanium crystal as a function of inverse temperature. For this sample T<363K, conduction is due mainly to the impurity carriers
(Extrinsic Region). For T>363K, conduction is due to electrons transferred to the conduction band (and the corresponding holes created in the valence band). This
is the Intrinsic region.
0
0.1
0.2
0.3
0.4
0.5
0.5 1.0 1.5 2.0-1(T x1000)
log
r10
2.5 3.0 3.5 4.0
0.6
0.7
0.8
0.9
1.0
.325
0.5
68
363K
Eg
2k
Eg
1T
Log re
-50.568 x 2.3026 x 8.6 x 10 x 2-30.325 x 10
( )
= 0.69 eV
=
22 Four Probe Experiment
Resistivity of Semiconductors by Four Probe Method at DifferentTemperatures and Determination of the Band-gap (Basic Model)
The experiment consists of the following:
(i) Four Probe Arrangement, FPA-02 : It has four individually collinear, spring loaded probes. The probes are mounted in a teflon bush, which ensures a good electricalinsulation between theprobes and maintainsthem at equal distance.The whole arrangementis mounted on a suitablestand and leads areprovided for the current& voltage measurements.
(ii) Oven FPO-02 (upto 200°C)
(iii) DFP-02 Sample (Ge-n type)
(iv) Thermometer DFP 02-THR (0 - 150°C)
(v) Four Probe Set-up, DFP-02
(a) Multirange Digital Millivoltmeter
Range
Resolution
Accuracy
Impedance
Display
(b) Constant Current Generator
Open Circuit
Current Range
Resolution
Accuracy
Load Regulation
Line Regulation
(c) Oven Power Supply : Built-in for the above oven.
D F P - 0 2
:
:
:
:
:
X1 (0-200mV)and X10 (0-2V)
100mV at X1 range
±0.1% of reading ±1 digit
1 Mohm
3½ digit, 7 segment LED
18V
0-20mA
10mA
±0.25%
±0.1% for no loadto full load
±0.1% for 10% changes
:
:
:
:
:
:
19
Resistivity of Semiconductor by Four Probe Method at DifferentTemperatures and Determination of the Band-gap (Advance Model)
Description of the experimental set-up
1. Probes Arrangement, FPA-03
It has four individually spring loaded probes. The probes are collinear and equally spaced. The probes are mounted in a teflon bush, which ensures a good electrical insulation between the probes. A teflon spacer near the tips is also provided to keep the probes at equal distance. The probe arrangement is mounted in a suitable stand, which also hold, the sample plate. To ensure the correct measurement of sample temperature, the RTD is embedded in the sample plate just below the sample. This stand also serves as the lid of the temperature controlled oven. Proper leads are provided for the current and voltage measurements.
2. DFP 03-SAMPLE
Germanium crystal in the form of a chip.
3. OVEN, FPO-03
This is a high quality temperature controlled oven suitable for Four Probe Set-up. The oven has been designed for fast heating and cooling rates, which enhances the effectiveness of the controller.
4. FOUR PROBE SET-UP, DFP-03
The set-up consists of three units housed in the same cabinet.
(i) Oven Controller
Platinum RTD (A class) has been used for sensing the temperature. A wheatstone bridge and an instrumentation amplifier are used for signal conditioning. Feedback circuit ensures offset and linearity trimming and a fast accurate control of the oven temperature.
23 Four Probe Experiment D F P - 0 3
20
FOUR PROBE SET-UP Model DFP-03
RANGE
ON OFF
SET TEMP RTD OVEN VOLTAGE CURRENT CURRENTADJ.
mV
mA
X1 X10OVEN CONTROL
K
C ENTI C E UI ME TS I FI Q P N
OORK ER E
D A OIGIT L MICR VOLTMETER
M d 1o el DMV - 00
C NS ANT NT O CEO T CURRE S UR
de S 0Mo l : CC - 1
PID CONTROLLED OVENModel : PID-200
S I I E PC E T F C Q I ME T
N I U N
R O KO R EE
CS IENTIFIC EQUIPMENTRO RKEEO
SCIENTIFIC EQUIPMENTROORKEE
24 Four Probe Set-up DFP - RM(Research Model)
Four Probe Set-Up for Measuring the Resistivity of Very Low toHigh Resistivity Samples at Different Temperatures
The experiment consists of the following:
The probe arrangement is mounted in a suitable stand, which holds the sampleplate and RTD (A class) sensor.
1. Probes Arrangement, FPA-RM:
21
Temperature Range : Ambient to 473KStability : ±0.5K
Measurement Accuracy : ±1K (typical)Oven : Specially designed for Four Probe Set-UpDisplay : 3½ digit, 7 segment LED with autopolarity and
decimal indication
(ii) Multirange Digital Voltmeter
Range : X1 (0-200mV) & X10 (0-2V)Resolution : 100mV at X1 rangeAccuracy : ± 0.1% of reading ± 1 digitDisplay : 3½ digit, 7 segment LED with autopolarity and
decimal indicationOverload Indicator : Sign of 1 on the left & blanking of other digits.
(iii) Constant Current Generator
Open Circuit Voltage : 18VAccuracy : ± 0.25% of the reading ± 1 digitCurrent range : 0 - 20 mAResolution : 10 mALoad regulation : ± 0.1% for 0 to full loadLine regulation : 0.1% for 10% changes
The experimental set-up is complete in all respect
2. PID Controlled Oven, PID-200
Specifications of the Oven
Temperature Range
Resolution
Short Range Stability
Long Range Stability
Measurement Accuracy
Oven, FRO-RM
3. Constant Current Source
a) Constant Current Source, CCS-01 (for low resistivity sample)
This current source is suitable for the resistivity measurement of low to mediumresistivity samples such as thin films of metals/alloys and semiconductors suchas Ge.
Specifications
Open Circuit Voltage
Current Range
Resolution
Accuracy
Display
Load Regulation
Line Regulation
b) Low Current Source, LCS-02 (for high resistivity samples)
4. D.C. Microvoltmeter, DMV-001 (Item No. 1).
: 12V
: 0-20mA & 0-200mA (continuously variable)
: 10mA at 0-20mA range
: ±0.25% of the reading
: 3½ digit, 7 segment LED
: ±0.1% for 0 to full load
: ±0.1% for 10% changes
Ambient to 200°C
0.1°C
±0.2°C
±0.5°C
±0.5°C (typical)
Specially designed for Four Probe Set-Up
3½ digit, 7 segment LED
:
:
:
:
:
:
:
22
This current source is especially suitable for the resistivity measurement ofpolymer films, conducting ceramics and Si crystals.
Specifications
Open Circuit Voltage
Current Range
Resolution
Accuracy
Display
Load Regulation
Power
: 15V (min.)
: 0-2mA, 0-20mA, 0-200mA & 0-2mA
: 1nA at 0-2mA range
: ±0.25% of the reading ±1 digit
: 3½ digit, 7 segment LCD with auto polarity
: ±0.1% for 0 to full load
: 3 x 9V batteries
Display
26 Magnetoresistance Set-up
Measurement of Magnetoresistance of Semiconductors
The resistance of the sample changes when the magnetic field is turned on. The phenomenon, called magneto- resistance, is due to the fact that the drift velocity of all carriers is not same. With the magnetic field on; the Hall voltage V = E t = |v x H| compensates y
exactly the Lorentz force for carriers with the average velocity; slower carriers are over compensated and faster one undercompensated, resulting in trajectories that are not along the applied field.
Introduction
The Four Probe Method is one of the standard and most widely used method for the measurement of resistivity. In its useful form, the four probes are collinear. The error due to contact resistance, which is significant in the electrical measurement on semiconductors, is avoided by the use of two extra contacts (probes) between the current contacts. In this arrangement the contact resistance may be high compared to the sample resistance, but as long as the resistance of the sample and contact resistance’s are small compared with the effective resistance of the voltage measuring device (potentiometer, electrometer or electronic voltmeter), the measured value will remain unaffected. Because of pressure contacts, and 2-way motion, the arrangement is specially useful for quick measurement on large samples at room temperature.
Easy resistivity mapping of large sample
2 – way movement with vernier scales(0.01mm)
Spring loaded contacts for firmconnections
Features
Four Probe Set-up for mapping the resistivity of large samples
Accessories required
1. Constant Current Source, CCS-01 (on page 22)
2. Low Current Source, LCS-02 (on page 22)
3. D.C. Microvoltmeter, DMV-001 (Item no. 1, on page 3)
23
25
M R X - 0 1
MAGNETORESISTANCE SET-UPModel : DMR-01
SCIENTIFIC EQUIPMENTROORKEE
mA
mV
Current Voltage
ON
Four Probe Set-up (Mapping) F P - 0 1This results in an effective decrease of the mean free path and hence an increase in resistivity. Here the above referred symbols are defined as: v = drift velocity; E = applied electric field; t = thickness of the crystal; H = Magnetic field. The set-up consists of the following:
(1) Four Probe arrangement, MRA-01
It consists of 4 collinear, equally spaced (2mm) and individually spring loaded probes mounted on a PCB strip. Two outer probes are for supplying the constant current to the sample and two inner probes for measuring the voltage developed across these probes. This eliminate the error due to contact resistance which is particularly serious in semiconductors. A platform is also provided for placing the sample and mounting the Four Probes on it.
(2) MR-Sample
Ge Crystal (n-type) dimensions: 10 x 10 x 0.5mm
(3) Magnetoresistance Set-up, DMR-01
This unit consists of a digital millivoltmeter and constant current power supply. The voltage and probe current can be read on the same digital panel meter through a selector switch.
24
(4) Electromagnet, EMU-50V (Item No. 7, on page 6)
(5) Constant Current Power Supply, DPS-50 (Item No. 7, on page 6)
(6) Digital Gaussmeter, DGM-202 (Item No. 9/8, on page 7/6)
(b) Constant Current Power Supply
Specifications:
Current
Accuracy
Load regulation
Line regulation
: 0-20mA (10mA min.)
: ±0.2% of the reading ±1 digit
: 0.1% for 0 to full load
: 0.2% for 10% variation
(a) Digital Millivoltmeter
Specifications: Range
Accuracy
: 0-200mV(100mV min.)
: ±0.1% (±1 digit)
Magnetoresistance of a Ge CrystalDr/ro vs. Magnetic Field (H)
Dr/r
o
0.036
0.032
0.028
0.024
0.020
0.016
0.012
0.008
0.004
0.0000.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5
Magnetic Field (KG)
5.0 5.5
27 Two Probe Setup
Two Probe Method for resistivity measurement of insulatorsat different temperatures
25
Oven(TPO-01)
Sensor
Display
Power
Specially designedfor Two Probe Set-Up
RTD (A class)
3½ digit, LED
150W
:
:
:
:
Temperature Range
Resolution
Short Range Stability
Long Range Stability
Measurement Accuracy
Ambient to 200°C
0.1°C
±0.2°C
±0.5°C
±0.5°C (typical)
:
:
:
:
:
3. High Voltage Power Supply, Model EHT-11 (Item No. 4, on page 4)
4. Digital Picoammeter, Model DPM-111 (Item No. 3, on page 4)
The experimental set-up is complete in all respect
The Two Probe Method is one of the standard and most commonly used method for the measurement of resistivity of very high resistivity samples - near insulators. The resistivity measurement of such samples is beyond the range of Four Probe Method.
DESCRIPTION OF THE EXPERIMENTAL SET-UP
1. Two Probes Arrangement, TPA-01
It has two spring load contact probes. These probes move in a pipe and are insulated by teflon washers. This probes arrangement is mounted in a suitable stand, which also holds the sample plate and RTD sensor. The stand also serves as the lid of PID Controlled Oven. Teflon coated leads are provided for connecting with High Voltage Power Supply EHT-11 and Digital Picoammeter DPM-111. With
-12this set-up assuming max. voltage = 1500V; current 100x10 A (max) and thickness of sample 1mm, the resistivity of the sample could be measured upto
1410 ohm.cm.
2. PID Controlled Oven, PID-200
This is a high quality temperature controlled oven suitable for Four Probe Set-up. The oven has been designed for fast heating and cooling rates, which enhances the effectiveness of the controller. While the basic design of the controller is around the PID configuration for its obvious advantages, wastage of power is avoided by using a Pulse Width Modulated (PMW) switch. This combination has the advantages of both on-off controller and linear PID controller. The result is a stable and accurate temperature control.
Platinum RTD has been used for sensing the temperature. A wheatstone bridge and an instrumentation amplifier are used for signal conditioning. Feedback circuit ensures offset and linearity trimming to a great degree of accuracy. The set and measured temperature are displayed on 3½ digit DPM through selector switch.
IGITAL MIC OVOLTM TERD
R E
M DE DMV 00O L - 1
N TA U E O CCO S NT C RR NT S UR E
D . 01MO EL NO CCS-
PID CONTROLLED OVENModel : PID-200
IME T
SCE NTI IC EQUIP NF
R EROO KE
SCI I E P NTENTIF C QUI ME
ROORKEE
SCIENTIFIC EQUIPMENTROORKEE
TPX-01
Fig. 1 : Block diagram of the ESR set-up
DPPH
OSCILLOSCOPE
HELMHOLTZ COIL
50 HzSWEEP UNIT
50 HzPHASE SHIFTER
X
Y
R.F.OSCILLATOR
DETECTORA.F.
AMPLIFIER
Electron Spin Resonance Spectrometer28
26
In this method, use is made of the Zeeman interaction of the magnetic dipoles associated with the nucleus or the electron, when placed in an external magnetic field. Accordingly, they are identified as NMR (Nuclear Magnetic Resonance) or ESR (Electron Spin Resonance). This form of spectroscopy finds many applications in many diverse branches of Physics, Chemistry and Biology.
This set-up is designed, keeping in mind the basic objectives of a classroom experiment. The observation of ESR in low magnetic field and in a radio frequency makes this apparatus very simple, inexpensive and within reach of every Post Graduate laboratory.
The Electron Spin Resonance experiment comes complete along with following accessories:
(i) Helmholtz coils with an attachment for the ESR-HC
(ii) ESR Sample: DPPH
(iii) R.F. Oscillator (10MHz to 19MHz)
The trace shows the resonancewhen the electron's precession
matches the oscillator frequency
ESR-104
Fig. A Glow Curve of the luminescent material(KCI crystal), the heating speed is 11K/min
T (K)
T =355KG
INT
EN
SIT
Y I (
nA
)
293273
10
0
20
30
40
50
60
70
80
90
313 333 353 373 393
29 Study of Thermoluminescence of F-centers
27
Study of Thermoluminescence of F-centers in Alkali Halide Crystals
Introduction
A colour centre is a lattice defect that absorbs visible light. These colour centers can be produced by heating the crystal in excess of alkali vapour or by irradiation. The colour centers formed due to trapped electrons or trapped holes can be destroyed (bleached) by illuminating with light or warming. Important information about the colour centers can be obtained from the plot of luminescence intensity vs. temperature, taken at a constant heating rate called glow curve. It may contain one or many glow peaks, depending upon whether there are one or several different kinds of traps.
The experiment consists of the following :
(a) Experimental set-up for creating Thermoluminescence
(1) Sample: KBr (TLS-KBr) and KCl (TLS-KCl) single crystal
(2) Thermoluminescence Temperature Meter, TL-02
• Digital Thermometer with RTD sensor
• Oven power supply
(3) Sample holder
(4) Thermoluminescence Oven (upto 423K), TLO-02
(5) Black Box, TL-BBx
(b) For measurement of Luminescence Intensity
(1) Photomultiplier tube 931A, TL-PMT
(2) PMT Housing with biasing circuit and connecting coaxial cables, TL-PMH
(3) High Voltage Power supply, EHT-11(Item no. 4, on page 4)
(4) Digital Nanoammeter, DNM-121 (Item no. 2, on page 3).
TLX-02 30 Hall Effect Experiment
28
The resistivity measurements of semiconductors cannot reveal whether one or two types of carriers are present; nor distinguish between them. However, this information can be obtained from Hall Coefficient measurements, which are also basic tools for the determination of carrier destiny and mobilities in conjunction with resistivity measurement.
The Hall effect experiment consists of the following:
a) Hall Probes (HPN-21, HPP-21)
Hall Probe (Ge Crystal)
Material : Ge single crystal n or p-type
Resistivity : 8-10 Wcm.
Contacts : Spring type (solid silver)
Zero-field potential : <1mV (adjustable)
Hall Voltage : 35-60mV/10mA/KG
b) Hall Effect Set-up, DHE-21, Consists of the following
(i) Digital Millivoltmeter
Range : 0-200mV(Resolution 100mV)
Accuracy : ±0.1% of reading±1 digit
Impedance : 1 Mohm
Display : 3½ digit,7 Segment LED
c) Electromagnet, EMU-50/ EMU-75 (Item no. 7/6, on page 6/5)
d) Constant Current Power Supply, DPS-50/ DPS-175 (Item no. 7/6, on page 6/5)
e) Digital Gaussmeter, DGM-202/ DGM-102 (Item no. 9/8, on page 7/6)
The experiment is complete in all respect.
(ii) Constant Current Generator
Currents : 0-20mA (Resolution 10mA)
Display : 3½ digit, 7 segment LED
Accuracy : 0.25%; ±1 digit
Load : 0.05% for no load toRegulation full load
Line : 0.05% for±10% changesRegulation
T C QU ES I N F E I M T
C E I I P N
ROOK EERSC E I I EQI MEN
I NT F C U P TO KRO R EE
HEX-21
Hall Effect enables the charge carrier concentration and mobility to be determined by experiment. Direction of the Hall Voltage in silver indicates negative charge carriers, which is in agreement with concepts of the model of the ‘free electron gas‘. Limitations of this model are shown by the so called ‘abnormal Hall Effect’ of tungsten. The experiment carried out under identical conditions for tungsten show the Hall Voltage to have about same magnitude but opposite direction as in silver.
This can be explained by the ‘Energy Band diagram’. The tungsten atom has 2 6 4 2…….5s 5p 5d 6s electronic structure.
When the atoms come close together to form the solid, the close lying states 5d and 6s broaden into bands, with s band broadening considerably more than the d band. This is because of the larger size of the s orbital. The figure schema-tically shows the allowed energies as a function of the interatomic distance. The number of allowed states is ten per atom in the d band and two in the s band. In tungsten there are six electrons to be shared between these two bands. The result is that at the interatomic distance in tungsten there are holes in the d band and electrons in the s band, making tungsten predominantly a hole conductor.
This sort of mixed (electrons and holes) conduction is a general characteristic of transition metals
The apparatus consists of the following:
Fig.: Allowed energies as a function of interatomic distance (d)
b) Hall Probe-Tungsten (HP-W)
Material : Tungten Strip(8 x 6 x 0.05 mm)
Contacts : Press type for current Spring Type for Voltage
Hall Voltage : ~15 mV/10A/10KG
a) Hall Probe-Silver (HP-Ag)
Material :Silver Strip(8 x 6 x 0.05 mm)
Contacts :Press type for currentSpring Type for Voltage
Hall Voltage :~17 mV/10A/10KG
31 Hall Effect in Metals HEM-01
29
E
5d
6s
1/d
HALL EFFECT IN METALSample thickness : 0.05 mm
Magnetic Field (KG)
30
c) High Current Power Supply, Model PS-20A
Range : 0-20A continuosly variableAccuracy : ±0.5%Regulation : ±0.5% for ±10% variation of mainsDisplay : 3½ digit, 7 Segment LED
d) Digital Microvoltmeter, DMV-001 (Item no. 1, on page 3)
e) Electromagnet, Model EMU-75T
Pole Peices : 75mm tappered to 25mm Mag. Field : 17KG 10mm airgap
f) Constant Current Power Supply, Model DPS-175 (Item no. 6, on page 5)
g) Gaussmeter, DGM-202/ DGM-102 (Item no. 7/6, on page 7/6)
The experiment is complete in all respect.
±5% at Energising Coils : Two of approx. 13W eachPower : 0-90Vdc, 3A, for coils in series
0-45Vdc, 6A, for coils in parallel
-60
-50
-40
-30
-20
-10
0
10
20
30
40
50
60
0 2 4 6 8 10 12 14 16 18 20
Ha
llV
olt
ag
e(
V)
I=19A
I=13A
I=7A
I=7A
I=13A
I=19A
Silver
I = Probe Current
m
Tungsten
32 Dependence of Hall Coefficient on Temperature
31
The Hall voltage for p-carriers has opposite sign from that for n-carriers. Therefore if a semiconductor with p-type doping is gradually heated up, more and more electrons from its valence band will go to conduction band. As a result hall voltage would fall rapidly with temperature and even become zero or change sign. At the point of zero Hall Coefficient, it is possible to determine the ratio of mobilities b=m /m . The Hall coefficient e h
inversion is a characteristic of only p-type semiconductors.
The set-up consists of the following. The Hall Probe comes complete with
1. Hall Probe (Ge: p-type), HPP-22
Ge single crystal with four spring type pressure contacts is mounted on a glass-epoxy strip. Leads are provided for connections with the current source andHall voltage measuring devices.
Oven
It is a small oven which could be easily mounted over the crystal or removed if required.
Temperature Sensor
Cromel-Alumel thermocouple with its junction at a distance of 1mm from the crystal.
2. Hall Effect Set-up :DHE-22
The set-up, DHE-22 consists of two sub set-ups, each consisting of further two units.
Measurement of Probe Current & Hall Voltage
This unit consists of digital millivoltmeter and constant current power supply. The Hall voltage and probe current can be read on the same digital panel meter through a selector switch.
(a) Digital Millivoltmeter
SpecificationsRange : 0-200mV (100mV min.)Accuracy : ±0.1% of reading ±1 digitImpedance : 1MW
(b) Constant Current Power Supply
SpecificationsCurrent : 0-20mA (10mA min)Accuracy : ±0.2% of the reading or ±1 digitLoad regulated : 0.05% for 0 to
full loadLine regulation : 0.05% for 10%
variation
Dependance of Hall Coefficient on TemperatureSample: Ge (p-type) medium doping (Hall Probe with small oven)
3-1
Ha
ll C
oe
ffic
ien
t x 1
0 (
cm
co
ulo
bm
)
28
26
24
22
20
18
16
14
12
10
8
6
4
2
0
-2
-40 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 115
0Temperature ( C)
HEX-22
SCIE EQUIPMENT
ORKEE
HALL EFFECT SET-UPModel : DHE-22
Hall Voltage (mV)
Probe Current (mA)
PROBECURRENT
HALLVOLTAGE
PROBEHEATER
CURRENTTHERMO
E.M.F.
ZE J. ON
Thermo e.m.t. (mV)
Heater Current (mA)
32
Apparatus for the Measurement of Susceptibility of Paramagnetic Solution by Quinck’s Tube MethodThe force acts on any substance when it is placed in a inhomogeneous magnetic field. This force depends on he susceptibility c, of the material, i.e., on ratio of intensity of magnetisation to magnetising field (I/H). Quantitatively it refers to the extent of induced magnetisation in unit field. The force acting on a substance, either of repulsion or attraction, can be measured with the help of an accurate balance in case of solids (Gouy's Method) or with the measurement of rise in level in a narrow capillary in case of liquids. From this measurement, the value of susceptibility can be calculated.
The apparatus consists of the following:
(a) Quinck’s tube, QTT-01, Quink Tube Stand (QTS-01)
(b) Sample: MnSO .H O4 2
(c) Electromagnet, EMU-50T / EMU-75T (Item no. 7/6, on page 6/5)
(d) Constant Current Power Supply, DPS-50 / DPS-175 (Item no. 7/6, on page 6/5)
(e) Digital Gaussmeter, DGM-202 (Item no. 9/8, on page 7/6)
(f) Travelling Microscope, TVM-02 (Item no 13, on page 9)
Note : For substances like MnSO .H O/ FeCl , EMU-50T is sufficient, but for substances like 4 2 3
water EMU-75T is required.
33 Quinck’s Tube Method
Measurement of Thermo emf and Heater current
The unit consists of a digital millivoltmeter and constant current power supply. The thermo emf of thermocouple and heater current can be read on the same DPM through a selector switch.(a) Digital Millivoltmeter
Specifications :Range : 0-20mVResolution : 10mV equivalent to 0.25°C in terms of thermo emfAccuracy : ±0.1% of reading ±1 digit
(b) Constant Current Power Supply
Specifications :Current : 0-1AAccuracy : ±0.2% of the reading ±1 digitLine regulation : 0.1% for 10% variationLoad regulated : 0.1% for 0 to full load
5. Electromagnet, EMU-50V (Item No. 7, on page 6)
6. Constant Current Power Supply, DPS-50 (Item No. 7, on page 6)
7. Digital Gaussmeter, DGM-202/ DGM-102 (Item No. 9/8, on page 7/6)
The experiment is complete in all respect.
QTX-01
In the Gouy method of susceptibility measurement, the solid sample in the form of a long cylinder (area of cross section A) is hung from the pan of a balance and is placed such that one end of the sample is between the pole-pieces of the magnet (field H ) and the other one is outside the field. The force exerted on the sample by the inhomogeneous magnetic field is obtained by measuring the apparent change (Dm ) in the mass of the sample. The susceptibility c is given by
2 c = 2Dmg/AH
If the sample is in the form of powder, it is filled in a long nonmagnetic tube which is then suspended from the pan of the balance.
The set up consists of the following:
(a) Scientific Balance, KSB-07
Capacity : 200 gms Sensitivity : 1/10 mg. by vernier Beam : Hard Bronze/ Brass Arrestment : Circular, falling away type Air Damping : Very quick and positive, beam coming to rest in 2-3 secCahinomatic Device : A gold plated chain is suspended from the beam with its other end
screwed on the rotating drum on which a scale graduated from 0 to 10 div each division representing 1mg is installed. By the movement of this scale before a vernier, reading upto 1/10th mg is taken
(b) Sample in the form of a long rod:
Set of 4 samples, 2 each of Ebonite and Wood
(c) Electromagnet, Model EMU-75TPole Peices : 75mm tappered to 25mm Mag. Field : 22KG mm airgap
(d) Constant Current Power Supply, Model DPS-175 (Item no. 6, on page 5)
(e) Gaussmeter, Model DGM-202 or DGM-102(Item no. 9/8, on page 7/6)
The experiment is complete in all respect.
at 5Energising Coils : Two of approx. 13W eachPower : 0-90Vdc, 3A, for coils in series
0-45Vdc, 6A, for coils in parallel
34 Gouy's Method GMX-01
33
Schematic Diagram Gouy’s Balance Experiment
A precise knowledge of various magnetic parameters of ferromagnetic substances is an important aspect of magnetic studies. These magnetic properties can be obtained from a magnetic hyteresis loop, which can be obtained by a number of methods. Some of these require samples in ring form, others can be used with films or even rock samples. Ring form samples are not always practical to make while in others, demagnetization effect sometime becomes important.
The present set-up can accept samples of thin wires. The effects of demagnetization, sample diameter and eddy currents have been taken into account.
Besides coercivity, saturation magnetization and retentivity, this set-up can also indicate the number of phases present.
The following magnetic properties of ferromagnetic sample can be studied, with this apparatus.
Coercivity Saturation magnetisation Retentivity Hyteresis loss
The number of phases present
The set-up is complete in all respect, including long solenoid, its power supply with a panel meter directly calibrated in gauss, sample holder with a pick-up coil and a set of samples: hard steel, soft iron and nickel (all in wire form). A measuring CRO will however be required.
35 Magnetic Hysteresis Loop Tracer
34
Fig. 1 J-H Loop Fig. 2 dJ/dt Loop 2 2Fig. 3 d J/dt Loop
LO ERHYSTERESIS OP TRACM el L Iod H T-II
SC E I U PMENI NT FIC EQ I T
OO KR R EE
HLT-111
36 Study of P-N Junctions
35
Study of the Energy Band-Gap and Diffusion Potential of P-N Junctions
This is an advanced level experiment to be performed on commercially available diodes viz. germanium or silicon diodes, various types of LED’s and also on the base-emitter/ collector-base junctions of a transistor. The results of the experiments not only give the device characteristics but also provide an insight into the properties of the materials used in the fabrication of the junction. In the set-up, all the necessary instrumentation is integrated as a result of which a minimum of external connections need to be made by the user. A CRO is the only accessory that is required.
The following studies can be carried out.
(i) Reverse saturation current l and 0
material constant h
(ii) Temperature coefficient of junction voltage dV/dt
(iii) Energy band-gap VG0
(iv) Junction capacitance
The experimental set-up consists of the following:
(1) Study of P-N Junctions, Model PN-1
(a) 3½ digit DPM for current/ temperature measurement.
(b) 3½ digit DPM for bias voltage/ junction voltage measurements.
(c) To connect the diode – one for experiment 1 & 2 and other for experiment 3.
(d) Two fixed frequency oscillators (5KHz & 20KHz) with the same output (200mV).
8
6
4
2
0
- 2
- 4
- 6
- 8
- 10
- 120.5 1 1.5 20
10
Fig. 1 Reverse Saturation Current
Junction Voltage (V)
h =
=
q
191.602 x 10- x 0.18-231.38 x 10 x 298 x 4.55
InI 0
InI
(I in
mA
)
I 0
= - 10.9
= -100.19 x 10 A
DV
Dln
I =
4.5
5
= 0.18
= 1.54
DV
kT
PN-1
DInI
Jun
ctio
n V
olta
ge
(V
)
0.8
0.7
0.6
290 310 330 350 370
Temperature (K)
Fig.2 Temperature Dependence
of Junction Voltage
SAMPLE: BC109
(BASE-EMITER JUNCTION)
=1.00 mA
(Constant for the set)
0.5
I
STUDY OF P-N JUNCTION MODEL PN-1
JUNCTIONCAPACITANCE
5 KHz 20 KHz
IN OUT
C.R.O.
BIAS VOLTAGE
BIAS JUNCTION
CURRENT
V
CURRENT TEMP
TEMP
ON
MAINS
ON
OVEN
mA
K
37 Semiconductors Diodes Characteristics
36
(2) Fast temperature controlled oven
with sensor, PNO-01
(3) PNJ-Sample
(a) Transistor BC109 (base-emitter)-Si
(b) Transistor AC128 (base-emitter)-Ge
(c) Diode: IN 5408/IN 4002-Si
The unit is supplied complete with a detailed instruction manual. Sufficient theoretical description is included for a proper understanding. This is followed by a step-by-step procedure and a typical set of readings and results.
The set-up is complete in all respect, except a CRO which is required for measurement of junction capacitance. Typical results obtained are shown in fig. no. 1, 2 & 3.
The set-up is
provided with a
booklet which
contains its theory of
operation,
description,
suggestions and
discussion of the
experiments that may
be performed with it.
Forward and reverse characteristics of Ge, Si diodes and LEDs
Study of Zener diode characteristics
CRO display of forward and reverse characteristics
CRO Display
80
70
60
50
40
30
202 4 6 8 100
Ca
pa
cita
nce
(pf)
Bias Voltage (V)
Depletion Layer Capacitance as a
Function of Reverse Bias Voltage
SAMPLE Si DIODE 1N4007
Fig. 3
Graph
24
20
16
12
8
4
0
Cu
rre
nt (m
A)
Voltage (V)
0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0
SDC-02
38
39
Study of a Transistor Amplifier (RC Coupled)Cum Feed-Back Amplifier
Study of Multivibrators
37
The following studies can be carried out with this set-up:
1. Study of the basic circuit of a RC Coupled Amplifier.
2. Frequency response of RC Coupled Amplifier.
3. Effect of negative feedback on the gain and frequency response of the amplifier.
4. Effect of positive feedback on the gain and frequency response of the amplifier.
5. Verification of the condition of oscillation.
6. Study of different classes of amplifier.
Accessories Required
(i) A.F. Oscillator (ii) True RMS A.C. Millivoltmeter, ACM-102(Item No. 5, on page 5)
The experiment consists of the following:
(i) Free Running vibrator
(ii) Univibrator
(iii) Bistable Multivibrators
(iv) Regulated Power Supply.
The free running multivibrator also serve as a pulse generator for the study of bistable multivibrator and univibrator.
Accessories Required
(i) Oscilloscope
RC-01
MV-01
40
41
Characteristics of Semiconductor Diodes
Study of Hybrid Parameters of a Transistor
38
The experimental set-up consists of the following:
1. Diodes: 1N4007, 1N34, 5.1V Zener and LED.
2. 3½ digit DPM for voltage measurement.
3. Precision resistances for measurement of forward and reverse current.
4. IC regulated power supply.
The following studies can be carried out with this set-up:
1. Forward and reverse characteristics of both Ge, Si diodes & LED’s.
2. Study of Zener diode characteristics.
The following studies can be carried out with this set-up.
1. Study of h parameter (input 11
impedance parameter).
2. Study of h parameter (output 22
admittance parameter).
3. Study of h parameter (forward 21
current transfer ratio).
4. Study of h parameter (reverse 12
voltage feedback ratio).
The experimental set-up is provided with a built-in power supplies and is complete.
Accessories Required:
(i) A.C. Millivoltmeter with built-in 1KHz Oscillator, ACM-103 (Item no. 5, on page 5)
(ii) Multimeter.
D-1
HP-01
42
43
44
Study of a Solid State Power Supply
Study of a Modulation and Demodulation with
Built-in Carrier Frequency Source (Solid State)
Study of Basic Operational Amplifier, Type-741
39
The following studies can be carried out with this set-up.
1. Study of rectification:
(a) Full wave rectification(b) Half wave rectification
2. Study of ac component (Ripples):
(a) Efficiency of various typeof filters – p, T type etc.
(b) The effect of load(c) The effect on regulation.
3. Regulation characteristics:
(a) The effect of load on regulation(b) The effect of change in main’s
voltage.
Accessories Required : (i) Multimeter; (ii) Oscilloscope
The following studies can be carried out with this set-up:
1. Carrier signal testing.
2. Variation of modulated wave with the modulation signal.
3. Study of detector circuit.
The experimental set-up is provided with a built-in power supply.
Accessories Required :
(i) Function Generator (Item No. 16)
(ii) Oscilloscope
The following studies can be carried out:
(i) Working of the basic circuit
(ii) Measurement of bias and offset currents
(iii) Study of inverting and non-inverting configurations
(iv) Study of Amplifier drift.
SSPS-01
MD-1
741-01
45
46
47
Study of 741 Applications
Study of Astable & Monostable Multivibratorusing Timer IC, Type-555
Study of an Integrated Circuit Regulator, Type-723
40
Features
Study of linear and non linear applications as:
– Integrator– Differentiator– Summer– Subtractor– Voltage to Current converter– Current to Voltage converter– Astable Mode– Precision rectifier
Built-in power supply
Built-in square wave and triangular wave Generator
Built-in current source.
The following studies can be carried out:
(i) Operation as a free running multivibrator
(ii) Operation as a monostable multivibrator
(iii) Operations as a preset time delay.
Accessories Required:
(i) Oscilloscope
The following studies can be carried out:
(i) Study of a voltage regulator
(ii) Study of a current regulator.
The experimental set-up consists of an IC 723 with facilities for convenient connections, an unregulated power supply, voltmeter, an ammeter and all the other components required to perform the experiments.
741-02
555
723
After his M.Sc. (Physics) from Allahabad University and Ph.D. from Calcutta University, Dr. Anand Prakash joined the California Institute of Technology, Pasedena, as a faculty member. During the period (1965-69), he worked on the crystal structures of various high pressure forms of ICE (ii to ix) by X-ray Diffraction and later by Neutron Diffraction at High Flux Beam Reactor at Brookhaven National Lab., N.Y. He also worked in the McMaster University, Canada on the crystal structure of alkoloids (1969-71). On returning, he joined the University of Roorkee, Roorkee. There he realized the need of new experiments in Physics which could bring insight into the subject and improve the quality of teaching. The present catalogue represents his 38 years of efforts in this direction. All the products here are designed and manufactured indigenously.
Dr. R. Mitra did his Ph.D. in the area of Control System in 1976 after completing his earlier education at IIT Kharagpur, University of Allahabad and University of Lucknow. Besides post dectoral work at DFVLR lab. in West Germany, he had been in the faculty of Dept. of Systems and Control, University of Technology, Baghdad and Thapar Engineering College, Patiala. For the past 32 years Dr. Mitra is with the Department of Electronics and Computer Engineering, IIT, Roorkee, teaching various courses in the areas of control systems, electronic instrumentation and microprocessor application. Dr. Mitra had been associated with us for over 30 years of teaching and research experience that we have been able to develop and perfect most of the equipment listed in this catalogue. Besides it is to his efforts that we have been able to diversity in the area of automatic control system, now providing about a dozen control experiments for engineering students.
Dr. Anand Prakash,Founder Chairman
Dr. R. Mitra,Chief Consultant
Control Laboratory, Instrumentation Lab, Measurement Lab Experimentsare also available
For further details contact our sister concern
261/16 Civil Lines, Roorkee -247 667, IndiaPh. : 91-1332-272852, Fax : 91-1332-274831E - mail : [email protected] us at http://www.sestechno.com
Formerly known as Scientific Equipment & Services, we have the pleasure of introducing ourselves as a reputed manufacturers of complete laboratory experiments and precision measuring instruments for BS and MS laboratories of Physics and Electronics.
For the last 39 years, we have been catering to the needs of over 1200 institutions in over 32 countries. We, while orienting ourselves to the need of laboratories have been trying our best to develop and introduce progressively more advanced laboratory experiments and general-purpose measuring instruments.
Our strict quality control ensures uncompromising quality of our instruments, low overheads and direct sales to the customers have enabled us to keep our rates very affordable.
The product catalogue highlights some of our manufacturing activities. The technical specifications given here are meant to give an accurate statement of performance of instruments manufactured by us.
We at SES Instruments Pvt. Ltd.(Formerly Scientific Equipment & Services), are greatful to the staff of IIT-Roorkee, particularly Professor K.C. Mittal, and Professor M.K. Srivastava, for their tremendous support all through these years by giving innovative suggestions and constructive criticism thereby helping in constantly adding new products and improving the existing ones.
All our products are manufactured under high level of quality control. They are warranted free from defects for a period of one year from the date of purchase. We will repair or replace as find suitable, any piece of defective equipment during this period. Even after this period after sales service is undertaken at very nominal charges.
Specifications of all products are subject to change without notice. SES Instruments Pvt. Ltd. reserves the right to make improvements to the products without incurring any obligation to incorporate these changes in products previously sold.
Equipment in this catalogue are designed to operate from 220V ±10%, 50Hz. All the equipment are also available for use with 110V ±10%, 60Hz power at no additional cost.