SIGMACHECKHANDHELDEDDYCURRENTCONDUCTIVITYMETER
USERMANUAL
DocNo:50024:Issue2
7/30/2019 SigmaCheck User Manual_02
2/26
2
TableofContents
1 Introduction...................................................................................................3 1.1 AbouttheSigmaCheck.....................................................................................................................3
1.2 Abbreviations.......................................................................................................................................3
1.3 Keys..........................................................................................................................................................41.4 Connectors............................................................................................................................................4
2 StandardPackageandaccessories..................................................................52.1 StandardPackage...............................................................................................................................5
2.2 RecommendedAccessories...........................................................................................................6
2.3 OptionalAccessories.........................................................................................................................6
3 Operationoftheinstrument...........................................................................73.1 Appearance...........................................................................................................................................7
3.2 Operationprinciple...........................................................................................................................7
3.3 Screenlayout........................................................................................................................................7
3.4 MenuScreen/Structure.................................................................................................................7
3.5 ReadingsScreen..................................................................................................................................8
3.5.1 Freeze................. .......... .......... .......... .......... .......... ......... .......... .......... .......... .......... .......... ......... .......... .....9
3.6 Home/InstrumentStatusScreen..............................................................................................9
3.7 Setup.....................................................................................................................................................10
3.7.1 Calibration.........................................................................................................................................10
3.7.2 Frequency...........................................................................................................................................10
3.7.3 Probe.....................................................................................................................................................11
3.7.4 Advanced.......... .......... ......... .......... .......... .......... .......... .......... ......... .......... .......... .......... .......... .......... ...11
3.7.5 Configure.......... .......... ......... .......... .......... .......... .......... .......... ......... .......... .......... .......... .......... .......... ...14
4 Charging.......................................................................................................18
5 UsingwithaPC.............................................................................................185.1 ConnectingtoPC..............................................................................................................................18
5.1.1 SigmaCheckPC..................................................................................................................................18
5.1.2 ETherFUU(FirmwareUpdateUtility)...................................................................................21
5.2 Usingaspreadsheettoanalysedata......................................................................................23
6 Setting-upprocedures..................................................................................24 6.1 Calibratingandmakingareading............................................................................................24
6.2 Usingthe2ddataloggingfunction..........................................................................................24
6.3 ALARMsetup.....................................................................................................................................25
7 CareandMaintenance..................................................................................25
8 TroubleShooting..........................................................................................25
9 TransportationandStorage..........................................................................25
10 Specifications............................................................................................26
7/30/2019 SigmaCheck User Manual_02
3/26
3
1 Introduction1.1 AbouttheSIGMACHECK
TheSIGMACHECKisdesignedtobeextremelyeasytousemakingitsuitableforuseby
shopfloorinspectorstoengineers.
TheSIGMACHECKmeetstherequirementsforconductivitymeasurementsinthe
aircraftmanufacturingandmaintenancefieldswhereconductivitymeasurements
areusedtoverifyproperalloy/temperordetectheatdamagedcomponents.The
SIGMACHECKisalsousefultodeterminethepurityofpreciousmetals(counterfeit
detection)suchasgoldbullionandcoins.
TheSIGMACHECKhasalarge,colour,LCDdisplaythatiseasilyseeninalllighting
conditions
TheSIGMACHECKisdesignedforconvenientsinglehandedoperationduetoits
sculptedhousinganditsconvenientsizeandweight;onlyofapound(350g.)and
measuringlessthan6tallx3widex1thick(163x80x25mm).Aremovable,
siliconerubberbootisincludedtoprotecttheinstrumentinroughenvironments.
ThestandardProbehasbeendesignedtocomfortablyfitthehandoftheuser.
TheSIGMACHECKhasastandardoperatingfrequencyof60kHzandauxiliary
frequenciesof:120,240,480and960kHzcanbeselectedwhennecessaryfortesting
thinmaterialswiththestandardprobe.
OtherfeaturesoftheETHERNDESIGMACHECKinclude:
FieldExchangeableProbesviaSDCard TrueUSBPCConnectivityforremotecontrolanddatalogging. TheUSBConnectionallowseasychargingoftheInstrumentwithouthaving
toswapthebatteries.
TheSIGMACHECKissuppliedwithaflameretardantABScase,adual(Hi-Low)
calibrationblock,probecable,charger,USBlead,removabledeskstandanduser
manual.Optional,smalldiameterprobewillbeoffered.
1.2 Abbreviations
C Centigrade
F Fahrenheit
IACS InternationalAnnealedCopperStandard
MicroSD MicroSecureDigitalFlashMemoryCard
Mil Thousandthsofaninch(0.001)US
MS/m MegaSiemenspermetre
Thou Thousandthsofaninch(0.001)UK
7/30/2019 SigmaCheck User Manual_02
4/26
4
1.3 Keys
TheSIGMACHECKinstrumenthas7keys.4arrowkeys(Up,Down,Left,Right),anOK
key(shownwithatick),aBackorCancelkey(shownwithacurlyarrow)andaPower
key.
Up
OK
Left Right
Down
1.4 Connectors
Thereare2ConnectorsontheSIGMACHECK
AMini-BUSBconnectorisontheleft.Thisisusedfor:
a) Chargingtheinstrumentifrechargeablebatteriesareinstalled.b) PoweringtheInstrument.c) ConnectingtheinstrumenttoaPCfordataupordownload.
Back/CancelPower
A5pinLemoconnectorisontherighthadsidefor
connectingprobes
7/30/2019 SigmaCheck User Manual_02
5/26
5
2 StandardPackageandaccessories2.1 StandardPackage
AllofthefollowingAccessoriesshouldbeavailablebeforeusingtheSIGMACHECK
instrument,althoughlimitedoperationcanbeperformedwithout.
NotpicturedisamainsconnectortowhichtheUSBcablecanbeconnectedtoifnot
connectedtoaPCforchargingand/oroperation.
DualsampleCalibrationBlock
MicroSDmemorycard
SigmaCheck
USBcable
Probe
Quantity PartNo Description
1 ISIG001 SIGMACHECKInstrument
1 ASIG001 DualconductivitystandardwithCuandStainless
Steel
1 ASIG002 Adjustableinstrumentstand
1 40222 SiliconeRubberBoot1 ALL05-L05-012-SIG ProbeCable
1 PSIG001 Probe13mmdia.60-480kHz
2 30239 2xErgonomicProbeSkirtsfortubes&easyProbe
handling
1 A090 USBCableforPCconnection/Charging
1 A092 UniversalUSBChargerAC100-240VAC
1 A089 MicroSDCard
1 40293 QuickReferenceGuide
1 60005 ManualandUtilityonUSBstick
1 AC001 SoftCarryingCase
7/30/2019 SigmaCheck User Manual_02
6/26
6
2.2 RecommendedAccessories
PartNo Description
ASIG004 PeliHardCasewithcustominsert.
ALL05-L05-012-SIG ProbeCable
PSIG001 Probe13mmdia.60-480kHz
2.3 OptionalAccessories
PartNo Description
PSIG002 7mmProbe240-960kHz
PSIG001R RightAngle13mmprobe60-480kHz
PSIG002R RightAngle7mmprobe240-960kHz
ASIG004 PeliHardCasewithcustominsert.ATBC-COPPER Accessory,TestBlock,Conductivity,Copper,(100.0-103.6%IACS)
ATBC-ALU1200
Accessory,TestBlock,Conductivity,AluminiumAlloy,1200-H4,(58.5-60.0
%IACS)
ATBC-ALU6082
Accessory,TestBlock,Conductivity,AluminiumAlloy,6082-T6(44.0-48.0
%IACS)
ATBC-ALU6061
Accessory,TestBlock,Conductivity,AluminiumAlloy,6061-T4(42.3-43.8
%IACS)
ATBC-ALU2014A-T6
Accessory,TestBlock,Conductivity,AluminiumAlloy,2014A-T6(38.0-42.0
%IACS)
ATBC-ALU2014A-T4
Accessory,TestBlock,Conductivity,AluminiumAlloy,2014A-T4(34.0-37.0
%IACS)
ATBC-ALU7075
Accessory,TestBlock,Conductivity,AluminiumAlloy,7075-T6(31.4-34.8
%IACS)
ATBC-ALU5083Accessory,TestBlock,Conductivity,AluminiumAlloy,5083(27.5-28.5%IACS)
ATBC-BRASS Accessory,TestBlock,Conductivity,Brass,CZ121(24.0-26.0%IACS)
ATBC-PBRONZE Accessory,TestBlock,Conductivity,PhosphorBronze(17.0-22.0%IACS)
ATBC-NICSILVER Accessory,TestBlock,Conductivity,NickelSilver,LC1291(8.5-9.5%IACS)
ATBC-STST303S Accessory,TestBlock,Conductivity,StainlessSteel,303S(2.0-3.0%IACS)
ATBC-TITANIUM Accessory,TestBlock,Conductivity,Titanium,6AL-4V(1.0-2.2%IACS)
ASIG003 5Testblockholder.
AC001 SoftCarryingCase
7/30/2019 SigmaCheck User Manual_02
7/26
7
3 Operationoftheinstrument3.1 Appearance3.2 Operationprinciple
TheSIGMACHECKisasimpletouseConductivityMeterwithLift-Offcapability.
TheSIGMACHECKcontainsmanymodernfeatures,forexampleaMicroSDcardis
includedintheinstrumentsothatlargeamountsofdatacanbestored.AUSB
connectorispresentsothattheinstrumentcancommunicateinreal-timewiththe
PCsoftware.
Onceconfigured,theSIGMACHECKcanveryquicklybepickedup,turnedon,calibrated
andthenaccurateandfastmeasurementscanbetakenofConductivityorLift-Off.
3.3 ScreenlayoutTheinstrumenthas2mainscreenlayouts,themenuscreenandthereadingsscreen.
TheMenuscreenallowsallaspectsoftheinstrumenttobeconfigured,fromthe
inspectiondetailstotheindividualuserspreference.TheReadingsscreenhas2
panes,atopandabottom.
3.4 MenuScreen/Structure
ThemenusystemontheSIGMACHECKisdesignedalongthe
linesofthefolder/filelayoutpresentonmanymodern
computers.Ifamenuitemcontainschildmenuitems
beneathit,itwillhaveaplus(+)symbolafterit,meaning
thatitcanbeexpandedtorevealmoremenuitems.Ifa
menuitemhasaminus(-)afteritstitlethenitschildmenu
itemsarealreadyvisible.ThisscreenshotshowstheTop
levelSIGMACHECKmenuwithits4submenus.Advancedand
Configurecontainmoresubmenus,hencetheir+.
TheUpandDownkeysmoreupanddownchildmenuitems
arethecurrentlevel.TomoveupaleveltheLeftorBack
keysshouldbeused.Toenteramenuitemssubitems,the
RightorOKkeysareused.
7/30/2019 SigmaCheck User Manual_02
8/26
8
3.5 ReadingsScreen
SelectingthetoplevelSIGMACHECKmenuitemchangestheinstrumentintoReadings
Mode.Herethescreenhas2configurablepanesandastatusbar.PressingtheUpor
Downkeystogglestheselectedpane,thisisvisiblebythetitleofthepanebeing
highlighted.PressingtheLeftorRightkeyschangesthedisplayoftheselectedpane
throughthefollowingpossibilities:
LiftOff
Conductivity
LiftOffBarGraph
ConductivityMeter
Time&Date
MetalSort
TheLowerPanecanalsobeOFF,whichexpandstheUpperPane.IfDataLoggingis
enabledthentheLowerpanewillONLYshowDataLogginguntilitisdisabled.
Examplesofthesevariousdisplaymodesareshownhere:
WiththeLowerPanesettoOFF.ExamplesofLift-Offbargraph,ConductivityMeter
andConductivity:
7/30/2019 SigmaCheck User Manual_02
9/26
9
TheStatusBarshows4AlarmLEDrepresentations,thetemperaturereadbythe
probe,aMicroSDcardiconandaBatteryMeterorUSBindicator.
Seethe3.7.4.1Alarmssectionbelowforanexplanationoftheirfunctionality.The
temperaturecanbedisplayedinCelsiusorFahrenheitdependingontheUnits
setting;seethe3.7.5.3Unitssectionbelow.TheMicroSDiconwillbewithorwithout
aredcross.Withoutaredcrossindicatesthatthecardispresentandfunctioning
correctly,withtheredcross,thereisafailureofthecard.Thebatterystatesare
Empty,25%,50%75%,FullorUSBConnected.
or
3.5.1 FreezeWheninReadingsmodepressingtheOKkeywillfreezethecurrentdisplay.Any
alarmsthatwereflashingwillremainilluminated.PressingOKagainwillremovethe
freezeandthereal-timereadingswillcontinue.Pressinganyotherkeywillalso
cancelthefreezeandthekeywillbeactioned.
TheFreezefunctionalitycannotbeusedwheninDataLoggingmode(asOKlogsa
reading)oriftheinstrumenthasnotbeencalibratedsincethelastpower-on(asOK
movestheSIGMACHECKtothecalibrationmenu).
PleasenotethattheFreezefunctionalityfreezesthedisplayonlyandNOTanyrealtimedatabeingoutputtedovertheUSB.
Toidentifywhenthescreenisfrozen,thescreentitletextofSIGMACHECKwill
changecolour.
3.6 Home/InstrumentStatusScreen
TheHomescreenshows3usefulpiecesofinformation,thecurrentProbefilethatis
loadedin,thecurrentFrequencythatisbeingusedandwhentheinstrumentshould
nextbecalibrated.
7/30/2019 SigmaCheck User Manual_02
10/26
10
TheselectedProbeshouldbethesameaswhatisactuallyconnectedtothe
instrument.Ifthesedifferthenthereisnoguaranteethatanyvaluesdisplayedwill
beaccurate.IfNoprobeInfoisdisplayed,thenaProbeneedstobeselectedfrom
theprobemenu,see3.7.3below.
Thefrequencyisdependentonwhatisrequiredfortheinspectionandwhatthe
probehasbeencalibratedtododuringmanufacture.
TheAnnualCalDuedateisthedatebywhichtheinstrumentandprobeshouldbe
re-calibrated.Thisdatewillflashonceithaspassed.Theinstrumentwillcontinueto
functionbutthereisnoguaranteethatanyvaluesdisplayedwillbeaccurate.
3.7 Setup
AllaspectsoftheSIGMACHECKcanbeconfiguredusingthemenus.Navigationofthe
menusystemisexplainedabovein3.4MenuScreen/Structure.
3.7.1 Calibration
Calibrationshouldbecarriedeachtimetheinstrumentisturnedon
andperiodicallythereafter.Thecalibrationprocedureisexplained
below,see6.1Calibratingandmakingareading.
Tocalibratetheinstrument,theCalibrationBlocks(CB1andCB2)
valuesMUSTbesettothevaluesoftheblocksattachedtothetopof
theinstrument.TochangeavalueuseUporDowntohighlightthe
desiredCalibrationBlock.UseLeftorRighttobegineditingthevalue.
LeftandRightwillchangethecurrentlyeditingcharacterandUpand
Downwillcyclebetween0and9.MovingLeftfromthemost
significantdigitwilladdanewdigituptoamaximumof3significant
figures.MovingRightfromasignificantfigureof0willremoveit.
Whenthevalueiscorrect,presstheOKbuttontosaveit,orBackto
canceleditingandreturnthevaluetoitsoriginalvalue.
SelectingthemenuoptionCalibratewillbeginthecalibrationprocedure.
PressOKorBacktoleavethismenu.
3.7.2 Frequency
Onenteringthismenuallavailablefrequenciesforthecurrently
loadedprobewillbelisted.Thecurrentfrequencywillbehighlighted.
Acrossafterthefrequencyindicatesthatthereisaproblemwiththe
probedetailsatthatfrequencystoredontheMicroSDcard.Atick
indicatesalliswellandthefrequencymaybeselected.
Left/UporRight/Downwillmoveupanddownthelistofavailable
frequencies.PressOKtoselectafrequencyorBacktoleavethe
currentselectedfrequency.
7/30/2019 SigmaCheck User Manual_02
11/26
11
3.7.3 Probe
Onenteringtheprobemenuthecurrentselectedprobewillbeshown.UpandDown
cyclethroughtheheadingsofS/N(SerialNumber),freq(frequency)andanynotes
thatarepresent.
MovingLeftorRightontheS/Nwillcyclethroughallavailableprobe
filesontheinstrumentandupdatetherestofthescreenwiththe
detailsofthatprobefile.
MovingLeftorRightonthefreqheadingwillmoveupanddownthe
availablefrequenciesforthisprobefile.PressingOKwhileafrequency
ishighlightedwillloadtheprobefileandconfiguretheinstrumentto
usethatfrequency.PressingOKwithnofrequencyhighlightedwill
loadintheprobefileandusethefirstfrequencyinthelist.
MovingLeftorRightwiththeNoteshighlightedwillscrollthenotesin
thecasethattherearemorecharactersthanfitontothescreen.PressingBackwillleavetheinstrumentwiththecurrentprobefileand
atthecurrentfrequency.
3.7.4 Advanced
3.7.4.1 Alarms
AvisualalarmcanbeconfiguredtotriggerwhenaConductivityorLift-Offexceedsor
dropsbelowapredeterminedvalue.Whenanalarmistriggeredandthevalueis
visibleonthereadingsscreenthereadingwillflashbetweentheValueandAlarm
colours(see3.7.5.2Colourbelow).InthealarmstatethecorrespondingLEDonthe
statusbarwillflash.TheLEDrelatestothevaluesbeingshownontheUpperorLowerpanesinthereadingsscreen.
ThefirstLEDindicatesthattheUpperPanevaluehasexceededitshighalarmtrigger
value.
7/30/2019 SigmaCheck User Manual_02
12/26
12
ThesecondLEDindicatesthattheUpperPanevaluehasdroppedbelowitslow
alarmtriggervalue.
ThethirdLEDindicatesthattheLowerPanevaluehasexceededitshighalarmtrigger
value.
TheforthLEDindicatesthattheUpperPanevaluehasdroppedbelowitslowalarm
triggervalue.
TheabovediagramillustratestheUpperpanevalueexceedingitshighalarmvalue
whiletheLowerpanevaluehasdroppedbelowitslowalarmvalue.
WhenenteringtheAlarmmenu,fouralarmsareshownusingthe
unitsspecifiedintheUnitsmenu.AHighandLowvaluefor
ConductivityandThicknessisavailable.
TochangeavalueuseUporDowntohighlightthedesiredAlarm.
UseLeftorRighttobegineditingthevalue.LeftandRightwillchange
thecurrentlyeditingcharacterandUpandDownwillcyclebetween0and9.MovingLeftfromthemostsignificantdigitwilladdanew
digituptoamaximumof3significantfigures.MovingRightfroma
significantfigureof0willremoveit.
Whenthevalueiscorrect,presstheOKbuttontosaveit,orBackto
canceleditingandreturnthevaluetoitsoriginalvalue.
PressOKorBacktoleavethismenu.
3.7.4.2 MetalSortTheMetalSortfunctionalityallows10differentrangesofConductivitytobeentered.
Ifconductivityisencounteredthatfitswithinoneoftheserangesthetextassociated
withthatrangewillbedisplayedontheMetalSortpane.
Thismenuitemallowstherangetobeentered(alowandhigh
conductivityvalue)andanamethatwillbeshown.Theunitsof
conductivitywilldependontheUnitsmenuitem.
UsingtheUpandDownkeys,thehighlightedoptionwillgofrom
NametoLowvaluetoHighvalueandthentothenextmetal,andso
onuptothe10thmetal.PressingOKwillbegineditingofthe
highlightedvalueorname,pressingOKagainwillsavethenew
name/valuewhilstpressingBackwillcanceleditingandreturnthe
name/valuetoitsoriginalvalue.Fordetailsoneditingvalues,see
3.7.4.1Alarmsabove.Editingofcharactersissimilar.LeftandRight
changethecharacterbeingeditedandUpandDowncyclethrough
theavailablecharacterslikeso:
0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZabcdefghijklmnopqrstuvwxyz
**Pleasenote,forametaltobeshownonthereadingspane,theMetalSortpane
mustbepresentontheUpperorLowerpane.**
7/30/2019 SigmaCheck User Manual_02
13/26
13
WhenallchangeshavebeenmadepressOKorBacktoreturntotheAdvanced
menu.
3.7.4.3 DataLoggingTheDataLoggingfunctionality,whenenabled,allowsConductivity
readingstobeloggedatthepressoftheOKbutton.Thereare2
locationstostoreDataLoggingdata,MemoryCardorRAM.The
differencesare:
RAM
Thenumberofreadingsislimitedtoabout8000(Rowsmultipliedby
Columns).
Oncepowerisremovedthereadingsarelost.
SigmaCheckPCcancontinuetooperatefully.DataLoggingdatacanbeextractedatanypointusing
SigmaCheckPC.
UsingtheLeftorRightkeys,previousdatacanbeoverwritten.
MemoryCard
WhileDataLoggingisenabled,anycardaccesswillstopdatalogging.
DatacannotbeuploadeduntilDataLogginghascompletedorbeenstopped.
Onceavaluehasbeenlogged,itcannotbechanged.
ManyseparatedataLoggingsessionscantakeplaceandalldataissaved.
ThesereadingsarestoredintheSIGMACHECKsmemoryuntiltheyareuploadedtoaPCviatheUSBcableandthePCSoftwarepackage.Theresultingfileisdesignedto
bereadablebymostSpreadsheetpackages,hencewhyontheDataLoggingmenua
numberofRowsandColumnsareshown.Whenenabled,theDataLoggingpaneis
shownintheLowerpaneofthereadingsscreen.ItshowsthecurrentColumn&Row
numbersandbelowthat,thelastconductivityvaluethatwaslogged.Whenall
ColumnsandRowsarefull,thescreenwillshowLoggingFull.Ifavaluehasbeen
takenincorrectlythepressingLeftorRightwillmovebackorforwardsbetween
readingsallowinganewreadingtobetakenandoverwritetheexistingvalue.When
movingbackwardsandforwards,thepreviousreadingatthislocationwillnotbe
shown.
Ifasimplelistofvaluesisallthatisrequired,thensimplysetthe
RowsorColumnsto1andtheothervaluetotherequirednumberof
readings.ThetotalnumberofreadingstakenwillalwaysbetheRows
multipliedbytheColumns.Aseachreadingistaken,theSIGMACHECK
incrementstheColumnfirst,thenwhentherowisfull,theRowis
incremented.Thereforetheresultingspreadsheetwillhavethefirst
readinginthetopleftcornerthenreadinglikeabooklefttoright,
toptobottom.
ForinformationondownloadingtheDataLoggerdata,see5.1
ConnectingtoaPCbelow.
7/30/2019 SigmaCheck User Manual_02
14/26
14
OntheDataLoggingmenu,theUpandDownkeyswillchangethehighlighted
heading.LeftorRightkeyswillbegineditingoftheRows/ColumnsortoggletheData
Loggingenabledoption.Foreditingofthevaluessee3.7.4.1Alarmsabove.
Pleasenote,ifDataLoggingisenabledhere,thelowerpaneonthereadingsscreen
willONLYshowDataLogging.
3.7.4.4 MemoryCard
TheSIGMACHECKcontainsaslotforaMicroSDmemorycard.Correct
functioningoftheinstrumentrequiresacardtobefitted(the
instrumentissuppliedwithacard).Thefilesandfolderstructureof
thecardcanbeseenusingthismenuitem.
OnselectingMemorythecontentsofthecardareshown.Foldersareshownwithalloftheircontentsindented.FoldersareintheText
colour,filesintheValuecolour(see3.7.5.2Colourabove).
Iftherearemorefiles/foldersthanfitontothescreen(evidentby
theverticallinetotheleftgoingoffofthebottomofthescreen)
thenpressingthedownkeywillscrolltheview.
IftheOKkeyispressedwhileafileishighlightedtheFileDetails
screenisshown.Itshowsthefilenamealongwithitscompletepath.
Thefilesize(-1ifthefileisinfactadirectory).Thereisthen2
options,DeleteandUpload.Deletewillpermantelydeletethefile.
UploadwillsendthefiletoSigmaCheckPCiftheSIGMACHECKis
connectedtoaPCviatheUSBandSigmaCheckPCisconnectedtothe
SIGMACHECK.Whenthefiletransferisinitiatedaprogressbarwillbe
shownonSigmaCheckPCmainscreen.Whencomplete,aSaveFile
windowwillopenonthePC.
PressBacktoreturntotheAdvancedmenu.
3.7.5Configure
TheConfiguremenucontainsthefollowingsub-menus:
Brightness Colour Units Precision Powersave Time&date
7/30/2019 SigmaCheck User Manual_02
15/26
15
3.7.5.1 Brightness
Thebrightnessmenuallowsthebrightnessofthescreentobe
changed.Loweringthebrightnesswillextendthebatterylifeofthe
SIGMACHECK.
PressingLeftorDownwilldecreasethebrightnessby10%
PressingUporRightwillincreasethebrightnessby10%.
PressingOKwillsavethecurrentbrightnesslevel.
PressingBackwillreturnthebrightnesslevelbacktoitsprevious
value.
3.7.5.2 Colour
4Coloursmaybechanged.
TheText,Background,ValueandAlarms.Thetextcolouristhecolourmostoftenusedformenuitemsandheadingsonthereadingsscreen.Thebackgroundcolouris
self-explanatory.TheValuecolouriswhatisusedforthenumbersonthereadings
screenandseveralmenuitems,theAlarmcolourshouldbenoticeablydifferentto
thevaluecolourthatisusedtoindicatethatavalueisinthealarmstate.Thecolours
usedforthemajorityofthescreenshotsinthismanualshowninthefirstscreen
shot.Theothersarepopularalternatives.
PressingUporDowncyclesbetweentheheadings.PressingLeftorRightcycles
throughtheavailablecolourswhichare:
BLACK,WHITE,RED,GREEN,YELLOW,BLUE,CYAN,GREY,PINK,CREAM
PressingOKwillsaveandusethecurrentcolourscheme.PressingBackwillgoback
tothepreviouscolourscheme.Itispossibletosetthetextandbackgroundcolours
tothesamecolourwhilecyclingthroughthecoloursbutitISNOTpossibletoaccept
thesecolourstobethesame.
7/30/2019 SigmaCheck User Manual_02
16/26
16
3.7.5.3 Units
Theunitscanbesettoanyofthefollowing:
Conductivity-%IACSorMS/m
TemperaturedegreesCelsiusorFahrenheit
Length(Lift-Off)mm,Thouormil
(Thou&mil=Thousandthsofaninch).
TheUpandDownkeyscyclebetweentheheadings,LeftandRight
toggletheUnitsforthatheading.
PressOKtoacceptthenewunitsorBacktoreturntothe
Configuremenuandleavetheunitsunaffected.
Theunitsselectedhereaffectallplaceswherethevaluesare
displayedontheinstrument,includingothermenus(i.e.Alarms).
TheunitswhendownloadingtothePCorviewingontheSIGMACHECK
PCSoftwareisfixedto:Conductivity-%IACS
TemperaturedegreesCelsius
Lengthmm
3.7.5.4 Precision
ThePrecisiononlyaffectshowmanydecimalplacesareshownonthescreenofthe
SIGMACHECK.TheConductivitycanhave0,1,2,3orAutodecimalplaces.
Temperature,0or1decimalplacesandLength0,1or2decimalplaces.Autochangestheprecisionofconductivityreadingsdependingonthevalue:
0to10%IACS=3decimalplaces.
10to100%IACS=2decimalplaces.
100andabove%IACS=1decimnalplace.
PressingtheUporDownkeyscyclesthroughtheheadings.
PressingLeftdecreasesthenumberofdecimalplacesand
pressingRightincreasesthenumber.
Thenumberbeloweachheadingisarepresentationofanumber
displayedtothecurrentnumberofdecimalplaces.
PressOKtoacceptthenewnumberofdecimalplacesandreturn
totheconfiguremenuorBacktoreturntothemenukeepingthe
previousnumberofdecimalplaces.
7/30/2019 SigmaCheck User Manual_02
17/26
17
3.7.5.5 PowerSave
Thereare2methodsofsavingpower.Theinstrumentcanbeset
toautomaticallydimthescreenafterasettimeperiodand/or
turnofftheinstrumentcompletelyafterasettimeperiod.These
timeperiodsrangeformOff(disabled)to60minutesin5minute
steps.
PressingtheUporDownkeyscyclesbetweentheheadings.
PressingtheLeftkeydecreasesthetimeinstepsof5from60to
Off.
PressingtheRightkeyincreasesthetimeinstepsof5fromOffto
60.
PressOKtousethenewvaluesorBacktousethepreviousvalues.
3.7.5.6 Time&Date
TheSIGMACHECKcontainsarealtimeclock.Itenablesthetimeand
datetobeviewedonapaneintheReadingsscreenandisalso
requiredforwritingdatatotheMicroSDcard.Thedateandtime
caneasilybesetandwillberememberedevenwhenthereareno
batteriesorUSBpower.
PressingLeftorRightcyclesbetweenthefields:
Day->Month->Year->Hour->Minute->Second.PressingUporDownwillincrementordecrementthehighlighted
value,exceptforsecondswhereUporDownwillsetthenumber
ofSecondsto0.
Whenfinished,pressingOKwillsetthetimeoftheSIGMACHECKtowhathasbeen
entered.PressingBackwillleavetheTime&dateunchanged.
3.7.5.7 Language
ThemajorityofthetextdisplayedbytheSIGMACHECKistranslatedintoseveral
differentlanguages.OnthelanguagemenupressingtheLeftorRightkeyscycles
throughthecurrentlyavailablelanguages.
PressOKtoacceptthenewlanguageorBacktoreturntotheConfiguremenuand
returnthelanguagetothepreviousone.
7/30/2019 SigmaCheck User Manual_02
18/26
18
4 Charging
TheSIGMACHECKmayeitherbechargedusingthesuppliedACcharger.Alternatively
usingtheUSBcabletheunitmaybechargedfroastandardPCUSBconnector.Note
whilsttheunitispowereduptheUSBchargerwillpowertheinstrumentbutchargingofthebatterieswillnotoccuruntiltheunitispowereddown.
AstheinstrumentusesAAbatteriesthennonrechargeablebatteriesmaybeused.
Thesewillnotbechargedasthereisprotectioncircuitrytopreventthis.
5 UsingwithaPC5.1 ConnectingtoPC
TheSIGMACHECKconnectstoaPCviaitsminiUSB-Bconnector.Thisconnectionprovidesallcommunicationsandchargingfunctionality.
Thereare2PCapplicationsthatusethisUSBconnection:
SigmaCheckPCandtheETherFUU(EtherFirmwareUpdateUtility)software
packages.
WhenaSIGMACHECKisfirstconnectedtoaPCtherequireddriverswillautomatically
beinstalled.SIGMACHECKusesWindowsdefaultdriverssonoadditionaldriversare
required.
5.1.1 SigmaCheckPC
Thisapplicationhasseveraluseswhenconnectedtoa
SIGMACHECK:
Displayreal-timevaluesofConductivity,Lift-OffandTemperature
DisplaytheUniqueID(nottheserialnumber)oftheSIGMACHECK.
FilehandlingoftheSIGMACHECKsmicroSDcard. UploadDataLoggerfilesfromtheSIGMACHECK. RemotecontroltheSIGMACHECK.
Upto4SIGMACHECKscanbesimultaneouslyconnected,butonly
onecanbeusedatatimebySigmaCheckPC.Theactiveunitcan
quicklyandeasilybechangedbetweenanyofthefourbyaclick
ofoneofthenumberedbuttonsalongthetopofthescreen.
7/30/2019 SigmaCheck User Manual_02
19/26
19
5.1.1.1 Real-Timedisplay
AssoonastheSIGMACHECKispowered,connectedtoSigmaCheckPCandthe
instrumentsplashscreenacknowledged,real-timevalueswillbedisplayedonthe
SigmaCheckPCscreen.AcrossthetopofthedisplayistheUniqueIDofthe
instrument,followedbyEtherNDEandSIGMACHECK.
Inlargertextbelowthatistheconductivity.Inlargetextinthelowerhalfofthe
screenistheLift-Off.
Acrossthebottomisthecurrentprobetemperature.
Acrossthemiddleisatextentryboxandaprogressbar.
Unlikeontheinstrument,theunitsherecannotbechangedandaresetto:
Conductivity -%IACS Temperature -degreesCelsius Liftoff-mm5.1.1.2 RemoteControl
AssoonastheSIGMACHECKisconnectedandtheUniqueIDdisplayed,itcanbe
controlledfromSigmaCheckPC.The7buttonsvisibleonthePCScreen(Up,Down,
Left,Right,OK,Power,Back),whenclickedhavethesameeffectontheinstrument
aspressingthephysicalbuttonsbyhand.PleasenotethatiftheSIGMACHECKisbeing
operatedcompletelyremotelythenabuttonmustbepressedaftertheUniqueIDis
showntoremovetheSIGMACHECKsplashScreen.Untilthisbuttonispressednoreal-
timedataisreceived.
ThePowerbuttonwillturntheremoteSIGMACHECKOFFbutwillnotturnONa
SIGMACHECK.
5.1.1.3 Uploading&Downloadingfilesto/fromaSIGMACHECK
Thereare2waysofgettingfilesonandofftheSIGMACHECKsMicroSDcard:
1. Removethecardfromtheinstrument(aslotispresentundertheCalibrationBlocks),placeintoaPCcardreaderandcopythefilesusingstandard
Windows/Maccapabilities.
2. ConnectingtheSIGMACHECKtoaPCusingthesuppliedUSBcablethenusingtheSigmaCheckPCfilehandlingfunctionalityviatheMicroSDCardbutton
(seebelow).
7/30/2019 SigmaCheck User Manual_02
20/26
20
5.1.1.3.1 SigmaCheckPCFilehandling
ConnecttheSigmaChecktoaPCusingtheUSBcable.WithSigmaCheckPCrunning
andconnectedtotheinstrument,clickontheMicroSDCardbuttonacrossthe
bottom.Awindowwillappearlikeso:
ThetopofthefiletreeshowsFileSystem.EverythingbelowthisiseitherafileorfolderthatispresentontheSIGMACHECKsMicroSDCard.Foldersthathaveany
contentsareshownwithasmallboxtotheleftoftheirfoldericon.Clickingonthis
boxtogglesitbetweenexpandedornotexpanded.
Ifafileordirectoryisselected,the3buttonsacrossthebottomareenabledandcan
beclicked.Theirfunctionalityisasfollows:
DeleteFileThehighlightedfilewillbedeleted.Directoriescannotbedeleted.
GetFileThehighlightedfilewillbedownloadedfromtheSIGMACHECK.Aprogress
barwillbeshownacrossthemainscreenofSigmaCheckPCtoshowtheprogressof
thisfiledownload.Oncecomplete,aSaveFilewindowwillappearandthefilecanbe
savedanywhereonthePC.SendFileAnOpenFilewindowwillappearandafilecanbeselectedfromthePC.
ClickingOpenwillsendthisfiletothedirectoryhighlighted(orifafileishighlighted,
toitsdirectory)ontheSIGMACHECK.TheprogressbarontheSigmaCheckPCmain
screenwillshowprogressofthisfiletransfer.
AfterDeletingorsendingafile,thescreenwillberefreshedshowingthenewstate
ofthefilesystemontheSIGMACHECKsMicroSDCard.
7/30/2019 SigmaCheck User Manual_02
21/26
21
Directoriescannotbecreatedordeleted.Thedirectorystructureisusedbythe
SIGMACHECKlikeso:
ProbesAllprobefilesmustresideinhere.
SettingsAllfilesrelatedtotheconfigurationoftheinstrumentmustresidehere.
DataLogAnyfilescreatedbytheDataLoggingfunctionality(assumingitis
configuredtowriteitsdatatothecardandnottoRAM)arewrittenhere.The
filenameiscreatedfromtheTime&Datethatthefilewascreatedinthefollowing
format:
YYYYMMDDHHMMSS.xml,
ie20120119103502=10:35am(and2seconds)onthe19thJan,2012.
Ifdownloadinganewprobefiletouse,pleaserememberthatitmustbeselected
fromtheProbesmenuontheinstrumentafterthedownload!
IfanyDataLoggingtotheMicroSdcardwasinprogresswhenclickingthe
SigmaCheckPCbuttonMicroSDCard,itwillbeterminated.Thedatawillnotbelost
andthefilewillbeclosedsafely.
5.1.1.4 DownloadingDataLoggerDatafromRAM.
DataLoggingcanbeconfiguredtoeithersavetoRAM(internalSIGMACHECKmemory)
ortooutputtotheMicroSDCard.
Thebenefitsofthetwoapproachesaredescribedinsection3.7.4.3.
ClickingtheRetrieveDatabuttonalongthebottomoftheSigmaCheckPCscreen
willbegintheuploadprocess.AslongastheSIGMACHECKisconnectedto
SigmaCheckPCthisprocesscanbeinitiated.EvenifdataLogginghasnotbeenstarted,datarelatingtothenumberofColumnsandRowswillbeuploaded.IfData
Loggingwasnotcompleted,theremainingreadingswillbeinanunknownstateand
willnotcontainvaliddata.
AftertheRetrieveDatabuttonispressed,datawillbeginuploading.Theprogress
bardoesnotfunctionforfileupload.Thetimetheuploadtakesisdependentonthe
numberofRowsandColumnsthatDataLoggingissettoo.WhenUploadiscomplete
afilesavewindowwillopen.SelectafilenameandlocationwheretosavetheXML
file.Forviewingthefile,see5.2UsingaSpreadsheettoanalysedatabelow.
5.1.2 ETherFUU(FirmwareUpdateUtility)
TheSIGMACHECKinstrumentfirmwarecanbeupdatedbythecustomerinthefield
allowingthemtotakeadvantageofnewfeaturesandimprovementsastheyare
releasedfromETHERNDE.TheprocessissimpleandrequirestheETherFUUsoftware
tobeinstalledonaPC,theMini-BUSBcableandofcoursetheSIGMACHECK
instrument.
AnupdatefilewillbeprovidedbyETHERNDE,thisfilewillbeaDFUfile(i.e.
v_1_1_0.dfu).RuntheETherFUUsoftware.Thescreenwillbeasso:
7/30/2019 SigmaCheck User Manual_02
22/26
22
First,theSIGMACHECKmustbeputintoFirmwareLoadermode.Todothis:
1. TurntheinstrumentOFF.2. HolddowntheDownkey.3. PressthePowerkey.4. WhentheSplashScreenisvisible,releasetheDownkey.5. TheSplashScreenwillshowEtherNDE,FirmwareLoader
WhentheSIGMACHECKisinFirmwareLoadermode,ETherFUUrunningonthePCand
theSIGMACHECKisconnectedtothePC,theSplashScreenwillshowUsbConnected
andETherFUUwillshow:
ConnectedtoEtherFirmwareLoader,butNOfileloaded.
Now,clicktheChoosebuttononETherFUUandusetheOpenfilewindowto
navigateandselecttheprovided*.dfufile.ETherFUUwillnowshow:
Filecorrectlyloadedandconnected
TheUpgradebuttonwillnolongerbegreyedoutandshouldbeclicked.The
softwarewillbedownloadedandprogresswillbeshownbytheprogressbar.
Whencompleted,theSIGMACHECKmustberestarted.CheckontheSplashscreenof
theSIGMACHECKthatthenewsoftwareversionisdisplayed.
7/30/2019 SigmaCheck User Manual_02
23/26
23
5.2 Usingaspreadsheettoanalysedata
DataLoggerdatathathasbeendownloadedviaSigmaCheckPCwillhavebeensaved
ontothePCina*.xmlfile(See5.1.1.3Uploading&Downloadingfiles&5.1.1.4
DownloadingDataLoggingDatafromRAM,above).
ThisfileformatisnotautomaticallyrecognisedbyWindowsasbeingaspreadsheet
file,sodonotopenthefilefromitsicon.Openaspreadsheetpackage,clickOpen
File,navigatetothefileandopenit.Thespreadsheetpackageshouldrecognisethe
fileformatandopenthefileasifitwasaregularSpreadsheetfile.Hereisan
examplefileopenedupinMSExcel.TheDataLoggerontheSIGMACHECKwas
configuredtohave2rowsand3columns:
Thevaluesshownare%IACSvalues.
Whenclosingthespreadsheetpackagewerecommendsavingthefilenotinthe
original*.xmlformat,butintheSpreadsheetpackagesdefaultformat.
7/30/2019 SigmaCheck User Manual_02
24/26
24
6 Setting-upprocedures6.1 Calibratingandmakingareading
TheSIGMACHECKshouldbecalibratedafterpoweronandatregularintervalsto
ensureaccuracy.Whatfollowsisaquickstep-by-stepguidetoconfiguringthe
SIGMACHECKandtakingareading.
SplashScreen
ReadingsScreen
MainMenuScreen
CalibrationScreen
PressPowerKeyfor2
seconds.Splashscreen
appears.Afteracouple
ofsecondsthereading
screenisshown.
Awarnignacrossthe
bottomreadsPress
toCalasnocalsince
power-on.
PressOKtoCal(goto
step5)ORpressBACK
togotothemenu
screen
RightorDownkeyto
accesstheMainMenu.
Tocalibrate,with
Calibration
highlighted,pressOK.
Setairpoint
SethighConductivity SetLowConductivity.
ReadingScreen
EnsurethattheCB1&
CB2valuesmatchthe
blocksontopofthe
instrument,thenwith
Calibratehighlighted,pressOK.
Holdprobeinairat
least100mm(4)away
fromanymetalor
magnets.WaitforBlue
bartostabilise.PressOK.
Placeprobeonleft
calibrationsamplewait
forbluebartostabilise
andpressOK
Placeprobeonright
calibrationsample
weightforbluebarto
stabiliseandpressOK.
Afterabriefpausetheinstrumentwillreturn
totheReadingScreen
andisreadytouse.
6.2 Usingthe2ddataloggingfunction
ThecompleteDataLoggingprocessisbrokenupintoseveralparts,eachisexplained
inprevioussections:
1. ConfiguringtheSIGMACHECKtoDataLog,seeDataLogging.2. DownloadingtheDataLoggerdatatoaPC,seeDownloadingDataLoggerData
7/30/2019 SigmaCheck User Manual_02
25/26
25
3. Viewingtheresultantfiles,seeUsingaspreadsheettoanalysedata.
6.3 ALARMsetup
TheAlarmfunctionalityisfullyexplainedintheAlarmssection.
7 CareandMaintenance
Usesoftcleanclothmoistenedwithasoapormilddetergenttocleanthecase,buttonsanddisplaywindow.
Donotuseanabrasivecleaningagentofanykind,asthiswilldamagetheinstrument.
Donotusesolventstocleantheinstrument,astheywillcauseirreparabledamagetocaseandcontents.
8 TroubleShooting9 TransportationandStorage
7/30/2019 SigmaCheck User Manual_02
26/26
10Specifications
InspectionTechnology EddyCurrent
OperatingFrequencies
60kHz,120kHz,240kHz,480kHz,960kHz(requiresprobe
suitablefortestfrequency)Display 240x320pixelscolourdisplay.LCDwithselectablebacklight.
ConductivityStandards
DualSampleConductivityReferenceBlocksmountedontop
ofunit.Removableforvalueverification,andensuring
thermalequilibrium.
ConductivityRange 0.5%IACSto120%IACS,0.28-70MS/m
Resolution
Selectableas1,2or3decimalplacesuptoamaximumof5
digitse.g.0.050-120.00
LiftOff
12.7mmprobecompensatedto0.030(0.7mm)7mmprobe
compensatedto0.015(0.35mm)
AbsoluteAccuracy
At20C.+/-0.5%ofreadingpluserrorinreference
standard.Probeinthermalequilibriumwithmetal.
Temperature
Measurement In-probesensor(accurateto0.5C)Range0Cto+50C
AutomaticCompensation Conductivitymeasurementsarecorrectedtothe20Cvalue.
EnvironmentalRange
0to95%relativehumidity,0Cto+50Cforreliable
operation
DataLoggerMemory
Removable2GBmicroSDCardallowingmanythousandsof
probemapstobestoredandeasilytransferredbetween
instruments.
Probes
12.7mmdiameterfor60kHz-480kHz.Probesare
interchangeablewithsimpleoperatorresettingprocedure.
Probesarefieldexchangeableanddonotrequirethereturn
tomanufacturerforcalibration.
PCConnectivity USBportforchargerandPCcommunications
Accessories
SettingsReferenceBlocks-Arangeofconductivityreferences
standardstraceabletoUSandEuropeanstandardsare
availableforin-fielduse.Uptofivecanbemountedonan
aluminiumanodisedholdingplate.
Size 163mmLongx80mmWidex25mmDeep
Construction&Storage
Highimpact,splashproof,mouldedUL94-5VAflame
retardantABScase.Protectiverubberboottoholdunit,
probes,probecable,operatormanualonUSB,and
removablestand.
Weight 350g(12oz.)includingbatteries
Power
Providedwith2x1.5VAANickelMetalHydrideBatteries,
Approx.upto6hrs.life.