Upload
doantuyen
View
219
Download
2
Embed Size (px)
Citation preview
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 1 of 14
Table of Contents
Page
1. Scope 2
2. Objective 2
3. Definitions 2
4. Responsibility 3
5. Procedure 3
6. Precautions 13
7. References 13
8. Appendices 13
9. Revision History 14
Original
Authorised Copy
for:
Reading Copy
Written by: Signature/Date:
Reviewed by: Signature/Date:
Approved by: Signature/Date:
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 2 of 14
1. Scope
This Standard Operating Procedure (SOP) applies to the staff and students
using the Hanna® Bench-Top pH Meter HI 8521 in the laboratories of the
Pharmacy Department, University of Malta.
2. Objective
To describe the procedure for installation, operation, calibration, maintenance
and troubleshooting of the Hanna® Bench-Top pH Meter HI 8521.
3. Definitions
3.1. ↑ oC and ↓
oC Keys: To set the temperature for Manual Temperature
Compensation.
3.2. Automatic Temperature Compensation (ATC): A feature in which
the temperature probe is able to detect the temperature of the solution
in which it is immersed in.
3.3. Bayonet Neill-Concelman (BNC) Electrode connector (1): A cable
connector used to connect the pH or an ORP electrode to the meter.
3.4. CAL Key: To enter the calibration mode.
3.5. CON Key: To confirm the calibration data.
3.6. Display: Digitally displays the pH, temperature and millivolt (mV)
results.
3.7. DC Power Socket (5): To connect the meter with the mains power
supply.
3.8. Manual Temperature Compensation (MTC): A feature in which the
temperature of the sample solution is manually inputted when this is
already known.
3.9. On/Off Key: To turn the meter on or off.
3.10. RANGE Key: To select between pH, millivolt or temperature results.
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 3 of 14
3.11. Reference Jack Socket (2): To connect a reference electrode to the
meter.
3.12. Reset Button (3): To be pressed whenever erroneous messages are
displayed due to strong electrical interference or when the meter is
disconnected from its power supply before being switched off properly.
3.13. Temperature Probe Socket (4): The socket at which the temperature
probe is connected to the meter.
4. Responsibility
4.1. The members of the Department of Pharmacy (staff and students) are
responsible for following this SOP.
4.2. The designated Laboratory Officer or Laboratory Assistant is
responsible for ensuring that this SOP is followed.
5. Procedure
5.1. Diagram of Front Panel
5.2. Diagram of Rear Panel
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 4 of 14
5.3. Installation
5.3.1. Connect the power cable to the DC Power Socket and to the
mains supply.
5.3.2. Connect the pH electrode and the temperature probe to the
BNC and temperature sockets respectively.
5.4. Operation
5.4.1. Measuring pH
5.4.1.1. Before use, ensure that the meter is being calibrated,
to obtain more accurate measurements (see section
5.5).
5.4.1.2. Press the On/Off Key to switch on the meter, which
will automatically enter into pH mode.
5.4.1.3. Submerge the pH electrode and the temperature
probe into the sample to be tested and keep them
close together, if automatic temperature
compensation is desired.
5.4.1.4. Disconnect the temperature probe from the meter if
manual temperature compensation is desired
whenever the temperature of the sample to be tested
is already known.
5.4.1.4.1. Press the RANGE Key until the display
changes to temperature mode.
5.4.1.4.2. Adjust the displayed temperature by
using the Up and Down Keys
respectively.
5.4.1.4.3. Press the RANGE Key again until the
display changes back to pH mode.
5.4.1.4.4. Submerge pH electrode into the sample
to be tested.
5.4.1.5. Shake the sample briefly and wait for reading to
stabilise before taking the reading.
5.4.1.6. Read off the displayed stable pH reading
compensated for temperature.
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 5 of 14
5.4.1.7. Rinse off the pH electrode with some solution from
the next sample if more than one sample is to be
successively tested.
5.4.2. Measuring Temperature
5.4.2.1. Press the On/Off Key to switch on the meter.
5.4.2.2. Press the RANGE Key until the display changes to
temperature mode.
5.4.2.3. Dip the temperature probe into the sample.
5.4.2.4. Allow 1-2 minutes for the reading to stabilise.
5.4.2.5. Read off the displayed stable temperature reading.
5.5. Calibration
5.5.1. Pour small quantities of pH 7.01 and pH 4.01 or pH 10.01
buffer solutions into two separate clean beakers and use another
two separate beakers for rinsing.
5.5.2. Press the On/Off Key to switch on the meter.
5.5.3. Rinse the pH electrode with some pH 7.01 buffer solution.
5.5.4. Immerse the pH electrode into the pH 7.01 buffer solution and
stir gently.
5.5.5. Ensure that the pH electrode is kept submerged approximately
4cm into the solution.
5.5.6. Ensure that the temperature probe is located as close to the pH
electrode as possible.
5.5.7. Press the CAL Key.
5.5.8. Wait until the pH symbol, next to the pH buffer value, stops
flashing to indicate that reading is stable.
5.5.9. Press the CON Key to confirm the calibration.
5.5.10. If [E 4] is flashing, ensure that the correct buffer solution is
being used.
5.5.11. Wait until [E 5] is displayed to indicate that the electrode is still
in the pH 7.01 buffer, but expecting a pH 4.01 or 10.01 buffer.
5.5.12. Rinse the pH electrode with distilled water followed by the
second buffer solution to be used.
5.5.13. Immerse the pH electrode into the second buffer solution (pH
4.01 or pH 10.01) and stir gently.
5.5.14. Wait until the pH symbol, next to the pH buffer value, stops
flashing to indicate that the reading is stable.
5.5.15. Press the CON Key to confirm the calibration.
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 6 of 14
5.5.16. If [E 4] is flashing, ensure that the correct buffer solution is
being used.
5.5.17. Wait until the meter returns to the normal operating mode to
ensure that the calibration process was successful.
5.6. Maintenance
5.6.1. Inspect the cables of the electrodes for any signs of broken
insulation.
5.6.2. Inspect the pH electrode for any cracks in its stem or bulb.
5.6.3. Rinse off any salt desposits with water.
5.6.4. Soak the pH electrode in Hanna HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour to perform general
cleaning.
5.6.5. Soak the pH electrode in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes if protein deposits are present
on the membrane and/or junction of the electrode.
5.6.6. Soak the pH electrode in Hanna HI 7074 or HI 8074 Inorganic
Cleaning Solution for 15 minutes if inorganic deposits are
present on the membrane and/or junction of the electrode.
5.6.7. Rinse the pH electrode with Hanna HI 7077 or HI 8077 Oil and
Fat Cleaning Solution if oil films or deposits are present on the
membrane and/or junction of the electrode.
5.6.8. Rinse the pH electrode thoroughly with distilled water after
performing any of these cleaning procedures.
5.6.9. Refill the reference chamber with fresh electrolyte and soak it
in HI 70300 or HI 80300 Storage Solution for at least 1 hour
before taking any measurements.
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 7 of 14
5.7. Trouble Shooting
Fault Possible Cause/s Solution
Meter fails to calibrate or
gives erroneous results pH electrode is defective
Replace the electrode and repeat the
calibration
Meter is slow in responding
or gives erroneous readouts
Electrode not working or
diaphragm is clogged
- Leave electrode in storage
solution after performing
diaphragm cleaning
- If problem is not solved,
replace electrode and
recalibrate pH scale
Meter does not accept 2nd
buffer solution
pH electrode defective or
bad buffer solution
- Clean the electrode
- Replace electrode and re-
calibrate
- Use fresh buffer solution
Reading is drifting
Temperature not stable
Wait longer since thermal
equilibrium is still not reached
Display shows [E 1] pH scale is out of range
- Ensure pH electrode is
connected
- Ensure pH of sample is
between 0-14
- Check the state of the electrode
and the level of its filling
electrolyte
Display shows [E 2] Temperature scale is out
of range
- Ensure T probe is properly
connected
- Ensure that T of sample lies
between -10oC and 100
oC
Display shows [E 3] mV scale is out of range
- Ensure that ORP probe is
properly connected
- Ensure mV is within +/- 1999
range
Display shows [E 4]
- Erroneous buffer
solution used for first
part of calibration
- Defective electrode
- Ensure that buffer solution used
is of pH 7 and replace if
necessary
- Replace electrode and re-
calibrate pH scale
Display shows [E 5]
- Erroneous buffer
solution used for
second part of
calibration
- Defective electrode
- Ensure that correct pH buffer
solution is being used (pH 4.01
or pH 10.01)
- Check electrode and replace if
necessary
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 8 of 14
5.8. Average time taken to carry out Procedure
Procedure Section Average Time (mins)
Installation N/A
Measuring pH with Automatic
Temperature Compensation
1 (for one pH reading)
Measuring pH with Manual
Temperature Compensation
2 (for one pH reading)
Measuring Temperature 2 (for one T reading)
Calibration 3 ( for two-point calibration)
Maintenance N/A
Troubleshooting N/A
5.9. Flow Charts
5.9.1. Installation
Start
Connect power cable to DC power
socket and to mains supply
Connect pH electrode and T probe to
BNC and T sockets respectively
End
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 9 of 14
5.9.2. Operation
before
Start
Measuring pH
Meter being
calibrated
Perform calibration before
use (see section 5.8.3)
Press On/Off Key
to switch on meter
Use ATC feature MTC
Submerge pH electrode and T probe into sample and keep
them close together
Press RANGE Key until display
changes to T mode
Disconnect T probe from meter
Adjust displayed T by using the
Up and Down keys respectively
Press RANGE Key again until
display changes back to pH mode
Submerge pH electrode into
sample to be tested
Shake sample briefly and wait
for reading to stabilise
Read off stable pH reading
Test another
sample
Rinse off electrode with next
sample before further use
1
Measuring T
Press On/Off key to
switch on meter
Dip T probe
Allow 1-2 minutes for
reading to stabilise
Read off displayed
stable T reading
End
2
1
Yes
No Yes
Yes Yes
Yes
Yes
No
No
No
No
No
Display in
T mode
Press RANGE Key
until display changes to T mode
Yes
No
2
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 10 of 14
5.9.3. Calibration
Start
Pour small quantities of pH 7.01 and pH 4.01 or pH 10.01 buffer
solutions into 2 separate beakers
Use another 2 separate
beakers for rinsing
Press On/Off Key to
switch on meter
Immerse pH electrode into
pH 7.01 buffer and stir gently
Submerged 4cm
into solution
Meter in pH
mode
Press CAL Key
[pH] symbol
stopped flashing
Press CON Key
[E 4] seen
flashing
1
Submerge accordingly
Place T probe as close to pH
electrode as possible
Press RANGE Key until display
changes to pH mode
Wait until it stops flashing
Using correct
buffer solution
Replace electrode
Use appropriate
buffer solution
1
Wait until [E 5] is displayed
2
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
Rinse pH electrode with some
pH 7.01 buffer solution
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 11 of 14
2
Rinse pH electrode
with distilled water
Rinse pH electrode with second buffer solution
Immerse pH electrode into
second buffer and stir gently
[E 4] seen
flashing [pH] symbol
stopped flashing
Press CON Key
Wait until meter returns to
normal operating mode
before further use
End
Wait until it stops flashing
Using correct
buffer solution Use appropriate
buffer solution
Replace electrode
1
1
No No
No
Yes
Yes
Yes
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 12 of 14
5.9.4. Maintenance
Start
Inspect cables of electrodes for any
signs of broken insulation
Inspect pH electrode for any
cracks in stem or bulb
Rinse off any salt deposits with water
General cleaning
Soak pH electrode in Hanna HI 7062
or HI 8061 General Cleaning Solution for ½ hour
Protein deposits
present
Soak pH electrode in Hanna HI 7073
or HI 8073 Protein Cleaning Solution for 15 minutes
Inorganic deposits present
Soak pH electrode in Hanna HI 7074
or HI 8074 Inorganic Cleaning
Solution for 15 minutes
Oil films / deposits present
Soak pH electrode in Hanna HI 7074
or HI 8074 Inorganic Cleaning
Solution for 15 minutes
Rinse pH electrode thoroughly
with distilled water
Refill reference chamber
with fresh electrolyte
Soak in HI 70300 or HI 80300 Storage
Solution for at least 1 hr before use
1 1 1
1
2
2
No No No
No
Yes Yes
Yes
Yes
End
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 13 of 14
6. Precautions
6.1. Always keep the pH electrode wet and rinse it thoroughly with the
sample to be measured before use.
6.2. In order to measure the pH accurately, the temperature factor must
always be taken into consideration.
6.3. If the sample temperature is quite different from the temperature at
which the pH electrode is kept, allow a few minutes for thermal
equilibrium to take place.
6.4. When performing a two-point calibration, use pH 7.01 and pH 4.01
when measuring acidic samples and pH 7.01 and pH 10.01 for alkaline
measurements. This provides more accurate readings to be taken.
6.5. Do not be alarmed when salt deposits are present on the electrode since
these normally form and can easily be rinsed off with water.
6.6. If any air bubbles form inside the glass bulb, “shake it down” firmly to
remove them.
6.7. Always store the electrode in its appropriate storage solution and never
store it in distilled or deionised water.
6.8. Avoid high humidity environments and static discharges since these are
detrimental for a stable pH reading.
6.9. Perform a calibration:
6.9.1. When pH electrode or temperature probe is replaced
6.9.2. When reference electrode is changed
6.9.3. At least once a month
6.9.4. After testing aggressive chemicals
6.9.5. When greater accuracy is required
6.9.6. After any cleaning procedure
7. References
Hanna® Instruments. Instruction Manual HI 8417 – HI 8519 – HI 8520 – HI
8521 Microprocessor Bench-Top pH Meters;1998.
8. Appendices
N/A
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT Ref. No. SOP/PD/312_01
STANDARD OPERATING PROCEDURE
HANNA® BENCH-TOP PH METER HI 8521
Valid for:
2 years from
approval
Page 14 of 14
9. Revision History
Version Number Amendments/ Reasons for change
01 Initial Release