S Static Compliance (C S ) Bobbi Jimenez Sue Trapp

Static compliance

Embed Size (px)

DESCRIPTION

lung compliance

Citation preview

Page 1: Static compliance

S

Static Compliance (CS)Bobbi Jimenez

Sue Trapp

Page 2: Static compliance

CS (ml/cmH2O)

Compliance- The ability of the alveoli to stretch and accept a volume of air per a given pressure gradient. Air flows from high to low pressures into the lungs.

Static compliance is determined during a time of no airflow by finding the plateau pressure.

During an I hold/breath hold, Plateau pressure can be measured. Plateau pressure is the pressure in the alveoli that disperses

through the Pores of Kohn, canals of lambert, and channels of martin when a breath is held.

Page 3: Static compliance

Compliance vs Elastance

• Compliance in its most general form:

• CL= ΔV

ΔP

• Elastance in its most general form:• E= ΔP

ΔV

• Elastance opposes compliance; Elastance is the ability to recoil or return to its original position.

Page 4: Static compliance

Equation

Cs= VtPlateau- PEEP

Page 5: Static compliance

Compliance’s Role in Ventilation

When the diaphragm contracts, a negative intrapleural pressure of approximately -2cmH2O is generated.

This establishes the pressure gradient for air to flow into the lungs, since the intrapleural pressure is negative compared to atmospheric pressure. Air is a mixture of gases, so it will flow from an area of high to an area of low pressure.

The lungs can then accept a volume of air.Normal Values: 60-100 ml/cmH2O ( on vent) 0.1-0.2 L/cmH2O

Page 6: Static compliance

More Normal Values

Pediatric patients 40 to 50ml/cmH20

Neonates 10 to 20 mL/cmH20

Static compliance with ventilator circuit: Tubing compliance varies depending on the type of diameter of tubing used. Typically tubing compensation (Ct) 2 to 3 mL/cmH2O

Cs=Vt-[(plateau-PEEP) x Ct]/(Plateau – PEEP)

Page 7: Static compliance

Compliance Changes

Decreased Cs Seen in Restrictive Processes:

PNA Pneumothorax Loss of surfactant Pulmonary Edema Atelectasis Pleural Effusions Neuromuscular Diseases

Page 8: Static compliance

Decreased Cs

Why is Cs decreased with Restrictive Processes? Restrictive process= Volume problem Volume DECREASES Plateau Pressure can INCREASE on vent because the

same amount of delivered Vt is trying to disperse into a smaller volume area.

Page 9: Static compliance

RESTRICTIVE PROCESSES LEADING TO DECREASED VT AND/OR INCREASED PLATEAU PRESSURES

Page 10: Static compliance

Cs= VtPlateau- PEEP

We can see that: 1) Plateau=Cs2) Vt= Cs

DECREASED STATIC COMPLIANCE

Page 11: Static compliance

Compliance Changes

Increased Cs Emphysema

Increased Air-trapping causes increased measured PEEP and greater dispersion of air in alveoli due to loss of elasticity, which causes a drop in plateau pressure.

Page 12: Static compliance

Cs= VtPlateau- PEEP

We can see that: 1) Vt=Cs2) Plateau=Cs3) Auto-PEEP=Cs

INCREASED STATIC COMPLIANCE

Page 13: Static compliance

Waveforms

Plateau pressure decreases from Peak pressure as gas dispersion occurs in alveoli

Page 14: Static compliance

Waveforms

LOW COMPLIANCE= LOW VTHIGH COMPLIANCE= HIGH VT

Page 15: Static compliance

SLOW RISE TO PEAK PRESSURE AND LOW PLATEAU PRESSURE

QUICK RISE TO HIGH PEAK PRESSURE AND HIGH PLATEAU PRESSURE

Page 16: Static compliance
Page 17: Static compliance

AIR-TRAPPING CAN LEAD TO AUTO-PEEP, WHICH INCREASES MEASURED PEEP, WHICH CAN LEAD TO INCREASED Cs, SIMILAR TO WHAT HAPPENS IN EMPHYSEMIC PATIENTS

Page 18: Static compliance

VOLUME-PRESSURE LOOP

DECREASED COMPLIANCE LEADS TO HIGHER PRESSURES AND DUMPING OF VOLUME SO THAT THE HIGH PRESSURE SETTING IS NOT EXCEDED.

Page 19: Static compliance

FLOW-VOLUME LOOP

RESTRICITVE PROCESSES LEAD TO DECREASED VT, WHICH CAUSE Cs TO DECREASE.