12
FETS—Meeting 1 st August 2007 Status Report of the Non—Destructive Emittance Instrument Photo Detachment Diagnostics for H- High urrent Applications at the Front End Test Stand FETS at 3MeV Rutherford Appleton Laboratory C. Gabor, J. Pozimski, A. Letchford, C. Prior The Front End Test Stand Collaboration

Status Report of the Non—Destructive Emittance Instrument

  • Upload
    oleg

  • View
    27

  • Download
    0

Embed Size (px)

DESCRIPTION

Status Report of the Non—Destructive Emittance Instrument. Photo Detachment Diagnostics for H- High Current Applications at the Front End Test Stand FETS at 3MeV, Rutherford Appleton Laboratory. C. Gabor, J. Pozimski, A. Letchford, C. Prior. The Front End Test Stand Collaboration. - PowerPoint PPT Presentation

Citation preview

Page 1: Status Report of the   Non—Destructive Emittance Instrument

FETS—Meeting 1st August 2007

Status Report of the

Non—Destructive Emittance InstrumentPhoto Detachment Diagnostics for H- High

Current Applications at the Front End Test Stand FETS at 3MeV,Rutherford Appleton Laboratory

C. Gabor, J. Pozimski, A. Letchford, C. Prior

The Front End Test Stand Collaboration

Page 2: Status Report of the   Non—Destructive Emittance Instrument

2

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Overview

• IntroductionPhoto detachment, slit—point transformation, movable detector

• Maximum Entropy as a Method for Emittance Reconstruction

• Experimental Results, measured at the Ion Source Development Rig (ISDR)

• SummaryOutlook, further development

Page 3: Status Report of the   Non—Destructive Emittance Instrument

3

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Photo—Detachment Principle

H -beam ,N eutra l a tom s produced in the tra nspo rt section

-

La se r, 1060 n m

H -beam-

N e utra l a to m s p ro d uc ed in th e tra ns po rt se c tio n

M agnet po letips

B X

xz

Region of interactionand separation of thebeam s

sca

n

S cin tilla to rin teraction w ith photons

y

+ Non—destructive+ Preferable for high power + On line measurements

Implemented Method: yy' in combination with movable angle detector

o Displacement & Spread of neutralized particles Angle profile I(y')o 2D Beam distributions along the neutrals drift

Slit—PointTransformation

Page 4: Status Report of the   Non—Destructive Emittance Instrument

4

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Problem of Data ReconstructionVery often incomplete and/ or noisy data where MaxEnt has its

strengthObject of Study Data

e.g.Density of electrons

Spectral intensityAn image: 2D, 3D

Measuring Instrument

T . {fi} = {dk + k} Traditional solution: {fi} = T-1 . {dk + k}

e.g.Line-integral

Fourier transformBlurred imageBeam profiles

e.g.Telescope,

spectrometer,emittance

scanner, .....

{ fi } i =1,…, Nobject

{dk} k=1,…, Ndata

Maximum Entropy: (MaxEnt) Data like beam profiles linked with a matrix to your test object (constraints) will be fitted with maximum—likelihood Entropy S is "an additional" search direction to fit data (max. this func. S !)

Page 5: Status Report of the   Non—Destructive Emittance Instrument

5

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Entrance Distribution and Beam Envelope

Page 6: Status Report of the   Non—Destructive Emittance Instrument

6

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Variable Number of Profiles using MaxEntA small number of profiles is sufficient for reconstruction.

(A) 3 profiles (B) 4 profiles(C) 8 profiles(D)Entrance emittance distribution

Problem:Not the number but the coverage of the phase space is crucial. That can be, e.g., provided by an optical focusing element.

http:// ~www.maxent.co.uk

Page 7: Status Report of the   Non—Destructive Emittance Instrument

7

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Comparison of rms--EmittancesImplemented Method: yy' in combination with movable angle

detector

Page 8: Status Report of the   Non—Destructive Emittance Instrument

8

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Ion Source Development Rig (ISDR)

Pepperpot emittances +

Beam profiles at various z—positions

Slit slit emittances

Seperate, movable slit in combination with scintillator

Beam parameters:H- @ 50mA17 keV extraction voltage35 keV total beam energyfield gradient n=1.4 (w)

Slit height: 0.3mmSlit length: 110mmDrift in between: 107mm

Page 9: Status Report of the   Non—Destructive Emittance Instrument

9

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Collimated Ion Beam (Pseudo Coloured)1.) Beam distribution at 260mm2.) Collimated beam parts after a separation drift of 107mm

Page 10: Status Report of the   Non—Destructive Emittance Instrument

10

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Pepperpot measurement "simulation"

-70

-50

-30

10

30

50

-70 -50 -30 -10 10 30 50

col imated beam, y=-30 mm colimated + drifted beam

colimated beam, y=+26m m colimated + drifted beam

drift=0.5m

- 6 0 - 4 0 - 2 0 0 2 0 4 0 6 0

0 . 0

0 . 2

0 . 4

0 . 6

0 . 8

1 . 0

1 . 2

P e p p e r p o t , y = + 2 6 m m

S c i n t i l l a t o r p r o f i l e

P e p p e r p o t , y = - 3 0 m m

S c i n t i l l a t o r p r o f i l e

In

te

ns

ity

/a

.u

.

A n g l e / m r a d

At -30mm and =26mm original collimated beamletsFor better separation of both distributions a drift of 0.5m was chosen.

Hardware limited resolution because of "low" number of pinholes which also effects the signal to noise ratio.Further problems with a restricted number of particles ( statistical restrictions), small variations of ion source performance

Page 11: Status Report of the   Non—Destructive Emittance Instrument

11

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

Summary• Photo detachment is the method for non destructive diagnostics • Highly perferable for intense ion beams (no heat loading on mechanical parts ...)• Presented emittance measurement method(s) are general

principles and not restricted to photo neutralization • Additional information about beam distribution are possible: Demonstrated at ISDR using a slit for beam collimation

and a scintillator with spatial resolution for trans. angle

• Outlook: Reconciling with the MEBT developing of diagnostic magnet design ( determinate resolution!)

• Concerning HIPPI: This contribution replaces RALs 3D code contribution (staff problems);

providing a deliverable as status report by end 2008

Page 12: Status Report of the   Non—Destructive Emittance Instrument

12

Christoph Gabor, ASTeCHIPPI Annual Meeting, CERN, 30.September 2008

References and Literature

G.N.Minerbo, "MENT: A maximum entropy algorithm for reconstructing a source from projection data", Computer Graphics and Image Processing 10 (1979), p. 48—68

O.R.Sander, G.N.Minerbo, R.A.Jameson, D.D.Chamberlin, "Beam tomography in two and four dimensions", 1979 LINAC--Conference Montauk, N.Y.

U.Rohrer, W.Rohrer, "Introduction of 2—dimensional beam tomography for monitoring transverse beam emittance at SIN", PSI Ann. Rep. 1982, NL 5—6

C.T.Mottershead (LANL, AT-6) "Maximum entropy beam diagnostic tomography", IEEE Trans. on Nucl. Sc., Vol. NS—32, No.5, 1985

D.S.Siva, "Data Analysis A Bayesian Tutorial", Oxford Science Publications (Clarendon Press Oxford), 1996, 2006

http://pc532.psi.chWeb page of Uli Rohrer, partly Co—author of MENTX

http:www.maxent.co.uk