Sterilbarrieresysteme Hart- vs. Weichverpackungen Christiaan Meijer Vortrag 43. Veranstaltung des Arbeitskreises Infektionsprophylaxe am 20.September 2016 in Potsdam und 21. September 2016 in Leipzig

Sterilbarrieresysteme Hart- vs. Weichverpackungen · Kammer einem mit Bacillussubtilis kontaminierten Aerosol ausgesetzt ... und 1000 hPA, kalte Hochdruckgebiete zwischen 1040 und

  • Upload
    others

  • View
    18

  • Download
    0

Embed Size (px)

Citation preview

  • SterilbarrieresystemeHart- vs.Weichverpackungen

    ChristiaanMeijerVortrag43.VeranstaltungdesArbeitskreisesInfektionsprophylaxe

    am20.September2016inPotsdamund21.September2016inLeipzig

  • Folie2

    Vortragsschwerpunkte

    • WelcheErkenntnisseliegenüberdieBarrierewirkung vonHart- undWeichverpackungenalsSterilbarrieresysteme vor?

    • WiesichernwirdieQualitätbiszurAnwendung?

    • WasistzueinemVergleichzwischenHart- undWeich-verpackungenbezüglichdesOrganisations- undArbeits-aufwandssowiederKostenzubemerken?

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Barrierewirkung vonHart-undWeichverpackungenimVergleich

  • Folie4

    EckpunktederDiskussionumdieBarrierewirkungvonHart- undWeichverpackungen

    • DieFrage„Hart- oderWeichverpackung“wirdoftengagiertdiskutiertundemotionalentschieden(„Philosophie“)

    • DieBarrierewirkung wirddabeigernbildhaftundsymbolträchtigbehandelt

    • Zuunterscheidensind• diegrundsätzlicheBarrierewirkung desSterilbarrieresystems• dieRobustheitdesSterilbarrieresystems gegenBeschädigungendurchmechanischeBeanspruchung/Handhabung• beimVerpackungsprozess• beimTransport• beiderLagerung(EinlagerungundEntnahme)

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie5

    ÜberprüfungderSterilbarrierewirkung

    • Robustheit• Empirisch(AnteilDefektebiszumZeitpunktderAnwendung)

    • Barrierewirkung• MethodenfüreineobjektiveÜberprüfungderSterilbarriere-wirkungwurden1999vondeBruijn etal.sowievonJunghannßetal.und2006vonDunkelbergetal.spezifiziert

    • NeueImpulsezurMessungderBarrierewirkung vonContainernkamenzuletztausFrankreichundderSchweiz(Wasserdichtig-keitsprüfung,2013,2014,2015)

    • Ende2015wurdenerstmalsimgrößerenUmfangvergleichendeErgebnissezurBarrierewirkung vonHart- undWeichverpackun-genuntereinerdynamischenAerosolbelastungveröffentlicht

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie6

    WesentlicherexperimentellerAnsatz:Sterilgut istLuftdruckänderungenausgesetzt

    • Temperaturbedingt(indirekt;TemperaturdifferenzzwischenSterilgut undUmgebungsorgtfürLuftaustausch)• AbkühlungnachderSterilisation• ÖrtlicheTemperaturdifferenzen(Lager,Transport)

    • Druckbedingt(direkt)• HöhenbedingteÄnderungendesatmosphärischenDruckswährenddesTransports(Aufzüge,LKW)

    • ÄnderungendesatmosphärischenDrucksimRaum(Belüftung/Klimatisierung,Öffnen/SchließenvonTüren)

    • WetterbedingteÄnderungendesatmosphärischenDruckswährendTransportundLagerung

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie7

    VergleichendeErgebnisseaus1999:niederländischer„finalpacktest“prüftFilterwirkung(Partikel)

    • Methode(RIVM,deBruin etal)• Sterilbarrieresystem wirdingeschlossenerKammereinemAerosolaus1,0-µm-Latexpartikelnausgesetzt

    • AusdemSterilbarrieresystem wirdLuftabgesaugt• DiePartikelinderabgesaugtenLuftwerdengezähltundinsVerhältniszurAusgangspartikelzahlgesetzt

    • 𝐹𝑖𝑙𝑡𝑒𝑟𝑤𝑖𝑟𝑘𝑠𝑎𝑚𝑘𝑒𝑖𝑡 = -./0123467839:;8-./01234?94:=0-./0123467839:;8

  • Folie8

    Ergebnissedesniederländischen„finalpacktests“

    0 10 20 30 40 50 60 70 80 90 100

    KonventionellesTextileinfach

    ModernesTextileinfach

    Krepppapiereinfach

    Vlieseinfach

    Vliesdoppelt(2x1)

    Vliesdoppelt(1x2)

    Container1

    Container1mitInnenverpackung

    Container2

    Container2mitInnenverpackung

    Container3

    Container3mitInnenverpackung

    Filterwirksamkeit in%vonSterilbarrieresystemen

    Filterwirksamkeitin%

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Quelle:(1)

  • Folie9

    MikrobiologischeÜberprüfungderBarriere-wirkungvonContainernin1999

    • Methode(Junghannß etal)• Sterilbarrieresystem wirdmitNährmedien(PetrischalenmitColumbia-AgarmitSchafsblutzusatz)ausgestattet

    • Sterilbarrieresystem wirdnachSterilisationingeschlossenerKammereinemmitBacillus subtilis kontaminiertenAerosolausgesetzt

    • AusdemSterilbarrieresystem wirdLuftabgesaugtumsodasAerosoldurchdasSterilbarrieresystem zuführen

    • NachderExpositionwerdendieNährmedienuntersterilenKautelenentnommen,bebrütetundausgewertet

    • 𝐾𝑒𝑖𝑚𝑟ü𝑐𝑘ℎ𝑎𝑙𝑡𝑒𝑣𝑒𝑟𝑚ö𝑔𝑒𝑛 = LMNOPQ ×100%

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie10

    ErgebnissedermikrobiologischenÜberprüfungderBarrierewirkung vonContainernin1999

    • Keimrückhaltevermögenvon16Containern• >99,99%

    • Keimrückhaltever-mögenvon5„worstcase“Containern(Ecken,Deckeldefor-miert,Deckeldichtungeingeschnitten)• >99,98%

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Quelle:(2)

  • Folie11

    MikrobiologischeÜberprüfungderBarrierewirkungvonContainernin2006

    • Methode(Dunkelbergetal)• Sterilbarrieresystem wirdmitNährmedien(PetrischalenmitSabouraud-Agar)ausgestattet

    • Sterilbarrieresystem wirdnachSterilisationmitzusätzlichen,offenenNährmedieningeschlossenerKammereinemmitSaccharomyces cerevisiae kontaminiertenAerosolmitwechselndemDruckausgesetzt

    • NachderExpositionwerdendieNährmedienderKammerunddesSterilbarrieresystems untersterilenKautelenentnommen,bebrütetundausgewertet

    • 𝐿𝑜𝑔𝑅𝑒𝑑𝑢𝑘𝑡𝑖𝑜𝑛𝑠𝑤𝑒𝑟𝑡 = 𝑙𝑜𝑔OP𝐾𝐵𝐸67839:;8 − 𝑙𝑜𝑔OP𝐾𝐵𝐸Z03/149.//13/3

  • Folie12

    ErgebnissedermikrobiologischenÜberprüfungderBarrierewirkung vonContainernin2006

    0 10 20 30 40 50 60 70 80 90

    DoppellagigeVliesverpackung

    NeueContainermitPermananetplastikfiltern

    GebrauchteContainermitPermanentplastikfiltern

    GebrauchteContainermitTextilfiltern

    GebrauchteContainermitPapierfilter

    AnzahlSterilbarrieresysteme mitundohneRe-Kontamnination

    Re-Kontamination keineRe-Kontamination

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Quelle:(3)

  • Folie13

    AktuellsteStudiezurLeistungvonHart- undWeichverpackungenalsSterilbarrieresysteme

    • „Sterility maintenance study:Dynamicevaluation of sterilizedrigidcontainers and wrapped instrument trays to preventbacterial ingress“• „StudiezurAufrechterhaltungderSterilität:DynamischeEvaluierungvonsterilisiertenfestenContainernundeingewickeltenInstrumentensiebenzurVermeidungdesEindringensvonBakterien“

    • StudiedurchH.L.Shaffer,AppliedResearchAssociates,Inc.(„ARA“,USA),etal.(H.L.Shaffer etal.,AmericanJournalofInfection Control43[2015]1336-41)

    • Veröffentlicht2015imAmericanJournalof Infection Control• StudiewurdemitfinanziellerUnterstützungdurchHalyardHealthcaredurchgeführt

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie14

    DynamischemikrobiologischeBelastungdurchkontaminierteAerosole

    • Sterilgut wird70MinutenlangineinerAerosolkammereinem kontaminiertenAerosolbeiwechselndenDrückenausgesetzt• 2x1,0PSIDruckdifferenz-Zyklus• 3x0,7PSIDruckdifferenz-Zyklus• 3x0,4PSIDruckdifferenz-Zyklus

    • DieDruckdifferenzenentsprecheninetwademUnterschiedzwischeneinemHoch- undeinemTiefdruckgebiet(1PSI≈69mbaroderhPa;mitteleuropäischeTiefdruckgebietezwischen990und1000hPA,kalteHochdruckgebietezwischen1040und1065hPA),sindalsonichtübertrieben

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Foto:Halyard Health

  • Folie15

    Micrococcusluteus als Testkeim

    • Mluteus (ATCC4896)• Aerobes,grampositivesBakteriumausderGattungMicrococcus

    • Gewählt,dainGrößeundFormvergleichbarmitStaphylococcus aureus

    • Aerosolkonzentrationbei255± 116KBE/dm3

    • Geringste,prüftechnischmachbareKonzentration

    • AerosolkonzentrationinderPraxisbeica.2KBE/dm3

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    JaniceCarr,https://commons.wikimedia.org/w/index.php?curid=11330617

  • Folie16

    GeprüfteHart- undWeichverpackungen

    • 111ContainerunterschiedlicherHersteller• 25neueContainer• 38Container<5Jahre• 48Container5– 9Jahre

    • 161weichverpackteSiebemitunterschiedlichemVerpackungsmaterial• 2SterilisationsvliestypeninDiagonalverpackung• 1SterilisationsvliesgemäßHerstellerangabeneingesetzt

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie17

    PrüfungderKontaminationindenSterilbarrieresystemen

    • PlatzierungundFixierungvon9 Aluminium-schalenmitjeweils1PolycarbonatmembraninjedemSiebbzw.Container

    • HerstellungdesSterilbarrieresystems• Sterilisation(extern)• Platzierungder(abgekühlten)Sterilgüter inderAerosolkammer

    • EntnahmederMembranenuntersterilenBedingungenimLabor

    • BebrütungderMembranenaufNährböden• PrüfungaufKeimwachstum• Positivbefund:keinSchutzgegenKontamina-tion durchluftgetrageneBakterien

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Fotos:Halyard Health

  • Folie18

    Ergebnisse(1):VliesverpackungenohneKontamination

    • Container• 12,6%ohneKontamination

    • 22,5%mit1– 9KBE• 46,8%mit10– 99KBE• 18,1%mit≥100KBE

    • Vliesverpackungen• 100%ohneKontamination

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    0

    20

    40

    60

    80

    100

    120

    140

    160

    180

    Container Vliesverpackungen

    0KBE 1- 9KBE 10- 99KBE ≥100KBEQuelle:(4)

  • Folie19

    Ergebnisse(2):KontaminationinAbhängigkeitvomContaineralter

    Sterilbarrieresysteme – Hart- vs.Weichverpackungeneven though identified and deemed “acceptable” and “in use” bythe supplying health care facility, had loose filter housings, mis-matched lids/bottoms, or dents/nicks on lids/bottoms. In addition,the seals of the rigid containers aged and lost some elasticitycompared with new seal material, and sustained nicks, cuts, orcreasing after years of use. However, even the majority of unusedcontainers showed bacterial ingress (72%), with high levels(>100 CFU) in some cases, which could be attributed to the designof the rigid container. The data also suggest that duration of use forrigid containers was at least partially accountable for theircompromised ability to maintain the sterility of their contentsunder the defined test conditions compared with sterilizationwraps. Figure 3 illustrates a trend suggesting that as rigid con-tainers age, the level of bacterial ingress increases. This would beexpected, as rigid containers sustain wear and tear from frequentuse.

    The sterilized instrument trays enclosed in sterilization wrapdemonstrated no detectable bacterial ingress. Unlike rigid con-tainers, these products are single-use items, negating issuesregarding age-based loss of performance. Although sterilizationwrap products must be inspected for tears and holes before andafter use, they are much less complicated devices than rigid con-tainers, which have numerous seals, filters, rivets, brackets, andlatches that can allow bacterial ingress poststerilization. In addi-tion, some wrap products are treated to create electret propertiesthat enhance small particle capture. Different grades of sterilizationwrap representing various levels of thickness and durability areused in the health care setting, based on the weight of the tray to bewrapped. No difference was found between the contaminationlevels of the different wrap grades; all demonstrated no detectablebacterial ingress.

    In accordance with standard hospital operations, various sterileindicators were used to validate and monitor the sterilizationconditions used in this study. Although these indicators demon-strate that an SPS has been properly sterilized, they are not inten-ded to monitor the poststerilization condition of the SPS before itsuse in the OR. All of the indicators used as part of this study indi-cated proper sterilization, yet M luteus contamination was stillfound in 87% of rigid containers. The clinical significance of thesefindings is not clear, given that the endpoint of this study was tomeasure bacterial contamination rates, not SSI rates, of SPSs in use.However, the term “sterile” is defined as no bacteria or microor-ganisms present, and by this definition, if a single organism

    ingresses into an SPS poststerilization, then the entire contents areto be considered nonsterile.

    Limitations of this study must be discussed to put the resultantdata into context with the test methodology. The aerosol challengeconcentration is expected to directly influence the contaminationlevel of the challenged devices. For this study, we selected thelowest viable challenge concentration possible that still provided aconsistent measurable endpoint. Bacterial aerosol concentration ina health care facility was measured by Greene et al,15 who found anaerosol concentration ofw2 CFU per liter of air, which isw100-foldlower than the challenge concentration used in this study.Assuming a linear relationship, if the challenge concentration ofthis study were lowered by 100-fold, then 18% of the rigid con-tainers would have still been contaminated. A follow-up study wasperformed on a limited number of SPSs (3 unused containers, 2containers of

  • Folie20

    Feststellungen,Bewertung,Bemerkungen(1)

    • „ImRahmendieserStudiekamesbeifestenContainernnachderSterilisationhäufig(89 %)zuBakterienbefall.EinigedergetestetenfestenContainerhatten,obwohlsievonderGesundheitseinrich-tung,vondersiestammten,als„akzeptabel“und„inGebrauch“eingeschätztwurden,loseFiltergehäuse,nichtpassendeDeckel/UnterteileundBeulen/KerbenandenDeckeln/Unterteilen.“

    • „AllerdingszeigtenauchdieMehrzahlderungenutzten[neuwerti-gen]ContainerKeimeinträge(72%),ineinigenFällenmithohenWerten(>100KBE),diedemDesignderContainerzugeschreibenwerdenkonnten.“

    • „DieErgebnisselassenvermuten,dassdieNutzungsdauerderContainerzumindestteilweiseverantwortlichfürihrebeeinträch-tigte Eignungwar,dieSterilitätihresInhalts[…]zubewahren.“

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Quelle:(4)

  • Folie21

    Feststellungen,Bewertung,Bemerkungen(2)

    • „DieinSterilisationsvliesenverpacktenInstrumentensiebezeigtenkeinenmessbarenKeimeintrag.“

    • „AndersalsContainerhandeltessichumEinwegprodukte,dienichtvoneinemaltersbedingtenLeistungsverlustbetroffensind.“

    • „KeinUnterschiedkonntezwischenunterschiedlichenVliesquali-tätenfestgestelltwerden;allezeigtenkeinenmessbarenKeimein-trag.“

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Quelle:(4)

  • Folie22

    Exkurs:FestlegungvonTestbedingungenzurPrüfungvonGebrauchsanforderungen

    • PhänomenundgeprüftesMerkmalsindzweiunterschiedlicheDinge!

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Phänomen Merkmal

    Evidenz

    Anliegen(Sorge)

    Begründung

    Messmethode

    Gebrauchs-anforderung

    Messgröße

    Quelle:AusarbeitungvonDr.MarkusWehrl undChristiaanMeijerfürNAMed 063-01-08GAOperationstextilien

  • Folie23

    Exkurs:FestlegungvonGrenzwertenzurPrüfungvonGebrauchsanforderungen

    • EineSchutzwirkungwirdüblicherweiseinFormeinesReduktions-faktors(derGefährdung)angegeben

    • DieEinhaltungeinesGrenzwertssolltemessbarsein(zumindestobjektivüberprüfbar)

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

    Messwert(alsoLeistung)

    untereNachweisgrenze

    Gebrauchsanforderung/Grenzwert

    untereBes

  • Folie24

    ZumVergleich:FranzösischeStudiestelltkeineKeimkontaminationbeiLagerungfest

    • CamilleFayardstelltfranzösischeStudiezurAuswirkungvonRestfeuchteaufdieSterilitätvonSterilgutaufdemSGSV-Kongress2016inBielvor(Präsentationaufwww.sgsv.ch)• BirgtRestfeuchteineinersterilenVerpackungwirklicheinpotenziellesRekontaminationsrisikofürOP-Siebe?

    • VorversuchezurErmittlungderVerdunstungsdauervon20mlund30mlWasserinHart-undWeichverpackungen• Nach10TageninbeidenVerpackungenkeineRestfeuchte• VerdunstungimContainerschnelleralsimVliestuch

    • PraxisnaheLagerbedingungen(LuftkeimmonitoringzeigtWerteüberdenGrenzwertenvonNFS90-351)• KeineexperimentellenLuftdruckänderungen

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie25

    UntersuchungsumfangderfranzösischenStudie

    • 6Containerund6SiebeinVliesverpackungmit15Keramik-Porzellan-PrüflingenaufleichterhöhtenPolypropylenschalenzumSammelnvonRestfeuchte

    • Jeweils1Positivkontrolle(ContainerohneFilter,Vliesperforiert)• Jeweils1Negativkontrollprobe(Keramik-Porzellan-PrüflingeinPlastik-Papier-Beutel)

    • AuswirkungRestwassernach0,1,3,7und14TagenLagerung• ÜberprüfunginflüssigemTSA-Nährboden• (Ggf.)RestwasserderPP-Schalen• Keramik-Porzellan-Prüflinge

    • BiokontaminationstestderLuftinderLagerzone

    Sterilbarrieresysteme– Hart- vs.Weichverpackungen

  • Folie26

    ErgebnisseundSchlussfolgerungenderfranzösischenStudie

    • UngünstigeLagerbedingungensorgenfürÜberexpositionderTestsets mitUmweltkeimen

    • Negativ-Kontrollennegativ• Positiv-Kontrollenpositiv(Koagulase-negativeStaphylokokkenbelegenmenschlichePräsenzinLagerzone)

    • Prüflinge• KeinePositiv-Befunde• 100%derSetssteril• KeineKontaminationungeachtetdesVerpackungstyps

    • „BeiVorhandenseinvonRestwassersindOP-SiebeinContainernoderdoppeltemVliestuchmindestens14Tagesteril“

    Sterilbarrieresysteme– Hart- vs.Weichverpackungen

  • Folie27

    DiskussionderexperimentellenBewertungderBarrierewirkung vonSterilbarrieresystemen

    • UnterschiedlicheexperimentelleAnsätze,dieeinemehroderwenigerstarkdifferenzierteLeistungsbewertungermöglichen• Ergebnis:WirkönnenimLaboroffenbarbesserevonwenigergutenSterilbarrieresystemen unterscheiden

    • BezugzurAnforderunginderPraxis• Wasistgutgenug?• Prüftechnisch:Gebrauchsanforderung,Grenzwert

    • BewertungderRobustheit?

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • QualitätssicherungbeiderAnwendungvonSterilbarrieresystemen

  • Folie29

    Qualitätssicherung:SicherungderSterilitätbiszurAnwendung

    • Herstellerangabenbeachten!• BeiderAufbereitung• InspektionFunktionskontrolle(Formhaltigkeit,Dichtung,etc.)

    • Wartung:ErsatzvonFilternundanderenVerschleißteilen(HinweisezuOriginalteilenberücksichtigen)

    • BeiderLagerung/vorderAnwendung• EinhaltungvonLagerbedingungenundfestgelegterLagerdauer

    • PrüfungderUnversehrtheitderVerpackungvorderAnwendung

    Sterilbarrieresysteme– Hart- vs.Weichverpackungen

    Quelle:Wagnersteriset Gebrauchsanweisung

  • Organisations- undArbeitsaufwandsowieKostenimVergleich

  • Folie31

    Routine-KonfigurationvonHart- undWeichverpackungen

    • Hartverpackung• Vlies• Container

    • Weichverpackung• Vlies• Vlies• Transportkorb

    • Anmerkung:BeideVerpackungssystemesehenzweiVerpackungenvor,diejeweilsaucheinzelnalsSterilbarrieresystem geeignetwären.BeidewerdennachwievorvonAnwendernmitBlickauf(doppelte)Transport- undLagersicherheitsowiediesichereasep-tischeEntnahmedersterilisiertenMedizinproduktegerngewählt.

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie32

    GemeinsamkeitenundUnterschiedebeiAufwandundKostenvonHart- undWeichverpackungen

    • ZumeistkeinenennenswerteUnterschiedebei• Transportlogistik• Sterilisation• Sterilgutlagerung

    • TendenziellUnterschiedebei• Produktauswahl,Beschaffung(Produktkosten,Arbeitsaufwand)• ValidierungdesVerpackungsprozesses(Prüfkosten,Arbeitsaufwand)• Lagerführung(Raum,Arbeitsaufwand)• Qualitätskontrolle,Pflege,Wartung,Reparatur(Kosten,Arbeitsaufwand)• AufbereitungderMehrweg-Komponenten(Container,Transportkörbe)• Müllentsorgung(Kosten,Arbeitsaufwand)• Verpackungsprozess(Produkt,Verbrauchsmaterial,Arbeitsaufwand)• WiederaufbereitungaufgrundbeschädigterVerpackung(Zykluskosten)

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie33

    VorbemerkungzumVergleichvonAufwandundKosten

    • NachfolgendesBeispielzuAufwandundKosten• basiertaufspezifischenVoraussetzungen• undkannhinsichtlichderdabeibetrachtetenFaktoreninteressantodersogarhilfreichsein

    • FürdieindividuelleEntscheidungvorOrtsinddiegenanntenWertenutzlos,weilnichtzwangsläufigübertragbar

    • Fürdieeigene,individuelleEntscheidungzählen• dieeigeneAusgangssituation(Containerbestand,Platz,Aufbereitungskapazität,etc.)

    • undZielsetzung(Fortführungoderggf.ÄnderungderSterilgutlogistik)

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie34

    EinBeispielfüreinenVergleichvonOrganisations-undArbeitsaufwandinZahlen

    • ImBeispiel:• ManuelleAufbereitungderContainer

    • BereitstellplatzfürKörbeundContaineridentisch

    • Validierungunberücksichtigt• BeiRückläufernnichtgesamteZyklus-Kostenberücksichtigt

    • (undeinigesmehr)• Kosten• Hartverpackung umetwa60%teureralsWeichverpackung

    Sterilbarrieresysteme– Hart- undWeichverpackungenimVergleich

  • FürdenHeimweg

  • Folie36

    Quintessenz?

    • Containersindrobust,bietendeswegenabernichtzwangs-läufigimmereinezufriedenstellendeBarrierewirkung gegenRe-Kontamination

    • ObunterdiesemGesichtspunktaufeineInnenverpackungverzichtetwerdenkann,isteinespannendeFrage

    • Vliesverpackungenschneidenexperimentellbesserab– aberwashilft‘s,wennsieaufdemWegzurAnwendungbeschädigtwerden?

    • DieFrageHart- oderWeichverpackungwirdalsosicherauchzukünftigaufgrundunterschiedlicherGewichtungderArgumenteunterschiedlich(undsehreinrichtungsspezifisch)beantwortetwerden

    • DabeispielenauchAufwandundKosteneinegroßeRolle– dieebenfallseinrichtungsspezifischermitteltundbewertetwerden

    Sterilbarrieresysteme – Hart- vs.Weichverpackungen

  • Folie37

    Quellen

    (1) DeBruin A.C.P.,Kastelein J.:Einfach- oderMehrfachverpackungvonMedizinprodukten:VerfahrensbewertungdurchForschung.Zentralsterilisation,1999,Ausgabe5,Seiten292- 303

    (2) Junghannß U.,WinterfeldS.,GabeleL.,Kulow U.:Hygienisch-mikrobiologischeundtechnischeÜberprüfungvonSterilisier-Containern.Zentralsterilisation,1999,Ausgabe3,Seiten154- 162

    (3) DunkelbergH.,Fleitmann-Glende F.:Measurementof the microbial barrier effictivenessof sterilization containers interms of the logreduction value for prevention ofnosocomial infections.AmericanJournalof Infection Control,2006,Ausgabe5,Seiten285- 289

    (4) Shaffer H.L.,Harnish D.A.,McDonaldM.,VernonR.A.,Heimbuch B.K.:Sterilitymaintenance study:Dynamicevaluation of sterilized rigidcontainers and wrappedinstrument trays to prevent bacterial ingress.AmericanJournalof Infection Control,2015,Seiten1336– 1341

    (5) Assessmentof ResidualMoisture and Maintenanceof Sterility inSurgical InstrumentSetsafterSterilization.Fayard C,LambertC,Guimier-Pingault C,Levast M,Germi R.InfectControlHosp Epidemiol.2015Aug;36(8):990-2