Upload
ahmad-atsari-sujud
View
230
Download
0
Embed Size (px)
Citation preview
7/24/2019 Sys Lab Magazine 595 433 En
1/1
The direct method
9 high accuracy thermo-hygrometers were located at the measuring positions
in Fig. 2., instead of the dew-point hygrometer and the thermometers [2].
For relative humidity, the stability and uniformity were evaluated in the same
way for temperature according to the G-20 in Ref.1.[1]
A comparison of two calibration methods for evaluating the performance of
climatic chambers was carried out. In this work, a 4.5 cubic foot PGC climatic
chamber was calibrated for its instability and uniformity both in terms of air
temperature and relative humidity.
In the first method, the absolute humidity inside the chamber was assumed to
be uniform, and therefore a dew-point hygrometer was used to measure only at
a corner of the chamber. The second method used 9 probes of high accuracy
thermo-hygrometer installed at different positions to evaluate the chamber
performance with out the uniform assumption as in the first method.
For both methods, the dew-point hygrometer and the thermo-hygrometers wascalibrated traceable to a NIMT 2 pressure primary generator. Due to the higher
accuracy of the dew-point hygrometer, the calibration uncertainty was lower
than that by the method using 9 probes of hygrometer.
For calibration results, the stability obtained from the two methods are in good
agreement within 0.1 %rh cover the measurement range, while the maximum
deviation of uniformity was larger than 0.5 %rh at the relative humidity of 92
%rh
COMPARISON OF CALIBRATION METHODS OF CLIMATIC CHAMBER
0.40
0.10
0.60
1.10
1.60
2.10
20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 100.0
Relative Humidit /%rh
Uniformity/%r
indirect
direct
T. Sinhaneti, T. Keawprasert, U. Norranim
Humidity Laboratory, Thermometry Metrology DepartmentNational Institute of Metrology (Thailand)
ABSTRACT
PRINCIPLE
COMPARISON RESULT
Fig. 1 illustrates variation of measurement temperature at different positions in a
climate chamber and an example for evaluation of stability and uniformity of the
23.05
23.1
23.15
23.2
23.25
0 5 10 15 20 25 30 35
Time (min)
Measurementtemperature(oC)
1
2
3
4
5
6
7
8
ref
1.50
1.00
0.50
0.00
0.50
1.00
1.50
20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 100.0
Relative Humidity/%rh
Stability/%rh
indirect
direct
The stability and the uniformity of climatic chamber when compare between
indirect and direct method deviated about +0.1%rh excepting the uniformity at
92%rh (~0.5%rh). Estimate measurement uncertainty in indirect method is lower
than that in direct method because of high accuracy of dew-point hygrometer and
the uniformity assumption in dew-point temperature. However the direct method
is still more convenient without complicating to calculate the relative humidity
via dew-point temperature. Finally, both method are capable to calibrate the
climatic chamber but the indirect method should be carried out due to lower
measurement uncertainty.
CONCLUSION
Figure 2. Schematic diagram for installing
the thermometers and the dew-point
thermometer
REFERENCES
CALIBRATION METHODS
Stability = tmax-tmin
= 23.17-23.14= 0.03oC = +0.015oC
Uniformity = Max (ti-t Ref)
= 23.2-23.07= 0.13oC = +0.065oC
es c am er in empera ure.
Figure 1. Illustration for determination of stability and uniformity for air temperature
Figure 4. Stability by comparison between indirect method and direct method.
Figure 3. Uniformity by comparison between indirect method and direct method
[1] G-20 Guidelines for Calibration and Checks of Temperature Controlled
Enclosures
[2] DKD-R 5-7 Guideline for calibration of Climatic Chamber.
t1
t2 t6
t5
t7t3
t4 t8
tdp.t ref.
The indirect method
9 thermometers and a dew-point
hygrometer were placed at measuring
positions as shown in Fig. 2.
The stability and uniformity in
temperature were determined
according to G-20 guideline [1].
The uniformity in relative humiditywas calculated via the measurement
temperature distribution and dew-point
temperature [2].