Table of contents Basis structure of SEM SEM imaging modes Comparison of ordinary SEM and FESEM Electron behavior Electron matter interaction Electron matter interaction o Elastic interaction o Inelastic interaction o Interaction volume and its measurement Backscattered electrons and imaging Secondary electrons and imaging

Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

  • Upload
    others

  • View
    11

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Table of contents Basis structure of SEM

SEM imaging modes

Comparison of ordinary SEM and FESEM

Electron behavior

Electron matter interaction Electron matter interaction

o Elastic interaction

o Inelastic interaction

o Interaction volume and its measurement

Backscattered electrons and imaging

Secondary electrons and imaging

Page 2: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Basic structure of SEM

• Two major component – Electron column and Control console

Various subsystem of SEM and their function is explained

Electron Column Control Console

Electron gun and 2 or more electron lenses to control the path of travelling electron in the vacuum column

Computers, Monitors and knobs to control various imaging parameters

Page 3: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Electron gun generate electron and accelerated them to an energy level

Basic structure of SEM

Electron gun and lenses

them to an energy level (0.1- 30keV)

• Electron lenses focuses the beam to smaller spot size to produce sharp images

Page 4: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Deflection system consists of electromagnatic coils.

• Causes the beam to move in a series of discrete points in a rectangular region to scan

Basic structure of SEM

Deflection system

rectangular region to scan complete specimen

• Excitation of coils also control the magnification

• In modern SEM, magnification is automatically compensated against working distance from sample

Page 5: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Detectors collect the signal generated from interaction of beam with specimen.

• Electronic of detectors convert the signal into digital images.

• Most often collected signal are SE (Secondary electrons) and BSE (Backscattered electrons) signal.

Basic structure of SEM

Electron Detectors

BSE (Backscattered electrons) signal.

Page 6: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• The very first control for generation of electron beam is Acceleration voltage and Emission current.

• Then condenser lens control – amount of beam current available and minimum beam size

Basic structure of SEM

Operator controls

available and minimum beam size

• Next is objective lens control – to focus the beam at a smaller diameter spot.

• Finally viewing screen controls like brightness, contrast etc.

Page 7: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

SEM imaging modes

• Electron probe size dia. (dp) - Dia. of electron beam focused at the specimen

• Electron probe current (ip) - Current impinging upon the specimen and

4 parameters controls different imaging modes in SEM

impinging upon the specimen and generating various signals

• Electron probe convergence angle (αp) - Half cone angle of electrons converging onto the specimen

• Electron beam accelerating voltage (Vo) kV

Page 8: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

SEM imaging modes

Mode I - Resolution Mode

• This mode is governed by the Electron probe size dia. (dp)

• Resolution refers to the size of finest detail that can be observed

• To image the finest details dp must be comparable with or smaller than the feature itself.

Resolution mode is only meaningful at high image magnifications • Resolution mode is only meaningful at high image magnifications (>10,000 X)

• Beam should contain sufficient current to exceed visibility threshold

Page 9: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

SEM imaging modesMode II – High Current Mode

• This mode is governed by the Electron probe current (ip)

• For the best image visibility and quality – Large ip• Unless a sufficient amount of current (required to overcome

random noise) is there details cannot be seen even if the spot size is small enough.size is small enough.

Smallest probe dia. – 15 nmMinimum current – 1 pA

Smaller probe dia. – 20 nmAdequate current - 5 pA

Larger probe dia. – 130 nmHigh current Mode – 320 pA

Noisy Perfect Blurred

Page 10: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

SEM imaging modesMode III – Depth-of-Focus Mode

• This mode is governed by the probe convergence angle (αp)

• For the best image quality – αp as small as possible

• With low beam convergence angle, beam dia. changes only a little over a long vertical distance, so surface features at different heights will all appear to be in focus at the same time.

αp= 15 mrad

αp= 1 mrad

Page 11: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

SEM imaging modesMode III – Low-Voltage Mode

• This mode is governed by the beam acceleration voltage

• At low accelerating voltage (≤ 5 kV), the beam interaction with the specimen is confined to the regions very close to the surface

• This provides an image rich in surface details compared to images at higher accelerating voltages.

V0= 5 keV

V0= 15 keV

V0= 30 keV

Sharp surface feature Poor surface feature Poor surface feature

Page 12: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Most of the older SEM uses Tungsten hairpin or thermionic Lanthanum Hexaboride (LaB6) Electron sources

• These thermionic sources are resistively heated to very high temperature for the emission and release of electrons

• Advantage: Less expansive source, no special vacuum requirement

Ordinary SEM

• Disadvantage: Low brightness, limited lifetime and large energy spread

Tungsten hairpin (LaB6)

Page 13: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

FE (Field emission) SEM• Field emission is alternate method where potential barrier for electrons is

lowered and narrowed by applying an negative potential at a very sharp wire tip (radius ≤ 100 nm) cathode.

• When the tip concentrated field reaches about 10 V/nm electrons emit• FE sources provide enhanced performance, reliability and lifetime.• Two class of field emitter (I) Cold field emitter (CFE) and (II) Schottky field

emitters (SFE) and Thermal field emitters (TFE)• CFE relies purely on the high applied field for electron emission independent

of the tip temperature.• TFE has the same property as CFE, but is operated at elevated temperature, • TFE has the same property as CFE, but is operated at elevated temperature,

helps to keep the tip clean, reducing noise and instability even in degraded vacuum condition.

• SFE guns are similar to other FE sources, but include a suppresser grid to eliminate unwanted thermionic emission from region outside of the tip.

CFE SFE

Page 14: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Comparison of ordinary SEM and FESEM

Comparison of various parameters for different electron sources at 20 kV

Better performance of FE sources are evident from the comparison

Page 15: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Elementary particle carrying a negative charge

• Accelerated electron not only acts as a particle but acts as wave too.

• Rest mass of electron = mo = 9.109 x 10-31 kg

• Charge of electron = e = -1.602 x 10-19 C

• For an accelerating electron, wave length of moving electron

Electrons

• For an accelerating electron, wave length of moving electron (considering the relativistic effects) is given as,

• Above equation in useful in calculating the acceleration potential for a particular wave length of electron waves.

20

02

12cm

eVeVm

h

Where, h = Planck Constant, V= Applied Voltagec = Velocity of light in vacuum = 2.998 x 108 m/s

Page 16: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Electron wave

• Waves in electron beam can be coherent or incoherent.

Coherent Waves of same wave lengths

Incoherent Waves of different wave lengths

• Electron beam from a SEM electron source is a bundle of coherent electron waves, before hitting a specimen.

• After interacting with specimen, electron waves can form either coherent or incoherent beams

Page 17: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Electron matter Interaction

• Electron entering a material interacts as a negatively charged particle with electric fields of specimen atoms.

• The positive charge of the atom in strongly concentrated at the nucleus, whereas negatively charge electron atoms are much more dispersed in the a shell structure.

+Positively Charged

Dispersed Negative electron -

- ---

• These interactions are classified in to two different types.• Elastic interactions: In this case no energy is transferred from

electron to sample. As result electron leaving the sample still has the original energy.

• Inelastic interactions: The energy of the incident electron is transferred to the sample atoms. Hence, after the interaction electron energy is reduced.

+Charged nucleus

electron charge

-- --

Page 18: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Elastic interactions deflects the electron beam along new trajectory, causing them to spread laterally.

• A strong elastic scatter very near to the nucleus may result in beam electron leaving the specimen via back scattering, called Backscattered electrons (BSE). These electrons provide an important

Elastic Interaction

These electrons provide an important class of information for SEM imaging.

• Probability of elastic scattering

• Increases strongly with atomic number (as Z2), as heavier atoms have much stronger positive charge at nucleus

• Decreases as electron energy increases (as 1/E2)

Page 19: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Inelastic Interaction

-

Incident Electron

Secondary Electron

• With the elastic scattering, beam electrons loose energy to specimen atoms in various ways.

• Such inelastic interactions gives rise to useful imaging signals such as Secondary Electrons (SE) and analytical signals such a X-rays.-

Diffused Electron

analytical signals such a X-rays.

• This energy loss happens gradually with distancee travelled inside the atom.

Page 20: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Interaction Volume

• Elastic and Inelastic scattering together distribute a single incident electron beam over a 3-Dimensional region inside the material.

• Region generated because of these phenomenon is called interaction volume.

Incident Electron Beam

volume.

• Size and shape of the interaction volume depends on the relative amount of elastic and inelastic scattering on electron beams.

• Direct measurement of interaction volume for materials with intermediate and high atomic number materials is not possible.

Scattered Electron Beam

Interaction Volume

Page 21: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Interaction Volume - Simulation

• Monte Carlo simulation is highly useful tool to simulate the scattering of electron beam in a material.

• To construct a Monte Carlo electron trajectory simulation, the effects of elastic and inelastic scattering are calculated from appropriate models to determine scattering angles, distances between scattering sites ("step length"), and the rate of between scattering sites ("step length"), and the rate of energy loss with distance traveled.

• Effect of different parameters like atomic number (Z), beam tilt angle, beam energy etc. can be studied using these simulations.

• An online version of this simulation can be found at following link

www.matter.org.uk/tem/electron_scattering.htm#

Page 22: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Electron RangeA simple measure of interaction volume

• Interaction volume is a complex 3-dimensional region which depends on beam energy, material parameters, and tilt angle (complement of angle between the beam and surface plane).

• Despite this complexity, the interaction volume can be simply described with a single parameter called "electron range.“

• The electron range is thus a simple value that is useful for rough comparisons and scaling various signal distributions.

• Within the electron range, the density of scattering events changes sharply with distance from the beam impact area.

• Analytical equations of the electron range are available in literatures.

Page 23: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• Using the Monte Carlo depiction of the interaction volume, consider a hemisphere constructed with a radius whose origin is the entry point of the beam into the specimen and which contains a specified fraction of the electron trajectories, for example, 90%. Radius of this hemisphere is called electron range.

Electron RangeA simple measure of interaction volume

called electron range.

67.1089.0

0276.0)( E

Z

AmRKO

Where, A is the atomic weight (g/mole), Zis the atomic number, ρ is the density (g/cm3) and E0 is the beam energy (keV)

Electron Range (RKO)

For using this formula, the specimen is assumed to be flat, thick enough to be electron-opaque and of sufficiently large lateral extent that there are no edges or boundary within electron range. Tilt angle is 0 degree.

Page 24: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - Imaging• Backscattered electrons (BSE) are beam electrons whose

trajectories have intercepted a surface usually, but not necessarily, the entrance surface and which thus escape the specimen.

• Backscattered electrons remove a significant amount of the total energy of the primary beam.

• Backscattering is quantified by the backscatter coefficient η• Backscattering is quantified by the backscatter coefficient ηwhich is defined as

B

BSE

B

BSE

i

i

where ηB is the number of beam electrons incident on the specimen and ηBSE is the number of backscattered electrons (BSE). The backscatter coefficient can also be expressed in terms of currents, where iB refers to the beam current injected into the specimen and iBSE

to the backscattered electron current passing out of the specimen.

Page 25: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - Imaging

• Backscattering increasesmonotonically as Atomic number (Z) increases. Hence it forms the basis foratomic number contrast.

Atomic number dependence of BSE

at E0=20keV

atomic number contrast.

• The slope of η vs Z is initially steep but decreases with increase in Z. Hence atomic number contrast between adjacent pair of elements is strong at low atomic number and weak at high atomic number.

3724 103.81086.1016.00254.0 ZZZ

Page 26: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - Imaging

• For beam energy 5-50 keVthere is only a small change (<10%) in the backscatter coefficient (η) w.r.t. beam energy.

Beam energy dependence of BSE

energy.

• At beam energies below 5 keV, η of light elements increases and those for heavy element decreases.

32 )(ln01491.0)(ln1292.0ln2235.01904.0

and )/9211.0(1382.0 where,

),(

ZZZC

Zm

CEEZ m

Page 27: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - Imaging

• Backscattering increasesmonotonically with tilt angle.

• This increase is the basis for topographic contrast in SEM, by which shape of object is

Title angle dependence of BSE

which shape of object is recognized.

• At very high tilt angle, which correspond to grazing incidence, values of η tends towards unity.

• For different elements at high values of θ , η tend to converge for all elements

Page 28: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - Imaging

• Angular distribution of BSE is defined relative to normal of incident surface

• η at an angle φ is given by,

Angular distribution of BSE – Normal beam incidence (0˚ tilt)

cos)( n

where, ηn is the value measurement along normal vector n.

• Thus maximum number of BSE is emitted along the normal to surface

• As the detector is placed at an angle φ n away from surface normal, number of BSE decreases.

• At shallow angle just above the surface, there are virtually no BSE trajectories.

• Thus Angular distribution of BSE detector relative to surface specimen will have a strong influence on its collection efficiency.

Page 29: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - Imaging

• For titled surface angular distribution becomes asymmetric.

• At higher tile angles (>45˚) distribution resembles a highly elongated ellipse, with the long axis at approx. same angle above the

Angular distribution of BSE – Non-normal beam incidence

at approx. same angle above the surface as the incident beam.

• Titling of specimen also results in asymmetry of interaction volume, this asymmetry also affects backscattering.

• For highly tilted surfaces, many of beam electrons skip off the first few atom layers and exit the specimen only after a few scattering events, so that the most are contained in a plane defined by the beam vector and the surface normal.

Page 30: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - ImagingEnergy distribution of BSE

• The energy distribution is a continuum extending from the incident beam energy to essentially no energy.

• Region I represents the high energy dump of BSEs that have lost less than 50% of E .

III

50% of E0.

• For most target of intermediate and high atomic number, the majority of BSEs will be found in region I.

• Region II is the gradually decreasing tail of energy distribution represents the electron that travel progressively greater distances, loosing progressively more energy within the specimen prior to backscattering

Page 31: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Backscattered Electrons - ImagingEnergy distribution of BSE

• The energy distribution peak becomes more distinct for higher atomic number targets.

• Cumulative integrals calculated from the BSEs energy distributions from the BSEs energy distributions show that BSEs retain at least 50% of the incident beam energy, with the fraction rising sharply for intermediate and high-atomic-number targets.

• Low-atomic number scatterersproduce far fewer high-energy BSEs.

BSE Energy distribution

Page 32: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Secondary Electrons - Imaging• Secondary electrons (SE) are loosely bound outer shell

electrons from the specimen atoms which receive sufficient kinetic energy during inelastic scattering of the beam electrons to be ejected from the atom and set into motion.

• Thus SE will propagate through the solid, and some will intersect the surface and escape.

• SE are defined purely on he basis of their KE, i.e. all emitted • SE are defined purely on he basis of their KE, i.e. all emitted electron less than a particular energy (say 50eV) are considered as SE.

• The total SE coefficient δ is given byB

SE

B

SE

i

i

where ηSE is the number of secondary electrons emitted from a sample bombarded by ηB beam electrons and I designates the corresponding currents.

Page 33: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• δ generally rises as the beam energy is lowered, as shown in table below, from increase in energy from 5keV to 50keV, δgradually decreases

Secondary emission as a function of energy

Secondary Electrons - Imaging

• SE generated while the incident beam passes through specimen surface are designated as SE1. The SE1 signal is inherently is high-resolution signal that preserves both the lateral spatial resolution of the focused beam and the shallow sampling depth of the SE.

Page 34: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

• SE are also generated by the BSE approaching the surface. These are designated as SE2. SE2 signal is a low-resolution signal.

Secondary Electrons - Imaging

• Compared to η , δ is relatively insensitive to atomic number.

• The emission on SE is very sensitive to the condition of the surface because the low KE severely limits the range of SEs.

• As the angle of tilt (θ) increases, δ increases.

Page 35: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Secondary Electrons - Imaging• η + δ – Total electron emission from sample (BSE +

SE2)

• Variation of η + δ with beam energy is shown.

• As energy decreases there comes a crossover point E2, where η + δ reaches unity. With further decrease in energy, η + δ increases above unity, i.e. more electron are emitted from the surface then supplied by the beam.

Upper crossover energy for various materials (Normal beam incidence)

by the beam.

• η + δ reaches a peak and then reduces on further decrease in energy, until the second crossover point E1 is reached, and then continues to decreases.

• Values of E1 are less than 1keV and is difficult to measure

• Values of E2 can be as high as 5-20, as listed

• Knowledge E2 point and η + δ behavior at low beam energy forms the basis for controlling charging in SEM imaging of insulators through careful choice of beam energy and specimen tilt.

Page 36: Table of contents - IIT KanpurSEM imaging modes Mode I - Resolution Mode • This mode is governed by the Electron probe size dia. (dp) • Resolution refers to the size of finest

Summary• Basis structure of a SEM, construction and functioning of its

important components are discussed.

• Different modes of imaging in the SEM is explained with the effect of different parameters on the quality of imaging in all modes.

• Difference between ordinary SEM and FESEM is discussed is briefbrief

• With a short review of electron behavior and properties, interaction of electron beam with material inside the SEM is explained.

• Various imaging signals resulting from the interaction of beam and specimen are discussed. Effects of different signal parameters and their importance for best quality of imaging is explained in details