TI00473C/07/EN/03.12
71164110
Technical Information
Liquistation CSF22
Automatic stationary sampler for liquid media
Integrated multiparameter controller with Modbus RS485
communication
Application
Liquistation CSF22 is a stationary sampler designed for the fully
automated sampling, defined distribution, and
temperature-controlled storage of liquid media.
The standard product version has one 0/4 to 20 mA analog
input, one binary input and one binary output.
The sampler is designed for use in the following indoor
applications:
• Municipal and industrial sewage treatment plants
• Laboratories and Water Conservancy Boards
• Monitoring of liquid media in industrial processes
Your benefits
• Rugged and reliable:
– Refrigerator with 304x housing
– Reliable sample temperature regulation
– Optimal air circulation in the lower enclosure
• Simple and easy to use:
– Swift menu guidance, navigator and large display
– Bottle tray for easy sample transportation
– Parts carrying medium are easy to disassemble, making
cleaning and maintenance simple
• Flexible:
– User friendly programs ranging from simple time programs
to event programs
– Functionality can be extended by connecting different
input parameters
• Communicative:
– Integrated data logger for recording measured values
– Service interface for data transmission
– Modbus RS485 communication
• Safe:
– Automatic diagnostic monitoring
– Optional battery backup system ensures uninterrupted
operation in the event of power failure
Liquistation CSF22
2 Endress+Hauser
Function system design
Liquistation CSF22 sampler A complete sampling unit comprises:
Liquistation CSF22, with the following depending on the version ordered:
• Controller with display, soft keys and navigator
• Peristaltic pump for sampling
• Plastic (PE) sample bottles for sample preservation
• Sample compartment temperature control for safe sample storage
• Suction line with suction strainer or filter
a0014369
Example of a Liquistation CSF22
1
2
3
4
5 6
7
8
9
1 Bottle distribution 6 Power supply
2 Distribution arm 7 Peristaltic pump
3 Medium detetection and suction-line connection 8 Door lock
4 Signal connection 9 Refrigerator display
5 Controller
Liquistation CSF22
Endress+Hauser 3
Mode of operation
a0011584
Sampling steps with a peristaltic pump
Sampling takes place in three steps:
1. Purge/rinse
The peristaltic pump runs in reverse and forces liquid back to the sampling point.
2. Intake
The peristaltic pump runs forward and draws in liquid. Once the liquid detection system detects the
sample, the pump is controlled by the flow and the specified sample volume is calculated automatically.
3. Purge
The pump runs in reverse again and forces the liquid back to the sampling point.
One advantage this system offers for obtaining a representative sample is the possibility of rinsing the suction
line several times:
Liquid is initially drawn in until the liquid detection system reacts, then the pump switches and forces the liquid
back to the sampling point. This process can be repeated a maximum of three times. The sample is then taken
as described.
The pump rollers deform the tubing, thereby causing a negative pressure and the suction effect.
The liquid detection system is based on a pressure sensor which detects the difference between a pipe that is
filled and not filled.
Thanks to a patented process for automatically detecting the suction height, the user does not have to enter
the suction height or suction line length. The self-calibrating software guarantees that the sample volume
remains constant.
1. 2. 3.
a0014451
Peristaltic pump
1
2
3
Pump tubing
Pump rollers
Liquid detection system (patented)
1 2 3
Liquistation CSF22
4 Endress+Hauser
Navigator and plain text
display
The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device
Controller interfaces The controller of the sampler has the following functions:
• Socket for M12 connector Modbus RS485 (5-pin, B-coded, female)
• Multi I/O connector
• Service connector
a0014360
Connection sockets
1 Socket for M12 connector (5-pin, B-coded, female)
2 Multi I/O connector
3 Service connector
Display Graphic display:
• Backlight with switch-off function
• Red background display for alarms alerts users to errors
• Transflective display technology for maximum contrast even in bright environments
a0013216
Easy operation
a0013222-en
Plain-text menu
12
3
a0013627
Example of program setup
Liquistation CSF22
Endress+Hauser 5
Sampling control
Intake speed with different
suction lines
a0014452
Intake speed in m/s at suction heights in m (vertical lift)
1 Intake speed as per Ö 5893 (Austrian standard)
2 Intake speed as per EN 25667, ISO 5667
a0014045
a. Flow curve
b. Time proportional sampling:
Constant time constant volume (CTCV)
A constant sampling volume is taken at steady intervals.
c. Flow proportional sampling:
Variable time constant volume (VTCV)
A constant sampling volume is taken at variable intervals.
d. Flow proportional sampling/time override:
Constant time variable volume (CTVV)
A variable sampling volume is taken at steady intervals.
e. Event-controlled sampling:
Sampling is triggered by an event.
Single and multiple samples can also be grouped in a program in
addition to the sampling methods listed. Furthermore, the software
allows interval sampling, switchover and event functions. The latter
permit up to 24 subprograms to be active simultaneously for a
variety of applications.
A sampling table makes it possible for users to program the bottle
assignment, time interval and sample volume.
Signals for external control can be connected via analog input and
binary input in the standard version of the product. Customized
text is entered to ensure the correct assignment of the inputs in the
memory.
1.2
1.0
0.8
0.6
0.4
0.22 3 4 5 6 7 8 m
6.6 9.9 13.1 16.4 19.7 23 26.3 ft
3.9
3.3
2.6
2.0
1.3
0.7
ft/s m/s
1
2
Liquistation CSF22
6 Endress+Hauser
Sample distribution CSF22 offers a range of bottle combinations and distribution versions. The sequencial version can be changed
or replaced easily without the need for special tools. In addition, the software makes it possible to configure
individual bottles and bottle groups as well as assign them to change based on a programmed event.
Note!
The composite version can not be upgraded to sequencial.
Bottle groups and distribution version depending on the order version:
Sample preservation The sample bottles are located in the sample compartment made of stainless steel.
All parts that transport liquid can be removed and cleaned easily without the need for additional tools.
a0015194
Sample bottles
Sample temperature
regulation
The temperature of the sample compartment can be adjusted using the controller. The factory setting is 4 °C
(39 °F). The current temperature is shown on the display.
a0014453
Display of cooling system
CSF22-**1** CSF22-**2** CSF-**3**
20 liter, PE
direct distribution1
2 liter, PE
direct distribution12
1 liter, PE
direct distribution24
Liquistation CSF22
Endress+Hauser 7
Maintainability • Independent, integrated ring memories (FIFO) or stack memories for recording:
- An analog value (e.g. flow rate, pH value, conductivity)
- Events (e.g. power failure)
- Sample statistics (e.g. sample volume, filling times, bottle assignment)
• Program memory: max. 100 programs
• Data logbooks:
– Adjustable scan time: 1 to 3600 s (6 h)
– Max. 8 data logbooks
– 150,000 entries per logbook
– Graphic display (load curves) or text display
• Calibration logbook: max. 75 entries
• Hardware version logbook:
– Hardware configuration and modifications
– max. 125 entries
• Version logbook:
– e.g. software updates
– max. 50 entries
• Operation logbook: max. 250 entries
• Diagnostics logbook: max. 250 entries
a0015032
Data logbook: Graphic display
Liquistation CSF22
8 Endress+Hauser
FieldCare
Configuration and asset management software based on FDT/DTM technology
• Complete device configuration when connected via FXA291 and service interface
• Logbooks can be downloaded in CSV format or binary format for "Field Data Manager" software
Field Data Manager
Visualization software and database for measuring, calibration and configuration data
• SQL database which is protected against manipulation
• Functions to import, save and print out logbooks
• Load curves to display measured values
a0016009
Field Data Manager: Load curves
Liquistation CSF22
Endress+Hauser 9
Input
Input types • 1 analog input
• 1 binary input
Binary input, passive
Span 12 to 30 V, galvanically isolated
Signal characteristics Minimum pulse width: 100 ms
Analog input, passive
Span 0/4 to 20 mA, galvanically isolated
Accuracy ±0.5 % of measuring range
Liquistation CSF22
10 Endress+Hauser
Output
Output types 1 binary output (standard):
Open collector, max. 30 V, 200 mA, galvanically isolated
Modbus RS485
Protocol-specific data
Power supply
Electrical connection --> For a detailed wiring diagram, see the Operating Instructions for Liquistation CSF22
Supply voltage 220 V AC ±10 %, 50 Hz
Caution!
A circuit breaker with a maximum rating of 10 A must be provided by the customer.
Power consumption 240 VA
Power failure • Real-time clock: lithium battery, type CR2032
• Power supply (optional): with additional charging control unit
Signal coding EIA/TIA-485
Data transmission rate 2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200 baud
Galvanic isolation Yes
Connectors M12 (5-pin, B-coded, female)
Protocol RTU/ASCII
Function codes 03, 04, 06, 08, 16, 23
Broadcast support for function codes 06, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 4 digital values (value, status),
diagnostic information
Supported features Address can be configured using switch or software
Liquistation CSF22
Endress+Hauser 11
Performance characteristics
Sampling methods • Time proportional sampling (CTCV)
• Flow proportional sampling (VTCV)
• Flow proportional sampling/time override (CTVV)
• Event sampling
• Single and multiple samples
• Sampling table
Sample volume 10 to 10000 ml (0.34 to 340 fl.oz.)
Dosing volume < 20 ml can vary in dosing accuracy and repeatability depending on application.
Dosing accuracy ±5 ml (0.17 fl.oz.) or 5 % of the set volume
Dosing volume < 20 ml can vary in dosing accuracy and repeatability depending on application.
Repeatability 5 %
Intake speed > 0.6 m/s (> 1.9 ft/s) for 10 mm (3/8") ID
Suction height Max. 8 m (26 ft)
Hose length Max. 30 m (98 ft)
Temperature control • Sample compartment temperature range: 2 to 20 °C (36 to 68 °F) ±2 °C
• Factory setting: + 4 °C (39 °F)
• Automatic defrost system
Liquistation CSF22
12 Endress+Hauser
Installation
Mounting conditions
a0014359
Mounting conditions
1. Correct
The suction line must be routed with a downward slope to the sampling point.
2. Incorrect
The sampler should never be mounted in a place where it is exposed to aggressive gases.
3. Incorrect
Avoid siphoning effects in the suction line.
4. Incorrect
The suction line should never be routed with an upward slope to the sampling point.
Note!
Note the following when mounting the sampler:
• Mount the sampler on a level surface.
• Protect the sampler from additional heating (e.g. from heaters).
• Protect the sampler from mechanical vibrations.
• Protect the sampler from strong magnetic fields.
• Do not mount the sampler directly over the inlet channel of the wastewater treatment plant (sulfur vapors!).
Liquistation CSF22
Endress+Hauser 13
Environment
Ambient temperature range 5 to 35 °C (41 to 95 °F)
Storage temperature -20 to 60 °C (0 to 140 °F)
Relative humidity 10 to 95%, not condensing
Degree of protection • Controller front panel: IP 65
• Controller with closed cover: IP 54
• Bottle compartment front: IP 54
Process
Sample temperature range 2 to 40 °C (36 to 104 °F)
Process pressure • Unpressurized, open channel (unpressurized sampling)
• Max. 0.8 bar piping (pressurized sampling)
Sample properties Sample media has to be free of abrasive substances.
Caution!
Pay attention to the material compatibility of the wetted parts.
Liquistation CSF22
14 Endress+Hauser
Mechanical construction
Dimensions
a0014362
Dimensions Liquistation CSF22 (depth = 637 mm (25.1"))
Weight 63 kg (139 lbs)
Material
Process connections Suction line ID 10 mm (3/8")
485 (19.1)
620 (24.4)
595 (23.4)mm (inch)
1154
(45.
4)
828
(32.
6)
Non-wetted parts
Cabinet housing Stainless steel 304x
For standard applications in wastewater treatment plants and
environmental monitoring
Sample compartment inner shell Stainless steel 304x
Wetted parts
Pump tubing Silicone
Distribution arm Plastic PE
Composite container/bottles Plastic PE
Suction line Plastic EPDM
Suction line connection Plastic PP
Liquistation CSF22
Endress+Hauser 15
Operability
Operating concept The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device
Display and operating
elements
Remote operation
a0016785
Easy operationa0013222-en
Plain-text menu
a0013639
• Liquid crystal display, backlighting
• 160 x 240 pixels
• 4 operating keys (soft key function) and navigator
• Menu-guided operation
a0016787
Modbus RS485
T Terminating resistor
FXA291
Modbus Master / Gateway
T T
Modbus RS485(RTU, ASCII)
CSF22
Liquistation CSF22
16 Endress+Hauser
Communication • 1 service interface, accessible via front panel connection
• Commubox FXA291 (accessory) required for communication with the PC
• M12 connector Modbus RS485 (5-pin, B-coded, female)
Software Field Data Manager
• Standardized user interface under Windows®
• Reading out the internal memory containing the measured flow rate, sample volume taken etc.
FieldCare
• Device settings saved in a database
• Configuration
Liquistation CSF22
Endress+Hauser 17
Ordering information
Product structure
Note!
Simply add the additional options to the order code you selected. Please contact your sales office if you have
any questions.
Note!
To receive a complete, operational sampling system, you have to order at least:
• 1 x CSF22 sampler
• 1 x sample distribution
• 1 x suction line with sample strainer
or
1 x suction line with sample filter
• 1 x signal cable to connect analog or binary inputs and outputs
Scope of delivery The scope of delivery comprises:
• 1 stationary sampler with:
– The ordered bottle configuration
– Optional hardware
• 1 "Commissioning" Operating Instructions
(In the preferred language if the "Default operating language" order option is selected. Otherwise, the Brief
Operating Instructions are supplied in English)
• Optional accessories
• 1 spare pump-tubing
Approval
AA Non-hazardous area
Sample distribution
1 1 x 20 liter (5.28 US gal.) PE
2 12 x 2 liter (0.53 US gal.), PE
3 24 x 1 liter (0.26 US gal.), PE
Sample temperature control
A Refrigerator indoor use, housing 1.4301 (AISI 304)
Additional option
1 Basic version
CSF22- Order code
Operating language (only one option may be selected)
AA English
AK Chinese
Accessory enclosed
N1 Suction line 10 mm (3/8") ID, EPDM, length 10 m (32 ft) + strainer 316(x)
N2 Suction line 10 mm (3/8") ID + filter, EPDM, length 10 m (32 ft)
N3 Signal cable to I/O connector
N4 Power pack for battery back-up sampling
Liquistation CSF22
18 Endress+Hauser
AccessoriesIn the following sections, you find the accessories available at the time of issue of this documentation.
For information on accessories that are not listed here, please contact your local service or sales
representation.
Accessories for Liquistation
CSF22
a0014726
Complete suction line
1 with stainer
2 with filter (protection removed)
Bottle tray + bottles + cap
71111155 Bottle tray + 12 x 2 liter (0.53 US gal.) PE wedge shaped bottle + cap
71111156 Bottle tray + 24 x 1 liter (0.26 US gal.) PE wedge shaped bottle + cap
Bottles + caps
71111170 20 liter (5.28 US gal.) PE + cap, 1 pc.
71111176 1 liter (0.26 US gal.) PE wedge shaped bottle + cap, 24 pcs.
71111178 2 liter (0.53 US gal.) PE wedge shaped bottle + cap, 12 pcs.
Complete suction line
71111234 Suction line ID 10 mm (3/8"), EPDM, length 10 m (33 ft), strainer V4A
71120800 Suction line ID 10 mm (3/8"), ID + filter, EPDM, length 10 m (33 ft)
Pump tubing customized
71114701 Pump tubing, 2 pcs.
71114702 Pump tubing, 25 pcs.
Accessories, electrics
71129420 Battery back-up (UPS), 1000 VA / 230 V
71130547 Signal cable to multi I/O connector
Communication, software
51516983 Commubox FXA291 + FieldCare Device Setup
71129799 Field Data Manager Software, Analysis
Liquistation CSF22
Endress+Hauser 19
Endress+Hauser Sales Center China
Endress+Hauser ShanghaiAutomation Equipment Co. Ltd.No. 458 Jiang Chuan Dong RoadShanghai 200241, P.R. China
Tel.: +86 21 24 03 96 00Fax: +86 21 24 03 96 03www.endress.comE-Mail: [email protected]