Upload
others
View
6
Download
0
Embed Size (px)
Citation preview
*Thanks for using the C-FILL obturation system intrument.For the sake of safely and effcient use of the product,please read this manual carefully.Ver2.0 Revision Date:2011-03-24
Standard Symbols
Take care of burn because of the high temperatureof the Pen-Tip and Needle during operation.
Dangerous Voltage Wihtin the product’s enclosure that may be of Suffcient magnitude to constitute riskof electric shock to persons.
ON(power connection)
OFF(power disconnection)
Alternating Current
CONTENT
Introduction
Safety Notice
Parts&Figures
Feature&Advantage
Instruction of Operation-Obturation Pen-Obturation Gun-Battery&Charger
Cleaning &Maintenance
Application of Gutta percha Obturation with C-FILL-Precautions-Down Pack with Obturation Pen-Downfill with Obturation Pen-Downfill with ObturationGun
Trouble & shooting
Technical
WarrantyPart List
Functional Nomenclature-Obturation Pen-Obturation Gun
1
1
1~2
3~4
5~6
7~8
9~1011~12
13~14
15
16
18~19
2021
22
23
23
24
17
Introduction
FILL
WARNING
Failure to follow warning instructions could result in serious physical injury or less of life.
Do not use loose or broken plug .It may be cause of fire or electrc shock.
Do not touch electric parts with wet hand.
Do not use multi plug in one outlet, which may cause of fire due to overheating.
Do not leave power cord and/or battery near of heating sources.Or it may be caused of fire or explosion and electric shock due to melting and coming out power cord.
Do not contact battery teminal plate with conductor.It may be caused explosion due to short circuit orgetting shock.
This device is medical appliance.Dentist ues only .
Ascertain voltage to be correct.Or it may cause of property damage.
When you disconnect plug. pull with a plug head nota cord.Or it may be occurred of snapping of a wire.
Be careful of heating injury from head injection partand pen tip.
Do not to disassemble.Contact maintenance center when it be necessary to be repair.
Make device to be cleaned in all time.It will give reduce failure and extend its life.
Do not drop or shock.Or it may be caused of failure.
Disconnect plug from outlet when it thunder lighting.
Keep away from children to avoid injury from getting pricked form sharp devices.
Connect power plug to outlet firmly not to be loose which may cause of fire.
Safety NoticeThis notice will provide you with thorough introduction to useour C-FILL in properly in order to prevent you from certain injury or damage.
FEATURE&ADVANTAGE
1.Exquisite and ergonomic design,easy to operate.2. Fantastic combination of filling method of vertical compaction and backfill.3. Excellent hermetic sealing with 3-dimensional filling (Total obturation of the entire canal including lateral canals)4. Easy and quick obturation with accurate and predictable filling.5. Digitizing temperature display contributes to convenient setting of temperaturel.
1 2
Parts&Figures
Obturation Pen
Obturation Gun
ChargerBattery(3EA)
Heat Insulator(5PCS)
Gun Needle(23,25 Gauge)
Pen Tip(F,FM)
Stand
(Pen)
(Gun)
Accessory
Multi-tools
Brush
Plunger head
Adapter Charging cord
43
Cleaning drill
Cleaning kit
Driver
注射针密封圈(4个)
Pen Tip------------------------------
Operating switch--------------------
Power/Select switch----------
operating signal lamp------------
---------------------- Battery
----------------- Charging jack
---------- Charging jack
----
--------- Battery cover
Functional Nomenclature(Obturation Pen)
1
2
4
3
5
6
76
Pen tip
7
5
4
3
2
1
The Pen Tip is designs for down packing with warmed Gutta Perchain the canal. Also,it can be used to cut off excess Gutta Percha or complete root canal filling by repeated use.
It can be operated in every direction.When operating,the signal lamp is wavering with sounds.
Operating signal lamp When press the operating switch,the Pen Tip is heated and the operating signal lamp is wavering.
Power/Select switchUse for turning on/off the power.Slect the operating accordance withthe size of Pen Tip.
BatteryThe battery is used commonly for Obturation Pen and Gun.Also battery can be recha geable just as it is built in with
Charging jackIt help to charge battery without disconnecting from device.
Battery coverIt will protect battery from dust and occasional disconnect battery from device.
Operating switch
charging cord.
6
Don' t use the Obturation pen when charging with the charging cord.
!CAUTION
5 6
Small Size(XF,F,FM)-Green LampBig size(M,ML)-Red Lamp
6 0
------------ BatteryCharging jack
-------- Battery cover
Temperature display
Plunger----------- Trigger
Heat insulator–--
Needle----------------------------------------
Plunger lever-----------
Plunger head-------------------
Gutta Percha slot------------
------------
----------
-------------
-----------
Temperature control button---------------------–-
Power swich--------------------–--
1
2
3
1211
13
7
10
6
5
4
9
8
NeedleInject melting Gutta Percha Bar from needle.(Provided 23,25 gauge each 3pieces)
Heat InsulatorIt is prevent from heat injury during Gun operation. Should be capping just before operation.
Temperature displayIt shows operating and seting temperature.
Gutta Percha slotGutta percha should be input only one piece in a one time.
Power switch Use for turning on/off the power.
Temperature control buttonTo control temperature of Gun between 50°C to 200 C.
Plunger leverTo be used for pulling/pushing a plunger.
Plunger headIt made with silicon material prevent backward flow of Gutta Percha.
PlungerTo inject melted Gutta Percha squeeze trigger makes move plunger 2mm forward.
Trigger It works for injecting melted Gutta Percha thru needle.
BatteryThe battery is used commonly for Obturation Pen and Gun.Also battery can be rechargeable just as it is built in with charging cord.
Charging jackIt help to charge battery without disconnecting from device.
Battery coverIt will protect battery from dust and occasional disconnect battery from device.
°
------------------------------------------------------
1
2
3
4
5
6
7
8
910
11
12
13
Functional Nomenclature(Obturation Gun)
7 8
The installation of iniection needle’s seal ring
Use the forceps to put the seal ring into the hole of the injectionneedle, and then install the injection needle to the fill gun.
Notice: put the concave surface of the seal ring to the outside.
Instruction of Operation(Obturation Pen)
Mounting the Pen Tip
Before using the Pen,mount thePen Tip to the Obturation Pen holder.
Operating switchIt can be operated in every direction .When operating,the signal lamp iswavering with sounds.
Turn on the power
Press the power/Select switchand select the operating accordancewith the mounted Pen Tip.
Turn off the power
After operating,press the powerswitch longer for turning off.
1 2
3 4
Obturation Pen Battery
Open the battery cover Change the battery
Take out the battery as the picture and change the battery.Charging jack must be outside way.
Charging jack
Close the battery coverAdapt the battery cover groove
cover switch
Stuck the switch,then push it tofrontward,when snip-snap sound showalready closed shown as the picture.
Hold the switch,and drag it to backward,then take down the battry cover .shown as the picture
After changing the new battery,close the battery cover.
9 10
After changing the new battery,close the battery cover.Fit an inside prominence of battery cover to depression point on a device shown as picture’s arrow mark.
Press one timeturn on the power
Press one more timeto select the pen tip typeSmall size,Green lampBig size,red lamp
Press longerturn off the power,turn off the lamp
(make sure that you have mounted the pen tip and effectiveness,or it can not be operated.)
Instruction of Operation(Obturation Gun)
Put on a Needle as shown in the picture.
Put on Gun Needle
Put on Gun Needle
Tightly fasten
Tightly fasten the Needle with multitool provided.
Put heat insulator into the arrowdirection.
CAUTION You must put heatinsulator on Obturation Gun just before operation as it works at high temperature.There is in danger of burning.
!
Bend Gun Needle
Bend the needle as picture shownin relation to the tooth being treated.
1 2
3 4
Pull out the plunger
Pull the plunger out as picture shown by turning the release lever left.
Gutta Percha slot open
Insert Gutta Percha Bar
Insert Gutta Percha Bar intothe loading chamber as picture shown.
5 6
7
!CAUTION
Push out the plunger
Press the power button
Press the power button until the temperature be shown.
+-
Pull a trigger slowly
It takes about three minutes to reach the adjusted tempeature.When pulling a trigger slowly after 3 minutes,the warmed Gutta percha flow out.
Adjust temperatureAdjust temperature with the tempera-ture control button.
The plunger goes forward per 2mm when the trigger is pulled.
You can adjust the temperature up to 200°C.“ ”:Low temperature“ ”:High temperature
Before you remove the used needle, please take the left Gutta Percha out completely.After then, remove the needle with multi tool provided. Please replace the plunger when it worn out or damaged as it is a disposable item.
Don’t try to operate when Gun is not properly heated.The device can be damaged.Please use new needle and heat insulator in every application.Patient can be cross contamination.
!CAUTION
8
109
11
11 12
Push the plunger into the heating chamber after inserting Gutta Percha Bar.
Gutta Percha Bar is inserted once a one.
Instruction of Operation(Obturation Gun Battery)
Open battery cover
Please replace the battery withspare one when it discharged.
Please cover switch press down and pull battery cover.
Replace battery
Replace the battery with new one,as picture shown and the battery ofcharging jack must be shown.
–-----Charging jack
Adapt the battery cover groove
After changing the new battery,close the battery cover.Fit an inside prominence of battery cover to depression point on a device shown as picture’s arrow mark.
Close the battery over
Charge(Charging Battery)
Connect the power.It is graded with DC5V,2.4A.And Li-ion battery mounts protective cri cuit with capacity of DC3.7V,2,200mAh and used for Gunand Pen commonly.
Charger and Battery
!CAUTIONPlease use main adapter only.It can be in danger of exploring and fire when you don’t use the main adapter.
It can be in danger of exploring and fire when it hit or shock by power.
Charging Battery
Put the battery on the charger.When the battery is inserted into charger,the lamp indicates red sign.And then,when the battery is fully charged,the lamp indicates green sign.
!CAUTIONPlease insert the battery properly intocharger to connect with charging jack.Please don’t connect with PC USB jack.There are in danger of serious damage ofPC and exploring of battery.
When fully charged,thecharging lamp turns intogreen light.You can charge two batteries at once.
When the plug in the power, the powerindicator lamp turns into green light.
charging lamp
Power indicator lamp
ChangingCord
Connect charging cord with charger and then,charge directly by connecting charging jack with Pen.
Take out the discharged battery of Pen and then,charge it by connecting with charger
ChangingCode
Charging Obturation Pen,Gun
Don’tconnect with PC USB Jack.There are in danger of serious damage of PC and exploring of battery.Please don’t use Pen or Gun when it is charged with charging code.
!CAUTION
1 2
3 4
21
3
13 14
Cleaning and Maintenance(Obturation Gun)
Pull the plunger out Pull the plunger out as picture shown by turning the release lever left.
Get move Needle
Ware out heat insulator.Using Mutl-tool for screw out of Gun from header position of Obturation Gun.
Take out battery Before cleansing work,remove thebattery from inner side of Obturation Gun.
It may be caused heating injury byturn on the power.
!CAUTION
Brush soaked into the alcoholic
Clean the cylinder using the brushssoaked into alcoholic.
Clean the cylinder using the brushesand the cleaning drill,soaked into thealcoholic again.Repeat there procedure until clean.
Turn the power off,Let the gun cool before cleaning it.Take care of burn.Maybe the reasin of doesn’t off.
Change of plunger headPlunger head is consumable part,whichis changeable when back-flow or blocking of Gutta percha.
It is recommended that the gun should be cleaned after every week use to prevent bulid-up of excess gutta percha.
!CAUTION
1 2
3 4
Application of Gutta percha Obturation with C-FILL(Precaution)
Pen tip,Gun Needle and heat insulatormust be sterilized before applicationwith autoclave unit.
Sterilization cycle is recommended at 121°Cfor 30minutes or 135°C for 10 minutes.
1 2
3 Once the pen is operated,according to its high-tension rubber switch,special care is needed to avoid any accidents by burning under controling.
The spring switch must be in turn off and be stay in the stand supplied specially.
15 16
Heating up obturation Gun after turn on thepower switch when the release lever left.
!CAUTION
Prepare fully charged batttery before treatment.After obturation charge it for the next.
5 6
Obturation Pen
Obturation Pen is designed for down packing with warmed Gutta perchain the canal.Also,it can be used to cut off excess Gutta Percha Point or complete root canal filling by repeated use.
Select the appropriate Pen Tip and insert into holder of the Pen.Available Pen Tip Tip sizes provided are XF,F,FM,M,and ML.
Obturation Gun
Obturation Gun is designed to inject warmed Gutta Percha that
is specially formulated into the root canal directly.It provides easy
and fast filling ability,but it is recommended to fill the apical part of
the canal with the pen first to avoid any overfilling or under filling and
the gun can be used for filling the remained part of the canal.
Down pack with Obturation Pen
Preparation-Gutta PerchaPoint
Select same size Gutta Percha Point with previous used file,If 06 Taper #30master apical file is used,select 06 Taper #30 Gutta Percha Point.
F FM M MLFM M
F FMM
MLFM
M
Preparation-Pen Tip
Select proper Pen Tip for preparedroot canel.
Dry root canal
Dry root canal using paper point.
Prepare 2~3plugger to compactGutta Percha,mark root canal lengthby rubber stopper.
Confirmation of tug back
Fit a Gutta Percha Point that matchesprepared root cannal to check it makesa tug back.
Length mark to the Pen Tip
Mark length on the selected Pen Tip,4~6mm short of working lengthby rubber stopper.
Preparation-Hand plugger1 2
3 4
5 6
17 18
Backfill with Obturation Pen
Heating & insertion
Heat and insert Pen Tip with GuttaPercha to 4~6mm short of theworking length.
Insertion of Gutta Percha Point
Coat 3mm from the tip with sealer,insert it to root canal with pumpingaction.
Removal of excess GuttaPercha
8
Cut the excess Gutta Percha by heated Pen Tip.
9 10 Compaction
Stop the heating and maintainapical pressure about 10 secondsfor prevention of dead space.
CompactRemoval of Pen Tip
Removal after heating the Pen Tip for 1~2 seconds to avoid Gutta Percha out with pen tip.
Compact the warmed Gutta Percha with the plugger.
Confirm obturation with a radiograph
11 12
7
Select the same size Gutta PerchaPoint used on vertical compact,cut 2~3mm the tip and insert into the rootcanal.
Insertion of gutta percha
Heating & Insertion
Heating and Insertion Pen Tip withGutta Percha to half of the working length.
Removal of excess Gutta Percha
Cut the excess Gutta Percha by heated Pen Tip.
Confirm obturation with a radiograph
Compact Gutta Percha with big plugger.
Compaction
1 2
3 4
19 20
Backfill with Obturation Gun
Gun Needle insertion Soften and Injection
Inser gun needle as it will go intothe canal space without binding.
Pull the trigger slowly and fill the guttapercha up to root canal.Gun needlewould be pushed by filled gutta percha naturally.
Compact Gutta Percha with big plugger,Harden the heated gutta percha withbigger plugger.
Compaction
Confirm obturation with a radiograph
Now the root canal is fullyobturation.
1 2
3 4 Complete filling
Trouble & shooting
When Gun,Pen dose not works
When Pen dose not work evenits operation button is pushed.
Push power button one time.Check battery is equipped firmly.Replace battery with full charged.
When Gutta Percha dose not inject during squeezing trigger.
When Gun temperature does not increase.
Check power on.(please recheckoperating light on)Replace Pen Tip with new one in case of tip failure.
Check power on.Check temperature display shows over 160°C.Check Gutta Percha bar exist in.Replace needle with new one in case of needle .failure
Check temperature control buttonto increase temperature in case of low setting.Replace battery with full charged.
Battery Do not dump an over-cycle battery with garbage.
or it cause environmental pollution.It is recommends
to waste in separate.
Pen Tip Do not dump with garbage.Or it cause environmental
pollution.It is recommends to waste in separate.
Do not dump with garbage.Or it cause environmental
pollution.It is recommends to waste in separate.
Needle
2221
Technical
Obturation Gun
Obturation Pen
Battery
Adapter
Charger
Operating Voltage
Heating time
Working temperature
Operating Time
Operating Voltage
Heating time
Working temperature
Operating Time
DC3.7V
3 minutes
50°C~200°C
Continuous for 1.5 Hours
DC3.7V
2 seconds
160°C
Continuous for 1.5 Hours
Li-ion battery
Charging Time
ICR18650-26(samsungSDI)3.7V,2,200mAh
Within 3 hours
Input
Output
100V~240VAC
DC5V,2.4A
Input DC5V
Part List
Obturation Gun
Obturation Pen
Charger
Adapter
Battery
Charging cord
Holder(Pen and Gun stand)
Cleaning kit(brush)
Mutil-tools
Heat insulator
Gun needle(23ga and )25ga
Pen Tip(F/FM)
Plunger head
1Set
1Set
1Set
6EA/1BOX
1EA
2EA
5EA
3EA
2EA
1EA
1EA
Warranty
COXO warrants the mechanical and electronic components of C-FILL
to be free of defects in materials and workmanship for period of one year
from the original date of purchase,in accordance with the COXO’s warranty
regulations.
If the C-FILL shows any defect within the specified warranty period that are
not excluded from this warranty.COXO shall,at his discretion,either replace
or repair C-FILL using suitable new or reconditioned parts.
In the case that other parts are used which constitute an improvement,
COXO may,at its discretion,charge the customer for the additional cost of
these parts.
In the warranty claim provides to be justified,the product will be returned to
the user freight prepaid.
Warranty claims other than those indicated above are expressly excluded.
Exclude warranty regulations
Warranty
Damages and defects caused by the following conditions are not covered by
this waranty:
1.Improper handing/disassemble/modifying,neglect or failure to operate
the unit in compliance with the instructions given in user or service manuals.
2.Damages and defects caused by force majeure or any other condition
that is the control of COXO company.
3.C-FILL whitening unit is guaranteed for 12 months from the date of
purchase.Theaccessories(Heat insulator,Pen tip,Gun needle,Gutta Percha,
Gutta Percha Point,Cleaning kit,Multi-tools, etc.)Are not covered by
warranty.
2423
Driver 1EA
Cleaning kit( )cleaning drill 1EA
1EA
2EA/1BOX