Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
The Development of Smart Sensors for
Aquatic Water Quality Monitoring
A thesis submitted to the University of Manchester for the degree of
Doctor of Philosophy in the Faculty of Engineering and Physical
Sciences
2014
CRAIG ALEXANDER
SCHOOL OF CHEMICAL ENGINEERING AND ANALYTICAL
SCIENCE
Craig Alexander
2
Contents
List of Figures ..........................................................................................................................................9
List of Tables..........................................................................................................................................16
Abstract ..................................................................................................................................................18
Declaration .............................................................................................................................................19
Copyright Statement ...............................................................................................................................19
Acknowledgements ................................................................................................................................20
List of Abbreviations ..............................................................................................................................21
Project Overview .................................................................................................................................. 24
1 Introduction ................................................................................................................................. 25
1.1 Aquarium Chemistry ................................................................................................................... 25
1.1.1 Analytes from the Nitrogen Cycle ................................................................................................ 26
1.1.1.1. Ammonia .................................................................................................................... 26
1.1.1.2. Nitrite .......................................................................................................................... 28
1.1.1.3. Nitrate ......................................................................................................................... 28
1.1.2 Aquarium pH ............................................................................................................................... 30
1.1.3 Chlorine .................................................................................................................................... 30
1.1.4 Dissolved Oxygen ....................................................................................................................... 31
1.1.5 Phosphate .................................................................................................................................... 32
1.1.6 Water Hardness ........................................................................................................................... 32
1.1.6.1. General Hardness ........................................................................................................ 32
1.1.6.2. Carbonate Hardness ..................................................................................................... 33
1.1.7 Summary of Parameters ............................................................................................................... 34
1.2 Current Methods of Analysis ....................................................................................................... 35
1.2.1 Test Kits .................................................................................................................................... 35
1.2.1.1. pH ............................................................................................................................... 36
1.2.1.2. Ammonia .................................................................................................................... 37
1.2.1.2.1. Nessler’s Reagent Method ........................................................................................... 37
1.2.1.2.2. Salicylate Method ........................................................................................................ 37
1.2.1.3. Nitrite .......................................................................................................................... 38
Craig Alexander
3
1.2.1.4. Nitrate ......................................................................................................................... 39
1.2.2 Commercial Sensors .................................................................................................................... 40
1.2.3 Benefits of the Proposed Device .................................................................................................. 40
1.3 Introduction to Chemical Sensors ................................................................................................ 42
1.3.1 Electrochemical Sensors .............................................................................................................. 43
1.3.1.1. Potentiometry .............................................................................................................. 43
1.3.1.2. Reference Electrodes ................................................................................................... 45
1.4 Ion-Selective Electrodes .............................................................................................................. 48
1.4.1 Selective Polymeric Membranes .................................................................................................. 48
1.4.2 Selectivity of ISEs ....................................................................................................................... 50
1.4.3 Polymer Membrane Electrodes .................................................................................................... 55
1.4.4 Miniaturisation of ISEs ................................................................................................................ 56
1.4.4.1. Coated-Wire Electrodes ............................................................................................... 56
1.4.4.2. Screen-Printed ISEs ..................................................................................................... 57
1.4.5 Sol-gels .................................................................................................................................... 59
1.4.5.1. Sol-gel Chemistry ........................................................................................................ 59
1.4.5.2. Sol-gel Ion-Selective Membranes ................................................................................ 61
1.4.6 ISEs for Aquarium-Significant Analytes ...................................................................................... 63
1.4.6.1. pH ............................................................................................................................... 63
1.4.6.2. Ammonia/Ammonium ................................................................................................. 65
1.4.6.3. Nitrite .......................................................................................................................... 67
1.4.6.4. Nitrate ......................................................................................................................... 69
1.5 Impedimetric Interdigitated Electrode Chemical Sensors ............................................................. 72
1.5.1 Electrical Impedance ................................................................................................................... 72
1.5.2 Interdigitated Electrodes .............................................................................................................. 74
1.5.3 IDEs for Chemical Sensing Applications ..................................................................................... 75
1.5.3.1. Ion-Selective Conductometric Microsensors ................................................................ 75
1.5.4 IDE Fabrication Techniques ........................................................................................................ 76
1.5.4.1. Photolithography ......................................................................................................... 77
1.5.4.2. Screen-Printing............................................................................................................ 79
Craig Alexander
4
1.5.5 Membrane Coating Techniques.................................................................................................... 80
1.5.5.1. Spin-Coating ............................................................................................................... 80
1.5.5.2. Dip-Coating ................................................................................................................ 81
1.5.5.3. Drop-Coating .............................................................................................................. 81
1.5.5.4. Screen-Printing............................................................................................................ 81
1.6 Aims and Objectives .................................................................................................................... 82
2 Materials and Methods................................................................................................................. 83
2.1 Preparation of Ion-Selective Membranes...................................................................................... 83
2.1.1 Materials .................................................................................................................................... 83
2.1.2 Polyvinyl Chloride Membranes.................................................................................................... 84
2.1.3 Sol-Gel Membranes ..................................................................................................................... 85
2.1.3.1. MTES ......................................................................................................................... 85
2.1.3.2. DEDMS/TEOS............................................................................................................ 85
2.1.4 Dielectric Screen-Printing Paste Membrane ................................................................................. 86
2.2 Construction of Ion-Selective Electrodes ..................................................................................... 87
2.2.1 Materials .................................................................................................................................... 87
2.2.2 Polymer Membrane Electrodes .................................................................................................... 87
2.2.3 Coated-Wire Electrodes ............................................................................................................... 88
2.3 Potentiometric Measurements ...................................................................................................... 89
2.3.1 Measurement Instrumentation ...................................................................................................... 89
2.3.2 Sensor Testing ............................................................................................................................. 89
2.3.2.1. Materials ..................................................................................................................... 89
2.3.2.2. Nitrate ISEs ................................................................................................................. 90
2.3.2.2.1. Stock Solutions ........................................................................................................... 90
2.3.2.2.2. Calibrations ................................................................................................................. 90
2.3.2.2.3. Selectivity Determination ............................................................................................ 91
2.3.2.3. Ammonium ISEs ......................................................................................................... 92
2.3.2.3.1. Stock Solutions ........................................................................................................... 92
2.3.2.3.2. Calibrations ................................................................................................................. 92
2.3.2.3.3. Selectivity Determination ............................................................................................ 93
Craig Alexander
5
2.3.2.4. pH ISEs ....................................................................................................................... 94
2.3.2.4.1. Stock Solutions ........................................................................................................... 94
2.3.2.4.2. Calibrations ................................................................................................................. 94
2.3.2.4.3. Selectivity Determination ............................................................................................ 95
2.3.2.5. Nitrite ISEs ................................................................................................................. 95
2.4 Fabrication of Interdigitated Electrodes ....................................................................................... 96
2.4.1 Lift-off Photolithography ............................................................................................................. 96
2.4.1.1. Materials ..................................................................................................................... 96
2.4.1.2. Sensor Design ............................................................................................................. 97
2.4.1.3. Fabrication Process ..................................................................................................... 98
2.4.1.4. Reduced Geometry IDEs ........................................................................................... 100
2.4.2 Screen-Printing .......................................................................................................................... 101
2.4.2.1. Materials ................................................................................................................... 101
2.4.2.2. Sensor Design ........................................................................................................... 102
2.4.2.3. Fabrication Process ................................................................................................... 104
2.4.2.3.1. Attempt One .............................................................................................................. 104
2.4.2.3.2. SP IDE Design 1 ....................................................................................................... 104
2.5 Construction of Ion-Selective Impedimetric Microsensors ......................................................... 106
2.5.1 Photolithographically-prepared IDE Sensors .............................................................................. 106
2.5.1.1. IDE Design 1............................................................................................................. 106
2.5.1.2. IDE Design 2............................................................................................................. 107
2.5.2 Screen-Printed IDEs .................................................................................................................. 107
2.5.2.1. SP IDE Design 1 ....................................................................................................... 107
2.6 Impedance Measurements .......................................................................................................... 108
2.6.1 Measurement Instrumentation .................................................................................................... 108
2.6.2 Sensor Testing ........................................................................................................................... 109
2.7 Ion-Selective Impedimetric Microsensors Methodology ............................................................ 110
2.7.1 Materials .................................................................................................................................. 110
2.7.2 Nitrate Sensors .......................................................................................................................... 110
2.7.2.1. Stock Solutions ......................................................................................................... 110
Craig Alexander
6
2.7.2.2. Pure Water Calibrations............................................................................................. 111
2.7.2.3. Background Effects ................................................................................................... 112
2.7.2.4. Selectivity ................................................................................................................. 112
2.7.2.5. Reproducibility.......................................................................................................... 114
Ammonium Sensors............................................................................................................................ 114
2.7.2.6. Stock Solutions ......................................................................................................... 114
2.7.2.7. Measurements in Water ............................................................................................. 114
2.7.2.8. Selectivity ................................................................................................................. 114
2.7.3 pH Sensors ................................................................................................................................ 115
2.7.3.1. Stock Solutions ......................................................................................................... 115
2.7.3.2. Calibrations ............................................................................................................... 116
2.7.4 Nitrite Sensors ........................................................................................................................... 117
2.7.4.1. Stock Solutions ......................................................................................................... 117
2.7.4.2. Measurements in Water ............................................................................................. 117
2.7.4.3. Selectivity ................................................................................................................. 117
3 Potentiometric Characterisation of Ionophores ........................................................................... 119
3.1 Nitrate-Selective Electrodes ....................................................................................................... 119
3.1.1 Nitrate PMEs ............................................................................................................................. 119
3.1.2 Nitrate CWEs ............................................................................................................................ 124
3.2 Ammonium-Selective Electrodes ............................................................................................... 129
3.3 Hydrogen Ion-Selective Electrodes ............................................................................................ 133
3.4 Conclusions for Chapter 3.......................................................................................................... 136
4 Nitrate Sensing using IDE Design 1 .......................................................................................... 140
4.1 Sensor Calibration ..................................................................................................................... 140
4.2 PVC Membrane Sensors ............................................................................................................ 147
4.2.1 Calibrations ............................................................................................................................... 147
4.2.2 Selectivity.................................................................................................................................. 155
4.3 Sol-gel Membrane Sensors ........................................................................................................ 157
4.3.1.1. DEDMS/TEOS Membrane ........................................................................................ 157
4.3.1.1.1. Calibrations ............................................................................................................... 157
Craig Alexander
7
4.3.1.1.2. Selectivity ................................................................................................................. 167
4.4 Conclusions for Chapter 4.......................................................................................................... 170
5 Nitrate Sensing using IDE Design 2 ........................................................................................... 174
5.1 Sensor Calibration ..................................................................................................................... 174
5.2 PVC Membrane Sensors ............................................................................................................ 177
5.2.1.1. Spin-coated Membranes ............................................................................................ 177
5.2.1.1.1. Calibrations ............................................................................................................... 177
5.2.1.1.2. Selectivity ................................................................................................................. 182
5.2.1.2. Drop-coated Membranes ........................................................................................... 183
5.2.1.2.1. Calibrations ............................................................................................................... 184
5.2.1.2.2. Selectivity ................................................................................................................. 202
5.3 Conclusions for Chapter 5.......................................................................................................... 207
6 Nitrate Sensing using SP IDE Design 1...................................................................................... 211
6.1 Issues Experienced with Initial Sensor Designs .......................................................................... 211
6.2 SP IDE Design 1 Calibration ..................................................................................................... 212
6.3 SP IDE Design 1 Selectivity ...................................................................................................... 224
6.4 Conclusions for Chapter 6.......................................................................................................... 227
7 pH, Ammonium and Nitrite Sensing using IDE Design 2 ........................................................... 230
7.1 Ammonium Sensors .................................................................................................................. 230
7.2 Nitrite Sensors ........................................................................................................................... 234
7.3 pH Sensors ................................................................................................................................ 238
7.4 Conclusions for Chapter 7.......................................................................................................... 244
8 Conclusions and Suggestions for Further Work ......................................................................... 246
8.1 Conclusions ............................................................................................................................... 246
8.2 Future Work .............................................................................................................................. 250
8.2.1 Individual Sensor Testing .......................................................................................................... 250
8.2.2 Membrane Characterisation ....................................................................................................... 252
8.2.3 Membrane Materials .................................................................................................................. 252
8.2.4 IDE Geometry ........................................................................................................................... 253
8.2.5 Screen-Printed Sensors .............................................................................................................. 254
Craig Alexander
8
8.2.6 Considerations for a Prototype Device ....................................................................................... 254
8.2.7 Sensor Longevity ....................................................................................................................... 255
8.2.8 Data Display .............................................................................................................................. 256
9 References ................................................................................................................................. 257
10 Appendices ................................................................................................................................ 265
10.1 Molar Equivalent Concentrations of Ions ................................................................................... 265
10.2 Details of Commercial Ionophores Used .................................................................................... 266
10.3 Potentiometric Data Acquisition System .................................................................................... 267
Final word count: 56,647
Craig Alexander
9
List of Figures
1.1 Algae accumulation caused by high levels of nitrate in an aquarium ........................................... 29
1.2 Structure of the two forms of phenol red which result in the observed colour change ................. .36
1.3 Structure of three forms of bromothymol blue which result in a colour change at varying pH
values…... ................................................................................................................................... 36
1.4 Structure of indosalicylate. .......................................................................................................... 38
1.5 Schematic representation of a double-junction Ag/AgCl reference electrode with
1 M LiCH3COO as outer filling solution. ..................................................................................... 47
1.6 (A) Potassium tetrakis-[3,5,-bis-(trifluoromethyl)phenyl]borate – used as an ionic additive in
cation-selective membranes; (B) tridodecylmethyl ammonium chloride (TDMACl)– used as an
ionic additive in anion-selective membranes ................................................................................ 49
1.7 Structures of three commonly-used plasticisers within ISEs (A) dioctyl sebacate (DOS),
(B) o-nitrophenyloctyl ether (NPOE), (C) dioctyl phthalate (DOP).. ............................................ 50
1.8 Example calibration curve for calculating the selectivity coefficent using the FIM ....................... 52
1.9 Schematic diagram of a typical PME set-up, showing the cell potential of an ISE being
measured against a SCE reference ............................................................................................... 55
1.10 Structures of three typical silicon alkoxide sol-gel precursors – (A) methyltriethoxysilane (MTES);
(B) tetraethoxysilane (TEOS); (C) diethoxydimethylsilane (DEDMS) ......................................... 60
1.11 Selection of the Hofmeister series showing the order of anion hydrophobicity ............................. 67
1.12 A typical AC sine wave voltage and current. ................................................................................ 72
1.13 Schematic diagram of a typical IDE design. ................................................................................. 74
1.14 Cross-section schematic representation of the electric field penetration depth from a
typical IDE when an AC voltage is applied. ................................................................................. 75
1.15 Schematic representation of an ISCOM ....................................................................................... 76
1.16 Photolithographic preparation of thin-film electrodes using (a) chemical etching and
(b) lift-off procedures .................................................................................................................. 78
1.17 Schematic diagram of screen-printing for the fabrication of thick-film electrochemical sensors. .. 79
2.1 Schematic representation of IDE Design 1 ................................................................................... 98
2.2 Microscope image (4x) of the digits of IDE Design 1, fabricated in-house using a mask-less,
lift-off photolithographic technique followed by e-beam deposition of gold metal ..................... 100
2.3 IDE Design 2 ............................................................................................................................. 101
2.4 Schematic representation of the screens used to produce 6 different IDE designs. ...................... 103
2.5 Schematic representation of SP IDE Design 1 ............................................................................ 104
2.6 Outline of the screen-printing fabrication process of SP IDE Design 1 ....................................... 105
2.7 Microscope image (4x) of the electrode digits of SP IDE Design 1 ............................................ 106
Craig Alexander
10
2.8 Front panel of the LabVIEW program used to control the Agilent E4980A LCR meter .............. 108
2.9 Schematic diagram of the experimental set-up for testing IDE ion sensors using the Agilent
E4980A LCR meter. .................................................................................................................. 110
3.1 Calibration graph of two nitrate-selective PMEs prepared using a PVC membrane containing
1% w/w TDAN as the ionophore. .............................................................................................. 119
3.2 Calibration graph of two nitrate-selective PMEs prepared using a PVC membrane containing
1% w/w NO3V as the ionophore ................................................................................................ 121
3.3 Selectivity determination of TDAN PME 1 using the FIM in 1000 ppm Cl- and
1000 ppm NO2-. ........................................................................................................................ .122
3.4 Selectivity determination of NO3V PME 1 using the FIM in 1000 ppm Cl- and
1000 ppm NO2-. ......................................................................................................................... 123
3.5 Calibration graph of two nitrate-selective CWEs prepared by dip-coating the electrode into a
PVC membrane ‘cocktail’ containing 1% w/w TDAN as the ionophore ..................................... 124
3.6 Calibration graph of four nitrate-selective sol-gel CWEs (two MTES and two DEDMS)
containing TDAN as the ionophore ............................................................................................ 125
3.7 Selectivity determination of TDAN PVC CWE 1 using the FIM in 1000 ppm Cl- and
1000 ppm NO2- .......................................................................................................................... 126
3.8 Selectivity determination of TDAN MTES CWE 1 using the FIM in 1000 ppm Cl- and
1000 ppm NO2- .......................................................................................................................... 127
3.9 Selectivity determination of TDAN DEDMS CWE 1 using the FIM in 1000 ppm Cl- and
1000 ppm NO2-. ......................................................................................................................... 128
3.10 Calibration graph of two ammonium-selective PMEs prepared using a PVC membrane
containing 1% w/w NH4I as the ionophore ................................................................................ 129
3.11 Selectivity determination of NH4I PME 1 using the FIM in 1000 ppm of the interfering cations
sodium, potassium, magnesium and calcium .............................................................................. 130
3.12 Calibration graph of two ammonium-selective PMEs prepared using a PVC membrane
containing 1% w/w NH4I as the ionophore, with the ionic additive omitted ............................... 131
3.13 Selectivity determination of NH4I PME 1 (ionic additive omitted) using the FIM in 1000 ppm
of the interfering cations sodium, potassium, magnesium and calcium.. ..................................... 132
3.14 Calibration graph of two hydrogen ion-selective PMEs prepared using a PVC membrane
containing 1% w/w HIII as the ionophore.. ................................................................................ 133
3.15 Calibration graph of two hydrogen ion-selective PMEs prepared using a PVC membrane
containing 1% w/w HV as the ionophore. .................................................................................. 134
3.16 Selectivity determination of HIII PME 1 using the FIM in 0.1 M of the interfering cations
sodium, potassium and magnesium.. .......................................................................................... 135
3.17 Selectivity determination of HV PME 1 using the FIM in 0.1 M of the interfering cations
sodium, potassium and magnesium.. .......................................................................................... 136
Craig Alexander
11
4.1 Conductance (A) and capacitance (B) spectra for the addition of NO3- to an uncoated
IDE Design 1 (1 Vrms input amplitude) ....................................................................................... 140
4.2 Conductance response of uncoated IDE Design 1 upon addition of 1–100 ppm nitrate at
six frequencies.. ......................................................................................................................... 141
4.3 Capacitance response of uncoated IDE Design 1 upon addition of 1–100 ppm nitrate at
six frequencies.. ......................................................................................................................... 142
4.4 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3- concentration for
uncoated IDE Design 1 at six frequencies.. ................................................................................ 143
4.5 Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3- concentration for
uncoated IDE Design 1 at six frequencies.. ................................................................................ 144
4.6 Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate concentration
for uncoated IDE Design 1 at six frequencies.. ........................................................................... 145
4.7 Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate concentration
for uncoated IDE Design 1 at three frequencies ......................................................................... 146
4.8 Conductance (A) and capacitance (B) spectra for the addition of NO3- to TDAN PVC spin-coated
IDE Design 1 (1 Vrms input amplitude).. ..................................................................................... 147
4.9 Conductance response of TDAN PVC spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.. ........................................................................................ 148
4.10 Capacitance response of TDAN PVC spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.. ........................................................................................ 149
4.11 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3- concentration
for TDAN PVC spin-coated IDE Design 1 at 2 MHz. ................................................................ 150
4.12 Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for TDAN PVC spin-coated IDE Design 1 at 2 MHz. .......................................... 150
4.13 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3-
concentration for TDAN PVC spin-coated IDE Design 1 at 2 MHz.. ......................................... 151
4.14 Measured baseline conductance value (G0) of TDAN PVC spin-coated IDE Design 1 in six
separate sample matrices at 2 MHz, prior to the addition of nitrate............................................. 153
4.15 Measured conductance response (Gm) of TDAN PVC spin-coated IDE Design 1 against
log NO3- concentration in six separate sample matrices at 2 MHz .............................................. 154
4.16 Baseline-subtracted conductance (Gm – G0) of TDAN PVC spin-coated IDE Design 1 against
log NO3- concentration at 2 MHz, in six separate sample matrices. ............................................ 154
4.17 Deionised water baseline-subtracted conductance (Gm – G0) of three TDAN PVC spin-coated IDE
Design 1 devices against log NO3- concentration at 2 MHz, for reproducibility determination.... 155
4.18 Selectivity determination of TDAN PVC spin-coated IDE Design 1.. ......................................... 156
4.19 Conductance (A) and capacitance (B) spectra for the addition of NO3- to TDAN DEDMS
spin-coated IDE Design 1 (1 Vrms input amplitude). ................................................................... 158
4.20 Conductance response of TDAN DEDMS spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies. ......................................................................................... 159
Craig Alexander
12
4.21 Capacitance response of TDAN DEDMS spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.. ........................................................................................ 160
4.22 Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate concentration
for TDAN DEDMS spin-coated IDE Design 1.. ......................................................................... 161
4.23 Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate concentration
for TDAN DEDMS spin-coated IDE Design 1 ........................................................................... 161
4.24 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3- concentration for
TDAN DEDMS spin-coated IDE Design 1 ................................................................................ 162
4.25 Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3- concentration for
TDAN DEDMS spin-coated IDE Design 1 ................................................................................ 163
4.26 Measured baseline conductance value (G0) of TDAN DEDMS spin-coated IDE Design 1 in six
separate sample matrices prior to the addition of nitrate. ............................................................ 165
4.27 Measured baseline capacitance value (C0) of TDAN DEDMS spin-coated IDE Design 1 in six
separate sample matrices prior to the addition of nitrate. ............................................................ 166
4.28 Selectivity determination of TDAN DEDMS spin-coated IDE Design 1 (conductive response).. 168
4.29 Selectivity determination of TDAN DEDMS spin-coated IDE Design 1 (capacitive response).. . 169
5.1 Conductance (A) and capacitance (B) spectra for the addition of NO3- to an uncoated
IDE Design 2 (1 Vrms input amplitude) ....................................................................................... 174
5.2 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3- concentration for
uncoated IDE Design 2 at six frequencies and varying input amplitudes. ................................... 175
5.3 Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3- concentration for
uncoated IDE Design 2 at six frequencies and varying input amplitudes .................................... 176
5.4 Conductance (A) and capacitance (B) spectra for the addition of NO3- to TDAN PVC spin-coated
IDE Design 2 (10 mVrms input amplitude). ................................................................................. 178
5.5 Conductance response of TDAN PVC spin-coated IDE Design 2 upon addition of 1–100 ppm
nitrate at six frequencies. ........................................................................................................... 179
5.6 Capacitance response of TDAN PVC spin-coated IDE Design 2 upon addition of 1–100 ppm
nitrate at six frequencies. ........................................................................................................... 180
5.7 Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate concentration
for TDAN PVC spin-coated IDE Design 2 at 2 MHz. ................................................................ 181
5.8 Deionised water baseline-subtracted capacitance (Cm – C0) of three TDAN PVC spin-coated IDE
Design 2 devices against log NO3- concentration, for reproducibility determination. .................. 181
5.9 Selectivity determination of TDAN PVC spin-coated IDE Design 2. ............................................ 182
5.10 Conductance response versus time of TDAN PVC drop-coated IDE Design 2 (10% w/v) upon
addition of 1–100 ppm nitrate at five frequencies. ...................................................................... 184
5.11 Capacitance response versus time of TDAN PVC drop-coated IDE Design 2 (10% w/v) upon
addition of 1–100 ppm nitrate at five frequencies. ...................................................................... 185
5.12 Conductance deionised water baseline settling response of a dry TDAN PVC drop-coated IDE
Design 2 (10% w/v) at five frequencies. ..................................................................................... 186
Craig Alexander
13
5.13 Capacitance deionised water baseline settling response of a dry TDAN PVC drop-coated IDE
Design 2 (10% w/v) at five frequencies. ..................................................................................... 187
5.14 Conductance response of TDAN PVC drop-coated IDE Design 2 (10% w/v) upon the addition
of 50 ppm NO3- to a deionised water background at five frequencies. ........................................ 188
5.15 Capacitance response of TDAN PVC drop-coated IDE Design 2 (10% w/v) upon the addition
of 50 ppm NO3- to a deionised water background at five frequencies. ........................................ 189
5.16 Conductance response of TDAN PVC drop-coated IDE Design 2 (5% w/v) upon the addition
of 50 ppm NO3- to a deionised water background at five frequencies. ........................................ 191
5.17 Capacitance response of TDAN PVC drop-coated IDE Design 2 (5% w/v) upon the addition
of 50 ppm NO3- to a deionised water background at five frequencies. ........................................ 192
5.18 Conductance response of TDAN PVC drop-coated IDE Design 2 (2% w/v) upon the addition
of 50 ppm NO3- to a deionised water background at five frequencies. ........................................ 193
5.19 Capacitance response of TDAN PVC drop-coated IDE Design 2 (2% w/v) upon the addition
of 50 ppm NO3- to a deionised water background at five frequencies. ........................................ 194
5.20 Conductance response versus time of TDAN PVC drop-coated IDE Design 2 (2% w/v) upon
addition of 1–100 ppm nitrate at four frequencies. ..................................................................... 196
5.21 Capacitance response versus time of TDAN PVC drop-coated IDE Design 2 (2% w/v) upon
addition of 1–100 ppm nitrate at four frequencies. ..................................................................... 197
5.22 Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate concentration
for TDAN PVC drop-coated IDE Design 2 (2% w/v) at two frequencies.................................... 198
5.23 Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate concentration for
TDAN PVC drop-coated IDE Design 2 (2% w/v) at 100 kHz. ................................................... 198
5.24 Conductance response versus time of NO3V PVC drop-coated IDE Design 2 (2% w/v) upon
addition of 1–100 ppm nitrate at four frequencies. ..................................................................... 199
5.25 Capacitance response versus time of NO3V PVC drop-coated IDE Design 2 (2% w/v) upon
addition of 1–100 ppm nitrate at four frequencies. ..................................................................... 200
5.26 Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate concentration
for NO3V PVC drop-coated IDE Design 2 (2% w/v) at three frequencies. ................................. 201
5.27 Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate concentration for
NO3V PVC drop-coated IDE Design 2 (2% w/v) at two frequencies. ........................................ 201
5.28 Selectivity determination of TDAN PVC drop-coated IDE Design 2 (2% w/v) using the FIM
(conductive response) ................................................................................................................ 202
5.29 Selectivity determination of TDAN PVC drop-coated IDE Design 2 (2% w/v) at 100 kHz using
the FIM (capacitive response) .................................................................................................... 203
5.30 Selectivity determination of NO3V PVC drop-coated IDE Design 2 (2% w/v) using the FIM
(conductive response). ............................................................................................................... 204
5.31 Selectivity determination of NO3V PVC drop-coated IDE Design 2 (2% w/v) using the FIM
(capacitive response). ................................................................................................................ 205
6.1 Microscope image (4x) of the carbon screen-printed IDE sensors produced at the University of
Bedfordshire.. ............................................................................................................................... 211
Craig Alexander
14
6.2 Conductance (A) and capacitance (B) spectra for the addition of NO3- to an uncoated SP IDE
Design 1 (1 Vrms input amplitude) ............................................................................................. .213
6.3 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3- concentration for
uncoated SP IDE Design 1 at six frequencies and varying input amplitudes. .............................. 214
6.4 Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3- concentration for
uncoated SP IDE Design 1 at six frequencies and varying input amplitudes ............................... 215
6.5 Conductance (A) and capacitance (B) spectra for the addition of NO3- to SP IDE Design 1 with
a layer of plasticised dielectric paste containing 1% w/w TDAN as the ionophore
(10 mVrms input amplitude). ....................................................................................................... 216
6.6 Comparison of the conductance response of SP IDE Design 1 when a membrane containing
1% w/w TDAN was printed over the sensing area with a ‘blank’ membrane which did not
contain any ionophore. ............................................................................................................... 217
6.7 Comparison of the conductance response of SP IDE Design 1 coated with a ‘blank’ membrane
containing no ionophore when zero, one or two ‘build-up’ layers of non-modified dielectric
paste were printed prior to the modified layer containing NPOE as the plasticiser. ..................... 218
6.8 Conductance spectrum for the addition of NO3- to TDAN SP IDE Design 1
(10 mVrms input amplitude).. ...................................................................................................... 219
6.9 Conductance response versus time of TDAN SP IDE Design 1 upon addition of 1–100 ppm
nitrate at four frequencies. ......................................................................................................... 219
6.10 Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate concentration
for TDAN SP IDE Design 1.. ..................................................................................................... 220
6.11 Deionised water baseline-subtracted conductance (Gm – G0) against log NO3- concentration for
TDAN SP IDE Design 1.. .......................................................................................................... 221
6.12 Measured baseline conductance value (G0) of TDAN SP IDE Design 1 in four separate sample
matrices prior to the addition of nitrate. ..................................................................................... 222
6.13 Deionised water baseline-subtracted conductance (Gm – G0) of TDAN SP IDE Design 1 against
log NO3- concentration in four separate sample matrices. ........................................................... 223
6.14 Baseline-subtracted conductance (Gm – G0) of three TDAN SP IDE Design 1 devices against log
NO3- concentration, for reproducibility determination. ............................................................... 224
6.15 Selectivity determination of TDAN SP IDE Design 1 using the FIM.. ........................................ 225
7.1 Conductance response versus time of NH4I PVC drop-coated IDE Design 2 (2% w/v) upon the
addition of 50 ppm NH4+ to a deionised water background at five frequencies. .......................... 231
7.2 Capacitance response versus time of NH4I PVC drop-coated IDE Design 2 (2% w/v) upon the
addition of 50 ppm NH4+ to a deionised water background at five frequencies. .......................... 232
7.3 Conductance response versus time of NO2I PVC drop-coated IDE Design 2 (2% w/v) upon the
addition of 5 ppm NO2- to a deionised water background at five frequencies. ............................. 235
7.4 Capacitance response versus time of NO2I PVC drop-coated IDE Design 2 (2% w/v) upon the
addition of 5 ppm NO2- to a deionised water background at five frequencies. ............................. 236
7.5 Conductance response versus time of HIII PVC drop-coated IDE Design 2 (2% w/v) between
pH 9.04–6.24 at five frequencies. ............................................................................................... 240
Craig Alexander
15
7.6 Capacitance response versus time of HIII PVC drop-coated IDE Design 2 (2% w/v) between
pH 9.04–6.24 at five frequencies. ............................................................................................... 241
7.7 Conductance response versus time of HV PVC drop-coated IDE Design 2 (2% w/v) between
pH 9.04–6.24 at five frequencies. ............................................................................................... 242
7.8 Capacitance response versus time of HV PVC drop-coated IDE Design 2 (2% w/v) between
pH 9.04–6.24 at five frequencies. ............................................................................................... 243
10.1 Calibration graph from a pH electrode obtained using a PHTX-22 pH/ORP preamplifier which was
powered using a linear lab supply at 10 V. .......................................................................................... 268
Craig Alexander
16
List of Tables
1.1 dH scale for water hardness ......................................................................................................... 33
1.2 Summary of the acceptable criteria of each of the most important analytes identified for
acceptable water quality within freshwater aquaria ...................................................................... 34
1.3 Comparison of the features of the proposed sensor device with the options currently available .... 41
1.4 Summary of some of the performance characteristics of H+-selective ionophores used within
pH-ISEs that have been described in the literature ....................................................................... 64
1.5 Summary of some of the performance characteristics of NH4+-selective ionophores..................... 66
1.6 Summary of some of the performance characteristics of NO2--selective ionophores ..................... 68
1.7 Summary of some of the performance characteristics of NO3--selective ionophores ..................... 70
2.1 Composition of each component required to produce PVC ion-selective membranes containing
1% w/w ionophore ....................................................................................................................... 84
2.2 Details of internal filling solutions used within PMEs for each target ion ..................................... 88
2.3 Volume of respective stock solutions required to produce 500 ml nitrate calibration solutions
with fixed 1000 ppm interfering ions, for selectivity coefficient determination ............................ 91
2.4 Volume of respective stock solutions required to produce 500 ml ammonium calibration solutions
with fixed 1000 ppm interfering ions, for selectivity coefficient determination ............................ 93
2.5 Volume of 0.1 M HCl required to add to 50 ml stock solution to produce 250 ml pH buffer
calibration solutions ..................................................................................................................... 94
2.6 Chemical composition of Megaposit SPR-220 positive photoresist .............................................. 96
2.7 Chemical composition of Megaposit MF-26A developer solution ................................................ 96
2.8 Geometric parameters of each IDE design shown in Figure 2.4 .................................................. 103
2.9 Composition of each phosphate buffer stock solution required to produce 0.1 M solutions of
each buffer at the required pH .................................................................................................... 116
3.1 Potentiometric selectivity coefficients of NH4I PME 1 .............................................................. 130
3.2 Potentiometric selectivity coefficients of NH4I PME 1 (ionic additive omitted) ......................... 132
3.3 Potentiometric selectivity coefficients of HIII PME 1 ................................................................ 135
3.4 Potentiometric selectivity coefficients of HV PME 1 ................................................................. 136
7.1 Selectivity determination of NH4I PVC drop-coated IDE Design 2 (2% w/v)
(conductive response) ................................................................................................................ 233
7.2 Selectivity determination of NH4I PVC drop-coated IDE Design 2 (2% w/v)
(capacitive response) ...................................................................................................................................... 234
7.3 Selectivity determination of NO2I PVC drop-coated IDE Design 2 (2% w/v)
(conductive response) ....................................................................................................... 237
Craig Alexander
17
7.4 Selectivity determination of NO2I PVC drop-coated IDE Design 2 (2% w/v)
(capacitive response) ................................................................................................................. 238
7.5 Final pH and TDS concentration of the diluted buffer solutions used to test the response of the
impedimetric pH sensors ........................................................................................................... 239
8.1 The resulting concentration, in ppm, of unionized ammonia present following additions of an
ammonium stock solution to a test sample at various pH values ................................................. 251
10.1 Concentration values of species stated in ppm throughout the thesis converted to the
molar equivalent ....................................................................................................................... 265
10.2 Details of the commercial ionophores used in the experimental work described throughout
this thesis .................................................................................................................................. 266
Craig Alexander
18
Abstract
Name of the University: The University of Manchester
Author: Craig Alexander
Degree Title: Doctor of Philosophy (PhD)
Thesis Title: The Development of Smart Sensors for Aquatic Water Quality Monitoring
Date: September 2014
The focus of this project was to investigate the use of interdigitated electrodes (IDEs) as
impedimetric ion-selective chemical sensors for the determination of several important
analytes found within a freshwater aquarium. The overall aim of this research was to
work towards a prototype sensing device that could eventually be developed into a
commercial product for sale to aquarium owners.
Polyvinyl chloride and sol-gels containing commercially-available ionophores for four
aquarium-significant ions (NH4+, NO2
-, NO3
- and pH) were prepared and investigated
for use within polymeric ion-selective membranes. Three separate IDE transducers were
produced using either photolithography or screen-printing microfabrication techniques.
A sinusoidal voltage was applied to the IDEs and an LCR meter was used to measure
changes in the conductance and capacitance of the ion-selective membrane layer
deposited over the electrode digits.
Each ionophore, when tested within potentiometric ion-selective electrodes (ISEs), was
found to be suitable for further investigation within IDE devices. Sol-gels were
investigated as a potential membrane material for a coated wire electrode; however,
poor response characteristics were observed. An IDE sensor fabricated in-house using
lift-off photolithography and spin-coated with a polymeric membrane was found to
produce non-selective responses caused by changes in the conductivity of the test
solution. IDE devices with reduced geometric parameters were purchased and coated
with a selective polymeric membrane. When the membrane was spin-coated,
non-selective responses were observed; therefore, drop-coating of the membrane
material was investigated. This initially resulted in an unacceptably long response time;
however, this effect was reduced by decreasing the membrane solution viscosity prior to
drop-coating. A fully-screen printed carbon IDE device was fabricated by incorporating
the ionophore into a support matrix based on a commercial dielectric paste. Matrix
interferences to the sensor response were reduced by printing ‘build-up’ layers over the
sensing area prior to the ion-selective membrane.
Two novel routes for monitoring the water quality of an aquarium, using IDE sensors
fabricated by either photolithography or screen-printing, have been demonstrated. Due
to the commercial aspect of this project, it is important to consider the final cost of
producing these sensors. Both of the techniques used to produce ion-selective sensors
require further experimentation to optimise the sensor response, prior to integration
within a multi-analyte sensing prototype.
Craig Alexander
19
Declaration
I declare that no portion of the work referred to in the thesis has been submitted in
support of an application for another degree or qualification of this or any other
university or other institute of learning.
Copyright Statement
i. The author of this thesis (including any appendices and/or schedules to this thesis)
owns certain copyright or related rights in it (the “Copyright”) and s/he has given The
University of Manchester certain rights to use such Copyright, including for
administrative purposes.
ii. Copies of this thesis, either in full or in extracts and whether in hard or electronic
copy, may be made only in accordance with the Copyright, Designs and Patents Act
1988 (as amended) and regulations issued under it or, where appropriate, in accordance
with licensing agreements which the University has from time to time. This page must
form part of any such copies made.
iii. The ownership of certain Copyright, patents, designs, trademarks and other
intellectual property (the “Intellectual Property”) and any reproductions of copyright
works in the thesis, for example graphs and tables (“Reproductions”), which may be
described in this thesis, may not be owned by the author and may be owned by third
parties. Such Intellectual Property and Reproductions cannot and must not be made
available for use without the prior written permission of the owner(s) of the relevant
Intellectual Property and/or Reproductions.
iv. Further information on the conditions under which disclosure, publication and
commercialisation of this thesis, the Copyright and any Intellectual Property and/or
Reproductions described in it may take place is available in the University IP Policy
(see http://documents.manchester.ac.uk/DocuInfo.aspx?DocID=487), in any relevant
Thesis restriction declarations deposited in the University Library, The University
Library’s regulations (see http://www.manchester.ac.uk/library/aboutus/regulations) and
in The University’s policy on Presentation of Theses.
Craig Alexander
20
Acknowledgements
I would like to take this opportunity to thank my supervisor Professor Peter Fielden for
all of his advice and continued support throughout this project, and also Professor Nick
Goddard for his role as co-supervisor. I would also like to thank Dr. Sara Baldock for
all of the invaluable help she has provided throughout my PhD. Thank you to all of the
other post-docs in the miniaturisation group, past and present, especially Dr. Bernard
Treves Brown, who spent many hours writing LabVIEW code and helping me to set up
equipment, Dr. Behnam Bastani, who advised in the preparation of sol-gels and using
the e-beam, and Dr. Jeremy Hawkes, who helped me to set up several pieces of
instrumentation. My thanks to Dr. Sung Quek for allowing me to use his LabVIEW
program to control the LCR meter.
I would like to thank Dr. Tim Harvey for all of his help as my industrial co-supervisor
during the early stages of the project, and Barry Jones from Compact Instruments who
has supported the project financially throughout.
I would also like to thank Dr. Barry Haggett from the University of Bedfordshire, and
Dr. Craig Banks and Dr. Jonathan Metters from Manchester Metropolitan
University/Kanichi Research, for the many hours they spent providing specialist help
producing screen-printed electrodes for this project.
Finally, I would like to thank my family, friends and girlfriend for all of their continued
support.
This project was supported by funding from the Engineering and Physical Sciences
Research Council (EPSRC).
Craig Alexander
21
List of Abbreviations
%RSD Relative standard deviation
µ-TAS Micro-total analysis system
AC Alternating current
AOB Ammonia-oxidising bacteria
APPA American Pet Products Association
AU Auxiliary Electrode
BHT Butylated hydroxytoluene
CAS Chemical Abstracts Service
CP Conducting polymers
CWE Coated-wire electrode
DEDMS Diethoxydimethylsilane
dH Degrees of hardness
DO Dissolved oxygen
EIS Electrochemical impedance spectroscopy
FIM Fixed interference method
FPM Fixed primary ion method
GH General hardness
HIII Hydrogen Ionophore III
HV Hydrogen Ionophore V
IDE Interdigitated electrode
ISAB Ionic strength adjustment buffer
ISCOM Ion-selective conductometric microsensor
ISE Ion-selective electrode
ISFET Ion-selective field effect transistor
IUPAC International Union of Pure and Applied Chemistry
KH Carbonate hardness
Craig Alexander
22
LDL Lower detection limit
MPM Matched potential method
MSM Mixed-solution method
MTES Methyltriethoxysilane
N-E Nicolskii-Eisenman
NH4I Ammonium Ionophore I
NO2I Nitrite Ionophore I
NO3V Nitrate Ionophore V
NOB Nitrite-oxidising bacteria
NPOE o-nitrophenyl octyl ether
NTC Negative temperature coefficient
OATA Ornamental Aquatic Trade Association
PEG Polyethylene glycol
PFMA Pet Food Manufacturers Association
PID Proportional-integral-derivative
PME Polymer membrane electrode
ppm Parts per million
PVC Polyvinyl chloride
RE Reference Electrode
rpm Revolutions per minute
SCE Saturated calomel electrode
SHE Standard hydrogen electrode
SP-ISE Screen-printed ion-selective electrode
SSM Separate-solution method
TAN Total ammonia nitrogen
TDAN Tetradodecylammonium nitrate
TDMACl Tridodecylmethylammonium chloride
Craig Alexander
23
TDS Total dissolved salts
TEOS Tetraethoxysilane
THF Tetrahydrofuran
UDL Upper detection limit
UV Ultraviolet
WE Working Electrode
Craig Alexander
24
Project Overview
The main objective of this PhD project was to develop ion-selective chemical sensors,
with the specific application of monitoring key ions within a tropical freshwater
aquarium. Both technical and financial assistance to the project was provided by
Compact Instruments Ltd., who sought to exploit the findings of this project to develop
a prototype multi-analyte sensing device, with the eventual outcome of producing a
commercial product for sale to home aquarium owners.
This thesis seeks to address the feasibility of producing such a device by establishing
which chemical species are the most important for aquatic water quality monitoring. It
will also discuss some of the approaches for testing that aquarium hobbyists currently
have available. The limitations of these existing methods will be discussed along with
potential technologies that could be utilised to produce an easy-to-use sensing device.
The benefits that this would offer over current methodologies will also be evaluated.
As the primary motivation of this project was to produce a device to be sold to
household consumers, this research focuses on low-cost technologies which could be
easily implemented on a commercial scale. The eventual product would also be used by
non-specialists; therefore, it was also of great importance to consider the final design of
the proposed sensor, to ensure that it could be operated by the user in a low maintenance
fashion without any prerequisite chemical knowledge.
Craig Alexander
25
1 Introduction
The keeping of fish as pets within the home is a longstanding and popular hobby in
many countries all over the world. Recent data from the Ornamental Aquatic Trade
Association (OATA) indicated that there are currently around 3–3.5 million households
in the UK who own either an ornamental aquarium or a pond, making fish the third
most popular pet category after dogs and cats1. The Pet Food Manufacturers Association
(PFMA) suggests that around 9% of UK households own a fish tank2 and there are
almost 15 million aquaria throughout Europe3. A survey conducted by the American Pet
Products Association (APPA) states that in 2013, 14.3 million homes in the USA kept
freshwater fish, with a further 1.8 million keeping saltwater fish4. Furthermore, it is
estimated that the fishkeeping industry is worth more than 1 billion US dollars each
year.
1.1 Aquarium Chemistry
Owning an aquarium can be a very rewarding pastime, providing that adequate care and
maintenance of the artificial habitat takes place regularly. Generally, a home aquarium
set up for the keeping of fish will fall into one of two categories: freshwater, which can
either be cold-water or tropical, or marine (saltwater). Each of these artificial
environments provides a complex and challenging matrix for the analysis of many
chemical species. Such species are very important for maintaining the health of the fish,
which can be sensitive to even slight changes in various physical or chemical properties
within the environment.
The purpose of this section is to discuss the key chemical constituents within a tropical
freshwater aquarium; often described collectively as its ‘water quality’. Various ionic
Craig Alexander
26
species dissolved within the aquatic habitat provide a delicate environment which must
be suitably monitored and controlled. A decline in the water quality of an aquarium can
easily spoil the enjoyment of an aquarium hobbyist; as even small changes can quickly
lead to issues such as murky water; stress, illness or even death of often expensive
aquatic organisms. Marine water has other specific considerations that will not be
addressed in this thesis.
Throughout this thesis, analyte concentrations will be quoted in units of parts per
million (ppm), equivalent to 1 mg of solute dissolved in a 1 litre solution, as this is
standard fishkeeping nomenclature. A table stating the molar equivalent concentrations
can be found in Appendix 10.1.
1.1.1 Analytes from the Nitrogen Cycle
Ammonia, nitrite and nitrate are three primary nitrogenous species that exist within an
aquarium; each of these has its own specific effects on water quality. Ammonia is
converted in a two-step oxidation to nitrite and then nitrate in a process called
nitrification, commonly referred to as the biological filtration of an aquarium5.
1.1.1.1. Ammonia
Ammonia is produced in an aquarium via protein metabolism in fish and is expelled
from the gills. A smaller quantity also comes from ammonification of organic nitrogen
which arises from faeces and uneaten food6.
When dissolved in an aqueous environment, ammonia exists in two forms, unionized
ammonia (NH3) and the ammonium ion (NH4+). These species are collectively referred
Craig Alexander
27
to as the total ammonia nitrogen (TAN) concentration6. The equilibrium between the
two forms of ammonia dissolved in water is shown in Equation 1.1.
𝐇𝟑𝐎+(𝒂𝒒) + 𝐍𝐇𝟑 (𝒂𝒒)
𝐇𝟐𝐎(𝒂𝒒) + 𝐍𝐇𝟒+
(𝒂𝒒) Equation 1.1
This equilibrium is heavily dependent on the pH and temperature of the solution in
which the ammonia is dissolved7. The pKa of the ammonium ion is 9.26
8; hence
dissolved unionized ammonia only becomes thermodynamically favourable in alkaline
conditions. If the TAN concentration, pH and temperature are known, it is possible to
calculate the free unionized NH3 concentration, as described by Thurston et al7.
Equation 1.2 shows the expression for calculating the fraction of unionized ammonia, f,
at a particular pH. This value is then multiplied by the TAN concentration to give the
concentration of unionized ammonia within the sample.
𝐟 = 𝟏 (𝟏𝟎𝐩𝐊𝐚−𝐩𝐇 + 𝟏)⁄ Equation 1.2
As the equilibrium is also temperature dependent, Thurston et al. produced an empirical
mathematical expression to relate the equilibrium constant with temperature, which can
simply be inserted into Equation 1.2 as the pKa term7. This is shown in Equation 1.3,
where T is the temperature (K).
𝐩𝐊𝐚 = 𝟎. 𝟎𝟗𝟎𝟏𝟖𝟐𝟏 + 𝟐𝟕𝟗𝟐. 𝟗𝟐 𝐓⁄ Equation 1.3
This expression can be used in the temperature range of 0–50 °C and the pH range of
6.0–10.0.
It is of upmost importance to monitor the concentration of ammonia within a freshwater
aquarium as it is one of the most common water quality problems, due to its toxicity to
fish9. This toxicity is predominantly caused by the unionized form of ammonia, whilst
Craig Alexander
28
NH4+ is effectively non-toxic within the aquatic environment
10, due to increased
permeability of NH3 across biological membranes11
.
Different species of fish exhibit varied tolerances to ammonia; however, it is generally
accepted that its levels in freshwater aquaria should be maintained as close to zero as
possible. OATA guidelines state that the unionized ammonia concentration should not
exceed 0.02 ppm12
. This value will be treated as the upper limit for NH3 throughout this
thesis.
The first step of nitrification is carried out by Nitrosomonas and other ammonia-
oxidizing bacteria (AOB), which break down ammonia to form nitrite, NO2- 9
.
1.1.1.2. Nitrite
Nitrite (NO2-) is the intermediate species formed during the nitrification process. The
anion is absorbed by the gills and diffuses into the bloodstream, where its toxic action is
believed to arise due to oxidation of haemoglobin, resulting in respiratory problems and
eventually death13
.
OATA guidelines state that the nitrite concentration within a freshwater aquarium
should not exceed 0.2 ppm.
Nitrobacter and other nitrite-oxidizing bacteria (NOB) facilitate the final stage of the
nitrification process, the oxidation of nitrite to form nitrate, (NO3-) 9
.
1.1.1.3. Nitrate
Nitrate, NO3-, is the final and least toxic product formed during the oxidation of
ammonia. Nitrate permeates poorly through the gills which corresponds to its relatively
low toxicity when compared with the two previously discussed nitrogenous species. At
Craig Alexander
29
toxic concentrations, nitrate transports through the fish’s gills and diffuses into the
bloodstream, where it has a similar toxic action to nitrite, affecting the oxygen-carrying
ability of haemoglobin, eventually resulting in death14
.
Although nitrate can be tolerated at much higher levels than either ammonia or nitrite
before there are any toxic effects on fish its presence can cause algae to grow, which
leads to oxygen depletion within an aquarium9. Figure 1.1 shows an algal bloom within
an aquarium caused by high nitrate levels.
OATA recommends that nitrate levels do not exceed 50 ppm in a freshwater aquarium.
Figure 1.1: Algae accumulation caused by high levels of nitrate in an aquarium, taken from
Roberts et al.9
As mentioned previously, the two-step oxidation of ammonia to form nitrite and then
nitrate is often referred to as biological filtration. When a new aquarium is first set-up a
‘conditioning’ period is required before any fish are added. During this period the levels
of AOB and NOB increase on surfaces within the tank, which increases the efficiency
of nitrification and reduces the ammonia and nitrite levels. Regular water changes are
Craig Alexander
30
required to eliminate nitrate from an aquarium as it will continue to accumulate as the
other two species are oxidized5. It is therefore important to ensure that all three of the
nitrogenous species are adequately monitored to prevent them from reaching dangerous
levels within the closed aquarium system.
1.1.2 Aquarium pH
pH is a measure of the activity, although in practical terms it is a measure of the molar
concentration of hydrogen cations in an aqueous solution, and is defined by
Equation 1.4.
𝐩𝐇 = −𝐥𝐨𝐠 [𝐇+] Equation 1.4
As mentioned is section 1.1.1.1, the pH of the aquarium has effects on other water
quality factors, therefore it is very important that it is monitored. pH is a much more
species-specific indicator of water quality as some tropical fish prefer a more acidic, or
alkaline, environment than others. Generally, a near-neutral pH is favourable in a
tropical freshwater aquarium with most species preferring water with a pH of 5.5–8.5 9.
However, rapid fluctuations in pH are more likely to cause harm to fish, however, rather
than a specific pH value.
1.1.3 Chlorine
Chlorine is routinely added to tap water around the world as a disinfectant to kill
waterborne pathogens. As tap water is commonly used to fill aquaria, the presence of
chlorine must be considered as this species is highly toxic to fish. When Cl2 is dissolved
in water it forms hypochlorous acid (HOCl) which equilibrates with hypochlorite anions
(ClO-) above pH 5.0, which damages the gill tissue of fish at very low concentrations
5.
Craig Alexander
31
No values are published for safe levels of chlorine within an aquarium. It is generally
accepted that it should be completely removed from the water before it is added to the
aquarium9.
Aquarium owners will generally treat tap water with a sodium thiosulfate-based
dechlorinating agent5 when setting up an aquarium and performing water changes,
making regular chlorine monitoring less critical than some of the other water quality
indicators.
The chloride anion (Cl-) is generally well tolerated by freshwater fish in the expected
range found in most tap water samples, up to a maximum of 250 ppm, so there is little
requirement for this to be regularly tested in an aquarium.
1.1.4 Dissolved Oxygen
A supply of dissolved oxygen (DO) within an aquarium is required not only for fish
respiration but also for respiration in bacteria and plants15
. The saturation of DO is
dependent on the physical and chemical properties of the aqueous environment. For
example, less oxygen can be dissolved as temperature or salinity increases. At a typical
aquarium temperature of 25 °C, 100% saturation of DO corresponds to 8.3 ppm16
. The
average saturation in an aquarium is around 70% (5.8 ppm at 25 °C)17
. OATA
guidelines recommend a DO concentration of at least 6 ppm is maintained within a
freshwater aquarium.
By ensuring than an aquarium is not overstocked with fish, and providing a source of
agitation such as an airstone, DO levels can be sustained at near saturation. Therefore it
is not critical for this water quality parameter to be regularly monitored.
Craig Alexander
32
1.1.5 Phosphate
The orthophosphate anion (PO43-
) has similar effects within an aquarium to nitrate as it
is not considered directly harmful to aquatic organisms in the levels likely to be
observed in an aquarium. PO43-
, however, acts as a fertiliser that can cause algal blooms.
This can occur at low PO43-
concentrations therefore, most literature states that the
concentration should not exceed 0.05 ppm18
. Aquarists should be particularly mindful of
the phosphate concentration within their aquarium when using phosphate-based buffers
to stablise aquarium pH; although most pH buffers intended for aquarium use no longer
contain phosphate5.
1.1.6 Water Hardness
Water hardness is a broad term that relates to the amount of dissolved minerals in the
water. Fishkeeping literature generally deals with water hardness as two separate
parameters; general hardness (GH), usually defined as the total concentration of
magnesium (Mg2+
) and calcium (Ca2+
) cations19
, and carbonate hardness (KH), which is
concerned with the amount of free carbonate (CO32-
) and bicarbonate (HCO3-) anions
within the aquarium.
1.1.6.1. General Hardness
Although other multivalent cations such as iron (Fe3+
), aluminium (Al3+
) and manganese
(Mn2+
) may be present in aquarium water, the main species that lead to ‘hard’ water are
Ca2+
and Mg2+
predominantly from carbonate salts (~80%) with the remainder from
sulfate and chloride salts5. For fishkeeping purposes, GH is commonly expressed in
units of ppm with respect to CaCO3, or more simply, using the degrees of hardness (dH)
scale, where 1 dH is equivalent to 17.8 ppm CaCO3. This method provides aquarists
Craig Alexander
33
with a simple and manageable way of monitoring water hardness within their aquarium
as opposed to using scientific nomenclature relating to concentrations of individual
ions. The definitions of what category of hardness a particular dH value belongs to vary
between different fishkeeping advice publications. Table 1.1 shows a typical dH scale
along with its corresponding equivalent CaCO3 concentration and hardness category.
dH Approximate equivalent
concentration of CaCO3/ppm Hardness category
0–4 0–70 Very soft
4–8 70–140 Soft
8–12 140–210 Medium-hard
12–20 210–350 Hard
>20 >350 Very hard
Table 1.1: dH scale for water hardness
Suitable levels of GH for most ornamental freshwater fish usually lie between
~6–18 dH, although it is very much a species-dependent parameter.
1.1.6.2. Carbonate Hardness
The KH (from the German spelling of carbonate, karbonat) of aquarium water provides
a pH buffering capacity, allowing for a more stable pH value. It is sometimes referred to
in fishkeeping literature as alkalinity, due to the ability of a solution with a high KH to
maintain an alkaline pH, although this is scientifically inaccurate. Aquaria with a low
KH will likely have a low pH, and also be susceptible to rapid drops or fluctuations in
pH due to a limited buffering capacity provided from the equilibrium between carbonate
and bicarbonate anions. KH is usually expressed in the same units as GH, where the
total concentration of CO32-
and HCO3- are converted to an equivalent value with respect
to CaCO3. Table 1.1 is therefore applicable for both GH and KH. KH values should not
drop below approximately 4 dH15
, particularly when live plants are kept. Low KH
Craig Alexander
34
affects the ability of aquarium water to dissolve carbon dioxide, which is required for
photosynthesis5.
1.1.7 Summary of Parameters
Table 1.2 provides a summary of each of the water quality factors discussed in this
section, along with the respective acceptable limits in a freshwater aquarium.
Analyte Acceptable concentration
within a freshwater aquarium
Ammonia <0.02 ppm (as NH3)
Nitrite <0.2 ppm
Nitrate <50 ppm
pH pH 5.5–8.5
(species dependent)
Chlorine/Chloramine Not detectable
Chloride Drinking water levels
(<250 ppm)
Phosphate <0.05 ppm (as PO43-
)
Dissolved Oxygen >6 ppm
General Hardness
6–18 dH
(100–300 ppm as CaCO3)
(species dependent)
Carbonate Hardness >4 dH
(>70 ppm as CaCO3)
Table 1.2: Summary of the acceptable criteria of each of the most important analytes identified for
acceptable water quality within freshwater aquaria
After reviewing both scientific and fishkeeping literature which discuss the key
chemical components present within a tropical freshwater aquarium, the focus of the
experimental work in this thesis will be on the determination of the aquarium pH and
the analysis of the three nitrogenous species as these appear to be the most crucial for
fish health. An aquarium hobbyist can be confident in the water quality of their
aquarium if these four components are known to be at acceptable levels. It is beneficial,
Craig Alexander
35
rather than essential, that others such as phosphate and water hardness are known;
therefore they are deemed outside the immediate scope of this work.
1.2 Current Methods of Analysis
As mentioned previously, the monitoring of aquarium water quality is very important to
fish health, and therefore various methods of analysis have been developed that are
available to aquarists. The purpose of this section is to discuss some of the options
currently available for water quality monitoring within aquaria, along with their
limitations, which are to be addressed throughout this thesis.
1.2.1 Test Kits
Currently, most fish keepers will carry out routine colourimetric chemical tests to check
the levels of various key analytes within their aquarium, usually in the form of test
strips or indicator solutions. These are often sold as packs containing multiple tests for
several ions; the most basic will typically test the pH, ammonia, nitrite and nitrate levels
as these are seen as the most critical. More expensive and specialist test kits can be
purchased for other analytes such as water hardness and phosphate.
The process for checking water quality parameters using the indicator solutions
provided with test kits usually involves removing a sample of water from the aquarium
and placing it into a small vial, adding the relevant indicator solution(s), and waiting to
observe a colour change. The coloured solution within the vial is held against a colour
chart which corresponds to an estimation of the concentration of the target species. This
value can then be checked against a given range from the literature to inform the user if
the levels within their tank are safe, or if any further action needs to be taken.
Craig Alexander
36
1.2.1.1. pH
pH indicators are weak acids or bases that are protonated and deprotonated at different
pH values, with each form existing as a different colour. The observed colour change
occurs gradually over a range of concentrations of hydrogen ions and therefore a
reasonable estimation can be made towards the pH of the tested sample.
Various pH indicators are used in aquarium test kits and each has more favourable
properties depending on whether the pH is being tested over a broad or narrow range, or
if marine water or freshwater is the test sample. Three common indicators include:
bromothymol blue (pH range 6.0–7.6) and phenol red (pH range 6.4–8.2), used for low
and high range freshwater pH testing, respectively; and m-cresol purple
(pH range 7.6–9.2), used in saltwater pH test kits. Figure 1.2 shows the two coloured
forms of phenol red observed upon addition of either OH- or H
+, and Figure 1.3 shows
the three species involved in the colour change of bromothymol blue.
Figure 1.2: Structure of the two forms of phenol red which result in the observed colour change
20
Figure 1.3: Structure of three forms of bromothymol blue which result in a colour change at
varying pH values21
Craig Alexander
37
1.2.1.2. Ammonia
Colourimetric testing for dissolved ammonia is usually based on either the Nessler or
Salicylate reactions, both of which are long-standing and well researched methods.
1.2.1.2.1. Nessler’s Reagent Method
The reaction of the Nessler reagent, an alkaline solution of mercury (II) iodide and
potassium iodide with the formula K2HgI4, was first suggested for the direct, qualitative
analysis of aqueous ammonia in 1856 by Julius Nessler. Nichols and Willets further
investigated the reaction scheme followed by this process, outlining any intermediates
formed along with the observed colour changes22
. In the presence of ammonia, a
yellow-coloured species is produced with an intensity that is proportional to the
concentration of ammonia in the sample. The reaction is shown in Equation 1.5.
𝟐𝐊𝟐𝐇𝐠𝐈𝟒 + 𝐍𝐇𝟑 + 𝟑𝐊𝐎𝐇 → 𝐇𝐠𝟐𝐎𝐈𝐍𝐇𝟐 + 𝟕𝐊𝐈 + 𝟐𝐇𝟐𝐎 Equation 1.5
1.2.1.2.2. Salicylate Method
The other colourimetric method for ammonia determination which is often used in
aquarium test kits is known as the salicylate method. This method is based on the
reaction between phenol and hypochlorite to form a blue-coloured indophenol. This is a
multi-step process, outlined below in Equations 1.6 and 1.7, which first requires the
formation of monochloramine via the reaction of ammonia with hypochlorite
(Equation 1.6). Monochloramine then reacts with salicylate to form 5-aminosalicylate
(Equation 1.7), which is oxidized in the presence of an iron catalyst to form
indosalicylate, shown in Figure 1.4.
Craig Alexander
38
𝐍𝐇𝟑 + 𝐎𝐂𝐥− → 𝐍𝐇𝟐𝐂𝐥 + 𝐎𝐇− Equation 1.6
Equation 1.7
Figure 1.4: Structure of indosalicylate
In the presence of an excess of the yellow-coloured catalyst, the blue indosalicylate
appears green. The intensity of the green colour observed is directly proportional to the
amount of ammonia present in the sample23
.
1.2.1.3. Nitrite
The most common nitrite indicators are based on the Griess reaction24
, a diazotisation
reaction which utilises any NO2- within the tank to form an azo-dye compound. The
colour of the azo-dye intensifies depending on the amount of free NO2- that is available
to react. The reaction scheme for this is outlined in Equation 1.8 and 1.9 25
. This two-
step mechanism proceeds via the reaction of nitrite with sulfanilamide (Equation 1.8)
followed by the addition of N-1-naphthylethylenediamine dihydrochloride (NED) to
form the azo-dye (Equation 1.9).
Craig Alexander
39
Equation 1.8
Equation 1.9
1.2.1.4. Nitrate
The determination of aquarium nitrate levels usually involves the reduction to nitrite
with cadmium metal prior to testing using the Griess reaction as above24
. Therefore one
limitation of this method is that the total concentration of both nitrate and nitrite is
observed.
Performing wet chemical tests to indicate the concentration of the target species offers
many sources of error, particularly for non-specialist users. The interpretation of the
results by comparing against colour charts is a subjective process, which could lead the
user to believe that their test is reading higher or lower than in actuality. Users may also
cause erroneous results due to poor adherence to a multi-step testing process, inefficient
shaking of the sample vial or not allowing sufficient time for the observed colour
Craig Alexander
40
change to occur. Finally, some of the chemicals involved are potentially harmful if
incorrectly handled.
1.2.2 Commercial Sensors
An alternative to using colourimetric test kits for aquarium water quality monitoring is
to use chemical sensors. Commercial electrochemical sensors, in the form of
potentiometric ion-selective electrodes (described later in sections 1.3 and 1.4), are
available to purchase for each of the most important target ions found within a home
aquarium (pH, NH4+, NO2
-, NO3
-). These are generally in the form of a handheld probe
that is submerged into the sample to obtain a reading, rather than an in situ device
offering continual monitoring. Sensors such as these are normally intended for
laboratory, rather than home, use and as such are very expensive, often costing several
hundreds of pounds for one probe. This prohibitive cost makes them a non-viable option
for most aquarium hobbyists. Such devices also need to be calibrated prior to use which
would require the aquarium owner to keep standard solutions of each target ion at
known concentrations; thus potentially adding user error to the analysis.
1.2.3 Benefits of the Proposed Device
The availability of a high quality, reliable and affordable sensor array for in situ
determination of the key aquarium water quality factors would eliminate the problems
associated with test kits, described in section 1.2.1, whilst providing a more viable
alternative for aquarium hobbyists to the expensive commercial probes that are
currently available.
A device providing continual monitoring with a suitable output would alert users
immediately if their aquarium required attention, rather than relying on routine
Craig Alexander
41
intermittent testing. Such a device would also provide an instantaneous output of
multiple parameters, as opposed to trial-and-error testing to deduce the cause of an
existing problem. The subjectivity of chemical testing could be removed by providing
the user with a quantitative or semi-quantitative result. Rather than interpreting the
results from a colour chart, the concentration of the species of interest would be
conveyed directly as a value, or by way of a ‘traffic-light’ system where the user is
alerted when the analyte of interest approaches levels where attention is required. By
incorporating a suitable transmitter and receiver unit into the device the results could
also be output to electronic equipment such as a standalone screen, computer or mobile
phone.
Another important benefit that the proposed device would offer is the complete removal
for the requirement of any toxic or harmful chemicals. Batch calibrations of the device
could be performed during manufacture, which would eliminate the need to provide
calibrant solutions. Table 1.3 summarises the benefits that the proposed sensor would
offer over the testing methods that are currently available.
Proposed sensor Test kits Commercial sensors
Continual monitoring
Free from chemicals
Inexpensive
Use without pre-calibrating
Quantitative, or semi
quantitative result Results outputted to a screen,
PC or mobile device Table 1.3: Comparison of the features of the proposed sensor device with the options
currently available
Craig Alexander
42
1.3 Introduction to Chemical Sensors
Chemical sensors have been defined as “a device which responds to a particular analyte
in a selective way through a chemical reaction and can be used for the qualitative or
quantitative determination of the analyte”26
. The primary function of a chemical sensor
is to provide real-time information about a test sample relating to the concentration of a
particular target analyte, or group of analytes27
. The purpose of this section is to
introduce some of the important features of chemical, specifically electrochemical,
sensors, which will form the main focus of the experimental work in this thesis.
Over the past few decades, much research has been carried out to attempt to miniaturise
traditional ‘bench-top’ laboratory techniques, including using chemical sensors in place
of conventional analytical processes. Using sensors offers several advantages for many
analytical applications compared with established macro-scale laboratory techniques,
such as the removal of complicated sample preparation or pre-treatment stages.
Chemical sensors also offer a more feasible route for on-site and in situ analysis whilst
also facilitating integration within a miniaturised device for single or multi-analyte
sensing28
.
A chemical sensor requires two main components: a recognition element, which imparts
the selectivity of the sensor enabling it respond to a particular analyte with minimal
matrix interferences; and a transducer, which converts a change in analyte concentration
into a measureable signal29
. A biosensor is a specific type of chemical sensor that
utilises a biological receptor, such as an enzyme or antibody, to facilitate a
highly-selective response to a particular target analyte.
Craig Alexander
43
Many different transducers for chemical sensors exist, including optical, piezo-electic
and thermal; however, for the purpose of this thesis the focus will be on
electrochemical transducers.
Chemical sensors must perform acceptably across a number of performance
characteristics for them to be successfully used for their desired application. These
include: dynamic range, precision, response time, operational lifetime, sensitivity, limit
of detection and selectivity against interfering species27, 29
.
1.3.1 Electrochemical Sensors
Three electrochemical processes that are often utilised in transducers for chemical
sensors are: potentiometry, voltammetry and electrical impedance29
. Voltammetric
sensors will not be considered in this thesis and impedimetric devices will be described
in detail throughout section 1.5.
1.3.1.1. Potentiometry
Potentiometric sensors involve the measurement of the potential difference between an
indicator electrode (EIND) and a suitable reference electrode (EREF) at zero current, as
shown in Equation 1.10 29
. Reference electrodes will be described in further detail in
section 1.3.1.2.
∆𝐄 = 𝐄𝐜𝐞𝐥𝐥 = 𝐄𝐈𝐍𝐃 − 𝐄𝐑𝐄𝐅 Equation 1.10
The Nernst equation defines the logarithmic relationship between the observed potential
and the ionic activities of the species in an electrochemical cell. Equation 1.11 shows a
typical half-cell reaction and Equation 1.12 shows the corresponding Nernst equation,
where E is the measured electrode potential (V); E° is the standard electrode potential
Craig Alexander
44
for the half-cell (V); R is the universal gas constant (8.3145 J mol-1
K-1
); T is the
temperature (K); n is the stoichiometry of the electrons involved, F is the Faraday
constant (96,485 C mol-1
); and aox and aR are the activities of the respective species.
𝐎𝐱 + 𝐧𝐞− 𝐑 Equation 1.11
𝐄 = 𝐄° +
𝐑𝐓
𝐧𝐅 𝐥𝐧 (
𝒂𝑶𝒙
𝒂𝑹)
Equation 1.12
For dilute solutions, the activities can be assumed to be equal to the molar
concentration29
. The activity of a species, X, is related to the molar concentration by the
activity coefficient, γ, as shown in Equation 1.13.
𝐚𝐗 = γ𝐗[𝐗] Equation 1.13
Assuming ax=[X], at room temperature (298 K) Equation 1.12 becomes:
𝐄 = 𝐄° +
𝟐. 𝟓𝟔𝟖 × 𝟏𝟎−𝟐
𝐧 𝐥𝐧 (
[𝐎𝐱]
[𝑹])
Equation 1.14
As conventional molar concentrations are usually expressed in powers of 10, Equation
1.14 can be multiplied by 2.303 to convert from a natural logarithm to a base 10
logarithm:
𝐄 = 𝐄° +
𝟎. 𝟎𝟓𝟗𝟐
𝐧 𝐥𝐨𝐠 (
[𝐎𝐱]
[𝑹]) Equation 1.15
[R] is generally a metal with a fixed concentration and therefore Equation 1.15 can be
further simplified:
Craig Alexander
45
𝐄 = 𝐄° +
𝟎. 𝟎𝟓𝟗𝟐
𝐧 𝐥𝐨𝐠([𝐎𝐱]) Equation 1.16
For experimental purposes where E° is not known and RT/F may differ from the
theoretical value, a more practical form of the Nernst equation can be written. This is
shown in Equation 1.17, where [C] is the concentration of the species of interest.
𝐄 = 𝐊 +
𝐒
𝐧 𝐥𝐨𝐠[𝐂] Equation 1.17
When a graph of E against log [C] is plotted, a straight line with a slope (S) and
intercept (K) is obtained. The values obtained experimentally can be compared with the
theoretical values, where K=E° and S=0.0592/n at 298 K 29
. A potentiometric sensor
can be deemed ‘Nernstian’ if the measured slope fits closely with the theoretical value,
i.e. for a monovalent cation at 298 K there would be a 59.2 mV increase in potential per
decade of change in ion concentration.
The construction of the indicator electrode can be prepared in such a way so that its
response to one particular ion is increased whilst minimising the interference observed
in the measured potential from any other ions present in the solution. These
potentiometric sensors are called ion-selective electrodes (ISEs) and will be discussed in
section 1.4.
1.3.1.2. Reference Electrodes
As mentioned previously, potentiometric sensors require a suitable reference electrode
against which the potential can be measured. It is not possible to measure the absolute
potential of a half-cell directly; instead it is the difference between the potential of the
half-cell of interest (the sensing, or indicator electrode) and the potential of a reference
Craig Alexander
46
electrode that is measured. An ideal reference electrode must therefore provide an
accurate and stable potential that is not influenced by the test solution30
.
The standard hydrogen electrode (SHE) involves the redox reaction of hydrogen gas at a
platinum electrode in an acidic solution and is used to compare the potential of
half-cells, allowing the compilation of a list of standard electrode potentials at known
temperatures. The SHE half-cell is shown in Equation 1.18. By convention, the SHE is
assigned a potential value of zero under standard conditions (298 K, 1 atm, 1 M H+).
𝟐𝐇+ + 𝟐𝐞− 𝐇𝟐(𝒈) Equation 1.18
Although the SHE has a great importance, it is not the most well-suited reference
electrode in practical situations. The two most common and commercially-available
electrochemical reference electrodes are the silver-silver chloride electrode (Ag/AgCl)
and the saturated-calomel electrode (SCE).
Ag/AgCl reference electrodes are produced by coating a layer of AgCl onto a piece of
Ag wire, which is submersed into a saturated or near-saturated (3 M) solution of KCl.
The electrode potential at 25 °C against the SHE is +0.223 V. The half-cell reaction is
shown in Equation 1.19 31
.
𝐀𝐠𝐂𝐥(𝒔) + 𝐞− 𝐀𝐠(𝒔) + 𝐂𝐥(𝒂𝒒)− Equation 1.19
The SCE is produced by placing mercury in contact with a paste consisting of
mercury (I) chloride and potassium chloride in a solution of saturated KCl. The half-cell
has a potential of +0.244 V against the SHE and proceeds according to Equation 1.20.
𝐇𝐠𝟐𝐂𝐥𝟐(𝒔) + 𝟐𝐞− 𝟐𝐇𝐠(𝒍) + 𝟐𝐂𝐥(𝒂𝒒)− Equation 1.20
Craig Alexander
47
The most practical reference electrode for use when obtaining potentiometric
measurements is a double-junction Ag/AgCl reference electrode. In many cases, a
simple single-junction reference is unsuitable for use, particularly when Cl- and/or K
+
ions are present in the sample solution. A double-junction reference electrode consists
of two compartments, one inside the other. The inner compartment contains the
Ag/AgCl electrode and is filled with a suitable electrolyte, such as saturated KCl. The
outer compartment is filled with an electrolyte that contains ions different from those
present in the sample solution, ensuring that any leakage from the outer compartment
does not affect the measurement. Lithium acetate is seen as a ‘universal’ outer
compartment filling solution within double-junction reference electrodes and is suitable
for use for the measurement of most ions. A schematic diagram of a double-junction
reference electrode is shown in Figure 1.5.
Figure 1.5: Schematic representation of a double-junction Ag/AgCl reference electrode with 1 M
LiCH3COO as outer filling solution32
Craig Alexander
48
1.4 Ion-Selective Electrodes
ISEs are potentiometric sensors which allow the selective determination of the activity,
and hence concentration, of one ionic species in the presence of others. The selective
ionic recognition is achieved using a membrane as the sensor surface, and the chemical
equilibrium which takes place across the membrane produces an electrode potential that
can be related to the concentration of the target ion by the Nernst equation33
.
1.4.1 Selective Polymeric Membranes
Various membrane materials have been used in potentiometric sensors to afford a
selective response, including modified glass, as used in pH-sensitive electrodes, and
crystalline materials such as LaF3 doped with EuF2, used within fluoride-selective
electrodes. Whilst both of these have their uses, they are limited in terms of the number
of ions that can be selectively determined using either of these methods34
. The
development of ISEs that incorporate a polymeric membrane containing an immobilised
ion-binding receptor has led to a vast amount of research in this field. In the literature
terms such as ion-exchanger or ion-carrier are used if the receptor is electrically charged
and neutral receptors are termed ionophores27
, for the purpose of this thesis, however,
‘ionophore’ will be used to refer to the receptor regardless of its charge.
Although others have been reported35,36,37
, the most widely used polymer material for
the preparation of ion-selective membranes is polyvinyl chloride (PVC)38
. The
selective-membrane ‘cocktail’ is prepared by dissolving PVC, along with an appropriate
plasticiser and the desired ionophore, into a suitable solvent, such as tetrahydrofuran
(THF)39
. When a neutral ionophore is used as the selective receptor, it is also necessary
to add a lipophilic ionic additive with a charge opposite to that of the target analyte to
Craig Alexander
49
prevent co-extraction of the counter-ion into the membrane40
. For cation-selective
membranes, alkali salts of tetraphenylborate derivatives (Figure 1.6A) are commonly
used as the ionic additive, whereas anion-selective membranes will generally include a
quaternary ammonium salt additive38
(Figure 1.6B).
A typical PVC membrane composition will consist of approximately 30–33% w/w
PVC, 60–66% w/w plasticiser and 1–10 % w/w ionophore41
with the ionic additive (if
required) present in a molar ratio of around 1:2, with respect to the ionophore34
.
Figure 1.6: (A) Potassium tetrakis-[3,5,-bis-(trifluoromethyl)phenyl]borate – used as an ionic
additive in cation-selective membranes, (B) tridodecylmethyl ammonium chloride (TDMACl) –
used as an ionic additive in anion-selective membranes27
The role of the plasticiser within an ion-selective membrane is not only to provide
mechanical plasticity to the membrane, but also to function as a solvent for the
membrane components; thus ensuring mobility of the ionophore within the membrane38
.
Plasticised polymers are essentially viscous liquids; and are commonly referred to in the
literature as liquid membranes. A plasticiser must exhibit good lipophilicity and be
compatible with the other membrane materials and the chosen solvent42
. The most
common plasticisers used in ISEs include phthalates, sebacates and
ortho-nitrophenyloctyl ether (NPOE)43
. The structures of some of these are shown in
Figure 1.7. Choosing the correct plasticiser for use within a particular ion-selective
membrane has been shown to improve both the selectivity and the detection limits
of an ISE44
.
Craig Alexander
50
Figure 1.7: Structures of three commonly-used plasticisers within ISEs (A) dioctyl sebacate (DOS),
(B) o-nitrophenyloctyl ether (NPOE), (C) dioctyl phthalate (DOP). Adapted from Buhlmann et al.34
1.4.2 Selectivity of ISEs
The selectivity of any chemical sensor is arguably its most important analytical
characteristic, as understanding how the sensor performs in the presence of other
species helps to ascertain whether the device is suitable for use in the intended
target sample45
.
An ideal ISE would respond only to the target ion within a solution that could contain
any number of potential interfering ions. In reality, this situation is highly unlikely,
especially where the sample matrix contains species with similar characteristics to the
target ion, such as size and charge, and hence the term ion-selective rather than
‘ion-specific’ is quoted. The potentiometric selectivity coefficient (kA,Bpot
) is used to
quantitatively evaluate the contribution that interfering ions have to the measured
potential46
. An additional term is inserted into the Nernst equation to account for the
effect of an interfering ion, to give the Nicolskii-Eisenman (N-E) equation, shown in
Equation 1.21, where aA and aB are the ionic activities and nA and nB are the charge
numbers of the target and interfering ions, respectively.
Craig Alexander
51
𝐄 = 𝐄° +
𝟐. 𝟑𝟎𝟑𝐑𝐓
𝐧𝐅 𝐥𝐨𝐠(𝒂𝐀 + 𝐤𝐀,𝐁
𝐩𝐨𝐭(𝒂𝐁)𝐧𝐀 𝐧𝐁⁄ ) Equation 1.21
The smaller the value of kA,Bpot
, the greater the ability of the ISE to distinguish the target
ion from the interfering ion. Its logarithm is generally taken to convert it to a more
convenient number. Like the Nernst equation, the N-E equation is also based on
activities of the ions, rather than the molar concentration. An ionic strength adjustment
buffer (ISAB) can be added to each sample solution used to adjust the total ionic
strength to a similar value, thus making the activity coefficient equal for each target ion
concentration33
. Therefore aA=CA and aB=CB, where CA and CB are the molar
concentrations of the target and interfering ions, respectively.
Several methods have been described for the experimental determination of kA,Bpot
. The
International Union of Pure and Applied Chemistry (IUPAC) recommended guidelines
outlining these methods were first published by Guilbault et al. in 197647
. These were
then revised in 1995 by Umezawa et al. to include the limitations of the N-E equation48
.
These procedures can generally be placed into two categories: mixed solution methods
(MSMs), where the electrode is placed into a test solution containing both primary and
interfering ions, and separate solution methods (SSMs), where the effect of CA and CB
are investigated independently.
Two common MSMs are the fixed interference method (FIM) and the fixed primary ion
method (FPM). The FIM involves measuring the potential of a solution containing fixed
CB with varying CA, whereas the FPM measures varying CB at fixed CA. The obtained
potential values are plotted against the log of the concentration of the target ion in the
case of the FIM, or the interfering ion for the FPM. The selectivity coefficient is
calculated by extrapolating the linear portion of the resulting plot to the intersection
Craig Alexander
52
which gives the value (CA for the FIM, CB for the FPM) to be entered into
Equation 1.22 48,49
; the FIM is illustrated graphically in Figure 1.8.
𝐤𝐀,𝐁
𝐩𝐨𝐭=
𝐂𝐀
(𝐂B)𝐧𝐀 𝐧𝐁⁄ Equation 1.22
Figure 1.8: Example calibration curve for calculating the selectivity coefficient using the FIM,
modified from Mikhelson et al.32
Another MSM is referred to as the two solution method (TSM). The potential of one
solution containing only the ion of interest at a known concentration is measured to give
EA, and the potential of a separate solution containing both the target and interfering
ions is also measured to give EA+B. The value of the potential difference between these,
ΔE = EA+B – EA, is entered into Equation 1.23 to calculate kA,Bpot
.
𝐤𝐀,𝐁𝐩𝐨𝐭
= 𝐂𝐀 (𝐞(𝚫𝐄𝐧𝐀𝐅 𝐑𝐓⁄ ) − 𝟏) (𝐂𝐁)⁄𝐧𝐀 𝐧𝐁⁄
Equation 1.23
The main benefit of using a MSM to calculate kA,Bpot
is that it simulates the actual
conditions in which the sensor is to be used, which gives a direct indication of the
Craig Alexander
53
performance of the sensor in the test sample, and in the case of the FIM and FPM there
is no risk of carry over contamination between measurements50
. However, there are also
SSMs that are accredited by IUPAC that are often the preferred method for determining
kA,Bpot
as they are generally simpler and allow the user to understand the behaviour of the
electrode when the target ion is not present.
The first SSM involves measuring one solution containing the target ion (A) at a known
concentration, and a separate solution containing the interfering ion (B) at the same
concentration as that of the target ion. The recorded potentials, EA and EB are entered
into Equation 1.24 to obtain kA,Bpot
.
𝐥𝐨𝐠 𝐤𝐀,𝐁
𝐩𝐨𝐭=
(𝐄𝐁 − 𝐄𝐀)
𝟐. 𝟑𝟎𝟑 𝐑𝐓 𝐧𝐀𝐅⁄+ (𝟏 − 𝐧𝐀 𝐧𝐁⁄ )𝐥𝐨𝐠 𝐂𝐀 Equation 1.24
A second SSM involves producing calibration plots of both the target and non-target
species. The concentrations of each ion that produce the same measured electrode
potential are entered into Equation 1.22 to obtain a value for the selectivity coefficient.
The limitations of using N-E equation-based methods for selectivity determination of
ISEs were critically evaluated after several papers reported discrepancies between
obtained kA,Bpot
values when different experimental conditions were used48
. The N-E
equation requires the ISE to exhibit a Nernstian response to both the target and
interfering ions, whilst it is also not particularly well suited when comparing ions of
different charge51
. As such, more recent research has focused around exploring methods
for obtaining selectivity information from ISEs that are not based on the N-E equation.
A matched potential method (MPM) for determining kA,Bpot
independent of the N-E
equation was first described by Gadzekpo et al52
. In this method the selectivity
Craig Alexander
54
coefficient is defined as the ratio of the activities (concentrations) of each species that
give rise to the same change in potential under the same conditions. The potential of a
background solution containing a known concentration of the primary ion (CA) is
measured followed by the addition of a further known concentration of the primary ion
(CA’); the resulting change in potential is then recorded. A solution of the interfering ion
(CB) is added to an identical background solution until the same change in the measured
potential is observed. kA,Bpot
is calculated according to Equation 1.25.
𝐤𝐀,𝐁
𝐩𝐨𝐭=
(𝐂𝐀 , − 𝐂𝐀)
𝐜𝐁 Equation 1.25
Despite numerous publications spanning almost 40 years addressing the topic of
potentiometric selectivity coefficient determination, it appears that there has yet to be a
fully definitive method that can be used in all cases. The revised IUPAC guidelines state
that methods based on the N-E equation are acceptable when both the target and
interfering species are of equal charge and produce Nernstian responses, whereas the
MPM is recommended if either or both of these criteria are not fulfilled48
. Attempts at
producing alternative methods have continued since the production of these
recommendations50,53,54
. When developing ion-selective sensors, the analyst must
choose a method that is the most appropriate for determining whether the sensor
provides adequate selectivity against the likely interfering ions found in the test sample
so it can be used with confidence for its intended purpose.
Craig Alexander
55
1.4.3 Polymer Membrane Electrodes
The conventional design of an ISE is often referred to as a polymer membrane electrode
(PME). In this design, an internal Ag/AgCl reference electrode is immersed into an
aqueous electrolyte, referred to as the inner filling solution, which is separated from the
sample by the ion-selective membrane55
. To produce a polymeric membrane that is
suitable for such a construction, the plasticised PVC-based ‘cocktail’ described in
section 1.4.1 is cast onto a flat surface, such as a glass petri-dish, to allow evaporation
of the solvent56
. The membrane that is produced is robust and dry to touch, but is
effectively a liquid due to the excess of added plasticiser. Once dry, an
appropriately-sized disc can simply be cut out and attached to one end of an inert
electrode body. The inner filling solution will generally contain the ion of interest at a
fixed dilute concentration, along with a Cl- salt to maintain a stable potential
39. This
sensing electrode and the chosen external reference electrode, as described in section
1.3.1.2, are connected to a suitable voltmeter and submerged into the test solution. A
diagram of the typical experimental configuration is shown in Figure 1.9.
Figure 1.9: Schematic diagram of a typical PME set-up, showing the cell potential of an ISE being
measured against a SCE reference57
Craig Alexander
56
1.4.4 Miniaturisation of ISEs
Due to aspects of its design, such as the requirement for an inner filling solution, the
standard PME is limited in terms of further miniaturisation. Miniaturisation of ISEs is
seen as beneficial where the intended use of the sensor is in a flow cell device or
micro-total analysis system (µ-TAS) for single or multi-ion sensing58
. Attempts at
miniaturisation have mainly been focused around replacing the liquid inner filling
solution/membrane interface with a solid-contact.
1.4.4.1. Coated-Wire Electrodes
One of the first attempts at miniaturising traditional ISEs was the development of the
coated-wire electrode (CWE). CWEs are produced by depositing the selective
polymeric material directly over the surface of the indicator electrode, thus eliminating
the need for an inner filling solution. These were first described by Freiser et al59,60
. The
membrane material is usually the same composition as is used in PMEs; however the
polymer is cast onto the surface of the electrode via dip-coating of the viscous solution.
The solvent evaporates from the surface, leaving behind a thin, polymeric ion-selective
film61
.
Various conductive materials have been used as the internal wire electrode, including
platinum, copper, silver and graphite62
. Numerous publications have described PVC
CWEs for the determination of a variety of ionic species, including cations such as
copper63
, silver64
, lead65
and palladium66
, as well as anions such as nitrate67
.
As well as being much easier to miniaturise than traditional PMEs, CWEs provide a
simpler and inexpensive route to producing ISEs. Although they have been shown to
Craig Alexander
57
perform comparatively well in terms of analytical characteristics, they do have
limitations with regards to drift and reproducibility33
.
1.4.4.2. Screen-Printed ISEs
Screen-printing technology provides a manufacturing route for the mass-production of
cost-effective, highly reproducible chemical sensors68
. The fabrication technique of
screen-printing for the production of thick-film chemical microsensors will be discussed
later in section 1.5.4.2. This section will focus on some of the research that has been
published on the development of screen-printed ISEs (SP-ISEs) as a means for
producing miniaturised, all-solid-state, potentiometric ion-selective devices.
Several SP-ISEs have been described in the literature. Some of these utilise a
screen-printed transducer coupled to a traditional ion-selective membrane, whereas
others have attempted to produce fully screen-printed devices that also incorporate a
screen-printed membrane layer. For a screen-printable membrane to be produced, the
selective components of the membrane (ionophore, plasticiser, additive) must be
incorporated into a suitable support matrix as conventional ion-selective membrane
solvents are not compatible with this technique due to their volatility69
.
Meyerhoff et al. examined the use of alternative membrane materials with improved
adhesion to silicon substrates for use within planar, all-solid-state ISEs. These were a
polyurethane (PU)-based matrix and a silicone rubber-based matrix, both of which
offered greater adhesion to the substrate than PVC without compromising the
electrochemical performance70
. This research eventually led to the production of a
selective polymer membrane matrix that was suitable for screen-printing by using
alternative, high boiling-point solvents as opposed to THF69,71
. SP-ISEs for K+, Ca
2+,
Craig Alexander
58
NH4+, Na
+, pH
71, and Cl
- 69 were prepared using this method, all of which showed
Nernstian responses with high reproducibility. The main disadvantage of this method is
the slightly more complicated membrane fabrication procedure involved. THF is still
required during the preparation of the membrane ‘cocktail’ to dissolve the selective
components, then the solvent which is suitable for screen-printing is added, before the
THF is allowed to evaporate.
Screen-printing of miniaturised, potentiometric transducers followed by manual
deposition of the polymeric, ion-selective membrane have been described by Walsh et
al.72
, Zielinska et al.73
and Musa et al.74
.
Koncki et al. produced fully screen-printed ISEs with excellent response characteristics
by incorporating the selective components into a commercially-available dielectric
screen-printing paste to produce the ion-selective membrane. In this case, the use of low
temperature inks also ensured that there is no degradation of any of the membrane
materials during the printing process. SP-ISEs for NH4+, K
+ and NO3
- were developed
using this method75,76
.
As miniaturised SP-ISEs still require the use of a suitable reference electrode to
measure the potential against, screen-printing technology has been utilised to produce
miniaturised reference electrodes. Koncki et al. also developed an all-solid-state, fully
screen-printed Ag/AgCl reference electrode77
. This could easily be combined with a
SP-ISE to produce a fully screen-printed potentiometric cell for use within a particular
analysis.
As screen-printing is potentially a low-cost, mass-production fabrication technique, it is
particularly suitable when the intended use of the developed product is for the
commercial market68
. Sensor designs can be printed onto inexpensive, flexible polymer
Craig Alexander
59
substrates which allows them to be used disposably76
; the sensing element could simply
be replaced if any degradation in the response was observed. This is a particularly
attractive feature for an aquarium sensor that is to be operated by a non-specialist user.
The sensors would not require an exceptionally long working lifetime, and the effects of
biofouling would effectively be negated. Therefore, the user would not be required to
carry out any regular cleaning, maintenance or calibration of the sensing device but
simply to remove and replace the disposable sensors at regular intervals.
Other attempts at miniaturising potentiometric ISEs have included using conducting
polymers (CP)78,79
and ion-selective field effect transistors (ISFETs)80
.
1.4.5 Sol-gels
Although it is still the most commonly used material, PVC does have several drawbacks
for use within polymeric selective membranes containing an ionophore. The operational
lifetime of a PVC-based ISE is often dictated by leaching of the plasticisers and
ionophores from within the membrane matrix into the test sample, causing the sensor to
lose its selectivity81,82
. Some of the plasticisers often used in PVC membranes also have
a highly toxic effect on aquatic organisms that would render their use within an
aquarium sensor particularly unsuitable83
. As such, alternative membrane materials have
been investigated, including sol-gels.
1.4.5.1. Sol-gel Chemistry
The sol-gel process provides a simple route to preparing hybrid inorganic-organic
materials with a wide range of potential applications84
. Sol-gels are prepared through
the hydrolysis and then condensation of metal alkoxides. One of the most common
examples is the preparation of silica gels from the sol-gel polymerisation of silicon
Craig Alexander
60
alkoxides85
. The typical reaction scheme proceeds according to Equation 1.26
(hydrolysis) and Equation 1.27 (water or alcohol condensation). The reaction takes
place in a mutual solvent such as ethanol in the presence of either an acid or base as the
catalyst84
.
𝐒𝐢(𝐎𝐑)𝟒 + 𝐧𝐇𝟐𝐎 → 𝐒𝐢(𝐎𝐑)𝟒−𝐧(𝐎𝐇)𝐧 + 𝐧𝐑𝐎𝐇 Equation 1.26
𝐒𝐢(𝐎𝐑)𝟒−𝐧(𝐎𝐇)𝐧 + 𝐒𝐢(𝐎𝐑)𝟒−𝐧(𝐎𝐇)𝐧 →
𝐒𝐢(𝐎𝐑)𝟒−𝐧(𝐎𝐇)𝐧−𝟏𝐎𝐒𝐢(𝐎𝐇)𝐧−𝟏(𝐎𝐑)𝟒−𝐧 + 𝐇𝟐𝐎 𝐨𝐫 𝐑𝐎𝐇 Equation 1.27
The viscosity of the resulting solution increases gradually as the sol-gel forms creating a
rigid, porous network. The sol-gel can be fully dried through evaporation of the
remaining alcohol and water to form a xerogel, allowing the simple preparation of
glass-like thin films86
. Numerous silicon alkoxide precursors can be used to produce
silica gels with different properties, some examples are shown in Figure 1.10.
Figure 1.10: Structures of three typical silicon alkoxide sol-gel precursors (A) methyltriethoxysilane
(MTES), (B) tetraethoxysilane (TEOS), (C) diethoxydimethylsilane (DEDMS)
Craig Alexander
61
The relatively mild reaction conditions required to produce sol-gels allows for the
incorporation of selective molecules into the matrix. This can be done either through
physical entrapment, by simply mixing them with the sol prior to its gelation, or by
covalently binding the receptor to functional groups within the silicate matrix to prevent
leaching into the sample86
. This has resulted in research being carried out into the use of
sol-gels for several chemical analysis applications, including as ion-selective
membranes within chemical sensors.
1.4.5.2. Sol-gel Ion-Selective Membranes
Kimura et al. first described utilising a sol-gel as an ion-selective membrane, by
incorporating the respective ionophores into a sol-gel glass based on a 3:1 ratio of the
precursors DEDMS and TEOS, for use within ISFETs for potassium and sodium
monitoring. Initially, just TEOS was used but this was found to produce a brittle
membrane; however, an increased amount of DEDMS in the mixture afforded a more
stable membrane which showed a highly selective response to both cations87-89
. These
membranes were further developed by covalently attaching the receptor molecules
within the sol-gel to improve the lifetime of a sensor90
. A similar membrane
composition was shown to be suitable for use within a conventional PME when an
appropriate support at the tip of a commercial ISE body was employed91
. Kimura et al.
have since used similar sol-gel compositions to produce chloride92
and calcium93
ISFETs, and modified the surface of a conventional pH glass electrode with a selective
sol-gel for Na+ and Cl
- sensing
94. A different research group employed a
DEDMS/TEOS-based membrane to produce a sulfate ISFET95
. Mixtures of TEOS and
DEDMS have become the most popular precursors for producing ion-selective sol-gel
membranes. This composition has since been used to prepare ISEs for copper96
, lead97
,
Craig Alexander
62
vanadyl98
, strontium99
and lanthanum100
ions where the viscous sol-gel solution was
coated over the surface of a graphite electrode.
Santos et al. tested three different sol-gel prescursors: MTES, TEOS and
tetramethoxysilane (TMOS), for their suitability as ion-selective membrane materials. It
was found that TEOS and TMOS produced brittle membranes, however MTES, when
mixed with a small quantity of polyethylene glycol (PEG) 6000, afforded a suitable
membrane that could be simply dip-coated onto the surface of a graphite electrode101
.
This sol-gel membrane was successfully employed for use within valproate-101
and
ampicillinate-102
selective electrodes for pharmaceutical applications.
Alternative sol-gel ion-selective membranes have been prepared by using
methacryloxypropyltrimethoxysilane (MAPTMS) and TMOS with immobilised
crown-ether ionophores103
, and by reacting either 1,4-butanediol or ethylene glycol with
(3-isocyanatopropyl)triethoxysilane to produce a sol-gel that has been used within
chloride-104, 105
, carbonate-106
and nitrite-107
selective electrodes. These membranes were
cast onto a flat surface to produce a membrane suitable for use within PMEs.
Craig Alexander
63
1.4.6 ISEs for Aquarium-Significant Analytes
A vast amount of research has been carried out over the past several decades into
producing ISEs with optimised performance characteristics for a wide range of ions in
various analytical situations. This includes numerous publications discussing sensors
utilising ionophore-doped polymeric membranes for the determination of the four
previously discussed key ionic analytes in freshwater aquaria. The purpose of this
section is to identify some of the ionophores for these four ions that have been quoted in
the literature, and hence could be used in the development of an ionophore-based
sensing device for use within freshwater aquaria.
1.4.6.1. pH
Arguably the most established and widely-used chemical sensor is the glass pH
electrode. Due to several reasons, such as its fragility and unsuitability for
miniaturisation, research has been carried out into producing pH sensors based on
polymeric membranes containing H+-selective ionophores
108. Examples of ionophores
for pH determination include, amongst others; tridodecylamine109
,
methyldioctadecylamine110
dioctylaniline111
, azobenzene derivatives112
, phenoxazine
derivatives113, 114
, alkyldibenzylamines115
, calix[4]arenes108, 116, 117
,
tetrabenzylalkylenediamines118
and metalloporphyrins119
. Table 1.4 provides a summary
of some of the key response characteristics from some of these pH ISEs.
Craig Alexander
64
log
𝐤𝐇
+,𝐂
𝐚𝟐
+𝐩
𝐨𝐭
-11.1
(FIM
)
-10.6
(FIM
)
-11.7
(FIM
)
-11.2
(FIM
)
-11.3
(FIM
)
Not
stat
ed
-11.1
(FIM
)
log
𝐤𝐇
+,𝐍
𝐚+
𝐩𝐨
𝐭
-10.4
(FIM
)
-10.3
(FIM
)
-12.3
(FIM
)
-10.9
(FIM
)
-9.8
(FIM
)
-10.3
(FIM
)
-11.1
(FIM
)
log
𝐤
𝐇+
,𝐊+
𝐩𝐨
𝐭
-9.8
(FIM
)
-10.0
(FIM
)
-10.8
(FIM
)
-10.5
(FIM
)
-9.9
(FIM
)
-8.3
(FIM
)
-10.4
(FIM
)
Res
po
nse
tim
e (s
)
0.4
0.5
0.8
< 1
0
50–60
15
50–60
Slo
pe
(mV
/pH
)
57.8
58.4
57.4
58.2
56.5
54.2
57.3
pH
ran
ge
4.5
–11.0
3.0
–11.0
1.7
–13.2
4.0
–12.0
2.0
–10.0
2.0
–11.0
1.5
–9.0
Ele
ctro
de
typ
e PM
E
PM
E
PM
E
PM
E
PM
E
PM
E
PM
E
Mem
bra
ne
ma
teri
al
PV
C
PV
C
PV
C
PV
C
PV
C
PV
C
PV
C
Ion
op
ho
re
Tri
dod
ecyla
min
e109
Met
hy
ldio
ctad
ecyla
min
e110
4,4
’-bis
[(N
,N-
dio
cty
lam
ino
)met
hyl]
azob
enze
ne
112
9-(
die
thyla
min
o)-
5-
oct
adec
ano
yli
min
o-5
H-
ben
zo[a
]ph
eno
xaz
ine
(ET
H 5
29
4)1
14
Oct
yld
iben
zyla
min
e115
p-t
ert-
buty
lcal
ix[4
]are
ne-
ox
acro
wn
-4108
N,N,N’,N’-
tetr
aben
zylm
ethy
len
edia
min
e118
Table 1.4: Summary of some of the performance characteristics of H+-selective ionophores used
within pH-ISEs that have been described in the literature
Craig Alexander
65
1.4.6.2. Ammonia/Ammonium
The biological and environmental importance of dissolved ammonia in various samples
has led to significant research being carried out into its real-time determination. Air-gap
electrodes for the direct analysis of dissolved NH3 have been developed, which involve
measuring the change in pH of an internal electrolyte caused by the diffusion of NH3
through a gas-permeable membrane120
, as have ionophore-based sensors for
NH4+ sensing.
The natural antibiotic nonactin is one of the earliest examples of a neutral ionophore for
use within ISEs, and provides an excellent response to ammonium121
. Although very
useful for NH4+ determination, nonactin has shown poor selectivity, particularly against
interfering K+ and Na
+ ions
122. As such, numerous publications have attempted to
produce alternative synthetic ionophores with improved selectivity; these include
19-crown-6 derivatives123, 124
, thiazole-containing benzo-crown ethers125
,
trisphenoxy-2,4,6-triethylbenzenes126
, THF-containing crown ethers127
and a
15-crown-5-functionalised carbosilane dendrimer128
. Some of the key performance
characteristics of these ionophores are summarised in Table 1.5.
Craig Alexander
66
log
𝐤𝐍
𝐇𝟒
+,𝐂
𝐚𝟐
+𝐩
𝐨𝐭
-2.6
2
(FIM
)
~ -
4.5
(SS
M)
-2.6
(SS
M)
-5.2
(MP
M)
~ -
2.5
(SS
M)
-2.7
3
(FIM
)
log
𝐤𝐍
𝐇𝟒
+,𝐌
𝐠𝟐
+𝐩
𝐨𝐭
-4.2
(FIM
)
~ -
3.8
(SS
M)
-4.5
(SS
M)
-4.7
(MP
M)
~ -
1.9
(SS
M)
-2.9
(FIM
)
log
𝐤𝐍
𝐇𝟒
+,𝐍
𝐚+
𝐩𝐨
𝐭
-2.1
4
(FIM
)
-3.5
2
(SS
M)
-3.7
(SS
M)
-2.8
(MP
M)
~ -
0.8
(SS
M)
-2.8
5
(FIM
)
log
𝐤𝐍
𝐇𝟒
+,𝐊
+𝐩
𝐨𝐭
-0.3
8
(FIM
-1.0
0
(SS
M)
-1.3
(SS
M)
-1.0
(MP
M)
-1.8
4
(SS
M)
-2.8
(FIM
)
Res
po
nse
tim
e (s
)
0.2
Not
stat
ed
Not
stat
ed
Not
stat
ed
30
6
Slo
pe
(mV
/dec
)
57.0
58.1
59.4
55.1
53.9
53.3
LO
D
(mo
l d
m-3
)
5 x
10
-6
5 x
10
-6
3 x
10
-6
2.5
x 1
0-6
1.5
8 x
10
-5
3.9
x 1
0-6
Ele
ctro
de
typ
e PM
E
PM
E
PM
E
PM
E
PM
E
PM
E
Mem
bra
ne
ma
teri
al
PV
C
PV
C
PV
C
PV
C
PV
C
PV
C
Ion
op
ho
re
Non
acti
n1
21
TD
19
C6
123
TD
B18
C6
125
1,3
,5-
Tri
s(b
enzy
lox
yp
hen
ox
ym
ethy
l)
-2,4
,6,-
trie
thylb
enze
ne1
26
1,4
,6,9
,11
,14
,16
,19
-
tetr
aoxo
cycl
oei
cosa
ne
127
15
-cro
wn
-5-f
un
ctio
nal
ised
carb
osi
lan
e d
end
rim
er128
Table 1.5: Summary of some of the performance characteristics of NH4+-selective ionophores
Craig Alexander
67
1.4.6.3. Nitrite
The development of receptors for use within ISEs has generally been more problematic
for anions than for cations. This is due to several factors, for example, anions are
commonly larger than cations, have more geometrical forms and are more susceptible to
changes in pH129
. When using a hydrophobic polymeric membrane as the recognition
element, it is much easier to sense some anions than others due to their relative
lipophilicity, i.e. more lypophilic anions will diffuse into the membrane phase more
readily than less liphophilic anions. The selectivity of anion-selective electrodes could
therefore be predicted according to the Hofmeister series, some of which is shown in
Figure 1.11.
𝐂𝐥𝐎𝟒− > 𝐈− > 𝐒𝐂𝐍− > 𝐍𝐎𝟑
− > 𝐂𝐥𝐎𝟑− > 𝐁𝐫− > 𝐍𝐎𝟐
− > 𝐂𝐥−
Figure 1.11: Selection of the Hofmeister series showing the order of anion hydrophobicity130
Research into producing anion-selective electrodes has therefore focused around
producing anti-Hofmeister responses, by producing ionophores that shift the target
anion further up the series, shown above, whilst discriminating between anions with
similar characteristics.
Several different ionophores with excellent nitrite selectivity have been reported. These
have mostly been cobalt complexes, including derivatives of Vitamin B12131-133
,
tetraphenylporphyrinato cobalt (III) acetate134
, Co(II)-salen135
and Co(II)-salophen136
.
Alternative nitrite ionophores have been described utilising indium porphyrins130
,
uranyl salophen derivatives137
and rhodium porphyrin and salophen complexes138
. A
summary of the performance characteristics of some of these ionophores is provided in
Table 1.6.
Craig Alexander
68
log
𝐤𝐍
𝐎𝟐
−,𝐂
𝐥−𝐩
𝐨𝐭
-4.6
(FIM
)
-3.9
-3.6
(MP
M)
-2.1
(SS
M)
Not
stat
ed
-4.4
6
(MP
M)
Not
stat
ed
-2.0
log
𝐤𝐍
𝐎𝟐
−,𝐍
𝐎𝟑−
𝐩𝐨
𝐭
-4.2
(FIM
-4.6
-4.3
(MP
M)
-2.9
(SS
M)
-2.4
(FIM
)
-4.0
(MP
M)
-3.4
1
(MP
M)
-2.1
Res
po
nse
tim
e (s
)
3.9
15
30
Not
stat
ed
Not
stat
ed
10
10
Not
stat
ed
Slo
pe
(mV
/dec
)
-57.3
-60.3
-60.3
-51.0
-56.2
-58.2
-59.8
-53.0
LO
D
(mo
l d
m-3
)
2.5
x 1
0-5
2 x
10
-8
8 x
10
-7
5 x
10
-6
Not
stat
ed
5 x
10
-7
8 x
10
-7
Not
stat
ed
Ele
ctro
de
typ
e PM
E
CW
E
(gra
phit
e)
PM
E
PM
E
PM
E
PM
E
PM
E
PM
E
Mem
bra
ne
ma
teri
al
PV
C
PV
C
PV
C
PV
C
PV
C
PV
C
Sol-
gel
Ion
op
ho
re
Aqu
acy
ano
cob
alt(
III)
-
hep
ta(2
-ph
enyle
thyl)
-
coby
rin
ate1
32
Tet
rap
hen
ylp
orp
hy
rin
ato
cob
alt(
III)
ace
tate
(p-i
sop
rop
yl
der
ivat
ive)
134
Ch
loro
(2-n
itro
-5,1
0,1
5,2
0-
tetr
aph
enylp
orp
hy
rin
ato
)
indiu
m130
Ura
ny
l sa
lop
hen
(4-n
itro
der
ivat
ive)
137
Co
(II)
-sal
en135
Co
(II)
-sal
op
hen
136
Ch
loro
-(5
,10
,15
,20
-
tetr
aph
enylp
orp
hy
rin
ato
)
cob
alt
(III
)107
Table 1.6: Summary of some of the performance characteristics of NO2--selective ionophores
Craig Alexander
69
1.4.6.4. Nitrate
The accurate determination of nitrate is very important in a number of biological,
toxicological and environmental situations139
. As such, a vast number of publications
spanning several decades appear in the literature relating to nitrate sensor development.
This includes numerous attempts at producing selective nitrate ionophores.
Some examples of nitrate ionophores that have been used within ISEs include;
quaternary phosphonium salts140
, quaternary ammonium salts141-143
, a tetra-coordinate
nickel(II) complex144
, bis(2-hydroxyanil)acetylacetone lead(II)145
, a tris(2-
aminoethyl)amine triamide derivative146
, bis(2-hydroxyacetophenone)ethylenediimine
vanadyl (IV)147
, meso-tetrakis[(2-arylphenylurea)-phenyl]porphyrins148
, a tetramethyl
cyclotetra-decanato-nickel(II) complex67
, [3.3.3.3]oxazane149
, Zn(II) complexes150
,
urea-calix[4]arenes151
and a cyclic bis-thiourea152
. The response characteristics of a
selection of nitrate ionophores can be found in Table 1.7.
Craig Alexander
70
log
𝐤𝐍
𝐎𝟑
−,𝐂
𝐥−𝐩
𝐨𝐭
-1.7
(FIM
)
-2.2
(FIM
)
-2.0
(FIM
)
-1.9
(SS
M)
-3.6
(FIM
)
-4.3
(FIM
)
-3.5
(FIM
)
log
𝐤𝐍
𝐎𝟑
−,𝐍
𝐎𝟐−
𝐩𝐨
𝐭
-2.9
(FIM
-2.0
(FIM
)
-1.8
(FIM
)
-1.6
(SS
M)
Not
stat
ed
-4.5
(FIM
)
-1.8
(FIM
)
Res
po
nse
tim
e (s
)
25
10
10
10
25
15
70
Slo
pe
(mV
/dec
)
-51.9
-59.6
-58.8
-54.7
-58.5
-57.8
-47.8
LO
D
(mo
l d
m-3
)
1 x
10
-5
2.5
x 1
0-6
1 x
10
-5
2.3
x 1
0-6
1 x
10
-6
5 x
10
-6
5.6
x 1
0-6
Ele
ctro
de
typ
e ISF
ET
PM
E
PM
E
PM
E
PM
E
CW
E
(pla
tinum
)
PM
E
Mem
bra
ne
ma
teri
al
PV
C
PV
C
PV
C
PV
C
PV
C
PV
C
PV
C
Ion
op
ho
re
Tet
rad
od
ecyla
mm
on
ium
nit
rate
142
5,7
,12
,14
-tet
ram
ethy
l-1
,4,8
,11
-
tetr
aaza
cycl
ote
tra-
dec
a-4
,6,1
1,1
3-
tetr
aen
e n
ick
el(I
I)144
Bis
(2-h
yd
rox
anil
)ace
tyla
ceto
ne
lead
(II)
145
Tri
s[2
-(4
-ter
t-
bu
tylb
enzo
yl)
amin
oet
hyl]
amin
e146
Bis
(2-
hy
roxy
acet
oph
eno
ne)
ethy
len
edii
min
e
van
adyl
(IV
)147
(6,8
,15
,17
-tet
ram
ethy
l-
7H
,16
H,5
,9,1
4,1
8-t
etra
azid
ob
enzo
[b,i
]-
cycl
ote
trad
ecan
ato
-(-2
)K4-
N,N
’,N
’’,N
’’’)
nic
kel
(II)
67
(9,1
1,2
0,2
2-
tetr
ahyd
rote
trab
enzo
[d,f
,k,m
][1
,3,8
,10
]t
etra
azac
ycl
ote
trad
ecin
e-10
-21
-
dit
hio
ne1
52
Table 1.7: Summary of some of the performance characteristics of NO3--selective ionophores
Craig Alexander
71
A review of the literature shows that a multitude of ionophores for the four previously
discussed ions have been synthesised and tested for their possible use within
potentiometric ISEs. Several of these are now available to purchase commercially,
along with relevant polymers, plasticisers, ionic additives and solvents153
. This allows
ionophore-based sensors to be produced without the requirement for specialist
chemical synthesis.
One of the major downfalls of potentiometric sensors is the requirement of a suitable
external reference electrode for them to function. Miniaturised reference electrodes have
been developed77
that can be used to produce a stable potential to measure against
whilst being simple to integrate within an appropriately-sized device for multi-analyte
sensing. However, if the reference electrode was to fail, this would negate the response
from each of the selective indicator electrodes. Reference electrodes are also susceptible
to potential drift over time which could lead to a false response. For the purpose of
home aquarium monitoring, it would be more beneficial to produce a sensor array that
measured the target analyte concentration directly. One method that would allow this is
to combine the recognition element of the sensor with an impedimetric transducer.
Craig Alexander
72
1.5 Impedimetric Interdigitated Electrode Chemical Sensors
The purpose of this section is to discuss the use of interdigitated electrodes (IDEs) as
impedimetric chemical sensors, including the underlying theory of operation and means
of fabrication, with regards to utilising such devices within sensors for aquatic water
quality monitoring.
1.5.1 Electrical Impedance
Electrical impedance (Z) describes the total opposition of an electrical circuit to an
alternating current (AC) when a voltage is applied. Simply, impedance is defined as the
ratio of the AC voltage (VAC) to the AC current (IAC)27
, as shown in Equation 1.28,
where j is the imaginary unit of impedance (√-1). A sinusoidal AC signal is
characterised by its amplitude (Vm or Im), frequency (f) or angular velocity (ω = 2πf) and
phase angle (ϕ). This is represented diagrammatically in Figure 1.12.
𝐙 =
𝐕𝐀𝐂
𝐈𝐀𝐂= |𝐙|𝐞𝐱𝐩(𝒋𝜙) Equation 1.28
Figure 1.12: A typical AC sine wave voltage and current. Adapted from Bănică et al.
27
Craig Alexander
73
The frequency of the sine wave has units of Hertz (Hz) and indicates the number of
cycles per second. The amplitude is often reported either as a peak-to-peak value (2Vm)
or a root mean square value (Vrms = Vm/√2).
Impedance is a complex parameter that is defined by both real and imaginary parts. For
the purpose of this thesis, conductance (G) and capacitance (C) measurements will be
taken to evaluate the sensor response. Conductance is related to impedance according to
Equation 1.29, where R is the resistance, X is the reactance, Y is the admittance
(which is the reciprocal of Z) and B is the susceptance.
𝒁 = 𝐑 + 𝒋𝐗, 𝐘 = 𝐆 + 𝒋𝐁 Equation 1.29
Conductance is measured in the SI unit of siemens (S), the reciprocal of one ohm.
Capacitance is defined as the ability of a body to store electrical charge. The SI unit of
capacitance is the farad (F). The impedance of a capacitor is dependent on the signal
frequency and is related to the reactance as shown in Equation 1.30.
𝐗 = (𝟐𝛑𝒇𝐂)−𝟏 Equation 1.30
Electrochemical impedance spectroscopy (EIS) is a powerful analytical tool that can be
used to provide information on the physical and chemical properties of a material. One
very useful application of EIS is as a transduction method for chemical sensors and
biosensors, by monitoring changes in the electrical properties that occur due to
interactions between a receptor and the target analyte154
. Electrical impedance
measurements are generally carried out by applying an AC signal between two
electrodes. A spectrum is obtained by performing these measurements at a number of
frequencies over a particular range.
Craig Alexander
74
One popular method that is often employed to obtain impedance measurements for
sensor applications is to use a pair of IDEs as the transducer.
1.5.2 Interdigitated Electrodes
IDEs consist of two co-planar electrodes with interlocking digits, forming a ‘comb-like’
structure, on an insulating substrate. A schematic representation of a typical IDE design
is shown in Figure 1.13.
Figure 1.13: Schematic diagram of a typical IDE design, reproduced from Mosaic Industries
155
An IDE is generally characterised by its geometric parameters including the width of
the individual electrode digits (W), the distance of the digit separation (S), the length of
the digits (L) and the number of digits at each electrode (N)154
. When excited with an
AC voltage, an electric field is generated between the adjacent electrode digits of
opposing charge156
. The characteristic penetration depth of the resulting electric field is
dictated by the IDE geometry. It is generally accepted that the majority of the electric
field is enclosed within a distance perpendicular to the surface of the electrodes equal to
that between the centres of adjacent digits, i.e. W + S 154
. This is illustrated in
Figure 1.14.
Craig Alexander
75
Figure 1.14: Cross-section schematic representation of the electric field penetration depth from a
typical IDE when an AC voltage is applied. Adapted from Bratov et al.157
IDEs are utilised in many different ways for a variety of purposes, such as in
telecommunications and biotechnology158
; however, it is their potential use as
transduction elements for chemical sensing applications that are of particular interest
with respect to the experimental work in this thesis.
1.5.3 IDEs for Chemical Sensing Applications
Numerous chemical sensors and biosensors have been described that utilise IDE
transducers159
. The sensing mechanism is usually achieved by placing a sensitive
membrane layer over the digits of the IDEs, or by immobilising receptors in the
interdigital spaces. EIS is used to monitor changes in the electrical properties due to
selective interactions between the receptors and the target analyte.
1.5.3.1. Ion-Selective Conductometric Microsensors
One IDE-based chemical sensor of particular relevance is the ion-selective
conductometric microsensor (ISCOM), first described by Cammann et al.156, 160
. An
ISCOM is produced by coating an IDE transducer with a thin polymeric membrane
containing an ionophore, as used in conventional ISEs. The IDE is used to measure the
bulk conductance (G) of the ion-selective membrane, where the magnitude of G relates
directly to the concentration of the target ion in the test solution156
. The main advantage
Craig Alexander
76
over a conventional ISE is that no external reference electrode is required for the
measurement. Camman et al. have described ISCOMs for the determination of K+, Ca
2+,
NH4+, Li
+ and pH
160. Jaffrezic-Renault et al. produced a NH4
+-selective conductometric
microsensor by depositing a nonactin-based PVC membrane over an IDE transducer161
,
and Port et al. produced ISCOMs for the environmental monitoring of caesium162
.
All ISCOMs that have been reported have appeared to perform comparably in term of
analytical characteristics when compared with their potentiometric counterparts. A
schematic diagram of an ISCOM is shown in Figure 1.15.
Figure 1.15: Schematic representation of an ISCOM
162
As well as not requiring an external reference electrode, IDE-based sensors have the
advantage of being easily miniaturised and mass-produced using microelectronic
fabrication techniques.
1.5.4 IDE Fabrication Techniques
The production of high-quality sensors with micrometer-sized features, such as the
digits of an IDE, requires an appropriate fabrication technique. Two common
microelectronic fabrication techniques that are often utilised for the production of
miniaturised chemical sensors are photolithographic methods and screen-printing.
Craig Alexander
77
1.5.4.1. Photolithography
Sensor fabrication using photolithography involves transferring the desired electrode
pattern into a photoresist layer by irradiating the substrate with ultraviolet (UV) light
through a suitable photomask stencil (although ‘mask-less’ methods are available as
alternatives)154
, to produce thin-film electrodes. This can be carried out either by
chemical etching or lift-off techniques. The processes of chemical etching (a) and lift-
off (b) are illustrated in Figure 1.16.
Chemical etching first involves the deposition of a thin-layer of the required metal, by
techniques such as vacuum evaporation or sputtering, onto a suitable substrate, which is
usually a glass or ceramic material. A photoresist chemical is then coated over the
surface of the metal, usually by way of spin-coating. A photomask is placed over the
substrate and exposed to UV light, which is then developed in a suitable solvent,
leaving behind the desired electrode pattern in non-UV exposed photoresist. The
uncoated metal layer is then chemically etched away and the photoresist removed, to
leave only the electrode design in metal163
.
When the lift-off technique is employed, the photoresist chemical is first coated onto the
surface of the substrate and the electrode pattern is UV-irradiated directly into this layer.
Upon development of the photoresist, the electrode pattern is formed directly on the
substrate. The desired metal is then deposited over the surface of the substrate,
following which the remaining photoresist layer can be removed, leaving behind only
the metal electrode design on the substrate163
.
Craig Alexander
78
Figure 1.16: Photolithographic preparation of thin-film electrodes using (a) chemical etching and
(b) lift-off procedures163
Photolithography provides a route to producing IDEs with excellent resolution, allowing
for very small digit widths and interdigital spacings. Photolithographically-prepared
IDEs with nano-scaled features have been described with digit widths and spacings of
several hundred nanometers164
; although, typical dimensions are in the range of
1–10 µm154
. This method for producing IDEs is not particularly cost-effective; and
requires a significant amount of specialist equipment. It is also largely limited to
inflexible substrates such as glass or ceramic, due to the harsh fabrication conditions
that are required, i.e. the use of various chemicals and high temperatures.
Craig Alexander
79
1.5.4.2. Screen-Printing
Screen-printing is a thick-film technology used for the production of planar chemical
sensors. The process of screen-printing involves forcing a thixotropic ink or paste
through a mesh screen, containing a stencil of the desired electrode design, onto a
contacting substrate with a squeegee68
. The composition of the inks can be altered to
produce conductive prints, by using inks which contain carbon-graphite or precious
metals such as gold or platinum, whilst dielectric pastes can be used to cover conductive
tracks and define sensing areas. Often one design will contain several layers, where
each screen design is printed sequentially to produce the final sensor165
. Once printed,
the inks are dried in an oven to remove the residual solvent. A schematic representation
of the screen-printing process is shown in Figure 1.17.
Figure 1.17: Schematic diagram of screen-printing for the fabrication of thick-film
electrochemical sensors68
Screen-printing provides a cost-effective route for high-throughput production of
multi-layered chemical sensors. The main costs involved are the screen production and
the purchase of the relevant inks and substrates, after which a large number of prints can
be performed at a relatively low running cost. Sensor designs can be printed onto
Craig Alexander
80
flexible, polymer substrates, providing that compatible inks are used68
. Therefore
screen-printing is a particularly attractive fabrication technique for disposable
sensing applications.
The main disadvantage of using screen-printing as the chosen fabrication technique,
particularly for the production of IDEs, is the much lower resolution that is achievable
compared with lithographic methods. Screen-printing is limited to a resolution of
around 200 µm154
, whilst the ink may expand slightly on drying which restricts designs
with features that are close together. As such, any impedimetric IDE device that is
produced via screen-printing technology would require a much thicker sensing layer to
contain the resulting electric field.
1.5.5 Membrane Coating Techniques
With the use of planar sensors which contain a membrane layer, such as ISCOMs, a
suitable coating technique must be selected to afford a homogenous and reproducible
deposition of the polymeric material, such as PVC or sol-gel, which forms the ionic
recognition layer of the sensor.
1.5.5.1. Spin-Coating
Spin-coating involves depositing a solution of a polymeric material onto the chosen
substrate and spinning at a high speed. This results in even spreading and evaporation of
the solvent to produce a uniform film166
. The thickness of the resulting film is
dependent on factors such as the viscosity of the solution and the chosen spin-speed.
Craig Alexander
81
1.5.5.2. Dip-Coating
Dip-coating generally consists of immersing the substrate into the polymer solution
followed by its vertical withdrawal, to produce a uniform film as the solvent evaporates.
The thickness of the resulting film depends on the immersion time, solution viscosity
and rate of withdrawal167
.
1.5.5.3. Drop-Coating
Both spin and dip-coating are wasteful techniques with regards to the polymer solution,
and hence the selective receptors which it contains. In spin-coating, a large quantity of
the polymeric solution is dispersed away from the target substrate; whereas with
dip-coating it is deposited onto both sides of a planar substrate, not just the desired area,
i.e. the sensing electrodes. An alternative method is to simply drop-cast the membrane
solution onto the substrate. Here the liquid simply spreads over the surface of the
substrate and the solvent evaporates to leave a thin-film. It is more difficult to control
the uniformity with the drop-casting technique but wastage of the polymer material is
minimised. Drop-coating is also limited to small areas of less than 1 cm2 27
. For coating
a large number of sensors on a commercial scale, it may also be possible for
drop-coating of the polymeric material to be automated by using small-volume liquid
robotic dispensing systems168
.
1.5.5.4. Screen-Printing
As mentioned in section 1.4.4.2, screen-printing has previously been utilised to deposit
the ion-selective membrane layer onto the conductive components of planar ISEs. This
provides a route for the manufacturing of fully screen-printed devices, and allows for a
controlled, accurate and reproducible deposition of the membrane material to produce a
Craig Alexander
82
homogeneous layer. The simplest method for this appears to be by incorporating the
selective components and plasticisers into a conventional, low-temperature curing,
dielectric screen-printing paste, as described by Koncki et al.75, 76
.
1.6 Aims and Objectives
The overall aim of this multi-disciplinary project was to develop low cost, easy to
fabricate ion-selective sensors for the determination of four aquarium-significant ionic
analytes (H+, NH4
+, NO2
- and NO3
-) with a view to producing a prototype commercial
device for use within artificial freshwater aquaria. The project aimed to create a novel,
multi-analyte sensing device by bringing together aspects of some established
electroanalytical chemistry techniques, such as ISEs, with some more recent
developments in chemical sensors, for example interdigitated impedimetric
microsensors, and micro-fabrication procedures such as screen-printing and
photolithography.
Craig Alexander
83
2 Materials and Methods
2.1 Preparation of Ion-Selective Membranes
2.1.1 Materials
The ionophores used throughout this work were: tetradodecylammonium nitrate
(TDAN), Nitrate Ionophore V (NO3V), Ammonium Ionophore I (NH4I), Hydrogen
Ionophore III (HIII), Hydrogen Ionophore V (HV) and Nitrite Ionophore I (NO2I).
These, along with high molecular weight PVC (viscosity 3.5–5.0 mPa.s), THF (reagent
grade ≥ 99.9%, containing 250 ppm butylated hydroxytoluene (BHT) as inhibitor), the
plasticiser o-nitrophenyloctyl ether (NPOE) and the ionic additives:
tridodecylmethylammonium chloride (TDMACl), sodium tetraphenylborate,
sodium tetrakis (4-fluorophenyl) borate dihydrate and potassium tetrakis
[3,5-bis(trifluoromethyl)phenyl] borate were purchased from Fluka’s Selectophore™
range for sensoric applications (Sigma Aldrich, Gillingham, UK). Bis (2-ethylhexyl)
sebacate (97%), also used as a plasticiser, was purchased from Acros Organics (Fisher
Scientific, Loughborough, UK). The ionic additive potassium tetrakis (4-chlorophenyl)
borate (98%) was purchased from Alfa Aesar (Johnson Matthey, Heysham, UK).
The sol-gel precursors, MTES (99%), DEDMS (97%) and TEOS (98%) were purchased
from Sigma Aldrich, as was PEG 600. Ethanol, used as the solvent for sol-gel
preparation, was analytical reagent grade and purchased from Fisher Scientific. 1.0 M
volumetric standard grade HCl was purchased from Sigma Aldrich, and was diluted to
0.1 M using deionised water (18.2 MΩ.cm), dispensed from an Elga Maxima system
(High Wycombe, UK), prior to use.
Dielectric screen-printing paste (Gwent Electronic Materials, Pontypool, UK) was
provided by Dr. Craig Banks’ research group at Manchester Metropolitan University.
Craig Alexander
84
2.1.2 Polyvinyl Chloride Membranes
PVC ion-selective membrane ‘cocktails’ were prepared according to previously
reported methods110, 117, 121, 132, 142, 152
. A typical PVC membrane composition contained
ionophore (1% w/w), high molecular weight PVC (31–33% w/w), plasticiser
(64–68% w/w) and, if required, a cationic or anionic additive at a molar ratio of
approximately 1:2, with respect to the ionophore (0.2–0.6% w/w). The combined mass
of the membrane components, totalling 500 mg, was dissolved into 5 ml THF to
produce a 10% w/v solution. Compositions of each component within specific
ion-selective membranes for each target ion can be found in Table 2.1. Details of the
chemical structures and IUPAC names of each of the commercial ionophores used can
be found in Appendix 10.2.
Target
Ion Ionophore Plasticiser Additive PVC
NO3-
5 mg tetradodecylammonium
nitrate
(TDAN)
330 mg bis (2-ethylhexyl)
sebacate N/A 165 mg
NO3-
5 mg Nitrate Ionophore V
(NO3V)
328 mg o-nitrophenyloctyl
ether (NPOE)
3 mg
tridodecylmethylammonium
chloride (TDMACl)
164 mg
NH4+
5 mg Ammonium Ionophore I
(NH4I) 338 mg NPOE
2 mg potassium tetrakis (4-
chlorophenyl) borate 155 mg
H+ 5 mg Hydrogen Ionophore III
(HI)
335 mg bis (2-ethylhexyl)
sebacate 2 mg sodium tetraphenylborate 158 mg
H+
5 mg Hydrogen Ionophore V
(HV) 330 mg NPOE
1 mg sodium tetrakis (4-
fluorophenyl) borate dihydrate 164 mg
NO2- 5 mg Nitrite Ionophore I
(NO2I) 328 mg NPOE
2 mg potassium tetrakis [3,5-
bis(trifluoromethyl)phenyl]
borate
165 mg
Table 2.1: Composition of each component required to produce PVC ion-selective membranes
containing 1% w/w ionophore. Each membrane was produced with a total mass of 500 mg
Craig Alexander
85
2.1.3 Sol-Gel Membranes
Two different sol-gel ion-selective membranes were prepared, based on ones previously
reported for use within CWEs97, 101
, using the precursor materials of MTES and a
combination of DEDMS and TEOS in a 3:1 ratio. The sol-gel membranes totalled 5.6
ml, and contained 20 mg ionophore along with ionic additive in the same ratio as for
PVC membranes, and were prepared according to the following general procedures.
2.1.3.1. MTES
3.6 ml MTES, 1.12 ml deionised water, 0.8 ml ethanol and 80 µl 0.1 M HCl were mixed
together for 15 minutes. 20 mg ionophore, along with additive (if required) and 40 µl of
PEG 600, to prevent the production of a brittle membrane, were added to the mixture
and stirred overnight.
2.1.3.2. DEDMS/TEOS
2 ml DEDMS, 0.72 ml TEOS, 2.24 ml ethanol and 0.64 ml 0.1 M HCl were mixed
together for 15 minutes. 20 mg ionophore, along with additive (if required) was added
and mixed for a further 10 minutes. The mixture was allowed to stand at 70 °C on a
hotplate for 1 hour prior to being left overnight under stirring.
All PVC and sol-gel membrane ‘cocktails’ were stored in a refrigerator at 4 °C and used
within 3 months of preparation.
Craig Alexander
86
2.1.4 Dielectric Screen-Printing Paste Membrane
A nitrate-selective membrane suitable for screen-printing was prepared by incorporating
ionophore (1% w/w) and plasticiser (22% w/w) into a commercial dielectric screen-
printing paste (77% w/w). 100 mg of the nitrate ionophore TDAN was weighed out and
2.2 g of NPOE was added. This was mixed for 10 minutes followed by the addition of
7.7 g of dielectric paste. NPOE was used as the plasticiser, as bis (2-ethylhexyl)
sebacate was observed to not mix homogenously with the dielectric paste.
All PVC, sol-gel and dielectric paste membrane ‘cocktails’ were also prepared without
ionophore or ionic additive present, as ‘blank’ membranes to be tested as a control.
Craig Alexander
87
2.2 Construction of Ion-Selective Electrodes
2.2.1 Materials
Internal Ag/AgCl reference electrodes for use within PMEs were prepared using
1.0 mm diameter silver wire (99.9%; Sigma Aldrich), which was cleaned using ethanol
and rinsed with deionised water prior to use. Sodium hypochlorite solution
(Sigma Aldrich) was reagent grade with 10–15% available chlorine. Electrode bodies
were fashioned from 13 mm outer diameter, 10 mm inner diameter acrylic tubing (RS
Components, Northamptonshire, UK). Lids for the ISE bodies were fashioned from lids
for volumetric flasks (size 10/19). BNC (F) to bare end leads (Pomona Electronics, WA,
USA) were purchased from Farnell (Leeds, UK).
2.2.2 Polymer Membrane Electrodes
Ion-selective membranes for use within a PME were prepared by casting the PVC
‘cocktail’ described in section 2.1.2 into a 60 mm glass petri dish. The dish was covered
with filter paper and left to stand overnight to allow the evaporation of THF from the
solution. A 12 mm disc of the resulting membrane was cut-out using a cork borer, and
affixed to one end of the electrode body using Glu & Fix all-purpose, clear, extra strong
adhesive (Bostik, Stafford, UK) and allowed to dry overnight. 10 ml of the appropriate
internal filling solution was added to the tube, a lid was attached to the exposed end,
and glued into place. Details of each specific internal filling solution are shown
in Table 2.2.
Craig Alexander
88
Target ion Internal filling solution
NO3- 0.01 M KCl + 0.01 M KNO3
NH4+
0.01 M NH4Cl
H+
pH 7.0 buffer solution* + 0.01 M KCl
Table 2.2: Details of internal filling solutions used within PMEs for each target ion.
*The composition of the pH buffer solution, used as internal filling solution for H+-ISEs, is
described in section 2.3.2.4.1
An internal Ag/AgCl reference electrode was prepared by soaking a 100 mm length of
silver wire in sodium hypochlorite for 15 minutes. One end was soldered onto a lead
connected to the central pin of a female BNC connector and placed inside the ISE body
through a 2 mm hole which had previously been drilled in the lid. The BNC connector
was attached to the measurement instrumentation using a 1 m co-axial BNC lead.
2.2.3 Coated-Wire Electrodes
A 20 mm Ag/AgCl reference electrode was prepared by soaking a silver wire in sodium
hypochlorite for 15 minutes. One end was attached to a BNC lead using a solder sleeve,
and the joint was covered in heat-shrink to leave 10 mm of the electrode exposed.
The ion-selective membrane was coated onto the Ag/AgCl electrode via dip-coating of
either the PVC or sol-gel solution. Each electrode was dipped into the viscous liquid a
total of ten times by hand, allowing several minutes between coatings. Following
dip-coating, PVC CWEs were dried in air at room temperature overnight and sol-gel
CWEs were dried in an oven at 70 °C overnight. All CWEs were rinsed with deionised
water prior to use.
All PMEs and CWEs were conditioned overnight in a 1000 ppm solution of the target
ion prior to testing, with the exception of the pH ISEs, which were conditioned
overnight in a pH 7.0 buffer solution (as described in section 2.3.2.4.1).
Craig Alexander
89
2.3 Potentiometric Measurements
2.3.1 Measurement Instrumentation
An attempt was made to build a bespoke 16-channel system for obtaining
potentiometric measurements using PHTX-22 pH/ORP preamplifiers (Omega,
Manchester, UK), controlled using an in-house written LabVIEW program (National
Instruments, Austin, TX, USA). This system is fully described in Appendix 10.3;
however, there were several irresolvable complications with the instrumentation and
software, therefore the results obtained are not included in this thesis. This system was
replaced with a commercial ELIT 4-channel ion analyser (NICO2000 Ltd., Middlesex,
UK), which was supplied with the relevant software, connected to a PC and used to
obtain all potentiometric data from the ISEs described in this thesis.
2.3.2 Sensor Testing
2.3.2.1. Materials
The salts used to produce stock solutions of target ions and interfering ions were:
potassium nitrate, potassium nitrite, potassium chloride, ammonium chloride, sodium
chloride, magnesium chloride hexahydrate and calcium chloride dehydrate. All of these
were of at least 99.0% reagent grade purity and purchased from various suppliers. Boric
acid (99.99%), citric acid monohydrate (99.5%) and disodium phosphate dihydrate
(Na2HPO4.2H2O) (>99.0%) were used to produce the pH buffer stock solution and were
purchased from Sigma Aldrich. 1.0 M volumetric standard NaOH was purchased from
Fluka and used as received. Ammonium sulfate (99.5% purity; VWR, Leicestershire,
UK) and copper (II) sulfate (>99.0% purity; Sigma Aldrich) salts were used to produce
ISABs. Double-junction silver/silver chloride reference electrodes were purchased from
Spectronic Camspec Ltd. (Leeds, UK). Lithium acetate dihydrate (reagent grade; Sigma
Craig Alexander
90
Aldrich) was used to produce the reference electrode outer filling solution at a
concentration of 0.1 M. All solutions were prepared using ultrapure deionised water.
Potentiometric experiments were conducted at ambient temperature. Temperature
measurements were recorded using an ELIT 8701 temperature probe, which was placed
into the test sample and connected to the 4-channel ion analyser. The prepared
electrodes were tested alongside commercial half-cells for each target ion, to confirm
the validity of the test sample solutions. Ammonium and nitrate-selective half-cells
were purchased from Scientific Laboratory Supplies (Nottingham, UK). A pH electrode
half-cell was purchased from VWR.
2.3.2.2. Nitrate ISEs
2.3.2.2.1. Stock Solutions
A 10,000 ppm NO3- stock solution was prepared by dissolving 16.31 g KNO3 in 1 litre
of deionised water. For selectivity experiments, a 10,000 ppm NO2- stock solution was
prepared by dissolving 18.50 g KNO2 in 1 litre of deionised water, and a 10,000 ppm
Cl- stock solution was prepared by dissolving 21.03 g KCl in 1 litre of deionised water.
Ammonium sulfate (2 M) was used as ISAB, and was prepared by dissolving 66.07 g in
250 ml deionised water.
2.3.2.2.2. Calibrations
Potentiometric nitrate ISEs were calibrated over the range 0.1–10,000 ppm NO3-.
500 ml working standard solutions of 1000, 100, 10, 1 and 0.1 ppm NO3- were prepared
from serial dilutions of the stock solution described in section 2.3.2.2.1, and 10 ml
ISAB was added to each standard solution prior to testing.
Craig Alexander
91
The electrode under test was placed into the first calibration solution, starting with
0.1 ppm NO3-, along with a double-junction Ag/AgCl reference electrode (0.1 M
lithium acetate outer solution) and the temperature sensor. The solution was stirred at a
constant rate. Once a stable measurement was obtained, the electrode was rinsed and
placed into the next calibration solution. This process was repeated until a reading had
been obtained for each of the six NO3- standard solutions. All potentiometric NO3
- ISEs
were prepared in duplicate and tested in parallel to establish reproducibility.
2.3.2.2.3. Selectivity Determination
Selectivity coefficients for nitrate against the interfering ions, chloride and nitrite, were
determined using the FIM. Selectivity data were obtained for nitrite and chloride as
these are seen as the most likely anion interferences in aquaria. The interfering anion
concentrations were fixed at 1000 ppm, and the test solutions were prepared by mixing
appropriate amounts of nitrate stock solution with interfering ion stock solution and
10 ml ISAB. This mixture was diluted to 500 ml with deionised water. The required
volumes of each stock solution are outlined in Table 2.3.
Required nitrate
concentration (ppm)
Volume of 10,000 ppm
NO3- stock solution (ml)
Volume of 10,000 ppm
interfering ion solution (ml)
0.1 0.005 50
1 0.05 50
10 0.5 50
100 5 50
1000 50 50
Table 2.3: Volume of respective stock solutions required to produce 500 ml nitrate calibration
solutions with fixed 1000 ppm interfering ions, for selectivity coefficient determination
The electrode under test was first placed into the solution with the lowest nitrate
concentration, along with a double-junction Ag/AgCl reference electrode and the
temperature sensor. The solution was stirred at a constant rate. Once a stable
Craig Alexander
92
measurement was obtained, the electrode was rinsed and placed into the next solution.
This process was repeated until the ISE had been tested in each of the six NO3- standard
solutions containing the fixed interfering anion.
2.3.2.3. Ammonium ISEs
2.3.2.3.1. Stock Solutions
A 10,000 ppm NH4+ stock solution was prepared by dissolving 29.70 g NH4Cl in 1 litre
of deionised water. For selectivity experiments, 1 litre stock solutions of each
interfering cation (K+, Na
+, Mg
2+ and Ca
2+) were prepared at a concentration of
10,000 ppm by dissolving 19.07 g KCl, 25.42 g NaCl, 83.63 g MgCl2.6H2O and 36.70 g
CaCl2.2H2O in deionised water. Copper (II) sulphate (1 M) was used as ISAB, and was
prepared by dissolving 159.60 g in 1 litre of deionised water.
2.3.2.3.2. Calibrations
Potentiometric ammonium ISEs were calibrated over the range 0.1–10,000 ppm NH4+.
Solutions of 1000, 100, 10, 1 and 0.1 ppm NH4+ were prepared from serial dilutions of
the stock solution described in section 2.3.2.3.1. 50 ml ISAB was added to each
standard solution prior to testing.
The electrode under test was placed into the first calibration solution, starting with
0.1 ppm NH4+, along with a double-junction Ag/AgCl reference electrode (0.1 M
lithium acetate outer solution) and the temperature sensor. The solution was stirred at a
constant rate. Once a stable measurement was obtained, the electrode was rinsed and
placed into the next calibration solution. This process was repeated until the ISE had
been tested in each of the six NH4+ standard solutions. All potentiometric NH4
+ ISEs
were prepared in duplicate and tested in parallel to establish reproducibility.
Craig Alexander
93
2.3.2.3.3. Selectivity Determination
Selectivity coefficients for NH4+ against the interfering ions potassium, sodium,
magnesium and calcium were determined using the FIM. The interfering cation
concentrations were fixed at 1000 ppm, and the test solutions were prepared by mixing
appropriate amounts of ammonium ion stock solution with interfering ion stock solution
and 50 ml ISAB. This mixture was diluted to 500 ml with deionised water. The required
volumes of each stock solution are outlined in Table 2.4.
Required ammonium
concentration (ppm)
Volume of 10,000 ppm
NH4+ stock solution (ml)
Volume of 10,000 ppm
interfering ion solution (ml)
0.1 0.005 50
1 0.05 50
10 0.5 50
100 5 50
1000 50 50
Table 2.4: Volume of respective stock solutions required to produce 500 ml ammonium calibration
solutions with fixed 1000 ppm interfering ions, for selectivity coefficient determination
The electrode under test was first placed into the solution with the lowest ammonium
concentration, along with a double-junction Ag/AgCl reference electrode and the
temperature sensor. The solution was stirred at a constant rate. Once a stable
measurement was obtained, the electrode was rinsed and placed into the next solution.
This process was repeated until the ISE had been tested in each of the six NH4+ standard
solutions containing the fixed interfering cation.
Craig Alexander
94
2.3.2.4. pH ISEs
2.3.2.4.1. Stock Solutions
A ‘universal’ pH buffer, as described by Ostling and Virtama169
, was prepared to test
the response of potentiometric pH-ISEs. A stock solution was prepared by weighing out
8.903 g disodium phosphate dihydrate, 7.00 g citric acid monohydrate and 3.54 g boric
acid into a 1 litre flask. 243 ml of 1 M NaOH was added and then filled to the mark with
deionised water. 1.0 M stock solutions of the interfering cations Na+, K
+ and Mg
2+ were
prepared for selectivity coefficient determination by dissolving the appropriate amount
of each respective chloride salt in 1 litre of deionised water.
2.3.2.4.2. Calibrations
Calibration standards of pH 2.5, 5.0, 7.0, 8.0, 9.0 and 11.5 were prepared by dispensing
50 ml of the buffer stock solution, described in section 2.3.2.4.1, into a 250 ml flask,
adding the required volume of 0.1 M HCl, as stated in Table 2.5, and filling to the mark
with deionised water.
Desired pH Required volume of 0.1 M HCl (ml)
2.5 158.125
5.0 115.450
7.0 83.775
8.0 71.750
9.0 61.200
11.5 26.875
Table 2.5: Volume of 0.1 M HCl required to add to 50 ml stock solution to produce 250 ml pH
buffer calibration solutions
Craig Alexander
95
The pH values of the resulting solutions were checked using a calibrated bench-top pH
meter (pH211 microprocessor pH meter, HANNA Instruments, Bedfordshire, UK) and
adjusted as necessary using either 0.1 M NaOH or 0.1 M HCl.
The electrode under test was placed into the first calibration buffer solution, starting
with pH 11.5, along with a double-junction Ag/AgCl reference electrode (0.1 M lithium
acetate outer solution) and the temperature sensor. Once a stable measurement was
obtained, the electrode was rinsed and placed into the next calibration solution. This
process was repeated until the ISE had been tested in each of the six buffer solutions.
All potentiometric pH ISEs were prepared in duplicate and tested in parallel to establish
reproducibility.
2.3.2.4.3. Selectivity Determination
Selectivity coefficients for H+ against the interfering ions potassium, sodium and
magnesium were determined using the FIM. Buffer solutions containing 0.1 M of the
interfering cation were prepared in the same manner as the 250 ml calibration standards,
as shown above in Table 2.5. However, 25 ml of the respective 1.0 M interfering cation
stock solution was also added prior to filling with deionised water.
2.3.2.5. Nitrite ISEs
Due to the cost of commercial nitrite ionophores, NO2- selective electrodes were not
prepared for potentiometric testing. Previously published research investigating the
response characteristics of the commercially-available ionophore NO2I found that a
PVC membrane containing 1% w/w of the ionophore resulted in log KPot values of -3.6
against interfering NO3- and -3.7 against interfering Cl
-, using the SSM
132.
Craig Alexander
96
2.4 Fabrication of Interdigitated Electrodes
2.4.1 Lift-off Photolithography
2.4.1.1. Materials
Photolithographic patterns were produced using a SF-100 ‘mask-less’ direct projection
exposure system (Intelligent Micro Patterning LLC, St. Petersburg, FL, USA). This
system directly projects a .bmp file onto the chosen substrate without the requirement
for a conventional photomask. Megaposit SPR-220 (Rohm Haas, Philadelphia, PA,
USA) is a commercial positive photoresist that was used in the IDE fabrication process.
Its chemical composition is shown in Table 2.6.
Chemical component CAS number % concentration
Ethyl lactate 97-64-3 30.0–52.0
Anisole 100-66-3 15.0–25.0
Diazo photoactive compound - 1.00–10.0
Cresol novolak resin - 14.0–40.0
Cresol 1319-77-3 0.01–0.99
2-methyl butyl acetate 624-41-9 1.00–5.00
n-amyl acetate 628-63-7 2.00–7.00
Organic siloxane surfactant - 0.01–0.10
Table 2.6: Chemical composition of Megaposit SPR-220 positive photoresist170
Megaposit MF-26A developer solution (Rohm Haas) was used as received without any
further dilution. Its chemical composition is shown in Table 2.7.
Chemical component CAS number % concentration
Water 7732-18-5 >95.0
Tetramethylammonium hydroxide 75-59-2 2.30
Polyglycol - <1.00
Table 2.7: Chemical composition of Megaposit MF-26A developer solution171
Craig Alexander
97
A KW-4A spin-coater (Chemat Technology, CA, USA), connected to a vacuum pump,
was used to evenly coat the substrate material with photoresist. A KW-4AH hot plate
(Chemat Technology) was used to bake the substrates to cure the thin-film photoresist
layer. E-beam deposition of metal layers was carried out using an Edwards Auto500
(Edwards, Crawley, UK). Gold and titanium metal pellets were 99.99% SuperVac®
grade (Testbourne, Hampshire, UK). Sulfuric acid (S.G. 1.83, >95% purity) was
purchased from Fisher Scientific. 30% hydrogen peroxide and acetone were purchased
from Sigma Aldrich.
2.4.1.2. Sensor Design
When designing the IDEs several factors with regards to the geometric and spatial
parameters had to be kept in mind. First of all, due to the in-house equipment available
for producing the electrodes using photolithography, the electrode geometry was limited
to fairly conservative feature sizes of at least 1 pixel (~15 µm) wide. As screen-printing
was also to be used as a fabrication technique for producing IDEs, the sensors were
designed with this in mind. This was to allow direct comparison of the sensing
performance of devices produced using the two different fabrication processes. The
electrodes required a suitable number of digits to produce a large enough sensing area,
but also needed to be of a suitable size so that at least four sensors could be placed into
an appropriately-sized prototype device. Due to the constraints of the SF-100
photolithography system which was used, the electrode design was limited to a total
area of 768 x 1024 pixels (approximately 11.5 x 15.4 mm).
Taking the above factors into account, the IDE sensors were designed with 25 pairs of
digits at each electrode. The digits were 9 mm long and 60 µm wide, with an inter-digit
spacing of 60 µm, giving a sensing area footprint of around 10 mm2. 60 µm was chosen
Craig Alexander
98
as it was hoped this was a large enough feature size to be obtainable using modern
screen-printing techniques, whilst being suitably small enough to produce an electric
field that would be sensitive to an ion-selective membrane deposited on top of the
electrodes. The final design can be seen in Figure 2.1. This sensor will be referred to as
‘IDE Design 1’ throughout this thesis.
Figure 2.1: Schematic representation of IDE Design 1 – Measurements not annotated are the width
of the conductive tracks (0.5 mm) and the distance between the end of each digit and the adjacent
conductive track (0.25 mm)
2.4.1.3. Fabrication Process
IDE Design 1 was fabricated in-house using a lift-off, ‘mask-less’ photolithography
technique followed by e-beam deposition of gold metal according to the following
procedure. 76 x 26 x 1 mm glass microscope slides (VWR) were cleaned by submersing
in a ‘Piranha’ solution, prepared by mixing 3 parts concentrated sulfuric acid with 1 part
hydrogen peroxide, for 15 minutes prior to use. The slides were rinsed with deionised
water and baked on a hotplate at 110 °C to dry. Approximately 2 ml of SPR-220
positive photoresist was deposited onto the clean glass slides, which were spin-coated at
3000 rpm for 30 seconds, to produce a photoresist layer around 7.5 µm thick172
. The
Craig Alexander
99
coated slides were ‘soft baked’ at 110 °C for 90 seconds, to ensure evaporation of any
solvents from the photoresist. Once baked, the slides required a rehydration period of at
least one hour, prior to patterning of the electrode design; however they were generally
allowed to hydrate overnight at this stage.
A .bmp file of the desired electrode design was loaded into a computer connected to the
SF-100 photolithography system. After ensuring the stage was level and the design was
focused, one of the coated microscope slides was placed under the light source.
Preliminary experiments suggested that an exposure time to the SF-100 light source of
60 seconds was sufficient. A total of three designs per microscope slide were patterned
into the photoresist. Each slide was left for at least 1 hour after exposure, following
which a post-exposure bake (PEB) was performed for 90 seconds at 110 °C. After the
PEB, each of the slides were allowed to rehydrate for at least one hour prior to
developing.
Once the patterned designs were suitably hydrated, each slide was placed into the
developer solution for approximately 2 minutes, with agitation, until no more of the
exposed photoresist was observed to be leaving the surface of the microscope slide.
After developing, each slide was rinsed thoroughly with deionised water and dried with
a nitrogen gun.
Following the development stage, the slides were placed face down into the Edwards
Auto500 for electron beam deposition of gold metal. After a suitable vacuum had been
obtained, a 4 nm titanium metal adhesion layer was deposited followed by a 50 nm
layer of gold metal.
Finally, the excess photoresist was removed by sonicating the slides in acetone for
approximately 2 minutes, leaving behind the gold electrode designs. Three devices were
Craig Alexander
100
produced per glass slide; the individual sensors were separated using a diamond scribe.
A microscope image of the electrode digits is shown in Figure 2.2.
Figure 2.2: Microscope image (4x) of the digits of IDE Design 1, fabricated in-house using a
mask-less, lift-off photolithographic technique followed by e-beam deposition of gold metal
2.4.1.4. Reduced Geometry IDEs
Gold IDEs with smaller features than the ones prepared in-house were provided by
MicruX Fluidics (Oviedo, Spain). This design, shown in Figure 2.3, consisted of 15
pairs of digits at each electrode, with a length of 5 mm and a digit width and spacing of
10 µm, on a Pyrex substrate. The sensor contained a reference and auxiliary electrode,
neither of which were connected during the impedimetric experiments. A layer of SU-8
negative photoresist was placed down to define the sensing area and protect the
conductive tracks. This sensor will be referred to as ‘IDE Design 2’ throughout this
thesis.
Craig Alexander
101
Figure 2.3: IDE Design 2. WE1 and WE2 refer to the two individual electrodes which form the
IDEs, RE and AE are the reference and auxillary electrodes, respectively, which were not
connected during the experiments. The red shaded areas highlight the exposed sensing and
connection areas that were not coated with SU-8 epoxy passivation layer173
2.4.2 Screen-Printing
2.4.2.1. Materials
A first attempt at producing screen-printed IDEs was carried out in collaboration with
Dr. Barry Haggett’s research group at the University of Bedfordshire. However, no
sensors that were suitable for testing were produced from this work. The issues
experienced with these sensors will be discussed in Chapter 6.
Screen-printing inks were purchased from Gwent Electronic Materials. The electrodes
were printed using a carbon-graphite ink for fine-line printing (C2110309D5), and the
Craig Alexander
102
sensing areas were defined using a low water ingress polymer dielectric
paste (D2020823D2).
The screens were purchased from DEK (Weymouth, UK). The mesh material of each
screen was polyester, and each screen contained six separate designs with varying
geometric features (Figure 2.4 and Table 2.8), with digit widths and spacings of either
60 or 100 µm. The screen for the IDE designs had a mesh thickness of 70 µm, a mesh
angle of 45° and a mesh count of 230. The screen for the protective dielectric layer had
a mesh thickness of 77 µm, a mesh angle of 45° and a mesh count of 175. Each screen
was mounted onto the frame with a tension of 22.5 N/cm.
96% alumina tiles (AD-96), with dimensions of 101.6 x 101.6 mm, and a thickness of
0.63 mm, were used as the substrate material for the first batches of screen-printed
electrodes, and were purchased from CoorsTek (Fife, UK). The substrate was laser-
scribed to allow the individual sensors to be easily separated.
Further screen-printed electrodes were provided by Dr. Craig Banks’ research group at
Manchester Metropolitan University. These designs printed effectively and will be
referred to throughout this thesis as ‘SP IDE Design 1’. The materials used in this work
were proprietary knowledge (Kanichi Research Ltd., Liverpool, UK).
2.4.2.2. Sensor Design
All sensor designs were produced using AutoCAD 2012 (version 18.2, Autodesk Ltd,
Farnborough, UK). The screen designs of the first attempt at producing screen-printed
IDEs are shown in Figure 2.4, while the geometric parameters of each individual sensor
are summarised in Table 2.8. Each screen contains the stencil of 6 different IDEs so that
Craig Alexander
103
a range of geometries could be tested, with digit widths and inter-digit spacings of either
60 µm or 100 µm, and digit lengths of 1, 5 or 10 mm.
Figure 2.4: Schematic representation of the screens used to produce 6 different IDE designs. The
parameters of each IDE A-F are summarised in Table 2.8. Figure 2.4(a) shows the electrode
geometry layer and Figure 2.4(b) shows the dielectric paste window which defines the sensing area
Design
Number of
digits at each
electrode
Digit width
(µm)
Inter-digit
spacing (µm)
Digit length
(mm)
A 12 60 60 5
B 12 100 100 5
C 12 60 60 1
D 25 60 60 5
E 25 100 100 5
F 25 100 100 10
Table 2.8: Geometric parameters of each IDE design shown in Figure 2.4
Due to the fabrication issues experienced with the first batch of screen-printed IDEs, to
be discussed later in Chapter 6, SP IDE Design 1 was constructed with much more
conservative feature sizes than the initial attempts.
SP IDE Design 1 was produced with 20 digits at each electrode, 10 mm long with a
width of 100 µm and an inter-digit spacing of 150 µm. The screen was produced so that
Craig Alexander
104
24 identical sensors were produced on a single substrate. The final electrode design is
shown in Figure 2.5.
Figure 2.5: Schematic representation of SP IDE Design 1. Measurements not annotated are the
width of the conductive tracks (1 mm) and the distance between the end of each digit and the
adjacent conductive track (0.5 mm)
2.4.2.3. Fabrication Process
2.4.2.3.1. Attempt One
The IDEs were produced as two-layer designs consisting of carbon electrodes and a
dielectric insulating window, using a microDEK 1760RS screen-printer (DEK). Each
layer was printed onto the alumina substrate, and cured at 80 °C in a fan oven for 30
minutes to remove residual solvent from the printed inks.
2.4.2.3.2. SP IDE Design 1
SP IDE Design 1 was produced using a three-layered design and is outlined in
Figure 2.6. Each sensor consists of a carbon electrode layer, an ion-selective membrane
layer and a dielectric passivation layer. Polyester screens were used; however, the exact
information is proprietary and therefore cannot be stated. A microDEK 1760RS
Craig Alexander
105
screen-printer (DEK) was used to produce the sensors, according to the
following procedure.
The IDE geometry was defined using a carbon-graphite ink formulation (carbon
conductive ink, Gwent Electronic Materials) that was screen-printed onto a
100 x 100 mm polyester flexible film substrate (Autostat, 250 µm thickness). This layer
was cured in a fan oven at 60 °C for 30 minutes (see Figure 2.6A). A microscope image
of the electrode digits is shown in Figure 2.7.
A layer of modified dielectric paste ion-selective membrane, described previously in
section 2.1.4, was printed as a rectangular block over the digits of the electrodes to form
the sensing area. This layer was cured in a fan oven at 90 °C for 60 minutes (see Figure
2.6B).
Finally, a layer of unmodified dielectric paste (Gwent Electronic Materials Ltd) was
printed to cover the connections and define the sensing area. After curing at 60 °C for
30 minutes the screen-printed sensor was ready to use (see Figure 2.6C).
Figure 2.6: Outline of the screen-printing fabrication process of SP IDE Design 1. (A) The first
layer represents the IDE geometry, connectors and conductive tracks. (B) The modified dielectric
paste ion-selective membrane layer printed over the electrode digits (dark blue). (C) The insulating
window printed to define the sensing area and cover the conductive tracks (light blue)
Craig Alexander
106
Figure 2.7: Microscope image (4x) of the electrode digits of SP IDE Design 1
2.5 Construction of Ion-Selective Impedimetric Microsensors
2.5.1 Photolithographically-prepared IDE Sensors
2.5.1.1. IDE Design 1
Gold IDEs were first cleaned with acetone, rinsed with deionised water and dried using
a nitrogen gun prior to use.
The ion-selective membrane was deposited onto the surface of the digits of the IDEs via
spin-coating. The desired membrane ‘cocktail’ (PVC or sol-gel; 50 µl) was dispensed
and spin-coated at 2000 rpm for 30 seconds. Sensors with PVC membranes were left
overnight to allow evaporation of any residual THF, whilst sensors with sol-gel
membranes were baked in an oven at 70 °C overnight.
Craig Alexander
107
Electrical connections to the IDEs were made by attaching a 0.5 mm non-insulated
copper bootlace ferrule (MultiComp, Farnell, UK) to the conductive pads using
CircuitWorks two-part, silver-loaded epoxy adhesive (Chemtronics, Kennesaw, GA,
USA), which was allowed to cure overnight. Two wires were soldered onto a female
BNC connector, which were inserted into the two ferrules and crimped tight. The
exposed conductive connectors of the sensor were covered with two-part, fast-drying
epoxy (Devcon, FL, USA). The conductive tracks were covered with electrical
insulating tape, which was also used to define a window around the sensing area.
2.5.1.2. IDE Design 2
IDE Design 2 was cleaned, using acetone, in the same way as IDE Design 1. The sensor
was supplied with cables attached to WE1 and WE2 (see Figure 2.3), RE and AE were
not connected during the experiments.
Ion-selective membranes were deposited onto the exposed sensing area via spin-coating
or drop-coating. For both coating techniques, 1 µl of the desired membrane ‘cocktail’
was dispensed onto the sensing area, with the spin-coated membranes spun at 2000 rpm
for 30 seconds. As for IDE Design 1, sensors with PVC membranes were left overnight
to allow evaporation of any residual THF, whilst sensors with sol-gel membranes were
dried in an oven at 70 °C overnight.
2.5.2 Screen-Printed IDEs
2.5.2.1. SP IDE Design 1
Screen-printed ion-selective microsensors had a membrane layer printed onto the IDE
digits during the fabrication process (see section 2.4.2.3) and therefore required no
additional modification prior to testing. Each individual sensor to be tested was simply
Craig Alexander
108
cut away from the master substrate using scissors. Electrical connections to the IDEs
were made in the same way as for IDE Design 1.
All IDE devices were stored in pure water for at least 24 hours prior to testing, to allow
sufficient soaking of the ion-selective membrane.
2.6 Impedance Measurements
2.6.1 Measurement Instrumentation
Impedimetric measurements were acquired using an Agilent E4980A LCR meter
(Agilent Technologies, Cheshire, UK), which measures two electrical properties
simultaneously across the chosen frequency range. It was controlled using an in-house
written LabVIEW program, the front panel of which is shown in Figure 2.8. The
conductive pads of the sensors were connected to the instrument using an Agilent
16085B terminal adaptor and a 1 m insulated BNC lead.
Figure 2.8: Front panel of the LabVIEW program used to control the Agilent E4980A LCR meter
Craig Alexander
109
The LCR meter used throughout this project is generally intended for use as a
front-panel controlled instrument, which would not be particularly suitable for the
experimental work described in this thesis. The LabVIEW program allowed the user to
select which impedance parameters of the sensor were to be monitored, along with the
frequency range, the number of steps along the range and the average factor of each
measurement. A function was also added to repeat the scan after a given interval, which
was particularly useful for the purpose of this work. At the completion of each scan, the
results were outputted as a Microsoft Excel workbook file.
2.6.2 Sensor Testing
The appropriate background solution (200 ml) was added to a beaker and stirred at a
constant rate. A beaker heater connected to a proportional-integral-derivative (PID)
controller was wrapped around the vessel to hold the test sample at the required
temperature. To prevent evaporation of the test sample solution, Parafilm all-purpose
laboratory film (Bemis Flexible Packaging, WI, USA) was placed over the top of the
beaker, to which a small incision was made to allow sensor access. The sensor to be
tested was submerged into the test vessel using a clamp and retort stand. Electrical
measurements over the full frequency range (20 Hz–2 MHz), at the selected input
amplitude, were taken at 10 minute intervals using the LCR meter until a stable baseline
was obtained, generally 1 hour. Once a steady baseline was reached, aliquots of the
target ion were added at various concentration intervals over the range of interest. Five
readings were taken at each concentration, two minutes apart, prior to the addition of the
next aliquot. Figure 2.9 shows a schematic diagram of the typical experimental set-up.
Specific methodology for each individual ion sensor is discussed in section 2.7.
Craig Alexander
110
Figure 2.9: Schematic diagram of the experimental set-up for testing IDE ion sensors using the
Agilent E4980A LCR meter
2.7 Ion-Selective Impedimetric Microsensors Methodology
2.7.1 Materials
Lithium nitrate (>99%) was purchased from Acros Organics. Magnesium sulfate,
disodium hydrogen phosphate and sodium dihydrogen phosphate monohydrate all had a
purity of at least 99.0% and were purchased from Sigma Aldrich. A low-range TDS
meter (ATP Instrumentation, Leicestershire, UK) was used to estimate the background
ionic strength of the test solutions. An API “Freshwater Master Test Kit” (Mars
Fishcare, PA, USA) was used to estimate the concentrations of target ions already
present within tap and aquarium water test solutions.
2.7.2 Nitrate Sensors
2.7.2.1. Stock Solutions
Unless otherwise stated, aliquots of the KNO3 stock solution described in section
2.3.2.2.1 were added to the test vessel to adjust nitrate concentration throughout. For
experiments where an alternative counter ion was required, a 10,000 ppm NO3- stock
solution was prepared by dissolving 11.12 g LiNO3 in 1 litre of deionised water.
Agilent E4980A LCR meter
Agilent 16085B terminal adaptor
Insulated 1 m BNC lead
Test solution
IDE under test
Retort stand and clamp
Craig Alexander
111
A stock solution of 10,000 ppm magnesium sulfate was produced by dissolving 5.00 g
in 500 ml of deionised water. From this, 10, 20 and 50 ml aliquots were placed into
separate 1 litre flasks and filled with deionised water, to produce 100, 200 and 500 ppm
solutions of MgSO4, respectively.
For selectivity experiments, the stock solutions of nitrite and chloride described
previously (section 2.3.2.2.1) were used.
Unless otherwise stated, all background solutions were stored in a water bath at 25 °C,
prior to testing.
2.7.2.2. Pure Water Calibrations
The nitrate sensor under test was submerged into a beaker containing 200 ml of
deionised water, stirring at a constant rate. Once a stable baseline was obtained, serial
additions of NO3- over the concentration range 1–100 ppm were added to the test vessel.
This concentration range was chosen as it is typical of that expected within a freshwater
aquarium, and covers twice the acceptable limit. NO3-
(1 ppm) was added and five
readings taken at 2 minute intervals until the next concentration (10 ppm) was added.
This process was repeated for 20, 30, 40, 50 and 100 ppm NO3-.
The effect of the counter ion on the sensor response was observed by repeating the
above experiment; however aliquots of the LiNO3 stock solution were added to the test
vessel over the same nitrate concentration range.
The effect of increasing the temperature of the test solution on the sensor response was
also observed by storing the background deionised water and NO3- stock solution at
30 °C prior to testing. The temperature of the PID-controlled beaker heater was also
increased to 30 °C.
Craig Alexander
112
2.7.2.3. Background Effects
Magnesium sulfate of varying concentrations was used as an ionic strength buffer to test
the effect of background ion concentration on the response of the impedimetric nitrate
sensors. MgSO4 solutions of 100, 200 and 500 ppm were tested. These background
concentrations were chosen as 100–200 ppm would be close to the expected TDS value
found in a freshwater aquarium, whilst 500 ppm should provide a background ionic
strength much higher than would be generally observed. As for the calibrations in pure
water, aliquots of nitrate added to the test vessel were over the concentration range of
1–100 ppm.
As many aquarium hobbyists will not have access to purified water, and therefore will
fill aquaria with tap water, it was deemed necessary to test the sensor response by
adding known concentrations of nitrate to tap water. Prior to testing, a handheld TDS
meter was used to estimate the background ionic strength of the sample. The calibration
was performed as described above (section 2.7.2.2). Known concentrations of nitrate
were also added to a sample of water taken from a live aquarium. Prior to testing the
sample, a handheld TDS meter was used to estimate the background ionic strength, and
the initial nitrate concentration in the aquatic water was approximated by with a
standard aquarium test kit and a nitrate ISE.
2.7.2.4. Selectivity
Selectivity of the impedimetric nitrate sensors was deduced using slightly modified
versions of two different mixed solution methods that are traditionally used when
determining selectivity coefficients of potentiometric ISEs (FIM and FPM).
Craig Alexander
113
The key anions found within a freshwater aquarium that could cause a potential
interference to a nitrate-selective sensor were identified as nitrite and chloride. As
discussed earlier in section 1.1.1.2, nitrite levels in an aquarium should not exceed 0.2
ppm. Chloride is generally found in tap water at a concentration of <250 ppm.
A value of 10 ppm nitrite was chosen as the background concentration for carrying out
the FIM. The sensor was added to pure water and, once a stable baseline was observed,
10 ppm NO2- was added to the test vessel. Five measurements were taken at 2 minute
intervals, followed by additions of nitrate in the concentration range 1–100 ppm as
before. For the FPM, the sensor was added to pure water until a stable baseline was
observed, followed by the addition of 50 ppm NO3-. Five measurements were taken at
2 minute intervals, followed by the addition of nitrite at a concentration of 1, 10 and
20 ppm.
A value of 50 ppm chloride was chosen as the background concentration for carrying
out the FIM. The sensor was added to pure water and, once a stable baseline was
observed, 50 ppm Cl- was added to the test vessel. Five measurements were taken at
2 minute intervals, followed by additions of nitrate in the concentration range
1–100 ppm as before. For the FPM, the sensor was added to pure water until a stable
baseline was observed, followed by the addition of 50 ppm NO3-. Five measurements
were taken at 2 minute intervals, followed by the addition of chloride at a concentration
of 10, 50 and 100 ppm.
The response of a sensor coated in a membrane prepared without ionophore or ionic
additive present was established by carrying out a calibration over the concentration
range of 1–100 ppm NO3- in pure water, as a control.
Craig Alexander
114
2.7.2.5. Reproducibility
To determine the device-to-device reproducibility of the impedimetric nitrate sensors,
each sensor was produced in triplicate and a pure water calibration using KNO3 over the
concentration range 1–100 ppm NO3- was performed.
Ammonium Sensors
2.7.2.6. Stock Solutions
Unless otherwise stated, aliquots of the NH4Cl stock solution described in section
2.3.2.2.1 were added to the test vessel to adjust ammonium concentration during sensor
calibration experiments throughout.
The interfering ion stock solutions of Na+, K
+, Mg
2+ and Ca
2+ are described in section
2.3.2.3.1. Unless otherwise stated, prior to testing, all background solutions were stored
in a water bath at 25 °C.
2.7.2.7. Measurements in Water
Due to time constraints, it was not possible to obtain a full data set for impedimetric
ammonium sensors. The ammonium sensor under test was submerged into a beaker
containing 200 ml of the deionised water, stirring at a constant rate. Once a stable
baseline was obtained, a single addition of NH4+ at a concentration of 50 ppm was
added to the test vessel.
2.7.2.8. Selectivity
Selectivity of the impedimetric ammonium sensors was deduced using both the FIM.
The cation interferences that were investigated were Na+, K
+ and Mg
2+.
Craig Alexander
115
The fixed interfering cation concentration was 50 ppm. The sensor was added to 200 ml
of deionised water, and once a stable baseline was observed, 50 ppm interfering ion was
added to the test vessel. Once a stable measurement was reached, 50 ppm of ammonium
was added to the test vessel.
The response of a sensor coated in a membrane prepared without ionophore or ionic
additive present was established by carrying out a measurement of 50 ppm NH4+ in a
deionised water baseline, as a control.
2.7.3 pH Sensors
2.7.3.1. Stock Solutions
For pH experiments, phosphate buffers of varying pH values were produced from stock
solutions of 0.1 M anhydrous disodium hydrogen phosphate (Na2HPO4) and 0.1 M
sodium dihydrogen phosphate monohydrate (NaH2PO4.H2O). Na2HPO4 (14.20 g) and
NaH2PO4.H2O (13.80 g) were dissolved in separate 1 litre flasks.
As stated in section 1.1.2, a typical aquarium would generally have a pH around neutral.
Therefore the chosen range for testing the impedimteric pH sensors was between
pH 6.0–8.5. Phosphate buffer solutions (0.01 M) were prepared by mixing the amounts
of 0.1 M stock solutions (described in section 2.7.2.1) shown in Table 2.9 and making
up to 500 ml with deionised water. The pH was confirmed using a bench top pH meter
and adjusted as necessary using drop-wise additions of the respective stock solutions.
Craig Alexander
116
Desired pH 0.1 M Na2HPO4 (ml) 0.1 M NaH2PO4.H2O (ml)
6.0 6.0 44.0
6.5 15.2 34.8
7.0 28.9 21.1
7.5 40.5 9.5
8.0 46.6 3.4
8.5 49.0 1.0
Table 2.9: Composition of each phosphate buffer stock solution required to produce 0.1 M solutions
of each buffer at the required pH
pH 8.5 is outside the published range of this buffer composition, and therefore required
more 0.1 M sodium hydroxide to adjust it to the correct pH value. Hence, the
background ionic strength of this buffer would be slightly higher than the rest.
For impedimetric measurements, a low conductivity solution is required. The final
sample solution was produced by taking a 20 ml aliquot of the buffer at the pH to be
tested and diluting to 200 ml with deionised water. This produced a buffer solution with
a concentration of approximately 1.0 mM. The final pH was recorded using a bench top
pH meter and the background ionic strength estimated using a handheld TDS meter,
prior to addition of nitrate over the concentration range 1–100 ppm. This process was
repeated for each of the buffers stated in Table 2.9.
Unless otherwise stated, prior to testing, all background solutions were stored in a water
bath at 25 °C.
2.7.3.2. Calibrations
Impedimetric pH sensors were calibrated over the range pH 6.0–8.5. The sensor under
test was added to the first calibration solution, starting with pH 8.5, with measurements
taken at 5 minute intervals for a total of 30 minutes. The electrode was rinsed with
Craig Alexander
117
deionised water and added to the next buffer solution, pH 8.0. This process was
repeated for pH 7.5, 7.0, 6.5 and 6.0.
2.7.4 Nitrite Sensors
2.7.4.1. Stock Solutions
Unless otherwise stated, aliquots of the KNO2 stock solution described in section
2.3.2.2.1 were added to the test vessel to adjust nitrite concentration throughout.
For selectivity experiments, the stock solutions of nitrate and chloride described
previously in section 2.3.2.2.1 were used. Unless otherwise stated, prior to testing, all
background solutions were stored in a water bath at 25 °C.
2.7.4.2. Measurements in Water
The nitrite sensor under test was submerged into a beaker containing 200 ml of
deionised water, stirring at a constant rate. Once a stable baseline was obtained, an
aliquot of NO2- stock solution was added to the test vessel to give a concentration of
5 ppm.
2.7.4.3. Selectivity
The selectivity of the impedimetric nitrite sensors was determined against interfering
nitrate and chloride anions at a fixed concentration of 50 ppm. A value of 50 ppm
nitrate was chosen as the background concentration for carrying out the FIM. A value
of 50 ppm chloride was chosen as the background concentration for carrying out the
FIM.
Craig Alexander
118
The response of a sensor coated in a membrane prepared without ionophore or ionic
additive present was established by carrying out a measurement of 5 ppm NO2-
in a
deionised water background, as a control.
Craig Alexander
119
3 Potentiometric Characterisation of Ionophores
This chapter evaluates the analytical performance characteristics of several
commercially-available ionophores experimentally for each of the identified key
aquarium ionic analytes, within conventional potentiometric ISEs. It was of great
importance to understand the behaviour of these ionophores within established chemical
sensing methodologies, to ascertain their potential suitability for use within a freshwater
aquarium monitoring device.
3.1 Nitrate-Selective Electrodes
3.1.1 Nitrate PMEs
Nitrate PMEs were prepared using PVC membranes containing 1% w/w of the
commercial ionophores TDAN and Nitrate Ionophore V (NO3V). The calibration
results over the range 0.1–10,000 ppm NO3- are shown in Figures 3.1 (TDAN) and
Figure 3.2 (NO3V). Selectivity data are shown in Figure 3.3 (TDAN) and
Figure 3.4 (NO3V).
Figure 3.1: Calibration graph of two nitrate-selective PMEs prepared using a PVC membrane
containing 1% w/w TDAN as the ionophore. The calibration results of a commercial
nitrate-selective half-cell tested in the same solutions are also shown
TDAN PME 1 y = -50.65x + 350.85
R² = 0.9972
TDAN PME 2y = -52.24x + 359.06
R² = 0.9981
Commercial nitrate half-cell y = -52.69x + 415.02
R² = 0.9951
0
50
100
150
200
250
300
350
400
450
500
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
TDAN PME 1
TDAN PME 2
Commercial nitratehalf-cell
Craig Alexander
120
The two prepared TDAN PMEs exhibited a linear range (R2 >0.995) over 1–1000 ppm
NO3- with near-Nernstian slopes of -50.7 and -52.2 mV/dec., respectively. The recorded
temperature for each measurement was between 21.1 and 22.5 °C, therefore the
temperature effects on the results can be considered negligible. The device-to-device
reproducibility of the TDAN PMEs was defined as the relative standard deviation
(%RSD) of the two slopes from the linear range of the calibration graphs, which was
calculated to be 2.19%. Extrapolation of the outliers to the linear section gives the lower
detection limits (LDLs) and upper detection limits (UDLs) for the ISE. These were
calculated for TDAN PME 1 as shown in Equations 3.1 and 3.2.
𝐋𝐃𝐋 (𝐩𝐩𝐦) = 𝟏𝟎(
𝟑𝟓𝟔.𝟓 𝐦𝐕−𝟑𝟓𝟎.𝟖𝟓−𝟓𝟎.𝟔𝟓
)
= 0.77 ppm NO3-
Equation 3.1
𝐔𝐃𝐋 (𝐩𝐩𝐦) = 𝟏𝟎(
𝟏𝟕𝟖.𝟓 𝐦𝐕−𝟑𝟓𝟎.𝟖𝟓−𝟓𝟎.𝟔𝟓
)
= 2540 ppm NO3-
Equation 3.2
The LDLs and UDLs for TDAN PME 2 were calculated as 0.93 and 2130 ppm NO3-,
respectively.
Craig Alexander
121
Figure 3.2: Calibration graph of two nitrate-selective PMEs prepared using a PVC membrane
containing 1% w/w NO3V as the ionophore
The two prepared NO3V PMEs exhibited a linear range (R2 >0.995) over 1–10,000 ppm
NO3- with near-Nernsitan slopes of -54.9 and -55.2 mV/dec., respectively. The recorded
temperature for each measurement was between 20.5 and 22.2 °C. The %RSD of the
two slopes was calculated as 0.36%. The LDLs were calculated as 0.57 ppm for NO3V
PME 1 and 0.56 ppm for NO3V PME 2. The UDLs were calculated as 8010 ppm for
NO3V PME 1 and 7980 ppm for NO3V PME 2.
NO3V PME 1y = -54.92x + 220.9
R² = 0.9965
NO3V PME 2 y = -55.2x + 222.3
R² = 0.9964
0
50
100
150
200
250
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
NO3VPME 1
NO3VPME 2
Craig Alexander
122
Figure 3.3: Selectivity determination of TDAN PME 1 using the FIM in 1000 ppm Cl
- and
1000 ppm NO2-. The response of a PME prepared with a ‘blank’ membrane containing no TDAN
is also shown
The potentiometric selectivity coefficients of TDAN PME 1 for interfering NO2- and Cl
-
anions were calculated as illustrated in Figure 1.8. For NO2-, extrapolating the outliers
to the linear section of the calibration curve gives a CA value of 30.3 ppm to enter into
Equation 1.22, and Cl- gives a CA value of 7.3 ppm. The CB value for the interfering
cations was fixed at 1000 ppm. The KPot values were calculated as shown in
Equations 3.3 and 3.4.
𝐊𝐍𝐎𝟑−,𝐍𝐎𝟐
− 𝐏𝐨𝐭 =
𝟑𝟎. 𝟑 𝐩𝐩𝐦
𝟏𝟎𝟎𝟎 𝐩𝐩𝐦= 𝟎. 𝟎𝟑𝟎𝟑 Equation 3.3
𝐥𝐨𝐠𝟏𝟎(𝐊𝐍𝐎𝟑−,𝐍𝐎𝟐
−𝐏𝐨𝐭 ) = −𝟏. 𝟓𝟐 Equation 3.4
Log10 (KNO3−,Cl−
Pot ) was calculated as -2.14. No Nernstian response behaviour was
observed from the ‘blank’ membrane, which was prepared without TDAN present
as the ionophore.
TDAN PME 1 y = -50.65x + 350.85
R² = 0.9972
0
50
100
150
200
250
300
350
400
450
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
Pure Nitrate
1000 ppm Cl-
1000 ppm NO2-
Blank membrane
Craig Alexander
123
Figure 3.4: Selectivity determination of NO3V PME 1 using the FIM in 1000 ppm Cl
- and 1000 ppm
NO2-. The response of a PME prepared with a ‘blank’ membrane containing no NO3V as the
ionophore or TDMACl as the ionic additive is also shown
The logarithmic potentiometric selectivity coefficients of NO3V PME 1 for interfering
NO2- and Cl
- anions were calculated as -1.32 and -2.19, respectively. No Nernstian
response behaviour was observed from the ‘blank’ membrane, which was prepared
without NO3V present as the ionophore or TDMACl present as the ionic additive.
NO3V PME 1y = -54.92x + 220.9
R² = 0.9965
0
50
100
150
200
250
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
Pure Nitrate
1000 ppm Cl-
1000 ppm NO2-
Blank membrane
Craig Alexander
124
3.1.2 Nitrate CWEs
CWEs, described in section 2.2.3, were prepared as nitrate sensors by dip-coating
Ag/AgCl electrodes by hand into either a PVC or sol-gel solution containing TDAN as
the ionophore. The calibration results for the PVC membrane electrodes are shown in
Figure 3.5, and the sol-gel membrane calibration results are shown in Figure 3.6. FIM
selectivity data are shown in Figure 3.7 (PVC CWE), Figure 3.8 (MTES CWE) and
Figure 3.9 (DEDMS CWE).
Figure 3.5: Calibration graph of two nitrate-selective CWEs prepared by dip-coating the electrode
into a PVC membrane ‘cocktail’ containing 1% w/w TDAN as the ionophore
Both prepared PVC CWEs exhibited reasonable linearity (R2 >0.96) over the
concentration range of 1–1000 ppm NO3-, with a slope that was much greater than the
expected Nernstian value of -59.2 mV/dec. The slope for TDAN PVC CWE 1 was
-77.09 mV/dec. and the slope for TDAN PVC CWE 2 was -71.47 mV/dec. The
recorded temperature for each measurement was between 21.2 and 22.8 °C. The %RSD
of the two slopes was calculated as 5.35%. The LDL for TDAN PVC CWE 1 was
calculated as 1.5 ppm, and for TDAN PVC CWE 2 it was calculated as 0.5 ppm. The
TDAN PVC CWE 2y = -71.47x + 337.48
R² = 0.9804
TDAN PVC CWE 1y = -77.09x + 353.91
R² = 0.9597
0
50
100
150
200
250
300
350
400
450
500
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
TDAN PVC CWE 1
TDAN PVC CWE 2
Craig Alexander
125
UDLs were calculated as 3240 ppm for TDAN PVC CWE 1 and 9530 ppm for
TDAN PVC CWE 2.
Figure 3.6: Calibration graph of four nitrate-selective sol-gel CWEs (two MTES and two DEDMS)
containing TDAN as the ionophore
The two sol-gel CWEs that were prepared from the silica precursor MTES containing
TDAN as the nitrate ionophore exhibited a linear range (R2 >0.99) between
10–1000 ppm NO3-. The recorded temperature for each measurement was between 21.3
and 22.1 °C. The %RSD of the two slopes was calculated as 0.58%; however, there
was a large discrepancy between devices of around 20 mV between the obtained
potential values at each measured nitrate concentration. TDAN MTES CWE 1 had a
slope of -48.25 mV/dec. and TDAN MTES CWE 2 had a slope of -48.65 mV/dec. The
LDLs were calculated as 8.4 ppm for TDAN MTES CWE 1 and 8.3 ppm for TDAN
MTES CWE 2. The UDLs were calculated as 2890 ppm for TDAN MTES CWE 1 and
5560 ppm for TDAN MTES CWE 2.
The two sol-gel CWEs that were prepared from the 3:1 ratio of the silica precursors
DEDMS and TEOS containing TDAN as nitrate ionophore exhibited a linear range
TDAN MTES CWE 1y = -48.25x + 286.17
R² = 0.9913TDAN MTES CWE 2y = -48.65x + 269.8
R² = 0.9936
TDAN DEDMS CWE 1y = -39.32x + 364.35
R² = 0.991
TDAN DEDMS CWE 2y = -43.91x + 380.75
R² = 0.9991
0
50
100
150
200
250
300
350
400
450
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
TDAN MTES CWE 1
TDAN MTES CWE 2
TDAN DEDMS CWE 1
TDAN DEDMS CWE 2
Craig Alexander
126
(R2 >0.99) between 10–10,000 ppm NO3
-. The DEDMS electrodes were tested
alongside the MTES electrodes and therefore the recorded temperature for each
measurement was the same as for the MTES CWEs above. The %RSD of the two
slopes was calculated as 7.8%. TDAN DEDMS CWE 1 had a slope of -39.32 mV/dec.
and TDAN DEDMS CWE 2 had a slope of -43.91 mV/dec. The LDLs were calculated
as 3.1 ppm for TDAN DEDMS CWE 1 and 4.5 ppm for TDAN DEDMS CWE 2. The
UDLs were calculated as 7760 ppm for TDAN DEDMS CWE 1 and 9750 ppm for
TDAN DEDMS CWE 2.
Figure 3.7: Selectivity determination of TDAN PVC CWE 1 using the FIM in 1000 ppm Cl
- and
1000 ppm NO2-. The response of a CWE prepared with a ‘blank’ membrane containing no TDAN is
also shown
The logarithmic potentiometric selectivity coefficients of TDAN PVC CWE 1 for
interfering NO2- and Cl
- anions were calculated as -0.65 and -1.38, respectively. A
Nernstian response was not observed from the ‘blank’ membrane, which was prepared
without TDAN present as the ionophore; however, there did appear to be some linearity
with increasing nitrate concentration. This suggested that the obtained response was not
exclusively due to increasing nitrate concentration within the test sample and some
TDAN PVC CWE 1y = -71.47x + 337.48
R² = 0.9804
0
50
100
150
200
250
300
350
400
450
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
Pure Nitrate
1000 ppm Cl-
1000 ppm NO2-
Blank PVC CWE
Craig Alexander
127
matrix interferences were observed. This therefore explained why a response that was
much greater than Nernstian was obtained for this ISE during the calibration
experiments.
Figure 3.8: Selectivity determination of TDAN MTES CWE 1 using the FIM in 1000 ppm Cl
- and
1000 ppm NO2-. The response of a CWE prepared with a ‘blank’ membrane containing no TDAN is
also shown
The logarithmic potentiometric selectivity coefficients of TDAN MTES CWE 1 for
interfering NO2- and Cl
- anions were calculated as 0.11 and -0.81, respectively. A
reasonably linear response, although not strictly Nernstian, was observed from a ‘blank’
MTES membrane which did not contain TDAN as the ionophore.
TDAN MTES CWE 1 y = -48.25x + 286.17
R² = 0.9913
0
50
100
150
200
250
300
350
400
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
Pure Nitrate
1000 ppm Cl-
1000 ppm NO2-
Blank MTES CWE
Craig Alexander
128
Figure 3.9: Selectivity determination of TDAN DEDMS CWE 1 using the FIM in 1000 ppm Cl
- and
1000 ppm NO2-. The response of a CWE prepared with a ‘blank’ membrane containing no TDAN is
also shown
The logarithmic potentiometric selectivity coefficients of TDAN DEDMS CWE 1 for
interfering NO2- and Cl
- anions were calculated as 0.55 and -0.73, respectively. The
‘blank’ DEDMS membrane which did not contain TDAN as the ionophore
demonstrated an almost equivalent response to the target nitrate ion. This suggested that
the selective ionophore within the membrane was not the sole cause of the response, and
that a non-selective response was achieved through matrix interferences.
Due to the poor analytical performance of the TDAN CWEs compared with the PMEs,
which were prepared with the same ionophore, further CWEs were not prepared using
NO3V or with ionophores for the other target analytes. Neither PVC nor the sol-gels
demonstrated good response characteristics when used as the membrane support matrix
for the ionophore TDAN. The CWEs demonstrated considerably inferior selectivity
coefficients against interfering nitrite and chloride anions compared with the PMEs. A
much more prominent response to the target nitrate ion was observed when a CWE with
a ‘blank’ membrane, which did not contain any ionophore, was tested. The CWEs also
TDAN DEDMS CWE 1y = -39.32x + 364.35
R² = 0.991
0
50
100
150
200
250
300
350
400
450
-2 -1 0 1 2 3 4 5
E (m
V)
log NO3- concentration (ppm)
Pure Nitrate
1000 ppm Cl-
1000 ppm NO2-
Blank DEDMS CWE
Craig Alexander
129
exhibited poor device-to-device reproducibility and less favourable detection limits than
the PMEs.
3.2 Ammonium-Selective Electrodes
Figure 3.10 shows the calibration results for an ammonium-selective PME over the
range 0.1–10,000 ppm NH4+, prepared with the NH4I PVC membrane described in
Table 2.1, containing potassium tetrakis (4-chlorophenyl) borate in a 1:2 mole ratio
(with respect to the ionophore), as the ionic additive. The FIM selectivity data for this
electrode against the interfering cations Na+, K
+, Mg
2+ and Ca
2+ are shown in
Figure 3.11.
Figure 3.10: Calibration graph of two ammonium-selective PMEs prepared using a PVC
membrane containing 1% w/w NH4I as the ionophore
The ammonium PMEs showed reasonable linearity (R2 >0.975) over the concentration
range 10–10,000 ppm NH4+, and exhibited sub-Nernstian slopes of 47.6 mV/dec. for
NH4I PME 1 and 41.5 mV/dec. for NH4I PME 2. The recorded temperature for each
measurement was between 20.9 and 22.7 °C. The %RSD of the slopes was calculated
as 9.61%. The LDLs were calculated as 6.4 ppm and 10.3 ppm NH4+, respectively.
NH4I PME 1y = 47.6x - 40.6
R² = 0.9873
NH4I PME 2y = 41.54x - 18.45
R² = 0.9776
-100
-50
0
50
100
150
200
-2 -1 0 1 2 3 4 5
E (m
V)
log NH4+ concentration (ppm)
NH4I PME 1
NH4I PME 2
Craig Alexander
130
Figure 3.11: Selectivity determination of NH4I PME 1 using the FIM in 1000 ppm of the interfering
cations sodium, potassium, magnesium and calcium
The potentiometric selectivity coefficients were calculated as before (section 3.1.1), for
the interfering cations K+, Na
+, Mg
2+ and Ca
2+. The calculated log KNH4
+,B+Pot values are
summarised in Table 3.1.
Interfering cation (B+) log 𝐊𝐍𝐇𝟒
+,𝐁+𝐏𝐨𝐭
K+ -0.08
Na+ -1.21
Mg2+
-1.90
Ca2+
-2.09
Table 3.1: Potentiometric selectivity coefficients of NH4I PME 1
It has previously been reported that although NH4I (nonactin) is a neutral ionophore,
adding an anion exchanger to the membrane can actually lead to a reduction in its
selectivity and potentiometric performance as an ISE174
. The membrane was therefore
prepared again, with the same ratios of NH4I, plasticiser and PVC; however, potassium
tetrakis (4-chlorophenyl) borate was not added to the polymeric ‘cocktail’ solution. The
calibration results for the PMEs prepared using this alternative membrane composition
are shown in Figure 3.12. The FIM selectivity data for interfering cations Na+, K
+, Mg
2+
NH4I PME 1y = 47.6x - 40.6
R² = 0.9873
-100
-50
0
50
100
150
200
-2 -1 0 1 2 3 4 5
E (m
V)
log NH4+ concentration (ppm)
Pure Ammonium
1000 ppm Na+
1000 ppm K+
1000 ppm Mg2+
1000 ppm Ca2+
Craig Alexander
131
and Ca2+
obtained using the alternative membrane composition are shown in
Figure 3.13.
Figure 3.12: Calibration graph of two ammonium-selective PMEs prepared using a PVC
membrane containing 1% w/w NH4I as the ionophore, with the ionic additive omitted. The
calibration results of a commercial ammonium-selective half-cell tested in the same solutions are
also shown
When the ammonium-selective membrane was prepared with the omission of potassium
tetrakis (4-chlorophenyl) borate, the PMEs exhibited good linearity (R2 >0.995) over the
concentration range 1–10,000 ppm NH4+, and near-Nernstian slopes of 50 mV/dec. for
NH4I PME 1 and 47.7 mV/dec. for NH4I PME 2. The recorded temperature for each
measurement was between 20.4 and 22.5 °C. The %RSD of the slopes was calculated
as 3.36%. The LDLs were calculated as 0.5 ppm and 0.8 ppm NH4+, respectively.
NH4I PME 2y = 47.69x - 45.86
R² = 0.995
NH4I PME 1y = 50.01x - 53.04
R² = 0.9965
Commercial ammonium half-celly = 55.81x + 93.1
R² = 0.9997
-150
-100
-50
0
50
100
150
200
250
300
350
-2 -1 0 1 2 3 4 5
E (m
V)
log NH4+ concentration (ppm)
NH4I PME 1
NH4I PME 2
Commercial ammoniumhalf-cell
Craig Alexander
132
Figure 3.13: Selectivity determination of NH4I PME 1 (ionic additive omitted) using the FIM in
1000 ppm of the interfering cations sodium, potassium, magnesium and calcium. The response of a
PME prepared with a ‘blank’ membrane containing neither NH4I nor ionic additive is also shown
The potentiometric selectivity coefficients for the alternative NH4I PME 1, prepared
with a membrane that did not contain potassium tetrakis (4-chlorophenyl) borate as
anion exchanger, are summarised in Table 3.2. There was no Nernstian response
observed from the ‘blank’ electrode with the membrane that did not contain NH4I as the
ionophore.
Interfering cation (B+) log 𝐊𝐍𝐇𝟒
+,𝐁+𝐏𝐨𝐭
K+ -1.25
Na+ -2.59
Mg2+
-2.60
Ca2+
-3.07
Table 3.2: Potentiometric selectivity coefficients of NH4I PME 1 (ionic additive omitted)
NH4I PME 1y = 50.01x - 53.04
R² = 0.9965
-100
-50
0
50
100
150
200
-2 -1 0 1 2 3 4 5
E (m
V)
log NH4+ concentration (ppm)
Pure Ammonium
1000 ppm Na+
1000 ppm K+
1000 ppm Mg2+
1000 ppm Ca2+
BLANK NH4I
Craig Alexander
133
3.3 Hydrogen Ion-Selective Electrodes
Figure 3.14 shows the calibration results for a hydrogen ion-selective PME prepared
with the commercial ionophore HIII and Figure 3.15 shows the calibration results for a
hydrogen ion-selective PME prepared with the commercial ionophore HV. Both
ionophores were prepared within a PVC membrane as described in Table 2.1, and tested
over the pH range 2.5–11.5. The FIM selectivity data for the HIII PME are shown in
Figure 3.16, and for the HV PME in Figure 3.17, against the interfering cations Na+, K
+
and Mg2+
.
Figure 3.14: Calibration graph of two hydrogen ion-selective PMEs prepared using a PVC
membrane containing 1% w/w HIII as the ionophore. The calibration results of a commercial pH
half-cell tested in the same solutions are also shown
The HIII PMEs demonstrated excellent linearity (R2 >0.99) over the full tested pH
range of 2.5–11.5, with slopes of -46.70 mV/dec. for HIII PME 1 and -51.11 mv/dec.
for HIII PME 2. The recorded temperature for each measurement was between 20.3 and
21.8 °C. The %RSD of the slopes was calculated as 6.38%. The LDLs were calculated
as pH 10.8 for HIII PME 1 and pH 11.4 for HIII PME 2. The UDLs were calculated as
pH 1.8 for HIII PME 1 and pH 2.6 for HIII PME 2.
HIII PME 1y = -46.698x + 371.5
R² = 0.9999
HIII PME 2y = -51.111x + 401.31
R² = 0.998
Commercial pH half-celly = -56.165x + 336.35
R² = 0.9997
-400
-300
-200
-100
0
100
200
300
400
0 2 4 6 8 10 12 14E (m
V)
pH
HIII PME 1
HIII PME 2
CommercialpH half-cell
Craig Alexander
134
Figure 3.15: Calibration graph of two hydrogen ion-selective PMEs prepared using a PVC
membrane containing 1% w/w HV as the ionophore
The HV PMEs demonstrated excellent linearity (R2 >0.99) over the pH range of
5.0–11.5, with sub-Nernstian slopes of -39.80 mV/dec. for HV PME 1 and -39.89
mv/dec. for HV PME 2. The HV electrodes were tested alongside the HIII ones and
therefore the recorded temperature for each measurement was the same as for the HIII
PMEs above. The %RSD of the slopes was calculated as 0.16%. The LDLs were
calculated as pH 11.6 for HV PME 1 and pH 11.7 for HV PME 2. The UDL were
calculated as pH 4.4 for HV PME 1 and pH 4.6 for HV PME 2.
HV PME 2y = -39.887x + 327.92
R² = 0.9977
HV PME 1y = -39.796x + 324.39
R² = 0.9959
-200
-150
-100
-50
0
50
100
150
200
250
0 2 4 6 8 10 12 14
E (m
V)
pH
HV PME 1
HV PME 2
Craig Alexander
135
Figure 3.16: Selectivity determination of HIII PME 1 using the FIM in 0.1 M of the interfering
cations sodium, potassium and magnesium. The response of a PME prepared with a ‘blank’
membrane containing neither HIII nor ionic additive is also shown
The potentiometric selectivity coefficients for HIII PME 1 were calculated as before, for
the interfering cations K+, Na
+ and Mg
2+. The calculated log KH+,B+
Pot values are
summarised in Table 3.3. There was no Nernstian response to the change in hydrogen
ion concentration observed from the ‘blank’ electrode with the membrane that did not
contain HIII as the ionophore.
Interfering cation (B+) log 𝐊𝐇+,𝐁+
𝐏𝐨𝐭
K+ -11.00
Na+ -11.14
Mg2+
-10.78
Table 3.3: Potentiometric selectivity coefficients of HIII PME 1
HIII PME 1y = -46.698x + 371.5
R² = 0.9999
-200
-150
-100
-50
0
50
100
150
200
250
300
0 2 4 6 8 10 12 14
E (m
V)
pH
HIII PME 1
0.1 M K+
0.1 M Na+
0.1 M Mg2+
Blank HIII
Craig Alexander
136
Figure 3.17: Selectivity determination of HV PME 1 using the FIM in 0.1 M of the interfering
cations sodium, potassium and magnesium. The response of a PME prepared with a ‘blank’
membrane containing neither HV nor ionic additive is also shown
The potentiometric selectivity coefficients for HV PME 1 were calculated as before, for
the interfering cations K+, Na
+ and Mg
2+. The calculated log KH+,B+
Pot values are
summarised in Table 3.4. There was no Nernstian response to the change in hydrogen
ion concentration observed from the ‘blank’ electrode with the membrane that did not
contain HV as the ionophore.
Interfering cation (B+) log 𝐊𝐇+,𝐁+
𝐏𝐨𝐭
K+ -11.09
Na+ -11.41
Mg2+
-11.21
Table 3.4: Potentiometric selectivity coefficients of HV PME 1
3.4 Conclusions for Chapter 3
Several commercially-available ionophores were tested within potentiometric ISEs to
assess their analytical performance relating to the determination of three
aquarium-significant ions. Where possible, two different ionophores were investigated
HV PME 1y = -39.887x + 327.92
R² = 0.9977
-200
-150
-100
-50
0
50
100
150
200
250
0 2 4 6 8 10 12 14
E (m
V)
pH
HV PME 1
0.1 M K+
0.1 M Na+
0.1 M Mg2+
Blank HV
Craig Alexander
137
for each target ion. Nitrite was not investigated using potentiometric ISEs due to the
cost of commercially-available nitrite ionophores.
For the nitrate anion, ion-selective membranes containing two different ionophores,
TDAN and NO3V, were tested within potentiometric ISEs. When tested within PMEs
with a PVC membrane layer, both ionophores responded in a near-Nernstian fashion
over a range which would make them suitable for determining the nitrate concentration
within a freshwater aquarium. The NO3V PME exhibited a wider linear range, a slope
that fitted the theoretical value more closely and a lower limit of detection than the
TDAN electrode. The NO3V PMEs also showed better device-to-device reproducibility.
The TDAN PME proved to be marginally more selective to nitrate over nitrite than the
NO3V PME (log10(KNO3−,NO2
−Pot ) = -1.52 versus -1.32), whereas the NO3V PME was
slightly more selective over chloride (log10(KNO3−,Cl−
Pot ) = -2.19 versus -2.14).
The ionophore TDAN was also tested within CWEs, using PVC and two different
sol-gels as the membrane support matrix. The PVC CWEs produced a slope that was
much greater than the theoretically expected value, and all of the response
characteristics were unfavourable when compared with the PMEs. Additionally, sol-gel
CWEs did not provide adequate analytical responses to nitrate. Inadequate response
characteristics from the CWEs may have arisen due to insufficient coating and/or poor
adhesion of the membrane material to the Ag/AgCl wire. As the electrodes were
dip-coated by hand, this may have led to an inhomogeneous coating onto the surface of
the electrode, which may have caused the lack of ion-selectivity and overall poor
analytical performance. To overcome this issue, a commercial dip-coating system, for
example a Chemat DipMaster175
, could be used to ensure that aspects of the dip-coating
process, such as immersion time and withdrawal speed are kept constant.
Craig Alexander
138
It was still of great interest to further investigate the use of sol-gels as an alternative
ion-selective membrane material, however it was considered that they may be more
suitable to coating a planar substrate rather than depositing over a wire electrode. The
use of silica gels as ion-selective membranes within impedimetric IDE devices will be
discussed in Chapter 4.
For ammonium, only one commercial ionophore was available to be purchased, NH4I,
which was tested within PMEs. Initially, PVC membranes were used which consisted of
the plasticiser NPOE and potassium tetrakis (4-chlorophenyl) borate as anion
exchanger. It was shown that the analytical performance of NH4+ PMEs, using NH4I as
the ionophore, was improved substantially when potassium tetrakis (4-chlorophenyl)
borate was omitted from the PVC ion-selective membrane. The PMEs prepared without
the ionic additive in the membrane demonstrated a wider linear range, a lower detection
limit, improved device-to-device reproducibility and improved selectivity coefficients
for each of the tested interfering cations (K+, Na
+, Mg
2+ and Ca
2+). Subsequent
ammonium-selective membranes required for further experimentation within
impedimetric IDE devices, described in Chapter 7, were therefore prepared with
potassium tetrakis (4-chlorophenyl) borate omitted from the membrane ‘cocktail’.
For pH determination, two commercial hydrogen ion-selective ionophores, HIII and
HV, were tested within PMEs with PVC membranes. Both ionophores exhibited
excellent linearity over the expected pH range of a freshwater aquarium, along with
very high selectivity coefficients for the tested interfering cations. HIII PME
demonstrated a wider linear range (2.5–11.5 versus 5.0–11.5), whereas HV did not
show linearity in the acidic region, below approximately pH 4.0.
Craig Alexander
139
As mentioned previously, one of the major disadvantages of potentiometric ISEs is the
requirement of an external reference electrode for them to function. Such a set-up would
not be suited for use within a home aquarium monitoring product, particularly with
sensors and reference electrodes containing liquid inner filling solutions. Therefore, the
main focus of this research was adjusted to producing ion-selective sensors which do
not require an external reference electrode. The proposed methodology was to use IDEs
to monitor changes in the electrical properties of an ion-selective membrane due to
selective responses between the target ions and the previously discussed
commercially-available ionophores.
As the intended final use of the prepared ion-selective membranes was as the
recognition layer within an impedimetric IDE device, it was important to bear in mind
that it has previously been reported that even when a particular membrane composition
has been shown to perform adequately within a potentiometric ISE, this does not
necessarily mean it will produce an equivalent selective response when used within an
ISCOM device156
.
It was decided that nitrate would be the principal target ion to be investigated using IDE
sensors. In the case of commercially-available nitrate ionophores, TDAN is
significantly less expensive than NO3V, costing around £30 per 100 mg as opposed to
around £250 per 100 mg, and therefore would offer a much more cost effective option
for nitrate sensing within a commercial product. TDAN also does not require an
additional ionic additive to be added to the membrane. Consequently, initial proof-of-
principle developmental work for nitrate sensing using IDEs was carried out using the
ionophore TDAN.
Craig Alexander
140
4 Nitrate Sensing using IDE Design 1
4.1 Sensor Calibration
It was first important to establish the behaviour of IDE Design 1 in the absence of an
ion-selective membrane to ensure that a response was obtained from changes in nitrate
concentration in the important range for aquarium purposes (1–100 ppm). Figure 4.1
shows capacitance (A) and conductance (B) spectra obtained from an uncoated sensor,
via addition of aliquots of NO3- stock solution to a deionised water background.
Figure 4.1: Conductance (A) and capacitance (B) spectra for the addition of NO3
- to an uncoated
IDE Design 1 (1 Vrms input amplitude). ‘Pure water’ refers to the initial baseline value and
10-60 mins are the values obtained during the baseline settling period
0.00E+00
1.00E-07
2.00E-07
3.00E-07
4.00E-07
5.00E-07
6.00E-07
7.00E-07
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 10 mins
Cp 20 mins
Cp 30 mins
Cp 40 mins
Cp 50 mins
Cp 60 mins
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
5.00E-04
1.00E-03
1.50E-03
2.00E-03
2.50E-03
3.00E-03
3.50E-03
4.00E-03
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Co
nd
uct
ance
(S)
Frequency (Hz)
G Pure water
G 10 mins
G 20 mins
G 30 mins
G 40 mins
G 50 mins
G 60 mins
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
141
Six frequencies across each spectra were evaluated: 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1
MHz and 2 MHz. Figure 4.2 shows the change in conductance against time upon
addition of nitrate at each chosen frequency, whilst Figure 4.3 shows the capacitance.
Figure 4.2: Conductance response of uncoated IDE Design 1 upon addition of 1–100 ppm nitrate at
six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
For the conductance, most of the six extrapolated frequencies, with the exception of the
lowest frequency at 100 Hz, showed a ‘step-wise’ increase to each of the added aliquots
of potassium nitrate stock solution. The response time of the sensor was almost
immediate (<1 minute), as very little drift was observed in any of the measurements.
A B
C
E
D
F
1 hour baseline
1 ppm NO3-
10 ppm NO3-
20 ppm NO3- 30 ppm NO3
-
40 ppm NO3-
50 ppm NO3-
100 ppm NO3-
Craig Alexander
142
Figure 4.3: Capacitance response of uncoated IDE Design 1 upon addition of 1–100 ppm nitrate at
six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
For the capacitance, the three lowest extrapolated frequencies (100 Hz, 1 kHz and 10
kHz) showed a ‘step-wise’ increase as the test solution conductivity was increased
through the addition of a potassium nitrate stock solution. At 100 kHz, a slight decrease
was noted at low nitrate concentrations (<20 ppm), followed by ‘step-wise’ increases
from 30–100 ppm. At 1 MHz and 2 MHz there was a decrease in the measured
capacitance as potassium nitrate aliquots were added to the test sample.
The response obtained from the deionised water baseline (G0 or C0) was subtracted from
average of the measured response obtained from the five data points obtained from each
nitrate addition (Gm or Cm). The change in conductance (Gm - G0) or capacitance
(Cm - C0) was plotted against log NO3- concentration. Figure 4.4 shows the conductance
A B
C D
E F
Craig Alexander
143
and Figure 4.5 shows the capacitance. By subtracting the baseline from the measured
response this also negated any electrical differences in the measurements caused by
cables or connectors. Therefore, the change in the electrical impedance observed is only
due to the addition of aliquots of target/interfering ion solutions to the initial
background solution.
Figure 4.4: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for uncoated IDE Design 1 at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
At 1 kHz, the addition of 100 ppm NO3- resulted in an increase in the measured
conductance from the baseline deionised water value of 0.87 mS. The Gm - G0 values for
100 ppm NO3- at the other extrapolated frequencies were 2.75 mS at 10 kHz, 3.36 mS at
100 kHz, 3.39 mS at 1 MHz and 3.41 mS at 2 MHz.
A B
C D
E F
Craig Alexander
144
Figure 4.5: Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3
-
concentration for uncoated IDE Design 1 at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
At 100 Hz, an increase in the measured capacitance of 436 nF was observed through the
addition of 100 ppm NO3-. At 1 kHz, this resulted in a Cm - C0 value of 169 nF, and at
10 kHz, Cm - C0 = 14.5 nF. At a frequency of 100 kHz, a slight decrease of
approximately 30 pF from the baseline deionised water value was observed up to
20 ppm NO3-, followed by an increase to 290 pF at 100 ppm NO3
-. Both 1 MHz and
2 MHz showed a decrease of 36 pF from the baseline response.
It has previously been reported for similar devices that plotting the logarithm of the
baseline-subtracted response against the logarithm of the ion concentration produces a
linear relationship176
. This information is shown in Figures 4.6 for G and Figure 4.7
for C.
A B
C D
E F
Craig Alexander
145
Figure 4.6: Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for uncoated IDE Design 1 at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
The log (Gm–G0) versus log NO3- graph resulted in a linear increase (R
2 >0.98) for each
of the extrapolated frequencies, with the exception of 100 Hz, where a linear response
was not observed.
A B
C D
E F
Craig Alexander
146
Figure 4.7: Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate
concentration for uncoated IDE Design 1 at three frequencies (negative values obtained at 100 kHz,
1 MHz and 2 MHz). A – 100 Hz. B – 1 kHz. C – 10 kHz
The log (Cm–C0) versus log NO3- graph resulted in a linear increase (R
2 >0.98) at
100 Hz, 1 kHz and 10 kHz.
It was established that IDE Design 1 was able to respond to changes in the conductivity
of the bulk test solution, through additions of KNO3 to a deionised water background.
The most analytically-relevant information appeared to arise from the change in
conductance, particularly at higher frequencies, as this is where the most marked
responses were observed. Further experimentation was carried out on this device, using
a membrane containing a commercially-available ionophore to coat the sensing area, in
an attempt to impart nitrate selectivity on the sensor. This would ascertain whether IDE
Design 1 was suitable to be used as a nitrate sensor within a device for monitoring the
water quality of freshwater aquaria.
A B
C
Craig Alexander
147
4.2 PVC Membrane Sensors
4.2.1 Calibrations
Figures 4.8A and 4.8B show the conductance and capacitance spectra, respectively, for
the addition of nitrate in the concentration range 1–100 ppm, to IDE Design 1 spin-
coated at 2000 rpm in a PVC membrane containing 1% w/w TDAN as the ionophore.
This particular device will be referred to as ‘TDAN PVC spin-coated IDE Design 1’.
Figure 4.8: Conductance (A) and capacitance (B) spectra for the addition of NO3
- to TDAN PVC
spin-coated IDE Design 1 (1 Vrms input amplitude). ‘Pure water’ refers to the initial baseline value
and 10-60 mins are the values obtained during the baseline settling period
0.00E+00
5.00E-08
1.00E-07
1.50E-07
2.00E-07
2.50E-07
3.00E-07
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 10 mins
Cp 20 mins
Cp 30 mins
Cp 40 mins
Cp 50 mins
Cp 60 mins
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
5.00E-04
1.00E-03
1.50E-03
2.00E-03
2.50E-03
3.00E-03
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Con
du
ctan
ce (S
)
Frequency (Hz)
G Pure water
G 10 mins
G 20 mins
G 30 mins
G 40 mins
G 50 mins
G 60 mins
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
148
When IDE Design 1 was spin-coated with a PVC membrane containing the ionophore
TDAN, it was more difficult to distinguish between each nitrate addition at many of the
extrapolated frequencies when compared to the results for the uncoated device, and the
‘step-wise’ response that was observed was not as prominent. Figure 4.9 shows the
change in conductance against time upon addition of nitrate at each chosen frequency,
whilst Figure 4.10 shows the capacitance.
Figure 4.9: Conductance response of TDAN PVC spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
A B
C D
E F
Craig Alexander
149
Figure 4.10: Capacitance response of TDAN PVC spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
When TDAN PVC was spin-coated on to IDE Design 1, the most prominent response
upon addition of the target nitrate anion to the test vessel was obtained from the
conductance at a frequency of 2 MHz. The deionised water baseline-subtracted response
for this frequency is shown in Figure 4.11.
A B
C D
E F
Craig Alexander
150
Figure 4.11: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for TDAN PVC spin-coated IDE Design 1 at 2 MHz
There was a total increase in the measured conductance at 2 MHz of 1.77 mS upon the
addition of 100 ppm NO3- to a deionised water baseline. The logarithmic
baseline-subtracted conductance against log NO3- concentration at 2 MHz is shown in
Figure 4.12.
Figure 4.12: Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for TDAN PVC spin-coated IDE Design 1 at 2 MHz
Craig Alexander
151
Taking the logarithm of the deionised water baseline-subtracted conductance at 2 MHz
and plotting it against log NO3- concentration confirmed the linearity of the response,
with a R2 value of 0.9931.
Although the logarithm of the baseline response versus the log of the nitrate
concentration was used to determine if the calibration was linear, the
baseline-subtracted values were more beneficial to compare responses from different
sensors and for comparing the performance of the same sensor in different sample
matrices. Figure 4.13 shows the baseline-subtracted conductance at 2 MHz for the
addition of nitrate in the concentration range 1–100 ppm for TDAN PVC spin-coated
IDE Design 1, comparing the aforementioned calibration experiment, using a potassium
nitrate stock solution to adjust nitrate concentration, with the response from an
alternative counter ion (lithium) nitrate salt and the effect of the response when the
temperature of the sample solution was increased to from 25 °C to 30 °C.
Figure 4.13: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for TDAN PVC spin-coated IDE Design 1 at 2 MHz. Three calibration experiments
are shown: the addition of 1–100 ppm NO3- with K
+ as the counter ion, the addition of 1–100 ppm
NO3- with Li
+ as the counter ion and the addition of 1–100 ppm NO3
- with K
+ as the counter ion at
30 °C
0
0.0005
0.001
0.0015
0.002
0.0025
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
KNO3 Calibration
LiNO3 Calibration
30 °C Calibration
Craig Alexander
152
When lithium nitrate was used to adjust the nitrate concentration in the test vessel there
was a decrease observed in the recorded measurements of around 20% when compared
with the use of potassium nitrate. At 50 ppm NO3-, which is the upper limit within a
freshwater aquarium, a change in the measured conductance of 1.39 mS was observed
when potassium nitrate was used to adjust the nitrate concentration. When lithium
nitrate was added, the Gm - G0 for 50 ppm NO3- was 1.14 mS. This suggested that the
ionic mobility of the counter ion was affecting the response from the sensor, as the less
ionically-mobile Li+
counter ion produced less of a response than when K+ was used.
Furthermore, when the experiment was conducted using KNO3 to adjust the nitrate
concentration at 30 °C, there was an increase of between 13–19% at each concentration
than when the same experiment was conducted at 25 °C. This suggested that changes
within the solution were being observed, as an increase in temperature would cause an
increase in the conductivity of the solution due to increased ionic mobility.
To establish the effect of changes in the background conductivity of the sample
solution, the sensor was tested through additions of nitrate to different test solutions of
varying ionic strength. Figure 4.14 shows the measured G0 values for each of the
tested environments.
Craig Alexander
153
Figure 4.14: Measured baseline conductance value (G0) of TDAN PVC spin-coated IDE Design 1 in
six separate sample matrices at 2 MHz, prior to the addition of nitrate
From Figure 4.14, it was evident that placing the sensor into an ionic solution increased
the observed baseline conductance value at 2 MHz. A 100 ppm solution of MgSO4
produced a G0 value 1.53 mS greater than the deionised water baseline. 200 ppm
MgSO4 was 1.95 mS greater and 500 ppm MgSO4 was 2.15 mS greater than the
deionised water baseline. A sample of tap water gave a TDS reading of 44 ppm when
taken with a handheld TDS meter, and increased the G0 by 0.89 mS. A sample of water
was taken from a live freshwater aquarium, the recorded TDS value was 75 ppm, there
was no detectable existing NO3- when tested with a commercial test kit, and the G0
increased by 1.49 mS compared to the G0 obtained for deionised water. The measured
conductance responses upon addition of NO3- over the range 1–100 ppm to each of
these background solutions are shown in Figure 4.15.
Deionised water
100 ppm MgSO4
200 ppm MgSO4 500 ppm MgSO4 Tap water (44 ppm TDS)
Aquarium water (75 ppm TDS)
Background sample matrix
G0
bas
eli
ne
val
ue a
t 2
MH
z
Craig Alexander
154
Figure 4.15: Measured conductance response (Gm) of TDAN PVC spin-coated IDE Design 1 against
log NO3- concentration in six separate sample matrices at 2 MHz
The corresponding baseline-subtracted conductance responses are shown in Figure 4.16.
Figure 4.16: Baseline-subtracted conductance (Gm – G0) of TDAN PVC spin-coated IDE Design 1
against log NO3- concentration at 2 MHz, in six separate sample matrices
As the overall background ionic strength of the sample solution was increased, the
sensitivity of the response of the sensor to changes in NO3- concentration was reduced.
This suggested that the sensor was heavily influenced by the background ionic strength
of the test solution.
0
0.0005
0.001
0.0015
0.002
0.0025
0.003
0.0035
0 0.5 1 1.5 2 2.5
Gm
(S)
log NO3- concentration (ppm)
Pure water calibration (G0)
100 ppm MgSO4
200 ppm MgSO4
500 ppm MgSO4
Tap water (44 ppm TDS)
Aquarium water (75 ppm TDS)
0
0.0002
0.0004
0.0006
0.0008
0.001
0.0012
0.0014
0.0016
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
Pure water calibration (G0)
100 ppm MgSO4
200 ppm MgSO4
500 ppm MgSO4
Tap water (44 ppm TDS)
Aquarium water (75 ppm TDS)
Craig Alexander
155
Figure 4.17 shows the deionised water baseline-subtracted conductance at 2 MHz for
the addition of NO3- to three separate TDAN PVC spin-coated IDE Design 1 devices.
Figure 4.17: Deionised water baseline-subtracted conductance (Gm – G0) of three TDAN PVC
spin-coated IDE Design 1 devices against log NO3- concentration at 2 MHz, for reproducibility
determination
The device-to-device reproducibility of the impedimetric devices was defined as the
%RSD of the three separate baseline-subtracted conductance measurements obtained for
50 ppm NO3- (upper limit for NO3
- in an aquarium). For TDAN PVC spin-coated IDE
Design 1 at 2 MHz, this was calculated as 17.15%.
4.2.2 Selectivity
The nitrate selectivity of TDAN PVC spin-coated IDE Design 1 was tested using the
FIM, however, the method deviated slightly from the one described in section 2.7.2.4,
as 1 ppm NO2- and 10 ppm Cl
- were added to the test vessel prior to the addition of the
target ion, not 10 ppm NO2- and 50 ppm Cl
- as is stated in Chapter 2. This was to
establish the effect of additions of low concentrations of the interfering ions on the
sensor response. The selectivity data are shown in Figure 4.18.
0
0.0005
0.001
0.0015
0.002
0.0025
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
TDAN IDE Design 1 - 1
TDAN IDE Design 1 - 2
TDAN IDE Design 1 - 3
Craig Alexander
156
Figure 4.18: Selectivity determination of TDAN PVC spin-coated IDE Design 1. Comparison of the
deionised water baseline-subtracted conductance response at 2 MHz to 1–100 ppm nitrate: when no
interfering anion is present, when a fixed concentration of 1 ppm nitrite is present and when a fixed
concentration of 10 ppm chloride is present. The response to 1–100 ppm nitrate of a ‘blank’
membrane, which did not contain TDAN as the ionophore, is also shown
At a frequency of 2 MHz, an increase in the obtained response was observed when
interfering nitrite and chloride anions were present in the test sample. At 100 ppm NO3-,
the baseline-subtracted conductance was 0.24 mS greater when 1 ppm nitrite was
present, and it was 0.37 mS greater when 10 ppm chloride was present, compared with
the deionised water baseline when no interfering anion was present. A large response
was observed from the ‘blank’ membrane which was prepared without ionophore; at
lower concentrations (<20 ppm) the response of the ‘blank’ was around 50% of the
‘active’ membrane, which did contain ionophore, however at high concentrations
(>50 ppm) the ‘blank’ response was over 100% of that of the ‘active’ membrane.
The response of IDE Design 1, when spin-coated with a PVC membrane containing
TDAN as a nitrate-selective ionophore, appeared to be too susceptible to changes in the
background ionic strength of the test solution, and therefore did not provide adequate
nitrate selectivity. The effect of the membrane composition on the sensor response was
0
0.0005
0.001
0.0015
0.002
0.0025
0.003
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
KNO3 Calibration
1 ppm Nitrite FIM
10 ppm Chloride FIM
Blank membrane
Craig Alexander
157
investigated by coating the sensing area of IDE Design 1 in an alternative membrane
material based on ionophore-doped sol-gels.
4.3 Sol-gel Membrane Sensors
Sol-gel membranes were investigated as an alternative membrane material for use
within impedimetric IDE devices for nitrate sensing, by incorporating the commercial
ionophore TDAN into the sol-gel mixture during its preparation, as described in section
2.1.3. A 3:1 ratio of the silica precursors DEDMS:TEOS were used as the sol-gel
membrane material. The sol-gels were spin-coated over the sensing area of IDE Design
1 at a speed of 2000 rpm. This particular device will be referred to as ‘TDAN DEDMS
spin-coated IDE Design 1’.
4.3.1.1. DEDMS/TEOS Membrane
4.3.1.1.1. Calibrations
Figures 4.19A and 4.19B show the conductance and capacitance spectra, respectively,
for the addition of nitrate in the concentration range 1–100 ppm, to IDE Design 1
spin-coated at 2000 rpm in a sol-gel membrane prepared from the silica pre-cursors
DEDMS/TEOS containing TDAN as the ionophore. The corresponding response versus
time data at each extrapolated frequency is shown in Figure 4.20 (G) and
Figure 4.21 (C).
Craig Alexander
158
Figure 4.19: Conductance (A) and capacitance (B) spectra for the addition of NO3
- to TDAN
DEDMS spin-coated IDE Design 1 (1 Vrms input amplitude). ‘Pure water’ refers to the initial
baseline value and 10-60 mins are the values obtained during the baseline settling period
0.00E+00
1.00E-08
2.00E-08
3.00E-08
4.00E-08
5.00E-08
6.00E-08
7.00E-08
8.00E-08
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 10 mins
Cp 20 mins
Cp 30 mins
Cp 40 mins
Cp 50 mins
Cp 60 mins
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
1.00E-03
2.00E-03
3.00E-03
4.00E-03
5.00E-03
6.00E-03
7.00E-03
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Co
nd
uct
ance
(S)
Frequency (Hz)
G Pure water
G 10 mins
G 20 mins
G 30 mins
G 40 mins
G 50 mins
G 60 mins
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
159
Figure 4.20: Conductance response of TDAN DEDMS spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
The corresponding logarithmic baseline subtracted conductance versus log nitrate
concentration produced a linear response at 100 kHz, 1 MHz and 2 MHz. These data are
shown in Figure 4.22. At 100 kHz, there was a total increase in the measured
conductance of 2.00 mS upon the addition of 100 ppm NO3- to a deionised water
background. At a frequency of 1 MHz, the addition of 100 ppm NO3- resulted in an
increase in the measured conductance from the deionised water baseline value of
4.82 mS, and at 2 MHz this value was 5.16 mS.
0.00E+00
1.00E-03
2.00E-03
3.00E-03
4.00E-03
5.00E-03
6.00E-03
7.00E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
G @ 2 MHz
0.00E+00
1.00E-03
2.00E-03
3.00E-03
4.00E-03
5.00E-03
6.00E-03
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
G @ 1 MHz
0.00E+00
5.00E-04
1.00E-03
1.50E-03
2.00E-03
2.50E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
G @ 100 kHz
0.00E+00
5.00E-05
1.00E-04
1.50E-04
2.00E-04
2.50E-04
3.00E-04
3.50E-04
4.00E-04
4.50E-04
5.00E-04
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
G @ 10 kHz
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
8.00E-05
9.00E-05
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
G @ 1 kHz
0.00E+00
2.00E-06
4.00E-06
6.00E-06
8.00E-06
1.00E-05
1.20E-05
1.40E-05
1.60E-05
1.80E-05
2.00E-05
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
G @ 100 Hz
A B
C D
E F
Craig Alexander
160
Figure 4.21: Capacitance response of TDAN DEDMS spin-coated IDE Design 1 upon addition of
1–100 ppm nitrate at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
The corresponding logarithmic baseline subtracted capacitance versus log nitrate
concentration produced a linear response at 1 kHz, 10 kHz and 100 kHz. These data are
shown in Figure 4.23. At 1 kHz, there was a total increase in the measured capacitance
of 21.05 nF upon the addition of 100 ppm NO3- to a deionised water background. This
frequency produced the least stable measurements as there appeared to be some drift
associated between the values obtained at each nitrate concentration, particularly upon
addition of 30, 40 and 50 ppm NO3-. At a frequency of 10 kHz, the addition of 100 ppm
NO3- resulted in an increase in the measured capacitance from the deionised water
baseline value of 9.78 nF, and at 100 kHz this value was 2.60 nF.
0.00E+00
5.00E-11
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
Cp @ 2 MHz
0.00E+00
5.00E-11
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
4.00E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
Cp @ 1 MHz
0.00E+00
5.00E-10
1.00E-09
1.50E-09
2.00E-09
2.50E-09
3.00E-09
3.50E-09
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
Cp @ 100 kHz
0.00E+00
2.00E-09
4.00E-09
6.00E-09
8.00E-09
1.00E-08
1.20E-08
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
Cp @ 10 kHz
0.00E+00
5.00E-09
1.00E-08
1.50E-08
2.00E-08
2.50E-08
3.00E-08
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
Cp @ 1 kHz
0.00E+00
5.00E-09
1.00E-08
1.50E-08
2.00E-08
2.50E-08
3.00E-08
3.50E-08
4.00E-08
4.50E-08
5.00E-08
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
Cp @ 100 Hz
A B
C D
E F
Craig Alexander
161
Figure 4.22: Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for TDAN DEDMS spin-coated IDE Design 1.
A – 100 kHz. B – 1 MHz. C – 2 MHz
The logarithmic baseline-subtracted conductance produced a linear response each of the
three extrapolated frequencies over the concentration range 1–100 ppm NO3-. At
100 kHz the R2 value was 0.9985, at 1 MHz it was 0.9918 and at 2 MHz it was 0.9902.
Figure 4.23: Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate
concentration for TDAN DEDMS spin-coated IDE Design 1.
A – 1 kHz. B – 10 kHz. C – 100 kHz
y = 0.652x - 3.6643R² = 0.9918
-4
-3.5
-3
-2.5
-2
-1.5
-1
-0.5
0
0 0.5 1 1.5 2 2.5
log
(Gm
-G
0) (S
)
log NO3- concentration (ppm)
y = 0.4892x - 3.6598R² = 0.9985
-4
-3.5
-3
-2.5
-2
-1.5
-1
-0.5
0
0 0.5 1 1.5 2 2.5
log
(Gm
-G
0) (S
)
log NO3- concentration (ppm)A B
y = 0.6759x - 3.6911R² = 0.9902
-4
-3.5
-3
-2.5
-2
-1.5
-1
-0.5
0
0 0.5 1 1.5 2 2.5
log
(Gm
-G
0) (S
)
log NO3- concentration (ppm)C
y = 0.7927x - 10.171R² = 0.9975
-10.4
-10.2
-10
-9.8
-9.6
-9.4
-9.2
-9
-8.8
-8.6
-8.4
0 0.5 1 1.5 2 2.5
log
(Cm
-C
0) (
F)
log NO3- concentration (ppm)
y = 0.45x - 8.9104R² = 0.9988
-9
-8.9
-8.8
-8.7
-8.6
-8.5
-8.4
-8.3
-8.2
-8.1
-8
-7.9
0 0.5 1 1.5 2 2.5
log
(Cm
-C
0) (
F)
log NO3- concentration (ppm)
y = 0.3631x - 8.3737R² = 0.9909
-8.5
-8.4
-8.3
-8.2
-8.1
-8
-7.9
-7.8
-7.7
-7.6
0 0.5 1 1.5 2 2.5
log
(Cm
-C
0) (
F)
log NO3- concentration (ppm)
A B
C
Craig Alexander
162
The logarithmic baseline subtracted capacitance produced a linear response at each of
the three extrapolated frequencies over the concentration range 1–100 ppm NO3-. At
1 kHz the R2 value was 0.9975, at 10 kHz it was 0.9988 and at 2 MHz it was 0.9909.
Figure 4.24 shows the baseline-subtracted conductance at the three extrapolated
frequencies for the addition of nitrate in the concentration range 1–100 ppm for TDAN
DEDMS spin-coated IDE Design 1. This figure compares the calibration experiment
using potassium nitrate to adjust the nitrate concentration with the response for the use
of lithium nitrate, and the effect on the observed response when the temperature of the
sample solution was increased from 25 °C to 30 °C. The effect of the capacitive
response under these conditions is displayed in Figure 4.25.
Figure 4.24: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for TDAN DEDMS spin-coated IDE Design 1. Three calibration experiments are
shown: the addition of 1–100 ppm NO3- with K
+ as the counter ion, the addition of 1–100 ppm NO3
-
with Li+ as the counter ion and the addition of 1–100 ppm NO3
- with K
+ as the counter ion at 30 °C
A – 100 kHz. B – 1 MHz. C – 2 MHz
0
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration
0
0.001
0.002
0.003
0.004
0.005
0.006
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration
0
0.0005
0.001
0.0015
0.002
0.0025
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration
A B
C
Craig Alexander
163
At 100 kHz, there was a decrease observed in the recorded conductance measurements
of around 60% when lithium nitrate was used to adjust the nitrate concentration,
compared with the corresponding value when potassium nitrate was used. At 50 ppm
NO3- a Gm - G0 value of 1.50 mS was observed when potassium nitrate was used to
adjust the nitrate concentration within the test vessel. This value decreased to 0.55 mS
when lithium nitrate was used. An increase in the temperature of the test solution from
25 °C to 30 °C resulted in an increase in the conductance response by approximately
5%. At 1 MHz, the obtained Gm - G0 value for 50 ppm NO3- was reduced from 2.87 mS,
when KNO3 was added, to 1.22 mS when LiNO3 was added. The Gm - G0 value for
50 ppm NO3- at 30 °C was 3.12 mS. At 2 MHz, the obtained Gm - G0 value for 50 ppm
NO3- was reduced from 2.97 mS, when KNO3
was added, to 1.28 mS when LiNO3 was
added. The Gm - G0 value for 50 ppm NO3- at 30 °C was 3.32 mS.
Figure 4.25: Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3
-
concentration for TDAN DEDMS spin-coated IDE Design 1. Three calibration experiments are
shown: the addition of 1–100 ppm NO3- with K
+ as the counter ion, the addition of 1–100 ppm NO3
-
with Li+ as the counter ion and the addition of 1–100 ppm NO3
- with K
+ as the counter ion at 30 °C
A – 1 kHz. B – 10 kHz. C – 100 kHz
0
1E-08
2E-08
3E-08
4E-08
5E-08
6E-08
7E-08
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
0
2E-09
4E-09
6E-09
8E-09
1E-08
1.2E-08
0 0.5 1 1.5 2 2.5
Cm
-C
o (F
)
log NO3- concentration (ppm)
0
5E-10
1E-09
1.5E-09
2E-09
2.5E-09
3E-09
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
164
At 1 kHz, there was a decrease observed in the recorded capacitance measurements of
around 65% when lithium nitrate was used to adjust the nitrate concentration, compared
with the corresponding value when potassium nitrate was used. At 50 ppm NO3-, a
Cm - C0 value of 16.85 nF was observed when potassium nitrate was used to adjust the
nitrate concentration within the test vessel. This value decreased to 5.62 nF when
lithium nitrate was used. An increase in the temperature of the test solution from 25 °C
to 30 °C resulted in a large increase in the capacitive response. At high nitrate
concentrations (>50 ppm) the response was over two times greater than what was
recorded from the sensor at 25 °C. At 10 kHz, the obtained Cm - C0 value for 50 ppm
NO3-
was reduced from 7.23 nF when KNO3 was added to 2.37 nF when LiNO3 was
added. The effect of temperature on the capacitive response was less marked than the
response observed at 1 kHz, and a slight decrease was observed compared with the
capacitive measurement at 25 °C. The Cm - C0 value for 50 ppm NO3- at 30 °C was
5.96 nF. At 100 kHz, the obtained Cm - C0 value for 50 ppm NO3- was reduced from
1.56 nF, when KNO3 was added, to 0.71 nF when LiNO3 was added. The Cm - C0 value
for 50 ppm NO3- at 30 °C was 1.55 nF; therefore, little effect on the capacitive response
at 100 kHz was observed from increasing the temperature of the test solution.
Figure 4.26 shows the measured G0 values at each of the three extrapolated frequencies
in four solutions of varying ionic strength. The C0 values are shown in Figure 4.27.
Craig Alexander
165
Figure 4.26: Measured baseline conductance value (G0) of TDAN DEDMS spin-coated IDE Design
1 in six separate sample matrices prior to the addition of nitrate
A – 100 kHz. B – 1 MHz. C – 2 MHz
Placing the TDAN DEDMS spin-coated IDE Design 1 into an ionic solution increased
the observed baseline conductance value at each of the extrapolated frequencies. At
100 kHz, a 100 ppm solution of MgSO4 produced a G0 value 1.59 mS greater than the
deionised water baseline. A 200 ppm solution of MgSO4 was 2.03 mS greater and
500 ppm MgSO4 was 3.01 mS greater than the deionised water baseline. A sample of
tap water gave a TDS reading of 55 ppm when taken with a handheld TDS meter, and
increased the G0 by 8.55 mS. A sample of water was taken from a live freshwater
aquarium, the recorded TDS value was 192 ppm and the result from the commercial test
kit showed approximately 5 ppm NO3- was present; the G0 value was increased by
2.18 mS when compared with the G0 obtained for deionised water.
At 1 MHz, a 100 ppm solution of MgSO4 produced a G0 value 3.17 mS greater than the
deionised water baseline. A 200 ppm solution of MgSO4 was 4.83 mS greater and
500 ppm MgSO4 was 7.96 mS greater than the deionised water baseline. The sample of
0
0.002
0.004
0.006
0.008
0.01
0.012
G0
bas
elin
e va
lue
at 2
MH
z
Background sample matrix
0
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.01
G0
ba
seli
ne
va
lue
at
1 M
Hz
Background sample matrix0
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.01
G0
bas
elin
e va
lue
at 1
00 k
Hz
Background sample matrix
A B
C
Craig Alexander
166
tap water increased the G0 by 8.98 mS and the sample of water taken from an aquarium
increased the G0 by 5.42 mS.
At 2 MHz, a 100 ppm solution of MgSO4 produced a G0 value 3.25 mS greater than the
deionised water baseline. A 200 ppm solution of MgSO4 was 5.25 mS greater and
500 ppm MgSO4 was 9.18 mS greater than the deionised water baseline. The sample of
tap water increased the G0 by 8.92 mS and the sample of water taken from an aquarium
increased the G0 by 5.88 mS.
Figure 4.27: Measured baseline capacitance value (C0) of TDAN DEDMS spin-coated IDE Design 1
in six separate sample matrices prior to the addition of nitrate
A – 1 kHz. B – 10 kHz. C – 100 kHz
An increase in the C0 values was also observed for TDAN DEDMS spin-coated IDE
Design 1. At 1 kHz, a 100 ppm solution of MgSO4 produced a C0 value 36.38 nF
greater than the deionised water baseline. A 200 ppm solution of MgSO4 was 64.59 nF
greater and 500 ppm MgSO4 was 110.49 nF greater than the deionised water baseline.
The sample of tap water increased the C0 by 2.51 nF and the sample of water taken from
an aquarium increased the C0 by 66.45 nF.
0
5E-10
1E-09
1.5E-09
2E-09
2.5E-09
3E-09
3.5E-09
4E-09
C0
bas
eli
ne
val
ue
at 1
00
kH
z
Background sample matrix
0
2E-09
4E-09
6E-09
8E-09
1E-08
1.2E-08
1.4E-08
1.6E-08
1.8E-08
2E-08
C0
bas
elin
e v
alu
e a
t 1
0 k
Hz
Background sample matrix0.00E+00
2.00E-08
4.00E-08
6.00E-08
8.00E-08
1.00E-07
1.20E-07
1.40E-07
C0
ba
selin
e v
alu
e a
t 1
kH
z
Background sample matrix
A B
C
Craig Alexander
167
At 10 kHz, a 100 ppm solution of MgSO4 produced a C0 value 6.94 nF greater than the
deionised water baseline. A 200 ppm solution of MgSO4 was 8.03 nF greater and
500 ppm MgSO4 was 16.48 nF greater than the deionised water baseline. The sample of
tap water increased the C0 by 2.14 nF and the sample of water taken from an aquarium
increased the C0 by 9.12 nF.
At 100 kHz, a 100 ppm solution of MgSO4 produced a C0 value 1.59 nF greater than the
deionised water baseline. A 200 ppm solution of MgSO4 was 2.18 nF greater and
500 ppm MgSO4 was 3.02 nF greater than the deionised water baseline. The sample of
tap water increased the C0 by 0.44 nF and the sample of water taken from an aquarium
increased the C0 by 2.42 nF.
As for TDAN PVC spin-coated IDE Design 1, a decrease in the sensitivity of the
response of the TDAN DEDMS spin-coated IDE Design 1 sensor was observed when
nitrate was added to a solution with increasing background ionic strength
(data not shown). This suggested that, similar to the results obtained for IDE Design 1
when a PVC ion-selective membrane was spin-coated over the sensing area, the
DEDMS/TEOS sol-gel membrane sensor was too susceptible to changes in the
background ionic strength of the solution, thus not providing adequate nitrate
selectivity. Selectivity experiments using solutions of likely interfering anions within a
freshwater aquarium were carried out to confirm the lack of selectivity towards nitrate.
4.3.1.1.2. Selectivity
As for the selectivity experiments conducted using TDAN PVC spin-coated IDE Design
1, the FIM was carried out using 1 ppm NO2- and 10 ppm Cl
- to establish the selectivity
Craig Alexander
168
of TDAN DEDMS spin-coated IDE Design 1. The deionised water baseline-subtracted
responses are shown in Figure 4.28 (G) and Figure 4.29 (C).
Figure 4.28: Selectivity determination of TDAN DEDMS spin-coated IDE Design 1 (conductive
response). A comparison of: The deionised water baseline-subtracted conductance responses to
1–100 ppm nitrate when no interfering anion is present, when a fixed concentration of 1 ppm nitrite
is present and when a fixed concentration of 10 ppm chloride is present. The response to 1–100 ppm
nitrate of a ‘blank’ membrane, which did not contain TDAN as the ionophore, is also shown
A – 100 kHz. B – 1 MHz. C – 2 MHz
At a frequency of 100 kHz, there was a notable increase in the obtained conductive
response when interfering nitrite and chloride anions were present, particularly at low
NO3- concentrations. At 1 ppm NO3
-, the deionised water baseline-subtracted
conductance was 0.08 mS greater when 1 ppm nitrite was present and the Gm - G0 value
was 0.52 mS greater than the pure nitrate measurement when 10 ppm chloride was
present.
At 1 MHz, a slight increase of around 0.05 mS was observed at NO3- concentrations
between 1–50 ppm when 1 ppm nitrite was present. At 100 ppm NO3- there was an
increase of 0.22 mS compared to when no interfering anion was present. When 10 ppm
0
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0 0.5 1 1.5 2 2.5
Gm
-G
0 (S
)
log NO3- concentration (ppm)
0
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0 0.5 1 1.5 2 2.5
Gm
-G
0 (S
)
log NO3- concentration (ppm)
0
0.0005
0.001
0.0015
0.002
0.0025
0 0.5 1 1.5 2 2.5
Gm
-G
0 (S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
169
chloride was present, the conductance response at each nitrate concentration was
increased by approximately 0.6 mS.
At 2 MHz, a slight decrease of around 0.05 mS was observed at NO3- concentrations
between 1–50 ppm when 1 ppm nitrite was present. At 100 ppm NO3- there was an
increase of 0.28 mS compared to when no interfering anion was present. When 10 ppm
chloride was present, the conductance response at each nitrate concentration was
increased by approximately 0.6 mS.
At all three extrapolated frequencies, a large response was observed from a ‘blank’
membrane which was prepared without the ionophore. At 100 ppm NO3-, the Gm - G0
value was greater for the ‘blank’ membrane than the membrane which contained
ionophore.
Figure 4.29: Selectivity determination of TDAN DEDMS spin-coated IDE Design 1 (capacitive
response). A comparison of: The deionised water baseline-subtracted capacitance responses to
1–100 ppm nitrate when no interfering anion is present, when a fixed concentration of 1 ppm nitrite
is present and when a fixed concentration of 10 ppm chloride is present. The response to 1–100 ppm
nitrate of a ‘blank’ membrane, which did not contain TDAN as the ionophore, is also shown
A – 1 kHz. B – 10 kHz. C – 100 kHz
0
5E-10
1E-09
1.5E-09
2E-09
2.5E-09
3E-09
3.5E-09
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
0
2E-09
4E-09
6E-09
8E-09
1E-08
1.2E-08
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
0
5E-09
1E-08
1.5E-08
2E-08
2.5E-08
3E-08
3.5E-08
0 0.5 1 1.5 2 2.5
Cm
-C
0 (F
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
170
As for the conductive responses, TDAN DEDMS spin-coated IDE Design 1 showed a
substantial increase in the capacitive response, at all three extrapolated frequencies, to
interfering nitrite and chloride anions. An almost equivalent response to NO3- was
observed from the device membrane when it was coated with a ‘blank’ membrane
which did not contain TDAN as the ionophore. This suggested that the response was
heavily influenced by changes in the bulk solution, and not from selective interactions
occurring within the membrane between the target ion and the ionophore. Device-to-
device reproducibility was not obtained for TDAN DEDMS spin coated IDE Design 1.
IDE Design 1 did not show adequate selectivity with either a spin-coated PVC
membrane or a spin-coated DEDMS/TEOS sol-gel membrane. The lack of selectivity
appeared to arise from the IDE transduction element, rather than due to the membrane
composition which was deposited over the sensing area. As such, a decision was made
not to investigate this particular sensor design further with either an alternative
polymeric membrane material or with an alternative nitrate-selective ionophore. When
using the FIM to determine selectivity of nitrate sensors, it was quite challenging to
deduce, at low interfering anion concentrations, whether the obtained response was
influenced by the presence of the interfering species. As such, the concentration of the
interfering anions was increased in subsequent selectivity experiments, as described in
section 2.7.2.4, so that the effect of the interfering species could be stated definitively.
4.4 Conclusions for Chapter 4
IDE Design 1 was fabricated in-house, using a lift-off photolithography technique
followed by e-beam deposition of gold metal, and tested for its feasibility as an
impedimetric nitrate-selective sensor for use within a freshwater aquarium through
spin-coating of the sensing area with a selective polymeric membrane containing a
Craig Alexander
171
commercially-available nitrate ionophore. TDAN was tested as the nitrate-selective
ionophore; PVC and silica gels from the precursors DEDMS/TEOS were used as the
membrane materials.
Conductance and capacitance spectra from the IDE devices were obtained using an LCR
meter by applying an AC voltage of 1 Vrms over a frequency range of 20 Hz–2 MHz.
Six frequencies across the spectra were extrapolated for further analysis and to establish
a frequency region of interest where an analytically-relevant response was obtained
upon addition of the target ion.
When a PVC membrane containing the ionophore TDAN was spin-coated over the
digits of IDE Design 1, a prominent response to increasing nitrate concentration was
observed from changes in the conductance at 2 MHz. When a DEDMS/TEOS sol-gel
containing the ionophore TDAN was used as the membrane material, a linear response
was observed from the change in capacitance at 1 kHz, 10 kHz and 100 kHz; and from
the change in conductance at 100 kHz, 1 MHz and 2 MHz.
It has been established that IDE Design 1, when coupled with a spin-coated, thin-film
polymer ion-selective membrane, was not suitable for use as a standalone nitrate sensor.
The geometric parameters of this particular design appeared to produce an electric field
with a penetration depth much greater than the thickness of the deposited membrane
layer. Therefore, upon addition of an ionic solution to the test vessel, the observed
response was largely generated by changes in the conductivity of the bulk sample, not
from selective interactions between the target anion and the immobilised ionophore
within the membrane layer. This conclusion can be drawn as a large response was
observed when aliquots of the target ion were added to a device that had been coated
with a ‘blank’ membrane that did not contain a selective ionophore. A response was
also observed from non-target, interfering anions such as chloride and nitrite, and a
Craig Alexander
172
large increase in the baseline measurement was observed as the ionic strength of the
background test sample solution was increased. To overcome this problem, and produce
a more selective sensor, either the thickness of the membrane layer would need to be
increased so that more of the generated electric field is contained within it, or the
geometric parameters of the IDE digits would need to be reduced so that the penetration
depth of the resulting electric field would be shorter and therefore contained within a
thin-film membrane.
The thickness of a spin-coated membrane layer could be increased by reducing the
spin-speed, although it was anticipated that reducing this to less than the current speed
(2000 rpm) would not increase the thickness sufficiently to produce a selective layer,
and it could also compromise the coating efficiency. An alternative coating process to
spin-coating could be used, such as drop-coating, although IDE Design 1 would not be
particularly well-suited to drop-coating due to the relatively large footprint of its
sensing area.
Due to the fabrication equipment that was available for producing microsensors, it was
not possible to produce IDEs with further reduced geometries in-house. A commercial
alternative was sought and IDE Design 2 was purchased from MicruX Fluidics. Its use
as an impedimetric nitrite sensor is described throughout Chapter 5. This has digit
widths and interdigital spacings of 10 µm, and therefore should produce an electric field
with a significantly shorter penetration depth when excited with an AC voltage than
IDE Design 1.
Another method for producing a thicker membrane layer is to screen-print the selective
components onto the IDE digits in a suitable support matrix, such as a dielectric screen-
printing ink. This was carried out by producing fully screen-printed IDEs, SP IDE
Craig Alexander
173
Design 1, and the results of using this device as an impedimetric nitrate sensor are
described throughout Chapter 6.
The in-house mircofabrication facilities that were available resulted in a laborious and
time-consuming fabrication process of IDE Design 1 that was only suitable for
producing relatively small batches of sensors. Small failures on individual devices, such
as broken digits or short-circuits caused by adjoining digits, would render it unsuitable
for use as an impedance sensor, and consequently the success rate of producing
electrodes that were available for testing was very low. This quite often led to several
days of manufacturing resulting in only a handful of useable devices. As a result of
these issues with fabrication and the results obtained with these sensors, a decision was
made to investigate alternative IDE sensor devices and no further testing was carried out
using IDE Design 1.
Craig Alexander
174
5 Nitrate Sensing using IDE Design 2
5.1 Sensor Calibration
The conductance (A) and capacitance (B) spectra for uncoated IDE Design 2 without a
membrane layer present responding to 1–100 ppm NO3- in a deionised water
background are shown in Figure 5.1.
Figure 5.1: Conductance (A) and capacitance (B) spectra for the addition of NO3
- to an uncoated
IDE Design 2 (1 Vrms input amplitude). ‘Pure water’ refers to the initial baseline value and
10-60 mins are the values obtained during the baseline settling period
0.00E+00
5.00E-08
1.00E-07
1.50E-07
2.00E-07
2.50E-07
3.00E-07
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 10 mins
Cp 20 mins
Cp 30 mins
Cp 40 mins
Cp 50 mins
Cp 60 mins
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
2.00E-04
4.00E-04
6.00E-04
8.00E-04
1.00E-03
1.20E-03
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Co
nd
uct
ance
(S)
Frequency (Hz)
G Pure water
G 10 mins
G 20 mins
G 30 mins
G 40 mins
G 50 mins
G 60 mins
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
175
Prior to further testing with IDE Design 2, an amendment to the LabVIEW program that
controlled the LCR meter was made so that the input amplitude could be altered. This
was seen as particularly beneficial for this project as if a response was observed using a
lower input amplitude it would allow for the production of a prototype device which
would require a much lower power consumption to function. Similar ISCOM-type
devices have also been reported using lower amplitudes than the 1 Vrms that was used
for IDE Design 1161, 176
. The deionised water baseline-subtracted results, at six
extrapolated frequencies, for uncoated IDE Design 2 upon the addition of NO3- over the
concentration range 1–100 ppm at amplitudes of 1 Vrms, 100 mVrms, 10 mVrms and
1 mVrms are shown in Figure 5.2 (G) and Figure 5.3 (C).
Figure 5.2: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for uncoated IDE Design 2 at six frequencies and varying input amplitudes
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
0
0.0002
0.0004
0.0006
0.0008
0.001
0.0012
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.0002
0.0004
0.0006
0.0008
0.001
0.0012
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.0001
0.0002
0.0003
0.0004
0.0005
0.0006
0.0007
0.0008
0.0009
0.001
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.0001
0.0002
0.0003
0.0004
0.0005
0.0006
0.0007
0.0008
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
-0.00005
0
0.00005
0.0001
0.00015
0.0002
0.00025
0.0003
0.00035
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
-0.00002
0
0.00002
0.00004
0.00006
0.00008
0.0001
0.00012
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C D
E F
Craig Alexander
176
An poor response to the increase in solution conductivity was observed when a reduced
input amplitude was used at low frequencies (below 10 kHz), compared with the 1 Vrms
experiment; however, the low frequency response was much less stable than at higher
frequencies, even at the higher input amplitude. Therefore, it would not be of analytical
importance to further investigate the low frequency results.
Figure 5.3: Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3
-
concentration for uncoated IDE Design 2 at six frequencies and varying input amplitudes
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
As for the conductance results, the capacitance response to the potassium nitrate
additions to deionised water were less pronounced at a low frequency (<1 kHz) when a
lower input amplitude was applied. However, a response was observed at 10 kHz,
100 kHz and 1 MHz when lower input amplitudes were used. At 2 MHz, the
capacitance response was erratic and did not produce an analytically-relevant response
at any of the tested amplitudes.
-1.4E-12
-1.2E-12
-1E-12
-8E-13
-6E-13
-4E-13
-2E-13
0
2E-13
4E-13
6E-13
8E-13
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
-2E-12
0
2E-12
4E-12
6E-12
8E-12
1E-11
1.2E-11
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
0
1E-10
2E-10
3E-10
4E-10
5E-10
6E-10
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
0
1E-09
2E-09
3E-09
4E-09
5E-09
6E-09
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
0
1E-08
2E-08
3E-08
4E-08
5E-08
6E-08
0 0.5 1 1.5 2 2.5C
m-
C0
(F)
log NO3- concentration (ppm)
-1E-07
-5E-08
0
5E-08
0.0000001
1.5E-07
0.0000002
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
A B
C D
E F
Craig Alexander
177
As a response in the change of conductance and capacitance was observed at lower
input amplitudes, which was seen as desirable so that a prototype device with a lower
power consumption could be produced, it was decided that subsequent experiments
using IDE Design 2 would be excited with an input AC amplitude of 10 mVrms. At
1 mVrms, the measurements were unstable and therefore this amplitude was not chosen
for further investigation.
Following the experiments conducted using IDE Design 1, it was realised that is was of
the upmost importance to ascertain the behaviour of a device that was coated with a
‘blank’ membrane that did not contain a selective ionophore. This was reflected in the
results that were obtained using IDE Design 2 as an impedimetric nitrate sensor.
5.2 PVC Membrane Sensors
5.2.1.1. Spin-coated Membranes
Initial experiments with IDE Design 2 as a potential nitrate sensor involved spin-coating
(2000 rpm) the sensing area of the device with 1 µl of a PVC membrane containing
1% w/w TDAN as the ionophore. This device will be referred to as ‘TDAN PVC
spin-coated IDE Design 2’.
5.2.1.1.1. Calibrations
Figures 5.4A and 5.4B show the conductance and capacitance spectra, respectively, for
the addition of nitrate in the concentration range 1–100 ppm, to TDAN PVC
spin-coated IDE Design 2.
Craig Alexander
178
Figure 5.4: Conductance (A) and capacitance (B) spectra for the addition of NO3
- to TDAN PVC
spin-coated IDE Design 2 (10 mVrms input amplitude). ‘Pure water’ refers to the initial baseline
value. Interference was observed at 50 Hz so this data has been omitted
Figure 5.5 shows the corresponding change in conductance against time data upon
addition of nitrate at each chosen frequency, whilst Figure 5.6 shows the
capacitance data.
0.00E+00
1.00E-09
2.00E-09
3.00E-09
4.00E-09
5.00E-09
6.00E-09
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
5.00E-06
1.00E-05
1.50E-05
2.00E-05
2.50E-05
3.00E-05
3.50E-05
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Co
nd
uct
ance
(S)
Frequency (Hz)
G Pure water
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
179
Figure 5.5: Conductance response of TDAN PVC spin-coated IDE Design 2 upon addition of
1–100 ppm nitrate at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
Adding 1–100 ppm NO3- to a deionised water background did not appear to produce a
linear response at any frequency from measuring the conductance of TDAN PVC
spin-coated IDE Design 2.
0.00E+00
5.00E-06
1.00E-05
1.50E-05
2.00E-05
2.50E-05
3.00E-05
3.50E-05
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
2.00E-06
4.00E-06
6.00E-06
8.00E-06
1.00E-05
1.20E-05
1.40E-05
1.60E-05
1.80E-05
2.00E-05
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
1.00E-06
2.00E-06
3.00E-06
4.00E-06
5.00E-06
6.00E-06
7.00E-06
8.00E-06
9.00E-06
1.00E-05
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
1.00E-06
2.00E-06
3.00E-06
4.00E-06
5.00E-06
6.00E-06
7.00E-06
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
5.00E-07
1.00E-06
1.50E-06
2.00E-06
2.50E-06
3.00E-06
3.50E-06
4.00E-06
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
2.00E-07
4.00E-07
6.00E-07
8.00E-07
1.00E-06
1.20E-06
1.40E-06
1.60E-06
1.80E-06
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E F
Craig Alexander
180
Figure 5.6: Capacitance response of TDAN PVC spin-coated IDE Design 2 upon addition of
1–100 ppm nitrate at six frequencies.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
The most analytically-relevant response from TDAN PVC spin-coated IDE Design 2
was from the capacitance response at a frequency of 2 MHz. The total addition of
100 ppm NO3- to a deionised water background resulted in a Cm – C0 of 1.15 pF. At
1 MHz, a linear increase to nitrate was observed up to 50 ppm NO3-, followed by a
slight decrease at 100 ppm. The logarithmic baseline-subtracted capacitance against log
NO3- concentration at 2 MHz is shown in Figure 5.7.
1.154E-10
1.156E-10
1.158E-10
1.160E-10
1.162E-10
1.164E-10
1.166E-10
1.168E-10
1.170E-10
0 20 40 60 80 100 120 140
Cap
acit
ance
(F)
Time (min)
1.144E-10
1.146E-10
1.148E-10
1.150E-10
1.152E-10
1.154E-10
1.156E-10
1.158E-10
1.160E-10
0 20 40 60 80 100 120 140
Cap
acit
ance
(F)
Time (min)
1.150E-10
1.155E-10
1.160E-10
1.165E-10
1.170E-10
1.175E-10
1.180E-10
1.185E-10
1.190E-10
0 20 40 60 80 100 120 140
Cap
acit
ance
(F)
Time (min)
1.28E-10
1.30E-10
1.32E-10
1.34E-10
1.36E-10
1.38E-10
1.40E-10
1.42E-10
1.44E-10
1.46E-10
1.48E-10
0 20 40 60 80 100 120 140
Cap
acit
ance
(F)
Time (min)
0.00E+00
5.00E-11
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
4.00E-10
4.50E-10
5.00E-10
0 20 40 60 80 100 120 140
Cap
acit
ance
(F)
Time (min)
0.00E+00
2.00E-10
4.00E-10
6.00E-10
8.00E-10
1.00E-09
1.20E-09
1.40E-09
1.60E-09
1.80E-09
2.00E-09
0 20 40 60 80 100 120 140
Cap
acit
ance
(F)
Time (min)
A B
C D
E F
Craig Alexander
181
Figure 5.7: Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate
concentration for TDAN PVC spin-coated IDE Design 2 at 2 MHz
Taking the logarithm of the deionised water baseline-subtracted capacitance at 2 MHz
and plotting it against log NO3- concentration produced a reasonably linear response
with a R2 value of 0.9773.
Figure 5.8 shows the deionised water baseline-subtracted capacitance at 2 MHz for the
addition of NO3- to three separate TDAN PVC spin-coated IDE Design 2 devices.
Figure 5.8: Deionised water baseline-subtracted capacitance (Cm – C0) of three TDAN PVC
spin-coated IDE Design 2 devices against log NO3- concentration, for reproducibility determination
y = 1.1092x - 14.091R² = 0.9773
-14.5
-14
-13.5
-13
-12.5
-12
-11.5
0 0.5 1 1.5 2 2.5
log
(Cm
-C
0)
(F)
log NO3- concentration (ppm)
-2E-13
0
2E-13
4E-13
6E-13
8E-13
1E-12
1.2E-12
1.4E-12
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
TDAN IDE Design 2 - 1
TDAN IDE Design 2 - 2
TDAN IDE Design 2 - 3
Craig Alexander
182
Poor device-to-device reproducibility was observed from TDAN PVC spin-coated IDE
Design 2. The third tested device showed a decrease below the baseline response up to
20 ppm NO3- before increasing at higher NO3
- concentrations. At 50 ppm NO3
-, the
%RSD was calculated as 50.90%.
5.2.1.1.2. Selectivity
As the previously tested device, IDE Design 1, was found to produce a large response to
the target nitrate anion when coated with a ‘blank’ membrane, it was deemed important
to ascertain the response of TDAN PVC IDE Design 2 when coated with a ‘blank’
membrane prior to investigating the effects of interfering anions on the response. The
comparison of the capacitance response at 2 MHz of a TDAN PVC spin-coated IDE
Design 2 device containing the ionophore with a device coated with a ‘blank’
membrane is shown in Figure 5.9.
Figure 5.9: Selectivity determination of TDAN PVC spin-coated IDE Design 2. Comparison of the
deionised water baseline-subtracted capacitance response at 2 MHz to 1–100 ppm nitrate of device
with an ‘active’ membrane, containing 1% w/w TDAN as the ionophore, and a device with a
‘blank’ membrane, which did not contain any ionophore
0
5E-13
1E-12
1.5E-12
2E-12
2.5E-12
3E-12
3.5E-12
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
KNO3 Calibration
Blank membrane
Craig Alexander
183
It was noted that a large response to nitrate was observed when a ‘blank’ PVC
membrane, which did not contain a nitrate-selective ionophore, was spin-coated onto
IDE Design 2. This suggested that the spin-coated membranes were not of an adequate
thickness to ensure a selective response; however, the observed response was largely
caused by non-selective changes in the conductivity of the test solution, rather than
target ion/ionophore interactions within the membrane. Due to the small overall
footprint of IDE Design 2, it was possible to simply drop-coat the membrane solution
over the sensing area, to produce a PVC membrane layer with increased thickness and
the added benefit of not wasting any of the ‘cocktail’ solution. Further testing of IDE
Design 2 was therefore carried out by depositing the PVC ion-selective membrane
‘cocktail’ by drop-coating.
5.2.1.2. Drop-coated Membranes
IDE Design 2 was prepared as a nitrate-selective sensor by drop-coating 1 µl of a
10% w/v PVC membrane ‘cocktail’ containing 1% w/w TDAN as the ionophore onto
the surface of the sensing area. This was to afford a thicker membrane layer and
therefore impart greater ion-selectivity on the sensor than the previously tested
spin-coated membrane electrodes, due to a larger proportion of the generated electric
field from the IDEs being enclosed within the membrane layer. This device will be
referred to as ‘TDAN PVC drop-coated IDE Design 2’.
Craig Alexander
184
5.2.1.2.1. Calibrations
Figures 5.10 and 5.11 show the measured conductance and capacitance, respectively,
for a TDAN PVC drop-coated IDE Design 2 against time for a calibration over
1–100 ppm NO3- in a deionised water background. Five extrapolated frequencies from
the resulting spectra are shown. There was no discernible response obtained at 100 Hz
therefore this data has not been displayed.
Figure 5.10: Conductance response versus time of TDAN PVC drop-coated IDE Design 2
(10% w/v) upon addition of 1–100 ppm nitrate at five frequencies.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
Craig Alexander
185
Figure 5.11: Capacitance response versus time of TDAN PVC drop-coated IDE Design 2
(10% w/v) upon addition of 1–100 ppm nitrate at five frequencies.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
It was evident from both the conductance and capacitance results that when a calibration
for IDE Design 2 containing a drop-coated membrane was carried out in the previously
described manner, insufficient time had elapsed for a stable measurement to be taken at
each NO3- concentration. As the membrane thickness has been increased due to
depositing it via drop-coating rather than spin-coating as before, the diffusion distance
for the ions to move to into the membrane layer will have also increased, which in turn,
will have resulted in a longer response time for the sensor. From the results shown in
Figure 5.10 and Figure 5.11, it was not possible to accurately deduce the
baseline-subtracted measurement at any frequency for either the conductive or
capacitive response due to the ‘drift’ associated with this longer response time. It was,
therefore, difficult to distinguish between the end of one NO3- concentration addition
and the beginning of the next. It was also noted that with this particular membrane
composition, a much longer pre-soaking period was required to ensure a stable baseline
Craig Alexander
186
was reached quickly during the measurements. Further experimentation was therefore
carried out to establish both the soaking time required and the response time of this
particular membrane composition when drop-coated onto IDE Design 2.
The baseline settling time of TDAN PVC drop-coated IDE Design 2 was established by
placing a newly-prepared sensor into a background sample of deionised water and
monitoring the response every two hours until a stable measurement was reached. These
data are displayed in Figure 5.12 (G) and Figure 5.13 (C).
Figure 5.12: Conductance deionised water baseline settling response of a dry TDAN PVC
drop-coated IDE Design 2 (10% w/v) at five frequencies
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz, the conductance response showed a slight increase over the initial 12 hours;
however, the measured response continued to fluctuate throughout the 96 hours over
which the experiment was conducted. At 10 kHz and 100 kHz the conductance response
0.00E+00
1.00E-04
2.00E-04
3.00E-04
4.00E-04
5.00E-04
6.00E-04
0 20 40 60 80 100
Co
nd
uct
ance
(S)
Time (hours)
0.00E+00
5.00E-05
1.00E-04
1.50E-04
2.00E-04
2.50E-04
3.00E-04
3.50E-04
4.00E-04
4.50E-04
5.00E-04
0 20 40 60 80 100
Con
du
ctan
ce (S
)
Time (hours)
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
0 20 40 60 80 100
Co
nd
uct
ance
(S)
Time (hours)
0.00E+00
1.00E-06
2.00E-06
3.00E-06
4.00E-06
5.00E-06
6.00E-06
7.00E-06
8.00E-06
9.00E-06
1.00E-05
0 20 40 60 80 100
Co
nd
uct
ance
(S)
Time (hours)
0.00E+00
5.00E-07
1.00E-06
1.50E-06
2.00E-06
2.50E-06
0 20 40 60 80 100
Co
nd
uct
ance
(S)
Time (hours)
A B
C D
E
Craig Alexander
187
was much more stable, although some ‘drift’ was still evident after 96 hours. The
conductance response at 1 MHz and 2 MHz was completely stable after approximately
36 hours.
Figure 5.13: Capacitance deionised water baseline settling response of a dry TDAN PVC
drop-coated IDE Design 2 (10% w/v) at five frequencies
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz, the capacitance response showed a slight increase over the initial 24 hours;
however, the measured response continued to fluctuate throughout the 96 hours over
which the experiment was conducted. At 10 kHz and 100 kHz the capacitance response
was much more stable, although some ‘drift’ was still evident after 96 hours. The
capacitance response at 1 MHz increased initially and showed near-stability after eight
hours. There was a slight decrease in the measured response; however, there was little
1.20E-10
1.22E-10
1.24E-10
1.26E-10
1.28E-10
1.30E-10
1.32E-10
1.34E-10
1.36E-10
1.38E-10
1.40E-10
0 20 40 60 80 100
Cap
acit
ance
(F)
Time (hours)
9.00E-11
1.10E-10
1.30E-10
1.50E-10
1.70E-10
0 20 40 60 80 100
Cap
acit
ance
(F)
Time (hours)
0.00E+00
5.00E-11
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
4.00E-10
0 20 40 60 80 100
Cap
acit
ance
(F)
Time (hours)
0.00E+00
1.00E-10
2.00E-10
3.00E-10
4.00E-10
5.00E-10
6.00E-10
7.00E-10
0 20 40 60 80 100
Cap
acit
ance
(F)
Time (hours)
0.00E+00
1.00E-10
2.00E-10
3.00E-10
4.00E-10
5.00E-10
6.00E-10
7.00E-10
8.00E-10
9.00E-10
0 20 40 60 80 100
Cap
acit
ance
(F)
Time (hours)
A B
C D
E
Craig Alexander
188
drift observed upon completion of the experiment. A similar response to the 1 MHz
results was also observed at 2 MHz.
The response time of TDAN PVC drop-coated IDE Design 2 was established by adding
an aliquot of 50 ppm NO3- to a deionised water background and monitoring the
response until a stable measurement was reached. These data are displayed in
Figure 5.14 (G) and Figure 5.15 (C).
Figure 5.14: Conductance response of TDAN PVC drop-coated IDE Design 2 (10% w/v) upon the
addition of 50 ppm NO3- to a deionised water background at five frequencies. The aliquot of nitrate
was added to the solution after 60 minutes
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
No observable response was seen from changes in the conductance at lower frequencies
(1 kHz, 10 kHz and 100 kHz) upon the addition of 50 ppm NO3- to a deionised water
background. At 1 MHz and 2 MHz, there was a noticeable increase of approximately
5.20E-04
5.30E-04
5.40E-04
5.50E-04
5.60E-04
5.70E-04
5.80E-04
5.90E-04
0 200 400 600 800 1000 1200 1400
Co
nd
uct
ance
(S)
Time (min)
4.40E-04
4.50E-04
4.60E-04
4.70E-04
4.80E-04
4.90E-04
5.00E-04
5.10E-04
0 200 400 600 800 1000 1200 1400
Co
nd
uct
ance
(S)
Time (min)
8.40E-06
8.60E-06
8.80E-06
9.00E-06
9.20E-06
9.40E-06
9.60E-06
9.80E-06
1.00E-05
1.02E-05
0 200 400 600 800 1000 1200 1400
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
5.00E-07
1.00E-06
1.50E-06
2.00E-06
2.50E-06
0 200 400 600 800 1000 1200 1400
Co
nd
uct
ance
(S)
Time (min)
1.01E-04
1.02E-04
1.03E-04
1.04E-04
1.05E-04
1.06E-04
1.07E-04
1.08E-04
1.09E-04
0 200 400 600 800 1000 1200 1400
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E
Craig Alexander
189
30 µS; however, the response did not settle and was still ‘drifting’ around 18 hours
following the addition of nitrate to the test solution.
Figure 5.15: Capacitance response of TDAN PVC drop-coated IDE Design 2 (10% w/v) upon the
addition of 50 ppm NO3- to a deionised water background at five frequencies. The aliquot of nitrate
was added to the solution after 60 minutes
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz, very little response was observed upon the addition of 50 ppm nitrate to a
deionised water background. At 10 kHz and 100 kHz the response was continuing to
drift around 18 hours following the addition of 50 ppm NO3-. At 1 MHz, there was a
distinct increase observed to the addition of nitrate. After two hours, the Cm – C0 was
approximately 3.2 pF, and the response remained stable for the remainder of the
experiment. A similar response was also seen at 2 MHz, after two hours an increase in
the baseline response of approximately 1.2 pF was observed and the response remained
stable for the remainder of the experiment.
1.500E-10
1.505E-10
1.510E-10
1.515E-10
1.520E-10
1.525E-10
1.530E-10
1.535E-10
1.540E-10
0 200 400 600 800 1000 1200 1400
Ca
pa
cita
nce
(F)
Time (min)
1.345E-10
1.350E-10
1.355E-10
1.360E-10
1.365E-10
0 200 400 600 800 1000 1200 1400
Ca
pa
cita
nce
(F)
Time (min)
3.60E-10
3.65E-10
3.70E-10
3.75E-10
3.80E-10
3.85E-10
3.90E-10
0 200 400 600 800 1000 1200 1400
Ca
pa
cita
nce
(F)
Time (min)
5.70E-10
5.75E-10
5.80E-10
5.85E-10
5.90E-10
5.95E-10
6.00E-10
6.05E-10
0 200 400 600 800 1000 1200 1400
Ca
pa
cita
nce
(F)
Time (min)
7.60E-10
7.70E-10
7.80E-10
7.90E-10
8.00E-10
8.10E-10
8.20E-10
8.30E-10
8.40E-10
8.50E-10
0 200 400 600 800 1000 1200 1400
Ca
pa
cita
nce
(F)
Time (min)
A B
C D
E
Craig Alexander
190
In an attempt to reduce the response time of the drop-coated membrane sensors, the
PVC ‘cocktail’ solution viscosity was reduced by decreasing the mass of the membrane
components that were dissolved in 5 ml THF. The previously tested 10% w/v ‘cocktail’
was reduced to 5% w/v by dissolving 250 mg of the membrane components into 5 ml
THF, as opposed to 500 mg. A 2% w/v ‘cocktail’ was prepared by dissolving 100 mg of
the membrane components into 5 ml THF. The membrane composition was prepared in
the same ratios of plasticiser: PVC, as described previously (Table 2.1; Chapter 2);
however, the mass of ionophore dissolved into 5 ml THF was kept constant (5 mg) to
ensure that the 1 µl of ‘cocktail’ that was drop-coated contained the same mass of the
ionophore despite the varying solution viscosity. Therefore, the 5% w/v ‘cocktail’
contained 2% w/w ionophore and the 2% w/v ‘cocktail’ contained 5% w/w ionophore.
A 1% w/v ‘cocktail’, prepared by dissolving 50 mg of the membrane components into
5 ml THF, produced a polymeric coating that was unsuitable for use as an ion-selective
membrane. The resulting solution, when cast onto a planar substrate, did not produce a
homogenous coating. It is likely that an insufficient amount of polymer was present
within the ‘cocktail’ to produce a suitable membrane layer.
Once drop-coated onto the electrodes, each membrane composition was allowed to soak
in deionised water for at least 48 hours prior to testing. The 50 ppm NO3- response time
data for IDE Design 2 containing a 5% w/v (2% w/w TDAN) drop-coated PVC
membrane are shown in Figure 5.16 (G) and Figure 5.17 (C).
Craig Alexander
191
Figure 5.16: Conductance response of TDAN PVC drop-coated IDE Design 2 (5% w/v) upon the
addition of 50 ppm NO3- to a deionised water background at five frequencies. The aliquot of nitrate
was added to the solution after 30 minutes
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz and 10 kHz, very little response was observed from the change in the
measured conductance upon the addition of 50 ppm nitrate to a deionised water
background. At 100 kHz there was a noticeable increase in the conductance response
from the deionised water baseline; however, this response did not adequately settle and
was continuing to drift around 24 hours following the addition of 50 ppm NO3-. A
similar conductance response was observed at a frequency of 1 MHz and 2 MHz.
8.00E-05
9.00E-05
1.00E-04
1.10E-04
1.20E-04
1.30E-04
1.40E-04
1.50E-04
1.60E-04
0 500 1000 1500 2000
Co
nd
uct
ance
(S)
Time (min)
8.00E-05
9.00E-05
1.00E-04
1.10E-04
1.20E-04
1.30E-04
1.40E-04
0 500 1000 1500 2000
Co
nd
uct
ance
(S)
Time (min)
3.80E-05
3.90E-05
4.00E-05
4.10E-05
4.20E-05
4.30E-05
4.40E-05
4.50E-05
4.60E-05
4.70E-05
0 500 1000 1500 2000
Co
nd
uct
ance
(S)
Time (min)
3.60E-06
3.70E-06
3.80E-06
3.90E-06
4.00E-06
4.10E-06
4.20E-06
0 500 1000 1500 2000
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
2.00E-07
4.00E-07
6.00E-07
8.00E-07
1.00E-06
1.20E-06
1.40E-06
0 500 1000 1500 2000
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E
Craig Alexander
192
Figure 5.17: Capacitance response of TDAN PVC drop-coated IDE Design 2 (5% w/v) upon the
addition of 50 ppm NO3- to a deionised water background at five frequencies. The aliquot of nitrate
was added to the solution after 30 minutes
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
A similar pattern was observed for the capacitance response time data for TDAN PVC
drop-coated IDE Design 2 (5% w/v). No observable response was seen from changes in
the conductance at 1 kHz. At 10 kHz, 100 kHz, 1 MHz and 2 MHz, there was a
noticeable increase in the capacitance response from the deionised water baseline;
however, this response did not adequately settle and was continuing to drift around
24 hours following the addition of 50 ppm NO3-.
The IDE Design 2 sensor that comprised of a 5% w/v PVC membrane ‘cocktail’ did not
show a marked improvement in the response time for either the conductance or
capacitance when compared with the previously tested 10% w/v membrane. Therefore,
the response time of a sensor drop-coated with the further reduced 2% w/v PVC
1.234E-10
1.236E-10
1.238E-10
1.240E-10
1.242E-10
1.244E-10
1.246E-10
1.248E-10
1.250E-10
0 500 1000 1500 2000
Ca
pa
cita
nce
(F)
Time (min)
1.245E-10
1.250E-10
1.255E-10
1.260E-10
1.265E-10
1.270E-10
1.275E-10
0 500 1000 1500 2000
Ca
pa
cita
nce
(F)
Time (min)
1.75E-10
1.80E-10
1.85E-10
1.90E-10
1.95E-10
2.00E-10
2.05E-10
0 500 1000 1500 2000
Ca
pa
cita
nce
(F)
Time (min)
2.75E-10
2.80E-10
2.85E-10
2.90E-10
2.95E-10
3.00E-10
3.05E-10
0 500 1000 1500 2000
Ca
pa
cita
nce
(F)
Time (min)
3.20E-10
3.30E-10
3.40E-10
3.50E-10
3.60E-10
3.70E-10
3.80E-10
3.90E-10
4.00E-10
0 500 1000 1500 2000
Ca
pa
cita
nce
(F)
Time (min)
A B
C D
E
Craig Alexander
193
membrane solution was investigated. The 50 ppm NO3- response time data for IDE
Design 2 containing a 2% w/v (5% w/w TDAN) drop-coated PVC membrane are shown
in Figure 5.18 (G) and Figure 5.19 (C).
Figure 5.18: Conductance response of TDAN PVC drop-coated IDE Design 2 (2% w/v) upon the
addition of 50 ppm NO3- to a deionised water background at five frequencies. The aliquot of nitrate
was added to the solution after 30 minutes
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz, there was an initial increase from the deionised water baseline conductance
response of approximately 1.5 µS when 50 ppm NO3- was added; however, the response
appeared unstable throughout the course of the experiment. At 10 kHz, there was an
initial increase of approximately 1.3 µS upon the addition of 50 ppm NO3- to the test
solution. This response was also erratic and continued to increase over the 24 hours
which the experiment was conducted. At 100 kHz, the conductance response continued
to drift around 24 hours following the addition of 50 ppm NO3-. At 1 MHz, the response
5.00E-05
7.00E-05
9.00E-05
1.10E-04
1.30E-04
1.50E-04
1.70E-04
0 200 400 600 800 1000 1200 1400 1600
Co
nd
uct
ance
(S)
Time (min)
5.00E-05
6.00E-05
7.00E-05
8.00E-05
9.00E-05
1.00E-04
1.10E-04
1.20E-04
1.30E-04
1.40E-04
0 200 400 600 800 1000 1200 1400 1600
Co
nd
uct
ance
(S)
Time (min)
3.30E-05
3.50E-05
3.70E-05
3.90E-05
4.10E-05
4.30E-05
4.50E-05
4.70E-05
4.90E-05
5.10E-05
0 200 400 600 800 1000 1200 1400 1600
Co
nd
uct
ance
(S)
Time (min)
1.05E-05
1.10E-05
1.15E-05
1.20E-05
1.25E-05
1.30E-05
0 200 400 600 800 1000 1200 1400 1600
Co
nd
uct
ance
(S)
Time (min)
3.00E-06
3.50E-06
4.00E-06
4.50E-06
5.00E-06
5.50E-06
6.00E-06
0 200 400 600 800 1000 1200 1400 1600
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E
Craig Alexander
194
appeared to become stable at a value of approximately 50 µS above the G0 value.
Approximately 90% of this value was reached within the first six hours. At 2 MHz, a
near-stable response was obtained at approximately 65 µS above the G0 value, which
was reached within around 6 hours.
Figure 5.19: Capacitance response of TDAN PVC drop-coated IDE Design 2 (2% w/v) upon the
addition of 50 ppm NO3- to a deionised water background at five frequencies. The aliquot of nitrate
was added to the solution after 30 minutes
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
For the capacitance response of TDAN PVC drop-coated IDE Design 2 (2% w/v), at
1 kHz, there was an initial increase from the deionised water baseline conductance
response of approximately 120 pF when 50 ppm NO3- was added; however, the
response appeared unstable throughout the course of the experiment. At 10 kHz, there
was an initial increase of approximately 11 pF upon the addition of 50 ppm NO3- to the
test solution, although this response continued to increase at the culmination of the
1.260E-10
1.265E-10
1.270E-10
1.275E-10
1.280E-10
1.285E-10
1.290E-10
0 200 400 600 800 1000 1200 1400 1600
Ca
pa
cita
nce
(F)
Time (min)
1.260E-10
1.270E-10
1.280E-10
1.290E-10
1.300E-10
1.310E-10
1.320E-10
0 200 400 600 800 1000 1200 1400 1600
Ca
pa
cita
nce
(F)
Time (min)
1.55E-10
1.60E-10
1.65E-10
1.70E-10
1.75E-10
1.80E-10
1.85E-10
1.90E-10
0 200 400 600 800 1000 1200 1400 1600
Ca
pa
cita
nce
(F)
Time (min)
2.85E-10
2.90E-10
2.95E-10
3.00E-10
3.05E-10
3.10E-10
3.15E-10
3.20E-10
3.25E-10
3.30E-10
0 200 400 600 800 1000 1200 1400 1600
Ca
pa
cita
nce
(F)
Time (min)
5.00E-10
5.50E-10
6.00E-10
6.50E-10
7.00E-10
7.50E-10
8.00E-10
0 200 400 600 800 1000 1200 1400 1600
Ca
pa
cita
nce
(F)
Time (min)
A B
C D
E
Craig Alexander
195
experiment; by 24 hours the capacitance response had reached approximately 35 pF. At
100 kHz, the capacitance response was reasonably stable at 24 hours following the
addition of 50 ppm NO3-, with a value approximately 27 pF greater than the C0 value.
At 1 MHz, the response appeared to become stable at a value of approximately 4.5 pF
above the C0 value, which was reached within 8 hours of the addition of 50 ppm NO3-.
At 2 MHz, there was an increase from C0 of approximately 2.0 pF; however, the
obtained response less stable than what had been observed at 100 kHz and 1 MHz.
The most distinct and stable responses from the addition of 50 ppm NO3- to a deionised
water background solution were seen when the least viscous PVC membrane ‘cocktail’
solution (2% w/v) was used. There was a notable increase in both the measured
conductance and capacitance of this particular membrane composition upon the addition
of 50 ppm NO3-, which also demonstrated the least amount of ‘drift’ in the response
over time. Although several hours were required for the response to be fully stable, 90%
of this final response was reached within 2 hours, and therefore it was deemed that this
particular composition was suitable for additional testing. Further investigation with
IDE Design 2 was therefore carried out using a drop-coated 2% w/v membrane
‘cocktail’ containing 5% w/w TDAN as the nitrate-selective ionophore.
The observed response time for this drop-coated membrane composition was
substantially longer than what had previously been observed when a spin-coated
membrane was used. Therefore, the methodology for conducting the calibration
experiments was adjusted to measure fewer target ion concentrations over the range
1–100 ppm NO3-, with the sensor exposed to each nitrate concentration for a longer
period of time. The subsequent experiments were conducted by adding 1, 20, 50 and
100 ppm NO3- sequentially to a deionised water background test solution and recording
Craig Alexander
196
the response from the sensor at five minute intervals for 1 hour. The resulting response
versus time data for TDAN PVC drop-coated IDE Design 2 (2% w/v) are shown in
Figure 5.20 (G) and Figure 5.21 (C). The response at 1 kHz was completely unstable for
both the measured conductance and capacitance, so this frequency was not
investigated further.
Figure 5.20: Conductance response versus time of TDAN PVC drop-coated IDE Design 2 (2% w/v)
upon addition of 1–100 ppm nitrate at four frequencies.
A – 10 kHz. B – 100 kHz. C – 1 MHz. D – 2 MHz.
At 10 kHz and 100 kHz, it was noted that the conductance response was not adequately
stable for the average measurement at each separate concentration to be recorded.
However, at 1 MHz and 2 MHz the changes in nitrate concentration were distinct and
appeared settled within the 1 hour period over which they were recorded. At 1 MHz,
upon the total addition of 100 ppm NO3-, the Gm – G0 value settled at approximately
57.7 µS. At 2 MHz, upon the total addition of 100 ppm NO3-, the Gm – G0 value settled
at approximately 70.1 µS.
5.48E-03
5.48E-03
5.49E-03
5.49E-03
5.50E-03
5.50E-03
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
1 ppm NO3-
5.59E-03
5.60E-03
5.61E-03
5.62E-03
5.63E-03
5.64E-03
5.65E-03
5.66E-03
5.67E-03
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
100 ppm NO3-
5.50E-03
5.51E-03
5.52E-03
5.53E-03
5.54E-03
5.55E-03
5.56E-03
5.57E-03
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
50 ppm NO3-
5.48E-03
5.48E-03
5.48E-03
5.48E-03
5.48E-03
5.48E-03
5.48E-03
5.48E-03
5.48E-03
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
2 hour baseline
A B
C D
20 ppm NO3-
Craig Alexander
197
Figure 5.21: Capacitance response versus time of TDAN PVC drop-coated IDE Design 2 (2% w/v)
upon addition of 1–100 ppm nitrate at four frequencies.
A – 10 kHz. B – 100 kHz. C – 1 MHz. D – 2 MHz.
At the higher frequencies of 1 MHz and 2 MHz, there was an initial decrease below the
C0 value of approximately 570 fF when 1 ppm NO3- was added to the deionised water
background solution. The response remained lower than C0 until 50 ppm NO3- had been
added. At 1 MHz, upon the total addition of 100 ppm NO3-, the Cm – C0 value settled at
approximately 2.9 pF. At 2 MHz, upon the total addition of 100 ppm NO3-, the Cm – C0
value settled at approximately 806 fF. There did not appear to be a stable capacitance
response at 10 kHz, the response of each addition of nitrate to the deionised water
background solution was indistinguishable. The most prominent capacitance response
for TDAN PVC drop-coated IDE Design 2 was obtained at a frequency of 100 kHz,
where each nitrate addition could be distinguished. Upon the total addition of 100 ppm
NO3-, the Cm – C0 value settled at approximately 27.4 pF.
The corresponding logarithmic baseline-subtracted response against log nitrate
concentration data are shown in Figure 5.22 (G) and Figure 5.23 (C).
1.034E-10
1.036E-10
1.038E-10
1.040E-10
1.042E-10
1.044E-10
1.046E-10
1.048E-10
1.050E-10
1.052E-10
0 50 100 150 200 250 300 350 400
Cap
acit
ance
(F)
Time (min)
1.025E-10
1.030E-10
1.035E-10
1.040E-10
1.045E-10
1.050E-10
1.055E-10
1.060E-10
1.065E-10
1.070E-10
0 50 100 150 200 250 300 350 400
Cap
acit
ance
(F)
Time (min)
8.00E-11
9.00E-11
1.00E-10
1.10E-10
1.20E-10
1.30E-10
1.40E-10
0 50 100 150 200 250 300 350 400
Cap
acit
ance
(F)
Time (min)
8.00E-11
9.00E-11
1.00E-10
1.10E-10
1.20E-10
1.30E-10
1.40E-10
1.50E-10
1.60E-10
1.70E-10
1.80E-10
0 50 100 150 200 250 300 350 400
Cap
acit
ance
(F)
Time (min)
A B
C D
Craig Alexander
198
Figure 5.22: Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for TDAN PVC drop-coated IDE Design 2 (2% w/v) at two frequencies.
A –1 MHz. B – 2 MHz.
The logarithmic baseline subtracted capacitance produced a linear response at each of
the two extrapolated frequencies over the concentration range 1–100 ppm NO3-. At
1 MHz the R2 value was 0.9921 and at 2 MHz it was 0.9887.
Figure 5.23: Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate
concentration for TDAN PVC drop-coated IDE Design 2 (2% w/v) at 100 kHz
Taking the logarithm of the deionised water baseline-subtracted capacitance at 100 kHz
and plotting it against log NO3- concentration confirmed the linearity of the response,
with a R2 value 0.9968.
The effect of changing the ionophore within the selective membrane layer was
investigated, to establish the affect this had upon the sensor response. A 2% w/v
membrane ‘cocktail’ containing 5% w/w of the ionophore NO3V was deposited over
y = 0.6483x - 5.6019R² = 0.9921
-6
-5
-4
-3
-2
-1
0
0 0.5 1 1.5 2 2.5
log
Gm
-G
o(S
)
log NO3- concentration (ppm)
y = 0.6858x - 5.6029R² = 0.9887
-6
-5
-4
-3
-2
-1
0
0 0.5 1 1.5 2 2.5
log
Gm
-G
o(S
)
log NO3- concentration (ppm)A B
y = 1.0663x - 12.658R² = 0.9968
-14
-12
-10
-8
-6
-4
-2
0
0 0.5 1 1.5 2 2.5
log
Cm
-C
o (F
)
log NO3- concentration (ppm)
Craig Alexander
199
the sensing area of IDE Design 2 via drop-coating, and a calibration over the nitrate
concentration 1–100 ppm was performed as before. This device will be referred to as
‘NO3V PVC drop-coated IDE Design 2’. The resulting response versus time data for
NO3V PVC drop-coated IDE Design 2 (2% w/v) are shown in Figure 5.24 (G) and
Figure 5.25 (C).
Figure 5.24: Conductance response versus time of NO3V PVC drop-coated IDE Design 2 (2% w/v)
upon addition of 1–100 ppm nitrate at four frequencies.
A – 10 kHz. B – 100 kHz. C – 1 MHz. D – 2 MHz.
At 10 kHz, there was an initial increase in the measured conductance upon the addition
of 1 ppm NO3- to the test vessel, which decreased gradually, settling approximately
12 µS above the G0 value. However, upon the addition of further aliquots of NO3
- stock
solution it was not possible to distinguish between 20, 50 and 100 ppm NO3-. At
100 kHz each addition of NO3-
was distinguishable, and appeared to settle over the
1 hour period that the measurement of each concentration was recorded. Upon the total
addition of 100 ppm NO3-, the Gm – G0 value settled at approximately 77.9 µS. At
1 MHz, upon the total addition of 100 ppm NO3-, the Gm – G0 value settled at
approximately 142.5 µS. At 2 MHz, upon the total addition of 100 ppm NO3-, the
Gm – G0 value settled at approximately 175.6 µS.
0.00E+00
5.00E-05
1.00E-04
1.50E-04
2.00E-04
2.50E-04
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.40E-04
1.60E-04
1.80E-04
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
8.00E-05
9.00E-05
1.00E-04
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
5.00E-06
1.00E-05
1.50E-05
2.00E-05
2.50E-05
3.00E-05
3.50E-05
4.00E-05
0 50 100 150 200 250 300 350 400
Co
nd
uct
ance
(S)
Time (min)
A B
C D
Craig Alexander
200
Figure 5.25: Capacitance response versus time of NO3V PVC drop-coated IDE Design 2 (2% w/v)
upon addition of 1–100 ppm nitrate at four frequencies.
A – 10 kHz. B – 100 kHz. C – 1 MHz. D – 2 MHz.
At 10 kHz, it was possible to distinguish between the measured capacitance values for
each addition of NO3- to the test vessel. Upon the total addition of 100 ppm NO3
-, the
Cm – C0 value settled at approximately 410 pF. It was also possible to distinguish each
addition of NO3- to the test vessel at 100 kHz.
Upon the total addition of 100 ppm NO3
-,
the Cm – C0 value settled at approximately 65.9 pF. At 1 MHz, there was a slight
decrease in the measured capacitance when 1 ppm NO3- was added to the deionised
water background solution, followed by an increase when the subsequent aliquots were
added. Upon the total addition of 100 ppm NO3-, the Cm – C0 value settled at
approximately 8.4 pF. A similar pattern was observed at 2 MHz; however, the response
was less stable at each concentration. Upon the total addition of 100 ppm NO3-, the
Cm – C0 value settled at approximately 5.3 pF.
The corresponding logarithmic baseline-subtracted response against log nitrate
concentration data are shown in Figure 5.26 (G) and Figure 5.27 (C).
1.29E-10
1.30E-10
1.31E-10
1.32E-10
1.33E-10
1.34E-10
1.35E-10
1.36E-10
0 50 100 150 200 250 300 350 400
Ca
pa
cita
nce
(F)
Time (min)
1.28E-10
1.29E-10
1.30E-10
1.31E-10
1.32E-10
1.33E-10
1.34E-10
1.35E-10
1.36E-10
1.37E-10
1.38E-10
1.39E-10
0 50 100 150 200 250 300 350 400
Ca
pa
cita
nce
(F)
Time (min)
9.00E-11
1.10E-10
1.30E-10
1.50E-10
1.70E-10
1.90E-10
2.10E-10
0 50 100 150 200 250 300 350 400
Ca
pa
cita
nce
(F)
Time (min)
1.00E-10
2.00E-10
3.00E-10
4.00E-10
5.00E-10
6.00E-10
7.00E-10
0 50 100 150 200 250 300 350 400
Ca
pa
cita
nce
(F)
Time (min)
A B
C D
Craig Alexander
201
Figure 5.26: Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for NO3V PVC drop-coated IDE Design 2 (2% w/v) at three frequencies.
A – 100 kHz. B – 1 MHz. C – 2 MHz.
The logarithmic baseline subtracted conductance produced a linear response at each of
the three extrapolated frequencies over the concentration range 1–100 ppm NO3-. At
100 kHz the R2 value was 0.9953, at 1 MHz it was 0.9928 and at 2 MHz it was 0.9875.
Figure 5.27: Logarithm of the change in capacitance (Cm – C0) against the logarithm of nitrate
concentration for NO3V PVC drop-coated IDE Design 2 (2% w/v) at two frequencies (negative
values obtained at 1 MHz and 2 MHz).
A – 10 kHz. B – 100 kHz.
The logarithmic baseline subtracted capacitance produced a linear response at each of
the two extrapolated frequencies over the concentration range 1–100 ppm NO3-. At
10 kHz the R2 value was 0.9940 and at 2 MHz it was 0.9988.
y = 0.3261x - 4.7573R² = 0.9953
-4.8
-4.7
-4.6
-4.5
-4.4
-4.3
-4.2
-4.1
-4
0 0.5 1 1.5 2 2.5
log
Gm
-G
o(S
)
log NO3- concentration (ppm)
y = 0.5312x - 4.8512R² = 0.9875
-6
-5
-4
-3
-2
-1
0
0 0.5 1 1.5 2 2.5lo
g G
m-
Go
(S)
log NO3- concentration (ppm)
y = 0.4874x - 4.845R² = 0.9928
-6
-5
-4
-3
-2
-1
0
0 0.5 1 1.5 2 2.5
log
Gm
-G
o(S
)
log NO3- concentration (ppm)A B
C
y = 1.0269x - 12.206R² = 0.9988
-14
-12
-10
-8
-6
-4
-2
0
0 0.5 1 1.5 2 2.5
log
Cm
-C
o (F
)
log NO3- concentration (ppm)
y = 0.3546x - 10.071R² = 0.994
-10.2
-10.1
-10
-9.9
-9.8
-9.7
-9.6
-9.5
-9.4
-9.3
0 0.5 1 1.5 2 2.5
log
Cm
-C
o (F
)
log NO3- concentration (ppm)A B
Craig Alexander
202
5.2.1.2.2. Selectivity
The FIM was used to establish the response of TDAN PVC drop-coated IDE Design 2
and NO3V PVC drop-coated IDE Design 2, through the addition of NO3- over the
concentration range 1–100 ppm in a deionised water background solution containing the
interfering anion at a fixed concentration. The deionised water baseline-subtracted
selectivity data for the ionophore TDAN are shown in Figure 5.28 (G) and Figure
5.29 (C). The selectivity data for the ionophore NO3V are shown in Figure 5.30 (G) and
Figure 5.31 (C).
Figure 5.28: Selectivity determination of TDAN PVC drop-coated IDE Design 2 (2% w/v) using the
FIM (conductive response). Comparison of: The deionised water baseline-subtracted conductance
responses to 1–100 ppm nitrate when no interfering anion is present, when a fixed concentration of
10 ppm nitrite is present and when a fixed concentration of 50 ppm chloride is present. The
response to 1–100 ppm nitrate of a ‘blank’ membrane, which did not contain TDAN as the
ionophore, is also shown
A – 1 MHz. B – 2 MHz.
For TDAN PVC drop-coated IDE Design 2 (2% w/v), the selectivity of the conductance
response was investigated at 1 MHz and 2 MHz, as these were the frequencies that
appeared to show the most analytically-relevant responses during the calibration
experiments. At 1 MHz, when 10 ppm NO2- was present, the baseline-subtracted
conductance for 1 ppm NO3- was 11.61 µS greater compared with when no interfering
anion was present. When 50 ppm Cl- was present, the baseline-subtracted conductance
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.40E-04
1.60E-04
1.80E-04
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.40E-04
1.60E-04
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
Craig Alexander
203
for 1 ppm NO3- was 34.46 µS greater compared with when no interfering anion was
present. A large response was observed for a ‘blank’ membrane that did not contain the
ionophore. At NO3- concentrations above 50 ppm, the response of the ‘blank’
membrane sensor was over four times the response from the membrane that did contain
the ionophore. A similar pattern was also observed for the response at 2 MHz.
Figure 5.29: Selectivity determination of TDAN PVC drop-coated IDE Design 2 (2% w/v) at
100 kHz using the FIM (capacitive response). Comparison of: The deionised water baseline-
subtracted capacitance responses to 1–100 ppm nitrate when no interfering anion is present, when
a fixed concentration of 10 ppm nitrite is present and when a fixed concentration of 50 ppm
chloride is present. The response to 1–100 ppm nitrate of a ‘blank’ membrane, which did not
contain TDAN as the ionophore, is also shown
The capacitance response of TDAN PVC drop-coated IDE Design 2 (2% w/v) was
investigated at 100 kHz. When 10 ppm NO2- was present, the baseline-subtracted
capacitance for 1 ppm NO3- was 6.13 pF greater compared with when no interfering
anion was present. When 50 ppm Cl- was present, the baseline-subtracted capacitance
for 1 ppm NO3- was 18.57 pF greater compared with when no interfering anion was
present. As for the conductance response, there was a substantial response observed
from a ‘blank’ membrane sensor that did not contain TDAN as the ionophore.
0
1E-11
2E-11
3E-11
4E-11
5E-11
6E-11
7E-11
8E-11
9E-11
1E-10
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
Craig Alexander
204
Figure 5.30: Selectivity determination of NO3V PVC drop-coated IDE Design 2 (2% w/v) using the
FIM (conductive response). Comparison of: The deionised water baseline-subtracted conductance
responses to 1–100 ppm nitrate when no interfering anion is present, when a fixed concentration of
10 ppm nitrite is present and when a fixed concentration of 50 ppm chloride is present. The
response to 1–100 ppm nitrate of a ‘blank’ membrane, which did not contain TDAN as the
ionophore, is also shown
A – 100 kHz. B – 1 MHz. C – 2 MHz.
For NO3V PVC drop-coated IDE Design 2, the selectivity of the conductance response
was investigated at 100 kHz, 1 MHz and 2 MHz, as these were the frequencies that
appeared to show the most analytically-relevant responses during the calibration
experiments. At 100 kHz, when 10 ppm NO2- was present, the baseline-subtracted
conductance for 1 ppm NO3- was 8.61 µS greater compared with when no interfering
anion was present. A reduction in the observed response was seen for the remaining
nitrate concentrations compared with when nitrite was not present. At 50 and 100 ppm
NO3-, the Gm – G0 was approximately 25 µS lower compared to when the interfering
anion was not present. When 50 ppm Cl- was present, the baseline-subtracted
conductance for 1 ppm NO3- was 53.80 µS greater compared with when no interfering
anion was present. However, there was very little response observed from a ‘blank’
membrane that did not contain NO3V as the ionophore or TDMACl as the cation
0.00E+00
5.00E-05
1.00E-04
1.50E-04
2.00E-04
2.50E-04
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.40E-04
1.60E-04
1.80E-04
2.00E-04
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
8.00E-05
9.00E-05
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
205
exchanger. At 20, 50 and 100 ppm NO3-, the response from the ‘blank’ membrane
sensor was <1% of that from the ‘active’ membrane sensor that did contain the
ionophore and cation exchanger. A similar response was observed at 1 MHz and
2 MHz; however, the ‘blank’ response was greater. At 50 ppm NO3-, the ‘blank’
response was approximately 31% that of the ‘active’ response at 1 MHz, and
approximately 46% at 2 MHz.
Figure 5.31: Selectivity determination of NO3V PVC drop-coated IDE Design 2 (2% w/v) using the
FIM (capacitive response). Comparison of: The deionised water baseline-subtracted capacitance
responses to 1–100 ppm nitrate when no interfering anion is present, when a fixed concentration of
10 ppm nitrite is present and when a fixed concentration of 50 ppm chloride is present. The
response to 1–100 ppm nitrate of a ‘blank’ membrane, which did not contain TDAN as the
ionophore, is also shown
A – 10 kHz. B – 100 kHz.
The capacitance response of NO3V PVC drop-coated IDE Design 2 (2% w/v) was
investigated at 10 kHz and 100 kHz. At 10 kHz, when 10 ppm NO2- was present, the
baseline-subtracted capacitance for 1 ppm NO3- was 60.08 pF greater compared with
when no interfering anion was present. When 50 ppm Cl- was present, the baseline-
subtracted capacitance for 1 ppm NO3- was 279.23 pF greater compared with when no
interfering anion was present. When a ‘blank’ membrane was tested, there was a
decrease below the C0 value of approximately 150 pF and there was very little
observable change to this value when additional aliquots of nitrate were added to the
test solution. At 100 kHz, when 10 ppm NO2- was present, the baseline-subtracted
0.00E+00
1.00E-11
2.00E-11
3.00E-11
4.00E-11
5.00E-11
6.00E-11
7.00E-11
8.00E-11
9.00E-11
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
-2.00E-10
-1.00E-10
0.00E+00
1.00E-10
2.00E-10
3.00E-10
4.00E-10
5.00E-10
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
A B
Craig Alexander
206
capacitance for 1 ppm NO3- was 8.88 pF greater compared with when no interfering
anion was present. When 50 ppm Cl- was present, the baseline-subtracted capacitance
for 1 ppm NO3- was 55.81 pF greater compared with when no interfering anion was
present. When a ‘blank’ membrane was tested, there was an initial increase of
approximately 6 pF when 1 ppm NO3- was added, this increased to approximately 10 pF
for 20 ppm NO3-; however, no further increase was observed with the subsequent nitrate
additions.
The response from a ‘blank’ NO3V membrane was much less prominent than that
observed from a ‘blank’ TDAN membrane prepared at the same solution viscosity. The
only difference between the two membranes was NPOE was used as the plasticiser with
NO3V whereas bis (2-ethylhexyl) sebacate was used with TDAN. This suggested that
the plasticiser NPOE may be better suited for use within this type of device, possibly
due to providing increased hydrophobicity to the membrane. It is necessary to further
investigate changing the membrane composition to optimise the response from the
sensors, such as investigating the use of different plasticisers.
Due to time constraints, it was not possible to further investigate IDE Design 2 as an
impedimetric nitrate sensor. Further experimentation is required to establish the effects
of changing the sample matrix composition and to establish the device-to-device
reproducibility. Several experiments were conducted to establish the response of this
device when coated with a PVC membrane containing ionophores for other target ions.
This was to establish whether this particular membrane composition was suitable with
IDE Design 2 when alternative ionophores were used for any of the other
aquarium-significant target ions. These results are discussed in Chapter 7.
Craig Alexander
207
5.3 Conclusions for Chapter 5
IDE Design 2, a commercially-available microfabricated IDE device, was purchased
and tested for its feasibility as a nitrate-selective sensor within a freshwater aquarium,
via spin-coating or drop-coating of the sensing area with a selective polymeric
membrane containing commercial nitrate ionophores. TDAN and NO3V were tested as
the nitrate-selective ionophores; PVC was used as the membrane material.
Prior to conducting any experiments, IDE Design 2 appeared to provide a more
favourable approach to producing impedimetric ion-selective sensors than the
previously discussed IDE Design 1, as it had reduced geometric parameters which in
turn would lead to a reduced penetration depth of the generated electric field.
Several experiments were conducted to excite the IDE device with a reduced input
amplitude than had been used with the previous device (1 Vrms). This was seen as
beneficial for producing a prototype device with bespoke circuitry to power the sensors.
It was concluded that an acceptable response was observed by applying an AC voltage
of 10 mVrms. Therefore, subsequent experiments were conducted using this amplitude.
When a PVC membrane containing 1% w/w of the nitrate ionophore TDAN was
deposited over the surface of the sensing area via spin-coating, a linear response was
obtained through nitrate additions to a deionised water background from the changes in
capacitance at 2 MHz. When a ‘blank’ membrane that did not contain the ionophore
was spin-coated over the IDE digits, a large response was observed which suggested
that the sensor was actually responding to non-selective changes in the conductivity of
the test solution, and not from selective interactions within the membrane. The response
of three spin-coated devices prepared in the same way demonstrated poor
Craig Alexander
208
device-to-device reproducibility when nitrate was added to each sensor. As such, further
testing was not carried out using TDAN PVC spin-coated IDE Design 2, and drop-
coating of the polymeric membrane material was investigated as an alternative
coating method.
Drop-coating was investigated as an alternative coating method in an attempt to produce
a thicker polymeric membrane layer over the sensing area and therefore impart greater
selectivity on the device. When the initial membrane ‘cocktail’ composition of 10% w/v
was drop-coated onto IDE Design 2, an unacceptably long response time to the target
nitrate anion was observed. This suggested that the resulting membrane, once cast, was
of a thickness that was too great in relation to the electrode geometry; the increased
diffusion distance for ions to move into the membrane layer produced a long sensor
response time. It is not likely that a particularly fast response time would be required for
a sensor for aquarium purposes, as the target analytes would generally accumulate over
time as opposed to reaching dangerous levels immediately. However, a response that
was prone to drift would be problematic, as it would be difficult to establish whether a
particular response was due to the presence of the target species or if the measurement
was unsettled.
In an attempt to improve the response time from the drop-coated membrane sensors, the
viscosity of the polymeric ‘cocktail’ that contained the selective ionophore was reduced
by decreasing the mass of the membrane components that were dissolved into 5 ml
THF. Reducing the membrane solution to a concentration of 2% w/v produced a sensor
with more acceptable response characteristics; therefore, this membrane composition
was investigated further.
Craig Alexander
209
When a 2% w/v PVC membrane containing 5% w/w of the ionophore TDAN was
drop-coated over the digits of IDE Design 2, a prominent response to increasing nitrate
concentration was observed from changes in the conductance at 1 MHz and 2 MHz, and
from changes in the capacitance at 100 kHz. Substantial interferences were observed
using this membrane composition from the interfering anions nitrite and chloride, and a
large response was observed when a ‘blank’ membrane was deposited over the sensing
area. NO3V was investigated as an alternative nitrate-selective ionophore within a
2% w/v drop-coated PVC membrane. This membrane composition produced an
analytically-relevant response at 100 kHz, 1 MHz and 2 MHz for the changes in the
measured conductance, and at 10 kHz and 100 kHz for changes in the measured
capacitance. A substantial reduction in the observed response from a ‘blank’ membrane
was observed; however, interference from chloride and nitrite anions was still seen. The
lack of response from a ‘blank’ NO3V membrane suggested that the hydrophobicity of
the deposited membrane is also an important factor to consider to ensure selectivity
from such devices.
For a microfabricated device such as IDE Design 2 to be used as an impedimetric nitrate
sensor, it is apparent that one of the major considerations has to be the thickness of the
polymeric membrane that coats the sensing area in relation to the electrode digit
geometry. By simply increasing the membrane thickness, this results in an unacceptably
long response time with a measurement that is prone to drift. By depositing a membrane
that is too thin over the surface of the electrodes a lack of ion-selectivity is observed due
to the penetration depth of the resulting electric field reaching into the bulk solution.
For the purpose of this work, it would have been beneficial to be able to characterise the
thickness of the deposited membrane to approximate whether it was sufficient to
Craig Alexander
210
enclose the generated electric field. During this work, an attempt was made to
characterise the membrane thickness using a ‘top-down’ scanning electron microscopy
(SEM) approach, however, the results that were obtained did not provide a definitive
answer. Suggestions for alternative routes for characterising membrane thicknesses are
provided in Chapter 8.
The best route for providing a compromise between an acceptable response time
without a loss of selectivity would be to investigate IDE sensors with even further
reduced digit geometry; however, it was not possible to obtain any devices with lower
electrode geometry than IDE Design 2 throughout this project.
IDE Design 2 was a commercial device that was purchased ‘off-the-shelf’ and was not
specifically designed for the purpose for which it was tested. This was due to time and
financial constraints, in an attempt to provide proof-of-principle data from a device with
reduced electrode geometry from the previously tested IDE Design 1. It was much less
expensive and much quicker to obtain this particular electrode device than it would have
been to design and obtain a bespoke sensor. Following on from this proof-of-principle
work, it would have been beneficial to produce a sensor with a custom design.
Unfortunately, time and financial constraints dictated that this was not possible for the
purpose of this project.
IDE Design 2 featured a reference electrode and auxiliary electrode, which, although
not connected during the experimental work, may have provided interference and
affected the obtained response. These additional electrodes would not be included in a
custom design. This particular device also consisted of fifteen digits at each electrode;
the sensitivity of the response could be improved by increasing the number of digit
pairs. Suggestions for further experimental work are discussed in Chapter 8.
Craig Alexander
211
6 Nitrate Sensing using SP IDE Design 1
6.1 Issues Experienced with Initial Sensor Designs
As mentioned previously in section 2.4.2, there were issues experienced during the
fabrication stages of the first screen-printed IDEs which resulted in no suitable devices
available for testing as ion-selective sensors. Figure 6.1A shows a microscope image of
one of the designs with 60 µm digit widths and inter-digit spacing, and Figure 6.1B
shows a design with 100 µm widths and inter-digit spacing.
Figure 6.1: Microscope image (4x) of the carbon screen-printed IDE sensors produced at the
University of Bedfordshire. (A) shows one of the IDEs printed with 60 µm feature sizes. (B) shows
one of the IDEs printed with 100 µm feature sizes
A
B
Craig Alexander
212
It is evident from Figure 6.1 that the electrodes comprising of either 60 µm or 100 µm
line width geometries were not printed successfully; therefore, no data were obtained
from any of these sensors. The feature sizes of both sensors were too small and the ink
had expanded upon drying which resulted in many short-circuits between adjacent
digits, rendering the device unsuitable for use as an impedimetric ion-selective sensor.
Testing was carried out using the larger SP IDE Design 1, which had been designed
with a narrower digit width (100 µm) and a larger space between adjacent digits
(150 µm), and this consequently prevented short-circuits occurring during expansion of
the ink upon drying. A microscope image of the digits of SP IDE Design 1 is shown in
Figure 2.7 (Chapter 2).
6.2 SP IDE Design 1 Calibration
As for the photolithographically prepared IDE Design 1 and IDE Design 2 devices, it
was important to ascertain the response to the target ion of a SP IDE Design 1 sensor
which did not contain a selective membrane layer over the IDE digits. An uncoated SP
IDE Design 1 sensor, which had not had the modified dielectric paste membrane layer
printed over the sensing area, was placed into a deionised water background until a
stable baseline was observed, followed by the addition of NO3- over the concentration
range 1–100 ppm. The resulting conductance (A) and capacitance (B) spectra are shown
in Figure 6.2.
Craig Alexander
213
Figure 6.2: Conductance (A) and capacitance (B) spectra for the addition of NO3- to an uncoated SP
IDE Design 1 (1 Vrms input amplitude). ‘Pure water’ refers to the initial baseline value and
10-120 mins are the values obtained during the baseline settling period
As for IDE Design 2, the response of an uncoated device was tested at a series of
reduced input amplitudes. The deionised water baseline-subtracted results upon the
addition of NO3- to an uncoated SP IDE Design 1 device at amplitudes of 1 Vrms,
100 mVrms, 10 mVrms and 1 mVrms are shown in Figure 6.3 (G) and Figure 6.4 (C).
0.00E+00
2.00E-08
4.00E-08
6.00E-08
8.00E-08
1.00E-07
1.20E-07
1.40E-07
1.60E-07
1.80E-07
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 10 mins
Cp 30 mins
Cp 60 mins
Cp 90 mins
Cp 100 mins
Cp 110 mins
Cp 120 mins
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.40E-04
1.60E-04
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Con
du
ctan
ce (S
)
Frequency (Hz)
G Pure water
G 10 mins
G 30 mins
G 60 mins
G 90 mins
G 100 mins
G 110 mins
G 120 mins
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
214
Figure 6.3: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for uncoated SP IDE Design 1 at six frequencies and varying input amplitudes.
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
An increase in the measured conductance was observed for 100 Hz, 1 kHz, 10 kHz and
100 kHz as the solution conductivity was increased, at each of the tested input
amplitudes. There was a decrease below the G0 value for the measured conductance at
1 MHz and 2 MHz.
-0.00003
-0.000025
-0.00002
-0.000015
-0.00001
-0.000005
0
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)log NO3
- concentration (ppm)
-0.00003
-0.000025
-0.00002
-0.000015
-0.00001
-0.000005
0
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.00002
0.00004
0.00006
0.00008
0.0001
0.00012
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.00002
0.00004
0.00006
0.00008
0.0001
0.00012
0.00014
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.00002
0.00004
0.00006
0.00008
0.0001
0.00012
0.00014
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.00002
0.00004
0.00006
0.00008
0.0001
0.00012
0.00014
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C D
E F
Craig Alexander
215
Figure 6.4: Deionised water baseline-subtracted capacitance (Cm – C0) against log NO3
-
concentration for uncoated SP IDE Design 1 at six frequencies and varying input amplitudes
A – 100 Hz. B – 1 kHz. C – 10 kHz. D – 100 kHz. E – 1 MHz. F – 2 MHz
There was no observable analytically-relevant response from the capacitance of
uncoated SP IDE Design 1. A decrease in the measured capacitance was observed as the
solution conductivity was increased through the addition of potassium nitrate to the test
vessel at 1 MHz and 2 MHz.
As a response in the change of conductance was observed at lower input amplitudes,
which was seen as desirable so that a prototype device with a lower power consumption
could be produced, it was decided that subsequent experiments using SP IDE Design 1
would be excited with an input AC amplitude of 10 mVrms, as this amplitude produced a
response that was analytically-valid. At an input amplitude of 1 mVrms an unstable
measurement was obtained.
-1E-10
-9E-11
-8E-11
-7E-11
-6E-11
-5E-11
-4E-11
-3E-11
-2E-11
-1E-11
0
0 0.5 1 1.5 2 2.5
Cm
-C0
(F)
log NO3- concentration (ppm)
-1E-12
-8E-13
-6E-13
-4E-13
-2E-13
0
2E-13
4E-13
0 0.5 1 1.5 2 2.5
Cm
-C0
(F)
log NO3- concentration (ppm)
-7E-12
-6E-12
-5E-12
-4E-12
-3E-12
-2E-12
-1E-12
0
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
-1.2E-10
-1E-10
-8E-11
-6E-11
-4E-11
-2E-11
0
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
-3E-10
-2E-10
-1E-10
0
1E-10
2E-10
3E-10
4E-10
5E-10
6E-10
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
-2.5E-08
-2E-08
-1.5E-08
-1E-08
-5E-09
0
5E-09
1E-08
1.5E-08
2E-08
0 0.5 1 1.5 2 2.5
Cm
-C
0(F
)
log NO3- concentration (ppm)
A B
C D
E F
Craig Alexander
216
Figures 6.5A and 6.5B show the conductance and capacitance spectra, respectively, for
the addition of nitrate in the concentration range 1–100 ppm, to SP IDE Design 1 which
had been screen-printed with a modified dielectric paste ion-selective membrane layer
containing 22% w/w NPOE as the plasticiser and 1% w/w TDAN as the ionophore.
Figure 6.5: Conductance (A) and capacitance (B) spectra for the addition of NO3
- to SP IDE Design
1 with a layer of plasticised dielectric paste containing 1% w/w TDAN as the ionophore (10 mVrms
input amplitude). ‘Pure water’ refers to the initial baseline value and 10-60 mins are the values
obtained during the baseline settling period. Interference at 50 Hz was observed; however this
frequency was not investigated
When this membrane layer was present, the most prominent analytical response from
the sensor was from the conductance at 1 kHz, 10 kHz and 100 kHz. As discussed
0.00E+00
2.00E-09
4.00E-09
6.00E-09
8.00E-09
1.00E-08
1.20E-08
1.40E-08
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Cap
acit
ance
(F)
Frequency (Hz)
Cp Pure water
Cp 10 mins
Cp 20 mins
Cp 30 mins
Cp 40 mins
Cp 50 mins
Cp 60 mins
Cp 1 ppm NO3-
Cp 10 ppm NO3-
Cp 20 ppm NO3-
Cp 30 ppm NO3-
Cp 40 ppm NO3-
Cp 50 ppm NO3-
Cp 100 ppm NO3-
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
8.00E-05
9.00E-05
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Co
nd
uct
ance
(S)
Frequency (Hz)
G Pure water
G 10 mins
G 20 mins
G 30 mins
G 40 mins
G 50 mins
G 60 mins
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
A
B
Craig Alexander
217
throughout Chapters 4 and 5, it was important to understand the behaviour of a sensor
comprising of a membrane layer coating the sensing area that did not contain any
selective ionophore. The deionised water baseline-subtracted conductance at 1 kHz,
10 kHz and 100 kHz of an IDE Design 1 device, consisting of a modified dielectric
paste membrane layer that contained TDAN as the ionophore, is compared with a
device consisting of a ‘blank’ dielectric paste membrane, without ionophore present, in
Figure 6.6.
Figure 6.6: Comparison of the conductance response of SP IDE Design 1 when a membrane
containing 1% w/w TDAN was printed over the sensing area with a ‘blank’ membrane which did
not contain any ionophore
A – 1 kHz. B – 10 kHz. C – 100 kHz
Above 1 ppm NO3-, the observed response for the ‘blank’ membrane SP IDE Design 1
device upon addition of the target anion was between 67–89% of that from the
ionophore-containing membrane at 1 kHz; this was around 60% at 10 kHz and between
65–90% at 100 kHz. This suggested that non-selective matrix interferences were
responsible for a large proportion of the observed response upon addition of NO3- to the
test solution.
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0.000035
0.00004
0.000045
0.00005
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0.000035
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000002
0.000004
0.000006
0.000008
0.00001
0.000012
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
218
In an attempt to reduce the response that was observed from a device which was coated
with a ‘blank’ modified dielectric paste ion-selective membrane, and therefore improve
the selectivity of the device by reducing matrix interferences, the thickness of the
membrane layer over the sensing area was increased by first printing and curing a
‘build-up’ layer of non-modified dielectric paste. A comparison of the response of three
‘blank’ membranes; one with no ‘build-up’ layer, one with a single ‘build-up’ layer and
one with two ‘build up’ layers, is shown in Figure 6.7.
Figure 6.7: Comparison of the conductance response of SP IDE Design 1 coated with a ‘blank’
membrane containing no ionophore when zero, one or two ‘build-up’ layers of non-modified
dielectric paste were printed prior to the modified layer containing NPOE as the plasticiser.
A – 1 kHz. B – 10 kHz. C – 100 kHz
It was observed at all three frequencies of interest that the conductive response was
substantially reduced when two ‘build-up’ layers of unmodified dielectric paste were
printed and cured over the digits of SP IDE Design 2 prior to the printing of the
modified dielectric paste containing plasticiser. Subsequent experiments were therefore
conducted with devices that had two ‘build-up’ layers printed with a third layer
comprising of modified dielectric paste containing 1% w/w of TDAN as the
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0.000035
0.00004
0.000045
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
-0.000001
0
0.000001
0.000002
0.000003
0.000004
0.000005
0.000006
0.000007
0.000008
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
219
nitrate-selective ionophore. This particular device will be referred to as ‘TDAN SP IDE
Design 1’. The conductance spectrum for this device is shown in Figure 6.8.
Figure 6.8: Conductance spectrum for the addition of NO3
- to TDAN SP IDE Design 1 (10 mVrms
input amplitude). ‘Pure water’ refers to the initial baseline value. Interference was observed at
50 Hz so this data has been omitted
The conductance response versus time data at each extrapolated frequency (1 kHz,
10 kHz and 100 kHz), upon the addition of nitrate over the concentration range
1–100 ppm, is shown in Figure 6.9.
Figure 6.9: Conductance response versus time of TDAN SP IDE Design 1 upon addition of
1–100 ppm nitrate at four frequencies. A – 1 kHz. B – 10 kHz. C – 100 kHz.
-2.00E-05
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07
Co
nd
uct
ance
(S)
Frequency (Hz)
G Pure water
G 1 ppm NO3-
G 10 ppm NO3-
G 20 ppm NO3-
G 30 ppm NO3-
G 40 ppm NO3-
G 50 ppm NO3-
G 100 ppm NO3-
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
G @ 100 kHz
0.00E+00
2.00E-06
4.00E-06
6.00E-06
8.00E-06
1.00E-05
1.20E-05
1.40E-05
1.60E-05
1.80E-05
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
G @ 10 kHz
0.00E+00
1.00E-06
2.00E-06
3.00E-06
4.00E-06
5.00E-06
6.00E-06
7.00E-06
0 20 40 60 80 100 120 140
Co
nd
uct
ance
(S)
Time (min)
G @ 1 kHz
A B
C
Craig Alexander
220
The corresponding logarithmic baseline-subtracted conductance versus log nitrate
concentration is shown in Figure 6.10.
Figure 6.10: Logarithm of the change in conductance (Gm – G0) against the logarithm of nitrate
concentration for TDAN SP IDE Design 1
A – 1 kHz. B – 10 kHz. C – 100 kHz.
Taking the logarithm of the deionised water baseline-subtracted conductance and
plotting it against log NO3- concentration confirmed the linearity of the response (1 kHz
R2 = 0.9988, 10 kHz R
2 = 0.9929 and 100 kHz R
2 = 0.9862).
Figure 6.11 shows the comparison between adjusting the nitrate concentration within
the test vessel using a KNO3 stock solution and a LiNO3 stock solution. The effect of
temperature on the sensor response was not determined, as it was deemed more
important to ascertain the ion-selectivity of the device before looking at peripheral
characteristics of the response.
y = 0.3635x - 5.1709R² = 0.9862
-5.3
-5.2
-5.1
-5
-4.9
-4.8
-4.7
-4.6
-4.5
-4.4
-4.3
0 0.5 1 1.5 2 2.5
log
(Gm
-G
0)
(S)
log NO3- concentration (ppm)
y = 0.3881x - 5.5975R² = 0.9929
-5.7
-5.6
-5.5
-5.4
-5.3
-5.2
-5.1
-5
-4.9
-4.8
-4.7
0 0.5 1 1.5 2 2.5
log
(Gm
-G
0)
(S)
log NO3- concentration (ppm)
y = 0.2591x - 5.8665R² = 0.9988
-5.9
-5.8
-5.7
-5.6
-5.5
-5.4
-5.3
0 0.5 1 1.5 2 2.5
log
(Gm
-G
0)
(S)
log NO3- concentration (ppm)A B
C
Craig Alexander
221
Figure 6.11: Deionised water baseline-subtracted conductance (Gm – G0) against log NO3
-
concentration for TDAN SP IDE Design 1. Two calibration experiments are shown: the addition of
1–100 ppm NO3- with K
+ as the counter ion and the addition of 1–100 ppm NO3
- with Li
+ as the
counter ion
A – 1 kHz. B – 10 kHz. C – 100 kHz.
When lithium nitrate was used to adjust the nitrate concentration in the test vessel, the
marked decrease compared with the potassium nitrate calibration that had been present
for previous devices was not observed. This suggested that the counter ion that was used
with the target ion was producing less of an effect on the observed sensor response. At
1 kHz, the lithium nitrate calibration was slightly higher than the potassium nitrate
calibration, which would not be expected if solution conductivity was heavily
influencing the sensor response. At 50 ppm NO3-, the %RSD between the two
calibration experiments was 7.87%. There was less of a marked difference between the
calibration experiments at 10 kHz and 100 kHz; %RSD between the two measurements
at 50 ppm NO3- was 2.22% at 10 kHz and 4.72% at 100 kHz.
The effect of the changes in background conductivity of the sample solution was
investigated by placing the sensor into solutions of varying ionic strength. Figure 6.12
0
0.000002
0.000004
0.000006
0.000008
0.00001
0.000012
0.000014
0.000016
0.000018
0 0.5 1 1.5 2 2.5
Gm
-G
0 (S
)
log NO3- concentration (ppm)
0
0.000001
0.000002
0.000003
0.000004
0.000005
0.000006
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0.000035
0.00004
0.000045
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
C
Craig Alexander
222
shows the measured G0 values for each of the tested environments at each of the three
frequencies of interest.
Figure 6.12: Measured baseline conductance value (G0) of TDAN SP IDE Design 1 in four separate
sample matrices prior to the addition of nitrate
A – 1 kHz. B – 10 kHz. C – 100 kHz.
An increase of G0 was seen at all three extrapolated frequencies as the ionic strength of
the background solution was increased. A 100 ppm solution of MgSO4 produced a G0
value that was 2.43 µS greater than the deionised water value at 1 kHz. A 200 ppm
solution of MgSO4 was 2.67 µS greater and 500 ppm MgSO4 was 3.09 µS greater than
the deionised water value. At 10 kHz the G0 values were 11.41 µS greater than
deionised water for 100 ppm MgSO4, 13.15 µS greater for 200 ppm MgSO4 and
14.87 µS greater for 500 ppm MgSO4. At 100 kHz the G0 values were 36.69 µS greater
than deionised water for 100 ppm MgSO4, 45.43 µS greater for 200 ppm MgSO4 and
57.71 µS greater for 500 ppm MgSO4.
0
0.00001
0.00002
0.00003
0.00004
0.00005
0.00006
0.00007
0.00008
0.00009
G0
bas
eli
ne
at
10
0 k
Hz
Background sample matrix
0
0.000005
0.00001
0.000015
0.00002
0.000025
G0
bas
eli
ne
at
10
kH
z
Background sample matrix0
0.0000005
0.000001
0.0000015
0.000002
0.0000025
0.000003
0.0000035
0.000004
0.0000045
G0
bas
eli
ne
val
ue
at
1 k
Hz
Background sample matrix
A B
C
Craig Alexander
223
The baseline-subtracted conductance responses at each frequency upon addition of NO3-
over the range 1–100 ppm to each of these background solutions are shown in
Figure 6.13.
Figure 6.13: Baseline-subtracted conductance (Gm – G0) of TDAN SP IDE Design 1 against log NO3
-
concentration in four separate sample matrices
A – 1 kHz. B – 10 kHz. C – 100 kHz.
At each of the extrapolated frequencies, the sensitivity of the sensor to each addition of
NO3- was reduced as the ionic strength of the background solution was increased.
Figure 6.14 shows the deionised water baseline-subtracted conductance at 1 kHz,
10 kHz and 100 kHz, for the addition of NO3- to three separate TDAN SP IDE
Design 1 devices.
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0.000035
0.00004
0.000045
0.00005
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000002
0.000004
0.000006
0.000008
0.00001
0.000012
0.000014
0.000016
0 0.5 1 1.5 2 2.5
Gm
-G
0 (S
)
log NO3- concentration (ppm)
0
0.0000005
0.000001
0.0000015
0.000002
0.0000025
0.000003
0.0000035
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
224
Figure 6.14: Deionised water baseline-subtracted conductance (Gm – G0) of three TDAN SP IDE
Design 1 devices against log NO3- concentration, for reproducibility determination
A – 1 kHz. B – 10 kHz. C – 100 kHz.
The device-to-device reproducibility, defined as the %RSD of the three obtained
baseline-subtracted conductance measurements at 50 ppm NO3-, was calculated as
27.20% at 1 kHz, 8.26% at 10 kHz and 10.93% at 100 kHz.
6.3 SP IDE Design 1 Selectivity
The FIM was used to establish the response of the TDAN SP IDE Design 1 sensor
through the addition of NO3- over the concentration range 1–100 ppm in a background
solution containing fixed interfering anions. The FPM was also used to deduce the
effect of adding an interfering ion to a deionised water background containing nitrate at
a fixed concentration (50 ppm). The selectivity data of TDAN SP IDE Design 1 are
shown in Figure 6.15.
0
0.000005
0.00001
0.000015
0.00002
0.000025
0.00003
0.000035
0.00004
0.000045
0.00005
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000002
0.000004
0.000006
0.000008
0.00001
0.000012
0.000014
0.000016
0.000018
0.00002
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000001
0.000002
0.000003
0.000004
0.000005
0.000006
0.000007
0.000008
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
225
Figure 6.15: Selectivity determination of TDAN SP IDE Design 1 using the FIM. Comparison of:
The deionised water baseline-subtracted conductance responses to 1–100 ppm nitrate when no
interfering anion is present, when a fixed concentration of 10 ppm nitrite is present and when a
fixed concentration of 50 ppm chloride is present. The response to 1–100 ppm nitrate of a ‘blank’
membrane, which did not contain TDAN as the ionophore, is also shown
A – 1 kHz. B – 10 kHz. C – 100 kHz.
When the FIM was used to establish the selectivity of TDAN SP IDE Design 1, at each
extrapolated frequency, an increase in the conductance response was observed when the
nitrate calibration was carried out as before, however 50 ppm Cl- or 10 ppm NO2
- were
present in the test sample.
At 1 kHz, there was an increase of 5.18 µS from the deionised water G0 value when 50
ppm Cl- was added to the test sample. When the FPM was used, and 50 ppm NO3
- was
added to the test sample, this resulted in an increase in the measured conductance of
4.35 µS. The addition of 10 ppm NO2- to the test vessel resulted in an increase in the
measured conductance of 2.72 µS, compared with 2.47 µS when 10 ppm NO3- was
added to a deionised water background solution.
0
0.00001
0.00002
0.00003
0.00004
0.00005
0.00006
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
-0.000001
0
0.000001
0.000002
0.000003
0.000004
0.000005
0.000006
0.000007
0.000008
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
0
0.000005
0.00001
0.000015
0.00002
0.000025
0 0.5 1 1.5 2 2.5
Gm
-G
0(S
)
log NO3- concentration (ppm)
A B
C
Craig Alexander
226
At 10 kHz, adding 50 ppm Cl- to the deionised water baseline resulted in a Gm - G0
value of 15.54 µS compared with 13.07 µS which was the Gm – G0 obtained from
50 ppm NO3-. The Gm – G0 value for 10 ppm NO2
- was 6.62 µS compared to 6.14 µS
which was obtained from 50 ppm NO3-.
At 100 kHz, adding 50 ppm Cl- to the deionised water baseline resulted in a Gm - G0
value of 37.37 µS compared with 29.82 µS which was the Gm – G0 obtained from
50 ppm NO3-. The Gm – G0 value for 10 ppm NO2
- was 18.03 µS compared with
14.10 µS which was obtained from 50 ppm NO3-.
As such a large response was obtained from the interfering anions at each extrapolated
frequency. This suggested that the TDAN SP IDE Design 1 was unable to adequately
distinguish between the different anions, and therefore did not provide adequate nitrate
selectivity.
Very little response was observed from the ‘blank’ membrane, which consisted of two
‘build-up’ layers of standard dielectric paste followed by a layer of the modified paste
that contained NPOE as the plasticiser but did not contain TDAN as the ionophore, at
any of the extrapolated frequencies. At 50 ppm NO3-, the ‘blank’ response at 1 kHz was
0.43% that of the conductance measurement when the ionophore was present; at 10 kHz
it was 8.26% and at 100 kHz it was 33.57%.
There was little response obtained when a ‘blank’ membrane, which did not contain any
ionophore, was deposited over the sensing area of the SP IDEs. This suggested that the
origin of the sensor response was from interactions within the membrane layer
containing an immobilised ionophore. As such a large response was obtained from
non-target species and an increase to the baseline G0 value was observed when solutions
with increasing ionic strength were used, this suggested that this particular ionophore,
Craig Alexander
227
TDAN, was not selective to nitrate within this membrane configuration. Unfortunately
it was not possible to obtain any further screen-printed devices with alternative
membrane parameters. Further work is required to produce sensors based on SP IDE
Design 1 that contain alternative plasticisers and/or ionophores within the membrane
layer. Suggestions for further experimental work are discussed in Chapter 8.
6.4 Conclusions for Chapter 6
An ion-selective impedimetric nitrate sensor based on carbon IDEs was fabricated using
a fully screen-printed process, which included a screen-printed ion-selective membrane
layer coating the sensing area, and tested for its feasibility as a nitrate-selective sensor
for use within a freshwater aquarium. A suitable membrane support matrix for
screen-printing was produced by incorporating a nitrate-selective ionophore (TDAN),
along with the plasticiser NPOE, into a commercial dielectric screen-printing ink. Initial
fabrication issues dictated that a larger geometry was required to produce IDEs via
screen-printing, and eventually it was found that a design consisting of electrode digits
with a width of 100 µm, separated by an interdigital space between adjacent digits of
150 µm, printed effectively and reproducibly.
When a single layer of the modified dielectric screen-printing ink was coated over the
IDEs as the ion-selective membrane layer, a large response to nitrate was observed with
a ‘blank’ membrane that did not contain the ionophore. This suggested that the
penetration depth of the resulting electric field was much greater than the thickness of
the ion-selective membrane. Therefore, non-selective changes in the conductivity of the
bulk solution were observed, rather than changes in the electrical properties of the
membrane due to selective interactions between the target anion and the ionophore
immobilised within it. In an attempt to overcome this issue, a ‘build-up’ layer of
Craig Alexander
228
standard dielectric paste was first printed and cured over the sensing area, followed by
the modified ink, to afford a thicker membrane layer. At certain key mid-range
frequencies (1 kHz–100 kHz) where a linear relationship between the impedimetric
response and nitrate concentration was observed, there was a reasonable reduction in the
magnitude of response from the ‘blank’ membrane. This suggested that the thicker
membrane was containing more of the electric field and therefore changes in sample
matrix conductivity were less significant to the response. By adding a second ‘build-up’
layer prior to the membrane, the ‘blank’ response was even further reduced. At a
frequency of 1 kHz the observed ‘blank’ response was less than 10% of the response
observed at each nitrate concentration compared with the sensor that contained TDAN
as the ionophore.
This layer-by-layer screen-printing approach to depositing the ion-selective membrane
layer of the devices appeared to provide a route for producing disposable, fully
screen-printed, all-solid-state nitrate sensors that do not require a reference electrode.
However, when the selectivity of the three-layer devices was tested, a strong
interference from both nitrite and chloride anions was observed. This suggested that the
TDAN ionophore within this particular membrane composition did not show a strong
selectivity for nitrate when implemented within this type of device, and therefore would
not be suitable for use within an aquarium nitrate sensor. Adding ‘build-up’ layers to the
device did not appear to significantly increase the response time of the electrodes, as the
selective components within the membrane were only held within the thinnest top layer,
so the overall diffusion distance for ions to move into the membrane layer was
not increased.
Craig Alexander
229
Due to the manufacturing costs involved in producing the screen-printed sensors at a
commercial rate, it was not possible to obtain any more IDE Design 1 sensors for
further testing. Additional investigation is required to produce sensors with alternative
ionophores and plasticisers contained within the membrane layer in order to produce a
greater selective response to the target ions. Screen-printing has been shown to be a
feasible route for producing ISCOM-type devices for impedimetric ion-selective
sensing without the requirement of an external reference electrode. It is also suitable for
producing disposable sensors on inexpensive substrates which is particularly attractive
for the purpose of a commercial aquarium sensor. As the sensors are fully
screen-printed, they are ready to use as received without any further modification. The
main drawback of utilising this technique for producing such devices is that a
reasonably large quantity of ionophore is required to produce a sufficient amount of
modified paste for screen-printing (100 mg ionophore per 10 g of membrane material),
some of which would be wasted as it is left behind on the screens following the printing
process. This is particularly unattractive for mass producing such devices as some of the
commercially-available ionophores for aquarium-significant ions are vastly expensive,
for example 100 mg of NO2I would cost in excess of £2300.
For commercial purposes, a comparison between fabrication processes is needed to
establish which method is the most cost-effective for mass-production. The options are
to produce a less expensive carbon IDE device which requires a larger amount of
ionophore using screen-printing, or to purchase commercial microfabricated IDEs, such
as IDE Design 2, which then require coating with a small amount of polymer
ion-selective membrane, and therefore require less ionophore. The final cost-per-device
will depend on how much ionophore is required and the initial cost of the individual
IDE transduction element.
Craig Alexander
230
7 pH, Ammonium and Nitrite Sensing using IDE Design 2
Due to time constraints, it was not possible to carry out all of the necessary experiments
to obtain full data-sets from IDE sensors for pH, ammonium or nitrite sensing. Several
experiments for these target ions were carried out using IDE Design 2 with a
drop-coated 2% w/v PVC membrane ‘cocktail’ containing 5% w/w of the
complimentary ionophore. The primary aims of these investigations were to ascertain
the response of a ‘blank’ membrane that did not contain the ionophore in comparison
with an ‘active’ membrane that did, and to establish the selectivity of the sensors against
interfering ions that are most likely to be present within an aquarium. Much further
research is required to fully validate the individual sensor performance for use within an
aquarium monitoring device. Suggestions for further experimental work are discussed
in Chapter 8.
7.1 Ammonium Sensors
An IDE Design 2 device was drop-coated with a 2% w/v PVC membrane ‘cocktail’
containing 5% w/w of the ionophore NH4I. An aliquot of 50 ppm NH4+ was added to a
deionised water background solution and the response was obtained at 10 minute
intervals over two hours. As with the nitrate experiments conducted using IDE Design 2
(Chapter 5), an input amplitude of 10 mVrms was used. This device will be referred to as
‘NH4I PVC drop-coated IDE Design 2’. The conductance versus time data at five
frequencies of interest are shown in Figure 7.1 and the capacitance versus time data are
shown in Figure 7.2.
Craig Alexander
231
Figure 7.1: Conductance response versus time of NH4I PVC drop-coated IDE Design 2 (2% w/v)
upon the addition of 50 ppm NH4+ to a deionised water background at five frequencies. The aliquot
of ammonium stock solution was added after 120 minutes.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
Each of the frequencies of interest showed a distinct increase from the G0 value upon
the addition of 50 ppm NH4+ to a deionised water background. At 1 kHz, the response
appeared to settle after around 90 minutes with a Gm – G0 value of approximately
3.0 µS. The response at 10 kHz also appeared to settle after around 90 minutes, with a
Gm – G0 value of approximately 3.8 µS. At 100 kHz, the response appeared to settle
after around 60 minutes with a Gm – G0 value of approximately 5.9 µS. The
conductance response at 1 MHz showed an immediate increase of approximately 38 µS
before gradually decreasing and settling after around 40 minutes with a Gm – G0 value
of approximately 26 µS. The conductance response at 2 MHz showed an immediate
0.00E+00
2.00E-05
4.00E-05
6.00E-05
8.00E-05
1.00E-04
1.20E-04
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
5.00E-07
1.00E-06
1.50E-06
2.00E-06
2.50E-06
3.00E-06
3.50E-06
4.00E-06
4.50E-06
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
2 hour deionised water baseline
50 ppm NH4+
0.00E+00
1.00E-06
2.00E-06
3.00E-06
4.00E-06
5.00E-06
6.00E-06
7.00E-06
8.00E-06
9.00E-06
1.00E-05
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
1.00E-06
2.00E-06
3.00E-06
4.00E-06
5.00E-06
6.00E-06
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E
Craig Alexander
232
increase of approximately 50 µS before settling after around 30 minutes with a Gm – G0
value of approximately 70 µS.
Figure 7.2: Capacitance response versus time of NH4I PVC drop-coated IDE Design 2 (2% w/v)
upon the addition of 50 ppm NH4+ to a deionised water background at five frequencies. The aliquot
of ammonium stock solution was added after 120 minutes.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz, the capacitance response increased from the C0 value by approximately
155 pF, however the response appeared unstable and prone to drift. The response at
10 kHz increased by approximately 24 pF; however, this also appeared unstable. At
100 kHz, the response appeared to settle after around 40 minutes with a Cm – C0 value
of approximately 16 pF. The capacitance response at 1 MHz showed an increase of
approximately 12 pF, settling after around 40 minutes. The capacitance response at
2 MHz showed an increase of approximately 11 pF, settling after around 40 minutes.
1.28E-10
1.30E-10
1.32E-10
1.34E-10
1.36E-10
1.38E-10
1.40E-10
1.42E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
1.28E-10
1.30E-10
1.32E-10
1.34E-10
1.36E-10
1.38E-10
1.40E-10
1.42E-10
1.44E-10
1.46E-10
1.48E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
1.35E-10
1.40E-10
1.45E-10
1.50E-10
1.55E-10
1.60E-10
1.65E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
0.00E+00
5.00E-11
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
4.00E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
A B
C D
E
1.28E-10
1.30E-10
1.32E-10
1.34E-10
1.36E-10
1.38E-10
1.40E-10
1.42E-10
1.44E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
Craig Alexander
233
Selectivity data for NH4I PVC drop-coated IDE Design 2 were obtained using the FIM,
through the addition of 50 ppm of the interfering cation to a deionised water
background. An aliquot of ammonium at a concentration of 50 ppm NH4+ was also
added to a device that had been prepared without the ionophore present. The
baseline-subtracted data at each extrapolated frequency are displayed in Table 7.1 (G)
and Table 7.2 (C).
Gm – G0 (µS)
Ion (50 ppm) Frequency
1 kHz 10 kHz 100 kHz 1 MHz 2 MHz
Ammonium 2.95 3.79 5.91 26.07 70.50
Potassium 1.05 1.29 1.83 41.33 98.86
Sodium 0.39 0.43 0.73 45.71 99.05
Magnesium 0.11 0.03 1.50 46.93 91.22
Ammonium
(blank membrane) 2.36 2.82 3.86 21.72 46.96
Table 7.1: Selectivity determination of NH4I PVC drop-coated IDE Design 2 (2% w/v) (conductive
response). Deionised water baseline-subtracted conductance upon the addition of the 50 ppm of:
The target ammonium ion and interfering cations potassium, sodium and magnesium. The response
from the addition of 50 ppm NH4+ to a device prepared without ionophore present is also shown
At the frequencies 1 kHz, 10 kHz and 100 kHz, the conductance response
characteristics of the sensor appeared to be favourable to ammonium when compared
with the response from each of the interfering cations. However, at the higher
frequencies of 1 MHz and 2 MHz there was an increased response to each of the three
tested interfering cations when compared with the response obtained from the target
ammonium ion. At each frequency of interest, a large response was observed from a
‘blank’ membrane that did not contain the ionophore.
Craig Alexander
234
Cm – C0 (pF)
Ion (50 ppm) Frequency
1 kHz 10 kHz 100 kHz 1 MHz 2 MHz
Ammonium 155.44 23.94 16.10 12.20 10.63
Potassium 51.17 15.72 14.88 11.72 7.28
Sodium 29.10 14.45 15.19 11.04 6.03
Magnesium 23.23 13.85 16.45 6.30 2.41
Ammonium
(blank membrane) 63.74 26.16 12.96 5.15 3.10
Table 7.2: Selectivity determination of NH4I PVC drop-coated IDE Design 2 (2% w/v) (capacitive
response). Deionised water baseline-subtracted capacitance upon the addition of the 50 ppm of: The
target ammonium ion and interfering cations potassium, sodium and magnesium. The response
from the addition of 50 ppm NH4+ to a device prepared without ionophore present is also shown
At 1 kHz, the capacitance response from NH4I drop-coated IDE Design 2 appeared to
be favourable to ammonium when compared with the response from each of the
interfering cations. The response from the ‘blank’ membrane that did not contain the
ionophore was around 40% that of the ‘active’ membrane at 1 kHz. At 10 kHz, the
capacitance response was much higher for ammonium compared with the interfering
cations; however, there was a substantial response from the ‘blank’ membrane. At
100 kHz there was a large response from each of the interfering cations and from the
‘blank’ membrane sensor. At 1 MHz and 2 MHz, substantial interference was observed
from potassium and sodium ions; however, the response from the ‘blank’ membrane
was less pronounced.
7.2 Nitrite Sensors
An IDE Design 2 device was drop-coated with a 2% w/v PVC membrane ‘cocktail’
containing 5% w/w of the ionophore NO2I. An aliquot of 5 ppm NO2- was added to a
deionised water background solution and the response was obtained at 10 minute
intervals over two hours. An input amplitude of 10 mVrms was used. This device will be
referred to as ‘NO2I PVC drop-coated IDE Design 2’. The conductance versus time
Craig Alexander
235
data at five frequencies of interest are shown in Figure 7.3 and the capacitance versus
time data at five frequencies of interest are shown in Figure 7.4.
Figure 7.3: Conductance response versus time of NO2I PVC drop-coated IDE Design 2 (2% w/v)
upon the addition of 5 ppm NO2- to a deionised water background at five frequencies. The aliquot
of nitrite stock solution was added after 60 minutes.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
Each of the frequencies of interest showed a distinct increase from the G0 value upon
the addition of 5 ppm NO2- to a deionised water background. At 1 kHz, there was a
response that settled after 10 minutes with a Gm – G0 value of approximately 2.0 µS.
The conductance response at 10 kHz, 100 kHz, 1 MHz and 2 MHz appeared to settle
after around 30 minutes. The Gm – G0 value at 10 kHz was approximately 17 µS. The
Gm – G0 value at 100 kHz was approximately 21 µS. The Gm – G0 value at 1 MHz was
approximately 21 µS. The Gm – G0 value at 2 MHz was approximately 20 µS.
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
2.00E-06
2.50E-06
3.00E-06
3.50E-06
4.00E-06
4.50E-06
5.00E-06
5.50E-06
6.00E-06
6.50E-06
7.00E-06
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
1 hour deionised water baseline
5 ppm NO2-
1.00E-05
1.50E-05
2.00E-05
2.50E-05
3.00E-05
3.50E-05
4.00E-05
4.50E-05
5.00E-05
5.50E-05
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
5.00E-06
1.00E-05
1.50E-05
2.00E-05
2.50E-05
3.00E-05
3.50E-05
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
0.00E+00
5.00E-06
1.00E-05
1.50E-05
2.00E-05
2.50E-05
3.00E-05
0 50 100 150 200 250
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E
Craig Alexander
236
Figure 7.4: Capacitance response versus time of NO2I PVC drop-coated IDE Design 2 (2% w/v)
upon the addition of 5 ppm NO2- to a deionised water background at five frequencies. The aliquot
of nitrite stock solution was added after 60 minutes.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
At 1 kHz, upon the addition of 5 ppm NO2-, the capacitance response increased from the
C0 value by approximately 1.12 nF, and the measurement appeared stable after 30
minutes. The response at 10 kHz increased by approximately 87.54 pF; and appeared
stable after 30 minutes. At 100 kHz, there was an increase to the C0 value of
approximately 2.2 pF; however, this response appeared unstable. The capacitance
response at 1 MHz and 2 MHz decreased below the C0 value when nitrite was added.
As chloride and nitrate are likely to be present at a higher concentration than the target
nitrite anion within an aquarium, it was important to establish the behaviour of NO2I
PVC drop-coated IDE Design 2 at high concentrations of the interfering anions.
Therefore, selectivity data was obtained through the addition of 50 ppm of the
1.276E-10
1.278E-10
1.280E-10
1.282E-10
1.284E-10
1.286E-10
1.288E-10
1.290E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
1.285E-10
1.286E-10
1.287E-10
1.288E-10
1.289E-10
1.290E-10
1.291E-10
1.292E-10
1.293E-10
1.294E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
0.00E+00
5.00E-11
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
0.00E+00
2.00E-10
4.00E-10
6.00E-10
8.00E-10
1.00E-09
1.20E-09
1.40E-09
1.60E-09
1.80E-09
0 50 100 150 200 250
Ca
pa
cita
nce
(F)
Time (min)
A B
C D
E
Craig Alexander
237
interfering anion to a deionised water background. Therefore, it was important to
establish the behaviour of the nitrite sensor when the interfering anions were present in
a greater concentration than that of the target ion. Nitrite at a concentration of 5 ppm
NO2- was also tested using a device that had been prepared without the ionophore
present within the membrane. The baseline-subtracted data at each extrapolated
frequency are displayed in Table 7.3 (G) and Table 7.4 (C).
Gm – G0 (µS)
Ion Frequency
1 kHz 10 kHz 100 kHz 1 MHz 2 MHz
Nitrite (5 ppm) 2.15 17.10 21.39 21.22 19.66
Chloride (50 ppm) 1.34 19.55 25.29 48.00 72.60
Nitrate (50 ppm) 1.2 19.30 25.29 47.64 63.22
Nitrite (5 ppm)
(blank membrane) 0.04 -0.75 9.26 42.84 43.79
Table 7.3: Selectivity determination of NO2I PVC drop-coated IDE Design 2 (2% w/v) (conductive
response). Deionised water baseline-subtracted conductance upon the addition of: 5 ppm of the
target nitrite ion and 50 ppm of the interfering anions chloride and nitrate. The response from the
addition of 5 ppm NO2- to a device prepared without ionophore present is also shown
At 1 kHz the conductance response characteristics of the sensor appeared to be
favourable to nitrite when compared with the response from each of the interfering
anions. There was also very little response from a ‘blank’ membrane that did not
contain the ionophore. At 10 kHz there was an increased response from chloride and
nitrate; however, there was little response from the ‘blank’ membrane. An increased
response was observed for chloride and nitrate at 100 kHz, and the ‘blank’ membrane
device showed an increased response compared to the lower frequency results. At the
higher frequencies of 1 MHz and 2 MHz there was an increased response of both tested
interfering anions when compared to the response obtained from the target nitrite ion.
There was also a substantial response, greater than that of the ‘active’ membrane, from
the ‘blank’ membrane that did not contain the ionophore.
Craig Alexander
238
Cm – C0 (pF)
Ion Frequency
1 kHz 10 kHz 100 kHz 1 MHz 2 MHz
Nitrite (5 ppm) 1117.56 87.54 2.19 - -
Chloride (50 ppm) 1101.79 126.32 10.09 - -
Nitrate (50 ppm) 1040.39 115.16 8.48 - -
Nitrite (5 ppm)
(blank membrane) 15.45 22.67 28.76 - -
Table 7.4: Selectivity determination of NO2I PVC drop-coated IDE Design 2 (2% w/v) (capacitive
response). Deionised water baseline-subtracted capacitance upon the addition of: 5 ppm of the
target nitrite ion and 50 ppm of the interfering anions chloride and nitrate. The response from the
addition of 5 ppm NO2- to a device prepared without ionophore present is also shown
The capacitance response of NO2I PVC drop-coated IDE Design 2 was not investigated
at 1 MHz or 2 MHz, as there was an observed decrease in the measurements when
5 ppm NO2- was added to a deionised water background. At 1 kHz, 50 ppm of each of
the target anions produced a response that was almost equivalent to 5 ppm of the target
nitrite ion. The ‘blank’ membrane produced a response that was <1% of that of the
‘active’ membrane. At 10 kHz, the interfering anions produced a greater response on the
sensor, and an increased response was observed from the ‘blank’ membrane device,
when compared with the 1 kHz response. At 100 kHz, the response from the interfering
anions was even more pronounced and the ‘blank’ membrane response was
substantially greater than the response observed from the ‘active’ membrane that
contained the selective ionophore.
7.3 pH Sensors
Impedimetric pH sensors were prepared by drop-coating a 2% w/v PVC ‘cocktail’
containing 5% w/w of the ionophore onto the sensing area of IDE Design 2. Two
different commercial ionophores were tested, HIII and HV. An input amplitude of
10 mVrms was used. These devices will be referred to as ‘HIII PVC drop-coated IDE
Design 2’ and ‘HV PVC drop-coated IDE Design 2’, respectively.
Craig Alexander
239
The sensors were tested by placing the sensor into a low conductivity buffer solution,
described in Chapter 2 (section 2.7.3.1; Table 2.9), and recording the response at five
minute intervals over 30 minutes. The measurements were carried out in six separate pH
buffer solutions over the likely range found within a freshwater aquarium
(pH 6.5–8.5). As mentioned in section 2.7.3.1, the buffer solutions were prepared by
diluting a 10 mM stock solution at the desired pH value by a factor of 10 in deionised
water. As such, the final pH of the test solution deviated from that of the stock solution
due to a reduced buffering capacity. The final pH of each tested sample solution and the
resulting TDS concentration are shown in Table 7.5.
pH of 10 mM
stock solution
Final pH of diluted
test sample
TDS of diluted
test sample (ppm)
6.00 6.24 68
6.50 6.76 60
7.00 7.24 79
7.50 7.89 86
8.00 8.56 87
8.50 9.04 77
Table 7.5: Final pH and TDS concentration of the diluted buffer solutions used to test the response
of the impedimetric pH sensors
The response versus time data for HIII PVC drop-coated IDE Design 2 are shown in
Figure 7.5 (G) and Figure 7.6 (C). The response versus time data for HV PVC
drop-coated IDE Design 2 are shown in Figure 7.7 (G) and Figure 7.8 (C).
Craig Alexander
240
Figure 7.5: Conductance response versus time of HIII PVC drop-coated IDE Design 2 (2% w/v)
between pH 9.04–6.24 at five frequencies.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
The conductance response of HIII PVC drop-coated IDE Design 2 does not appear to
show an analytically-relevant change over the pH range 9.04–6.24 at any of the
extrapolated frequencies. None of the measurements in any of the tested solutions
appeared to produce a stable response over the 30 minutes that they were submerged
within each test solution. Therefore, further analysis of the conductance results obtained
using this device was not possible.
5.96E-03
5.97E-03
5.98E-03
5.99E-03
6.00E-03
6.01E-03
6.02E-03
6.03E-03
6.04E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
5.85E-03
5.86E-03
5.87E-03
5.88E-03
5.89E-03
5.90E-03
5.91E-03
5.92E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
5.79E-03
5.80E-03
5.80E-03
5.81E-03
5.81E-03
5.82E-03
5.82E-03
5.83E-03
5.83E-03
5.84E-03
5.84E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
5.760E-03
5.765E-03
5.770E-03
5.775E-03
5.780E-03
5.785E-03
5.790E-03
5.795E-03
5.800E-03
5.805E-03
5.810E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
5.74E-03
5.75E-03
5.76E-03
5.77E-03
5.78E-03
5.79E-03
5.80E-03
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
pH 9.04
pH 8.56
pH 7.89
pH 7.24
pH 6.76pH 6.24
A B
C D
E
Craig Alexander
241
Figure 7.6: Capacitance response versus time of HIII PVC drop-coated IDE Design 2 (2% w/v)
between pH 9.04–6.24 at five frequencies.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
The capacitance response of HIII PVC drop-coated IDE Design 2 produced a more
stable measurement at each tested pH than the conductance response at several
frequencies. At 1 kHz, there was no discernible response from the capacitance as the
sensor was placed into solutions with varying pH. At 10 kHz, 100 kHz, 1 MHz and
2 MHz, the capacitance response decreased as the pH decreased. This was not as
expected as an increase in the concentration of hydrogen ions within the test solution
would result in an increase in the measured conductance as H+ moved into the selective
membrane layer.
The effect of changing the ionophore on the response was investigated by drop-coating
a 2% w/v PVC membrane containing the ionophore HV onto the sensing area of
IDE Design 2.
1.080E-10
1.085E-10
1.090E-10
1.095E-10
1.100E-10
1.105E-10
1.110E-10
1.115E-10
1.120E-10
1.125E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
1.090E-10
1.100E-10
1.110E-10
1.120E-10
1.130E-10
1.140E-10
1.150E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
1.00E-10
1.05E-10
1.10E-10
1.15E-10
1.20E-10
1.25E-10
1.30E-10
1.35E-10
1.40E-10
1.45E-10
1.50E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
4.00E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
6.00E-10
6.50E-10
7.00E-10
7.50E-10
8.00E-10
8.50E-10
9.00E-10
9.50E-10
1.00E-09
1.05E-09
1.10E-09
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
A B
C D
E
Craig Alexander
242
Figure 7.7: Conductance response versus time of HV PVC drop-coated IDE Design 2 (2% w/v)
between pH 9.04–6.24 at five frequencies.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
The conductance response of HV PVC drop-coated IDE Design 2 demonstrated a
reduction in the measured response as the solution pH was decreased at 10 kHz,
100 kHz, 1 MHz and 2 MHz. The observed responses at each frequency were more
stable than the conductance response when HIII was used as the ionophore. As for the
results obtained when HIII was the ionophore, the observed change decreased as the
hydrogen ion concentration in the test solution was increased.
1.00E-04
1.20E-04
1.40E-04
1.60E-04
1.80E-04
2.00E-04
2.20E-04
2.40E-04
2.60E-04
0 50 100 150 200
Con
du
ctan
ce (S
)
Time (min)
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
8.00E-05
0 50 100 150 200
Con
du
ctan
ce (S
)
Time (min)
5.00E-06
7.00E-06
9.00E-06
1.10E-05
1.30E-05
1.50E-05
1.70E-05
1.90E-05
2.10E-05
2.30E-05
2.50E-05
0 50 100 150 200
Con
du
ctan
ce (S
)
Time (min)
3.00E-06
3.50E-06
4.00E-06
4.50E-06
5.00E-06
5.50E-06
6.00E-06
6.50E-06
7.00E-06
0 50 100 150 200
Co
nd
uct
ance
(S)
Time (min)
A B
C D
E
6.00E-05
8.00E-05
1.00E-04
1.20E-04
1.40E-04
1.60E-04
1.80E-04
0 50 100 150 200
Con
du
ctan
ce (S
)
Time (min)
Craig Alexander
243
Figure 7.8: Capacitance response versus time of HV PVC drop-coated IDE Design 2 (2% w/v)
between pH 9.04–6.24 at five frequencies.
A – 1 kHz. B – 10 kHz. C – 100 kHz. D – 1 MHz. E – 2 MHz.
The capacitance response of HV PVC drop-coated IDE Design 2 showed a reduction in
the measurement as the pH of the test solution was decreased.
There was a limitation in the method that was used for measuring the response from the
pH sensors, as it required the complete removal of the device from one test sample
before being rinsed and placed into the next. This could have caused impedance effects
from the movement of the coaxial cable connecting the sensor under test to the LCR
meter. As such, the measurements obtained were unreliable and it cannot be definitively
stated that the observed response was a result of changes in the H+ concentration within
the test sample. One possible alternative method for testing impedimetric pH sensors,
without the requirement of withdrawing and submerging the sensor into different
sample solutions, would be to use an exponential dilution system. The sensor under test
1.330E-10
1.335E-10
1.340E-10
1.345E-10
1.350E-10
1.355E-10
1.360E-10
1.365E-10
1.370E-10
1.375E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
E
1.36E-10
1.37E-10
1.38E-10
1.39E-10
1.40E-10
1.41E-10
1.42E-10
1.43E-10
1.44E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
1.40E-10
1.50E-10
1.60E-10
1.70E-10
1.80E-10
1.90E-10
2.00E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
1.00E-10
1.50E-10
2.00E-10
2.50E-10
3.00E-10
3.50E-10
4.00E-10
4.50E-10
5.00E-10
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
6.00E-10
7.00E-10
8.00E-10
9.00E-10
1.00E-09
1.10E-09
1.20E-09
1.30E-09
1.40E-09
1.50E-09
0 50 100 150 200
Cap
acit
ance
(F)
Time (min)
A B
C D
Craig Alexander
244
would be placed into a sample solution of known pH at the highest or lowest required
hydrogen ion concentration. Using a peristaltic pump at a known flow rate, a constant
addition of a pH buffer could then be added to either steadily increase or decrease the
hydrogen ion concentration within the test sample over the desired pH range over time.
Due to time constraints it was not possible to further investigate IDE Design 2 for its
use as a possible pH sensor.
7.4 Conclusions for Chapter 7
Several experiments were conducted to investigate the use of IDE Design 2 as an
impedimetric sensor for ammonium, nitrite and pH.
When 50 ppm NH4+ was added to a deionised water background, an increase in both the
conductance and capacitance was observed from NH4I PVC drop-coated IDE Design 2
at each of the extrapolated frequencies. The selectivity of this device was investigated
through the addition of 50 ppm of the interfering cations potassium, sodium and
magnesium to a deionised water background, without the target ion present. At lower
frequencies of 1 kHz and 10 kHz there was a decrease in both the conductance and
capacitance response of each of the investigated interfering cations compared with the
response obtained for a solution containing only 50 ppm ammonium. However, there
appeared to be a substantial conductance and capacitance response when 50 ppm NH4+
was added to a solution and tested with a device that contained a ‘blank’ membrane. At
increased frequencies, both the interfering cation and ‘blank’ membrane responses
became more prominent. This suggested that the membrane may have been too thin for
this electrode geometry and therefore changes in the conductivity of the bulk solution
were observed.
Craig Alexander
245
Nitrite was added to a deionised water background and the resulting change in
conductance and capacitance was monitored using NO2I drop-coated IDE Design 2. An
increase in the response of the sensor was observed from the capacitance at all of the
extrapolated frequencies and from the conductance at frequencies <100 kHz. The
selectivity of this device was investigated through the addition of 50 ppm of the
interfering anions chloride and nitrate to a deionised water background, without the
target ion present. At a frequency of 1 kHz, there was a decrease in both the
conductance and capacitance response of each of the investigated interfering cations
compared with the response obtained for a solution containing only 5 ppm nitrite. At
this particular frequency very little response was obtained from the ‘blank’
PVC membrane.
Both the ammonium and nitrite sensors constructed from IDE Design 2 appeared to
demonstrate favourable response characteristics with regards to ion-selectivity at lower
frequencies. Therefore, further investigation is required into the frequency-dependence
of the response of IDE-based sensors, as the interference from the bulk solution was
most marked at the highest investigated frequencies.
Experiments conducted using IDE Design 2, which was coated with a 2% w/v PVC
membrane containing two different ionophores for H+ showed a decrease in the
response as the concentration of hydrogen ions within the test sample was increased.
This was not as expected and may have arisen due to a limitation with the method that
was used, which required the removal of the sensor from the test solution. Further
experimentation is required into the use of IDE devices coated with a polymeric
membrane as impedimetric pH sensors.
Craig Alexander
246
8 Conclusions and Suggestions for Further Work
8.1 Conclusions
The main focus of this thesis was to investigate the use of IDEs coated in a selective
membrane as impedimetric chemical sensors. Three separate devices based on IDEs
were fabricated using either photolithography or screen-printing and tested as
impedimetric ion-selective sensors for the determination of key ionic analytes found
within a freshwater aquarium.
The initial experimental work focused on the characterisation of commercially-available
ionophores for each of the identified aquarium-significant ions within conventional
potentiometric ISEs, to establish their feasibility for use within a freshwater sensing
device. Calibration and selectivity data were obtained from PMEs containing the
ionophores of interest and the results obtained deemed that each membrane composition
would be suitable for further testing within an IDE-based sensor. CWEs for nitrate were
also produced by coating the polymeric membrane over the surface of a Ag/AgCl
electrode, but these did not demonstrate acceptable response characteristics when
compared with PMEs. Sol-gels were investigated as an alternative membrane material
to the conventional PVC membranes; however, they exhibited poor response
characteristics that may have arisen from inadequate adhesion to the electrode.
IDE Design 1 was fabricated in-house using a ‘mask-less’ photolithography technique
followed by e-beam deposition of a 50 nm layer of gold metal. The sensing area of this
device was coated in a polymeric membrane containing a commercially-available nitrate
ionophore via spin-coating. The results from this particular device suggested that
non-selective changes in the bulk conductivity of the test solution were observed. It was
concluded that the electrode geometry of this device, with regards to the digits, was too
Craig Alexander
247
large and therefore the resulting penetration depth of the generated electric field was not
contained within the membrane layer and reached into the bulk test solution. Due to the
lack of selectivity and arduous fabrication process involved with this device, further
investigation was not carried out. Alternative devices were sought which featured either
reduced electrode geometry and/or an increased thickness of the selective
membrane layer.
To produce an impedimetric sensor with reduced electrode geometry, IDE Design 2 was
constructed from a commercial microfabricated IDE transducer. Initially the sensing
area of the device was coated with a PVC membrane via spin-coating; however, a large
response was observed when a device with a ‘blank’ membrane was tested. In an
attempt to overcome this and produce a thicker membrane layer, the polymeric material
was deposited via drop-coating. This also had the advantage of reducing the volume of
membrane solution that is wasted compared with spin-coating. Drop-coating of the PVC
membrane resulted in an increased response time of the sensing device due to an
increased diffusion distance. To improve the sensor response time, the PVC ‘cocktail’
viscosity was reduced by decreasing the mass of polymer and plasticiser that was
dissolved into the THF membrane solvent. A PVC membrane cocktail consisting of
2% w/v of the membrane components appeared to produce a sensor with a more
acceptable response time, so this composition was investigated further. The
nitrate-selective ionophore TDAN was investigated initially; however, when this
ionophore was tested it did not demonstrate adequate ion-selectivity and a response was
observed when a ‘blank’ membrane was deposited. An alternative ionophore, NO3V,
was investigated. The results from this membrane composition also demonstrated
interference from non-target anions; however, the response from the ‘blank’ membrane
was substantially lower than for the TDAN membrane. The difference between the two
Craig Alexander
248
‘blanks’ was the choice of plasticiser within the membrane. NPOE, which was used
with the ionophore NO3V, may have imparted greater hydrophobicity on the PVC
membrane and therefore a lower response was observed when no ionophore was
present.
Several experiments were conducted using IDE Design 2 as a sensor for the other
identified aquarium significant ions: ammonium, nitrite and pH. The sensing area of the
device was drop-coated with a 2% w/v PVC membrane containing 5% w/w of the
respective ionophore. For ammonium, the response of the sensor was ascertained by
measuring the response of a single concentration of the target ion (50 ppm). The
selectivity of the sensor was investigated by measuring the obtained response from
50 ppm of the interfering cations potassium, sodium and magnesium. For nitrite, the
response of the sensor was ascertained by measuring the response of a single
concentration of the target ion (5 ppm). The selectivity of the sensor was investigated by
measuring the obtained response from 50 ppm of the interfering anions chloride and
nitrate. It was observed that the ion-selectivity of both the ammonium and nitrite
sensors was greatest at the lowest investigated frequencies. The responses from the
devices that were coated in ‘blank’ membrane were also lowest in the low frequency
region. The results obtained from IDE Design 2 devices coated in an H+-selective
membrane were inconclusive, although this appeared to be as a result of the method that
was used, not from the devices themselves.
SP IDE Design 1 was produced as a fully screen-printed device which consisted of
carbon IDEs coated with a selective membrane layer. The ion-selective membrane layer
was modified to be suitable for screen-printing by mixing a commercial dielectric
screen-printing paste with NPOE as the plasticiser and TDAN as a
Craig Alexander
249
nitrate-selective ionophore. An initial attempt at producing screen-printed carbon IDE
transducers resulted in devices with many short-circuits as the digits were printed too
close together. An IDE geometry of 100 µm digit width with a 150 µm space between
adjacent digits was found to print effectively. When a single layer of the modified
dielectric paste ion-selective membrane layer was printed over the sensing area, strong
non-selective matrix interferences were observed. These were reduced by printing
‘build-up’ layers of non-modified paste prior to the final sensing layer. This
significantly reduced the observed response from a ‘blank’ membrane, which did not
contain any ionophore. When TDAN was used as the nitrate-selective ionophore, a
strong interference was observed from both chloride and nitrite anions, which would
render this particular membrane composition unsuitable for use within a freshwater
aquarium. Availability issues dictated that it was not possible for further screen-printed
IDE devices to be produced.
Due to the commercial aspect of this project, it is very important to consider the final
cost of producing individual ion-selective sensors based on IDEs, using either
photolithographic or screen-printing fabrication techniques. Screen-printing is generally
seen as the more cost-effective route for the high-throughput, mass-production of
sensing devices; however, due to the cost of commercially-available ionophores, this
may not be the case for this specific application. Both fabrication techniques require
further investigation into producing the best possible ion-selectivity. If both fabrication
techniques prove to produce devices which are viable for use within a freshwater
aquarium monitoring device, then the final cost-per-device needs to be established. This
would be calculated by ascertaining the cost of the transduction element together with
the cost of the recognition layer.
Craig Alexander
250
The use of IDEs as impedimetric sensors for aquarium water quality monitoring
provides several advantages over currently available methods. The subjectivity of wet
chemical testing is removed and the fabrication techniques involved allow for a more
cost-effective alternative to commercial ISEs. IDE devices are also well-suited to
disposable sensing without the requirement of an external reference electrode.
This thesis has described the first use of fully screen-printed ISCOM-type devices
which included both a screen-printed transduction element and a screen-printed
ion-selective recognition layer. This brought together research conducted by
Jaffrezic-Renault et al.161
, Camman et al.156
and Port et al.162
into the development of
ISCOM devices, with the research into screen-printable ion-selective membranes for
potentiometric applications, as described previously by Koncki et al.76
To the author’s
knowledge, this research also describes the first use of ISCOM-type devices with
ionophore-doped silica gels prepared via the sol-gel method as an alternative
ion-selective membrane material.
A substantial amount of further work is required to fully optimise and validate the
individual sensors before moving to the next stage of producing a prototype device. The
purpose of the following section is to discuss some of the potential experimental work
required to achieve this, along with considerations for the design of a prototype
multi-analyte aquarium sensing device.
8.2 Future Work
8.2.1 Individual Sensor Testing
Further investigation into individual ion sensors is required, such as for ammonium. As
mentioned previously in section 1.1.1.1, the ratio between the concentrations of
unionized and ionized ammonia is heavily dependent on both the pH and the
Craig Alexander
251
temperature of the environment. It is therefore important that calibration experiments
are carried out in a background buffer solution at several pH values in the expected pH
range of an aquarium. It is important to establish a suitable concentration range which
provides a wide range of unionized, toxic ammonia at each pH value, around the
acceptable upper limit for aquaria of 0.02 ppm NH3. Table 8.1 shows the unionized
ammonia concentration in the test vessel following the additions of a NH4+ stock
solution over the TAN concentration range 0.25–50 ppm, at pH values between pH 6.0
and 8.5. This would allow the observation of the response of the sensor using the same
TAN concentrations, but at varying ratios of non-toxic NH4+ and toxic NH3.
Ammonium concentration added (ppm)
pH 6.0 pH 6.5 pH 7.0 pH 7.0 (30 °C)
pH 7.5 pH 8.0 pH 8.5
0.25 0.00014 0.00045 0.00141 0.00199 0.0044 0.01342 0.03801
0.5 0.00028 0.00089 0.00282 0.00398 0.00881 0.02683 0.07603
1 0.00057 0.00179 0.00564 0.00796 0.01762 0.05366 0.15205
2 0.00113 0.00358 0.01128 0.01593 0.03523 0.10732 0.3041
5 0.00283 0.00895 0.02819 0.03982 0.08808 0.26831 0.76026
10 0.00567 0.0179 0.05639 0.07965 0.17616 0.53662 1.52052
50 0.02834 0.0895 0.28193 0.39823 0.88079 2.6831 7.60258
Table 8.1: The resulting concentration, in ppm, of unionized ammonia present following additions
of an ammonium stock solution to a test sample at various pH values. The values highlighted in
green are below the acceptable aquarium limits of 0.02 ppm NH3, whereas the values highlighted
in red are above this limit. Unless otherwise stated these concentrations relate to the equilibrium
of the two forms of ammonia at 25 °C
As the measured ammonia concentration is pH-dependent, it is important to ensure that
the ammonium measurement is linked to the pH measurement so that it can be corrected
accordingly. Likewise, pH is temperature dependent and therefore would require the pH
measurement to be linked to a temperature sensor. A suitable algorithm, taking into
account both the pH and temperature of the aquarium, would need to be incorporated
into any software that was written for a commercial device, to convert the measured
NH4+ concentration into its respective NH3 concentration, as this is the value that is
Craig Alexander
252
important for aquarium users to understand. It is also of upmost importance that the
effects of changing the pH within the aquatic environment on the response of sensors
for the other ions, such as nitrate and nitrite, are investigated.
8.2.2 Membrane Characterisation
It has been shown that one of the most important considerations for utilising IDE-based
sensors, which consist of a membrane layer coating the sensing area, is the thickness of
the deposited membrane in relation to the IDE geometry. Unfortunately, throughout this
project it was not possible to characterise any of the deposited membrane layers to
ascertain an exact thickness measurement. It would be very useful to be able to
accurately determine the physical characteristics of the deposited polymeric membrane
so that this information could be used to predict the effects that matrix interferences are
likely to have on the response of the sensor. By using specialist thin-film thickness
measurement equipment, such as the optical instruments offered by Filmetrics177
, or
microscopy techniques, such as atomic force microscopy (AFM), it may be possible to
accurately deduce the thickness of the ion-selective membrane layer.It has also been
shown that it is possible to use alternative coating techniques to increase the thickness
of the ion-selective membrane over the sensing area of the IDEs to contain more of the
generated electric field; however, this results in an increased response time of the sensor
due to an increased diffusion distance.
8.2.3 Membrane Materials
The use of sol-gels as an alternative ion-selective membrane material requires further
investigation. Sol-gels based on silica precursor materials were investigated throughout
this project; however, the results obtained were inconclusive as to their potential
feasibility within an impedimetric aquatic water sensing device. The inconclusive
Craig Alexander
253
results were largely caused by an unsuitable transduction element as opposed to the
selective membrane recognition layer. Further investigation into optimising the
composition of the silica gels is required. The ability to accurately characterise the
thickness of a deposited sol-gel layer, as described earlier, would aid this as it would
allow the user to predict the suitability of the sol-gel as an ion-selective membrane
based on its thickness relative to the membrane geometry. It would be possible to adjust
the viscosity of the sol-gel solution, and therefore the thickness of the final cast
membrane, by adjusting parameters such as the mass of precursor used and the volume
of solvent used during its preparation.
8.2.4 IDE Geometry
It would be most beneficial to investigate the response of an IDE sensor which had the
smallest possible geometry with respect to the digit widths and interdigital spacing. This
would allow the transduction element of the sensor to be coupled with an ion-selective
membrane layer that was very thin without compromising either the ion-selectivity or
response time. For example, if a microfabricated IDE transducer was designed with
digit widths and spacings of 1 µm, then a thin-film membrane of 2 µm would be
required to enclose the resulting electric field when excited with an AC voltage. Such a
thickness would be very achievable using conventional polymeric coating techniques.
Unfortunately, throughout this project, it was not possible to obtain any sensors with an
electrode geometry lower than IDE Design 2, which was purchased from MicruX
Fluidics, and consisted of digit widths and interdigital spacings of 10 µm. IDE Design 2
also appeared to demonstrate poor sensitivity which may have arisen due to the low
number of digits at each electrode (15). It would be advantageous to investigate how the
number of digits at each electrode affects the response of the sensors. This would also
Craig Alexander
254
be benefitted by having the smallest possible digit geometry, as a larger number of
digits could fit into a small sensing area.
8.2.5 Screen-Printed Sensors
Screen-printing has also been demonstrated as an appropriate fabrication technique for
producing ion-selective sensors based on IDEs. A layer-by-layer approach was used to
produce fully screen-printed devices, which included a screen-printed ion-selective
membrane layer by incorporating ionophores and plasticisers into a commercial
dielectric screen-printing paste. Further research is required into adjusting the
components within the recognition layer to maximise the selectivity of the sensor
response, by investigating several combinations of dielectric pastes, plasticisers and
ionophores.
8.2.6 Considerations for a Prototype Device
As the overall aim of this project was to develop sensors which could be implemented
within a prototype commercial device, it was important to consider a suitable design.
The individual sensors for each target ion need to be optimised with regards to the
electrode geometry, membrane composition, optimum frequency and input amplitude.
Once this information has been determined, it would be beneficial to produce bespoke
circuitry to power multiple sensors simultaneously along with a suitable housing for the
prototype. The IDE sensors could be produced as a replaceable transduction element
that is placed into the housing containing the necessary circuitry to operate them.
Screen-printing would likely provide the more cost-effective route for the transduction
element; however, the relatively large sensing area would require more of the often
expensive ionophore to be present within the membrane layer. A commercial IDE
Craig Alexander
255
device which has been fabricated using photolithography would likely be more
expensive than a screen-printed IDE device; however, the smaller overall footprint of its
sensing area would only require a very small amount (1 µl or less) of membrane
material to be fully coated.
A screen-printed device would have the advantage that it could be produced
‘ready-to-use’ with no further modification required, and it is possible to use flexible
substrates which would be better suited to disposable sensing applications. However,
the possibility of automating the deposition of a polymeric membrane on to an IDE
device using robotics could be investigated. The small surface area of a microfabricated
device could ensure that the sensing areas are produced on to an appropriate substrate,
such as glass, and then mounted into a suitable holder which would act as the
replaceable transduction element of the prototype.
8.2.7 Sensor Longevity
As the LCR meter that was used during this project was a single-input device, it did not
lend itself particularly well for use in determining the longevity of the impedimetric
sensors. It would be favourable to do this with a bespoke prototype device that is
capable of monitoring several sensors alongside one another. The sensors could be
placed into a test sample solution containing the target ions at a known, fixed
concentration and the response would be monitored over time. Degradation in the
observed signal would determine the operational lifetime of the sensors.
The operational lifetime of the sensors would be dictated by several factors, such as
leaching of the selective components from the membrane material and biofouling.
Either of these could be overcome by producing a disposable sensing element which
Craig Alexander
256
would require replacement at regular intervals. The device-to-device reproducibility of
the individual sensors could also be more accurately determined using bespoke
circuitry. Once reproducibility had been established, it could be possible to batch
calibrate devices, which would mean individual sensors would not require the final user
to calibrate them.
8.2.8 Data Display
One final consideration is how the results are to be displayed to the user. One option is
to use a screen which is connected directly to the device which could be positioned on
the outside of the user’s aquarium. Another option would be to use an appropriate
transmitter to wirelessly send the measurements to a suitable receiver unit, such as a PC
or smartphone with appropriate software installed.
Craig Alexander
257
9 References
1. http://www.ornamentalfish.org/aquanautmarket/ukmarket.php (accessed: 22/11/2013)
2. Pet Food Manufacturers Association, Pet Population 2013. Available at: http://www.pfma.org.uk/pet-
population/ (accessed: 20/11/2013)
3. European Pet Food Industry, Facts and Figures 2012. Available at: http://www.fediaf.org/facts-figures/
(accessed 20/11/2013)
4. American Pet Products Association, Pet Industry Size and Ownership Statistics 2013/14. Available at
http://www.americanpetproducts.org/press_industrytrends.asp (accessed: 20/11/2013)
5. K. J. Ruff, Freshwater Aquarium Chemistry - A Comprehensive Guide for Tropical Fish Hobbyists,
AquaChem Publishing, 2009.
6. R. Grommen, I. Van Hauteghem, M. Van Wambeke and W. Verstaete, Aquaculture, 2002, 211, 115-
124.
7. K. Emerson, R. C. Russo, R. E. Lund and R. V. Thurston, J. Fish. Res. Board Can., 1975, 32, 2379-
2383.
8. J. J. Keaffaber, R. Palma and K. R. Williams, J. Chem. Educ., 2008, 85, 225-230.
9. H. Roberts and B. Palmeiro, Vet. Clin. N. Am.: Exotic, 2008, 11, 359-374.
10. R. Baird, J. Bottomley and H. Taitz, Water Res., 1979, 13, 181-184.
11. F. B. Eddy, J. Fish Biol., 2005, 67, 1495-1513.
12. Ornamental Aquatic Trade Association (OATA) - Water Quality Criteria, Version 2.0, March 2008.
Available at http://oata.wedidthis.co.uk/wp-content/uploads/2012/08/Water-Quality-Criteria.pdf
(accessed 22/10/2013)
13. H. Kroupova, J. Machova and Z. Svobodova, Vet. Med., 2005, 50, 461-471.
14. J. A. Camargo, A. Alonso and A. Salamanca, Chemosphere, 2005, 58, 1255-1267.
15. R. R. Hensen, OFI Educational Publication 1 - Water Quality in the Ornamental Aquatic Industry,
Ornamental Fish International, 2006.
16. Dissolved Oxygen Fact Sheet - Available at
http://www.waterboards.ca.gov/water_issues/programs/swamp/docs/cwt/guidance/3110en.pdf (accessed:
06/12/2013)
17. Oxygen in the Aquarium - Available at http://www.algone.com/articles/technical-aquarium-
information/oxygen-in-the-aquarium (accessed: 6/12/13)
18. J. H. Tullock, Water Chemistry for the Marine Aquarium, Barron's Educational Series, Inc., 2002.
19. J. Saurina, E. Lopez-Aviles, A. Le Moal and S. Hernandez-Cassou, Anal. Chim. Acta, 2002, 464, 89-
98.
20. Phenol red acid/base equilibria. Available at http://commons.wikimedia.org/wiki/File:Phenol_red_-
_acid_base_equilibria_-_2.gif (accessed: 11/12/2013)
21. UV-Vis Spectrometry - Available at http://ishigirl.tripod.com/pchem/pka.html (accessed: 11/12/2013)
Craig Alexander
258
22. C. W. M.I. Nichols, J. Am. Chem. Soc., 1934, 56, 769-774.
23. H. Verdouw, C. J. A. Van Echteld and E. M. J. Dekkers, Water Res., 1978, 12, 399-402.
24. K. J. Ruff, Freshwater Test Kit Review, AquaChem Publishing, 2010.
25. Biotek® Application Note - Nitric Oxide Determination using the PowerWave 200 Scanning
Microplate Spectrophotometer and KC4 Data Reduction Software - Available at
http://www.biotek.de/resources/docs/PowerWave_200_Nitric_Oxide_Determination.pdf (accessed:
11/12/2013)
26. R. W. Catterall, Chemical Sensors, Oxford University Press, 1997.
27. F.-G. Bănică, Chemical Sensors and Biosensors: Fundamentals and Applications, John Wiley & Sons
Ltd., 2012.
28. U. E. Spichiger-Keller, Chemical Sensors and Biosensors for Medical and Biological Applications,
Wiley-VCH, 1998.
29. B. R. Eggins, Chemical Sensors and Biosensors, John Wiley & Sons Ltd., 2002.
30. C. M. A. Brett, Electroanalysis, Oxford University Press, 1998.
31. F. J. H. D.A. Skoog, S.R. Crouch, Principles of Instrumental Analysis, Thomson Brooks/Cole, 2007.
32. K. N. Mikhelson, Ion-Selective Electrodes - Lecture Notes in Chemistry, Springer-Verlag, Berlin,
2013.
33. R. W. Cattrall, Chemical Sensors, Oxford University Press, 1997.
34. P. Buhlmann, Ion-Selective Electrodes with Ionophore-Doped Sensing Membranes, John Wiley and
Sons Ltd., 2012.
35. A. K. Singh, G. Bhattacharjee, M. Singh and S. Chandra, Electroanal., 1997, 9, 1005-1008.
36. E. Malinowska, V. Oklejas, R. W. Hower, R. B. Brown and M. E. Meyerhoff, Sensor. Actuat., B-
Chem, 1996, 33, 161-167.
37. S. Y. Yun, Y. K. Hong, B. K. Oh, G. S. Cha, H. Nam, S. B. Lae and J. I. Jin, Anal. Chem., 1997, 69,
868-873.
38. R. D. Johnson and L. G. Bachas, Anal. Bioanal. Chem., 2003, 376, 328-341.
39. A. Craggs, G. J. Moody and J. D. R. Thomas, J. Chem. Educ., 1974, 51, 541-544.
40. U. Schaller, E. Bakker, U. E. Spichiger and E. Pretsch, Anal. Chem., 1994, 66, 391-398.
41. F. Faridbod, M. R. Ganjali, R. Dinarvand and P. Norouzi, Afr. J. Biotechnol., 2007, 6, 2960-2987.
42. R. Eugster, T. Rosatzin, B. Rusterholz, B. Aebersold, U. Pedrazza, D. Ruegg, A. Schmid, U. E.
Spichiger and W. Simon, Anal. Chim. Acta, 1994, 289, 1-13.
43. Cristina Mihali and Nora Vaum (2012). Use of Plasticizers for Electrochemical Sensors, Recent
Advances in Plasticizers, Dr. Mohammad Luqman (Ed.), InTech. Available from:
http://www.intechopen.com/books/recent-advances-in-plasticizers/use-of-plasticizers-for-
electrochemical-sensors (accessed: 19/12/2013)
Craig Alexander
259
44. Mohsen M. Zareh (2012). Plasticizers and Their Role in Membrane Selective Electrodes, Recent
Advances in Plasticizers, Dr. Mohammad Luqman (Ed.), InTech. Available from:
http://www.intechopen.com/books/recent-advances-in-plasticizers/plasticizers-and-their-role-in-
membrane-selective-electrodes (accessed: 19/12/2013)
45. E. Bakker, R. Meruva, E. Pretsch and M. Meyerhoff, Anal. Chem., 1994, 66, 3021-3030.
46. E. Lindner and B. D. Pendley, Anal. Chim. Acta, 2013, 762, 1-13.
47. G. G. Guilbault, R. A. Durst, M. S. Frant, H. Freiser, E. H. Hansen, T. S. Light, E. Pungor, G.
Rechnitz, N. M. Rice, T. J. Rohm, W. Simon and J. D. R. Thomas, Pure Appl. Chem., 1976, 48, 127-132.
48. Y. Umezawa, K. Umezawa and H. Sato, Pure Appl. Chem., 1995, 67, 507-518.
49. K. Umezawa, P. Buhlmann, Y. Umezawa, K. Tohda and S. Amemiya, Pure Appl. Chem., 2000, 72,
1851-2082.
50. P. Kane and D. Diamond, Talanta, 1997, 44, 1847-1858.
51. E. Bakker, R. K. Maruva, E. Pretsch and M. E. Mayerhoff, Anal. Chem., 1994, 66, 3021-3030.
52. V. P. Y. Gadzekpo and G. D. Christian, Anal. Chim. Acta, 1984, 164, 279-282.
53. F. Deyhimi, Talanta, 1999, 50, 1129-1134.
54. H. X. Wang and M. Pu, Chin. Chem. Lett., 2002, 13, 355-358.
55. A. Radu, T. Radu, C. McGraw, P. Dillingham, S. Anastasova-Ivanova and D. Diamond, J. Serb.
Chem. Soc., 2013, 78, 1729-1761.
56. G. J. Moody, R. B. Oke and J. D. R. Thomas, Analyst, 1970, 95, 910-918.
57. E. Bakker, P. Bühlmann and E. Pretsch, Chem. Rev., 1997, 97, 3083-3132.
58. P. W. Alexander, T. Dimitrakopoulos and D. B. Hibbert, Electroanal., 1998, 10, 707-712.
59. R. W. Cattrall and H. Freiser, Anal. Chem., 1971, 43, 1905-1906.
60. H. James, G. Carmack and H. Freiser, Anal. Chem., 1972, 44, 856-857.
61. H. Freiser, J. Chem. Soc., Faraday Trans., 1986, 82, 1217-1221.
62. A. Abbaspour, A. Izadyar and H. Sharghi, Anal. Chim. Acta, 2004, 525, 91-96.
63. A. Shokrollahi, A. Abbaspour, M. Ghaedi, A. N. Haghighi, A. H. Kianfar and M. Ranjbar, Talanta,
2011, 84, 34-41.
64. H. R. Zare, M. Salavati-Niassary, F. Memarzadeh, M. Mazloum and N. Nasirizadeh, Anal. Sci., 2004,
20, 815-819.
65. M. M. Ardakany, A. A. Ensafi, H. Naeimi, A. Dastanpour and A. Shamlli, Sensor. Actuat., B-Chem,
2003, 96, 441-445.
66. M. Aghamohammadi and N. Alizadeh, Anal. Chim. Acta, 2003, 480, 299-306.
67. M. M. Ardakani, A. Dastanpour and M. Salavati-Niasari, J. Electroanal. Chem., 2004, 568, 1-6.
68. J. P. Metters, R. O. Kadara and C. E. Banks, Analyst, 2011, 136, 1067-1076.
Craig Alexander
260
69. R. W. Hower, J. H. Shin, G. S. Cha, R. K. Meruva, M. E. Meyerhoff and R. B. Brown, Sensor.
Actuat., B-Chem, 1996, 33, 168-172.
70. G. S. Cha, D. Liu, M. E. Meyerhoff, H. C. Cantor, A. R. Midgley, H. D. Goldberg and R. B. Brown,
Anal. Chem., 1991, 63, 1666-1672.
71. H. D. Goldberg, R. B. Brown, D. P. Liu and M. E. Meyerhoff, Sensor. Actuat., B-Chem., 1994, 21,
171-183.
72. S. Walsh, D. Diamond, J. McLaughlin, E. McAdams, D. Woolfson, D. Jones and M. Bonner,
Electroanal., 1997, 9, 1318-1324.
73. R. Zielińska, E. Mulik, A. Michalska, S. Achmatowicz and M. Maj-Zurawska, Anal. Chim. Acta,
2002, 451, 243-249.
74. A. E. Musa, F. J. Del Campo, N. Abramova, M. A. Alonso-Lomillo, O. Domínguez-Renedo, M. J.
Arcos-Martínez, M. Brivio, D. Snakenborg, O. Geschke and J. P. Kutter, Electroanal., 2011, 23, 115-
121.
75. R. Koncki, S. Głab, J. Dziwulska, I. Palchetti and M. Mascini, Anal. Chim. Acta, 1999, 385, 451-459.
76. L. Tymecki, S. Glab and R. Koncki, Sensors, 2006, 6, 390-396.
77. Ł. Tymecki, E. Zwierkowska and R. Koncki, Anal. Chim. Acta, 2004, 526, 3-11.
78. U. Lange, N. V. Roznyatovskaya and V. M. Mirsky, Anal. Chim. Acta, 2008, 614, 1-26.
79. J. Bobacka, Electroanal., 2006, 18, 7-18.
80. C. Jimenez-Jorquera, J. Orozco and A. Baldi, Sensors, 2010, 10, 61-83.
81. U. Oesch and W. Simon, Anal. Chem., 1980, 52, 692-700.
82. P. Bühlmann, Y. Umezawa, S. Rondinini, A. Vertova, A. Pigliucci and L. Bertesago, Anal. Chem.,
2000, 72, 1843-1852.
83. C. A. Staples, R. Guinn, K. Kramarz and M. Lampi, Hum. Ecol. Risk Assess., 2011, 17, 1057-1076.
84. M. Collinson, Crit. Rev. Anal. Chem., 1999, 29, 289-311.
85. A. M. Buckley and M. Greenblatt, J. Chem. Educ., 1994, 71, 599-602.
86. M. M. Collinson, TrAC, Trends Anal. Chem., 2002, 21, 30-38.
87. K. Kimura, T. Sunagawa and M. Yokoyama, Chem. Lett., 1995, 967.
88. K. Kimura, T. Sunagawa and M. Yokoyama, Chem. Commun., 1996, 745.
89. K. Kimura, T. Sunagawa and M. Yokoyama, Anal. Chem., 1997, 69, 2379-2383.
90. K. Kimura, T. Sunagawa, S. Yajima, S. Miyake and M. Yokoyama, Anal. Chem., 1998, 70, 4309-
4313.
91. K. Kimura, S. Yajima, K. Okamoto and M. Yokoyama, J. Mater. Chem., 2000, 10, 1819-1824.
92. K. Kimura, H. Takase, S. Yajima and M. Yokoyama, Analyst, 1999, 124, 517-520.
93. S. Yajima, S. Uchida, Y. Hori and K. Kimura, Supramol. Chem., 2010, 22, 420-424.
Craig Alexander
261
94. K. Kimura, S. Yajima, H. Takase, M. Yokoyama and Y. Sakurai, Anal. Chem., 2001, 73, 1605-1609.
95. J. Liu, X. Wu, Z. Zhang, S. Wakida and K. Higashi, Sensor. Actuat., B-Chem, 2000, 66, 216-218.
96. M. Mazloum Ardakani, M. Salavati-Niasari, M. Khayat Kashani and S. M. Ghoreishi, Anal. Bioanal.
Chem., 2004, 378, 1659-1665.
97. M. M. Ardakani, M. K. Kashani, M. Salavati-Niasari and A. A. Ensafi, Sensor. Actuat., B-Chem,
2005, 107, 438-445.
98. M. Arvand, M. Moghimi and M. A. Bagherinia, Electroanal., 2007, 19, 2571-2576.
99. M. A. Zanjanchi, M. Arvand, N. O. Mahmoodi and A. Islamnezhad, Electroanal., 2009, 21, 1816-
1821.
100. A. Islamnezhad and M. Valizadeh, Anal. Bioanal. Electrochem., 2012, 4, 386-398.
101. E. M. G. Santos, A. N. Araújo, C. M. C. M. Couto and M. C. B. S. M. Montenegro, Anal. Bioanal.
Chem., 2006, 384, 867-875.
102. N. Zárate, A. N. Araujo, M. Conceiçao, B. S. M. Montenegro and R. Pérez-Olmos, Am. J. Anal.
Chem., 2011, 2, 491-499.
103. A. Jiménez-Morales, J. C. Galván and P. Aranda, Electrochim. Acta, 2002, 47, 2281-2287.
104. W. Kim, S. Chung, S. B. Park, S. C. Lee, C. Kim and D. D. Sung, Anal. Chem., 1997, 69, 95-98.
105. W. Kim, D. D. Sung, G. S. Cha and S. B. Park, Analyst, 1998, 123, 379-382.
106. B. H. Lee, Y. B. Shim and S. B. Park, Anal. Chem., 2004, 76, 6150-6155.
107. H. Zhou, M. E. Meyerhoff, K.-S. Bi and S. B. Park, J. Korean Electrochem. Soc., 2009, 12, 335-
341.
108. A. Demirel, A. Doğan, E. Canel, S. Memon, M. Yilmaz and E. Kili, Talanta, 2004, 62, 123-129.
109. P. Schulthess, Y. Shijo, H. V. Pham, E. Pretsch, D. Ammann and W. Simon, Anal. Chim. Acta,
1981, 131, 111-116.
110. H.-L. Wu and R.-Q. Yu, Talanta, 1987, 34, 577-579.
111. V. V. Egorov and Y. F. Lushchik, Talanta, 1990, 37, 461-469.
112. R. Yuan, Y.-Q. Chai and R.-Q. Yu, Analyst, 1992, 117, 1891.
113. V. V. Cosofret, T. M. Nahir, E. Lindner and R. P. Buck, J. Electroanal. Chem., 1992, 327, 137-146.
114. E. Lindner, V. V. Cosofret, R. P. Kusy, R. P. Buck, T. Rosatzin, U. Schaller, W. Simon, J. Jeney, K.
Toth and E. Pungor, Talanta, 1993, 40, 957-967.
115. D.-H. Cho, K.-C. Chung and M.-Y. Park, Talanta, 1998, 47, 815-821.
116. V. V. Egorov and Y. V. Sin'kevich, Talanta, 1999, 48, 23-38.
117. X. J. Liu, B. Peng, F. Liu and Y. Qin, Sensor. Actuat., B-Chem, 2007, 125, 656-663.
118. D.-H. Cho, K.-C. Chung, S.-S. Jeong and M.-Y. Park, Talanta, 2000, 51, 761-767.
119. T. Inoue, T. Baba and A. Yuchi, Electroanal., 2011, 23, 536-542.
Craig Alexander
262
120. J. Georges, Clin. Chem., 1979, 25, 1888-1890.
121. M. S. Ghauri and J. D. R. Thomas, Analyst, 1994, 119, 2323.
122. J. Chin, J. Oh, S. Y. Jon, S. H. Park, C. Walsdorff, B. Stranix, A. Ghoussoub, S. J. Lee, H. J. Chung,
S. M. Park and K. Kim, J. Am. Chem. Soc., 2002, 124, 5374-5379.
123. K. Suzuki, D. Siswanta, T. Otsuka, T. Amano, T. Ikeda, H. Hisamoto, R. Yoshihara and S. Ohba,
Anal. Chem., 2000, 72, 2200-2205.
124. S. Sasaki, T. Amano, G. Monma, T. Otsuka, N. Iwasawa, D. Citterio, H. Hisamoto and K. Suzuki,
Anal. Chem., 2002, 74, 4845-4848.
125. H.-S. Kim, H. Y. Park, H. J. Oh, K. K. Young, J.-H. Choi, D.-H. Lee, G. S. Cha and H. Nam, Anal.
Chem., 2000, 72, 4683-4688.
126. H.-S. Kim, D.-H. Kim, K. S. Kim, J.-H. Choi, J. H. Shim, I. S. Jeong, G. S. Cha and H. Nam, J.
Inclusion Phenom. Macrocyclic Chem., 2003, 46, 201-205.
127. H. L. Jin, T. H. Kim, J. Kim, S. S. Lee and J. S. Kim, Bull. Korean Chem. Soc., 2004, 25, 59-62.
128. S. Chandra, R. Buschbeck and H. Lang, Talanta, 2006, 70, 1087-1093.
129. P. D. Beer and P. A. Gale, Angew. Chem., Int. Ed., 2001, 40, 486-516.
130. D. Gao, J. Gu, R.-Q. Yu and G.-D. Zheng, Analyst, 1995, 120, 499.
131. P. Schulthess, D. Ammann, B. Krautler, C. Caderas, R. Stepanek and W. Simon, Anal. Chem., 1985,
57, 1397-1401.
132. R. Stepánek, B. Kräutler, P. Schulthess, B. Lindemann, D. Ammann and W. Simon, Anal. Chim.
Acta, 1986, 182, 83-90.
133. U. Schaller, E. Bakker, U. E. Spichiger and E. Pretsch, Talanta, 1994, 41, 1001-1005.
134. M. Shamsipur, M. Javanbakht, A. R. Hassaninejad, H. Sharghi, M. R. Ganjali and M. F. Mousavi,
Electroanal., 2003, 15, 1251-1259.
135. M. R. Ganjali, M. Rezapour, M. R. Pourjavid and M. Salavati-Niasari, Anal. Sci., 2003, 19, 1127-
1131.
136. M. R. Ganjali, S. Shirvani-Arani, P. Norouzi, M. Rezapour and M. Salavati-Niasari, Microchim.
Acta, 2004, 146, 35-41.
137. W. Wroblewski, Z. Brzozka, D. M. Rudkevich and D. N. Reinhoudt, Sensor. Actuat., B-Chem.,
1996, 37, 151-155.
138. M. Pietrzak and M. E. Meyerhoff, Anal. Chem., 2009, 81, 3637-3644.
139. T. R. L. C. Paixão, J. L. Cardoso and M. Bertotti, Talanta, 2007, 71, 186-191.
140. M. G. Mitrakas, C. A. Alexiades and V. Z. Keramidas, Analyst, 1991, 116, 361.
141. L. Ebdon, J. Braven and N. C. Frampton, Analyst, 1991, 116, 1005.
142. L. Campanella, C. Colapicchioni, G. Crescentini, M. P. Sammartino, Y. Su and M. Tomassetti,
Sensor. Actuat., B-Chem., 1995, 26-27, 329-335.
143. P. Sutton, J. Braven, L. Ebdon and D. Scholefield, Analyst, 1999, 124, 877-882.
Craig Alexander
263
144. A. Asghari, Talanta, 2003, 61, 557-563.
145. A. Asghari, M. K. Amini, H. R. Mansour, M. Salavati-Niasari and M. Rajabi, Anal. Sci., 2003, 19,
1121-1125.
146. J. Ortuno, R. Exposito, C. Sanchezpedreno, M. Albero and A. Espinosa, Anal. Chim. Acta, 2004,
525, 231-237.
147. M. M. Ardakani, M. Salavati-Niasari and M. Jamshidpoor, Sensor. Actuat., B-Chem., 2004, 101,
302-307.
148. H. K. Lee, K. Song, H. R. Seo and S. Jeon, Talanta, 2004, 62, 293-297.
149. P. Kong Thoo Lin, A. N. Araujo, M. C. B. S. M. Montenegro and R. Perez-Olmos, J. Food Chem.,
2005, 53, 211-215.
150. R. Mahajan, R. Kaur, H. Miyake and H. Tsukube, Anal. Chim. Acta, 2007, 584, 89-94.
151. B. Schazmann and D. Diamond, New J. Chem., 2007, 31, 587.
152. A. Watts, V. Gavalas, A. Cammers, P. Andrada, M. Alajarin and L. Bachas, Sensor. Actuat., B-
Chem., 2007, 121, 200-207.
153. Selectophore® Sensoric Applications - Available at http://www.sigmaaldrich.com/analytical-
chromatography/analytical-reagents/sensoric-applications.html (accessed: 23/01/2014)
154. R. V. Harrison, Chemical Sensors: Properties, Performance and Applications, Nova Science
Publishers, Inc., New York, 2010.
155. Cell Constant of Interdigitated Electrodes. Available at http://www.mosaic-
industries.com/embedded-systems/instrumentation/conductivity-meter/microfabricated-planar-
interdigitated-electrodes-cell-constant (accessed: 07/02/2014)
156. A. A. Shul'ga, B. Ahlers and K. Cammann, J. Electroanal. Chem., 1995, 395, 305-308.
157. A. Bratov, N. Abramova, J. Ramón-Azcón, A. Merlos, F. Sánchez-Baeza, M. P. Marco and C.
Domínguez, Electrochem. Commun., 2008, 10, 1621-1624.
158. A. V. Mamishev, A. Sundara-Rajan, F. Yang, Y. Du and M. Zahn, IEEE Sens. J., 2004, 92, 808-
845.
159. N. Jaffrezic-Renault and S. V. Dzyadevych, Sensors, 2008, 8, 2569-2588.
160. K. Cammann, B. Ahlers, D. Henn, C. Dumschat and A. A. Shul'ga, Sensor. Actuat., B-Chem., 1996,
35-36, 26-31.
161. B. Khadro, P. Namour, F. Bessueille, D. Leonard and N. Jaffrezic-Renault, Int. J. Environ. Anal.
Chem., 2009, 89, 11-19.
162. I. D. Nickson, C. Boxall and S. N. Port, IOP C. Ser. Mat. Sci. Eng., 2010, 9, 012044.
163. H. Suzuki, Electroanal., 2000, 12, 703-715.
164. P. Van Gerwen, W. Laureyn, W. Laureys, G. Huyberechts, M. Op De Beeck, K. Baert, J. Suls, W.
Sansen, P. Jacobs, L. Hormans and R. Mertens, Sensor. Actuat., B-Chem., 1998, 49, 73-80.
165. S. Laschi and M. Mascini, Med. Eng. Phys., 2006, 28, 934-943.
Craig Alexander
264
166. K. Norrman, A. Ghanbari-Siahkali and N. B. Larsen, Annu. Rep. Prog. Chem., Sect. C, 2005, 101,
174-201.
167. C. J. Brinker, G. C. Frye, A. J. Hurd and C. S. Ashley, Thin Solid Films, 1991, 201, 97-108.
168. Micro10x™ Robotic Dispenser - Available from http://www.hudsonrobotics.com/products/liquid-
handling/micro10x/ (accessed: 12/09/2014)
169. S. Ostling and P. Virtama, Acta phys. Scandinav., 1946, 11, 289-293.
170. Megaposit SRP220 Material Safety Data Sheet - http://kni.caltech.edu/facilities/msds/SPR220.pdf
(accessed: 10/04/2014)
171. Megaposit MF-26A Material Safety Data Sheet -
http://sindhu.ece.iisc.ernet.in/nanofab/twikii/pub/Main/LaserWriter,Microtech/MF-
26A_Developer__msds.pdf (accessed: 10/04/2014)
172. Megaposit SPR-220 Series Photoresists Data Sheet -
http://microchem.com/PDFs_Dow/SPR%20220%20DATA%20SHEET%20R%26H.pdf (accessed:
06/08/2014)
173. Micrux Technologies - Thin-film interdigitated array microelectrodes -
http://www.micruxfluidic.com/archivos/MicruX_thin-film_IDA1_electrodes.pdf (accessed: 06/08/2014)
174. M. S. Ghauri and J. D. R. Thomas, Anal. Proc., 1994, 31, 181-183.
175. Chemat Scientific Dip Coating. Available at -
http://www.chemat.com/chematscientific/DipCoater.aspx (accessed: 01/09/2014)
176. U. Trebbe, M. Niggemann, K. Cammann, G. Fiaccabrino, M. Koudelka-Hep, S. Dzyadevich and O.
Shulga, Fresenius. J. Anal. Chem., 2001, 371, 734-739.
177. Filmetrics Thin-Film Thickness Measurement. Available at - http://www.filmetrics.com/ (accessed:
12/09/2014)
Craig Alexander
265
10 Appendices
10.1 Molar Equivalent Concentrations of Ions
Ion Concentration (ppm) Concentration (mol/l)
Recommended upper limits for freshwater aquaria
Ammonia (NH3) 0.02 1.18 x 10-6
Nitrite (NO2-) 0.2 4.35 x 10
-6
Nitrate (NO3-) 50 8.0 x 10
-4
Phosphate (PO43-
) 0.05 5.27 x 10-7
Calcium carbonate
(CaCO3) 100–300 1.0 x 10
-3 – 3.0 x 10
-3
Stock solutions
Ammonium (NH4+) 10,000 0.56
Nitrite (NO2-) 10,000 0.22
Nitrate (NO3-) 10,000 0.16
Chloride (Cl-) 10,000 0.28
Potassium (K+) 10,000 0.26
Sodium (Na+) 10,000 0.44
Magnesium (Mg2+
) 10,000 0.41
Calcium (Ca2+
) 10,000 0.25
Table 10.1: Concentration values of species stated in ppm throughout the thesis converted to the
molar equivalent. The upper limits of the key water quality factors are shown, along with the stock
solutions that were prepared of each target and interfering ion
Craig Alexander
266
10.2 Details of Commercial Ionophores Used C
hem
ical
stru
ctu
re
IUP
AC
Nam
e
Tet
radodec
yla
mm
oniu
m n
itra
te
9,1
1,2
0,2
2-T
etra
hydro
tetr
aben
zo[d
,f,k
,m][
1,3
,8,1
0]
tetr
aaza
cycl
ote
trad
ecin
e-10,2
1-d
ithio
ne
Nonac
tin
N,N
-Dio
ctad
ecylm
ethyla
min
e
Cal
ix[4
]-az
a-cr
ow
n
Cyan
oaq
ua-
cobyri
nic
aci
d-h
epta
kis
-(2
-
phen
yle
thyle
ster
)
Com
mercia
l N
am
e
of
Ion
op
hore
N/A
Nit
rate
Ionophore
V
Am
moniu
m I
onophore
I
Hydro
gen
Ionophore
III
Hydro
gen
Ionophore
V
Nit
rite
Ionophore
I
Target
Ion
NO
3-
NO
3-
NH
4+
H+
H+
NO
2-
Table 10.2: Details of the commercial ionophores used in the experimental work described
throughout this thesis. The commercial name of each ionophore is shown, along with its
corresponding target ion, IUPAC name and chemical structure
Craig Alexander
267
10.3 Potentiometric Data Acquisition System
A system was designed to obtain data from potentiometric ISEs. Sixteen PHTX-22
pH/ORP preamplifiers (Omega, Manchester, UK) were arranged in two banks of eight.
Each bank was connected to a connector block (BNC-2110, National Instruments,
Austin, TX, USA), with 1 BNC lead per amplifier, and then to a 16-bit National
Instruments data acquisition card using a shielded cable (SHC68-68-EPN). For
availability reasons two different data acquisition cards were used: a PCIe-6320
and a PCI-6221.
Each 16-bit data acquisition card was operated in referenced single-ended mode, that is,
the measurement was taken between system ground and the preamplifier output voltage.
The reference electrodes were attached to system ground via the BNC connector blocks;
a double banana socket was inserted into each connector block. Therefore, four separate
reference electrodes could be connected, allowing for measurements to be taken from
up to four test samples simultaneously.
Initially, all 16 of the preamplifiers were powered using a linear lab supply, set to
provide 0 and 10 VDC. Unfortunately it was discovered that the amplifier units are
designed to drive the reference electrode to approximately 2.5 V above their negative
supply, which meant that experimentally obtained potentials lower than the reference
potential could not be measured with the reference electrode grounded. This is
illustrated in Figure 10.1, with a calibration graph from a pH electrode, with the
obtained measurements ‘clipping’ near 0 V.
Craig Alexander
268
Figure 10.1: Calibration graph from a pH electrode obtained using a PHTX-22 pH/ORP
preamplifier which was powered using a linear lab supply at 10 V
The offset is amplifier dependent, so groups of amplifiers sharing the same power
supply all have different reference electrode potentials, and hence only one working
electrode and one reference can be used in each sample solution. To use the array as
intended, all the reference voltages needed to be connected together, so a floating power
supply was needed for each preamplifier.
The first attempt to provide a floating supply for each amplifier was to connect two 9 V
batteries to each one; however, the power consumption of the amplifiers was too great
and the batteries required replacement with impractical frequency. A circuit board was
constructed with sixteen isolated DC/DC converters (NMR101C, Murata Power
Solutions, MA, USA) and output filters (a high-frequency inductor, IM02EB470K,
Vishay Dale, Selb, Germany) to supply each amplifier separately.
Temperature measurements were made using four negative temperature coefficient
(NTC) thermistors (B57831M871A3, EPCOS, Munich, Germany), which were
Craig Alexander
269
purchased from Farnell. These were wired in series with an 887 ohm 0.1% resistor and
each circuit was excited with a 5 V supply from a third National Instruments data
acquisition card (12-bit, PCI-6024E), which was also used to measure the voltage across
the thermistors.
The thermistors were calibrated by measuring the resistance at a series of known
temperatures between 25 °C and 50 °C, and fitted to the β parameter equation, shown in
Equation 10.1, where R0 is the resistance at T0 (25 °C).
R = R0 e−β(
1T0
− 1T)
Equation 10.1
The three data acquisition cards were synchronized using the National Instruments
Real-Time System Integration (RTSI) bus, with signals provided by the PCI-6024E
(M Series Synchronization with LabVIEW and NI-DAQmx, National Instruments
(http:// http://www.ni.com/white-paper/3615/en/). The data were acquired from the
cards (two cards for potential measurements and one for temperatures) and saved to a
table in a mySQL database (Oracle Corp., Redwood, CA, USA), under the control of a
program written using LabVIEW (National Instruments). A further program was written
to retrieve, display, and save the data in individual files.
Although the system worked much better to obtain potentiometric data from ISEs once
each amplifiers was provided with its own floating power supply, there were several
issues with obtaining the data from the database, and several times analytical data was
lost or unable to be recalled. Numerous attempts were made to overcome these issues
but the system proved very problematic and due to time constraints, a decision was
made to move to a commercial instrument, and an ELIT 4-channel ion analyser
was purchased.
Craig Alexander
270
Had more time been available to fully optimise the bespoke 16-channel system, it would
have been very beneficial to use it for this project. It was a much more cost-effective
way of producing a multi-channel analysing system than purchasing a commercially
available one. As all the results from the system were outputted to a database, this
would have made it very easy to store and quickly locate large amounts of data. This
system would also have been better suited to long-term monitoring, as all of the data
points between two time-stamps could have been recalled, which would have been
particularly beneficial for deducing the longevity of potentiometric sensors and also for
testing them alongside other sensors in a test sample, such as an aquarium, to
approximate the target ion concentration.