NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USA Telephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected] MAPPING H64 Package Updated September 22, 2016 R4 G 34 42 43 41 44 40 45 35 33 39 46 38 47 37 48 36 17 29 18 28 19 27 32 26 20 30 21 25 22 24 31 23 G R3 R1 G 64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49 15 16 13 14 11 12 9 10 7 8 5 6 3 4 1 2 G R2 NEURONEXUS H64 Connector SPECIFICATIONS Connector Mating Connector (2x) Omnetics 36 Position Dual Row Male Nano Miniature Connector (4 guideposts) (2x) Omnetics 36 Position Dual Row Female Nano Miniature Connector (4 guideposts) G = Ground R = Reference p.1

NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

R4

G

34

42

43

41

44

40

45

35

33

39

46

38

47

37

48

36

17

29

18

28

19

27

32

26

20

30

21

25

22

24

31

23

G

R3

R1

G

64

63

62

61

60

59

58

57

56

55

54

53

52

51

50

49

15

16

13

14

11

12

9

10

7

8

5

6

3

4

1

2

G

R2

NEURONEXUS

H64 Connector

SPECIFICATIONS

Connector

Mating Connector

(2x) Omnetics 36 Position Dual Row Male Nano Miniature Connector (4 guideposts)

(2x) Omnetics 36 Position Dual Row Female Nano Miniature Connector (4 guideposts)

G = GroundR = Reference

p.1

Page 2: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

Reference Channel Configuration (Gen. 4)

To use the Probe Reference site, follow these instructions: 1. Do not cut the wire loop jumper.2. Cut both Wire 1 and Wire 2. This prevents ground loops from forming.3. Reference channels R1 and R2 connect to the Probe Reference site. Reference channels R3 and

R4 can be connected to an external reference via Wires 3 and 4.

To disable the Probe Reference site, follow these instructions: 1. Cut the wire loop jumper. Do not cut the reference wires yet. 2. With the wire loop jumper cut, each of the reference wires can be connected to an external reference

source.

The H64 Gen. 4 package has 1 insulated wire loop jumper (see above) and 5 colored insulated wires. The Ground wire is green. Wires 1, 2, 3, and 4 correspond to the Reference pins on the Om-netics connectors (Wire 1 connects to R1, etc.). Please read fully before making your desired changes - it may not be possi-ble to reconnect the wire loops once they have been cut.

Below are 2 possible wiring configurations. You must choose one option and act accordingly or a ground loop may form.

Probe Ref. Jumper

Wire 2

Wire 1Ground

Wire 3

Wire 4

Note: For proper grounding, please use the correct wiring configuration for your probe

p.2

Page 3: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

Reference Channel Configuration (Gen. 3)

Probe Ref. Jumper

Wire 2

Wire 1Ground

To use the Probe Reference site, pursue only one of the following 4 options: a) To feed the Probe Reference into all 4 Reference pins, cut Wires 1 and 4, twist Wires 2 and 3 together, and leave the

jumpers uncut.b) Cut Wires 1 and 2, and leave the jumpers uncut. Channels R1 and R2 serve as the Probe Reference, and 1 wire reference

can be fed to both channels R3 and R4.c) Cut Wires 1, 2, and Wire Reference Jumper 2. Channels R1 and R2 serve as the Probe Reference, and Channels R3 and R4

serve as independent wire references (via Wire 3 and Wire 4, respectively).d) Cut Wire 1 and both Wire Reference Jumpers. Channel R1 serves as the Probe Reference, and Channels R2, R3, and R4 serve

as independent wire references (via Wire 2, Wire 3, and Wire 4, respectively).

The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground wire is designated with black shrink wrap. Wires 1, 2, 3, and 4 correspond to the Reference pins on the Omnetics connectors (Wire 1 connects to R1, etc.). Please read fully before making your desired changes - it may not be possible to reconnect the wire loops once they have been cut.

Below are 7 possible wiring configurations. You must choose one option and act accordingly or a ground loop may form.

To disable the Probe Reference site, cut the Probe Reference jumper, then take only one of the following 3 actions: e) For two wire reference signals, leave the Wire Reference jumpers uncut. Wires 1 and 2 feed into Channels R1 and R2,

and Wires 3 and 4 feed into Channels R3 and R4.f) For three wire reference signals, cut Wire Reference Jumper 1. Wires 1 and 2 (Channels R1 and R2) serve as independent

wire references, and Wires 3 and 4 (Channels R3 and R4) serve as one combined wire reference.g) For four independent wire reference signals, cut all jumpers.

Wire 3

Wire 4

Wire Ref. Jumper 2

Wire Ref. Jumper 1

Note: For proper grounding, please use the correct wiring configuration for your probe

p.3

Page 4: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

Reference Channel Configuration (Gen. 2)

Wire Loop Jumper

To use only the Probe Reference site, follow these instructions: 1. Do not cut the wire loop jumper.2. Cut both reference wires (Wire 1 and Wire 2). This prevents ground loops from forming.3. All the connector reference channels (R1, R2, R3, and R4) connect to the Probe Reference site.

To use an external headstage reference, follow these instructions: 1. Cut the wire loop jumper. Do not cut the reference wires yet. 2. With the wire loop jumper cut, connector channels R1 and R3 now connect to the Probe Reference,

and connector channels R2 and R4 can be connected to an external headstage reference with Wire 2 (closer to connectors).

3. To prevent a ground loop from forming, cut Wire 1 (closer to probe).

Wire 1

Wire 2

The H64 Gen. 2 package has 1 bare wire loop jumper (see above) and 3 bare wires. The Ground wire is designated with black shrink wrap. Wire 1 (closer to probe) is connected to channels R1 and R3. Wire 2 (closer to connectors) is connected to channels R2 and R4. Please read fully before making your desired changes - it may not be possible to reconnect the wire loops once they have been cut.

NeuroNexus recommends taking one of two reference configuration options. You must choose one option (see below) and act accordingly or a ground loop may form.

Ground

Note: For proper grounding, please use the correct wiring configuration for your probe

p.4

Page 5: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

R4

G

34

42

43

41

44

40

45

35

33

39

46

38

47

37

48

36

17

29

18

28

19

27

32

26

20

30

21

25

22

24

31

23

G

R3

R1

G

64

63

62

61

60

59

58

57

56

55

54

53

52

51

50

49

15

16

13

14

11

12

9

10

7

8

5

6

3

4

1

2

G

R2

NEURONEXUS

Reference Channel Configuration (Gen. 1)

Wire 1

Wire 2

IMPORTANT: Check our catalog to see if your probe model has a probe reference (PR) site.If your design has a PR site, and you plan on using it:1. Cut Wire 1 (see above)2. Make sure that the PR site is completely implanted3. Reference channels R2 and R4 (see below) now function as the Probe Reference

If you do not intend to use the PR site, cut Wire 1, connect Wire 2 to your external reference source, and use reference channels R1 and R3 as your external reference channels.

Ground

Note: For proper grounding, please use the correct wiring configuration for your probe

p.5

Page 6: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

2 1511

5

1 1614

3

6 129

8

4 1310

7

18 3127

21

17 3230

19

22 2825

24

20 2926

23

34 4743

37

33 4846

35

38 4441

40

36 4542

39

50 6359

53

49 6462

51

54 6057

56

52 6158

55

A4x4-tet p.6

Page 7: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

98

107

116

125

134

143

152

161

25242623272228212920301931183217

41404239433844374536463547344833

57565855595460536152625163506449

A4x16

Probe Reference

p.7

Page 8: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

5

4

6

3

7

2

8

1

13

12

14

11

15

10

16

9

21

20

22

19

23

18

24

17

29

28

30

27

31

26

32

25

37

36

38

35

39

34

40

33

45

44

46

43

47

42

48

41

53

52

54

51

55

50

56

49

61

60

62

59

63

58

64

57

A8x8

ProbeRef.

p.8

Page 9: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

8

7

6

5

4

3

2

1

16

15

14

13

12

11

10

9

24

23

22

21

20

19

18

17

32

31

30

29

28

27

26

25

40

39

38

37

36

35

34

33

48

47

46

45

44

43

42

41

56

55

54

53

52

51

50

49

64

63

62

61

60

59

58

57

A8x8-Edge p.9

Page 10: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

1234

8765

9101112

16151413

17181920

24232221

25262728

32313029

33343536

40393837

41424344

48474645

49505152

56555453

57585960

64636261

Buzsaki64 / Buzsaki64L p.10

Page 11: NEURONEXUSneuronexus.com/wp-content/uploads/2018/07/H64-Maps-20160922.… · The H64 Gen. 3 package has 3 bare wire loop jumpers (see above) and 5 white insulated wires. The Ground

NeuroNexus Technologies, Inc. ©2016 | 655 Fairfield Court, Suite 100, Ann Arbor, Michigan USATelephone: +1.734.913.8858 | Fax: +1.734.786.0069 | [email protected]

MAPPINGH64 Package

Updated September 22, 2016

12345

109876

1112131415

2019181716

2122232425

3029282726

3132333435

4443424140

37

38

39

36

4546474849

5453525150

5556575859

6463626160

Buzsaki64sp / Buzsaki64spL p.11