The LyoS™ Control System FOR LYOSTAR™ 3 LYOPHILIZERS

The LyoS™ Control System

Embed Size (px)

Citation preview

Page 1: The LyoS™ Control System

The LyoS™ Control System

FOR LYOSTAR™ 3 LYOPHILIZERS

Page 2: The LyoS™ Control System

Welcome to LyoS™

Page 3: The LyoS™ Control System

Welcome to LyoS™

• Pronounced \‘lī-,ōs\. • Windows®-based platform. • Leverages GE Proficy HMI/SCADA iFIX. • Provides visualization, data acquisition

and supervisory capabilities directly to an Allen-Bradley CompactLogix PLC.

Page 4: The LyoS™ Control System

Key Features

• Automated Leak and Function Testing • Automatic Backfill / Automatic Stoppering • Barometric Endpoint Testing (PRISE) • Pirani vs. Capacitance Manometer

differential testing (PRCM)

Page 5: The LyoS™ Control System

Sophisticated Instrumentation Set

• Sixteen (16), Type T product temperature-monitoring thermocouples.

• Two (2) Capacitance Manometers (product and condenser chambers).

• Pirani vacuum transducer (product chamber).

Page 6: The LyoS™ Control System

Optional Cycle Optimization Tools

• SMART™ Freeze Dryer Technology (optional).

• ControLyo™ Nucleation On-Demand Technology (optional).

Page 7: The LyoS™ Control System

Fully Validateable

• Fully configurable and fully validatable for regulated pharmaceutical and biotechnology applications.

• SP Scientific offers onsite control system IQ/OQ validation support and a more comprehensive factory-initiated System Integration Testing (SIT) program.

Page 8: The LyoS™ Control System

Documentation and Support

• All original software is provided with the LyoS™ control system.

• A backup copy of the LyoS™ control system is provided.

• PLC source code is available. Requires a signed Confidentiality Disclosure Agreement (CDA).

Page 9: The LyoS™ Control System

Control System Hardware / Software Comparison

Control System Feature LyoStar™ 3 LyoStar™ II

Praxair ControLyo™ Nucleation Technology

Optional Not Available

SMART™ Freeze Dryer Technology

Optional (Available June 2011)

Optional

Programmable Logic Controller (PLC)

Allen-Bradley CompactLogix

Opto 22

Vacuum Sensor Configuration (Standard)

2 Capacitance Manometers 1 Pirani Vacuum Gauge

1 Capacitance Manometers 2 Pirani Vacuum Gauge

Page 10: The LyoS™ Control System

Control System Software Feature Comparison

Control System Feature LyoStar™ 3 LyoStar™ II

PVG/CM Differential Feedback Testing

Standard Offering. Not Available.

Barometric Endpoint / Pressure Rise Testing

Standard Offering. Not Available.

Vacuum Ramping During Primary Drying

Standard Offering. Not Available.

Vacuum Preseal Standard Offering. Not Available.

Product Driven Mode Standard Offering. Not Available.

Vacuum Control Using Secondary Gauge

Standard Offering. Not Available.

Page 11: The LyoS™ Control System

Basic Navigation (Buttons)

The circular icons located at the top middle and top right corner of the screen provide access to additional screens and functions.

Stop

Back

Play

Alarm Summary

Print

Maintenance

Data Export

Trend

Page 12: The LyoS™ Control System

Basic Navigation (Buttons)

• Play. Start or continue an automatic cycle. • Stop. Terminate an automatic cycle. • Back. Return to the previous screen.

Page 13: The LyoS™ Control System

Basic Navigation (Alarm Summary Button)

• The Alarm Summary button provides access to a detailed list of all active alarms.

• The button displays the current number of unacknowledged alarms.

• Alarms are automatically saved to the computer’s hard drive.

Page 14: The LyoS™ Control System

The Alarm Summary Page

Page 15: The LyoS™ Control System

The Alarm Summary Page (Alarm Print)

• Select the Alarm Print button to print

saved alarms. • The Alarm Print dialog appears with a list

of dates. • Alarms are printed by date. Dates are

formatted as yy/mm/dd.

Page 16: The LyoS™ Control System

The Alarm Print Dialog

• To print the Alarm Summary, select a date and click Print Alarm.

• To close the Alarm Print dialog, click Cancel.

Page 17: The LyoS™ Control System

Basic Navigation (Maintenance Button)

The Maintenance button opens the Maintenance menu dialog. • Enable Component Mode. • Open the Configuration page. • Open the Calibration page.

Page 18: The LyoS™ Control System

The Maintenance Menu (Component Mode)

• Component Mode allows you to manually control system components listed on the Synoptic page such as the heater, fluid pump, vacuum pump, compressors, and associated valves.

• Click the Component button to enable. The Synoptic page will open.

Page 19: The LyoS™ Control System

The Maintenance Menu (Configuration)

• Select the Configuration button to open the Configuration page.

• Configure settings used by the software to determine how the system will operate, as well react to certain conditions.

• PLC version and system serial number are displayed at the top of the page.

Page 20: The LyoS™ Control System

The Configuration Page

Page 21: The LyoS™ Control System

The Configuration Page (ControLyo™)

• Systems configured with ControLyo™ Nucleation Technology will include additional settings.

Page 22: The LyoS™ Control System

The Maintenance Menu (Calibration)

• Select the Calibration button to open the Calibration page.

• Calibrate the lyophilizer’s temperature sensors (i.e., shelf, condenser, ambient, interstage and product probes).

Page 23: The LyoS™ Control System

The Calibration Page

Page 24: The LyoS™ Control System

Basic Navigation (Trend Button)

Use the Trend button to launch the Historical Trend screen. • Monitor real-time and historical cycle data. • Customize active charts. • Create new charts.

Page 25: The LyoS™ Control System

The Historical Trend Screen

Page 26: The LyoS™ Control System

Basic Navigation (Data Export Button)

Use the Data Export button to open the Historical Data Export page. • Export historical data into a comma-

separated value (.CSV) file that can be viewed in any spreadsheet application.

Page 27: The LyoS™ Control System

The Historical Export Page

Page 28: The LyoS™ Control System

Basic Navigation (Print Button)

• Print. Prints the current full-screen view to

the available printer.

Print

Page 29: The LyoS™ Control System

Basic Navigation (Cycle Status)

The Cycle Status pull-down is accessible from the top left corner of the screen.

Cycle Status Pull-Down

Page 30: The LyoS™ Control System

Basic Navigation (Cycle Status)

The Cycle Status pull-down displays the real-time status of critical system parameters.

Page 31: The LyoS™ Control System

Basic Navigation (Cycle Status)

Select the Probe Avg button to open the Product Probes pop-up window.

Page 32: The LyoS™ Control System

Basic Navigation (Tabs)

The LyoS™ control system includes a series of navigation tabs located across the bottom of the screen. To navigate to a particular page, simply click the tab of the page you wish to access.

Page 33: The LyoS™ Control System

The Synoptic Page

• A process flow diagram of the complete lyophilization system.

• Individual components may be monitored and/or controlled from this screen.

• Critical system parameters are displayed. • Activate or deactivate individual

components with a click of the mouse.

Page 34: The LyoS™ Control System

The Synoptic Page

Page 35: The LyoS™ Control System

The Synoptic Page (ControLyo™)

• Systems configured with ControLyo™

Technology include additional process flow items.

Page 36: The LyoS™ Control System

The Freeze-Dry Page

• The Freeze-Dry Recipe page allows you to view and edit current recipe parameters, and load a previously saved recipe.

• This page provides access to the Recipe Manager function, which allows you to create and save recipes.

Page 37: The LyoS™ Control System

The Freeze-Dry Page

Page 38: The LyoS™ Control System

The Freeze-Dry Page (ControLyo™)

Systems configured with ControLyo™ Technology include additional fields below the Freezing/Thermal Treatment steps.

Page 39: The LyoS™ Control System

The SMART™ Page

• SMART™ technology utilizes manometric temperature measurement (MTM).

• Automated measurements of the product temperature at the sublimation interface are used to determine ideal parameters for preventing product collapse or meltback during primary drying.

Page 40: The LyoS™ Control System

The SMART™ Page

Coming June 2011.

Page 41: The LyoS™ Control System

The Leak Test Page

• The Leak Test page is intended to be used when verifying the vacuum integrity of the system.

• Leak Test results are stated in terms of the system’s Rate of Rise, as well as its Volume–Based Leak Rate (VBLR).

Page 42: The LyoS™ Control System

The Leak Test Page

Page 43: The LyoS™ Control System

The Function Test Page

• Verify the system’s performance against specifications or previous test data.

• Test results include cold and warm probe test data; shelf, condenser and vacuum pull-down rates; lowest shelf and condenser temps, lowest vacuum temps., and shelf heat-up rate.

Page 44: The LyoS™ Control System

The Function Test Page

Page 45: The LyoS™ Control System

The Semi-Auto Page

• The Semi-Auto page may be accessed to activate or deactivate the semi-auto functions, including Shelf (Shelf Control to setpoint or Heat Only to setpoint), Condenser, Vacuum Control to setpoint, Isolation Valve, Backfill, Vacuum Release and Defrost.

Page 46: The LyoS™ Control System

The Semi-Auto Page

Page 47: The LyoS™ Control System

The Semi-Auto Page (ControLyo™)

• Systems configured with ControLyo™ Nucleation Technology will include the following Semi-Auto functions.

Page 48: The LyoS™ Control System

Additional Layout Items

From left to right, the top of the page layout also includes the following: • SP Scientific logo. • Current security access level. • Question Mark and Information icons. • Date and time.

Page 49: The LyoS™ Control System

Security Access Levels

• The control system contains a multi-level, password-protected security system.

• The standard security access levels are identified as Guest, Operator and Supervisor.

Page 50: The LyoS™ Control System

Question Mark Icon

• The Question Mark icon indicates a Help entry point.

• When selected with the Synoptic page active, illustrated components are tagged with descriptive callouts. This feature may be particularly useful when operating your lyophilizer in Component Mode.

Page 51: The LyoS™ Control System

Question Mark Icon

Descriptive Callout Tags

Question Mark Icon

Page 52: The LyoS™ Control System

Information Icon

Selecting the Information icon opens a window displaying a Vapor Pressure over Ice table.

Page 53: The LyoS™ Control System

Date and Time

The current date and time, according to the Windows® operating system clock, is displayed at the top right corner of the control system layout.