Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
THE OPTICS OF THEESO 3.6 METRE TELESCOPE
R.N. WILSON
This report is pub1ished by the
European Southern Observatory
Te1escope Project Division
c/o CER N
1211 Geneva 23
Switzer1and
, * * *
Copyright © ESO 1974
THE OPTICS OF THE 3.6 m TELESCOPE
by
R.N. Wilson
P R E F ACE
The European Southern Observatory (ESO) is theresult of a scientific collaboration for astronomicalresearch in the Southern Hemisphere between the sixEuropean member countries, Belgium, Denmark, FederalRepublic of Germany, France, Netherlands and Sweden.
Situated 600 km north of Santiago de Chile, theobservatory is located on a mountain named La Silla,2400 m above sea level in the southern tip of the AtacamaDesert.
According to the prescriptions of the ESO Convention, signed in 1962, the principal instrument of theobservatory will be the 3.6 m telescope, currently beingbuilt and due to be commissioned in 1976. This instrumentwill be installed on the highest summit of La Silla.
Designed as a general-purpose instrument, it willbe used for furthering a wide range of research programmesin the visible and infrared part of the spectrum. Researchwith the observatory's existing telescopes has underlinedthe need for such a large instrument.
The ESO Telescope Project Division, establishedon the CERN site in Geneva following an agreement onco-operation reached between the two organizations, isresponsible for the design and construction of theobservatory's newest and largest telescope. This Divisionhas now begun to issue aseries of Technical Reportstreating the various design and construction aspects ofthe different parts of the telescope. The present reportis the second of the series and will be followed in thenear future by others.
S. Laustsen
Leader of TP Division, ESO
Contents
1. Purpose of present report
2. Basic parameters
2.1 Aperture
2.2 Relative apertures at the three foci
2.3 Geometry of coude system
3. Optica1 design
3.1 Cassegrain focus
3.2 Prime focus correctors
3.3 Coude focus
4. Manufacture of optics
1
1
1
2
4
5
5
11
13
13
Primary mirror4.1
4.1.1
4.1. 2
4.1. 3
The blank
Methods of working
Testing
13
13
14
15
4.1.3.1
4.1.3.2
4.1.3.3
4.1.3.4
4.1.3.5
Hartmann tests without a compensation system
Null tests with a compensation system
Wave Shearing Interferometer (WSI) tests
Auxi1iary tests of the te1escope opticsby the method of LYTLE
Conc1usions regarding the primary
16
20
20
21
22
4.2.1
4.2.2
4.2.3
4.2 Secondary mirrors
The blanks
Methods of working
Testing
24
24
24
25
4.3 Coude plane mirrors
4.4 QRC corrector
4.5 Prime focus correctors
Acknow1edgments
References
3030
3131
32
- 1 -
The Optics of the 3.6 m Telescope
1. Purpose of present report
It is the purpose of this report to give a resume
of the astronomical/technical arguments leading to the
choice of the telescope parameters; to discuss in
some detail the optical design considerations which
fixed the final form of the mirrors; and to give an
account of the manufacture and test results with
final values of the parameters. No discussion is
given here of auxiliary optics for acquisition or
guiding: this will be dealt with in aseparate report.
2. Basic parameters
Most modern reflecting telescopes have basic
parameters laid out more or less according to the
recommendations of Bowen (1) with prime focus,
Cassegrain and coude observing stations working at
relative apertures of about f/3, f/8 and f/30
respectively. The ESO 3.6 m telescope has parameters'
in close agreement with these recommendations.
An excellent discussion of the considerations
involved in fixing the basic parameters has been
given by Fehrenbach (2). The main reasons for the
decisions will simply be summarized here:
2.1 Aperture
The original aperture of 3 m was upgraded to
3.5 m to allow sufficient space for the prime focus
cage. Because the blank as finally cast was somewhat
larger, the final free diameter is 3.57 m.
- 2 -
2.2 Relative apertures at the three foci
In the Bowen conception, the most important
observing station is the Cassegrain with a relative
aperture of the order of f/B. An analysis for the
ESO telescope confirmed that the most efficient
relative aperture from the point of view of limiting
magnitude, sky background and exposure time was about
f/G to f/B. Even at f/G, if this relative aperture
were used at the prime focus, the tube length and
dome size would have been prohibitive. Thus the
Cassegrain form, whose basic and most important
feature is its strong telephoto-effect, emerges as
the natural solution for providing a short tube length
combined with the required relative aperture. Por a
practical image position behind the primary and an
acceptable obstruction ratio of the secondary, the
primary mirror would have to be excessively steep to
yield a relative aperture at the Cassegrain focus of
f/G. With f/B at the Cassegrain focus, a relative
aperture of the primary mirror of about f/3 gives an
acceptable obstruction ratio.
The acceptability of an f/3 primary will depend
on the feasibility of its manufacture which will
depend on its form as weIl as its relative aperture;
also on the astronomical usefulness, which in turn
will depend on the field size obtainable with good
optical correction and acceptable complexity of,
correctors). This is discussed in more detail below.
Adequate optical solutions exist but it is neverthe
less debatable whether a prime focus station is
justified in view of the mechanical complications
- 3 -
associated with a prime focus cage. However, as with
all other telescopes of this size, it was decided to
retain the prime focus observing station. Manufacturers
were confident that good quality primaries of this
relative aperture were technically feasible, although
very difficult.
The relative aperture of the coude was fixed at
f/30. The efficiency of a coude spectrograph,
properly dimensioned, only depends on the absolute
aperture of the telescope, not on its relative aperture.
Por a given grating size the longer the focal length
of the telescope, the longer will be the spectrograph
collimator unless a focal reducer is used. Thus,
there is no interest in increasing the focal length
of the telescope at the coude focus further than
necessary. A relative aperture of f/30 gives a
secondary of comparable size to that of the Cassegrain
so that it is conveniently placed in the tube for
switching from one system to the other. In the last
few years, however, evidence has been accumulating
that high-reflectance dielectric mirrors bring such
advantage in light gain that they must be seriously
considered. Since such coatings are at present only
possible on smaller mirrors and only smaller mirrors
are suitable for changing with the spectral region,
an additional coude module working at f/ISO is now
under consideration.
The finished mirrors of the basic f/3, f/B, f/30
layout have, according to Hartmann measurements of the
primary and final calculations of the Cassegrain and
coude systems based on them, relative apertures for
- 4 -
the free diameter of 3,57 m as follows:
Prime focus without corrector:
Cassegrain focus without corrector:
Cassegrain focus with corrector:
Coude focus:
f/3.04f/8.09f/8.4l
f/31.79.
These values have been subsequently confirmed by
auxiliary optical tests.
Thus the relative apertures at the prime and
Cassegrain foci are very close to those originally
prescribed. The reduction of the relative aperture
by about 6% in the coude case is for purely technical
reasons in the manufacture of the secondary. This
increase in focal length can be accommodated in the
present concept of the coude spectrograph.
2.3 Geometry of coude system
The geometry of the coude system is largely
dictated by the limited number of telescope mountings
feasible for a telescope of this size. The types
most frequently considered are the fork type, the
horseshoe yoke type (Palomar) and the mixed fork-horse
shoe type. The latter was chosen for the ESO telescope.
Such a mounting requires three plane mirrors for the
coude system, the last being moveable if different
coude labo~atories are to be fed. (A two plane mirror!
system is in fact possible and has been used in the
2.2 m fork-mounted telescope of the Max-Planck
Institute (3). However, this arrangement has a
moveable first plane mirror, leading to some complica-
- 5 -
tion of the tube structure, and a vertically mounted
spectrograph). Fig. 1 shows schematically the final
geometry of the coude system.
As stated above, a complementary coude system
of the Richardson type, using small dielectric mirrors,
is also being studied.
3. Optical design
3.1 Cassegrain focus
This focus being considered astronomically the
most important, its requirements fixed the form of all
the mirrors. However, the decision has important
implications for the other two foci, particularly for
the prime focus.
The two mirror Cassegrain arrangement achieves
its optimum form as a Ritchey-Chretien aplanatic
telescope. Most other modern telescopes of similar
size are strict Ritchey-Chretiens or minor modifi
cations thereof. The only case against such a
solution is that the prime focus is useless without
a corrector because of the overcorrection of the
R.C. primary, and that the coude focus has consider
ably more field coma than the classical telescope,
which may be a disadvantage in offset-guiding. Apart
from the ·R.C. and classical (paraboloidal primary)
telescopes, the only other candidate of interest isthe Dall-Kirkham telescope using a spherical secondary
with an ellipsoidal primary. The D.-K. telescope
is a very interesting solution if only a single
Cassegrain observing station is required and only
- 6 -
axial imagery, since it is technically much easier
from the point of view of manufacture and testing
of the secondary and is normally much less sensitive
to decentering. The advantages and disadvantages
of the three basic types may be summarized as in
the table over-leaf.
Type
Ritchey-Chretienor Quasi-RitcheyChretien
Classical(paraboloidalprimary)
Dall-Kirkham
- 7 -
Advantages
Optimum performance atCassegrain focus.Simplicity of correctorsfor residual aberrationsof mirror system.
Corrected axial imagein prime focus without corrector. Coudefield virtually freefrom coma. Somewhateasier to manufacturethan R.C.
Ease of manufacture ofone secondary.Intensitivity to decentering.
Disadvantages
Hyperboloidal primarygiving uncorrected axialimage at prime focuswithout corrector. Comain coude field. Moredifficult to manufacturethan classical orDall-Kirkham.
Useful field at primefocus extremely smallat f/3 without corrector.Field corrector at primefocus more difficultthan for R.C. primary.Cassegrain focus suffersfrom significant coma atf/8. Field corrector ismore difficult than forR.C. telescope.
Large amount of coma inCassegrain field.Complexity of correctorrequired which mayeliminate the advantage ofintensitivity tode-centering.
Large amount of coma incoude field.
Ellipsoidal primary givinguncorrected axial image atprime focus withoutcorrector. Undercorrectionof primary is unfavourablefor the provision of fieldcorrectors at prime focus.Only one of the secondariescan be made spherical.
- 8 -
It is clear that the Dall-Kirkham solution is not
suited to a general purpose instrument with three
observing stations.
The only optical dis advantages of the Ritchey
Chretien are its coma in the coude field and the
uncorrected axial image of the primary. There seems
to be no clear evidence that the former is serious in
practice, particularly if automatic guiders are used
which lock on to the centre of gravity of the comatic
image. The latter is hardly a practical objection if
the primary relative aperture of f/3 is retained,
since the field corresponding to 1 arcsec tangential
coma is only 96 arcsecs or 5 mm in diameter. If it is
accepted that this field is too small to be useful
in practice so that a field corrector is anyway
necessary, then the overcorrection of the R.C. primary,
which facilitates this field correction, means that
the R.C. or quasi R.C. solution is optically preferable
for both the Cassegrain and Prime Focus observing
stations.
It was considered that the manufacture of a R.C.
telescope with the above parameters was technically
feasible; the decision was therefore taken to assume
the R.C. solution as the basis for the detailed
design.
A ver~ good account of the basic optical design
consideratfons has been given by Köhler (4) (5)
Further discussions were given by Baranne (6) (7)
and by the present author (8) (9)
- 9 -
The problem of correctors for the primary and
Cassegrain foci of a telescope of this type has been
the subject of extensive research during the last
15 years or so. At the time the basic design of the
Cassegrain system of the ESO telescope was decided,
hardly any of the currently available solutions had
been investigated in detail. The simplest form of
corrector, the field-flattening lens, still leaves
the astigmatism of the Ritchey-Chretien system
uncorrected. Köhler (4) (5) discovered that this can
be corrected as weIl if the negative corrector lens
is moved a short distance away from the image plane.
However, it is very important to make clear that the
retention of the coma correction requires some modi
fication of the form of the mirrors. The axial beam
width is such that the spherical aberration is hardly
affected. Thus, this corrector fulfilled the speci
fication set at the time that correction with the
corrector should be achieved over a field of about
± 0.250 and that axial correction should be maintained
when the corrector is removed. This specification
permitted adeparture from the aplanatic (R.C.) form
of the mirror system, which is thus a quasi R.C.,
not a strict R.C.
The spot-diagrams of Fig. 2 show the performance
of the Cassegrain system with and without corrector,
in the latter case assuming optimally bent plates.
These spot-diagrams are, in fact, calculated for the
final system based on measurements of the finished
primary. Two points are worthy of special note:
- 10 -
- The departure of the quasi R.C. form from the strict
aplanatic R.C. form gives a comatic spot-diagram
at, say, ± 0.200 of about one third of that of the
equivalent classical telescope. In fact, the strict
R.C., because of its uncorrected astigmatism, gives
a spot-diagram almost as big at this field but it
has the advantage of being symmetrieal. Doublet
correctors are now well-known (10) (11) (9) which
give very good results over the small angular fields
required in the Cassegrain foci of such large
telescopes - because of limitations of plate size
while retaining the strict R.C. form of the mirrors.
Given present day knowledge, it may be that astriet
R.C. form would be preferred; however, the spot
diagrams of the ESO telescope without corrector in
the Cassegrain focus show that the correction must
still be rated as good and is much superior to that
of a classical telescope with the same geometry
(see ref. 9, figs. 6,7 and 8).
- The size of the individual spot-diagrams with correc
tor remains extremely small ( 0.3 arcsec ) if
optimum focusing is performed for three regions over
the range 365 nm to 1014 nm. The most serious
error by far of the singlet corrector is its un
corrected chromatic difference of magnification
which amounts to about 2 aresec between 365 nm
and 1014 nm at a field of ± 0.250 and grows linearly
with the field. If the spectral region is divided
up into three regions by plate-filter combinations,
the resolution is thus about 0.7 arcsec over each
region. The disadvantage of the lateral colour
error of the singlet corrector may be offset by
- 11 -
more favourable performance with regard to
ghost images: apart from reflexions from
the photographic plate, the singlet only yields
1 ghost image whereas a doublet yields 6.
A mechanical disadvantage of the singlet
corrector is its short back focal distance of
160 mm, whereas doublet correctors tend to have
larger back focal distances. However, the
problem of fitting in guide probes in the space
available in the ESO solution seems soluble.
Fig. 3 shows the final Optical Data Sheet for the
quasi Ritchey-Chretien focus. This, like the spot
diagrams, was based on available Hartmann measurements
of the final form of the primary mirror. The aspheric
constants in this Data Sheet are expressed in two
different ways, according to the following formulae:
x =2y
2 r+ • • • • • • • • • • • • • • • • • (1)
x =2y
2 r+ 4y + 6
y ...(2).
16 r 5
3.2 Prime focus correctors
The potentia1ities of two basic types of correctors
permitting fie1ds of about 10
diameter were investi
gated by Glatzel and Köhler (4) (5) at Car1 Zeiss,
Oberkochen, and by Baranne at Marseille (6) (7)
The former work was based on a system of three aspheric
plates ana1ysed by Meinel (12) and further developed
- 12 -
by Schulte (13). Baranne's work was based on earlier
work of Paul (14) and used 3 lenses of much smaller
diameter, one containing an aspheric surface. An
exhaustive analysis of prime focus correctors has beencarried out by Wynne (10) (lI) (15) Wynne has
designed 3-lens correctors for the Kitt Peak and
Anglo-Australian telescopes. The above three
corrector types - plate type, Paul - Baranne type and
Wynne types - were compared by the author (9) for the
Max-Planck telescope who came to the conclusion that
the Wynne type is the most favourable from the point
of view of performance, and ease of manufacture.
However, final decisions on a corrector covering this
order of field size have not yet been taken.
Apart from the corrector covering about
10 diameter, two others are envisaged covering a more
modest field with fewer elements.
A doublet corrector combined with the quasi
R.C. is capable of good correction over a field ofoabout + 0.3. The overcorrection of the primary brings
its greatest advantage, however, with the third
corrector envisaged, namely a Gascoigne plate (16)
(17) (18); for this corrector is not possible with
a parabolic primary. The Gascoigne plate can provide
good correction of spherical aberration and coma for
fields of about ± 0.14~ a most attractive field for
use with electronographic cameras. The ESO telescopeI
is particul1arly favourable to this solution because
its primary is somewhat more eccentric than the strict
R.C. form would prescribe. The higher the eccentricity
the larger the corresponding Gascoigne plate and the
- 13 -
better its optical performance.
A subsequent technical report will give detailed
results for the prime focus correctors.
3.3. Coude focus
Fig. 4 shows the Optical Data Sheet of the coude
system, based on the Hartmann measurements of the para
meters of the finished primary.
The uncorrected third order tangential coma is,
of course, a linear function of the field. For a field
of ± 0.050, it amounts to 0.65 arcsec.
4. Manufacture of the optics
4.1 Primary mirror
4.1.1 The Blank
Zero expansion ceramic materials were not
available at the time the blanks were ordered. It was
decided to use the material of lowest expansion
available at that time, namely fused quartz.
Fehrenbach (2) gives quite a full account of the
ordering of the prime mirror blank from the CORNING
Glass Works at its reception by REOSC S.A. in May 1967
who had been awarded the contract for working the
optical elements.
After the preliminary milling work for the central
hole and the production of the necessary radius of
- 14 -
about 21.7 m, smoothing was carried out to produce
a spherical surface. It was decided to polish this
surface provisionally in order to test the material.
This revealed that the surface was not suitable for
polishing because of some 25 000 to 30 000 bubbles
present in or near it. After discussions with
representatives of Corning, it was decided that the
disc should be returned to the U.S.A. to be remelted
and covered with a new layer of quartz. The blank
left France in September 1968. In February 1969,
CORNING stated that the remelted blank contained
3 holes which would affect the optical surface.
Further discussions between the two firms revealed
that only one hole would remain after the preliminary
roughing and that this could be plugged by REOSC
using a special technique. The blank was therefore
provisionally accepted by an ESO-Commission and the
president of REOSC in June 1969.
Work was recommenced at REOSC in September 1969.
The orifice of the hole was milled out and polished
to a hemispherical form. A hemispherical p1ug was
similarly prepared from quartz supplied by Corning
from the same me1t and cemented in with synthetic
resin.
4.1.2 Methods of working
The aspheric surface was produced and checked
to an accur~cy of a tenth of a micron by means of an
electro-mechanical test method developed by REOSC (19)
The amount of deformation of a hyperbolic concave
mirror from the starting sphere will depend on the form
of deformation acceptable to the manufacturer. If a
- 15 -
zero point is accepted at an intermediate zone, the
amount of deformation can be reduced, but most
manufacturers prefer a monotonically increasing
deformation-function from the centre to the edge even
if more material has to be removed. On this basis,
the deformation from the vertex curvature sphere is
0.144 mm, so that the relative accuracy of the electro
mechanical measurements was about 0,069%.
An important feature of the technique of working
aspheric surfaces at REOSC is the use throughout of
full-size tools possessing the necessary flexibility
due to their frame construction in wood. This
technique had been successfully employed on a number
of f/3 mirrors, as weIl as Schmidt plates. The use of
full-size tools clearly has advantages from the point
of view of avoiding zonal errors and ripple if the
asphericity can still be achieved. The asphericity
is produced and controlled by the form and area of
the lapping surface.
An account of the working and testing at REOSC
is given by Bayle (20)
4.1.3 Testing
The testing of a concave mirror is always much
easier than that of a convex mirror, because a concave
surface can always be tested in autocollimation at its
centre of curvature. Because of its aspheric form, the
image has, of course, a large amount of spherical
aberration - about 171,5 mm longitudinal aberration or
a wavefront aberration for "optimum focusing" of about
2320'A (A,= 500 nm) for the free aperture of 3.57 m.
- 16 -
Tests were performed using four basic methods:
4.1.3.1 Hartmann tests without a compensation system
This method has the advantage that no auxi1iary
optics are necessary but the corresponding disadvantage
of not being a "null" method: the test must estab1ish
that the correct amount (within a defined to1erance)
of third order spherica1 aberration is present apart
from the absence of zones. Because of its dis
continuous nature with regard to the aperture function,
it is 1ess suitab1e for checking the smoothness of the
surface.
Fig. 5 is a reproduction of figures from the
REOSC report to ESO on the primary. At the right, the
slopes of the surface are given as a function of
diameter; at the 1ef~ the integration giving the
departure in microns from the "best fit" function.
According to this curve, the maximum wavefront error
(max. to min.) is 0.14 A (A= 500 nm).
The Hartmann analysis of an uncompensated
primary of this eccentricity is a difficu1t technica1
process. Apart from the steepness of the wavefront
aberration curve, the problem of centering is consi
derab1e. A sma11 amount of decentering introduces
coma to an extent which makes precise interpretation
of the Hartmann resu1ts difficu1t.
REOSC came to the conc1usion that the most
probable va1ues for the parameters are:
=
=
21 700 nun
1, 15175.
- 17 -
An independent evaluation of the REOSC Hartmann
plates was carried out at ESO by Behr who pointed out
that the longitudinal aberration deduced from the
Hartmann plates will depend not only on r 1 and E. 1
but also on the distance of the source from the
vertex centre of curvature of the mirror. He con
firmed this by ray-tracing; results further confirmed
by Baranne from ray-tracing and first order theory,
and by the author from ray-tracing and by an analytical
treatment. It is easy to establish the following
general formula, including all orders, for the
longi tudinal aberration A s '2 resul ting from a
shifted source (shift = crs) relative to that,
bS'l,resulting from a source placed at the centre of
curvature.
!:.l s ' 2 = 1 - 2 fO: ~{dS ~2 {OS ~3
+ 4 ~--r- ~ - 8 r-r~+ ••• (3)
This formula gives very good agreement with the
ray-tracing results. As an example for r l = 21 700 mm,
~ = 3.66 m and tfs = 100 nun, A s'l = 178,623 nun and
~ s' 1 - ~s' 2 = 1,624 nun for the (uncorrected) data
supposed at that time. The error is thus almost
1% for d s = 100 nun.
- 18 -
If we restrict the above formula to the third
order term, then by a simple transformation:
~s ' 2ßs' 1
1 - 2 (4 )..........
in which u l and u2
are the ray inclinations to the axis.
Further, to third order accuracy, for the corresponding
wavefront aberrations:
w '2w '1
2
-- ! ::::; l 6s' 2 ••••••••••••••••••••• (5)
Equations (4) and (5) are simply statements of the
well-known fact that the wavefront aberration remains
unchanged by such a source shift to third order
accuracy, but that the longitudinal aberration varies
with the square of the aperture angle.
Using the three parameters r l' E, 1 and cf s, i t is
possible to derive various solutions from the Hartmann
results corresponding to a "best fit" to the slope
height function. Figs. 6 and 7 show several such
possible solutions set up by Behr. However, without
independent knowledge of rl
or el
, it was not possible
to decide definitely between these possibilities.
Behr suggested as the most probable solution:
=
=
21 687,5 rnrn
- 1,1550.
- 19 -
These values were used in the final ealeulations of the
QRC and eoude systems(Figs. 3 and 4). However, the
auxiliary (LYTLE) optieal tests (see paragraph 4.1.3.4)
indieated a slightly greater value for r l , in elose
agreement with the REOSC value of 21 700 mm. Behr's
analysis gives as eorresponding value
EI = - 1,1567, somewhat higher than the REOSC value of
81 = - 1,15175. This diserepaney ean be aeeounted for
by the effeet diseussed above, of the souree position
relative to the vertex eentre of eurvature. This effect
ehanges the REOSC value to Cl = - 1,1583. Thus the
alternative solution of Behr, whieh seems the most probable
in the light of subsequent tests, is:
=
=
21 700 mm
1,1567 ± 0,0005.
It should be noted that the form of Figs. 6 and 7 (Behr)
is in exeellent agreement with Fig. 5 (REOSC).
REOSC also used the Hartmann results to analyse the energy
eontained in a given angular diameter, based on geometrieal
opties without the effeet of diffraetion. The eontraetual
requirement was that 75% of the energy should fall within
a diameter of 0.4 aresee (QRC) and 0.5 aresee (eoude).
Aeeording to the first REOSC report, 97% of the energy
is eontained within a diameter of 0.5 aresee and 77%
within 0.25 aresee. But a subsequent, more refined analysis
by REOSC gives 98.6% within 0.4 aresee and 100% within 0.5
aresee (Fig. 8). They also give values for the r.m.s.
error (~/33) and the Strehl intensity ratio (96%). The
values of this analysis do not take aeeount of residual
astigmatie errors, but the evidenee is that these are
small (see paragraph 4.1.3.4).
4.1.3.2
- 20 -
Null tests with a compensation system
An OFFNER type compensator giving negligible
theoretical residual aberration for the combination
(lateral aberration only l.Spdiameter), was used for
a Null-Test examination of the image of a pinhole
source 0.3 arcsec diameter. The image of the source
was found by REOSC not to exceed 0.4 arcsec diameter
from which it was concluded that the transverse
aberration for optimum focus and optimum adjustment
of the compensator did not exceed 0.1 arcsec.
Fig. 9 shows a FOUCAULTGRAM of the image
of a slit of width 0.1 arcsec. This shows that the
surface is very smooth. Two zones are observable,
the outer one with a diameter of about 2550 mm being
the stronger. This corresponds to the maximum change
of slope of about 0.35 arcsec detected in the
Hartmann measurements. Integration gave ~/36 as
the maximum wavefront error in this region of the
mirror.
It should be emphasized that no clear
information was drawn from this test regarding the
first order (vertex radius) and third order form
(Schwarzschild constant) of the primary. In principle,
such information can be obtained by tight tolerancing
of the radii of the compensator and careful measure
ment of th~ position of the principal compensation
1ens requited to achieve optimum correction.
4.1.3.3 Wave Shearing Interferometer (W.S.I.) Tests
W.S.I. tests of the primary were performed
by G. Monnet using a He-Ne Laser. These tests were
- 21 -
able to confirm the smoothness of the primary mirror
surface. The determination of the asphericity was
in agreement with that deduced from Hartmann measure
ments but the accuracy was at least an order of
magnitude lower.
4.1.3.4 Auxiliary tests of the telescope optics
by the method of LYTLE
The main purpose of these tests was to check
the secondary mirrors, but a subsidiary aim was to
provide evidence of freedom from astigmatism of the
primary and to determine independently its vertex
radius.
The results of the LYTLE tests will be
given in detail in a future Technical Report. So far
as the primary mirror is concerned, the following
results emerged:
The primary is tested in this method in
double pass. The zones shown in the
Foucaultgram were clearly visible but
the very good general smoothness was
confirmed.
In double pass and in combination with
the coude secondary, the astigmatism was
less than 0.1 arcsec.
The value deduced for the vertex radius
of curvature was
= 21 705 mm + 10 mm.
- 22 -
4.1.3.5 Cone1usions regarding the primary
The tests by diverse methods deseribed above
have given very e1ear information regarding the
primary mounted in its ee11 with vertiea1 axis.
All tests have indieated a very high quality
of smoothness.
The information on basie parameters ean be
sUmmed up as folIows:
Method andEvaluation r 1 (mm) Cl
Hartmann 21 700 - 1,15175
- REOSC (ignoring effeetof
souree position)
Hartmann 21 687,5 - 1,1550
- Behr or 21 700 - 1,1567
Lyt1e 21 705
- Wilson
,
I
- 23 -
From this eVidence, it is proposed to assume
for further calculations on prime focus correctors
the follo~ing values for the prime mirror parameters
as the most probable:
= 21 700 mm + 10 mm
= 1,1567 + 0,0005 (for a givenvalue of r l ).
The LYTLE tests did not give explicitly a value
for EI. However, the fact that the aberration was
correctly compensated for the correct positions of
the compensation lenses in the QRC and coude cases
confirms very weIl the values given by Behr.
Small errors in the prime focus correctors
resulting from errors in the mirror parameters are
not in themselves very serious since the correctors
can be adjusted in position relative to the focus
to correct the spherical aberration; if the shift
from the theoretical position is small, the field
correction (coma) will not be appreciably affected.
Nevertheless, it is highly desirable to have values
as accurate as possible for the parameters; for
otherwise the mechanical design of the prime focus
adapter is complicated by the provision of sufficient
reserves for adjustment of the optical elements.
In summary, there is clear evidence that the
primary mirror is extremely smooth and that its
parameters are very close to those given in the
optical data sheets (Figs. 3 and 4). There seems
- 24 -
little doubt, so far as its performance in its cell
with vertical axis is concerned, that the contractual
specification of 75% (geometrical) energy concen
tration within a diameter of 0.4 arcsec in the QRC
focus has been more than fulfilled: the mirror
is probably significantly better than this.
Some test material with horizontal axis is
available and suggests that the quality is not
markedly inferior. However, the technical problems
due to air turbulence are so considerable that
final judgement of the performance of the mirror and
cell with non-vertical axis will have to be reserved
until the telescope is in operation.
The primary mirror was formally accepted by
ESO in February 1972.
4.2 Secondary mirrors
4.2.1 The blanks
It was decided to use low expansion material
also for the secondary mirrors. The blanks are thus
also of fused quartz, supplied by HERAEUS.
4.2.2 Methods of working
Similar methods of working were employed at
REOSC for the secondaries as for the primary.
Although these mirrors are much smaller, the diffi
culty of manufacture is at least as great,
probably greater, than that of the primary. The
difficulty is due partly to the much higher
eccentricity of the secondaries (Schwarzschild
- 25 -
constants E1 - 6.91 for the QRC and - 2.18 for
the coude), and partly to the convex form, which
is fundamentally both more difficult to test and
also to aspherize. A monotonie aspherization
function is not possible.on a convex hyperbolic
surface. The maximum deformation from the sphere
of departure is about 22jV in the QRC case,
161' in the coude case.
4.2.3 Testing
The testing of the convex secondaries is one
of the major problems in the manufacture of modern
reflecting telescopes. REOSC had proposed and
used (21) during manufacture a method of great
elegance and offering many advantages, namely the
double pentaprism method. The use of a moveable
pentaprism for testing a telescope was probably
first suggested by Wetthauer and Brodhun in1920 (22)
Methods for testing convex secondaries will be
analysed in a future ESO Technical Report. The
advantages of the pentaprism method derive basically
from the fact that it is a functional test,
simulating the performance of the primary and
secondary mirrors in the actual telescope. This is
a very important advantage, for it means there are
only two mechanical tolerances to be respected and
these are easily held: the source position relative
to the vertex of the primary and the separation of
the mirrors. Any error in the parameters of the
primary is automatically compensated by the secondary.
- 26 -
Thus the image, if placed at the source position
will automatically be free from spherical aberration.
The limits to the mechanical tolerances mentioned
above are thus set by the introduction of coma into
the QRC image when using the corrector and by the
mechanical possibilities for shifting the final
image from its theoretical position. The following
tolerances were established on this basis:
Source position: QRC: 0 mm
+ 50 mm
Coude: + 390 mm.
This tolerance is set by spherical aberration intro
duced by focusing with the secondary mirror to
shift the final image. The effect on coma is
negligible.
Mirror separation: + 28.7 mm introduces 0.1 arcsec
of coma at + 0.250 of the QRC
field. Since this is technically
very easy, a tolerance of
+ 10 mm was prescribed.
Although errors in the parameters of the primary
will not introduce spherical aberration, they will
neverthe~ess change the coma and, to a lesser exten~
the astigmatism, at each of the three foci. This
does not matter for the coude focus (coma fully
uncorrected) or the QRC focus without corrector
(residual coma small but not negligible), but it
will affect the performance of the QRC focus with
- 27 -
corrector and the prime focus correctors since
these correctors are calculated for assumed mirror
parameters.
REOSC presented a complete analysis of the
pentaprism test results. Figs. 10 and 11 show the
slope-height functions of the wavefront emerging
from the QRC combination for field heights 0 mm,
105 mm and 140 mm. Figs. 12 and 13 show the
geometrical energy concentrations calculated from
these values for a mean of three meridians.
These values are weIl within the contractual require
ment of 75% to be included within a diameter of
0.4 arcsec.
In the coude case the field supplement to the
diameter is small. Figs. 14 and 15 show the slope
height function and the geometrical energy
concentration. Again the concentration is weIl
within the contractual tolerance of 75% within a
diameter of 0.5 arcsec.
Although the double-pentaprism test procedure
has real advantages from the point of view of
ensuring freedom from spherical aberration in the
secondary images, it suffers from certain weaknesses:
- Examination of the whole pupil at one time
is not possible, so that detection of
errors of high spatial frequenc~ such as
ripple)is very difficult. Furthermore,
it is difficult to make measurements
corresponding to the edge of the free
aperture of the primary.
- 28 -
- Sinee the measurement over eaeh meridian is
normally referred to its own optimum referenee
sphere, deteetion of astigmatism is not easy.
The preeision of measurement estimated by
Espiard and Favre (21) is about + 10-6 rad
or + 0.2 aresee. While it is not easy to
aehieve greater preeision with other methods,
this uneertainty nevertheless makes an
independent test desirable.
- The aetual position of the eoude foeus is
many meters behind the primary, a position
whieh is not aeeessible for the souree for
tests with a vertieal axis. REOSC overeame
this diffieulty by using a third, eoneave
mirror plaeed in front of the primary as a
foeal redueer. This mirror is theoretieally
a hyperboloid of moderate eeeentrieity.
Although theoretieally the use of the third
mirror is of no eonsequenee, it nevertheless
introduees a third element with inevitable
additional sourees of error.
- The exeess diameter of the QRC seeondary
neeessary for the field ean only be tested
in eonjunetion with the QRC eorreetor using
an off-axis souree and tilted pentaprism.I
This eomplieation may also introduee errors;
on the other hand, the eorreetor is tested
in funetion with the mirrors.
- 29 -
For these reasons, it was decided that auxiliary
optical tests would be valuable. The method
chosen should complement the pentaprism method,
its prime aim being to provide a test of
smoothness. If possible, it should also give
information regarding astigmatism and the
parameters of the primary mirror. Following an
analysis (to be given in a future technical report),
the choice fell on the method proposed by J.D.
Lytle (23). A full account of these tests is given
in aseparate report. Only the principal results
will be quoted here. The tests were successful
in all three respects:
The coude secondary is about as smooth
as the primary; the QRC secondary is
even better. The principal zone on the
primary appeared doubled in height in
the LYTLE test and gave an excellent
comparison measure. Slight zones could
be identified on the secondaries with
reference to the slope-height curves of
Figs. 11, 12 and 15, also the turned up
edge in the QRC case.
The coude LYTLE combination indicated a
residual astigmatism less than 0.1 arcsec.
The QRC LYTLE test was less conclusive
in this respect but could detect no
astigmatism attributable to the primary
or secondary.
- 30 -
- A value for the ver tex radius or curvature
of the primary of 21 705 ± 10 mm was deduced.
This gave excellent confirmation of the value
deduced from the Hartmann tests.
In view of this very satisfactory confirmation
of the pentaprism test results, the secondary mirrors
were formally accepted in December 1973.
4.3 Coude plane mirrors
The coude plane mirrors are not yet completely
finished. Testing has been performed in the usual
way by the Common-Ritchey method of insertion in
the beam illuminating a high quality concave
spherical mirror from its centre of curvature.
Because of turbulence, a horizontal arrangement does
not allow a sufficiently accurate assessment of
flatness. Hence, the mirrors are being tested in
the final stages in the vertical test tower of
the primary, the spherical mirror being placed near
the side of the tower with axis horizontal and
the plane mirror under test deflecting the axis
vertically.
It is expected that the plane mirrors will be
accepted by the Spring of 1974.
4.4. QRC corrector
This singlet negative corrector of free diameter
310 mm is a relatively simple element to manufacture.
The material is UBK7. Being near the image plane,
- 31 -
the tolerances regarding form and homogeneity are
simple to meet. Acceptance, which will take place
shortly, will be based on test plate results, on
Foucault tests of the front (concave) surface
and on the combined pentaprism test with the primary
and secondary mirrors, described above.
4.5 Prime focus correctors
None of these is finally designed at this stage.
The manufacture will be the subject of new contracts.
Acknowledgrnents
The author would like to express his sincere
gratitude to the firm of REOSC for their excellent
cooperation and generous provision of material; to the
CARL ZEISS foundation, where many of the calculations were
performed, for the spot diagrams reproduced in this
report; and to his colleague, Professor A. Behr, for his
collaboration, particularly with regard to the evaluation
of the Hartmann tests of the primary.
References
- 32 -
1.
2.
I.S. Bowen:
Ch. Fehrenbach:
Publ. Astron. Soc. Pacific 11,114 (1961)
Proceedings of ESO/CERN Conferenceon "Large Telescope Design",1971, p. 99
3. K. Bahner:
4. H. Köhler:
5. H. Köhler:
6. A. Baranne:
7. A. Baranne:
Sterne u. Weltraum 12 (4),103 (1973)
ESO-Bulletin No. 2 (1967)
Appl. Opt. 2, 245 (1968)
Publ. de l'Observatoire de HauteProvence 8 (22), 75
In "The Construction of LargeTelescopes" Proc. lAU 27,Academic Press, 22 (1966)
8.
9.
R.N. Wilson:
R.N. Wilson:
Appl. Opt. 2, 253 (1968)
Proceedings of ESO/CERN Conferenceon "Large Telescope Design",1971, p.131
10. C.G. Wynne: Appl. Opt. !, 1185 (1965)
11. C.G. Wynne: Astrophys. J. 152, 675 (1968)
12. A.B. MeineI: Astrophys. J. 118, 335 (1953)
13. D.H. Schulte: Appl. Opt. 2" 313 (1966)
14. M. Paul: Rev. d'Opt. 14, 169 (1935)
15. C.G. Wynne: "progress in Optics", Vol. 10,139 (1972)
II
16. S.C.B. Gascoigne: Observatory 85, 79 (1965)
17. S.C.B. Gascoigne: Quart. J. Roy, Astron. Soc. 2.,98 (1968)
18. S.C.B. Gascoigne: Appl. Opt. 12, 1419 (1973)
19. J. Espiard:
20. A. Bayle:
21. J. Espiard andB. Favre:
22. Wetthauer andBrodhun:
23. J.D. Lytle:
- 33 -
Proceedings of ESO/CERN Conferenceon "Large Telescope Design", 1971,p. 219
Ref. (19) p. 229
Nouv. Rev. d'Opt. 1, 395 (1970)
zeitsehr. f. Instrumentenkunde40, 96 (1920)
Appl. Opt. 9, 2497 (1970).
7664,4
1239
15000
Tube Axis
Polar Axis
//J
~p~\,.':>
MI ////
/_7-
//'
//
M4
M3 .-1-M2
oU1N~
ooM~
-'s 4250 + 15604 + 7000 - 1239 25615 mm
FIG. 1 Geometry of Coude System (schematic)
o,... ()
u0
CUCD C(IO '0 () @ III
UIn 1-4tI") ttl::- Ln.
0
11
~r--
0- () (] G) 0) 0ID11
::r 1-4In CU
+ICUij....
pO
CU.-t
0U1-4
tI") 0-, 0 0 0 0 ....ID UInID
oo::r --o- o o 0- 0 0 - 00
~ ~ ~0 .-t 0 0 0Ln . .-t . "It \0N .-t N r-- .-t . III. M . 0 . Ln ....0 .-t 0 .-t 0 r-- ~
+1 +1 +1 +1 .::+1 +1
~0 Ln
Q'I .N .-t. Ln0 .-t
+1 +1
Fig. 2a(i) Spot diagrams showing the theoretical image quality ofthe QRC system~ corrector (flat field). Optimumfocus for green (approx. best mean focus).
------------_._----_._-_. ,..----_.__.._-_ ... - ..... -..__......_._ ..- .__.._ ..._._-----
: ----i1~:(:{{W 0(T)
- C:(({H 0 o
0
C:{(Wü CiifiiiÜ (((iO C[),.... 0- ~&\\\::r0
.~:~.
::r
UaJtIlU1-/n:l
0 fffiO C!O CDCD , 'j CID 0 lt'l
Ln ~;\\\::.
(T)11,' 0
::r
~,....
If0 (I' ~
1-/- , i et (j) aJca : 4J,,'::r aJLn ~
•.-1'0aJr-iU1-/
•.-1C)
0 an(T) Clu Cli @caLnca
00 Cl::r - t-.. 0 t- 0 • - o· 8-0
I~ ~ ~ ~
0 lt'l 0 0 0 0 0
'" . lt'l . r-i . oqo u:lN r-i N r-i N r-- r-i . tIl. lt'l . M . 0 . lt'l '.-I0 r-i 0 r-i 0 r-i 0 r-- X+1 +. +. +. +1 +1 +1 +1 ~
Fig. 2a(ii) Spot diagrams showing the theoretical image gualityof the QRC system~ corrector (flat field).Optimum focus for red.
ooIJ:I10/l1
• 0 • 0 -0 '0
I~- '0 -0 '0 ·0 0::r0':r
UQ)lJlU~
0.11l
CD -0 '0 .~.) 0 8 Itl,-IJ:I ./l1 0:r
11
~I"'-
0~- G G B G) @,- Q)
10 +J::r Q)IJ:I
~•.-l"CIQ)r-fU~•.-lU0
/l1 (> @,-10IJ:I10
0~
"'" \0r-f . lJl. Itl ..-l0 I"'-
~+1 +1
oo::r --o- 00--
I
~o Itl
0'1 •N r-f• Itlo r-f
+1 +1
0 (J- 0 0-
I I~ ~
0 r-f 0 0Itl . r-f .N r-f N I"'-. M . S0 r-f 0
+1 +1 +1 +1
00 o
Fig. 2a(iii) Spot diagrarns showing the theoretical image qualityof the ORC system~ corrector (flat field).Optimum focus for blue.
+ 0.29 0 ····6:-·.":: 0° • • • :: , ••'. .'
+ 145.2 '.::::.::::.::::::.......
TI
+ 0.25 0
~.... {:", ';..125.7 '.' .'+ mm ':'".....:'
"
+ 0.210 •+ 102.7 mmt\•.
+ 72.6 mm
Axis oCirc1e Diameter
= 70jU = 0.50 arcsec
Fig. 2 b Spot diagrams showing the theoretica1 image qua1ity of theQRC system without the corrector for p1ates bent to optimumcurvature ( r =3384 mm)
~
I1338,8894,121 687,5
\'lith lens
) - -15 (cl = + 1,899386.10 r = - )) c 2 = 0 ( r = - )
t: 1 = - 1,15500
E2 = - 1
\ 1439,8 ...without lens
I \~ 7491,0
r 8761,8 ro- 25,0 160,)
/1456,1I
10 735,9(17)
- 4 -13 (c l = + 5,966 23.10 r =_ I -20 (c 2= - 1,~1090.10 r
Cl = - 6.906381
C2=+3 1 ,197
Asphericconstants:
Radius ofcurvature
Free diameter :field ± 0,2880
Diameter ofaxial beam
1184,0
1107,9
3570,0
3570,0
307,9
20,94
309,5
19,10
302,9
o
f =- 30 035,9 111:11
f = 28 884,8 10111
with correetor
,",'i thout eorreetor
k = 8,413
k = 8,091
FIG. 3 Optieal Data Sheet for the QRC foeus
Radius of curvature 7 030,30 21 687,5, (plane rnirrors)
,
\ 7664,42·1 .1- 33 279,5 I --..Jr t 25 615,0 I
Aspheric cl = + 4,250695.10- 13 ( r = - ) cl = + 1,899386.10- 15 ( r = - )constants:
c 2 = - 5,4867.10-21(r = - ) C2 = 0 ( r = - )
&1 = - 2,181603 E1 = - 1,15500
E2 = + 0,508 ~2 = - 1
Free diameterfield + 3 arcmin.: 1062,2 3570,0 19B,1
Diameter ofaxial beam:
1048,7 3570,0 o
f' = 11 J 506 mm k = J 1 ,791•
FIG. 4: Optical Data Sheet for the Coude focus
:Y~ "",,
l.4Il-~ f-" r-... I,.,..lI'" l':' ",\ :Yl'" ,,\1'''l'\ "",,
I-'" .- 1-- -, '-''"~ ,,\ ~ \.'\~ ~ 1'\."" ['\ ~ i:Il-- f= I-~- -I- ~ -~ " f-,.~' l:\.: ,"" Y~ , "- ~I' ,
10 r0~ I'\. "",. ,,\I' l'\: " r'\.' l"\: 1'\.""
11 I 11 1 II 11 I I 11 1 11 1I I I I
rrI I I II I I I I I
n~
Central obstruction
>-........W0\
~
>-.............w
~
.....oUlCD
.....l\JW~
.....~
oCD
..........UlO\CDl\JW l\J
..........UlCD
.....\0W\0
l\J.....o.....
l\Jl\JCDl\J
l\J~
UlCD
l\J0\Wl\J
l\JCDo.....
Wol\JUl
Wl\J~
l\J
W W W~UlUl
UlOUlWO\O
Io
n.l\J
I I I I I I I I I I I I I11 1 1 I 11 1 11 1 11 I 11 1 I
,""l\.:: ,,\ I'\: 0-: ",\ :Y0-: ,""~~r0 /" .......... ~ I~t'0~~ in "- " ..$ / .......... ....... -- 1-'", \. " .- I- , ... ::-:,.- 6-k- ~ l- V _.
:::.....:: P-'~ - t:":::: f$- 1-- ~-
~~~ " -0:~~ -0:~~~~ Iir .......... QY F-:::: r---0-~~ r.....'\ :'\,' -0: ,""~ -0: r'\.""l" r....."" "~
Central obstruction
Fig. 5 At the right
At the left
error of finished primary as a slope - height function
error of finished primary as a wavefront error - height function
with reference to the "best fit" aspheric surface.
(Figures reproduced from REOSC report).
(i) 0':6
0~4
0~2
0ll'l 0'.'10r-i 22.0N 6:
11 22.5 0~2
I-l 23.00~'4
0~6
(ii) 0~4
0~2
72.0 0"J: 72.5
73.0 0~2
8 0~4l'r-iN
110':6
I-l0~'8
0~10
~
G>1<1'1 I
v r-.... jLE 1/ - lQ~
Iw..... f"-~ Im Im
/ I", ..... -Im... L-... lt::I / ~
~1/ v I":Cl
IP' ~ 1"--- --+-.. j--
a< Im. 1'1il Im Ir>
v ~j-- j-.
IlD, 1'--< J I\.u;
,"'- 9-- ~ Ie fd>
-L:I0:::..:.:...:5o..- ~c ~_=1:.:..•.:::.5 _L::ll!:....!.~8
-'Il.
I"",./ ~N ~- r\
-'" '\t!'.. 1/ r\IJ" Je:: 'I-- N '"1-
111V I)l" -". N j\ -, ) Q)
lil 1/ Ie... " v I" J(J1 f".. 11' Im ""
I"\. / \ 1'E< !\1',,( 1\ I\. I W
1""... l-n(
'" I'l"\ 1\I" I '"~
(11i) 0':6
Height above axis (m)
LA..k
Ie;)-V ....... fl..... Im..- 1"---!_ i". 1/
V I)" t--- f'm 1\Etl1/ ~ ~ I'--1--m(jJ "- I/ 1'-..
1ll \ ~ RI'--t--It> 1\
" '\ ,r,G>""", Ii'
"\ ~
1\ IY/ IIV
IW
1.5
11..~ _. ..J.1-=1~.~81.0
I 1.0
10.5
0"
0~4
0~2
0~4
0~2
0,'6i
0':8
J. 122.0122.5
Fig. 6 (a) : Behr's analysis of Hartmann plates of primary withmirror axis vertical. Solutions for Cl = - 1.15175.
9Im ~
m
I,...- i--" ..... -G 1/ "- .JIw- 1:> hi IJIl 9
/ I'i'. 'l-. ;11\ 6.. ~
-'" --r- t'-., VI '\l ,.- I.- rw., •1/ % $... ~ !Il.. ~'tJt--..-,.7 ~ / [\. 'J '\
(Jf "- I>-- -f.... T~
~ II'l/l
rk--
Ir.;:
rlil
~ 1./ f),.,
/ .......V-- --r-A I IICi)-'?' ........ --d ~ /D- 'E&"
/ ........"" I ~ / 'lI:l'"
-(j; Iv I)U r--..... '$ K I~ 1l....<17V ~h- I ""./
~
~ "" I0;tJ
...m. ~
V:>-- ---(J / \.
I~ ~ ~ t--.. ~
; Im-[..:.:;J I'" I---l~ P\ " I~.A"\
~ 1/ 1\ '" /-ijj1/ IJtT ,..... r--.. ~ Ittl(j)V !'Ei "\ rm r{~ "- v 1"\ Q)
llJ 7•~ (<V ~
111\ 1\i\. I ~~
(i) 0~8
0~6
0~4
0~2
0"
0It'l 22.5 0~2
::: J' . 23.0N •
1123.5 0~4
J.l0~6
(H) 0~6
0~4
0~2
8 72.0 0"....r-l 72.5N 6: 72.6 0~2
11 73.0J.l
0~'4
(Hi) 0~8
0%
--u 0~4Q)UlUJ.l 122.0 0~'210 0
~ c): 122.5l:: r-l 0"0 N 123
•.-1+J 1110 0~2J.l J.lJ.lQ)
0~4~J.l
0~'610r-l::s0'
~
Height above axis
10.5
,0.5
(m)
11.0
1.0
11.0
I 1. 5
1.5
I 1. 5
1.8
1.8
Fig. 6 (b) Behr's analysis of Hartmann plates of primary withmirror axis vertica1. Solutions forCl = -1.15675.(Fig. 6 (b) (ii) is considered the most probablesolution) .
(1) On4
0~12
0~10
0~8
0~6
0~4
0LO 0~210r-iC'l
11 0"22.0
I-l22.5 0~'2
J" 23.023.5 0~'4
24.00~'6
l.m9PI' '"
/' I~ 1-.e ß
~~~y ./;JJl w
/ ? I1/ / I
j I""'" ........., / ~I(r.' .G>-
I~. ~E,...- -1:/ / .,/
..",-
.-E y/ V / 1/' .d ~ / I
w~~ / -1/ \.
1/ 1% it\ I", /wGi / IQ-...... .....N!S I'-..V Ia< / J:'> 1/ 1- L.,(Ld /rJ'U """"'!'-... Im ~
'(±) i'-.-' /
V ~N Il-- 1/III /"'-..
- -
..1
.Al. 11+1
-...... l$-N ß..,... / ~
.Al. "" '\ .......-~ I---... h-1/ ßY ~ r...... l'e I~ '" /
-"~1/ r-a / ~C1 1'-../ "' "' ~-
~ ~
"" I'd>
~/
11> /~ ./:D" fe"/ "- ./ }.,
--'IV - "1Ii IllY' "Il5 ~
-:;: 7/- --e--..; / "./' 'W~
..I ""..........
..~
jIr 19- '" "\ 7'" / '""-N 1/ '\v ~.l
eJ~
(11) 0': 8
0~'6
0~4
00~20
I'r-iC'l 0"
11
I-l72.5
J" 730~2
73.5 0~4
UQlUlUI-lIII (111) 0~4
I:: 0~20•.-4 0+J LOIII I' 0"I-l r-i
~=123.0I-l C'lQl 123.5~
11 0~2;
I-lI-l
0~4IIIr-i~Cl
~
He1ght above ax1s (m)
0.5
0.5
0.5
1.0
,1.0
11.0
1.5
1.5
1.5
1.8
1.8
Fig. 6 (e) Behr's analysis of Hartmann plates of primarywith mirror axis vertieal. Solutions forEI = - 1.16175.
l-
I!
)i\ J.~ I (Wo~ 1\ ~1I",
ft 1/ ~ ~ 1\1...... h;.. Jt\ I""
I 1'-.. 1/ jf. r" W / ) I( " V I}f)"t- f'.. u 1/ IW Jt) 1\1,..,.. 9- l-.r. I",\ II1/ "" r-~
~
Iw ~1- A
\-r,
\ 1-
Ii''" / 1\1",-~ f',.
J _ ~r-- I:tl lp..... - G
" V "" " A \ '$~ ~ I(~" V IJtT
""" 7 - ~ 1/ 1\ ~ I",
11" I'\tl1/ IJ&-I-- "l "J'.. ~ ktl If'-o...
1\I"ID / 1/ l'a R 1-j4 J \ '" ~ I ... J
I\.i ~ 1/ 1\ IGi'.. f'-.. ~ Ij;)
-.v I'E'" ~
IV~
- 111V 1'-.. J \
/,...-"", / CO) rrv- t--. l/ 1\~ f'.,. '" 1\I" / LV' ." 17 I'Q;)..
"-I--~ I \ r'al
I' ~ / 1: i'. 1/ 1\ .'Elt-. !Al I)l'
r\; "I, Ii?I'.:- r\7
r'm. ~
'U~ j
IV
(i) 0~6
0~4
0~2
0ll'l 0"~ 6 19.0N ~ 19.5
0~211
1-40~4
(11) 0%
0~4
0~2
68.50 0"g 6: 69.0.-l 69.5N
0~'211
1-40~'4
0~6
0~8
(11i)
t.l0:'6
<lltflt.l 0~41-4nl
&118.50, 0~2l:: ~ 119.00·ri .-l
0".j.l Nnl1-4 11l-I<ll 1-4 0~2
~1-4 0~4nl.-l::1 0~6Öl
~
I 0.5
I 0.5
0.5
1.0
1.0
11.5
11.5
11.8
I 1.8
1 1. 8
Height above axisFig. 7 :
(m)Behr's analysis of Hartmann plates of primary withmirror axis horizontal. So1utions for cl = -1.15675.(Fig. 7 (ii) is considered the most probable solution).
100%
ion
Radiusarcsec)
l....-' :- .......--
~v~
v/' "'0ht ra ct SIPE cj fj Ce t
1//
//
1//
1//
11/1
(/
1/
50%
o 0~2 0~3
Fig. 8 Finished prirnary - geornetrica1 energy concentration(Figure reporduced frorn REOSC report.
-....--lt---reflexion parasitedue au systemecorreeteur
".~--,,:......;;.::.,--+--obturationcentrale
Fig. 9
Meridian C
0~2
0~1
0"
0~1
0~2
'a\ ~
\ / \ y ...... "'-I/ \ 1
\ I ~~ \\ / \ 11 ~
\ \ }..,
I~ 1\
I I
o 10 10 10o co co cor- r- co CTI
10 10co 1010 r-.... ....
10 10co co"'" lt'l.... ....
10 10co coN M.... ....
10 10co coo r-l.... ....
I
.... '\..\ / \ / I\.
/ \ ~ "4
\ \ I \y \ /
~~ I w
\ v1'\ /
0~2
0~2
0"
0~'1
0~1
Meridian B
10co10
10cor-
10 10co coco CTI
10coo....
10co........
10 10co coN M.... ....
10 10co co"'" lt'l.... ....
10 10co 1010 r-.... ....
0~'3
.....0
0'~2Q)l/l
Era 0~1.....c:: 0"0.....j.lra
0~1I-tI-tQ)
~ 0~2
I-tl'dr-l='t1>~
I
/1\1/ \
/ ['...I 1\ / '0-
e \ / .....
'" V 1\ /1 I( \ \
/ \\ /
:r- 1\'I
w
Meridian A
Height above axis (mm)
Fig. 10: QRC Secondary - S10pe - height error function of thetest wavefront - Axis (Reproduced from REOSC report).
o ~ \0 \0OCO co cor-. r-. co 0\
\0 \0 \0co co coo r-l Nr-l r-l r-l
\0 \0 \0co co \0Ln \0 r-.r-l r-l r-l
0~3
0~2
0~1
0"
0~1
0~2
, I
lil
/
1/
V 1'\ / ...... 1/) 17 r\ 1/
1/ " / \V /..I ''Hf \1/w
Fie1d + 140 mm
0\0 \0 \0 \0OCO co co cor-. r-. co 0\ 0
r-l
\0 \0co cor-l Nr-l r-l
\0coMr-l
\0 \0co co~ Lnr-l r-l
..... 0~2uQ)Ul
0~'1uHrtl
0"s::0
or-f on.j.J
rtlHH 0~2Q)
~H
0~3
rtlr-l~C'l
~
I I I
1\I \I
~ 1\ 1/ i"- 11 \ I)'"
1\ J l\. l"-t--<V~ f.-.
i\ I11
\
Fie1d + 105 mm
Height above axis (mm)
Fig. 11: QRC Secondary - Slope - height error function ofthe test wavefront. - Meridian C - (Reproducedfrom REOSC report).
Radius in Concentrationsarcsec
0~0518,88 + 11,88 + 16,66 = 15,51 %3
0~'1035,93 + 69,15 + 52,13 = 52,40 %3
0~1557,57 + 69,15 + 80,05 68,92 %3
0~1996,72 + 90,57 + 75,57 87,62 % Average for3 three
0~2096,72 + 90,57 + 86,40 91,23 % meridians
3
0~208 96,72 + 90,57 + 92,81 93,37 %3
0~21596,72 + 100 + 92,81 96,51 %3
0~305 ------------------- 100 %
Radius,arcsec)
O~ 30~20~1
-...11:
~
/
//
/'V
V}~
//
//
/:V
V./ J
o
50%
100%
Fig. 12 QRC Secondary - Geometrica1 energy concentration in axialimage of QRC combination. (Reproduced from REOSC report)
%V
./V-
I,.-- i--"'"
/"J
/
r7/
/
/
1/
1/
1//
11 (
%_I--I--I--
V' ~-./v
,/./\
,/
./
7
1/V
l/
1/
1//
1/
/ (
100
50%
o
100
50%
o
0~1
0~1
0~2
0~2
0~3
0~3
Radius,arcsec)
Radius,arcsec)
Fig. 13 QRC Secondary - Geometrica1 energy concentration atfie1d + 105 mm (above) and + 140 mm (be1ow) of ORCcombination (Reproduced fro~ REOSC report).
\0 0 r-- Lnco r-- \0 \0\0 r-- co '"
M r-l\0 \0o r-lr-l r-l
'" \0 LnLn Ln '<I'N M '<I'r-l r-l r-l
r-l CO \OLnLn '<I' N\OLn \0 r--r--r-l r-l r-lr-l
0~2
0~1.....uaJ 0"UluHtU
0~'1......l::0 0~2•.-1.jJtUH 0~3HaJ
~ 0~4
HtU
r-l~C'>
~
I I I I I I I I I I
--~ 11 l'r I~ I \
\ 1"'- 11/ \ / ~ 1\/ In \
I>J \ / \ I1'1 1\ I
1/ Illl hL/
V
Height above axis (mm)
Fig. 14 Coud~ Secondary - Slope - height error of thetest wavefront (Reproduced from REOSC report)
l.US,
sec)
E9--~k/V
/y
V DY
././(
//
V1/
//
//
(Rad/V are
s::oorl.j..lIII1-1.j..ls::(])us::ou
50%
100%
o on 0~2 0~3 0~4
Fig. 15 Coude Seeondary - Geometrieal energy eoneentration in axial image of Coudeeombination (Reprodueed from REOSC report).