Upload
others
View
7
Download
0
Embed Size (px)
Citation preview
PIRO® Manual, Version 1.5.2
1
THE PIRO®
PIPETTING ROBOT
OPERATION MANUAL
Version 1.5.2
PIRO® Manual, Version 1.5.2
2
All rights reserved. No parts of this work may be reproduced in any form or by any means -
graphic, electronic, or mechanical, including photocopying, recording, taping, or information
storage and retrieval systems without the written permission of the authors.
Products that are referred to in this document may be either trademarks and/or registered
trademarks of the respective owners. The publisher and the author make no claim to these
trademarks.
While every precaution has been taken in the preparation of this document, the publisher
and the author assume no responsibility for errors or omissions, or for damages resulting
from the use of information contained in this document or from the use of programs and
source code that may accompany it. In no event shall the publisher and the author be liable
for any loss of profit or any other commercial damage caused or alleged to have been
caused directly or indirectly by this document.
PIRO® Manual, Version 1.5.2
3
Intended Use The PIRO® Pipetting Robot is designed for the use of accurately transporting liquids within
the PIRO® system. With no exception, the instrument can only be used by trained personnel
and only with liquids and substances appropriate for the instrument. The standard deviations
for pipetting volumes determined by DORNIER-LTF can only be achieved by using the
pipetting tips recommended by DORNIER-LTF. Any use other than mentioned above does
not apply to the intended use of the system and DORNIER-LTF shall not be liable for any
damage resulting from such unintended use. An intended use also includes the compliance
with any safety and environmental laws and regulations as well as the instructions in the
manual to prevent any damage to the user and the environment. Any liability will also
extinguish when no or inadequate maintenance of the system has been performed. In doubt,
please contact DORNIER-LTF.
Intended Use of any Accessories on the PIRO® Pipetting Robot
Any accessories used on the PIRO® must have been recommended from DORNIER-LTF
and can only be used according to the description given by DORNIER-LTF. Any additional
equipment or accessory not recommended by DORNIER-LTF will not be regarded as an
intended use. For any damage resulting from such use, DORNIER-LTF will not take any
liability. The connection between the PIRO® and any accessory may only be done according
to the regulations outlined in the documents delivered or outlined in this manual. For any
damage resulting from inadequate connection of accessories DORNIER-LTF will not take
any liability. In doubt, please contact DORNIER-LTF GmbH.
Changes to the PIRO® Pipetting Robot
The PIRO® as a complete system is tested according to the safety regulations and can only
be used according to the intended use. Any changes to the PIRO® can only be performed
with the permission of DORNIER-LTF and only in the scale the permission has been granted
for. If the PIRO® will be changed without the permission of DORNIER-LTF, any liability and
warranty will be extinguished. For any damage resulting from changes to the PIRO®
DORNIER-LTF will not take any liability. In doubt, please contact DORNIER-LTF GmbH.
Original Parts from DORNIER-LTF
All parts used in the PIRO® are tested for safety and are optimised for the use of the PIRO®
in connection with all other parts used. Faulty parts can only be exchanged using original
parts from DORNIER-LTF. If original parts are not used and inadequate spare parts are
used for the replacement of faulty original parts, for example electrical parts, danger to staff
could arise from such a replacement. Non original parts could also influence the
performance of the PIRO® and its pipetting quality. For any damage to the user or the
instrument resulting from the use of non-original parts to the PIRO®, DORNIER-LTF will not
take any liability and the warranty will be extinguished. In doubt, please contact DORNIER-
LTF GmbH.
PIRO® Manual, Version 1.5.2
4
Contents
1 The PIRO® Pipetting Robot ............................................................................................ 8
1.1 Unpacking the PIRO®.............................................................................................. 8
1.2 Box Contents ........................................................................................................ 10
1.3 The PIRO® Hardware ............................................................................................ 10
1.4 Pipetting Heads .................................................................................................... 11
1.4.1 Changing the Pipetting Head ......................................................................... 11
1.5 UV Radiation Procedure ....................................................................................... 11
1.6 HEPA-Filter ........................................................................................................... 11
1.7 Positioning the Robot ............................................................................................ 11
1.8 Lid ......................................................................................................................... 12
1.8.1 Functionality of the Lid ................................................................................... 12
1.8.2 Cleaning Acrylic Plexiglas .............................................................................. 12
1.9 Calibrations ........................................................................................................... 12
1.10 Safety ................................................................................................................... 12
1.11 Tips ....................................................................................................................... 13
1.12 Specifications ........................................................................................................ 13
1.13 Level Sensing ....................................................................................................... 14
1.14 The Computer ....................................................................................................... 14
1.15 Connecting the Robot ........................................................................................... 14
1.16 PIRO® Software Installation .................................................................................. 14
1.16.1 Updating the Software ................................................................................... 16
2 Introduction to the PIRO® Software .............................................................................. 17
2.1 Downloads ............................................................................................................ 17
2.1.1 PIRO® Software ............................................................................................. 17
2.1.2 PIRO® Manual ............................................................................................... 17
2.1.3 Other Documents ........................................................................................... 17
2.2 Starting the Software ............................................................................................ 17
2.3 Virtual Mode.......................................................................................................... 17
3 The PIRO® Workspace ................................................................................................ 18
3.1 Deck Layout .......................................................................................................... 18
3.2 Racks and Adaptors .............................................................................................. 18
3.2.1 Orientation of Rack and Plates ...................................................................... 19
3.2.2 Tip Racks ....................................................................................................... 19
3.2.3 Tips ................................................................................................................ 19
3.2.4 Tip Disposal ................................................................................................... 19
PIRO® Manual, Version 1.5.2
5
3.2.5 Racks and Adaptors....................................................................................... 21
3.2.6 Plate Positions 4, 8, 12, 16 ............................................................................ 22
3.2.7 Tubes ............................................................................................................ 23
3.2.8 Setting up Plates in the Software ................................................................... 23
3.3 Position Offsets .................................................................................................. 24
3.4 Plate Functions ..................................................................................................... 25
3.5 Exchanging Existing Plates ................................................................................... 25
4 Main Menus ................................................................................................................. 26
4.1 File Menu .............................................................................................................. 26
4.2 Layout Menu ......................................................................................................... 27
4.2.1 Clipboard ....................................................................................................... 27
4.2.2 Tools .............................................................................................................. 27
4.2.3 Zoom ............................................................................................................. 28
4.2.4 Extras ............................................................................................................ 28
4.3 Options Menu ....................................................................................................... 29
4.3.1 Language Options ......................................................................................... 29
4.3.2 Robot Options ................................................................................................ 30
4.3.3 Graphics Options ........................................................................................... 31
4.4 Robot Menu .......................................................................................................... 31
4.4.1 Robot Controls ............................................................................................... 31
4.4.2 Report ............................................................................................................ 32
4.4.3 UV Program ................................................................................................... 33
4.5 Deck Window ........................................................................................................ 34
5 Feature Tree and Feature Table .................................................................................. 35
5.1 Feature Tree ......................................................................................................... 35
5.1.1 Triangles ........................................................................................................ 35
5.1.2 Info ................................................................................................................ 35
5.1.3 Warning ......................................................................................................... 36
5.1.4 Error .............................................................................................................. 36
5.1.5 Add ................................................................................................................ 36
5.1.6 Remove ......................................................................................................... 36
5.2 Feature Table ....................................................................................................... 36
5.2.1 Well Position Description ............................................................................... 36
5.2.2 Changing Names ........................................................................................... 37
5.2.3 Viscosity ........................................................................................................ 37
5.2.4 Mixing Options ............................................................................................... 37
PIRO® Manual, Version 1.5.2
6
5.3 Plate Submenu ..................................................................................................... 38
5.3.1 Plates ............................................................................................................ 38
5.3.2 Tips ................................................................................................................ 38
5.3.3 Waste ............................................................................................................ 38
5.4 Transports ............................................................................................................ 38
5.4.1 Changing the Order of Transports .................................................................. 39
5.4.2 Setting up Transports ..................................................................................... 42
5.4.3 Auto Assignment of Transports in Empty Wells .............................................. 43
5.4.4 Function Colour Coding ................................................................................. 44
5.4.5 Transports and Selecting Wells ..................................................................... 44
6 Setting up a Run .......................................................................................................... 45
6.1 Diluent .................................................................................................................. 45
6.1.1 Setting up Diluents ......................................................................................... 45
6.2 Reagents .............................................................................................................. 45
6.2.1 Setting up Reagents ...................................................................................... 46
6.3 Mixes .................................................................................................................... 47
6.3.1 Setting up Mixes ............................................................................................ 48
6.3.2 Pipetting Mixes into Mixes.............................................................................. 49
6.3.3 The Pipetting Order........................................................................................ 49
6.4 General Orientation on Plates ............................................................................... 50
6.5 Samples ................................................................................................................ 51
6.5.1 Setting up Samples ........................................................................................ 51
6.5.2 Sample Options in the Deck Window ............................................................. 51
6.5.3 Skip Missing Samples .................................................................................... 52
6.5.4 Sample Groups .............................................................................................. 52
6.5.5 Sample Group Table ...................................................................................... 53
6.5.6 Creating Sample Groups from Samples located on different Plates ............... 54
6.5.7 Import ............................................................................................................ 55
6.5.8 Intelligent Pipetting ........................................................................................ 59
6.5.9 Barcode Reading ........................................................................................... 64
6.5.10 Normalisation of Sample Concentrations ....................................................... 65
6.5.11 Dilution Series of Samples ............................................................................. 67
6.6 Reactions and Fixed Reactions ............................................................................. 69
6.6.1 Setting up Reactions ...................................................................................... 69
6.6.2 Reaction Feature Table ................................................................................. 70
6.6.3 Setting up Fixed Reactions ............................................................................ 71
PIRO® Manual, Version 1.5.2
7
6.6.4 Special Features of Fixed Reactions .............................................................. 72
6.6.5 Reaction Export ............................................................................................. 73
6.6.6 Sample-Multi-Dispensing ............................................................................... 74
6.7 Standards ............................................................................................................. 76
6.7.1 Setting up Standards ..................................................................................... 76
6.7.2 Ready to use Standard .................................................................................. 76
6.7.3 Creating a Standard ....................................................................................... 76
6.7.4 Standard Options in the Deck Window ........................................................... 77
6.7.5 Standard Group ............................................................................................. 78
6.8 Reserved .............................................................................................................. 79
6.8.1 Setting up Reserved Area .............................................................................. 79
6.9 Special Functions of the PIRO® Software .............................................................. 80
6.9.1 Automatic Tube Switching (ATS) ................................................................... 80
6.9.2 Automatic Fluid Transfer (AFT) ...................................................................... 81
7 Quick starting the PIRO® .............................................................................................. 82
8 Appendices .................................................................................................................. 83
8.1 PIRO® Alarm List .................................................................................................. 83
PIRO® Manual, Version 1.5.2
8
1 The PIRO® Pipetting Robot The PIRO® PI-petting RO-bot is a precise pipetting system for performing a vast variety of pipetting operations. Due to its high flexibility, the PIRO® can be programmed to set up multiple mixes, standards, dilutions, and samples, using multiple groups of samples in many plate and tube formats. It automatically calculates the volume of mixes, reagents and other components required for each experiment. Import functions are implemented as well as export functions, which can be used for further processing. Reports are generated before and after each run and are automatically saved for future reference. For maximum precision the PIRO® uses high quality non-conductive tips with three tip volume sizes (50 µl, 200 µl, and 1000 µl) in order to achieve a perfect combination of volume capacity and pipetting speed. An innovative sensor technology provides a highly precise level sensing precision of very low volumes down to 5 µl. The tip moves to just below the surface of the liquid, rather than the bottom of the tube preventing carryover of material on the outside of the tip. Specially optimized pipetting performance steps increase accuracy and precision of all pipetting steps. The PIRO® software has been designed to be user friendly. The progress of any run can be followed in real time. On the screen, the software highlights the position of the pipetting head while moving. A progress bar and a continuous table update show the status of the current run. The screen layout corresponds to the plates set up on the robot. Any desired change can be performed by simply pointing to the corresponding image of the hardware on the screen. The PIRO® software can also be run in virtual mode which means runs can be set up, progress followed, and files stored, even if the robot is switched off or disconnected. The PIRO® software can be used on computers not directly attached to the PIRO® and due to the small size of PIRO® files they can easily be transferred to the computer running the PIRO® instrument. For further information or assistance, please do not hesitate to contact your distributor or the DORNIER-LTF team. DORNIER-LTF GmbH, Rickenbacherstraße 107, 88131 Lindau a.B., Germany.
Tel: +49 (0) 8382-273089-10, Fax: +49 (0) 8382-273089-29, www.dornier-ltf.com.
For general enquiries about DORNIER-LTF products and sales, contact: [email protected].
1.1 Unpacking the PIRO®
The packaging of the PIRO® has been carefully designed in order to ensure safe shipping
conditions. Please do not discard the packaging in case the PIRO® needs to be transported
again. The unpacking procedure is outlined below, to repack simply reverse the instructions.
PIRO® Manual, Version 1.5.2
9
Open the box and remove the foam protection from the top of the PIRO®.
Remove the sides of the cardboard box by lifting it up from the PIRO®. Lift the PIRO® on to a
table and carefully remove the plastic protection, whilst taking care not to scratch the lid of
the PIRO®. Remove the two small foam inserts located below the pipetting arm.
Remove the top two pieces of foam protection and cardboard box from the PIRO®. The
remaining piece of foam protection on the bottom of the PIRO® can now be removed.
Before connecting the PIRO® to the computer make sure, the deck is completely empty. The
PIRO® can now be connected and is ready for use.
PIRO® Manual, Version 1.5.2
10
1.2 Box Contents
The PIRO® shipping box contains the following items listed below unless ordered differently:
1 x PIRO® (P100100, or /200, 300, 400)
1 x IEC Mains Cable (E100312)
1 x USB-cable (E100291)
1 x Software CD
1 x Manual (P100343)
Various racks consisting of:
1 x waste channel-1 (P100322)
1 x waste rack (P100321)
4 x tips racks (P100317)
2 x 96-well plate racks (P100318)
3 x standard racks (P100319)
3 x tube racks (P100323, P100487,
P100371)
5 x 96 tips (50 µl) (P100328)
5 x 96 tips (200 µl) (P100329)
For more details, please see delivery note
A computer is required to operate the PIRO® and can be ordered with the PIRO® if required.
The box content may vary, depending on the specifications of the order. Please check the
content while unpacking.
1.3 The PIRO® Hardware
An overview of the PIRO® hardware is shown below. The dimensions of the PIRO® are as
follows: 600 x 515 x 470 mm.
PIRO® Manual, Version 1.5.2
11
1.4 Pipetting Heads
There are two choices of pipetting heads for the PIRO® available and each head has its own
electronics and calibration data:
1) Pipetting Head 200 2) Pipetting Head 1000 - in preparation
1.4.1 Changing the Pipetting Head
To change a pipetting head turn the
power off and move the y-arm
forward and the pipetting head
upwards to have an optimal
handling space. Hold the pipetting
head with your left hand and press
the unlock button on the bottom of
the head with your right hand index
finger and move the pipetting head
out of the lock. To install a new
pipetting head reverse the
procedure making sure both pins
are sitting in the appropriate holes
and fit properly to the connector.
The calibration data are stored in
the pipetting head and no new calibration has to be performed.
1.5 UV Radiation Procedure
The UV radiation procedure is implemented to allow the user to use UV in the PIRO® for
sterilization at the end of a run. The procedure can only be performed with a closed hood to
avoid any damage to the user. Though the hood is manufactured using UV-blocking acrylic
Plexiglas preventing transmission of UV light to the outside, for safety reasons do not look
into the UV light. Opening the hood will stop the UV radiation procedure immediately and the
procedure has to be restarted again.
In order to use the UV light it is best to remove all racks and disposables from the deck, so
the UV can access the complete deck area. To start the procedure, please see 4.3.2 and
4.4.3. The PIRO® arm will then move over the deck to avoid shadow effects.
1.6 HEPA-Filter
The PIRO® can be delivered with or without a HEPA filter. If a PIRO® has been installed
without a HEPA filter, it can also be upgraded by the distributor at a later stage. To active the
HEPA filter press the box ‘Activate HEPA filter’ located in the
‘Options’ menu. Once a PIRO® run is started the HEPA filter will then be activated.
1.7 Positioning the Robot
The PIRO® can be positioned on any bench, as long as the bench is reasonably stable and
not moving by the movement of the PIRO®. Make sure the lid can be opened completely as
otherwise scratches or damages of the lid might occur. The temperature and humidity should
be reasonably consistent and according to the specification (see 1.12). It is important to
make sure all four feet of the PIRO® evenly touching the surface. As outlined in the below
PIRO® Manual, Version 1.5.2
12
chapter (see 3.2.4.1), if the waste channel-1 should be used but not the waste console (see
3.2.4.3), the front feet of the PIRO® have to be placed just before the edge of the table,
therefore allowing to place a waste bin below the waste channel-1.
1.8 Lid
The PIRO® is equipped with a lid to protect the reactions and reagents within the PIRO® but
also to protect the user. The lid is manufactured using UV-blocking acrylic Plexiglas.
1.8.1 Functionality of the Lid
The PIRO® will only work with a closed lid. Opening the lid during a run pauses the
instrument in order to protect the user. When the lid is closed, the PIRO® run has to be
actively resumed and the pause is recorded in the post-run report.
1.8.2 Cleaning Acrylic Plexiglas
Care should be taken when cleaning the lid. Use a soft cloth (e.g. cotton) never use paper
towels as they will leave streaks on the surface. Plexiglas has a very soft surface which can
easily be scratched. Dust can be blown off with a hairdryer or washed off using water and a
soft cloth. Then use a recommended cleaner to complete the job. Note: a mild solution of
dish detergent and warm water takes off stubborn dirt easily without harming the acrylic
Plexiglas. Please, never, use cleaners containing ammonia. Some recommended cleaning
methods are given below:
• Use only products specifically recommended for cleaning acrylic such as Novus
#1, Plexiclear or Brillianiz,
• Fill a spray bottle with an equal mixture of water and isopropyl alcohol. Use the
type of alcohol available in the first aid section of your grocery store,
• Mix 2/3 cup white distilled vinegar, 1 1/3 cups distilled water, and 5 drops
lavender essential oil, 5 drops rosemary essential oil in a spray bottle and clean
the Plexiglas surface.
1.9 Calibrations
The PIRO® does not need any calibration during the installation of the instrument as for all
plates and vessels in the plate database, the calibration has been performed in the factory
and the user can start working with the instrument straight after the installation.
.
All plates, racks, and positions implemented in the PIRO® software are pre-calibrated. If
alternative plate types with unusual formats need to be used, please contact the distributor,
as at this point of time, calibration of such unusual formats must be implemented by
DORNIER-LTF. As soon as new plate settings are available they will be implemented in a
software update for any user to be downloaded at following web page: www.dornier-ltf.com.
If the calibration of plates with formats available in the software needs to be adapted, please
see 3.3, which describe position offsets.
1.10 Safety
The PIRO® pipetting robot has moving parts, which the user needs to be aware of when
handling the PIRO®. For your own safety and to prevent damage of the instrument, please
take note of the following safety requirements:
Always use the packaging provided with the robot when shipping the instrument
PIRO® Manual, Version 1.5.2
13
Do not interfere with the PIRO® while running (e.g. moving plates or tubes)
Do not lift the PIRO® by yourself to prevent damage to the user or the PIRO®
Keep the lid of the PIRO® closed whenever possible
NEVER apply any force (by hand or otherwise) in any direction to the robot arm
NEVER place any items on the y-arm of the instrument
NEVER put your hand under the pipetting head while it is descending
Only use recommended tips and tip racks
The PIRO® is for indoor use only
Use the Stop/Pause button located in the tool bar of the software to stop/pause the
robot before removing plates or tubes
1.11 Tips
The tips used on the PIRO® are clear, sterile, and filtered. Conductive tips are not needed to
perform level sensing (see 1.13). If 50 µl tips are incorrectly placed in a 200 µl tip position,
the pipetting head may become contaminated and will require a service.
Tips other than those recommended by DORNIER-LTF can be used on the PIRO® provided
that the tips have a compatible hub size and are of the same volume. However, DORNIER-
LTF does not guarantee the precision or accuracy of pipetting, if tips others than those
recommended are used. Performance may seriously be affected and any new tip must
undergo evaluation before using.
1.12 Specifications
Pipetting Tips
96 racked filtered tips, 50 μl, 200 μl, and
1000 μl (clear, non-conductive tips)
Pipetting Volume Range
0.5 μl - 180 μl using the 200 μl pipetting
head
X-, Y-, Z-axis Drive Resolution
Less than 0.0065 mm
Pipetting Speed
10 μl – 200 μl/sec, software adjustable
HEPA-Filter (optional)
High Efficiency Particulate Air filter system
provides positive pressure
UV Sterilization (optional)
UV light system for deck sterilization under
closed lid conditions
Precision, Accuracy using a 200µl head
CV < 1 %, 5 μl to 180 μl (dry-well transfer)
CV < 5 %, 2 μl to 5 μl (wet-well transfer)
CV < 10 %, 0.5 μl to 2 μl (wet-well transfer)
Temperature Range
4°C - 35°C
Electrical
100 - 250 VAC, 50 / 60 Hz, 60 Watts
Dimensions (W x D x H)
600 x 515 x 470mm
Weight
45 kg
Relative Humidity
40 - 70 %
Noise emission
42 dba
PIRO® Manual, Version 1.5.2
14
1.13 Level Sensing
For maximum precision, the PIRO® is equipped with a liquid level sensor. The use of the
level sensor helps to prevent carryover of material on the outside of the tip. The liquid level
sensor is a very sensitive electronic device and is capable of detecting very small amounts
of liquid. The PIRO® level sensor relies on detection of pressure changes and allows the
robot to automatically detect when the pipetting tip is touching the surface of the liquid and
can therefore be used with non-conductive tips.
In a tip not touching fluid, no pressure can be measured. As soon as the tip reaches the
liquid level surface, pressure is built up. This build-up of pressure is measured by a pressure
sensor in the pipetting head, which then allows the pipetting head to lower the tip to a level
to aspirate the set volume. If a vessel is completely empty, a minimal z-axis value will be
recorded and an error message is displayed, allowing the user to add fluid to the tube.
In order for the above to work properly, there are two requirements for the tips: firstly, they
need to fit perfectly on the pipetting head and secondly, need to have a defined pore size of
the filter inside the tips.
1.14 The Computer
A personal computer running Microsoft Windows operating system is required to control the
PIRO® pipetting robot. The PC will be provided by DORNIER-LTF or the distributor. If a PC
is supplied by the user the following minimum specifications are required to run the PIRO®
software: Intel Core i3 2100, 2x 3.10 GHz or similar or more, 4 GB RAM, 250 GB Hard disc,
USB-drive, Graphic card: 1 GB memory, 1366 x 768 Pixel (widescreen), Monitor:
Widescreen. The software can be run with computers with lower specifications, but complex
calculations or processes might be slowed down, therefore influencing the graphical
presentation. In order to enhance the performance, the ‘use graphics card acceleration’ box
should ticked.
1.15 Connecting the Robot
To connect the PIRO® and the computer, please use the USB and the power cable supplied.
Connect the PIRO® to the power supply using the power cable socket at the rear of the
instrument. The USB connection on the PIRO® is found at the back of the instrument. Make
sure the PIRO® is switched off before connecting both the USB and power cables. The
PIRO® is now ready to be switched on.
1.16 PIRO® Software Installation
PIRO® Manual, Version 1.5.2
15
For error free usage of the PIRO® software, make sure that all other software is shut down
and that the screen saver and power save mode are disabled. Insert the software CD or
USB stick and double click on the setup.exe icon to start the installation process. When
installing the PIRO® software for the first time, the setup will check if the required Microsoft
.NET framework is installed. If there is no Microsoft .NET framework or an older version
installed, the software will ask to download and install it before continuing with the PIRO®
software installation. Make sure the computer is connected to the Internet. Click ‘Install’ and
wait until it is finished. After installation of the Microsoft .NET framework, the setup of the
PIRO® software continues. Follow the on screen instructions.
Welcome to the DORNIER-LTF PIRO® Setup Wizard: This screen starts the installation
procedure. Click ‘Next’ to continue.
Select Installation Folder: This screen allows the user to select the installation folder. Click
‘Browse’ to alter the folder the software is installed in. ‘Disk Cost’ shows how much disk
space is needed and how much is available on existing hard drives. Choose between
‘Everyone’ and ‘Just me’ to select access rights for the software. Click ‘Next’ to continue with
the installation process.
Confirm Installation: Once the installation settings have
been selected, the software is installed by clicking ‘Next’.
Selecting ‘Back’ will take the setup back one step. To stop
the installation select ‘Cancel’.
Installation Complete:
Once the installation is
completed, click ‘Close’
and the software can
now be started. It is not
necessary to reboot the computer before starting the
PIRO® software for the first time. If the PIRO® is switched
on, a PIRO® run can be performed. If the PIRO® is
switched off the software will run in virtual mode (see 2.3).
PIRO® Manual, Version 1.5.2
16
1.16.1 Updating the Software
Updating the software follows a similar procedure to the first installation. When the setup is
started, it will detect the previous version of the PIRO® software, which is automatically
removed before the new version is installed.
Select ‘Install’. A small window appears showing
that another version has been detected, click
‘Uninstall’ and wait until the process is done. Installation of the new software is performed as
described above.
PIRO® Manual, Version 1.5.2
17
2 Introduction to the PIRO® Software The PIRO® software has been designed to be easy to use. For most applications there are
multiple ways of working to make the software as intuitive and easy to use as possible. The
following chapters will give a detailed overview of the software.
The PIRO® software has been designed for general pipetting operations, but its initial focus
is the setup of PCR and qPCR reactions. The software supports a variety of PCR and qPCR
plate formats and it is being continually updated with new plate types. If for any reason a
plate format cannot be supported, please contact your distributor.
2.1 Downloads
2.1.1 PIRO® Software
If the software has not been delivered with the PIRO® instrument, it can be downloaded from
the following web page: www.dornier-ltf.com. In order to do so, the user has to register and
once approved, will be able to download the software. We recommend registering for the
website, as DORNIER-LTF will provide regular updates (e.g. new features, new plate
libraries, fixing reported bugs). It is the users’ responsibility to keep the software updated to
ensure the best possible performance of the PIRO®.
2.1.2 PIRO® Manual
The most up-to-date version of the PIRO® manual can also be downloaded from the above
web page, once registered. It is provided as a pdf and free of charge.
2.1.3 Other Documents
In the future it is planned to add application notes and other useful material to the restricted
user area. It is therefore worthwhile regularly checking this area for updates.
2.2 Starting the Software
To start the software, double-click the PIRO® icon. The software will open the
default project (see 4.1). If the computer is connected to the PIRO® and the
PIRO® is switched on, the software will show it is connected on the bottom
right of the main screen. If the PIRO® is not connected, then ‘Not
connected’ is displayed and the PIRO® software will run in virtual
mode (see 2.3).
2.3 Virtual Mode
The virtual mode option allows the user to run the software without a connection to the
PIRO®. This is a useful option as it allows running files or templates to be set up, saved, or
analysed on a remote computer. Files can then be transferred to the computer the PIRO® is
attached to and run.
PIRO® Manual, Version 1.5.2
18
3 The PIRO® Workspace
Shown above is the main area of the PIRO® workspace. This includes the feature tree (see
5.1), file menu (see 4.1), layout menu (see 4.2), options menu (see 4.3), and robot menu
(see 4.4). The feature tree is the top left window and contains detailed information about the
current run. When an item in the feature tree is selected, more details are shown in the
feature table which is located below the feature tree (see 5.2). The right side of the screen
shows the main workspace of the PIRO®.
3.1 Deck Layout
The figure below shows the empty deck layout of the PIRO®. The numbering starts in the top
left corner with number 1 and is followed through to number 16 on the bottom right corner of
the deck. Various racks and adaptors can be positioned on the PIRO® deck and these
options are discussed in more detail in the section below.
3.2 Racks and Adaptors
The PIRO® system is supplied with a standard selection of racks, but more racks and
adaptors are available. For further information, please check the customer area on our web
page or check the PIRO® software under ’Plates’ for an indication of available racks and
PIRO® Manual, Version 1.5.2
19
adaptors. All PIRO® loading racks are manufactured from stainless steel and/or high quality
aluminum, which is trowalised to give the rack a hard wearing finish. The racks are fully
autoclavable. To keep the racks cold, place the racks in a fridge or freezer before use.
DORNIER-LTF manufactured adaptors can be washed in bleach or soapy water. Please
note that some solvents or salts may stain the aluminum surface.
3.2.1 Orientation of Rack and Plates
PIRO® racks and plates have been designed so that they can only be placed on the deck in
the correct orientation using different sized holes on the bottom of each rack or plate.
3.2.2 Tip Racks
Tip racks can be placed in any suitable deck position. Right-clicking over the frame of any
plate will open a window allowing tips to be marked as available or unavailable, the tip type
to be changed, and removal of the tip rack.
Tip racks can be placed in the deck positions 1 to 12. However, in positions 4, 8, and 12,
only the first two rows of the tips can be reached by the pipetting head (see 3.2.6). Positions
13 to 16 are smaller and are not designed for use with tip racks.
3.2.3 Tips
The PIRO® has been calibrated for use with filtered tips from DORNIER-LTF and will
therefore guarantee the most accurate performance. For any tips, other than those available
in the PIRO® software, DORNIER-LTF will not take any responsibility with regards to the
performance. Tips not tested on the PIRO® might be too large, therefore allowing air to be
aspirated and inaccurate volumes pipetted or even fall off the pipetting head. It is strongly
recommended that only DORNIER-LTF tips are used with the PIRO®.
The software has options for 50 µl, 200 µl, and 1000 µl tips and all tips can be used with the
pipetting head 200. However, 1000 µl tips will only aspirate a maximum of 180 µl due to the
size of the pipetting head 200. If volumes up to 1000 µl need to be pipetted, then the
pipetting head 1000 needs to be used.
Before starting a run it is recommend indicating in the software the availability of the tips on
the deck. If the PIRO® however does not find any tip, it will move to the next tip position until
a tip has been found to continue the run.
3.2.4 Tip Disposal
The PIRO® has various options for the tip disposal: the external waste disposal through the
waste channel-1 or internal waste disposal into a waste rack.
PIRO® Manual, Version 1.5.2
20
3.2.4.1 The Waste Channel
The waste channel-1 is a metal tube
that leads to an opening located at
position 16 on the front right corner of
the PIRO®. Tips ejected into the waste
channel-1 will fall out of the bottom of
the PIRO® and can be collected externally either by placing a waste
container such as a bin or draw below the waste channel-1 or attaching a bag to the
underside of the waste channel-1 to catch the tips. DORNIER-LTF also provides a special
PIRO® table which has a built in section for collecting tips ejected through the waste
channel-1.
3.2.4.2 The Waste Racks
Tips may also be discarded into a waste rack positioned on the deck. The waste rack is equipped with a lid.
The metal waste rack is lined with a bag and can be discarded at the end of each run.
3.2.4.3 The Waste Console with Drawer
A further option to manage the tip waste is a
drawer positioned below the PIRO®. The console
is equipped with two drawers for the collection of
waste in the drawer at the right side and storing
accessories in the drawer on the left side. For
more details of the console please contact your
distributor.
PIRO® Manual, Version 1.5.2
21
3.2.5 Racks and Adaptors
Positions 1 to 12 are designed to hold SBS format racks. This include tip racks, 96- and 384-
well plates, adaptors, waste boxes, and many special plate formats (e.g. Eco-plates,
SmartCycler, Rotor-Gene plates).
3.2.5.1 Plate Racks
Three different SBS standard racks are shown above. From left to right one tip rack, Plate
Rack-1, SBS Grid 96 Wells, and Plate Rack-1, SBS Standard are shown.
Tip racks take all DORNIER-LTF tips. The standard and 96-well plate racks can
accommodate any SBS plate that fits within the footprint. The 96-well plate racks can also
take 96-well unskirted plates.
3.2.5.2 Adaptors
Adaptors are designed to fit on the 96-well and universal plate racks or other racks provided.
DORNIER-LTF provides a range of adaptors, such as the Adapter SBS 32 Well.
Adaptors are also available from alternative manufacturers such as Illumina (Eco-plate) or
Qiagen (33-well plate) which take 32 x 1.5 ml tubes. Additional plate formats are regularly
added to the PIRO® software. Once registered, software updates are available from
www.dornier-ltf.com.
3.2.5.3 Overview of Racks and Adapters
For an overview of on how racks, adapters, and plastic consumables fit together, please see
the document “Rack Adapter Compatibility List” located on the customer area of our web
page.
PIRO® Manual, Version 1.5.2
22
3.2.5.4 Tube racks (Positions 13 to 16)
Positions 13 to 16 at the bottom of the PIRO® deck are smaller than positions 1 to 12 and
have been designed to hold tube racks. Tube racks available take a variety of different
volumes like 0.2, 0.5, 1.5, 2, or 5 ml tubes. Other tubes and formats will be added to the
PIRO® software over time, as requested. For instructions on downloading PIRO® software
updates, please see 2.1. Some of the tube racks (TR) available from DORNIER-LTF are
shown below:
Tube racks are used to hold tubes containing diluents, reagents, mixes, standards, and
samples on the PIRO®’s work space.
If required, tube racks could also be set up on position 1 to 12 of the deck to allow for more
flexibility of using plates on different positions.
At position 16, there is a hole in the deck. Position 16 can be used in
conjunction with the waste channel-1 and allows used tips to be ejected
outside the PIRO® during a run. The PIRO® front feet have been
mounted about 15 cm inwards, allowing for the PIRO® to be placed
over the front of the table so a waste bin can be positioned below
position 16 of the deck. If waste tips are being collected in a waste rack
during a run, then position 16 can be used to hold a tube rack instead.
3.2.6 Plate Positions 4, 8, 12, 16
The PIRO® has 16 positions available on the deck. The availability of
positions 4, 8, 12, and 16 is limited due to the space needed for
positioning of the pipetting head. The light grey area shows the area
available for use, the dark grey area indicates where the pipetting head
is unable to reach. These positions can be used for the waste rack and
also for SBS racks where only a few positions are in use.
TR-28; 0.2 ml tapered
TR-5; 5 ml
TR-8; 2 ml V-bottom
TR-8; 2 ml free standing
TR-8; 1.5 ml tapered
TR-10; 0.5 ml tapered
PIRO® Manual, Version 1.5.2
23
3.2.7 Tubes
Many different types of tube can be used on the PIRO®. Some of the tubes that are
compatible with the PIRO® are shown below, along with the racks required to hold them:
3.2.8 Setting up Plates in the Software
Plates are added to the deck by right clicking on the arrester on the deck
position. A window will open containing multiple options. Hovering over ‘Add
plate’ opens a window allowing to select from ‘tip racks’, ‘plate racks’, or ‘tube
racks’. Hovering over each of these options brings up a different catalogue of plate and rack
types. Clicking on a plate or rack will add it to the deck.
TR-28; 0.2 ml tapered TR-8; 2 ml V-bottom
TR-8; 2 ml free standing
TR-10; 0.5 ml tapered TR-5; 5 ml tube
TR-8; 1.5 ml tapered
PIRO® Manual, Version 1.5.2
24
Details of the selected plate can be found by clicking on the plate (outside any position
indicating a well). The information appears in the feature table on the left side of the screen.
The ‘Properties’ tab contains details about the selected plate and the ‘Positions’ tab contains
information about the position, function, content, and other options available for each well on
the selected plate. Plates can also be added to the plate deck using the toolbar. Detailed
information on using the toolbar to add plates to the deck is given in the below chapter (see
4.2.2).
3.3 Position Offsets
The function ‘Position Offsets’ allows
the user to change the position of the
pipetting head relative to the plates or
tubes used. This is generally not
necessary if the exact plate described
in the software is used, but if a similar
plate is used, this option allows
adjusting the x, y, and z-values of the
pipetting head.
Having this option is useful for various
reasons:
Defining x- and y positions inside
the vessel for dispensing liquid,
for example closer to the wall
Defining z limitations for the tip if
a contact to a sediment should be
avoided
Defining z limitations to reduce
dead volume
Similar plates to what is in the plate library could be calibrated accordingly
PIRO® Manual, Version 1.5.2
25
In order to calibrate a plate, chose the plate where the adjustment should be done. Press the
‘real-time processing’ button and adjust positions by pressing the ‘up’ and ‘down’ arrows in
the x -, y-, or z offset tick boxes, depending on the direction it needs to be changed. Positive
and negative values could also be typed into the appropriate boxes. Repeat the processes
until the values are correct and the tip is above the desired position of the well. Press the
‘real-time processing’ button again to leave the ‘Position Offsets’ feature.
If plates are similar but not exactly the same, the depth of a well might be different compared
to the plate used. Similar to above, change the z values to move the head up or down. Make
sure the tip does not touch the bottom of the well, as otherwise the performance would be
influenced.
Following the adaptation of the x -, y-, or z values, the offset values are changed. To indicate
that a plate has adjusted x-, y-, or z-values, the plate now has a yellow rectangle in front of
the plate description in the feature tree (see red circle).
In order to use the plate with the adjusted values again, save the file as a template and use
the template. Be aware that the settings are only valid for the position of the plate where the
values have been adjusted for. If the same original plate is chosen from the plate library
again, the values need to be adjusted for this particular plate at its actual position again.
Moving a plate with altered x-, y-, or z values to another position will maintain the changes at
the new position. This means that position offset changes are attributes of the appropriate
plate.
It is also possible to perform these adjustments while the robot is running. The changes will
affect the adjustments immediately after pressing ‘Enter’, so any change can be observed
during the pipetting process.
WARNING: Please take caution not to introduce position alterations that might cause tip
crashes next to or inside the wells.
3.4 Plate Functions
Plate functions can be chosen independently of the plate and plate position. It is therefore
possible to have several different functions (e.g. reactions, samples, mixes) on the same
plate. This has various advantages; the speed of a run can be increased by choosing
positions close to each other, the transport ways can be optimised, and storage for future
runs can be facilitated.
3.5 Exchanging Existing Plates
The PIRO® software allows the plate type to
be changed even when the plates are in
use. This can be useful if the wrong plate
has been selected. This option can also be
used to change from 96- to 384-well plates
but not to plates where the number of wells
in the plate of origin is higher. In order to
change a plate, open the plate menu and choose a plate to replace the existing plate.
PIRO® Manual, Version 1.5.2
26
4 Main Menus
4.1 File Menu
The file menu tab is used to load and save projects as well as to import and export samples
and reactions.
New: Selecting ‘New’ enables opening a
new ‘Default project, ‘Empty project’, or
‘Template’. If no default project has been
saved then this option will be unavailable.
‘Default Project’ loads a project which has
been saved as a default project. Any
project may be saved as a default project.
Besides the deck layout and transports,
the default project will also save other settings like multi-
dispensing, level sensing etc. An ‘Empty project’ opens a file with
the deck completely empty and ‘From Template’ opens a
previously saved template.
Open Project: Selecting this function will open a previously saved
project.
Save Project: This function will save the current setup as a
project file. Choose a folder where to save the project and press
Save. Once saved, another click on ‘Save Project’ will overwrite
the previous project file.
Save As: A project can either be saved as a ‘Default project’ or a
‘Template’. Selecting ‘Save as’ opens a new window in which the
project file is to be saved in and can be selected or altered.
Template: Opening a template will open the previously saved template. Templates can only
be altered in a limited fashion, which ensures that runs can be repeated in controlled
conditions. There is no limit on the amount of project or template files that can be saved.
Project: A project contains all deck definitions (e.g. sample, mix, reaction setups) as well as
all defined options (e.g. multi-dispensing, level sensing) and all included settings. A project
name will be created automatically before executing a run. Once saved it can be easily
reopened and used for the same or similar runs.
For Import and Reaction Export please see 6.5.7 and 6.6.5, respectively.
Recent Files: This allows the user to quickly open
any of the last ten files that have been run on the
PIRO®. Files which have been saved, but not run do
not appear in the ‘Recent files’ list.
PIRO® Manual, Version 1.5.2
27
Exit: Selecting ‘exit’ closes the PIRO® software. If any changes have been made, the
software will ask to save them. Click ‘Yes’ to save, ‘No’ to exit or ‘Cancel’ to leave the
software open.
Right mouse clicking over any of the main
menu options will open the window shown on
the right.
Add to Quick Access Toolbar: Tabs will be added to the toolbar below the main tabs.
Show Quick Access Toolbar above/below the Ribbon: Will move the toolbar either above
or below the ribbon.
Minimize the Ribbon: Will minimize the ribbon; the ribbon can be restored by right clicking
in the menu ribbon.
4.2 Layout Menu
4.2.1 Clipboard
Paste, Cut, Copy: As the names say, the ‘Paste‘, ‘Cut‘, and ‘Copy‘ functions allow the user
to paste, cut, and copy. These functions can be used with the reagent
and mix function. This allows additional mixes to be created quickly by
copying and pasting mixes that have already been created and then
altering the content of the mix. This is particularly useful for users who create runs with many
mixes that contain only slight variation (e.g. magnesium chloride, primer, probe, or DNA
template titrations).
4.2.2 Tools
Pointer: The pointer is the default setting and allows free movement around the PIRO®
screen. The pointer function can be selected by clicking on ‘Pointer’ in the tools
menu or by double clicking in a grey section of the deck.
Delete: This option is used to delete plates from the deck to which no function has been
assigned. It is also used to delete the function of the selected area on a plate. If a
transport has been assigned to the function to be deleted, an error message will
appear indicating that the transport needs to be removed before the function can be deleted.
Move: This option allows plates to be moved from one position to another. Plates can be
moved even if various functions have been assigned to positions within the plate.
In addition, reactions, mixes, reagents, and diluents can also be moved, as well
as the whole deck for a better overview. Samples cannot be moved. To deselect the move
cursor function, press ‘Move’ again or click the ‘Pointer’ icon.
PIRO® Manual, Version 1.5.2
28
Undo, Redo: Those options allow the user to undo or redo any action taken during the run
setup. This is particularly useful if previous actions need to be checked or
corrected. This function can be used to go back to the very first change made
after opening the file. Even if the file has been saved in between the setup, the
undo button will still undo all the actions taken right back to when the file was initially
opened.
Plates: Selecting this function opens a catalogue of all the plates and racks available in the
PIRO® software. Plates will be continually added to the software and the latest
version of the PIRO® software, containing all the plates currently supported, can be
found in the restricted area of our web page: http://dornier-ltf.com/. The
plates menu can also be accessed by right clicking over the circle in the
main deck layout (for more information on selecting plates, please see 3.2).
Functions: In order to use the ‘Functions’ button, empty wells without
functions have to be selected. Once selected, the ‘Functions’
icon can be used to display a list of different functions that can
be assigned to the wells. The figure on the right side shows all
function options available.
4.2.3 Zoom
Zoom: The Zoom function allows the user to zoom into an area of interest and can be done
with the buttons shown or with the mouse wheel. This is particularly
useful when setting up 384-well plates or similar plates with small
distances. Pressing the ‘Fit to screen’ button restores the screen to
its original size.
4.2.4 Extras
Show Transports: Using ‘Show Transports’ shows the transports of the selected wells.
Different arrows are displayed, depending on the type of transport. A thin arrow
is shown, if the transport has been set up from one tube to another tube (one-
to-one). A thin arrow at the beginning with a thick ending indicates a transport
from one tube to many tubes (one-to-many). A thick arrow at the beginning and
the end will show a group transport. Many thin arrows ending up in one tube shows setup of
a mix or many diluents combined in one well. For more details on transport, please also see
5.4.
PIRO® Manual, Version 1.5.2
29
Show Plate IDs
In order to give an ID to a plate choose ‘Properties’ as shown above and fill in the orange
field as shown below. This can be done manually as well as using a bar code reader (see
6.5.9), so each plate can have its own ID.
In order to see the ID on the screen, use the ‘Show plate ID’ from the menu.
Zip Log File and Project
Using the “Zip log file and project” function will automatically zip the project and log file which
can then be saved anywhere on the computer. This might be important for analyzing files.
4.3 Options Menu
4.3.1 Language Options
The software is designed to be used in many languages. Using the drop down menu shows
what languages are available and any user can choose the language suitable. At this point
of time, the English, German, and Russian languages are implemented, but this will increase
over time. If an alternative language is required, please contact your distributor to see for
options.
PIRO® Manual, Version 1.5.2
30
4.3.2 Robot Options
Use Multi-Dispensing
Turning on the multi-dispensing option enables multiple dispenses
to be performed from a single tip, saving both time and tips. If ‘Force
Tip Change’ is also selected, ‘Force Tip Change’ will take precedence over the multi-
dispensing option.
If the multi-dispensing function in the options menu is turned on, the PIRO® software also
has the option to turn off multi-dispensing function for selected plates. Select the plate where
the multi-dispensing options should be turned off, untick the box in the ‘Properties’ window
as shown below (orange circle) and no multi-dispensing will be done on this plate.
Enable Level Sensing
The PIRO® is equipped with a real-time liquid level sensor. When this
option is selected, the robot automatically locates the surface of the
liquid. We recommend that the level sensing option is enabled before starting a run. For
more details see 1.13. As shown above, level sensing for selected plates can also be turned
off if the level sensing has been enabled.
Level Sensing each (times)
This allows for choosing the amount of times the level
sensing can be performed when pipetting any new tube.
Typing in the number ‘1’ will perform level sensing each time, using e.g. ‘3’ will perform the
level sensing every third time.
Force Tip Change
With this option selected, tips will be changed after every
operation. When ‘Force tip change’ and ‘Use multi-dispensing’
are selected, ‘Force Tip Change’ has priority over multi-dispensing (see above). If mixes are
distributed to empty wells, which is by default done with the same tip and ‘Force tip change’
is enabled, the mix will be distributed with changing a tip every time. This could be of
advantage when wells already contain liquid and cross contamination should be avoided.
PIRO® Manual, Version 1.5.2
31
UV Program Duration
This option allows the user to choose the time in minutes
how long a UV program should be performed. The default
and recommended time is 15 minutes. For further details please see 1.5 and 4.4.3.
HEPA Filter
To active the HEPA filter press the box ‘Activate HEPA filter’. Once a
PIRO® run is started the HEPA filter will then be activated.
4.3.3 Graphics Options
Use Graphics Card Acceleration
This option allows for computers with lower performance to
be optimised. In order to do so, tick the box above. For
computers with specification mentioned above (see 1.14), this box should stay unticked,
otherwise an incorrect presentation might occur.
4.4 Robot Menu
4.4.1 Robot Controls
Stop
Will immediately stop and abort the run.
Pause
Pressing the pause button pauses the run. A window appears that informs the user how to
continue. This window also allows for the option to move the pipetting head to the home
position by ticking the box on the bottom left
corner of the window.
The run can also be paused by lifting the lid of the
PIRO®. When the lid is opened, the robot will
finish the operation it is currently performing
before pausing the run and moving to the home
position. Once paused, the run can be restarted
by pressing ‘Resume’.
The pause functions allows for various options, tips could be reloaded (see 3.2.2) and
changes would need to be marked in the software or plate position offsets could be adjusted
(see 3.3).
PIRO® Manual, Version 1.5.2
32
Start
Choosing ‘Start’ initiates the run to be started. Before a run is started, the following window
containing a checklist will appear:
Various colours are used to display potential issues arising from starting the run. Information
displayed in red indicates an error that needs to be corrected before the run can be started.
An error could be that no waste box or waste channel has been selected. The tick box
before the red information cannot be ticked. Orange warnings show potential errors, for
example if a reaction does not contain a mix. This tick box can still be ticked and the run
started. Black information show for example if a sample ID has not been given and if not
needed the boxes can be ticked and the run started. In the same window a Pre-Run Report
can be accessed showing all the operation to be done in the run set up. ‘Start Robot’ will
finally start the run. All boxes must
be ticked before the run can be
started.
4.4.2 Report
Show Post-Run Report
Using ‘Show Post-Run report’
creates a ‘Post-Run Report’ with all
details of the run. In order to save
the ‘Post-Run Report’ open the
report and select ‘Save file’, choose
a folder and click ‘Save’.
The user can define the file access via document security.
Document Security: Filling in a password will protect the document from any unauthorized
access. The created pdf file can only be opened with the defined password.
PIRO® Manual, Version 1.5.2
33
Document Info: Information about
the run can be entered here, if
needed.
Output Range: Defines the output
of the document.
Document Restriction: The author of the pdf
document can restrict further handling and
editing in a pdf viewer or editor (e.g. Adobe
Reader, Acrobat etc.) by entering a password in
the document restriction box.
The document restriction options that can be
selected are:
Allow printing
Allow content copying or extraction
Allow content editing
Allow annotations editing
If the document security and document restriction passwords are the same, then the
document restrictions will automatically be listed when the document is opened. Choosing
separate password for document security and restrictions will prevent this from happening.
4.4.3 UV Program
UV Program Start
To start the UV-radiation procedure, choose the time how long the procedure
should take (see 4.3.2). Press the icon shown on the left and start the UV
procedure. For further information, please also see 1.5.
PIRO® Manual, Version 1.5.2
34
4.5 Deck Window
Many of the PIRO® functions features can be accessed by right-clicking over the main deck
area. The deck window is shown on the right and cross references to other way to access
the feature are given.
Project: See 4.1. New
projects can be opened,
either empty, default, or
others.
Zoom: See 4.2.3.
Cut, Copy, Paste: See
4.2.1.
Show Transports: See
4.2.4.
Add Plate: See 4.2.2. This
function is similar to the
plates function in the
toolbar. However, only
plates that can be added to
the position selected are
shown.
Remove Plate: see 4.2.2.
When a plate is selected,
this option can be used to
remove the plate.
Add Function: See 4.2.2.
The additional greyed out
options shown in the deck
window are explained in the
appropriate chapters which
are the options about
samples, fixed reaction,
standards, and
normalisation.
PIRO® Manual, Version 1.5.2
35
5 Feature Tree and Feature Table
5.1 Feature Tree
The feature tree is the window located on
the top left corner of the main screen. It
features and summarizes the complete
setup of a PIRO® run. It also shows
symbols e.g. information, warnings, and
errors, as explained below.
A general function when highlighting any of
the main headings in the feature tree e.g.
samples, mixes, standards, is that at the
same time, the appropriate wells are
highlighted on the screen.
The ‘sample feature tree’ refers only to the
main heading and does not include the
subfolders, for example where the sample
groups are located. This is also the case for
mixes, reagents, standards, etc., see figure
on the right.
The explanation of symbols in the feature
tree is given below:
5.1.1 Triangles
Triangles appear in two colors: when white, more information can be found
under the triangle; when black, no extra information can be found by clicking on
the triangle. Clicking the black triangle next to the project heading will close the all folder
submenus including plate, tips, and waste folders. This allows for a better view of the actual
setup of the project.
5.1.2 Info
The info button appears in front of a task to indicate that there may be some information
missing from the task. If the user
decides the missing information is
not needed, the run can be started
regardless of the missing
information.
PIRO® Manual, Version 1.5.2
36
5.1.3 Warning
A warning sign is displayed
when tasks are missing
which are usually required to set
up a PCR run. In the displayed case, no mix has been added to the reaction. However, it
could be that a project has been designed not needing a mix, e.g. the run is simply
transferring liquids from one plate to another. In order to allow for such tasks, a warning
message is displayed for information only and the run may still be started.
5.1.4 Error
The error symbol indicates
that there is a problem with
the setup and prevents the run
from being started until it has
been fixed. Errors can include a
defined mix to which no volume has been added, a missing transport or any other serious
issue.
Hovering over the main mixes heading will only indicate that there is an error, the actual
error message will then be found in the mix subfolder with the error warning displayed in
front of the mix that has the error associated with it. In the above example it is mix 2. Once
the error has been fixed, the error message for this particular mix will disappear. If this is the
only mix error, then the error symbol in the main mix heading will also disappear, if there are
other mix errors then they must be fixed before the error message on the main mix heading
will disappear.
5.1.5 Add
The add symbol appears for example when sample groups should be added to
existing sample groups. This symbol is only used in the window when right-clicking
over a main feature tree item.
5.1.6 Remove
The remove symbol appears when right-clicking over a submenu of the main feature
tree and allows sample groups, mixes etc. to be removed from the feature tree.
5.2 Feature Table
The ‘feature table window’ is located below the feature tree and displays information
according to the task chosen in the feature tree. Described below are general features used
in various feature tables. Well position description, changing of names, and viscosity are
used in several feature tables, detailed descriptions of all other feature tables can be found
below.
5.2.1 Well Position Description
In any feature table a well position will be found but cannot changed. Well positions can only
be changed in the table layout. The first number of the well position description indicates the
plate position on the deck (see 3.1) and the second number gives the position of the well on
the particular plate. A position description 6.A5 therefore means plate 6 and position A5.
PIRO® Manual, Version 1.5.2
37
5.2.2 Changing Names
The name of any well with an assigned function can be altered in the feature tables. The
default names are the function followed by a number. To change a name, simply double-
click on the name and type in a different name. In some feature tables there is also an option
to choose names from a drop down menu.
In order to avoid confusion, the PIRO® software does not allow a mix and a reaction to have
the same name. A suggestion would be to add an ‘M’ (mix), ‘R’ (reaction) or a number
following the name, so the names could be clearly differentiated.
5.2.3 Viscosity
The PIRO® software allows
the robot pipetting speed to
be altered. Highly viscous
fluids containing glycerol or
other viscous substances
should be pipetted more slowly for optimal accuracy when pipetting.
The standard viscosity setting is ‘Normal’, which will pipette most liquids accurately. If a
liquid contains a high amount of alcohol (over 50 %) then the ‘Low viscous’ setting should be
used. Liquids containing glycerol as a
concentration of 20 % or greater should be
pipetted using the ‘High viscous’ setting.
5.2.4 Mixing Options
The PIRO® software allows the user to define
the mixing options. Various options in the
different menus are available either by
aspiration or dispensing. By typing a number
in the appropriate field tells the software how
often the fluid will be mixed. Shown in the
figure on the right is the mixing of reagent 4
and 3. Both are mixed two times on aspiration
and dispensing. In addition, the mix menu
also allows for a mixing option at the very
end, so the created mix is mixed properly
before dispensed in the reaction.
Mixing can also be performed
choosing the ‘Options’ after all
reagents have been added to the
mix.
PIRO® Manual, Version 1.5.2
38
5.3 Plate Submenu
5.3.1 Plates
The plates feature tree gives information on all the
plates on the deck, including the tip racks. In the
example shown is a 50 µl tip rack named ‘96-tips Rack
50 µl’, the capacity of the rack is 96 and the availability
is 87 out of 96 tips. More details on the plate including
name, category, number of positions and more can be
found by clicking the plate of interest. The plate
feature table will open showing the plate details.
Once plates are set up and checked, it is
recommended to close that feature tree for a better
overview of the other processes.
5.3.2 Tips
The tips feature tree indicates the total
number of tips available in the current setup. The tips
feature table displays an overview of the tip type, total
capacity of tips, availability, usage, and the remaining
number of tips on each tip rack.
5.3.3 Waste
Clicking ‘Waste’ on the main screen will indicate where a waste channel-1 or a
waste box is positioned. There is no waste feature table available.
5.4 Transports
This option in the feature tree allows the user to check all transports set up in a particular
run.
The ‘Commands’ tab in the transport
feature table displays a summary of all the
actions set up in a particular run. The
number in the first column indicates the
order in which the action is performed the
second column gives the name, the third
column the ‘Type’ of action to be taken, and
the fourth column gives the details of the action. The fifth column gives the volume and the
sixth column how often this task I performed.
PIRO® Manual, Version 1.5.2
39
5.4.1 Changing the Order of Transports
While setting up a transport as well as after all transports have been setup, the order of the
transports can be changed.
As an example a run is setup with
two sample groups, two fixed
reactions, and two mixes. The
current order is Mix 2, Mix 1,
Sample group 1, and Sample
groups 2. The new order should be
as follows: Mix 1, Mix 2, Sample
group 2, and Sample groups 1 (see
the screenshots on the right).
In order to achieve this, choose ‘Transports’ in the feature tree, press the tick box ‘Enable
sorting the transport list manually’ and the box is highlighted in red.
To change the order of the
transports, select ‘Manually sorted’
and use the ‘Up’ and ‘Down’ buttons
to move the transports up or down.
Save the project or start the run and
the transports will be performed as
indicated in the ‘Manually sorted’
table. There will be no changes in
the ‘Transports’ table, as the
‘Manually sorted’ table has priority
over the ‘Transport’ table. The
‘Commands’ will list the transports
accordingly and cannot be changed.
Special Commands:
This option allows adding special
commands, which currently includes
a wait and a reset function. To
access this option press ‘Add’ and a
window appears to specify the parameters. Once done, the wait function has been included
in the list and can be edited or removed as the buttons are active.
PIRO® Manual, Version 1.5.2
40
A further special command is a function called ‘Reset reaction contents’. This function will
“forget” previously performed pipetting steps and assumes the reaction vessel is now empty.
This function makes sense to use when for example a first mix has been pipetted into the
reactions. The user now wants to add a second mix in the same reaction but wants to allow
for the multi-dispensing option to be performed saving tips. Normally the PIRO® software
would not allow this task, as there is already one mix in the reactions, therefore potentially
contaminating the reactions. Using the ‘Reset reaction contents’ function tells the PIRO® that
the reactions tubes are “empty” and will therefore allow for multi-dispensing. Using this
function will not reset sample pipetting operations.
Add new Transports: When adding new transports in the run, the details in the ‘Manually
sorted’ mode will be shown in orange and italic. Use the ’Update’ and ‘Confirm’ buttons to
finally include those reactions into the transport list.
PIRO® Manual, Version 1.5.2
41
The reactions can now be resorted again and the final reaction list looks like as shown
below:
When changing from ‘Transport’ to the ‘Manually sorted’ option all automated safety rules
build into the software will be overruled, as no feasibility checks are done. It is the users’
responsibility to ensure plausibility of the new transport order.
PIRO® Manual, Version 1.5.2
42
5.4.2 Setting up Transports
Brief overview: Setting up transports using the PIRO® software always follows the same
principle. Highlight the wells to be used, assign a function, and then drag the selected wells
into the destination required.
In detail: Select the wells, assign a function (see 4.2.1), and drag them to a new area. In the
case shown above, positions 9.1 to 9.6 have been assigned as samples and dragged to
plate 10. The software automatically assigns the corresponding number of wells in plate 10
to the reaction function.
In position 9.23 to 9.28, a reagent function has been assigned by the user and reagents
have been dragged into the mix position 13.3.
To set up a diluent, select the well, choose the dilution function, and drag the diluent into the
mix. The mix can now be dragged into the reaction. Once the mix has been dragged to the
reaction a one-to-many arrow will appear, showing the mix has been transferred to the
reaction area. The last figure on the bottom right shows two arrows.
The lower arrow shows the transfer of the mix to the reaction (one-to-many) and the upper
arrow shows the transport of the samples into the reactions. The upper arrow is a one-to-one
arrow, so each sample is transferred into a different well. For more details on transport
arrows, see 5.4.1.
PIRO® Manual, Version 1.5.2
43
5.4.3 Auto Assignment of Transports in Empty Wells
The PIRO® software has several automated procedures which facilitate the setup of
transports. When a function has been chosen but no function of the target wells has been
defined, the PIRO® software will auto assign a function for the user, saving steps during the
setup of the transports.
Empty well: If an empty well or empty wells are chosen and moved, a sample function is
created in the well of origin and a fixed reaction created in the destination well.
Samples: If a sample or samples are dragged into an empty well, a fixed reaction will be
created according to the number of samples moved.
Reagents, Standards: If a reagents or standards are highlighted and dragged into an empty
well, a mix function is created and all reagents are transported into this mix.
Diluents: If a diluent or diluents are highlighted and dragged into an empty well, a mix
function is created and all diluents will be moved into this mix.
Mix: If a mix or various mixes are highlighted and dragged into an empty well, a fixed
reaction is created and the mix or mixes transported into this well.
Reaction or Reserved: If any of these three functions are highlighted and moved to an
empty well, no function in the target is created and a ‘forbidden’ sign will appear, indicating
that this is not a predefined setup option.
PIRO® Manual, Version 1.5.2
44
5.4.4 Function Colour Coding
The PIRO® software assigns a different colour to each function making it
easy to identify the function of individual wells.
Reagent PURPLE
Mix ORANGE
Diluent BLUE
Sample YELLOW
Reaction RED
Standard GREEN
Reserved GREY
No function WHITE
Examples of each function are shown below:
An empty circle: the well is empty.
An empty circle with colour on the outside:
a function has been assigned to the well
A grey filled circle with colour on the outside:
the well has been assigned both a function and a
transport.
A highlighted circle: the well has been selected.
A grey filled circle with colour on the outside
and a highlighted circle: a function as well as a
transport has been assigned and the reactions are highlighted.
An inner white circle: in addition to the functions
described above, the well has been selected in the
feature tree.
A line through the grey circle: Normalised sample
A line at the outside: Diluted sample, dilutions steps
are indicated by the amount of lines
Grids: Automatic Tube Switching
5.4.5 Transports and Selecting Wells
When wells should be selected for assigning a function,
position the cursor outside the first well to be used and
press the left mouse button. Move the cursor over the
wells to be used, release the left mouse button and the
wells are selected. If additional wells should be added
to the selection, use the ‘Ctrl’ button to select further wells.
In order to transport a selection, press inside the
selected well and move the selection to the position
required. The coloured outside of a well cannot be used
for the transport, as it is simply indicating what kind of
function is assigned to the particular well.
PIRO® Manual, Version 1.5.2
45
6 Setting up a Run
This chapter gives details on how to set up a PIRO® run, using diluents, reagents, standards,
mixes, or samples. The feature tree and feature table are explained of each of the above
options. It also describes on how to set up templates for intelligent pipetting, perform cherry
picking of samples, or set the orientation of the plates. Serial dilutions and the normalisation
function are explained as well as the special functions like automatic tubes switching or
automatic fluid transfers.
6.1 Diluent
A diluent is usually used to dilute mixes in order to make them up to the final reaction
volume. Water is often used as the diluent, but the PIRO® software
allows other diluents to be used and multiple diluents can be assigned to
a run. This can be particularly useful when developing an assay, as it
allows multiple grades of water or different buffers to
be tested in one run.
The diluent feature table displays the well position, name, and volume of
each diluent.
6.1.1 Setting up Diluents
Select the wells to be used as diluents, and right-click over the selected
wells. Select ‘Add Function’ and choose ‘Diluents’. Diluents can also be
assigned using the function button in the layout menu. The diluent will
now be marked in blue.
6.2 Reagents
Selecting ‘Reagents’ in the feature tree highlights all the wells on
the deck, where reagents have been assigned. In this example,
ten reagents have been selected and three are shown in the
table. The number following ’Reagents’ indicates the number of
reagents set up in this run file (10). This does not necessarily
mean that all reagents are used in the run and no warning or
error message will be displayed if wells have been assigned as
reagents, but are unused.
As with any other function, reagents can be located in any position on the PIRO® deck. In
most cases, reagents are stored and used in 1.5 ml to 5 ml tubes and are therefore suitable
to be used in tube racks located from position 13 to 15. Reagents can also be assigned to a
32-well plate, which takes 1.5 ml tubes or any other suitable plate. The reagent table shows
the location (well), the name, and viscosity of each reagent (see 5.2.3).
PIRO® Manual, Version 1.5.2
46
6.2.1 Setting up Reagents
In order to set up reagents, select the wells to be used for reagents, then use the function
button in the layout menu (see 4.2) or right click over the selected wells and choose
‘Reagents’.
In this example, five
reagents on plate 9 have
been selected. To create a
mix using all five reagents,
simply select the reagents
and drag them into the
desired well. In this
example position 13.3 was
used and the PIRO®
software automatically
assigns this well as a mix.
The result will look like as
shown in the figure above
on the right.
Once the reagents have
been transferred to a mix,
the mix feature table will
open and the volume of
reagents required for each
reaction needs to be
specified. Once the
volumes have been
entered, the reagent setup
is completed.
Add/Remove: The ‘Add’
and ‘Remove’ functions
below the ‘Content of this
mix’ table can be used to
add or remove reagents
from the mix.
PIRO® Manual, Version 1.5.2
47
6.3 Mixes
The ‘mix
feature tree’
displays the
main heading of the mixes
as well as the submenus,
which are the various mixes
used in this project. The
number following ‘mixes’
indicates how many mixes
have been set up. Selecting
mixes in the feature free will
highlight all mixes set up on
the deck. Selecting one of
the mixes in the submenu
will highlight that particular
mix on the deck, as well as
all transports set up if the
transport button has been
activated. The number in the
submenu following each mix
will indicate how many
reagents have been used to
set up this mix.
Mix Submenu: This menu
allows for the name (see
5.2.2), ID, and viscosity (see
5.2.3) to be chosen. The
‘Volume added to the
reactions:’ means the volume
added to the reactions not
including the volume of the
sample. The sample volume
and the ‘Volume added to the
reactions:’ together is the
final reaction volume. ‘‘The
content of this mix’ shows the
reagents and diluents
included in this mix. Names
and volumes can be chosen
as well as the mixing option
for that particular mix (see 5.2.4).
Volume Calculation Submenu: This menu allows various volume calculations to be
selected. ‘Create Exact Volume’ allows the user to create a mix volume that is independent
PIRO® Manual, Version 1.5.2
48
of the volume calculated by the reaction set up. This option does not calculate if sufficient
volume to complete the run has been entered. ‘Calculate Volume from Reaction Count’
automatically calculates the volume of reagent required based on the reactions that will be
performed during the run. ‘Volume is Predefined / Premixed’, allows a mix to be used which
had previously been made either by the PIRO® run or manually.
Options Submenu: At this point of time the only option available under ‘Options’ is ‘Mixing
at end’. This allows the user mixing the mix created by the PIRO® after all reagents have
been added.
If ‘Volume is Predefined / Premixed’ is enabled, then in ‘Options’ the following window is
enabled, which allows ‘Mixing prior to usage’.
6.3.1 Setting up Mixes
Select a mix by assigning a mix function to a well, as for example shown below, see 6.6.1.
Drag the diluents and reagents into the well. Clicking Mix 1 in the feature tree displays the
transports in this project. To complete the setup, enter the ‘Volume added to Reaction’ and
the various diluent and reagent volumes to the feature table.
The mix is completed when no error message appears in front of the ‘Mix 1’ of the feature
tree, as shown above right.
PIRO® Manual, Version 1.5.2
49
6.3.2 Pipetting Mixes into Mixes
In most cases a mix is created and distributed in various reactions. But there are some
cases where the mix should be transferred in a further mix or mixes prior to the distribution
of the final mix into the reaction vessel. Shown below is an example and the way to set this
up in the PIRO® software.
On plate 9, 3 x 4 reagents have been set up. Reagents 1 to 4 have been transferred into mix
3 on plate 13, reagents 5 to 8 have been transferred into mix 2, and reagents 9 to 12 have
been transferred into mix 1. By simply dragging mix 3 into mix 2 and mix 2 into mix 1, the
final mix (mix 1) to be transferred into the reaction vessel has been created. Mixes do not
have to be prepared in a previous run in order to mix them with other mixes.
6.3.3 The Pipetting Order
The PIRO® has been assigned rules regarding the order of pipetting. When setting up a mix,
the largest volume will be pipetted first. If several reagents in a mix have the same volume,
the order will depend on the order in which the user entered them into the mix. For changing
the pipetting order, please see 5.4.1.
PIRO® Manual, Version 1.5.2
50
6.4 General Orientation on Plates
If a specific orientation of e.g. various reactions is preferred it is necessary to choose the
orientation before setting up the run. Select the plate, in the case shown below, a 384-well
plate has been chosen. Open the drop down menu and choose ‘Horizontal’. The samples on
the plate will now be ordered horizontally.
This is independent of the orientation of the samples, they could be vertically as well as
horizontally orientated and the result on the 384-well target plate is horizontally orientated
reactions.
This is valid for ‘reactions’ (see 6.6.1), so the complete plate is orientated in a chosen
orientation. If ‘fixed reactions’ are used, than the orientation can be chosen plate
independently and several ‘fixed reactions’ could have different orientations on the same
plate (see 6.6.4.1).
PIRO® Manual, Version 1.5.2
51
6.5 Samples
The sample feature tree displays the main heading of the
samples as well as the submenus, which are the sample groups.
The number following the ‘Samples’ indicates how many sample
groups have been set up by the user. Selecting
‘Samples’ highlights all samples set up on the deck.
Selecting a sample groups will highlight the
particular sample group on the deck, as well as
showing all transports assigned to that
particular sample. Right-clicking over ‘samples’ will open the option to add another sample
group. Right-clicking over a sample group will open the option to remove a sample group.
Sample groups can only be deleted, if the sample group has not been assigned to a
transport. If a transport has been assigned, the transport must be removed first by deleting it
in the feature tree. Once all transports have been removed, the sample group can be
deleted.
6.5.1 Setting up Samples
In order to set up samples for a run, select the wells by marking the area where the samples
should be located in (see 5.4.1). Having chosen ‘Samples’, the sample feature table is
showing the samples, inducing the position as described in the columns ‘Well’ and ‘Name’.
In this table, the position of the samples cannot be changed, but by clicking in the cells of
each of the other columns, the names can be altered as well as the ID and the concentration
filled in. If no ID has been assigned, the run will still start, but the software will display an info
sign (see 5.1.2), reminding the user to enter an ID, if required.
A special feature when setting up samples is that no target function has to be assigned if the
samples are to be transferred into a reaction. By dragging samples into an empty well, the
software automatically creates as many fixed reactions as samples have been transferred
(see 5.4.1).
6.5.2 Sample Options in the Deck Window
Right-clicking over
‘Samples’ opens the
following options in the
deck menu (see 4.5),
allowing for the below
actions to be taken:
Add samples to existing sample group: If several sample
groups have already been set up, choose the sample group, the
selected sample should be added to.
Remove samples from existing sample group: Individual
samples can be removed from a sample group.
Create sample group: A new sample group can be created.
PIRO® Manual, Version 1.5.2
52
Remove from sample group: The button below the sample table can be used to remove
samples from a sample group.
6.5.3 Skip Missing Samples
This function allows skipping samples even if the level sensing of the PIRO® does not detect
any liquid in the sample well. The run will continue with the next sample position without
asking for confirmation. If this function is ticked off, the robot will stop at the position where
no liquid has been detected and prompts the user to react, either ‘Ignore’ or ‘Continue’ the
run. This has been introduced as in some cases not all wells contain liquid but the PIRO®
should finish the run without intervention of the user.
Reasons for that could be that the extraction procedure was missing a sample which should
be ignored, or that in purpose a well has been left empty in order to keep the layout of the
samples, or for other user specific reasons. If this option was not turned on, then the PIRO®
would wait for a confirmation to either ignore or continue the run, the user would actually
need to be near the
PIRO® to confirm the
next action taken. With
the option turned on,
the PIRO® will finished
the run without asking
for confirmation
However, in the report
the PIRO® will record
the samples not
pipetted and the user
has a record of the
samples pipetted. This
function only makes
sense to use when level sensing is turned on.
In order to use this function, select ‘Project’ in the feature tree and tick the box ‘Skip missing
samples’. This function is related to the complete project. The PIRO® will now skip those
samples which do not have any liquid.
With this function turned on, the PIRO® only skips empty sample wells but does not skip
empty mix or reagent wells.
6.5.4 Sample Groups
A sample group is a group of samples. Multiple sample groups can be created in one run.
The same samples can be assigned to multiple sample groups, allowing different tasks to be
performed with the same sample. For example, seven samples can be set up; then five of
the seven samples are tested with mix 1 and four of the seven samples tested with mix 2.
PIRO® Manual, Version 1.5.2
53
Assigning samples to sample groups can be achieved in several different ways: mark empty
wells or select samples on the plate, right-click over the selected wells and choose ‘Create
Sample Group’ (see 6.5.2). The project table will then show a new sample group. A second
option is to right-click over ‘samples’ in the project table and choose ‘Add sample group’ (see
6.5). The desired samples then need to be added to the new sample group.
The figure above shows a table including
sample group 1 to sample group 8 and the
ticked boxes indicate the groups each
sample or various samples belongs to.
6.5.5 Sample Group Table
Various options are available in each
sample group table. The name of the group
and the samples can be changed in the
sample group table. The viscosity of the
sample can be addressed and the ID and
concentration can be entered. When a
sample is highlighted, the ‘Remove from
sample group’ button becomes active and a
sample can be removed from the sample
group.
Ticking the box ‘Return remaining quantity
for multi-dispensing’ means that the
remaining volume of the sample in the tip used is returned into the vessel where the sample
has been taken from. Depending on the height where the sample has been dispensed
during the ‘sample-multi-dispensing’ function, minimal remaining liquid on the outside of the
tip could be transferred back into the original sample vessel. The unticked box will discard
the remaining sample volume.
PIRO® Manual, Version 1.5.2
54
6.5.6 Creating Sample Groups from Samples located on different Plates
The PIRO® software is providing the option of creating sample groups from samples located
on different plates. This is of interest, when samples have been stored in different tube
formats or plates as shown below. 0.2 ml, 0.5 ml, 1.5 ml, 2 ml, and 5 ml tubes have been
used to create a sample group and have been transferred to a 96-well plate see below.
A sample group can also be created when the amount of samples is higher than the
available spaces on a given plate. Shown below are two 33-well plates. All 66 samples have
been allocated to one sample group and transferred into a 96-well plate. This also gives the
option of a sample group consisting of 4 x 96 samples to be transferred in one 384-well
plate.
PIRO® Manual, Version 1.5.2
55
6.5.7 Import
6.5.7.1 Import sample data
Existing sample data can easily be imported via the ‘Import’ in the file menu. Besides sample
names, ID, concentrations also indicators to determine further processing of the imported
samples can be imported (see 6.5.8). Select ‘Import sample data’/ ‘CSV Import’ and the
sample import window will open.
Prior to the sample import, a CSV file needs to be created and could contain the following
information:
Well position, name, ID, and concentration of
the sample. If the file is open in another
application during the import process, an error
message will appear saying ‘The entered file
cannot be opened’. Close the CSV file and
proceed.
Features of the file to be imported:
• The minimal file content consists of one column, e.g. the sample names or IDs. If no
column number for the ID is given, no import can take place but choosing the same
column number allows the import of the CSV file.
• Colum headers are not a necessity but allows for easier identification. There is no given
format for the headers of the columns to be used.
• Plate positions in the target plate are not mandatory, but allow the user to direct the
samples in certain positions. This could also be done by specifying the starting well
under ‘Start at position’ see below.
• In the CSV file there is no definite order of columns required this can be specified in the
boxes behind the sample import details.
• The CSV file can also contain columns not to be used for the import and would simply
not be specified in the boxes behind the sample import details.
• TXT files could also be imported but the file ending needs to be changed to .CSV
Prior to the sample import please make sure of the following points:
PIRO® Manual, Version 1.5.2
56
• The target plate provides enough space for the amount of samples to be imported
• The target positions are not in use by other samples or sample groups
• The names differ to the samples already present in the template
Open the file menu, select ‘Import’ / ‘Import sample data’ and click ‘CSV Import’. The sample
import filter window will open.
Select the browse button and open the folder containing the CSV file. Choose a
file by double clicking or select ‘Open’ to import the file.
Use the following encoding: Choose the geographic location.
Separator: Choose ‘Tabulator’, ‘Semicolon’, ‘Comma’ or ‘Other’. If for ‘Other’ is selected,
enter the separator used in the file to be imported. Tick the box ‘File has column headers’ if
the file has column headers.
The file contains column headers: This box should be selected if the columns have
column headers. If this box is ticked the first row of the CSV file will be ignored.
PIRO® Manual, Version 1.5.2
57
Import Range: Define the import range and type in the starting row. If a limit is to be set, tick
the ‘Limit to’ box and fill in the last row. Not ticking the box imports all rows defined in the
CSV file. Using the function ‘Rows’ allows defining specific rows to be imported, comma
separate the rows (e.g. 2, 4, 6) and press ‘Import’.
Sample Positions: Select how the
imported samples are to be
positioned in a defined plate setup.
Choose between:
‘Auto arrange within sample
function’, arranges all samples in one or more defined samples positions.
‘Read positions from file [column]’ imports all samples into predefined positions.
Choose between ‘Plate from Import File’ and other predefined plates.
‘Start at position [e.g: 5.a1]’ automatically arranges all samples from a defined
starting position.
Import Sample Name, ID, and Concentration need to be ticked and the column number of
the CSV file specified if any of those values are to be imported. If only one column should be
imported, then choose the same column number twice.
Sample Groups: Select ‘Sample Groups’ if imported samples are defining a new sample
group or if integration into existing sample groups should be performed:
‘Create new group’: Will add a new sample group to the setup
‘Add samples to
existing sample
group’: Select from
preexisting sample
groups
‘Read sample
groups from file’:
see 6.5.8.1
‘Insert samples
groups from file’:
see 6.5.8.1
If all settings are entered
correctly and no red cell or
text indicates an error, the
‘Import’ button becomes
active and sample data can
be imported. ‘Close’,
cancels the sample import
function.
PIRO® Manual, Version 1.5.2
58
6.5.7.2 Import volume data
This feature will allow in the future importing volume data and is currently under
development.
PIRO® Manual, Version 1.5.2
59
6.5.8 Intelligent Pipetting
The intelligent pipetting option allows a run to be set up directly from a database and once
set up, can be used repetitively for further runs simply by changing the import file. The three
steps required are explained below:
Set up a PIRO® template (see 6.5.8.1)
Setting up the CSV file for Import (see 6.5.8.2)
Importing the CSV file (see 6.5.8.3)
In the example below three parameters have been chosen to be tested (Virus-1, Virus-2, and
Virus-3) using three different mixes and sample groups.
6.5.8.1 Setting up a PIRO® Template for Intelligent Pipetting
a) Choose and select the plate layout to be used for the run. Mark an area and select
samples from the function menu (see 6.5.1). In this example the complete plate Nr. 9
has been selected. Unmark the samples. Right mouse click over samples in the
feature tree and select three different sample groups (see 6.5), which do not contain
any samples. Rename the sample groups accordingly; in this example the names
used are Virus-S1, Virus-S2, and Virus-S3. The sample groups need to have the
same name as the names in the database, see below.
b) Make sure that samples, mixes, and reactions do not have identical names, in the
example shown the samples are called Virus-S1, Virus-S2, and Virus-S3, the
reactions are called Virus-R1, Virus-R2, and Virus-R3 and the mixes are called Mix
Virus-1, Mix Virus-2, and Mix Virus-1.
c) Set up three fixed reactions areas (see 6.6.3), in the example shown below, plate 10
has been used. Name the fixed reactions accordingly; in this example the names
used are Virus-R1, Virus-R2, and Virus-R3.
PIRO® Manual, Version 1.5.2
60
d) Set up three mixes (see 6.3.1) and name them accordingly, in our example the mixes
have been named Mix Virus-1, Mix Virus-2, and Mix Virus-1. In the reaction feature
table field called ‘Volume added to reaction’, add the volume to be used for each mix.
Continue until no error message is displayed in the mix feature tree anymore and
drag the mixes in the appropriate fixed reactions.
e) Select a fixed reaction. Press ‘Add’ to add a sample group to be used with this fixed
reaction. Select from the drop down menu and choose the appropriate sample group.
The file should look like as shown below:
PIRO® Manual, Version 1.5.2
61
f) Save this file as a template.
6.5.8.2 Setting up the CSV-File for Import
To import the data create a file as shown in the example below:
The first column indicates the position where the sample should be imported to, the second
column indicates the names, and the third column indicates the ID of the sample. Columns
4, 5, and 6 show the tests which should be done for each of the samples. For sample ‘a’ all
three test should be done, for sample ‘b’, only Virus 2 should be tested and so on. As
mentioned above, it is important that the headings of columns 4, 5, and 6 are identical to the
sample group names in the PIRO® template. Once this Excel table is set up it is necessary
to save the file as a CSV file and it is not opened in any other program while importing (see
6.5.7).
On the bottom of the file above are also shown the total reactions of each test. This could be
helpful when setting up the amount of fixed reactions in the template.
Position Name ID Virus-S1 Virus-S2 Virus-S3
5.1 a 111 1 1 1
5.2 b 222 1
5.3 c 333 1 1
5.4 d 444 1 1
5.5 e 555 1
SUM 2 4 3
PIRO® Manual, Version 1.5.2
62
6.5.8.3 Importing a CSV-File
To import the CSV file open the above created template, select ‘Sample import’, ‘CSV
import’ and the sample import window will open (see 6.5.7). Select the file to be imported,
and select ‘Read sample groups from file’. Fill in the columns from the CSV file where the
description is given what test should be done with what sample. Press import and the PIRO®
run file is setup. Future run files can simply be set up by opening a template, importing the
samples from the CSV file and press start run.
PIRO® Manual, Version 1.5.2
63
6.5.8.4 Import Samples in Consecutive Order
In the above example, three different fixed reaction areas have been created in order to
separate the three different tests from each other (see 6.5.8.1c). Another option is to have
the reactions aligned in a consecutive order. For that to be achieved, proceed as outlined
above (see 6.5.8.1) but instead of setting up fixed reactions, simply set up one reaction area
(see 6.6.1) enough for all reactions to be done. In the
feature tree, right mouse click over ‘Reaction’ and
choose ‘Add reaction’. Importing the samples (see
6.5.8.3) will now order all the reactions in a consecutive order.
PIRO® Manual, Version 1.5.2
64
6.5.9 Barcode Reading
The PIRO® offers the option of
barcode reading the sample IDs. This
firstly facilitates the user with naming
the samples rather than typing in
every sample by hand. Secondly,
sample IDs are often needed for LIMS
integration. In order to do that, use a
manual hand scanner which can be
supplied by DORNIER-LTF and
connect it to the free USB port. Open the sample list of a project, click into the table of
sample name or sample ID and scan the barcode. No further software is required.
If the barcode reader has been supplied by DORNIER-LTF make sure the correct country
code setting are used. For more information the manual of the barcode reader has been
uploaded in the customer area of DORNIER-LTF.
The PIRO® also offers the option of barcode reading the Plate IDs, as shown in the figure
above. Select the orange field above and use the barcode reader to read the Plate ID.
In addition, ID for standards could also be given, for more details see 6.7.3.
PIRO® Manual, Version 1.5.2
65
Position ID conc
5.a1 a 1
5.a2 b 2
5.a3 c 3
5.a4 d 4
5.a5 e 5
5.a6 f 6
5.a7 g 7
5.a8 h 8
6.5.10 Normalisation of Sample Concentrations
The PIRO® software allows samples to be normalised. Samples with different concentrations
are diluted accordingly so all the samples will result with the same concentrations. The
samples are normalised within intermediate wells and can then be used for setting up
reactions or other procedures required in the same run. Normalized samples are shown with
a black line through a circle (see 5.4.4).
6.5.10.1 Import Sample Data
Sample concentrations can either be imported or directly typed into the software. Shown
below is an example of a CSV-file, where the position, ID, and concentration have been
given. By importing the CSV-file (see 6.5.7), the concentrations are imported and a new
sample group can be created.
6.5.10.2 Setting up a Normalisation
The next step is to create the normalisation procedure. Firstly, assign a diluent anywhere on
the deck, as without a diluent, no normalisation will take place. Next, select the samples to
be normalised, drag them over an empty well, right mouse click, and choose ‘Create
normalisation’ from the pop-up menu.
A new window appears showing the ‘Sample group 1 – Normalisation 1’ in the feature tree.
Fill in the requested information in the feature table, including a diluent as well as ‘Minimum
sample volume’ and ‘Minimum usable volume’. The ‘Minimum sample volume’ is the volume
of sample to be used from Sample Group 1 (undiluted samples) to do the normalisation with
and ‘Minimum usable volume’ is the volume required from the normalised samples for further
reactions.
PIRO® Manual, Version 1.5.2
66
Once the above information has been filled in, a reaction and mix can be created and the
normalised samples can be used to set up a run.
In the example above, the ‘Normalise to lowest concentration’ option has been used, which
means that the normalisation has been performed to the lowest sample concentration. The
other option is to ‘Normalise to target concentration’, where a different concentration can be
used to normalise the samples to. Samples with a lower concentration will then be ignored
but a warning message appears in the feature tree (see below on how to set that up).
The ‘Positions’ tab in the feature
table gives an overview of the
details on the dilutions to be
done. ‘Source conc.’,
‘Normalised conc.’, ‘Sample
volume’, ‘Diluent volume’, ‘Total
volume’ and ‘Diluent ratio’ are
indicated.
The ‘Options’ tab in the feature table allows for samples not to be aliquoted, as they have the
final concentration already. To allow for that, tick the box ‘No aliquot for undiluted samples’.
If the complete sample group will be used for further reactions, the undiluted sample will then
be used from the original sample well.
PIRO® Manual, Version 1.5.2
67
‘Mixing at the end’ allows
introducing a mixing step after
the normalisation by simply
indicating a number of times to
mix.
Samples could also be filtered
by a given concentration. To do
so, tick the box following ‘Filter
samples’ and set a range of
concentrations to be used. In the
example below, all samples with
concentrations above 7.00 and
below 2.00 have been excluded
from the normalisation step.
The orientation of the normalised samples can either be chosen automatically or set
manually.
Using the deck menu, two further options are available:
‘Add positions to
normalisation’ allows the
user to add more position
than currently available and
‘Remove position from normalisation’ allows the user to remove positions.
6.5.11 Dilution Series of Samples
With samples dilution series can be performed.
Before starting set up a diluent, as otherwise an
error message will appear. A dilution series cannot
be done without a diluent. Select the samples to
be diluted and drag them into an empty well. Right
mouse click over the new well and select ‘Create
dilution’.
PIRO® Manual, Version 1.5.2
68
If further dilutions should be created, repeat the above procedure, ending up in various
dilutions of the original sample, as shown above. With each dilution step an additional line is
added to the circle of the sample on the screen indicating the number of dilution steps done.
In the above example, five dilutions have been created. Names are also created
automatically as shown below, but can be changed if necessary.
When the dilution has been set up a window appears. Click on the ‘Dilution’ window to add
the information requested. The name can be changed, a concentration unit given and a
diluent chosen if several diluents have been created when setting up the run. In the lower
part of the window, the dilution ratio can be given as well as the final volume to be created.
The ‘Positions’ window shows the details of the dilutions to be done.
The ‘Options’ window
allows for specific
details to be changed
including filtering of
samples as well as
choosing the
orientation of the
samples. This
window is similar to
the window described
in normalisation (see
6.5.10.2).
PIRO® Manual, Version 1.5.2
69
6.6 Reactions and Fixed Reactions
Reaction wells are defined as carrying the final result of the pipetting steps in a certain
project. Reactions are the end of a pipetting project and are normally taken to be analyzed in
a further instrument.
Fixed Reactions: The PIRO® has an additional definition for reactions, called fixed
reactions. Fixed reactions help the user the setup of reactions,
and have their own special rules in a reaction area. If a
reaction has been set up as a reaction or fixed reaction can be
seen in the feature tree following the name of a particular
reaction; in the example shown above reaction 3 has been set up as a fixed reaction. There
are two numbers following a fixed reaction, the first number indicates how many fixed
reactions are in use and the second number indicates how many fixed reactions have been
set up. The orientation of fixed reactions on the plate can be changed, replicates can be
treated differently, and options to assign various mixes are realized. For more information on
fixed reactions, please see 6.6.2.
The reaction feature tree displays the main heading of the reactions and fixed reactions as
well as the submenus, which are the various reactions used in the project. The number
behind ‘Reactions’ indicates how many reactions have been set up. Clicking ‘Reactions’
highlights all reactions set up on the deck. Clicking on one of the reaction tabs in the
submenu only highlights those particular reactions on the deck, as well as all transports set
up for those reactions, if the transport button in the transport menu has been activated. The
number behind each reaction in the submenu indicates how many reactions have been used
in this run file.
6.6.1 Setting up Reactions
To set up a reaction area, mark the wells to be used
and either use the function button in the tools menu
or right-click over the chosen area and select ‘Add
Function’ then ‘Reactions’. A reaction area has now
been created.
In order to add samples to the reaction area, select
empty wells or define samples and drag them into the defined reaction area as shown
above. When using this function, be aware that the number of samples cannot exceed the
number of reactions. If ignored, samples will be missing in the reaction. The same procedure
can also be applied when cherry picking samples, meaning samples from various different
locations are chosen. Simply use the ‘Ctrl’ button on the keyboard, select the samples of
interest and drag them into the reaction area.
PIRO® Manual, Version 1.5.2
70
A sample volume now has to be entered into the feature table, as indicated by the red cell as
well as by the error message appearing in the feature tree. Once a sample volume has been
entered, the error message will disappear and in the case of the example shown below, a
warning, (see 5.1.3) message will be displayed, as no mix has been added to the reaction as
yet.
6.6.2 Reaction Feature Table
The reaction feature table has
three different submenus, reaction,
positions, and options.
Reaction: This window specifies
the ‘Name’ of the reaction, the
‘Reaction volume’, the ‘Content of
the reaction’, including the mixing
options available. All parameters
could be changed in this window,
except the ‘Reaction volume’,
which is the sum of the mix and the
sample per reaction.
When changing the volume of a mix, it also has to
be changed in the mix window. A warning sign will
appear if the two values do not match.
Position: This window gives details of the well
position, the ID, and the concentration of the
reaction, if required.
PIRO® Manual, Version 1.5.2
71
Options: This window gives several
options to the user. Ticking ‘User defined
orientation’ allows the user to automatically
let the software choose the default
orientation of the reactions. The orientation
options (see 6.6.4.1), are only available for
the ‘fixed reactions’. ‘Replicates’ can
simply be added by changing the number
in the replicate box.
‘Allow mix without samples’ is an option to
be used when no sample will be added to
the final reaction. As the software is mainly
set up for PCR users, a mix without a
sample is defined as an error, but this
option can be ticked to allow the reaction to be set up without the addition of a sample.
6.6.3 Setting up Fixed Reactions
The first option to set up fixed
reactions is by dragging samples or
selected empty wells into the desired
fixed reaction area (see 5.4.1). This
creates as many fixed reactions as samples have been dragged into the fixed reaction area.
The second option is to highlight an area where the fixed reaction is supposed to be then
right-click over the well and select ‘Create fixed reaction’. The same menu also allows
‘adding positions to an existing fixed reaction’. The third option to set up fixed reactions is to
choose ‘Options’ in the reaction table and tick the box called ‘Is fixed reaction’. This
transforms a reaction into a fixed reaction.
PIRO® Manual, Version 1.5.2
72
6.6.4 Special Features of Fixed Reactions
6.6.4.1 Orientation of Fixed Reaction
When using fixed reactions, the orientation of the samples in the reactions can be changed,
so for example horizontally orientated samples can be changed to vertically orientated
samples.
6.6.4.2 Replicates when using Fixed Reactions
If ‘fixed reactions’ are chosen the number of replicates to each ‘fixed reaction’ can differ. If
only ‘reactions’ have been used, all ‘reactions’ have to have the same number of replicates.
‘Fixed reactions’ therefore allow for more flexibility when setting up reactions in general.
6.6.4.3 Fixed Reactions and Intelligent Pipetting
Fixed reactions should also be used when a sample import or Intelligent Pipetting step is
performed. Due to the separation in fixed reactions each fixed reaction can have its own mix
or sample group assigned. If however, the reactions should be in a consecutive order, then
‘reactions’ instead of ‘fixed reactions’ should be used. For further details in regards to the
intelligent pipetting please see 6.5.8.4.
PIRO® Manual, Version 1.5.2
73
6.6.5 Reaction Export
Sample data and reaction results can be
exported as a CSV file. To export a
reaction, open the file menu, select
‘Reaction export’ and click on ‘CSV
Export’.
Two further options are available, which
is the direct export to samples files used
in the LightCycler and the Rotor-Gene.
In case multiple reaction plates have
been prepared by the PIRO®, different
export files for individual plates and
formats can also be selected.
This opens the ‘Reaction Export Filter’
Click the tab with the dots to
open the hard drive directory, name the
document, select where to save the file,
and select ‘Export’.
Use the following encoding: Choose
the preferred option:
Separator: Choose between ‘Tabulator’,
‘Semicolon’, ‘Comma’, or ‘Other’. For
‘Other’ type the separator to be used.
Tick the boxes if column headers and
plate info have to be exported as well.
Click on ‘Export’ to finish or ‘Close’ to
cancel.
An example of a reaction export is
shown below:
PIRO® Manual, Version 1.5.2
74
6.6.6 Sample-Multi-Dispensing
Warning: The ‘Sample-Multi-Dispensing’ feature could lead to cross-contamination of a
reaction. It is therefore important to verify prior to any run using the ‘Sample-Multi-
Dispensing’ feature if the outcome of the use of this feature is according to the aim of the
user.
The ‘Sample-Multi-Dispensing’ feature has been developed to multi-dispense samples into
reactions. According to the PIRO® rules, a reaction including various reagents is prepared
and the PIRO® will then use a new tip for every sample. This is the most secure way not to
allow for potential cross-contamination as the PIRO® will normally not multi-dispense
samples.
The advantage of using ‘Sample-Multi-Dispensing’ feature is that a large amount of tips
could be saved, so more cost effective runs could be done. Two options are available:
‘Enable Sample-Multi-Dispensing for replicates’: Using this option will multi-dispense
samples into reactions, where the reactions have the same mix, meaning that replicates are
prepared. If the tip containing the sample will touch the reaction, only a minimal amount of
mix on the outside of the tip could potentially be transferred into the next reaction. If this is
considered to be a contamination needs to be decided by the user.
‘Enable Sample-Multi-Dispensing for all reactions’: Using this option will multi-dispense
samples independent of the reaction content. This means that a sample will also be multi-
dispensed into reactions with different reaction content, therefore potentially contaminating
from the outside of the tip the next reaction with mix from the previous reaction. It therefore
needs to be taken great care if such a setup makes sense to apply for the user.
PIRO® Manual, Version 1.5.2
75
The advantage is that if the same sample should go into for example eight different reactions
that seven times less tips are used. This is a considerable amount of tips to be saved.
This function has been implemented as a PIRO® user has extensively tested this setup for
cross-contamination and no cross-contamination has been observed in the users’ setup.
DORNIER-LTF therefore decided to implement this option.
In addition and to allow for further flexibility, the PIRO® software has also the option to set
different x, y, and z-offsets when using the ‘Sample-Multi-Dispensing’ function. The general
offset values are normally pre-calibrated when the consumables have been set in the
software or if new consumables different to the software are used, the user has set such
values.
Using the ‘Sample-Multi-Dispensing’ function now also allows changing the x, y, and z-
offsets only for the ‘Sample-Multi-Dispensing’. This means that the mix will be pipetted
according to the initial settings, but the samples will be pipetted according to the new x, y,
and z-offsets chosen. As an example, the sample could therefore be dispensed above the
reaction or on the side of the wall of the reaction vessel. This will allow for contamination free
pipetting when using the ‘Sample-Multi-Dispensing’ function. However, the user has to take
great care that such settings are according to the expected results.
PIRO® Manual, Version 1.5.2
76
6.7 Standards
The standard feature tree displays the main heading of the standards
as well as the submenus, which are the various standards used in this
project. The number next to ‘Standards’ indicates how many
standards have been set up. Clicking ‘Standards’ highlights all the standards set up on the
deck. Using one of the standard tabs in the submenu highlights the standards in the
submenu on the deck, as well as all transports set up. The number next to each standard in
the submenu indicates how many standards have been used in the run file.
6.7.1 Setting up Standards
A diluent should be set up before setting
up any standards (see 6.1.1), as standards
cannot be set up without a diluent even in
virtual mode. If no diluent is set up, the
software only allows a ‘Ready to use
standard’ to be used.
To set up standards mark the wells to be
used, right click over the marked well and
choose ‘Create standard’ see below.
6.7.2 Ready to use Standard
A ‘Ready to use standard’ is one which
has been created prior to the run. The
‘Ready to use standard’ only allows the
concentration to be changed.
6.7.3 Creating a Standard
The PIRO® robot can be used to create
standards. Un-tick the box ‘Ready to use
standard’ and enter the required
information.
A master standard can either be placed on the same plate or rack, but can also be provided
on a different plate or a different sized tube. This means that the master sample does not
have to be aliquoted.
Minimum Volume to create [µl]: A volume of 50 µl is the default setting in the PIRO®
software, but the volume to be created can be adjusted. If the value is changed, the software
will automatically recalculate the volumes needed.
Diluent: If several diluents have been set up on the deck, the software allows the
appropriate diluent to be selected when setting up standards. This enables different diluents
to be evaluated in the same run.
Mixing: The ‘mixing’ option allows the number of times the mix should be mixed to be
defined by the user. This is particularly important for standards as the following standard
dilution has to be provided with a homogenous standard.
PIRO® Manual, Version 1.5.2
77
Change Tips: This option allows for tips to be changed when creating the standards.
Standard IDs: This field allows for IDs to be given, manually or with a barcode reader.
Standard Table: The standard table shows the details of the standards to be created.
Shown in the first column is the well position, the second column shows the dilution ratio to
be used. The original standard is called the ‘Master’ standard.
Concentration units: A concentration can be entered in the concentration column if
desired.
Diluent Added [µl], Sample Added [µl], Sample Taken [µl], and Created Vol [µl] indicate
the volumes the PIRO® will transfer during the run. The volumes can be altered by the user if
required.
6.7.4 Standard Options in the Deck Window
‘Create standard’ creates a
standard.
‘Add Positions to Existing
Standard’ adds a position to
an existing standard.
Create Standard Group: see 6.7.5
PIRO® Manual, Version 1.5.2
78
‘Add Standard Positions to Existing Standard Group’ adds standard positions to an
existing standard group.
‘Remove Standard Positions from Existing Standard Group’ removes standards from an
existing standard group.
6.7.5 Standard Group
The use of the standard group function has been developed to
provide an option to select only standards of interest as not all
standards are always used in a particular run, even though they have been created by the
PIRO®. The software therefore allows creating a group of standards derived from the
standards created originally. In many cases the user has a high concentrated master
standard but only wants to use the lower diluted standards. In other cases, the user would
like to use every second standard dilution, as the wide concentration range should be
covered. Performing a one-step 1:100 dilution is less accurate than doing a two-step 1:100
standard dilution.
To set up a standard group, create a standard as shown below. In our example, 12
standards have been created.
PIRO® Manual, Version 1.5.2
79
Now only choose the standards desired and drag them into a fixed reaction, in our example,
only every second standard should be used. A new standard group will automatically be
created. Any combination of standards can be chosen to create a new standard group.
6.8 Reserved
This option allows the user to reserve certain wells on the deck to prevent them from being
used. This function may be used if parts of a plate have been used during a previous run. In
some instruments, certain areas might not work
identically compared to others, which could be
excluded from usage as well. Certain output files
should look identically, so reserved areas could be
used without given up the option of an automatic
orientation provided by the software. In the
example shown on the right, positions C1, D1, and
E1 have been marked as reserved. Five standards
from plate 9 have been transferred to plate 10, starting at position A1. The
standards have automatically been added to positions A1, B1, F1, G1,
and H1. Positions C1, D1, and E1 are not used as they have been
marked as reserved wells.
6.8.1 Setting up Reserved Area
To set up a reserved area, select the wells to be used and use ‘Functions’
to create a reserved area. The reserved feature tree will then display the
number of reserved wells in the project as well as the position and names
of the reserved wells.
PIRO® Manual, Version 1.5.2
80
6.9 Special Functions of the PIRO® Software
6.9.1 Automatic Tube Switching (ATS)
The PIRO® has the option to automatically switch the tube when empty and use a second
alternative tube with the same content. This is possible for reagents, diluents, and mixes.
6.9.1.1 Automatic Tube Switching for Reagents and Diluents
Assign a position with the function ‘Reagent as described (see 6.2.1) and the feature window
opens. Two windows are shown titled ‘Assigned wells’ and ‘Unassigned wells’. Choose the
location where the alternative tube should be located by pressing the ‘+’ sign and
the available well positions will appear. In this case, position 13.2 has been
chosen. Press the arrows to transfer the well 13.2 to the window called ‘Assigned
wells’ and the position is defined. Alternative wells could be assigned as many as positions
are available. The alternative tube is displayed with a grid.
6.9.1.2 Automatic Tube Switching for Mixes
The same principal can be applied to mixes, however, before doing that choose ‘Volume is
predefined / premixed’. Open the ‘Options’ window and move the well to ‘Assigned wells’ as
described above and an alternative well containing the same mix is defined.
Switching level: The volume level when switching a tube can be defined by the user and is
measured in mm. Be aware that the level must be within the definitions of the tube size
which the software will recognize. If 1 mm is chosen, then the user needs to make sure the
calibration of the tubes is done correctly.
PIRO® Manual, Version 1.5.2
81
6.9.2 Automatic Fluid Transfer (AFT)
The “Automatic Fluid Transfer” option allows the user to carry over remaining fluid from a
reagent or mix into an alternative well with the same content. An ‘alternative well’ is
described above (see 6.9.1), meaning the same reagent or mix in a second or third tube. If
the ‘automatic tube switching’ is turned on, the remaining fluid from the first tube will be
aspirated and transferred into the second tube with the same content and so on, therefore
not wasting any fluid.
As with the ‘Automatic Tube Switching’ option, for mixes the option ‘Volume is predefined /
premixed needs to be chosen.
PIRO® Manual, Version 1.5.2
82
7 Quick starting the PIRO®
Start the PIRO® software.
Choose the desired plates by using ‘Plates’ in the Layout Menu.
Highlight the reagents and use the ‘Function’ button in the Layout Menu to assign the
function. Drag the reagents into a well where the mix is supposed to be.
Assign a diluent function to a well and drag the diluent into the mix.
In the feature table, add the reagent volumes to the mix and specify the ‘Volume added to
reaction’.
Mark the wells containing the samples; assign a sample function and drag the samples into
the position where the reactions are supposed to be. The software will automatically assign
a reaction area. Specify the sample volume in the feature table.
Select the mix and drag into the reaction area.
PIRO® Manual, Version 1.5.2
83
8 Appendices
8.1 PIRO® Alarm List
Alarms Description Consequence
There are no tips with the needed volume left. User needs to
replace the empty tip rack.
Not enough tips for the project Change tips and push "resume"
There is not enough fluid left in the vessel to finish the
transport. Please make sure that there is enough fluid in each
vessel. You can continue or stop the run.
Not enough fluid for the project User decision, push "resume" or "stop"
The tip is lost during a job execution. Please check the project
for contamination. Robot run is stopped.
Tip lost You must check for contamination and
start the project from the beginning
The current job cannot be completed successfully because
there is no tip attached to the robot head. Maybe it has been
lost. Robot run is being stopped
Tip lost You must check for contamination and
start the project from the beginning
An unexpected error happened during job execution. Unable
to continue the robot run. Please check the project for
contamination
unexpected error
You must check for contamination and
start the project from the beginning
Disconnected while executing command Communication is interrupted Check USB cable. In addition you must
check for contamination and start the
project from the beginning
Communication lost Communication is interrupted. Check USB cable. In addition you must
check for contamination and start the
project from the beginning
Unable to reach x axis position Axis are blocked, light barriers are not
working or motor does not work
Call distributor
Unable to reach y axis position Axis are blocked, light barriers are not
working or motor does not work
Call distributor
Unable to reach z axis position Axis are blocked, light barriers are not
working or motor does not work
Call distributor
Tip switch is always active PIRO can not drop the tip or switch is
not working
Try to drop the tip manually and start
the project again. Change pipetting head
Pump limit switch is always active Limit switch is not working Change pipetting head
Pump switch not found Pump switch is not working Change pipetting head
Pump Capacity is too low Pipetting head has not enough volume
for the job
You must check for contamination and
start the project from the beginning. In
rare case change the pipetting head
The robot pump reports an invalid volume. Please check if the
pump is connected properly. If the pumpe is connected
properly, please contact the distributor. Pump serial number
________
Pipetting head is not working or lost its
memory
Change pipetting head
PIRO® Manual, Version 1.5.2
84
INDEX
A
Add ....................................................................... 36
Auto Assignment of Transports in Empty Wells ... 43
Automatic Tube Switching .................................... 80
C
Calibrations .......................................................... 12
Changing Names ................................................. 37
Changing the Order of Transports ........................ 39
Checklist .............................................................. 32
Cherry Picking ...................................................... 69
Cleaning Acrylic Plexiglas .................................... 12
Clipboard Menu .................................................... 27
Colour Coding ...................................................... 44
Computer ............................................................. 14
Copy ..................................................................... 27
Create Sample Group .......................................... 51
Cut ....................................................................... 27
D
Deck Layout ......................................................... 18
Deck Window ....................................................... 34
Default Project...................................................... 26
Delete ................................................................... 27
Diluent .................................................................. 45
Dilution Series ...................................................... 67
Document Info ...................................................... 33
Document Restriction ........................................... 33
Document Security ............................................... 32
Downloads ........................................................... 17
E
Enable sorting the transport list manually ............ 39
Error ..................................................................... 36
Exchanging Existing Plates .................................. 25
Exit ....................................................................... 27
F
Feature Table ....................................................... 36
Feature Tree ........................................................ 35
File Menu ............................................................. 26
Fixed Reactions ................................................... 69
Force Tip Change ................................................ 30
Functionality of the Lid ......................................... 12
Functions ............................................................. 28
H
HEPA-Filter .......................................................... 11
I
Import ................................................................... 55
Info ....................................................................... 35
L
Language ............................................................. 29
Layout Menu ......................................................... 27
Level Sensing ....................................................... 30
Level Sensing each (times) .................................. 30
Lid......................................................................... 12
M
Manual Orientation ............................................... 71
Mix Volume Calculation Submenu ........................ 47
Mixes .................................................................... 47
Mixing Options ...................................................... 37
Move ..................................................................... 27
Move to home position ......................................... 31
Multi-Dispensing ................................................... 30
N
New ...................................................................... 26
O
Open Project ........................................................ 26
Options Menu ....................................................... 29
Output Range ....................................................... 33
Overview of Racks and Adapters ......................... 21
P
Paste .................................................................... 27
Pause ................................................................... 31
Pipetting Heads .................................................... 11
PIRO® Manual ...................................................... 17
PIRO® Workspace ................................................ 18
Plate Functions ..................................................... 25
Plate Submenu ..................................................... 38
Plates ............................................................. 28, 38
Pointer .................................................................. 27
Position Offsets .................................................... 24
Positioning the Robot ........................................... 11
Post Run Report ................................................... 32
PIRO® Manual, Version 1.5.2
85
Project .................................................................. 26
Q
Quick starting the PIRO ....................................... 82
R
Racks ................................................................... 21
Racks and Adaptors ....................................... 18, 21
Reaction Export.................................................... 73
Reactions ............................................................. 69
Reagents .............................................................. 45
Recent Files ......................................................... 26
Remove ................................................................ 36
Remove Samples ................................................. 51
Reserved Area ..................................................... 79
Robot Menu ......................................................... 31
S
Safety ................................................................... 12
Sample Groups .................................................... 52
Sample-Multi-Dispensing ..................................... 74
Samples ............................................................... 51
Save As ................................................................ 26
Save Project ......................................................... 26
Separator ............................................................. 56
Setting up Transports ........................................... 42
Show Transports .................................................. 28
Software Installation ............................................. 14
Software Updates ................................................. 16
Special Racks ...................................................... 22
Standard Group.................................................... 78
Standards ............................................................. 76
Start ..................................................................... 32
Start Robot ........................................................... 32
Starting the Software ............................................ 17
Stop ...................................................................... 31
T
Template .............................................................. 26
The Waste Hole .................................................... 20
The Waste Rack ................................................... 20
Tip ........................................................................ 38
Tip Disposal .......................................................... 19
Tip Racks ............................................................. 19
Tips....................................................................... 19
Tools Menu ........................................................... 27
Transports ............................................................ 38
Triangles ............................................................... 35
Tubes ................................................................... 23
U
Undo, Redo .......................................................... 28
UV Program Duration ........................................... 31
UV Radiation Procedure ....................................... 11
V
Virtual Mode ......................................................... 17
Viscosity ............................................................... 37
W
Warning ................................................................ 36
Waste ................................................................... 38
Waste Console with Drawer ................................. 20
Well Position Description ...................................... 36
Z
Zoom .................................................................... 28