The Right GC Column - Selection Essentials

The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Embed Size (px)

Citation preview

Page 1: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

The Right GC Column- Selection Essentials

Page 2: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Column

FlowController

Regulators

Air

Hydro

gen

Carr

ier

Gas

Mol-SieveTraps

Fixed

InjectionPort Detecto

r

Electrometer

Recorder/Integrator

Restrictors

Cylinders or Generators

Typical Gas Chromatographic System

Picking the appropriate stationary phase and optimum dimensions for the column will give the greatest resolution in the shortest analysis time.

Page 3: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Four Primary Selection Areas

Stationary Phase Type

Column Internal Diameter

Stationary Phase Film Thickness

Column Length

Page 4: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Resolution

• N = (gas, L, rc)

• k = (T, df, rc)

= (T, phase)

RNk

ks = 4 11

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 5: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Resolution

RNk

ks = 4 11

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

N = (gas, L, rc)

k = (T, df, rc)

= (T, phase)

Page 6: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Stationary Phase - Common Types

•Siloxane polymers

•Poly(ethylene) glycols

•Porous polymers

Page 7: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Liquid Phase

Carrier Gas

Porous Layer Open Tube (PLOT)

Wall Coated Open Tube (WCOT)

Solid Particles

Carrier Gas

Capillary Column Types

Page 8: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

HH

HO- - C-C-O -- H

H H n

Polyethylene glycol backbone

Stationary Phase Polymers

Page 9: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

k2 = partition ratio of 2nd peakk1 = partition ratio of 1st peak

Why Is Stationary Phase Type Important?

=k2

k1

Influence of

Page 10: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Selectivity

Relative spacing of the chromatographic peaks

The result of all non-polar, polarizable and polar interactions that cause a stationary phase to be more or less retentive to one analyte than another

Page 11: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Optimizing Selectivity

•Match analyte polarity to stationary phase polarity

• Like dissolves like(oil and water don’t mix)

•Take advantage of unique interactions between analyte and stationary phase functional groups

Page 12: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Compounds - Properties

Compounds Polar Aromatic HydrogenBonding Dipole

Toluene no yes no induced

Hexanol yes no yes yes

Phenol yes yes yes yes

Decane no no no no

Naphthalene no yes no induced

Dodecane no no no no

Page 13: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

100% Methyl Polysiloxane (boiling point column?)

Strong DispersionNo DipoleNo H Bonding

0 2 4 6 8 10 12 14 16

12

3

4

5

6

1. Toluene 110o 2. Hexanol 156o

3. Phenol 182o

4. Decane (C10) 174o

5. Naphthalene 218o

6. Dodecane (C12) 216o

Page 14: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

5% Phenyl

Strong DispersionNo DipoleWeak H Bonding

0 2 4 6 8 10 12 14 16

0 2 4 6 8 10 12 14 16

100% Methyl

12

34

56

5% Phenyl

12

34

5,6

Strong DispersionNo DipoleNo H Bonding

1. Toluene2. Hexanol3. Phenol4. Decane (C10)5. Naphthalene6. Dodecane (C12)

Compounds Polar Aromatic HydrogenBonding Dipole

Toluene no yes no inducedHexanol yes no yes yesPhenol yes yes yes yesDecane no no no noNaphthalene no yes no inducedDodecane no no no no

?

Page 15: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

50% Phenyl

Strong DispersionNo DipoleWeak H Bonding

50% Phenyl

0 2 4 6 8 10 12 14 16

100% Methyl

12

34

5

6

0 2 4 6 8 10 12 14 16

1 2 4 3 6 5

Strong DispersionNo DipoleNo H Bonding

1. Toluene 110o

2. Hexanol 156o

3. Phenol 182o

4. Decane (C10) 174o

5. Naphthalene 218o

6. Dodecane (C12) 216o

Compounds Polar Aromatic HydrogenBonding Dipole

Toluene no yes no inducedHexanol yes no yes yesPhenol yes yes yes yesDecane no no no noNaphthalene no yes no inducedDodecane no no no no

?

Page 16: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

14% Cyanopropylphenyl

Strong DispersionNone/Strong Dipole (Ph/CNPr)Weak/Moderate H Bonding (Ph/CNPr)

0 2 4 6 8 10 12 14 16

0 2 4 6 8 10 12 14 16

100% Methyl

12

34 5

6

1 2 4 63

514% Cyano-propylphenyl

Strong DispersionNo DipoleNo H Bonding

1. Toluene2. Hexanol3. Phenol4. Decane (C10)5. Naphthalene6. Dodecane (C12)

Compounds Polar Aromatic HydrogenBonding Dipole

Toluene no yes no inducedHexanol yes no yes yesPhenol yes yes yes yesDecane no no no noNaphthalene no yes no inducedDodecane no no no no

?

Page 17: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

50% Cyanopropyl

Strong DispersionStrong DipoleModerate H Bonding

0 2 4 6 8 10 12 14 16

0 2 4 6 8 10 12 14 16

100% Methyl

12

34

56

4 16 2

5 350%Cyanopropyl

Strong DispersionNo DipoleNo H Bonding

1. Toluene2. Hexanol3. Phenol4. Decane (C10)5. Naphthalene6. Dodecane (C12)

Compounds Polar Aromatic HydrogenBonding Dipole

Toluene no yes no inducedHexanol yes no yes yesPhenol yes yes yes yesDecane no no no noNaphthalene no yes no inducedDodecane no no no no

?

Page 18: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

100% Polyethylene GlycolStrong DispersionStrong DipoleModerate H Bonding

0 2 4 6 8 10 12 14 16

41 6

2

5 3100% PEG

0 2 4 6 8 10 12 14 16

100% Methyl

12

34

56 Strong Dispersion

No DipoleNo H Bonding

1. Toluene2. Hexanol3. Phenol4. Decane (C10)5. Naphthalene6. Dodecane (C12)

Compounds Polar Aromatic HydrogenBonding Dipole

Toluene no yes no inducedHexanol yes no yes yesPhenol yes yes yes yesDecane no no no noNaphthalene no yes no inducedDodecane no no no no

?

Page 19: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Selectivity is important but not everything…

Inertness and Bleed can be critical factors in column selection.

Temperature limits will play a role as well.

Page 20: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Stationary Phase Bleed

• A thermodynamic equilibrium process that occurs to some degree in all columns, and is proportional to the mass amount of stationary phase inside the capillary tubing/carrier gas flow path

• Polysiloxane backbone releases low molecular weight, cyclic fragments

• Is negligible in low temperature, O2-free, clean GC systems

• Increased by increased temperature, oxygen exposure, or chemical damage

Page 21: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Bleed: Why Does It Happen?“Back Biting” Mechanism of Product Formation

+

Repeat

Si Si Si Si Si Si SiO O O O O O OH

CH3 CH3 CH3 CH3 CH3 CH3 CH3

CH3 CH3 CH3 CH3 CH3 CH3 CH3

OHSiSi O Si O Si O

CH3 CH3 CH3 CH3

CH3 CH3 CH3 CH3

SiO O

O

CH3H3C

H3C

H3C

SiSiCH3

CH3

OO O O O

OSi

HO

H3C

CH3 CH3 CH3 CH3

Si SiSi Si Si SiCH3

CH3

CH3

CH3CH3 CH3

CH3

CH3

CH3

Cyclic products are thermodynamically

more stable!

Page 22: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

DB-5ms Structure

CH3

CH3

CH3 CH

3

CH3 CH

3

Si

Si

Si

Si

O

O

O

ODB-5 Structure

DB-55% Phenyl

CH3

CH3

CH3

CH3

CH3

CH3 CH

3

CH3

Si

Si

Si

SiO

O

DB-5ms Structure

DB-5ms 1.Increased stability 2.Different selectivity 3.Optimized to match DB-5

Page 23: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Solid line: DB-5ms 30 m x .25 mm I.D. x .25 m Dashed line:DB-5 30 m x .25 mm I.D. x .25 mOven: 60o C isothermalCarrier gas: H2 at 40 cm/sec

1: Ethylbenzene2: m-Xylene3: p-Xylene4: o-Xylene

Difference in Selectivity

Page 24: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Four Types Of Low Bleed Phases

Phases tailored to “mimic” currently existing polymers-Examples: DB-5ms, DB-35ms, DB-17ms, DB-225ms

Phases unrelated to any previously existing polymers-Examples: DB-XLB

Optimized manufacturing processes-DB-1ms, HP-1ms, HP-5ms

Hand selected columns

Page 25: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Benefits of Low Bleed Phases PAH Sensitivity Using DB-35MS

1. Naphthalene 2. Acenaphthylene 3. Acenaphthene 4. Fluorene 5. Phenanthrene 6. Anthracene 7. Fluoranthene 8. Pyrene 9. Benz[a]anthracene10. Chrysene11. Benzo[b]fluoranthene12. Benzo[k]fluoranthene13. Benzo[a]pyrene14. Indeno[1,2,3,-c,d]anthracene15. Dibenz[a,h]anthracene16. Benzo[g,h,i]perylene

Columns: 30 m x 0.32 mm x 0.35 um.Carrier: H2, constant flow, 5 psi at 100 oC.Injector: 275 oC, splitless, 1 ul , 0.5-5ppm.Oven: 100 oC to 250 oC (5 min.) at 15 oC/min.,; then to 320 oC (10 min.) at 7.5 oC/min.Detector: FID, 320 oC.

5 10 15 20 25 min.

DB-35MS

1

2

3

4

5 6

7

8

9 10

11

12 13 14

1516

Commercially Available 35% phenyl column

Benzo[ghi]peryleneS/N = 120

Benzo[ghi]peryleneS/N = 15

Page 26: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Benefits of Low Bleed Phases DB-35ms vs Standard 35% Phenyl

200000

400000

600000

800000

1000000

1200000

1400000

10.00 12.00 14.00 16.00 18.00 20.00 22.00

Standard 35% Phenyl

DB-35ms

Benzo[g,h,i]perylene, 1ng

Page 27: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

M/Z ->

Abundance

50

78

96

135

157

207

223

253

276

331

346

377

405

439

50 100 150 200 250 300 350 4000

10000

20000

30000

40000

50000

60000

70000

80000

90000

100000

110000

120000

130000

140000

150000

Scan 1118 (20.560 min): 3901004.D

Standard 35%Phenyl

M/Z ->

Abundance

78

119

138207

239

274

276

315377

50 100 150 200 250 300 350 4000

10000

20000

30000

40000

50000

60000

70000

80000

Scan 1138 (20.640 min): 3901004.D

DB-35ms

Higher Spectral Purity

Page 28: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Polarity vs Stability/Temperature Range

Polarity StabilityTemperature Range

Page 29: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Stationary Phase Selection

• Existing information• Selectivity/Polarity• Critical separations• Temperature limits• Application designed

Examples: DB-VRX, DB-MTBE, DB-TPH, DB-ALC1, DB-ALC2, DB-HTSimDis, DB-Dioxin, HP-VOC, etc.

Choose the column phase that gives the best separation but not at the cost of robustness or ruggedness.

Page 30: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Resolution

• N = (gas, L, rc)

• k = (T, df, rc)

= (T, phase)

RNk

ks = 4 11

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 31: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Resolution

• N = (gas, L, rc)

• k = (T, df, rc)

= (T, phase)

RNk

ks = 4 11

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 32: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Column Diameter - Theoretical Efficiency

k’ = 5

30 m

20 m

10 m

5 m

I.D. (mm)

0.05

0.10

0.20

0.25

0.32

0.45

0.53

0.18

n/m

23,160

11,580

5830

4630

3660

2840

2060

6,660

N ~ 112,000

N ~ 112,000

N ~ 112,000

N ~ 112,000

Total Plates

Page 33: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Column Diameter and Carrier Gas Flow

•Lower flow rates: Smaller diameter columns

•Higher flow rates: Larger diameter columns

Low flow rates : GC/MSHigh flow rates: Headspace, purge & trap

Page 34: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Diameter Summary

•If you decrease the inside diameter:•Efficiency Increase•Resolution Increase•Pressure Increase•Capacity Decrease•Flow rate Decrease

Page 35: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Film Thickness and Retention: Isothermal

Constant DiameterNormalized to 0.25 µm

Thickness (µm) Retention Change0.10 0.400.25 1.001.0 4.003.0 12.05.0 20.0

Page 36: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Film Thickness and Resolution

RWhen solute k > 5

Rdf

When solute k < 5

df

or T

or T

(early eluters)

(later eluters)

Page 37: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Film Thickness and Capacity

0.32 mm I.D.Like Polarity Phase/Solute

Thickness (µm) Capacity (ng)

0.10 50-100

0.25 125-250

0.50 250-300

1 500-1000

3 1500-3000

5 2500-5000

Page 38: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Film Thickness and Bleed

More stationary phase = More degradation products

Page 39: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Film Thickness and Inertness

1.0

active inactive active inactive

3.0

active inactive

0.25

Page 40: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Film Thickness Summary

If you increase the film thickness:

•Retention Increase•Resolution (k<5) Increase•Resolution (k>5) Decrease•Capacity Increase•Bleed Increase•Inertness Increase•Efficiency Decrease

Page 41: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Resolution

• N = (gas, L, rc)

• k = (T, df, rc)

= (T, phase)

RNk

ks = 4 11

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 42: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Resolution

• N = (gas, L, rc)

• k = (T, df, rc)

= (T, phase)

RNk

ks = 4 11

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 43: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Column Length and Resolution

Length X 4 = Resolution X 2

Rn L

t L

Page 44: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Column Length and Cost

15m30m

60m

£ £ £ £ £ £ £

Page 45: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Length Summary

If you Increase Length:

•Efficiency Increase•Resolution Increase•Analysis Time Increase•Pressure Increase•Cost Increase

Page 46: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Summary - Four Primary Selection Areas

Stationary Phase Type

Column Internal Diameter

Stationary Phase Film Thickness

Column Length

Page 47: The Right GC Column - Selection Essentials. Cylinders or Generators Typical Gas Chromatographic System Picking the appropriate stationary phase and optimum

Still Can’t Decide Which Column to Use????? …Get in touch with us!!!