3M Health Care AcademySM

© 3M 2017. All Rights Reserved 1

The Science of Speed:How Fast Biological Indicators have Changed Quality Control in Reprocessing

Janet PrustDirector – Global Standards

Page 2: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 2. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Devil’s CascadeRainy Lake

Northwestern Ontario

Page 3: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 3. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Hexagenia Mayfly

Page 4: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 4. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Common Loon

Page 5: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 5. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Tullibee Cisco

Page 6: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 6. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Freshwater Mussels - Ontario

Page 7: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

What do these four species have in common?

Bioindicators (indicator species) • Species or a group of species whose function, population or status can reveal

the qualitative status of the environment in which it lives.

What environment are they bioindicators for? • Clean water. Clean water indicators are sensitive to pollution and are

a good indicator of water quality changes.

Page 8: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 8. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Viability is reflective of it’s environment

Common Biological Indicator

Page 9: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 9

Learning Objectives

1. Review how biological indicators work

2. Understand how rapid readout biological indicators work

3. Discuss the biological indicators role in sterilization quality control.

Page 10: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Director – Global Standards

3M Health Care Employee

Medical Solutions Division

Janet Prust - Disclosure

Page 11: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 11. All Rights Reserved.25 September 2019© 3M 3M Confidential.

Page 12: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved© 3M 2016. All Rights Reserved3M Health Care AcademySM

12

Biological Indicator Basics

Page 13: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 13

Biological Indicator – Definition:

A test system containing viable microorganisms providing a defined resistance to a specified sterilization process.

ISO/TS 11139: 2018. Sterilization of health care products — Vocabulary. International Organization for Standardization.

Page 14: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Types of Biological Indicators

Spore stripsInoculated carriers Liquid BIs

Page 15: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Self-contained Biological Indicators

Cap

Filter

Growth Media

with pH Indicator

Spore Strip Sleeve

Page 16: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 16

The Microbiology of Biological Indicators

Page 17: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Bacterial Spores

Some types of bacteria can form spores to protect themselves from a hostile environment, such as lack of nutrients

• e.g. some Bacillus sp., Clostridium sp.

• Like nuts or seeds - no biological activity - dormant

Very tough and hard to kill – challenges the sterilization process

Page 18: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Bacterial Spores

Geobacillus stearothermophilus Bacillus atropheus

Too small to see . . . about 1-2 micrometers (.000001 meters)

Page 19: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 19

Growth Media

• Mixture of sugars and other nutrients

• Formula optimized for the biological indicator microorganisms

Page 20: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Bacterial Spores – Activation and Germination

Google Images

A spore “sprouts”A spore “sprouts”Like a seed sprouts . . .

Page 21: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Bacterial cell

Spore

Biochemical and physical transformation steps

Bacterial Spores – Activation and Germination

Page 22: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Bacterial Growth

Bacterial cell

Page 23: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved© 3M 2016. All Rights Reserved3M Health Care AcademySM

23

Biological Indicator Standards

Page 24: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 24

ISO 11138 Series

11138 – 2: Ethylene oxide

11138 – 3: Moist heat (steam)

11138 – 4: Dry heat

11138 - 5: Low temperature steam formaldehyde

CAN/CSA-ISO 11138-1:17 Scope

This document specifies general requirements for production, labelling, test methods and performance characteristics of biological indicators, including inoculated carriers and suspensions, and their components, to be used in the validation and routine monitoring of sterilization processes.

This document specifies basic and common requirements that are applicable to all parts of ISO 11138. Requirements for biological indicators for particular specified processes are provided in the relevant parts of ISO 11138. If no specific subsequent part is provided, this document applies.

Page 25: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 25

Biological Indicator Performance Requirements

Page 26: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Key BI Performance Parameters

Population (number of test organisms)• Number of organisms that must be killed to get a negative BI• The more organisms, the longer it takes to inactive the BI• Measured by distributing the spores on plates and counting• Expressed in exponents (e.g. 106 = 1,000,000)

D-value (resistance)• A measure of how resistant the organisms are to the process

(how hard to kill)• Measured as the time to reduce the population by 90%• Expressed in time units (usually minutes)

Page 27: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved© 3M 2016. All Rights Reserved3M Health Care AcademySM

27

Biological Indicator Incubation Time

Page 28: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 28

Biological Indicator IncubationProcess for providing the spores in the biological indicator with the perfect conditions to support the growth of the organisms

- Nutrients (food)- Optimal temperature- Optimal pH

The incubation conditions are designed to be able to recover a very small number of organisms that remain alive in the biological indicator after exposure to the sterilization process, if there is a sterilization process failure.

Page 29: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 29

Biological Indicator Incubation

These biological processes happen during incubation . . .

Page 30: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 30

Biological indicator incubation time is the time required to make a final determination that the BI is negative• If the biological indicator turns positive at any point, the test is

complete and appropriate action must be taken.

Biological Indicator Incubation Time

Page 31: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 31

Biological Indicator Incubation Time

“An incubation period is commonly recognized to be 7 days for established sterilization processes . . .”

ISO 11138 (2017) Sterilization of health care products – Biological indicators –Part 1: General requirements

Page 32: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 32

Reduced Incubation Time (RIT)

Reduced Incubation Time is any established biological indicator incubation time that is less than 7 days• For example: 48 hours, 24 hours, 3 hours, 1 hour or 24 minutes

• The biological indicator manufacturer compares (validates) the BI response at the RIT to the BI response after 7 days, against established criteria.

How is this validation accomplished?

Page 33: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 33

RIT Test Methods

Pharmacopeias- No methodology defined in USP, EP

ISO Standards- No specific methodology defined in ISO 11138 series standards- No specific methodology defined in ISO 14161- Standard in Draft – ISO CD 16342

U.S. Food and Drug Administration (FDA)- Method defined in FDA 510k Guidance document

Page 34: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 34

Fractional cycles and MPN

Page 6

Page 35: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 35

Fractional cycles and MPN

Page 36: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 36

U.S. FDA RIT VALIDATION

http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm071261.htm

Reduced incubation time validation is part of the legal requirements for biological

indicators used in hospitals in the United States.

Page 37: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 37

http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm071261.htm#a2

U.S. FDA –RIT Methodology

30+’s → MPN = 0.357• Range 0-3 spores/BI

80+’s → MPN = 1.609• Range 0-7 spores/BI

AAMI WG 4 Presentation to U.S. FDA; May 26, 2016

Page 38: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 38

Positive Biological Indicators• Any spores that survive the sterilization cycle (a sterilization cycle failure)

are exposed to the growth media, then germinate (change from spores to cells), and start to grow:

• As they germinate and grow, they create many chemical reactions, that can be detected as a “signal”, that the biological processes are underway.

Page 39: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 39

Positive Biological Indicator “signals”

TurbidityBacterial cells scatter light

Media Color ChangepH shift caused by metabolic by-products

Page 40: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved© 3M 2016. All Rights Reserved3M Health Care AcademySM

40

3M Rapid Readout Technology

Page 41: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 41

3M Rapid Readout Technology

• Introduced in 1991

• Now used to monitor 3 different sterilization technologies (steam, ethylene oxide, hydrogen peroxide)

• Electronic detection of a fluorescent signal produced by the action of an enzyme system active in the germination and outgrowth phases of the test organism- Geobacillus stearothermophilus (steam, hydrogen peroxide): alpha-

glucosidase- Bacillus atropheus (ethylene oxide): beta-glucosidase

Page 42: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 42

Glucosidase Enzyme Systems

Bacterial glucosidase enzymes play an important role in uptake and utilization of nutrients

“The enzyme activity observed by sterilization assurance assays appears likely to come from heat stable enzyme in the spore core and enzyme(s) synthesized in spore outgrowth.”

Setlow P, Korza G, Setlow B. Analysis of alpha-glucosidase enzyme activity used in a rapid test for steam sterilization assurance. J Appl Microbiol 2016; 120: 1326-1335.

Vihinen M, Mäntsälä P (1989) Microbial amylolytic enzymes. CritRev Biochem Mol Biol 24:329–418

Page 43: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved43

3M™ Attest™ Rapid Technology

Enzyme acts on a non-fluorescent substrate in the media, resulting in a fluorescent end product, which is

detected by the Auto-reader (“signal”)

Non-fluorescent substrate (MUG)

Fluorescent end-product (4-MU)

Page 44: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved44

Rapid Attest™ Technology

Enzyme acts on a non-fluorescent substrate in the media, resulting in a fluorescent end product, which is

detected by the Auto-reader (“signal”)

Non-fluorescent substrate (MUG)

Fluorescent end-product (4-MU)

Page 45: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 45

Positive Biological Indicator “Signals”

TurbidityBacterial cells scatter light

Media Color ChangepH shift caused by metabolic by-products

Page 46: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 46

Positive Biological Indicator “Signals”

TurbidityBacterial cells scatter light

Media Color ChangepH shift caused by metabolic by-products

FluorescenceFluorescent response generated by metabolic enzyme action

Page 47: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 47

Optimized for Signal Production and Detection

Page 48: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 48

Optimized for Signal Production and Detection

Media optimized to produce fluorescent signal

Page 49: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 49

Optimized for Digital Production and Detection

Physical design to concentrate the signal

Advanced optics and software to detect the signal

Page 50: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 50

Rapid Readout Biological Indicators Use Worldwide

Malaysia JapanChinaTaiwanAustraliaNew ZealandIndiaThailandIndonesiaKoreaSingaporePhilippinesVietnam

United StatesCanadaMexicoBrazilColombiaChilePeruArgentinaUruguayCaribbean Region

SpainItalyGreeceBelgiumFranceNetherlandsSaudi ArabiaTurkeyPolandSouth AfricaUKIrelandRussiaIsrael

Page 51: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved© 3M 2016. All Rights Reserved3M Health Care AcademySM

51

Use of Biological Indicators in Sterilization Quality Control

Page 52: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Sterilization Process Monitoring

Goal of the sterilization process?• To kill all microorganisms!

Page 53: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Biological Indicators

Chemical Indicators

Physical Monitors

Sterilization Process Monitoring Tools

Page 54: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Physical Monitors

•Information from sensors in the sterilizer chamber walls or drain (e.g. temperature, pressure)

•Can confirm that the correct cycle was selected

•Provides a record of the cycle • Sensor placement

Page 55: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Chemical Indicators

Test system that reveals change in one or more predefined process variables based on a chemical or physical change resulting from exposure to a processISO/TS 11139:2005 Sterilization of health care products — Vocabulary

Bowie-Dick Tests

Pack indicatorsProcess indicators

Page 56: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Biological indicatorsBiological indicators are recognized by most authorities as being closest to the ideal monitors of the sterilization process because they measure the sterilization process directly by using the most resistant microorganisms (i.e., Bacillus spores), and not by merely testing the physical and chemical conditions necessary for sterilization. Since the Bacillus spores used in biological indicators are more resistant and present in greater numbers than are the common microbial contaminants found on patient-care equipment, the demonstration that the biological indicator has been inactivated strongly implies that other potential pathogens in the load have been killed.

CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008

Page 57: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Biological indicators

Biological indicators are recognized by most authorities as being closest to the ideal monitors of the sterilization process because they measure the sterilization process directly by using the most resistant microorganisms (i.e., Bacillus spores), and not by merely testing the physical and chemical conditions necessary for sterilization. Since the Bacillus spores used in biological indicators are more resistant and present in greater numbers than are the common microbial contaminants found on patient-care equipment, the demonstration that the biological indicator has been inactivated strongly implies that other potential pathogens in the load have been killed.

CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008

Page 58: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Biological indicators

Biological indicators are recognized by most authorities as being closest to the ideal monitors of the sterilization process because they measure the sterilization process directly by using the most resistant microorganisms (i.e., Bacillus spores), and not by merely testing the physical and chemical conditions necessary for sterilization. Since the Bacillus spores used in biological indicators are more resistant and present in greater numbers than are the common microbial contaminants found on patient-care equipment, the demonstration that the biological indicator has been inactivated strongly implies that other potential pathogens in the load have been killed.

CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008

Page 59: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Process Challenge Devicesprocess challenge device (PCD): Item designed to constitute a defined resistance to a sterilization process and used to assess performance of the process.

ISO/TS 11139: 2005. Sterilization of health care products — Vocabulary. International Organization for Standardization.

Page 60: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 60

Use of PCDs

• Biological indicators should be inside of the load, but this is not feasible

• PCDs are easily retrieved without disturbing any of the load items

• PCDs make it more difficult for the sterilant to reach the BI, like the BI was located inside of the load

Page 61: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 61

Frequency of use – biological indicators

Steam sterilization

Test CAN/CSA-Z314.-18 U.S.: ANSI/AAMI ST 79 (2017) Comprehensive guide to steam sterilization and sterility assurance in health care facilities

Routine Efficacy Monitoring BI PCD. Each day the sterilizer is used for each type of cycle

BI PCD. Weekly, preferably daily; Each cycle type used.

Load release – non-implant Physical parameters. BI PCD with CI provides increased sterility assurance

Optional. PCD with BI, Type 5, Type 6.

Load release - implants Every load. PCD with BI + Type 5 or 6 CI. Quarantine until BI result is available.

Every load. PCD with BI + Type 5 or 6 CI. Quarantine until BI result is available.

Page 62: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Considerations for monitoring every load with a BI/PCD:• Uniform standard of care for all patients

• Highest level of quality control

• Reduce the possibility of monitoring mistakes (implant vs. non-implant)

• Fast readout BIs facilitate frequent monitoring

Note: 53% of US Tier A hospitals monitor every load with BI PCD.

Source: 3M survey 2017, 2018.

Page 63: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Considerations for monitoring every load with a BI/PCD:• Uniform standard of care for all patients

• Highest level of quality control

• Reduce the possibility of monitoring mistakes (implant vs. non-implant)

• Fast readout BIs facilitate frequent monitoring

7 days 2 days 3 hours 1 hour Minutes!

Page 64: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 64

Steam sterilization quality issues . . .

• Most steam sterilization failures are related to problems with the steam quality, rather than time or temperature (e.g. air, non-condensable gases, wet steam, dry steam)

• Physical monitors and most chemical indicators cannot detect these problems

• Poor steam quality reduces the killing power of the process. Biological indicators directly measure the effectiveness of the process, so they do detect steam quality problems.

Page 65: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 65

SUMMARY1. Biological indicators are the direct measure of lethality of

the sterilization process, because they measure the effect of the process on spores.

2. Biological indicators can detect cycle quality problems.

3. Biological indicators incubation times have been reduced by detecting biological signals faster.

4. Fast incubation times make BIs easier to use, and enable better quality control by increased monitoring frequency.

Page 66: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved

Spin-Off Knowledge Assessment Game!Step 1: Open your mobile phone browser and go to:

www.C3ping.com

Step 2: Enter code: 6965

Step 3: Wait for game to start

Step 4: Select a team

Step 5: Enter your first or a nickname

Page 67: The Science of Speed: How Fast Biological Indicators have ......ISO 11138 Series 11138 –2: Ethylene oxide 11138 –3: Moist heat (steam) 11138 –4: Dry heat 11138 -5: Low temperature

3M Health Care AcademySM

© 3M 2017. All Rights Reserved 67

Thank you for your time and attention.