--Turbidity is the optical property that causes light to be
lines through the sample.
.
7/28/2019 Turbiditate
2/27
MUNICIPAL
Water Treatment Plants
Wastewater Treatment Plants
INDUSTRIAL
Metals and Mining
Chemical Processing
Pulp and Paper
Power
Food and Bevera e
Pharmaceuticals
7/28/2019 Turbiditate
3/27
Plant Goal: Provide a Source of Safe and Reliable Drinking
.
Harmful Microorganisms Are Removed by a
Turbidity Measures the Adequacy of the Filtration
Why Measure Turbidity?
on or an mprove an c ency
Detect Filter Breakthrough Meet Regulatory and Governmental Requirements
7/28/2019 Turbiditate
4/27
--Two Design Methods For Reporting Turbidity:
.
International Standards (ISO 7027)
Comparison of USEPA 180.1 and ISO 7027 Method
Item EPA 180.1 ISO 7027
Light Source Tungsten LampLED or Other
Filtered Source
Wavelength 400-600 nm 860 30 nm
Characteristics
Long Warm Up
Time Low Stray Light
More sensitive to
Smaller Particles
Less Sensitive to
Smaller Particles
Color Interferences
Interferences
7/28/2019 Turbiditate
5/27
---- Different Units
NTU: Nephelometric Turbidity Units
FTU: Formazin Turbidity Units
FNU: Formazin Nephelometric Units
These Units, NTU, FTU, and FNU are Equivalent
Formazin is the Polymer Suspension Used to CalibrateTurbidity Sensors
Turbidity is Determined in a Nephelometric Instrument,
and Measures the Light that is Scattered at 90 from theLight Source
7/28/2019 Turbiditate
6/27
Total Suspended SolidsTotal Suspended Solids -- DefinitionDefinitionTotal Suspended SolidsTotal Suspended Solids -- DefinitionDefinition
Definition:
A Measure of the Total Mass of Particles in a Sample
Determination of TSS: Filtering a Volume of Sample and Weighing the Mass of the
Dried Residue Retained on the Filter
e aue s acua e rom e ur yMeasurement
the User
,
7/28/2019 Turbiditate
7/27
Turbidity is not the same as total suspended solids (TSS)
Both Turbidity and TSS Are Defined by the Method Used to MeasureThem
Filterable Material in the Sample
Turbidity is an Optical Measurement; it Depends on the Number of
For 100 mg of solid
particles in 1 L
TSS is 100 m /L.
TSS is st ill 100 mg/L.
Turbidity is different:
since there are more
Turbidity has
a certain value
particles and smaller
in size.
in half
7/28/2019 Turbiditate
8/27
Turbidity Depends on:
Number of Particles Per Volume
ze o ar c es
Wavelength of Light Used for Measurement
Color of Particles
Refractive Index of Particles Presence of Air Bubbles
7/28/2019 Turbiditate
9/27
Wine of Beer Was Determined by
Source and Look for SmallParticles.
7/28/2019 Turbiditate
10/27
Size: Smaller Thank 1/10 the Size: Approximately 1/4 the
Description: Symmetric
Description: Scattering Concentratedin Forward Direction
Large Particles
7/28/2019 Turbiditate
11/27
Light
Light Scattered
by Particles
Transmitted
Source
Light
Scattered
Light at 90
DetectorTurbidity is the measure of the amount of light scattered by
particles in a sample. A beam of light passes through a samplecontaining suspended particles. The particles interact with the
light and scatter the light. A detector at 90measures theintensity of the scattered light.
7/28/2019 Turbiditate
12/27
u yu y -- ..Collimator
Transmitted
Light
Scattered Li ht at 90
LightSource
Optical Filter
DetectorEPA Method 180.1 requires:
1. Tungsten Lamp Light Source with Filament Temperature Between 2200 - 2700 K3. Detector Response Between 400 and 600 nm
4. Scattered Light be Measured at 90 30
5. Total Path Length Through Sample Less Than 10 cm
7/28/2019 Turbiditate
13/27
u yu y --
TransmittedLight
LightScattered Li ht at 90
DetectorISO Method 7027 requires:
1. The Wavelength of the light to be 860 nm 30 nm2. The Measuring angle to be 90 2.5
3. Light Source Using a Laser, LED, or Tungsten Filament Lamp with Filter
7/28/2019 Turbiditate
14/27
Complete System Sin le or Dual Sensor In ut
Debubbler/Measuring Chamber
USEPA Method 180.1 or ISO Method ISO 7027
0 - XXX NTU Range
Resolution X.XXX NTU
Analyzer
Suitable for Wastewater / Discharge / Industrial /rocess a er pp ca ons
7/28/2019 Turbiditate
15/27
Turbidity Sensor
Debubbler/Measurement Chamber
7/28/2019 Turbiditate
16/27
Three Turbidity Calibration Methods:. ser repare o u on
2. Commercially Available Standard Solution3. Comparison with a Referee Turbidity Instrument
Note: The USEPA Does Not Recommend Calibrating an On-line Turbidityme er y omparson w a enc e e
Dry Calibration Standard Available for Checking the Performance
Total Suspended Solids Calibration Method:
Enter an Equation into the Analyzer to Convert Turbidity to a TSSReading
7/28/2019 Turbiditate
17/27
Turbidity Sensor
Debubbler/Measurement Chamber
7/28/2019 Turbiditate
18/27
Not Not
Recommended Good BEST METHOD!
Sample Line
Sample Line Sample Line
(water)
(water)
(water)
(water)
Sediment Sediment Sediment Sediment
Process PipeProcess Pipe Process PipeProcess Pipe
7/28/2019 Turbiditate
19/27
Quantity of analyzers depends
on w e er e wa er use s
in a closed circuit or from a
once-through natural source
Turbidity
7/28/2019 Turbiditate
20/27
--Polymer
Chlorine
Addition
Turbidity monitors the
River WaterOr GroundWater
present
Sludge
To CoolingSystem
TClT
ar er
Turbidity monitorsfor particulatebreakthrou h
7/28/2019 Turbiditate
21/27
CausticFerric
Tank
Clarifier
or e
Tank
Industrial
Waste Water
further
T
CleanWater
rea men
pH
Flocculants are used to remove
suspended solids from water. slime Optimum removal depends on pH.
Turbidity is used to control addition ofto landfill
7/28/2019 Turbiditate
22/27
Overview
Efficiency reductions and equipment failures due
to corrosion and deposition cost the power
indust ry billions of dollars every year.
Good water chemistry can help to
mitigate some of these effects
and is best achieved with
online liquid analysis
and control.
ApplicationsDemineralization (Make-up Water)
Condensate Returns
Cooling Water Systems
Scrubbers
Waste Treatment
MeasurementsConductivity, Cation Conducitivity,
pH & ORP, Dissolved Oxygen,
Chlorine, Ozone, Turbidity,
Silica, Sodium, Phosphate,
Hydrazine
7/28/2019 Turbiditate
23/27
--
Plant IntakeSettling basinFilter wastewater backwash
Filter bed effluent (individual and combined)
7/28/2019 Turbiditate
24/27
Two USEPA Rules to Monitor Turbidity for Drinking Water Treatment Plants--. ,
Combined Filter Effluent (CFE) Water Turbidity:
7/28/2019 Turbiditate
25/27
--Second Rule ..
2. Plants Serving
7/28/2019 Turbiditate
26/27
--Regulations for Turbidity Requirements for Drinking WaterTreatment Plants Outside the USA
These Plants Have Either ...
Adopted the USEPA Standards,
Developed Their Own Regulations, o Follow the World Health Organization (WHO)
Recommendations of 1 NTU