21
Measurement And Instrumentations DR. Eng. Samir Elshamy DR. Eng. Samir Elshamy 1

Types of Error

  • Upload
    samir

  • View
    222

  • Download
    1

Embed Size (px)

DESCRIPTION

error in measurements

Citation preview

Page 1: Types of Error

Measurement And

Instrumentations

DR. Eng. Samir Elshamy

DR. Eng. Samir Elshamy1

Page 2: Types of Error

Definition of TermsReadability of an instrument:This term indicates the closeness with which the scale of the instrument

may be read; an instrument with a 12-cm scale would have a higher readability than an instrument with a 6-cm scale and the same range.

The least count:is the smallest difference between two indications that can be detected on the instrument scale.

DR. Eng. Samir Elshamy2

Page 3: Types of Error

The sensitivity of an instrument:is the ratio of the linear movement of the pointer on an analog instrument to the change in the measured variable causing this motion.For example, a 1-mV recorder might have a 25-cm scale length. Its sensitivity would be 25 cm/mV

DR. Eng. Samir Elshamy3

Page 4: Types of Error

Accuracy and Precision

DR. Eng. Samir Elshamy4

Page 5: Types of Error

The accuracy of an instrument

indicates the deviation of the reading from a known input. Accuracy is

frequently expressed as a percentage of full-scale reading, so that a 100-kPa

pressure gage having an accuracy of 1 percent would be accurate within ±1

kPa over the entire range of the gage.

DR. Eng. Samir Elshamy5

Page 6: Types of Error

The precision of an instrument

indicates its ability to reproduce a certain reading with a given accuracy. As an

example of the distinction between precision and accuracy, consider the

measurement of a known voltage of 100 volts (V) with a certain meter. Four

readings are taken, and the indicated values are 104, 103, 105, and 105 V. From

these values it is seen that the instrument could not be depended on for an

accuracy of better than 5 percent (5 V), while a precision of ±1 percent is

indicated since the maximum deviation from the mean reading of 104 V is only

1 V. DR. Eng. Samir Elshamy6

Page 7: Types of Error

DR. Eng. Samir Elshamy7

Page 8: Types of Error

Calibration

The calibration of all instruments is important, for it affords the opportunity to check the instrument against a known standard and subsequently to reduce errors in accuracy. Calibration procedures involve a comparison of the particular instrument with either(1)a primary standard, (2) a secondary standard with a higher accuracy than the instrument to be calibrated(3) a known input source.

DR. Eng. Samir Elshamy8

Page 9: Types of Error

For example, a flow meter might be calibrated by

(1) comparing it with a standard flow-measurement facility

of the National Institute for Standards and Technology (NIST),

(2) comparing it with another flow meter of known accuracy,

(3) directly calibrating with a primary measurement such as

weighing a certain amount of water in a tank and recording the

time elapsed for this quantity to flow through the meter.

DR. Eng. Samir Elshamy9

Page 10: Types of Error

The Generalized Measurement System منظام القياس المعم

Most measurement systems may be divided into three parts:1. A detector-transducer stage, which detects the physical variable

and performs either a mechanical or an electrical transformation to convert the signal into a more usable form.

2. Some intermediate stage, (Signal conditioning)which modifies the direct signal by amplification, filtering, or other means so that a desirable output is available.

3. A final or terminating stage, which acts to indicate, record, or control the variable being measured. The output may also be digital or analog. DR. Eng. Samir Elshamy10

Page 11: Types of Error

Bourdon-tube pressure gage

1) The bourdon tube is the detector-transducer stage because it converts the pressure signal into a mechanical displacement of the tube.

2) The intermediate stage consists of the gearing arrangement, which amplifies the displacement of the end of the tube so that a relatively small displacement at that point produces as much as three-quarters of a revolution of the center gear.

3) The final indicator stage consists of the pointer and the dial arrangement, which, when calibrated with known pressure inputs, gives an indication of the pressure signal impressed on the bourdon tube.

DR. Eng. Samir Elshamy11

Page 12: Types of Error

DR. Eng. Samir Elshamy12

Page 13: Types of Error

DR. Eng. Samir Elshamy

Analysis of Experimental DataIn this section we present a discussion of some of the types of

errors that may be present in experimental data and begin to

indicate the way these data may be handled.

13

Page 14: Types of Error

DR. Eng. Samir Elshamy

Types of Errors

Random

Systematic

Gross error

Environmental

Observational

Instrumental

14

Page 15: Types of Error

DR. Eng. Samir Elshamy

An experimental error is an experimental error. If the experimenter knew what

the error was, he would correct it and it would no longer be an error. In other

words, the real errors in experimental data are those factors that are always

vague to some extent and carry some amount of uncertaintyعدم اليقين.

Our task is to determine just how uncertain a particular observation may be

and to devise a consistent way of specifying the uncertainty in analytical form.

A reasonable definition of experimental uncertainty may be taken as the

possible value the error may have. This uncertainty may vary a great deal

depending on the circumstances of the experiment. Perhaps it is better to

speak of experimental uncertainty instead of experimental error because the

magnitude of an error is always uncertain. Both terms are used in practice,

however, so the reader should be familiar with the meaning attached to the

terms and the ways that they relate to each other.

Error Analysis on a Commonsense Basis تحليل الخطأ علي أساس المنطق

15

Page 16: Types of Error

DR. Eng. Samir Elshamy16

Consider the calculation of electric power from

P = EIwhere E and I are measured asE = 100 V ± 2 VI = 10 A ± 0.2 AThe nominal value of the power is 100×10 = 1000W. By taking the worst possible variations in voltage and current, we could calculatePmax = (100 + 2)(10 + 0.2) = 1040.4 WPmin = (100 − 2)(10 − 0.2) = 960.4 WThus, using this method of calculationthe uncertainty in the power is= +4.04 % & −3.96 %.

Page 17: Types of Error

DR. Eng. Samir Elshamy17

Static error (A) :

is defined as the difference between the best measured value

and the true value A=A1-A

Where A1=measured value & A =true value

Relative static error(uncertainty)= (A/A)

Page 18: Types of Error

DR. Eng. Samir Elshamy18

The result R is a given function of the independent variables x1, x2, x3, . . . , xn. Thus, R = R(x1, x2, x3, . . . , xn)Let ωR be the uncertainty in the result and ω1, ω 2, . . . , ωn be the uncertainties in the independent variables.

If this relation is applied to the electric-power relation of the previous section, the expected uncertainty is 2.83 percent instead of 4.04 percent.

Page 19: Types of Error

DR. Eng. Samir Elshamy19

Example 3.1 :UNCERTAINTY OF RESISTANCE OF A COPPER WIRE.

The resistance of a certain size of copper wire is given as

R = R0[1 + α(T − 20)]

where R0 = 6 ± 0.3 percent is the resistance at 20◦C,

α = 0.004◦C−1 ± 1 percent is the temperature coefficient of

resistance, and the temperature of the wire is T = 30±1◦C.

Calculate the resistance of the wire and its uncertainty.

Page 20: Types of Error

DR. Eng. Samir Elshamy20

Page 21: Types of Error

Thanks for your kind attention

DR. Eng. Samir Elshamy21