> Freezing down to -86°C > Innovative evaporator technology UF I Ultra Deep Freezers 292.9707.85 09/2015 BIOMEDICAL REFRIGERATION Innovative and reliable refrigeration solutions www.bmedicalsystems.com Your global partner to save lives. G G B M e d i c a l S y s t e m s - P A T E N T E D

UF I Ultra Deep Freezers - sakuf-trading.comsakuf-trading.com/brochures/BMedical/UF-Ultra-Deep-Freezers.pdf · GMP Clean Room Class B / ISO 6 (ISO EN 14644-1) GMP Clean Room Class

  • Upload
    others

  • View
    20

  • Download
    0

Embed Size (px)

Citation preview

Page 1: UF I Ultra Deep Freezers - sakuf-trading.comsakuf-trading.com/brochures/BMedical/UF-Ultra-Deep-Freezers.pdf · GMP Clean Room Class B / ISO 6 (ISO EN 14644-1) GMP Clean Room Class

> Freezing down to -86°C> Innovative evaporator technology

UF I Ultra Deep Freezers

292.

9707

.85

09/2

015

BIOMEDICAL REFRIGERATIONInnovative and reliable refrigeration solutions

www.bmedicalsystems.com

Your global partner to save lives.

S G G

B M

ed

ica l Syste ms - PATENT

ED

Page 2: UF I Ultra Deep Freezers - sakuf-trading.comsakuf-trading.com/brochures/BMedical/UF-Ultra-Deep-Freezers.pdf · GMP Clean Room Class B / ISO 6 (ISO EN 14644-1) GMP Clean Room Class

02 | 03 BIOMEDICAL REFRIGERATION

B Medical Systems UF I -86°C Ultra Deep Freezers

Your global partner to save lives.

UF 455 G / GG

UF 755 G / GG

Ultra Deep models for freezing down to -86°C

The models UF 455 G / GG & UF 755 G / GG reflect the highest and most uncompromising requirements in state-of-the-art technology and economy.

Designed as a cascade system with two hermetically sealed compressors, the cooling system is optimally designed with respect to energy consumption as well as to the development of waste heat and noise.

Thanks to their innovative and unique evaporator system and the application of most modern VIP technology the UF models ensures a balanced tem-perature stability throughout the entire refrigeration chamber and this at ambient temperatures up to +32°C (even up to +43°C with the water-cooled ver-sion).

Due to the highly efficient compressors and the fan motor, the UF 755 GG for example consumes 35% less energy compared to its predecessor. Meaning that it has the lowest energy consumption on the whole market. By means of excellent thermodynamic features of the new refrigerants and of the optimized compressors, the cooling performance has been re-improved what imports low compressor running times and more safety for the valuable preparations.

The UF 755 GG is the first really environmentally friendly ultra deep freezer market-wide with GS cer-tification. In addition this product is equipped with ATEX-conform components for highest user safety.

Control of the storage temperature and documen-tation of the temperature changes can – depend-ing on the application – be carried out via an optional temperature recorder (in form of a cir-cular chart recorder) or via the optional DCU. This through the Monitoring & Visualization Software DMN, which – in addition to the temperature record-er history – also records an event and activity history.

Max. number of Cryo-Racks : 15 Pieces

Max. number of Cryo-Racks : 25 Pieces

Mixed inventory with different Cryo-Rack systems for different box heights possible.

(maximum floor space 136 x 136 mm)

Cryo-Rack Box height 50 mm for 20 Boxes

Cryo-Rack Box height 75 mm for 15 Boxes

Arrangement Possibilities

Model UF 455 G UF 455  GG UF 755 G UF 755  GGGMP Clean Room Class B / ISO 6 (ISO EN 14644-1)

GMP Clean Room Class A / ISO 5 (ISO EN 14644-1), with water cooling

B Medical Systems Electronic

Key-operated power switch (power ON/OFF)

Safety door lock

Digital temperature indicator (display: 0.1 digits)Self-contained alarm system with integrated battery takes over the alarm functionand temperature value measurements in case of power failure for at least 48 hoursAcoustic/visual alarm signal in case of temperature alarm and power failureAll relevant data of temperature alarm and power failure alarm are stored in the alarm history. Such as date and time of start and end, min. max and average temperatureAlarm function test: simulation of a temperature rise or drop in order to test the alarm functionality

Control via self-diagnostic system

Door opening alarmRemote transmission alarm signal (via potential-free contact) in case of temperature alarm (change-over contact)Remote transmission alarm signal (via potential-free contact) in case of power failure (change-over contact)Automatic closing of the front door below a door opening angle of 90°

Interior made from stainless steel

Separate internal doors for prevention of cold loss when the front door opens

Climate class (ambient temperature range) SN (+10°C to +32°C)

Smooth castors with stabilizers for optimum flexibility of movementRS 485 interface for the display of all operating and control functions (hardware and software settings) via DMN monitoring software on a peripheral device (computer)DMN Software package

DCU - Data Communication Unit standard optional - not available

S G G S G G

CERTIFIED

REINRAUMKLASSIFIZIERUNGCLASSIFICATION DES SALLES BLANCHES

CLEAN ROOM CLASSIFICATION

The Safety Standards developed by B Medical Systems define certain significant technical features of a product. These ensure the safe storage of the preparations as well as the trend-setting safety of the user.

The B Medical Systems Gold Safety Standard efficiently complements the safety require-ments of the B Medical Systems Silver Safety Standard and therefore exceeds even the official standards. Gold models are denomi-nated with a “G”.

The new “green” models (denominated with a “G”) convince by their technical optimiza-tions in terms of economy and environmen-tal protection. Characteristic features are :

use of natural gases as refrigerants

40-60% less energy consumption

up to 40% less power needed

over 80% less heat ejection

In addition, the new “green” models stand out because of improved hold over times thanks to optimized door insula-tion and drastically reduced noise level for more workplace convenience.

Subject to change without prior notice. Some of the accessories shown in the pictures are optional.

Page 3: UF I Ultra Deep Freezers - sakuf-trading.comsakuf-trading.com/brochures/BMedical/UF-Ultra-Deep-Freezers.pdf · GMP Clean Room Class B / ISO 6 (ISO EN 14644-1) GMP Clean Room Class

04 | 05 BIOMEDICAL REFRIGERATION

B Medical Systems Electronic B Medical Systems UF I Technical Data

Your global partner to save lives.

UF 455 G UF 455 GG UF 755 G UF 755 GGGross volume 453 l 753 lNet volume 440 l 733 lExternal dimensions H x W x D 1450 x 910 x 950 mm 1990 x 910 x 950 mmExternal dimensions without door H x W x D 1450 x 910 x 865 mm 1990 x 910 x 865 mmInner dimensions H x W x D 812 x 730 x 760 mm 1352 x 730 x 760 mmNet weight with standard equipment 271 Kg 257 Kg 337 Kg 324 KgSet temperature (preset) -82°C -82°CSet temperature (setting range) can be adjusted in steps of 0.5°C -20°C to -86°C -20°C to -86°CTemperature cold / warm alarm limit (preset) -90°C / -65°C -90°C / -65°CControl sensor PT1000 2-WIRE 1/3DIN CL.B PT1000 2-WIRE 1/3DIN CL.BPrecision (at -80°C) ± 0.2°C ± 0.2°CDisplay sensor PT1000 2-WIRE 1/3DIN CL.B PT1000 2-WIRE 1/3DIN CL.BPrecision (at -80°C) ± 0.2°C ± 0.2°CVoltage 220-240 V - 50 Hz (16 A) 220-240 V - 50 Hz (16 A)Power 1100 W 900 W 1200 W 900 WEnergy consumption 17.5 kWh/24h 12 kWh/24h 19 kWh/24h 13 kWh/24hHeat emission 560 kcal/h 425 kcal/h 680 kcal/h 465 kcal/hCompressor running time 60% 55% 65% 60%Noise level at 1m height & 1m distance 59 dB(A) 53 dB(A) 60 dB(A) 54 dB(A)Accu data / function time of the control panel when power failure 12V - 7 AH / 48h 12V - 7 AH / 48hClimate class (ambient temperature range) SN (+10°C to +32°C) SN (+10°C to +32°C)Relative humidity at ambient temperature ≤ 75% ≤ 75%Defrosting technique manuel manuelRefrigerant type R404a/R508b R290/R170 R404a/R508b R290/R170Door insulation 65 mm PU + 20 mm VIP 65 mm PU + 20 mm VIPCasing insulation (polyurethane) 70 mm PU + 20 mm VIP 70 mm PU + 20 mm VIPHold over time (from -82°C to -60°C) 120/180 min. 120/210 min.Safety class I IEMC directive 2004 / 108 / EEC 2004 / 108 / EECLow voltage directive 2006 / 95 / EEC 2006 / 95 / EECMaterial inner body and inserts Stainless steel (V2A - 1.4301) Stainless steel (V2A - 1.4301)Material internal doors ABS ABSMaterial outer casing & door STO2Z-AZ150 STO2Z-AZ150Color outer casing White (similar RAL 9010) White (similar RAL 9010)Colour internal doors Grey (similar to RAL7035) Grey (similar to RAL7035)Color contrasts Blue (similar RAL 5002) Blue (similar RAL 5002)

Interior Equipment & Options (Concerning further information on accessories please see our separate leaflet “Racking & Storage Systems”)

Stainless steel compartment division / inserts 3 / 2 5 / 4 Seperate internal doors in order to minimize the loss of cold 3 5 Cryo Racks H50 15 (3 F & 5 R) 25 (3 F & 5 R) Cryo Racks H75 15 (3 F & 5 R) 25 (3 F & 5 R) ST-Racks, max. 15 25 GMP – clean room classification A / ISO 5 with water coolingGMP – clean room classification B / ISO 6DCU LAN/WLANDMN Software package Ambient temperature sensorPotential-free contact in case of power failure Integrated inlet for external sensor (installed by customer) Condenser filter Smooth castors with stabilizers for optimum flexibility of movement Integrated temperature recorder in form of a circular chart recorder Integrated, for 24h or 7 days Integrated, for 24h or 7 days CO2 BackUp / Emergency cooling system internal internal Ice scraper External water cooling via connection to customer‘s on-site water supplyDoor hinge Right / Left / / Wooden packaging for ocean transport / export

standard / optional / – not available All values were measured at +25°C ambient temperature and without load (with inertial mass).

The new and innovative B Medical Systems Electronic (operation and control panel) assures thanks to its password protected settings menu optimum protection for your stored preparations. The menu structure of the modern and user-friendly graphic display offers a simple and intuitive utilization.

The new B Medical Systems Electronic also offers :

> A wide range of adjustment and diagnostic facilities as well as additional protection / warning operations (via external alarm operations, histories and individual dis-play signals).

> An optional PT 100 sensor inlet to show the sensor’s temperature data on the display as well as forwarding and further processing via a 4 ... 20 miliA outlet.

> An optional 20 miliA outlet to transmit temperature data of a sensor connected to the electronic.

> Connection facilities for additional (optional) tempera-ture sensors.

> DMN (Data Monitoring Network) and the (optional) DCU (Data Communication Unit) allows illustration of texts on the product’s display.

LED “Power on„ Power switch, key-operation

O = OFF / I = ON

Back button

LED “Alarm„ Navigation buttons Settings menu Extra menu

CO2 BackUp / Emergency cooling system (internal)

Backup cooling for the additional protection of tempera-ture-sensitive goods in storage against sudden tempera-ture rises within the working space. The emergency cool-ing system injects CO2 (boiling point/sublimation point: -78.92°C) into the working space if the internal tem-perature within the working space rises above a certain programmable alarm value (CO2 On). The injection stops automatically when the temperature within the working space has reached another certain programmable tem-perature value (CO2 Off).

Inclusive additional operation and control panel for all operational and control functions of the emergency cool-ing system inclusive foil-protected keyboard and digital temperature display. The emergency cooling system is powered by an integrated battery.

The On and Off temperatures within an area can be freely set within a range of -20°C to -75°C.

Default setting : CO2 On : -60°C

CO2 Off : -70°C

The CO2 injection is stopped automatically when the front door is opened.

Power switch ON/OFFRed LED

Green LEDDisplay temperature

Red LED

Green LEDNavigation buttons

Subject to change without prior notice. Some of the accessories shown in the pictures are optional.

S G G

Page 4: UF I Ultra Deep Freezers - sakuf-trading.comsakuf-trading.com/brochures/BMedical/UF-Ultra-Deep-Freezers.pdf · GMP Clean Room Class B / ISO 6 (ISO EN 14644-1) GMP Clean Room Class

06 | 07 BIOMEDICAL REFRIGERATION

B Medical Systems UF I -86°C Features / Equipment / Options (extract)

Your global partner to save lives.

Surface : Temperature Contour : Temperature

Example of “Stable running (temperature stability)”

For all models, temperature distribution graphs for the stages “Cool Down (Cooling)”, “Stable Running (temperature stability)” and “Hold Over (temperature graphs, for example in case of power failure)” are available upon inquiry for various voltages. Determined at 5 measuring points per level or compartment (empty).

Protection against cold losses

The separate interior doors (with PU foam insulation) allow for the targeted removal of preparations, thus protecting the remaining stored products against cold losses caused physically by the opening of the front door. The combination of a heated high-grade sili-con seal (designed for temperatures down to -110°C) along the entire door frame and the separate interior doors with PU foam insula-tion ensures the absolute tightness of the system and prevents from ice forming in the device.

Example

Model : UF 755 GG (230V/50Hz)

Ambient temperature : +25°C

Rel.humidity : 45%

Set point : -82°C

Cold alarm : -91°C

Warm alarm : -65°C

Inner cabinet wall = evaporator

PU foam

Vacuum Insulation Panel

Outer casing

VIP (Vacuum Insulation Panel)

Evaporator Technology

Innovative “Rollbond” evaporator technology of the highest grade (stainless steel) for constant temperatures and uniform temperature distribution.

Temperature recorder (in form of a circular chart recorder) (optional, integrated for UF model range)

Water cooling, external (ex factory) (optional, for UF model range)

ST-Rack, Stainless steel rack

Remote temperature and power failure alarm (standard)

Cryo-Rack Stainless steel rack for Cryo-Box H50, H75 & H100

Ice scraper (standard)

Subject to change without prior notice. Some of the accessories shown in the pictures are optional.

Page 5: UF I Ultra Deep Freezers - sakuf-trading.comsakuf-trading.com/brochures/BMedical/UF-Ultra-Deep-Freezers.pdf · GMP Clean Room Class B / ISO 6 (ISO EN 14644-1) GMP Clean Room Class

17, Op der HeiL - 9809 Hosingen, Luxembourg

Tel. : + 352 92 07 31-1Fax : + 352 92 07 31-300

[email protected]

B Medical Systems S.à r.l.

B Medical Systems Software

All r

ight

s res

erve

d. A

ll in

dica

tions

with

out g

uara

ntee

and

exc

ludi

ng a

ll lia

bilit

y du

e to

faul

ty, i

ncom

plet

e or

out

date

d in

form

ation

. All

tech

nica

l dat

a su

bjec

t to

chan

ge w

ithou

t prio

r noti

ce.

> Interface connection of B Medical Systems appliances to an existing network.

> The DCU offers direct connection to the Ethernet, even wireless, to the serial BUS RS 485, as well as to the central building control system (4 ... 20 mA).

> Possibility of connection of actors (4 ... 20 mA out).

> Digital IN/OUT (customer-specific use of these connections is programmable).

> The integrated USB port allows stored data to be written to an external memory stick.

> Recording and storage of relevant data of the appliance.

> The DCU replaces the paper temperature recorder.

> The DCU works with all Gold electronics from 2000 on.

> All data are recorded and saved in the data base of the DMN and are available for analysis at any time.

> Possibility of connection of several additional self-sufficient temperature sensors (up to 4 PT1000 & 2 PT100).

Your essential advantages :

One integrative system for collecting all temperature relevant appliances and ambients.

Many different connection facilities allow flexible upgrades for individual projects.

DCU - Data Communication UnitHardware module that notes all operating conditions and passes them through to a central data base – via local network, on which devices are connected.

DMN & DCU in combination offer a highly flexible system that is adaptable to specific customer requirements> Complete & legally safe documentation of temperature data> Comprehensive applications and diagnostic possibilities nts.

DMN - Data Monitoring NetworkUniversal software for collection, long-term recording and visualization of temperature data.

> Complete activity list (password protected).

> Integrated event and activity history of all appliance components.

> Graphical visualisation of all temperature curves.

> Connection to existing or third-party appliances via network technology (LAN, WLAN, WAN).

> Simultaneous data monitoring and recording.

> Possibility for specific and individually configurable alarm forwardings, e. g. via email, SMS (with optional GSM module) or via DECT.

> Simple and intuitive utilization.

> Essential price advantage compared to a traditional circular chart recorder and its spare parts.

> Free of charge for all Gold & Silver ranges.

> Real-time temperature output for third-party software.

Your essential advantages :

Access to the data within your entire network via one central database.

Economy of time and money as regular changes of recorder paper, ink and battery is not necessary.

Ethernet LAN

USB

CON1: DC inletCON2: Electronic

CON3: -----CON4-6: Add. sensors

WLAN (option)

CON7: RS 232CON8: RS 485