Ultrasound BasisUltrasound Basis

Michel SlamaMichel SlamaAmiensAmiens

Page 2: Ultrasound Basis Michel Slama Amiens
Page 3: Ultrasound Basis Michel Slama Amiens

EchocardiographEchocardiographyy

Page 4: Ultrasound Basis Michel Slama Amiens

UltrasoundsUltrasounds 4 critical points4 critical points

FréquencyFréquency EnergyEnergy ImpédanceImpédance PRFPRF

Page 5: Ultrasound Basis Michel Slama Amiens

FrequencyFrequency

Page 6: Ultrasound Basis Michel Slama Amiens

Wavelength.Wavelength. Audible sound 15 Hz-20,000 Audible sound 15 Hz-20,000

Hz.Hz. Ultrasound Above 20 KHz.Ultrasound Above 20 KHz. Medical Ultrasound- 1-20 MHz.Medical Ultrasound- 1-20 MHz. Velocity= Wavelength x Velocity= Wavelength x

FrequencyFrequencyv = v = x Hz x Hz

PHYSICS OF PHYSICS OF ULTRASOUNDULTRASOUND

Page 7: Ultrasound Basis Michel Slama Amiens

Propagation speedFrequency

FREQUENCY

Ultrasound in medicine from 1 to 10 MHz

Page 8: Ultrasound Basis Michel Slama Amiens

Velocity of soundVelocity of sound Dependent on Dependent on

physical make up of physical make up of material.material.

Velocity is inversely Velocity is inversely proportional to proportional to compressibilitycompressibility

Directly Directly proportional to proportional to densitydensity

In body it is almost In body it is almost constant.constant.

Page 9: Ultrasound Basis Michel Slama Amiens

MessagesMessages

Propagation velocity is high in bonesPropagation velocity is high in bones Propagation velocity is low in the airPropagation velocity is low in the air In human tissue 1540 m/sIn human tissue 1540 m/s

Page 10: Ultrasound Basis Michel Slama Amiens

Axial resolutionAxial resolution

Page 11: Ultrasound Basis Michel Slama Amiens

Axial resolutionAxial resolution

Small wavelenght means high frequency

Page 12: Ultrasound Basis Michel Slama Amiens

Frequency and axial Frequency and axial resolutionresolution

λλ = C/F = C/F C in human tissu is 1540 m/sC in human tissu is 1540 m/s Frequency 2 MHz, Frequency 2 MHz, λλ = 0.77 mm. = 0.77 mm. Frequency 5 MHz, Frequency 5 MHz, λλ = 0.31 mm. = 0.31 mm. Frequency 10 MHz, Frequency 10 MHz, λλ = 0.15 mm. = 0.15 mm.

Higher frequency, Higher axial Higher frequency, Higher axial resolutionresolution

Page 13: Ultrasound Basis Michel Slama Amiens

MessageMessage

Higher the frequency is, Higher the frequency is, higher is the axial resolutionhigher is the axial resolution

Page 14: Ultrasound Basis Michel Slama Amiens

EnergyEnergy

Page 15: Ultrasound Basis Michel Slama Amiens

EnergieEnergie

An ultrasound is emitted with a An ultrasound is emitted with a certain energy certain energy

This energy is lost in tissuesThis energy is lost in tissues Conversion of US to HeatConversion of US to Heat

Page 16: Ultrasound Basis Michel Slama Amiens

ATTENUATION OF ATTENUATION OF ULTRASOUNDULTRASOUND

Exponentially with depth of travel,Exponentially with depth of travel, Dependent on Absorption and Dependent on Absorption and

ScatterScatter

On average for every 1 MHz, there On average for every 1 MHz, there is 1 db/cm loss, eg. 3.5 MHz loss in is 1 db/cm loss, eg. 3.5 MHz loss in 2 cm of tissue is 7 db.2 cm of tissue is 7 db.

Page 17: Ultrasound Basis Michel Slama Amiens
Page 18: Ultrasound Basis Michel Slama Amiens

MessageMessage

High frequency probe has low High frequency probe has low penetrationpenetration

Low frequenc probe has high Low frequenc probe has high penetration.penetration.

Page 19: Ultrasound Basis Michel Slama Amiens

MessageMessage

A balance should be made between A balance should be made between Best Resolution ( High frequency) Best Resolution ( High frequency) and propagation depth ( Lower and propagation depth ( Lower frequency)frequency)

Page 20: Ultrasound Basis Michel Slama Amiens

ImpedanceImpedance

Page 21: Ultrasound Basis Michel Slama Amiens
Page 22: Ultrasound Basis Michel Slama Amiens
Page 23: Ultrasound Basis Michel Slama Amiens

MessageMessage

To see any structure this structure To see any structure this structure should have a different impedance with should have a different impedance with the previous structure the previous structure

Page 24: Ultrasound Basis Michel Slama Amiens

AngleAngle

Page 25: Ultrasound Basis Michel Slama Amiens

Perpendicular

reflexion

Reflexion Diffusion

Diffusion

Page 26: Ultrasound Basis Michel Slama Amiens

MessageMessage

To get the best signal it is better To get the best signal it is better to work perpendicularely to to work perpendicularely to studied structures when using studied structures when using echocardiographic techniqueechocardiographic technique

Page 27: Ultrasound Basis Michel Slama Amiens

Echocardiographic Echocardiographic modesmodes

Page 28: Ultrasound Basis Michel Slama Amiens

M-ModeM-Mode

Page 29: Ultrasound Basis Michel Slama Amiens

M-ModeM-Mode

Page 30: Ultrasound Basis Michel Slama Amiens

M-ModeM-Mode

Page 31: Ultrasound Basis Michel Slama Amiens

M-ModeM-Mode

LV

S

PW

LV

S

PW

Probe Probe

LV

S

PW

Page 32: Ultrasound Basis Michel Slama Amiens

M-ModeM-Mode

Shortening Fraction = (LVEDD – LVESD)/LVEDDShortening Fraction = (LVEDD – LVESD)/LVEDDVelocity of Fiber Shortening = SF/Ejection TimeVelocity of Fiber Shortening = SF/Ejection Time

Page 33: Ultrasound Basis Michel Slama Amiens

Echocardiography: principlesEchocardiography: principles

ProbeProbe

Left ventricleLeft ventricle

2-D2-D

ProbeProbe

M-modeM-mode

Page 34: Ultrasound Basis Michel Slama Amiens

512

Image Analysis

Better lateral

resolution proximal

than distal

Page 35: Ultrasound Basis Michel Slama Amiens

Image Analysis

Shorter sector planeHigher axial resolution

Page 36: Ultrasound Basis Michel Slama Amiens

Pulse Repetition Pulse Repetition Frequency Frequency

Page 37: Ultrasound Basis Michel Slama Amiens

Pulse

Page 38: Ultrasound Basis Michel Slama Amiens

EchocardiographyEchocardiography

Analysis of close structures permits Analysis of close structures permits high PRF and therefore high image high PRF and therefore high image rate.rate.

In contrast deep structure cannot be In contrast deep structure cannot be analysed high image rateanalysed high image rate

Page 39: Ultrasound Basis Michel Slama Amiens

DopplerDoppler

Page 40: Ultrasound Basis Michel Slama Amiens

VelocityVelocity Cardiac DopplerCardiac Doppler

F0

F1

Blood FlowV

V = (F1 - F0) x C/ 2 x F0 x cos a

Probe

Page 41: Ultrasound Basis Michel Slama Amiens
Page 42: Ultrasound Basis Michel Slama Amiens

DopplerDoppler Doppler ultrasound beam should in Doppler ultrasound beam should in

the alignement of the blood flowthe alignement of the blood flow Possible under estimation of the true Possible under estimation of the true

velocityvelocity No over estimationNo over estimation

Page 43: Ultrasound Basis Michel Slama Amiens

VelocityVelocity

Systolic flow recorded at aortic valve level 0 m/s

V max = 1 m/s

VTI : velocity time integral

Page 44: Ultrasound Basis Michel Slama Amiens

DopplerDoppler

Page 45: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler

One cristal which sends and receives One cristal which sends and receives alternatively the ultrasound wavealternatively the ultrasound wave

Page 46: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler To analyse correctly an ultrasound To analyse correctly an ultrasound

wave we need at lest to analyse ¼ of wave we need at lest to analyse ¼ of this wave.this wave.

Page 47: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler Consequences: Consequences:

High velocity of flow = high frequencyHigh velocity of flow = high frequencyLow velocity High

velocity

Page 48: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler Flows with high velocities cannot be Flows with high velocities cannot be

analysed by using pulsed Doppler. analysed by using pulsed Doppler. All physiological flows have low All physiological flows have low

velocity (1 m/s) and can be analysed velocity (1 m/s) and can be analysed by using pulsed Dopplerby using pulsed Doppler

In contrast pathological flows have In contrast pathological flows have high velocities and CANNOT be high velocities and CANNOT be analysed by using Pulsed Doppler.analysed by using Pulsed Doppler.

Page 49: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler Consequences: Consequences:

Close structureClose structure

Page 50: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler Consequences: Consequences:

Deep structureDeep structure

Page 51: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler Deep and fast flows CANNOT be Deep and fast flows CANNOT be

analysed by using pulsed Doppleranalysed by using pulsed Doppler Close flow CAN be analysed by using Close flow CAN be analysed by using

pulsed Doppler even if this flow is pulsed Doppler even if this flow is fastfast

Page 52: Ultrasound Basis Michel Slama Amiens

Pulsed DopplerPulsed Doppler AdvantagesAdvantages

Spacial Spacial resolutionresolution

Analyse of Analyse of physiological physiological flowsflows

Disadvantages Flows with

high velocities and deep cannot be analysed

Page 53: Ultrasound Basis Michel Slama Amiens

Continuous wave DopplerContinuous wave Doppler

Continuous emission of ultrasound Continuous emission of ultrasound beambeam

Continuous reception Continuous reception

Page 54: Ultrasound Basis Michel Slama Amiens

Continuous wave DopplerContinuous wave Doppler AdvantagesAdvantages

Analysis of flows Analysis of flows with high frequencywith high frequency

Analysis of close Analysis of close and deep flowsand deep flows

DisavantagesDisavantages Spacial ambiguitySpacial ambiguity

Page 55: Ultrasound Basis Michel Slama Amiens

Color DopplerColor Doppler

Page 56: Ultrasound Basis Michel Slama Amiens

Color DopplerColor Doppler

Page 57: Ultrasound Basis Michel Slama Amiens
Page 58: Ultrasound Basis Michel Slama Amiens

Color DopplerColor Doppler

Page 59: Ultrasound Basis Michel Slama Amiens

DopplerDoppler Pulsed Pulsed

DopplerDoppler

Continuous Continuous

DopplerDoppler

Color DopplerColor Doppler

0 m/s0 m/s

PositivesPositivesVelocitiesVelocities

NegativesNegativesVelocitiesVelocities

Page 60: Ultrasound Basis Michel Slama Amiens

Tissue Doppler ImagingTissue Doppler Imaging

Page 61: Ultrasound Basis Michel Slama Amiens

Tissue Doppler Imaging

Intensity

Frequency

Page 62: Ultrasound Basis Michel Slama Amiens
Page 63: Ultrasound Basis Michel Slama Amiens
Page 64: Ultrasound Basis Michel Slama Amiens

Dynamics of Dynamics of fluidsfluids

Page 65: Ultrasound Basis Michel Slama Amiens

Doppler : non-invasive "Swan-Doppler : non-invasive "Swan-Ganz"Ganz"

P1P1 P2P2

VV

P2 - P1 = 4 x V2

Page 66: Ultrasound Basis Michel Slama Amiens

Velocity= 3 m/s

TR

Page 67: Ultrasound Basis Michel Slama Amiens

VD

OD

Velocity= 3 m/s

TR

Page 68: Ultrasound Basis Michel Slama Amiens

2 mmHg

38 mmHg

Gradient = 36 mmHg dP = 4 V² Velocity= 3 m/s

TR

Page 69: Ultrasound Basis Michel Slama Amiens

IM

2 m/s

MR

Page 70: Ultrasound Basis Michel Slama Amiens

40 mm Hg

76 mmHg

36

IM

2 m/s

Gradient 16 mmHg

MR

Page 71: Ultrasound Basis Michel Slama Amiens

0 mm Hg

196 mmHg

196

IM

7 m/s

MR

Page 72: Ultrasound Basis Michel Slama Amiens
Page 73: Ultrasound Basis Michel Slama Amiens

V

P1

P2

Page 74: Ultrasound Basis Michel Slama Amiens

ConclusionsConclusions

Echocardiography : imageEchocardiography : image Pulsed and continuous Doppler : Pulsed and continuous Doppler :

hemodynamicshemodynamics Color Doppler : flowsColor Doppler : flows Tissue Doppler : myocardial Tissue Doppler : myocardial

functionfunction

Page 75: Ultrasound Basis Michel Slama Amiens

Other TechniquesOther Techniques DTIDTI ContrasteContraste Color kinesisColor kinesis

Page 76: Ultrasound Basis Michel Slama Amiens

Different Doppler Different Doppler techniquestechniques

Pulsed DopplerPulsed Doppler

Continuous Continuous

DopplerDoppler

Color DopplerColor Doppler

0 m/s0 m/s

Positives Positives VelocitiesVelocities

NegativeNegativevelocitiesvelocities

Page 77: Ultrasound Basis Michel Slama Amiens

Doppler Pulsé

Page 78: Ultrasound Basis Michel Slama Amiens

Doppler Continu

Page 79: Ultrasound Basis Michel Slama Amiens
Page 80: Ultrasound Basis Michel Slama Amiens
Page 81: Ultrasound Basis Michel Slama Amiens
Page 82: Ultrasound Basis Michel Slama Amiens
Page 83: Ultrasound Basis Michel Slama Amiens
Page 84: Ultrasound Basis Michel Slama Amiens
Page 85: Ultrasound Basis Michel Slama Amiens
Page 86: Ultrasound Basis Michel Slama Amiens
Page 87: Ultrasound Basis Michel Slama Amiens
Page 88: Ultrasound Basis Michel Slama Amiens