TQC Work Verify
Ultrasound Ultraprobe 10,000 InstructionsValve Inspection ABCD Sometimes E Method:
Rev. 1.0 3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES
2-5 min UE Systems Training
*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*
Needed: Ultraprobe 10,000 and CF card;Stethoscope Module; Rubber Focusing Probe;Headphones
1Extension Rods for Stethoscope Module whenneeded for Hard To Reach areas.2
Move Cursor to Function Bar by PushingSens.Knob 2X and Press Enter for Setup Menu3
If Using a Route, move Route to Flash Card, thenload to Ultraprobe by using Data Transer. Refer toInstructions "Loading Route."
4
Page 1 of 6
TQC Work Verify
Ultrasound Ultraprobe 10,000 InstructionsValve Inspection ABCD Sometimes E Method:
Rev. 1.0 3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES
2-5 min UE Systems Training
*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*
Be sure Instrument Setup is in Automatic, ifBaselining Manual. Load Group (Route) toUltraprobe.
5If you are PERFORMING the BASELINE Readings,proceed to NEXT INSTRUCTION, if not READ thePicture Below.
6
Rotate for Application Select, Push Sens.Knob.Rotate to Valve, push Sens.Knob (Enable Valve).Press Enter to Exit.
7A clean memory is a must to keep files organized.It is Normal to Delete Records when changingapps.
8
Page 2 of 6
TQC Work Verify
Ultrasound Ultraprobe 10,000 InstructionsValve Inspection ABCD Sometimes E Method:
Rev. 1.0 3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES
2-5 min UE Systems Training
*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*
Rotate Sens. Knob back to Display Screens, PushSens. Knob, Rotate to ABCD, Push Sens.Knob toSelect/Exit.
9Rotate to Module Type Select, push Sens.Knob.Rotate to Stethoscope (STM); Push Sens toSelect/Exit
10
Rotate to Instrument Setup, Push Sens.Knob.Rotate for "Manual", Push Sens.Knob to Select/Exit11
Rotate to Enter Inspector ID [under "More" in SetupMenu], push Sens.Knob. Enter 3 Alpha-NumericCharacters use Sens.Knob.
12
Page 3 of 6
TQC Work Verify
Ultrasound Ultraprobe 10,000 InstructionsValve Inspection ABCD Sometimes E Method:
Rev. 1.0 3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES
2-5 min UE Systems Training
*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*
Press Enter 2X to Save/Exit and return to ValveABCD Display.13
IF Route has been baselined and established referto NOTE. Note does not hold true fortroubleshooting WITHOUT having a route loaded.
14
Push Sens.Knob 1X to move cursor to Frequency.Rotate to 25 kHz.15
Push Sens. 1X more [if cursor at S= push 2X]moving Cursor to Function Bar. Rotate Sens untilStore "A" in prep for readings.
16
Page 4 of 6
TQC Work Verify
Ultrasound Ultraprobe 10,000 InstructionsValve Inspection ABCD Sometimes E Method:
Rev. 1.0 3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES
2-5 min UE Systems Training
*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*
Push Sens.Knob 1X more to return cursor to S=(Operational Home Spot for Adjusting Sensitivity).17
Note Approximate Test Points for the ABCD,sometime E Method of testing Valves.18
NOTE: Choose Test Point A for the Valve19
Store Record: Adjust Sens. for dB reading ofPosition A,Once Stored, A dB appears in position'A' and advances to 'B.' Repeat for position B,C,D.
20
Page 5 of 6
TQC Work Verify
Ultrasound Ultraprobe 10,000 InstructionsValve Inspection ABCD Sometimes E Method:
Rev. 1.0 3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES
2-5 min UE Systems Training
*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*
Once" Finished Store D", Unit Advances to 'StoreRecord' to Permanently Store the Entire Record ofA,B,C,D points. Push 'Enter' to Store.
21Press "ENTER" To Store all Points to Record andto Advance to next Valve/Record of A,B,C,D points.22
Visual Instructions exist for Storing a Record.23
ALWAYS WEAR the HEADPHONES.24