1
URINE M-SPIKE QUANTIFICATION IN MULTIPLE MYELOMA : EVALUATION OF A QUANTIFICATION ON ELECTROPHORETIC PROFILE WITH URINE PROFILE AND BENCE JONES IMMUNOFIXATIONS H. Caillon 1 , A. Laude 1 , M. Melki 3 , P. Moreau 2 , T. Dejoie 1 1 Biochemistry Laboratory, University Hospital of Nantes, France 2 Clinic of Hematology, University Hospital of Nantes, France 3 Sebia, Lisses, France Presented at EuroMedLab Athens 2017. Poster code : M035 Background In multiple myeloma (MM), detection and quantification of monoclonal component (M-spike) by urine protein electrophoresis (UPE) and immunofixation are essential for diagnosis and response evaluation according to the International Myeloma Working Group (IMWG) criteria. Up to now, quantification of urine M-spike was only possible on electrophoretic profile with Hydragel High Resolution (HR) (Sebia). Recently, Sebia has developed the possibility to quantify M-spike from the electrophoretic track on urine agarose gel immunofixations (Urine Profile [UP] and Bence Jones [BJ]) with the Phoresis software 8.63. The purpose of this study was to evaluate these two methods of quantification and to determine if they could be used alternatively from HR. Conclusion This study shows that UP, BJ and HR are overall comparable for urine M-spike quantification. The availability of a quantification tool with UP and BJ gels provides the opportunity for many laboratories to measure urine M-spike with accuracy methods as long as the MM follow-up. For only a few samples, profiles can be slightly different between HR and UP/BJ leading to the classical recommendation in clinical biology, to mostly follow- up disease with the same method. Methods We included 90 urine samples in this study, with several profiles: Bence Jones proteinuria only, Bence Jones with glomerular or tubular proteinuria or both, Bence Jones with several degrees of polymerization (see the table beside). Each sample was tested within 2 days on HR, UP and BJ. Urine M-spike was measured with the Phoresis software 8.63 with the orthogonal method. Finally, 67 of the 90 samples exhibited a measurable M-spike. Concordance of UP and BJ quantification with the referent method HR was evaluated, using a correlation graph and a Bland and Altman diagram with deviants analysis (using SFBC recommendations). 3 operators made the M-spike quantification on HR, UP and BJ for each sample exhibiting a measurable monoclonal component, to determine the intra-operator CV for each method. Results Concordance of M-spike quantification between UP and HR, and BJ and HR respectively was excellent (correlation coefficients 0,98 for both). Example : Quantification : 0,43 g/L 0,38 g/L 0,47 g/L Only one deviant was observed for BJ-HR comparison and two for UP-HR comparison. Differences were due to the resolution characteristics in proteins separation (degrees of polymerization for Bence Jones proteinuria, association of Bence Jones with tubular microproteins) according to each gel. Inter-operator CV was quite low for HR, similar with BJ but higher for UP. Samples n (%) Samples with measurable monoclonal component (n,%) BJ proteinuria only 32 (35%) 26 (39%) BJ proteinuria with several degrees of polymerization 12 (13%) 10 (15%) BJ proteinuria with glomerular or tubular proteinuria or both 26 (29%) 20 (30%) BJ proteinuria with the complete immunoglobulin 10 (11%) 10 (15%) Complete immunoglobulin only 5 (6%) 1 (1%) Normal proteinuria 5 (6%) - Total 90 67 On 20 samples HR BJ UP Inter-operator CV for M- component quantification(%) 5,4 5,6 8,9

URINE M-SPIKE QUANTIFICATION IN MULTIPLE MYELOMA ... Euromedlab... · proteinuria only, Bence Jones with glomerular or tubular proteinuria or both, Bence Jones with several degrees

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

Page 1: URINE M-SPIKE QUANTIFICATION IN MULTIPLE MYELOMA ... Euromedlab... · proteinuria only, Bence Jones with glomerular or tubular proteinuria or both, Bence Jones with several degrees

URINE M-SPIKE QUANTIFICATION IN MULTIPLE MYELOMA : EVALUATION OF A QUANTIFICATION ON ELECTROPHORETIC PROFILE WITH URINE PROFILE AND BENCE JONES IMMUNOFIXATIONS

H. Caillon 1, A. Laude 1, M. Melki 3, P. Moreau 2, T. Dejoie 1

1Biochemistry Laboratory, University Hospital of Nantes, France 2Clinic of Hematology, University Hospital of Nantes, France – 3Sebia, Lisses, France

Presented at EuroMedLab – Athens 2017. Poster code : M035

Background

In multiple myeloma (MM), detection and quantification of monoclonal component (M-spike) by urine protein electrophoresis (UPE) and immunofixation are essential for diagnosis and response evaluation according to the International Myeloma Working Group (IMWG) criteria. Up to now, quantification of urine M-spike was only possible on electrophoretic profile with Hydragel High Resolution (HR) (Sebia). Recently, Sebia has developed the possibility to quantify M-spike from the electrophoretic track on urine agarose gel immunofixations (Urine Profile [UP] and Bence Jones [BJ]) with the Phoresis software 8.63. The purpose of this study was to evaluate these two methods of quantification and to determine if they could be used alternatively from HR.

Conclusion This study shows that UP, BJ and HR are overall comparable for urine M-spike quantification. The availability of a quantification tool with UP and BJ gels provides the opportunity for many laboratories to measure urine M-spike with accuracy methods as long as the MM follow-up. For only a few samples, profiles can be slightly different between HR and UP/BJ leading to the classical recommendation in clinical biology, to mostly follow-up disease with the same method.

Methods

We included 90 urine samples in this study, with several profiles: Bence Jones proteinuria only, Bence Jones with glomerular or tubular proteinuria or both, Bence Jones with several degrees of polymerization (see the table beside). Each sample was tested within 2 days on HR, UP and BJ. Urine M-spike was measured with the Phoresis software 8.63 with the orthogonal method. Finally, 67 of the 90 samples exhibited a measurable M-spike. Concordance of UP and BJ quantification with the referent method HR was evaluated, using a correlation graph and a Bland and Altman diagram with deviants analysis (using SFBC recommendations). 3 operators made the M-spike quantification on HR, UP and BJ for each sample exhibiting a measurable monoclonal component, to determine the intra-operator CV for each method.

Results

Concordance of M-spike quantification between UP and HR, and BJ and HR respectively was excellent (correlation coefficients 0,98 for both).

Example :

Quantification : 0,43 g/L 0,38 g/L 0,47 g/L

Only one deviant was observed for BJ-HR comparison and two for UP-HR comparison. Differences were due to the resolution characteristics in proteins separation (degrees of polymerization for Bence Jones proteinuria, association of Bence Jones with tubular microproteins) according to each gel. Inter-operator CV was quite low for HR, similar with BJ but higher for UP.

Samples n (%) Samples with measurable

monoclonal component (n,%)

BJ proteinuria only 32 (35%) 26 (39%)

BJ proteinuria with several degrees of polymerization

12 (13%) 10 (15%)

BJ proteinuria with glomerular or tubular proteinuria or both

26 (29%) 20 (30%)

BJ proteinuria with the complete immunoglobulin

10 (11%) 10 (15%)

Complete immunoglobulin only 5 (6%) 1 (1%)

Normal proteinuria 5 (6%) -

Total 90 67

On 20 samples HR BJ UP

Inter-operator CV for M-component quantification(%)

5,4 5,6 8,9