User Guide KI 734x Series Two way Loss & ORL Tester . www.kingfisher.com.au. Two-way Loss Tester. T he fastest and easiest two-way loss tester Real time display of bi-directional average multi- link loss on both instruments 3 year warranty & calibration cycle No warm up required - PowerPoint PPT Presentation
Citation preview
User Guide KI 734x Series
Two way Loss & ORL Tester
www.kingfisher.com.au
Two-way Loss Tester
The fastest and easiest two-way loss tester
Real time display of bi-directional average multi- link loss on both instruments 3 year warranty & calibration cycle No warm up required 360 hours battery life & interchangeable connectors 1200 storage space for dual test results
Fast & Easy For Bi-directional Testing
Step 1: Set referencesStep 2: Press the “Auto test” button
Real time bi-directional average multi- link loss display on both instruments
Step 1 - Set Reference At Both Ends
Two-way port
Meter port
Test lead
End A
End B
An optic fiber link
Source off
Source Return Loss Power Meter Auto test
- 52.73 dBm
1310nm
Source on
Auto test Menu Abs / Rel Hold
- 4.01 dBm
1310nm - +
-4.00 dBm
Source on
Auto test Menu Set Ref Abs / Rel Hold
- 0.00 dB R
1310nm - +
-4.00 dBm
Source off
Source Return Loss Power Meter Auto test
- 52.73 dBm
1310nm
Pushing “AUTOTEST”
Pushing “Abs / Rel”
Alternates between 1310 and 1550 nm if 1310 & 1550 nm lasers are installed
Holding “Set Ref” for 3 seconds to set reference for multi-
Pushing “Menu” to be back to the turn-on display
Set Reference Key Press Sequence
An optic fiber link
Step 2 - Measure Insertion And Return Loss
End A
End B
Source off
Source Return Loss Power Meter Auto test
- 52.73 dBm
1310nm
Pushing “Auto test” at one End.
Source on
Auto test Menu Hold
- 34.011310nm - +
-4.00 dBm
Alternates between 1310 and 1550 if 1310 & 1550 nm lasers are installed
- + Source on
Auto test Menu Hold
- 54.17 - +
-4.00 dBm
1310nm
Testing is completed for the link.
A ◄► BdB
A ◄ BdB RL
STORE
Pushing “STORE” to store average Loss, A-to-B Loss, B-to-A
Loss, ORL at end A, ORL at end B of multi-
Displays the average bi-directional link loss of multi-Displays Loss/Optical Power from A-to-B (‘A ►B’)Displays Loss/Optical Power from B-to-A (‘A◄ B’)
Displays ORL at A end (‘A ►’)Displays ORL at B end (‘◄ B’)
Back to the average bi-directional link loss display
Press ‘Select’
Press ‘M
enu’Measure Insertion And Return Loss -
Continue
End A
End B
An optic fiber link
Communication During Testing
Source on
Auto test Menu Hold
- 34.011310nm - +
-4.00 dBm
A ◄► BdB
After completing one link test,
pushing “Hold” will freeze thedisplay on both units and the
unitat End B beeps twice. This
tells End B that a test has been completed.
Source on
Auto test Menu Hold
- 34.01 - +
-4.00 dBm
A ◄► BdB
Then, End B connects to the next link and pushes “Hold”. This
releasesthe display on both units and the
unit at End A beeps once. This tells
End A that he is ready for the next test.
Increases the productivity in high fiber-count systems.
Set Ref In Autotest
Source on
Auto test Menu Set Ref Hold
- 34.011310nm - +
A ◄►BdB
Displays average link loss for multi-.Hold “Set Ref” for 3 seconds will set references on both units for multi-.
View Ref Value In Autotest
Source on
Auto test Menu Abs / Rel Hold
- 34.011310nm - +
A ►BdB R
Displays link loss and reference value for multi- for A-to-B.
Source on
Auto test Menu Abs / Rel Hold
- 32.241310nm - +
A ◄ BdB R
Displays link loss and reference value for multi- for B-to-A.
-8.20 dBm
-7.80 dBm
Displays the average bi-directional link loss of multi-Displays Loss/Optical Power from A-to-B (‘A ►B’)Displays Loss/Optical Power from B-to-A (‘A◄ B’)
Displays ORL at A end (‘A ►’)Displays ORL at B end (‘◄ B’)
Back to the average bi-directional link loss display
Press ‘Select’
Press ‘M
enu’
- +
Unit A reference value
Unit B reference value
Pass/Fail Assessment
In Two-way testing mode, press ‘Level’ to enter Pass/Fail display mode.To exit press ‘Level’ again, or press ‘Select’.
Source on
<<< Auto test Menu Hold
PASS1310nm - +
A
Level
Select
Insertion Loss Pass/Fail Threshold Set UpMulti- of Insertion Loss Pass/Fail criteria can
be set.
Shift Default Level
Cancel Set
Shift Source off
SEt Menu
15.731310nm - +
dB
Select another two-way operational wavelength to set up Average Insertion Loss Pass/Fail threshold.
Enter Power Meter mode.Select two-way operational wavelength.
Use ‘Shift’ + ‘Level’ to set Average Insertion Loss Pass/Fail threshold, push ‘Set’ to store.
- +
Return Loss Pass/Fail Threshold Set UpMulti- of Return Loss Pass/Fail criteria can be
set.
Shift Default Level
Cancel Set
Select another wavelength to set up Return Loss Pass/Fail threshold.
Use ‘Shift’ + ‘Level’ to set Return Loss Pass/Fail threshold, push ‘Set’ to store.
- +
Shift Source off
SEt Menu
-10.571310nm - +
dB RL
Enter Return Loss mode.Select wavelength.
The Contents of User Memory
Average loss for multi-
Reference of both units for multi-
ORL for multi-
Nominal output power for multi-
Location info by flashing symbol of either A or B
KITS™ testing software