user manual

www.ActiveForever.com

2

www.ActiveForever.com

3

Contents

Chapter 1 1 Intended Use, Contraindications and General Precautions

Chapter 2 3 Description of the Inogen One® G3 Oxygen Concentrator 3 ImportantPartsoftheInogenOne®G3OxygenConcentrator 4 UserInterfaces 5 Input/OutputConnections 6 PowerSupplyOptions 8 InogenOne®G3Accessories

Chapter 3 11 Operating Instructions 11 GeneralInstructions 15 AdditionalOperatingInstructions 18 BatteryOperatingInstructions 19 BatteryCareandMaintenanceChapter 4 21 Inogen One® G3 Oxygen Concentrator Audible and Visible Signals (including Alarms)

Chapter 5 29 Troubleshooting Chapter 6 31 Cleaning, Care and Maintenance 31 CannulaReplacement 31 CaseCleaning 32 FilterCleaningandReplacement 33 InogenOne®G3ColumnChangeProcedure 36 OtherServiceandMaintenance 36 DisposalofEquipmentandAccessories

Chapter 7 37 Symbols Used on Concentrator and Accessories

Chapter 8 39 Inogen One® G3 System Specifications

English

www.ActiveForever.com

4

www.ActiveForever.com

1

1 Intended Use, Contraindications and General Precautions

IntendedUse,ContraindicationsandGeneralPrecautionsChapter1

Intended Use

TheInogenOne®G3OxygenConcentratorisusedonaprescriptivebasisbypatientsrequiringsupplementaloxygen.Itsuppliesahighconcentrationofoxygenandisusedwithanasalcannulatochanneloxygenfromtheconcentratortothepatient.TheInogenOne®G3maybeusedinhome,institution,vehicle,onanairplaneandvariousmobileenvironments.

TheexpectedlifefortheInogenOne®G3OxygenSystemis5years,withtheexceptionofthesievebeds(metalcolumns)whichhaveanexpectedlifeof1yearandthebatteries,whichhaveanexpectedlifeof500fullcharge/dischargecycles.

CAUTION USAFederallawrestrictsthisdevicetosalebyorontheorderofaphysician.Mayalsobeapplicableinothercountries.

CAUTION Useofacannulaotherthanahighflowcannula(e.g.Salter1600Q)mayrestrictoxygendeliveryand/orattachmenttonozzlefittings.

wArNING Availabilityofanalternatesourceofoxygenisrecommendedincaseofpoweroutageormechanicalfailure.Consultyourequipmentproviderfortypeofback-upsystemrecommended.

CAUTION Itistheresponsibilityofthepatienttomakeback-uparrangementsforalternativeoxygensupplywhentraveling;Inogenassumesnoliabilityforpersonschoosingnottoadheretomanufacturerrecommendations.

! C

English

www.ActiveForever.com

2

Contraindications

wArNING ThisdeviceisNOTINTENDEDtobelifesustainingorlifesupporting. CAUTION Undercertaincircumstances,theuseofnon-prescribedoxygentherapy

canbehazardous.Thisdeviceshouldbeusedonlywhenprescribedbyaphysician.

CAUTION Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealarmsorcommunicatediscomfort.Ifthepatientshowsanysignsofdiscomfort,aphysicianshouldbeconsultedimmediately.

CAUTION TheInogenOne®G3isnotdesignedorspecifiedtobeusedinconjunctionwithahumidifier,nebulizerorconnectedwithanyotherequipment.Useofthisdevicewithahumidifier,nebulizerorconnectedwithanyotherequipmentmayimpairperformanceand/ordamagetheequipment.DonotmodifytheInogenOne®G3Concentrator.Anymodificationsperformedontheequipmentmayimpairperformanceordamageequipmentandwillvoidyourwarranty.

! C

! C

! C

General Precautions

wArNING Thedeviceproducesenrichedoxygengaswhichacceleratescombustion.DoNOTALLOWSMOKINGOROPENFLAMESwithin

10feetofthisdevicewhileinuse.

wArNING DonotsubmersetheInogenOne®G3oranyoftheaccessoriesinliquid.Donotexposetowaterorprecipitation.Donotoperateinexposedrain.Thiscouldleadtoelectricalshockand/ordamage.

CAUTION Donotuseoil,grease,orpetroleum-basedproductsonorneartheInogen

One®G3.

CAUTION NeverleavetheInogenOne®G3inanenvironmentwhichcanreachhightemperatures,suchasanunoccupiedcarinhightemperatureenvironments.Thiscoulddamagethedevice.

www.ActiveForever.com

3

Important Parts of the Inogen One® G3 Oxygen Concentrator

Description of the Inogen One® G3 Oxygen Concentrator2

On/Off Button

Display Back-light ButtonBreath

Detection Light

Display

Audible Alarm Button

Alert/Alarm Light

Flow Control

DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

English

DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

www.ActiveForever.com

4

User Controls

ON / OFF ButtonPressoncetoturn“ON”;Pressandholdforonesecondtoturn“OFF”.

Audible Alarm Button PressingthisbuttonwilltoggletheInogenOne®G3’sbreathdetectionaudiblealertonandoff: 1.DefaultMode.WhentheInogenOne®G3ispoweredup,thebreath detectionaudiblealertisdisabled.Thedisplay’smodeindicationarea willshowabelliconcrossedoutwithanXindefaultmode. 2.BreathDetectionAlertMode.TheInogenOne®G3willalertwith audibleandvisualsignalsfor“nobreathdetected”whenthismode isenabledandnobreathhasbeendetectedfor60seconds. At60seconds,thedevicewillenterintoautopulsemodeandonce anotherbreathisdetected,thedevicewillexitautopulsemodeand delivernormallyoninspiration.Thedisplay’smodeindicationareawill showabellicon,flashingredlightanddisplaymessagewhenthe alertisenabled. 3.Ifpowerislost,breathdetectionalertmoderesetsto defaultmode.Flow Setting Control Buttons Usethe–or+flowsettingcontrolbuttonstoselectthesettingasshownonthedisplay.Therearefoursettings,from1to4.

Display Backlight ButtonPushon;automaticallyturnsoffafter10seconds.

User Interfaces

DisplayThisscreendisplaysinformationregardingflowsetting,powerstatus,batterylifeanderrors.IfyouwouldliketochangethelanguageontheInogenLCDScreencontactInogenClientServices.

www.ActiveForever.com

5DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

User Interfaces (continued)

Indicator LightsAredlightindicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Aflashinglightishigherprioritythannon-flashing.

Audible SignalsAnaudiblesignal(beep)indicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Morefrequentbeepsindicatehigherpriorityconditions.

Input / Output Connections

Particle FiltersThefiltersmustbeinplaceattheintakeendoftheconcentratorduringoperationtokeepinputairclean.

Cannula Nozzle FittingThenasalcannulaconnectstothisnozzleforInogenOne®G3outputofoxygenatedair.

DC Power InConnectionforexternalpowerfromtheACpowersupplyorDCpowercable.

USB PortUsedforservicingonly.

English

www.ActiveForever.com

6

Power Supply Options

Single and Double rechargeable Lithium Ion BatteriesThebatterywillpowertheInogenOne®G3withoutconnectiontoanexternalpowersource.Whenfullycharged,asinglebatterywillprovideupto4hoursofoperation;adoublebatterywillprovideupto8hoursofoperation.ThebatteryrechargeswhenproperlyinstalledintheInogenOne®G3andtheconcentratorisconnectedtoACorDCpower.Rechargingtimeisupto4hoursforasinglebatteryand8hoursforadoublebattery.Seeinformationinthe“BatteryCareandMaintenance”section.

Power SupplyOverviewTheInogenOne®G3ACpowersupply(BA-301)isusedtopowertheInogenOne®G3concentratorfromanACpowersource.

DescriptionTheInogenOne®G3ACpowersupplyisspecificallydesignedforusewiththeInogenOne®G3OxygenConcentrator(IO-300).TheACpowersupplyprovidestheprecisecurrentandvoltagerequiredtosafelypowertheInogenOne®G3andisdesignedtooperatefromspecifiedACpowersources.WhenusedwithACpowersources,thepowersupplyautomaticallyadaptstoinputvoltagesfrom100Vto240V(50-60HZ)permittingusewithmostpowersourcesthroughouttheworld.

TheACpowersupplywillchargetheInogenOne®G3BatterieswhenusedwithACinputpower.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G3Batterywhenusedonanaircraft.TheACpowersupplyisusedwiththefollowingcomponents:

•PowersupplywithattachedpoweroutputcabletoconnecttotheInogenOne®G3•ACpowerinputcable

TheDCpowercable(BA-306)isspecificallydesignedforusewiththeInogenOne®G3OxygenConcentrator(IO-300).TheDCpowerinputcableconnectsdirectlytotheautomobilecigarettelighterorauxiliaryDCpowersupply.RefertoChapter3foradditionaloperatinginstructionsforusewithDCpowersource.

www.ActiveForever.com

7DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

English

*Actualproductappearancemayvary.

AC Power Plug*

Power Output Plug (Plug into Concentrator)

LED Indicator Light*AC Input

InogenOneACPowerSupplyModel#BA-301

InogenOneG3DCPowerCableModel#BA-306

DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft

Power Output Plug (Plug into Concentrator)

wArNING Donotusepowersuppliesorpowercablesotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedpowersuppliesorpowercablesmaycreateasafetyhazardand/orimpairequipmentperformance.Donotwrapcordsaroundpowersupplyforstorage.Donotdrive,dragorplaceobjectsovercord.Keepchildrenandpetsawayfromcord.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.

! C

www.ActiveForever.com

8

Inogen One® G3 Accessories

wArNING Donotusepowersupplies/adaptersoraccessoriesotherthanthosespecified inthisusermanual.Theuseofnon-specifiedaccessoriesmaycreateasafety hazardand/orimpairequipmentperformance.

Nasal CannulaAnasalcannulamustbeusedwiththeInogenOne®G3toprovideoxygenfromtheconcentrator.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.

wArNING Toavoiddangerofchokingandstrangulationhazard,keeptubingawayfromchildrenandpets.

NOTE Increasingthecannulalengthmayreducetheperceivednoiseduringoxygenbolusdelivery.

NOTE Whenusingacannula25ft.inlengthwiththeInogenOne®G3,anincreaseinflowsettingmayberequired.

Carry Bag (CA-300)TheCarryBagprovidesaprotectivecoverwithahandleandadjustableshoulderstraptoenableyoutocarrytheInogenOne®G3.TheInogenOne®G3canbeoperatedusingbatterypowerduringtransportwiththeCarryBag.ThecarrybagisdesignedsothattheInogenOne®G3Concentratorisloadedintothecarrybagfromthebottomofthebag.Theexpandablezipperedbottomisdesignedwithanadjustablezipperforusewithasingleordoublebatteryandeasyaccessforchangingbattery.Theexpandablesectionisnotintendedforuseasastoragecompartment.

! C

! C

www.ActiveForever.com

9

Inogen One® G3 Optional Accessory

External Battery Charger (BA-303)1.PlugtheExternalBatteryChargerACpowersupplycord

intoanelectricaloutlet.2.PlugtheExternalBatteryChargerACpowersupplyinto

thebatterycharger.3.SlideyourchargerontotheInogenOneG3Battery byclickingandlockingintothecharger.4.Whenthebatteryisinthecorrectposition,asolid redlightwillindicatethatthebatteryischarging.5.Whenthegreenlightilluminates,thebatteryis fullycharged.

CAUTION AvoidtouchingtherecessedelectricalcontactsoftheExternalBatteryCharger;damagetocontactsmayaffectchargeroperation.

NOTE Thesecontactsarenotpoweredunlessabatteryisinplaceandcharging.

NOTE TocompletelyremovepowerfromtheExternalBatteryCharger, removetheplug.

Inogen One G3 BackpackAlternative/optionalwayofcarryingyourInogenOne®G3,handsfree,morecomfort,outofyourwaywithextrapocketsforadditionalaccessories.

ToorderpleasecallInogenClientServices.

DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

English

www.ActiveForever.com

10

www.ActiveForever.com

11OperatingInstructionsChapter3

English

Operating Instructions3

! C

ExhaustGeneral Instructions

1. Place the Inogen One® G3 in a well ventilated location.Airintakeandexhaustmusthaveclearaccess.LocatetheInogenOne®G3insuchawaythatanyauditoryalarmsmaybeheard.

wArNING AvoiduseoftheInogenOne®G3inpresenceofpollutants,smokeorfumes.DonotusetheInogenOne®G3inpresenceofflammableanesthetics,cleaningagentsorotherchemicalvapors.

CAUTION Donotobstructairintakeorexhaustwhenoperatingthedevice.

Blockageofaircirculationorproximitytoaheatsourcemayleadtointernalheatbuildupandshutdownordamagetotheconcentrator.

2. Ensure both particle filters are in place.

CAUTION DonotoperatetheInogenOne®G3withoutbothparticlefiltersinplace.Particlesdrawnintothesystemmaydamagetheequipment.

3. Install the battery. InserttheInogenOne®G3batterybyslidingbatteryintoplaceuntilthelatchreturnstotheupperposition.

Intake

www.ActiveForever.com

12

CAUTION TheInogenOne®G3batteryactsasasecondarypowersupplyintheevent ofaplannedorunexpectedlossoftheACorDCexternalpowersupply.When

operatingtheInogenOne®G3fromanACorDCexternalpowersupply,aproperlyinsertedInogenOne®G3batteryshouldbemaintainedintheunit.Thisprocedurewillensureuninterruptedoperationandwilloperateallalarmsandalertsintheeventofalossoftheexternalpowersupply.

4. Connect the Power Supply. ConnecttheACinputplugtothepowersupply.ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplugtotheInogenOne®G3.ThegreenLEDonthepowersupplywillbeilluminatedandabeepwillsoundfromtheconcentrator.

CAUTION Ensurethepowersupplyisinawellventilatedlocationasitrelieson aircirculationforheatdissipation.Thepowersupplymaybecomehotduring operation.Makesurethepowersupplycoolsdownbeforehandling.

CAUTION Thepowersupplyisnotwaterresistant.

CAUTION Donotdisassemblethepowersupply.Thismayleadtocomponentfailure and/orsafetyrisk.

CAUTION Donotplaceanythinginthepowersupplyportotherthanthesuppliedwallcord.AvoidtheuseofelectricalextensioncordswiththeInogenOne®G3.

Ifanextensioncordmustbeused,useanextensioncordthathasanUnderwritersLaboratory(UL)Markandaminimumwirethicknessof

18gauge.Donotconnectanyotherdevicestothesameextensioncord.

NOTE Undercertainconditions(seeTechnicalSpecifications)thepower supplymayshutdown.ThegreenLEDwillblinkorwillnolongerbe

illuminated.Ifthisoccurs,disconnectthepowersupplyforatleast10secondsandreconnect.

www.ActiveForever.com

13OperatingInstructionsChapter3

NOTE WhenthepowersupplyisdisconnectedfromtheACoutlet,alsodisconnectitfromtheconcentratortoavoidunnecessarybatterydischarge.

5. Connect the nasal cannula tubing to the nozzle fitting.NozzlefittingislocatedonthetopoftheInogenOne®G3.Useofasinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.Additionaltitrationmaybeneededtoensureproperoxygendeliverywhenusingaparticularcannula.

CAUTION Toensureoxygenflow,ensurethatthenasalcannulaisproperlyconnectedtothenozzlefittingandthatthetubingisnotkinkedorpinchedinanyway.

CAUTION Replacethenasalcannulaonaregularbasis.Checkwithyourequipmentproviderorphysiciantodeterminehowoftenthecannulashouldbereplaced.

6. Turn on your Inogen One® G3 by pressing the ON/OFF Button.AsingleshortbeepwillsoundaftertheInogenlogoisdisplayed.“PleaseWait”willappearwhiletheconcentratorstartsup.Thedisplaywillindicatetheselectedflowsettingandpowercondition.Followingabriefstart-upsequence,awarmupperiodupto2minuteswillinitiate.Duringthistimeperiodtheoxygenconcentrationisbuildingtobutmaynothavereachedspecification.AdditionalwarmuptimemaybeneededifyourInogenOne®G3hasbeenstoredinextremelycoldtemperatures.

7. Set the Inogen One® G3 Concentrator to the flow rate prescribed by your physician or clinician. Usethe+or–settingbuttonstoadjusttheInogenOne®G3tothedesiredsetting.Thecurrentsettingcanbeviewedonthedisplay.

English

www.ActiveForever.com

14

! C

8. Position the nasal cannula on your face and breathe through your nose. TheInogenOne®G3willsensetheonsetofinhalationanddeliveraburstofoxygenataprecisetimewhenyouinhale.TheInogenOne®G3willsenseeachbreathandcontinuetodeliveroxygeninthismanner.Asyourbreathingratechanges,theInogenOne®G3willsensethesechangesanddeliveroxygenonlyasyouneedit.Attimes,ifyouinhaleveryquicklybetweenbreathstheInogenOne®G3mayignoreoneofthebreaths,givingtheappearanceofamissedbreath.ThismaybenormalastheInogenOne®G3sensesandmonitorsthechangesinyourbreathingpattern.TheInogenOne®G3willnormallysensethenextbreathanddeliveroxygenaccordingly.

Agreenlightwillflasheachtimeabreathisdetected.Makecertainthatthenasalcannulaisproperlyalignedonyourfaceandyouarebreathingthroughyournose.

wArNING Ifyoubegintofeelillorareexperiencingdiscomfortwhileusingthisdevice,consultyourphysicianimmediately.

CAUTION TheInogenOne®G3isdesignedtoprovideaflowofhighpurityoxygen.

Anadvisoryalarm,“OxygenLow”,willinformyouifoxygenconcentrationdrops.Ifalarmpersists,contactyourequipmentprovider.

GeneralToremovepower,unplugtheinputcordfromitssource(i.e.,ACwalloutlet,DCcarcigarettelighteradapter).

CAUTION Ensurethepowersupplyispoweredfromonlyonepowersource(ACorDC)atanygiventime.

www.ActiveForever.com

15OperatingInstructionsChapter3

Additional Operating Instructions

For Use In Home with AC PowerForoperationofthepowersupplyusinganACpowersource,followtheseinstructions:

1. ConnecttheACinputtothepowersupply.2. ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplug

totheInogenOne®G3.ThegreenLEDwillbeilluminated,indicatingthe powersupplyhasinputpower.

Traveling with your Inogen One® G3 System

TheInogenOne®G3Systemmakestravelbyboat,carortrainmoreconvenientforoxygenusersthaneverbefore.Nowyougetthesamequalityperformanceandconveniencewhileonthegothatyou’reusedtoreceivingfromyourInogenOne®G3athome.HerearesomeusefulandimportantinstructionsformaximizingperformanceandconveniencewhenusingyourInogenOne®G3whiletraveling.

Youshouldbeginplanningforatripwithachecklistofitemstoremember.Thischecklistshouldinclude:

✓ ACpowersupplyandDCpowercable✓ ExtraBattery(s)ifrequired✓ Importantphonenumbers,suchasthoseofyourphysicianandhomehealthcare provider,orprovidersintheareayouwillbetraveling✓ Planforbackupoxygenintheeventofaprolongedpoweroutageor mechanicalfailure

For Use In Automobile/rV/Boat/AircraftForoperationusingaDCpowersource,followtheseinstructions:

1.ConnecttheDCpoweroutputplugtotheInogenOne®G3.2.ConnecttheDCpowerplug(CigaretteLighterAdapter)to

thepowersource.3. Theplugshouldinsertintothesocketwithoutexcessive

forceandsecurelyremaininplace.

English

www.ActiveForever.com

16

4. Checkthedevicedisplayscreentoconfirmthatanexternalpowersupplyisconnected. AniconofabatterywithalightningboltoranACpowerplugwillbedisplayedonthe

screen,indicatingthatexternalpowerisconnectedandfunctioningproperly.

WARNING EnsurethattheautomobilepowersocketisadequatelyfusedfortheInogenOne®G3powerrequirements(minimum15Amp).Ifthepowersocketcannotsupporta15Ampload,thefusemaybloworthesocketmaybedamaged.

WARNING ThetipoftheCigaretteAdapterPlugbecomesHOTwheninuse.Donottouchthetipimmediatelyafterremovalfromanautocigarettelightersocket.

CAUTION Ensuretheautomobilepowersocketiscleanofcigaretteashandtheadapterplugfitsproperly,otherwiseoverheatingmayoccur.

CAUTION Donotusethepowersupplywithacigaretteplugsplitterorwithan extensioncable.ThismaycauseoverheatingoftheDCpowerinputcable.

CAUTION DonotjumpstarttheautomobilewiththeDCpowercableconnected.Thismayleadtovoltagespikeswhichcouldshutdownand/ordamagetheDCpowerinputcable.

CAUTION WhenpoweringtheInogenOne®G3inanautomobileensurethevehicle’sengineisrunningfirst,beforeconnectingDCcableintocigarettelighteradapter.Operatingthedevicewithouttheenginerunningmaydrainthevehicle’sbattery.

CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.

Traveling By AirTheFAAallowstheInogenOne®G3onboardallU.S.aircraft,hereareafewpointstomakeairtraveleasy.

Planning Your FlightWhenflyingwiththeInogenOne®G3,youmustinformtheairlinethatyouwillbeusingyourInogenOne®G3onboardtheaircraft.Youmustalsokeepasignedstatementfromyourphysicianwithyouthatincludes:

! C

! C

www.ActiveForever.com

17OperatingInstructionsChapter3

English

•Yourabilitytosee/hearalarmsandappropriatelyrespond.

•Whenoxygenuseisnecessary(alloraportionofthetrip).

•Maximumflowratecorrespondingtothepressureinthecabinundernormaloperatingconditions.

•Anewstatementwillnotbenecessaryeachtimeyoufly,butthestatementshouldbeavailableduringeveryflight.

•Someairlinesmayequiptheiraircraftwithonboardelectricalpower.YoumayhaveanopportunitytorequestaseatwithapowerportwhichcanbeusedtopoweryourInogenOne®G3.However,availabilityvariesbyairline,typeofaircraftandclassofservice.Youshouldcheckwithyourairlinesforavailabilityandalwaysplanonhavingsufficientbatterypowerforthedurationofyourflight,plusaconservativeestimateofunanticipateddelays.

•YourDCPowerCableisequippedwithacommonlyusedcigarettelighteradapter.However,aircraftsusedifferentpowerportconfigurationsanditisdifficulttodeterminewhichtypeofpowerportcompatibilityyouraircraftmaysupply;it’sagoodideatopurchaseanadaptersuchasMagellan’sEmPowerSocketConnectorItem#EA270.

Before Your FlightHerearesomethingstokeepinmindthedayyourflightdeparts:

•EnsureyourInogenOne®G3isclean,ingoodconditionandfreefromdamageorothersignsofexcessivewearorabuse.

•BringenoughchargedbatterieswithyoutopoweryourInogenOne®G3forthedurationoftheflightplusaconservativeestimateofunanticipateddelays.

•Regional/Commuterairlinesdonotofferonboardelectricalpower.Ifyourtravelplanscallforflightsonregionalairlines,youwillneedsufficientbatterypowerforthedurationoftheflight,plusaconservativeestimateofunanticipateddelays.

CAUTION Airlinesmaynotbeequippedtoprovidebackupoxygen.

•Arriveattheairportearly.AirportsecurityscreeningpersonnelmayrequireextratimetoinspectyourInogenOne®G3.

www.ActiveForever.com

18

•Whilewaitingtoboardyourflight,youmaybeabletoconservebatterypowerbyusingtheACPowerSupplytopoweryourInogenOne®G3fromanelectricaloutletintheairporttermi-nalifavailable.

•YoushouldinformtheairlineyouwillbeusingyourInogenOne®G3.Haveyourphysicianletterwithyouandreadyforinspectionifrequested.

During Your Flight1. Ifusingairlinepowerport,removetheBatteryfromtheInogenOne®G3Oxygen

Concentrator.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G3Batterywhenonboardaircraft.

2. ConnecttheDCPowerPlugforavailableairlinepower.Checkwithairlinepersonneltoensurecompatibility.

•Duringtaxi,takeoffandlanding,stowyourInogenOne®G3undertheseatinfrontofyou.YourInogenOne®G3willfituprightundermostairlineseats.However,ifitdoesn’tfityoumayturnitonitssidewithventsfacingup.

•ItisnotnecessarytoturnoffyourInogenOne®G3duringtaxi,takeoffandlandingifyourphysician’swrittenstatementrequiresyoureceiveoxygenduringtheseperiods.

CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.TheInogenOne®G3hasbeenverifiedtoprovideoxygentospecificationupto10,000ft.(3048m).Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.

After Your Flight•Remembertorechargeadditionalbatteriesyoumayhaveusedpriortoyournextflight.

Traveling By Bus, Train or BoatContactyourcarriertofindoutaboutpowerportavailability.

Battery Operating InstructionsEnsurethatthebatteryisinplaceandcharged.DisconnecttheInogenOne®G3fromitspowersource.WhiletheInogenOne®G3isoperatingonbatterypower,thebatterywilldischarge.Thedisplaywillindicatetheestimatedremainingpercentage(%)orminutesofuse.

www.ActiveForever.com

19

! C

English

OperatingInstructionsChapter3

Whentheconcentratordetectsthatthebatterylifeislow,withlessthan10minutesremaining,alowpriorityalertwillsound.Whenthebatteryisempty,thealertwillchangetoahighpriority.

Whenbatterylifeislow,dooneofthefollowing:• PlugtheInogenOne®G3intoanACorDCpowersourceusingtheACpowersupplyor DCcable.•ReplacethebatterywithachargedbatteryafterturningofftheInogenOne®G3 (bypressingtheON/OFFbutton).Toremovebatterypressandholdthebatterylatch buttonandslidebatteryofftheconcentrator.• Ifthebatteryisdrained,chargethebatteryorremoveitfromtheconcentrator.

IftheInogenOne®G3isbeingpoweredbytheACpowersupplyorDCpower,batterieswillchargeduringoperation.LeavingyourInogenOne®G3pluggedinpastthefullchargetimewillnotharmtheconcentratororthebattery. wArNING Itistheresponsibilityofthepatienttoperiodicallycheckthebatteryand replaceasnecessary.Inogenassumesnoliabilityforpersonschoosingnotto adheretomanufacturersrecommendations.

Normal Battery ChargingToensurethatyourbatteryisproperlycharging,inspectthatthecorrectACandDCpoweroutputplugadapterisbeingusedandthattheadapterisproperlyinsertedintothepoweroutlet.Observethedisplayorlightsthatindicatechargingstatus.

NOTE Whenstartingtochargeafullydischargedbattery,thechargingprocessmay startandstopduringthefirstfewminutes.

Battery Care and MaintenanceYourInogenOne®G3LithiumIonBatteryrequiresspecialcaretoensureproperperformanceandlonglife.UseonlyInogenOne®G3BatterieswithyourInogenOne®G3Concentrator.

Keep DryAlwayskeepliquidsawayfrombatteries.Ifbatteriesbecomewet,discontinueuseimmediatelyanddisposeofbatteryproperly.

www.ActiveForever.com

20

Effect of Temperature on Battery PerformanceTheInogenOne®G3singlebatterypowerstheInogenOne®G3Concentratorupto4hoursundermostenvironmentalconditions.Toextendtherun-timeofyourbattery,avoidrunningintemperatureslessthan41˚F(5˚C)orhigherthan95˚F(35˚C)forextendedperiodsoftime.

Battery Time remaining ClockTheInogenOne®G3continuouslydisplaysbatterytimeremaining.Thisdisplayedtimeisonlyanestimateandtheactualtimeremainingmayvaryfromthisvalue.

Please Follow These Important Guidelines to Maximize Battery Performance and Life:• Storebatteryinacool,dryplace.Storewithachargeof40-50%.• Ifusingmultiplebatteries,makesurethateachbatteryislabeled(1,2,3orA,B,C,etc.) androtateonaregularbasis.Batteriesshouldnotbeleftdormantformorethan 90daysatatime.

www.ActiveForever.com

21

Mode

PowerText

Display Icons

TheInogenOne®G3displayisdividedintothreeareas.Theupperleftcornerofthedisplayshowsthebreathdetectionalertstatus.Thelowerleftcornerindicatespowersourceandbatterychargelevel.Therightsideofthedisplaycontainstextinformation,suchasflowsetting,batterytimeremaininganderrornotifications.

Power Status Icons

Theseiconsareexamplesofthoseshowninthedisplay’spowerstatuswindowwhentheInogenOne®G3isoperatingonbatterypower.

Inogen One® G3 Oxygen Concentrator Audible and Visible Signals

AudibleandVisibleSignalsChapter4

! C

! C

! C

! C

Meaning

Batteryisempty.

Batteryhaslessthan10%chargeremaining.Thisiconflashes.

Batteryhasapproximately40%to50%chargeremaining.

Batteryisfull.

Icon

4English

www.ActiveForever.com

22

Power Status Icons (continued)TheiconsbelowareexamplesofthoseshownwhentheInogenOne®G3isoperatingfromanexternalpowersupplyandchargingthebattery.Thelightningboltindicatesthatanexternalpowersupplyisconnected.

! C

! C

! CMeaning

Batteryischargingwithchargelevelbetween60%and70%.

Thebatteryisfullychargedandischargingasnecessarytomaintainitscharge.

Batteryischargingwithchargelevellessthan10%.

TheInogenOne®G3isoperatingfromanexternalpowersourcewithnobatterypresent.

Icon

Mode IconsThesearetheiconsshowninthedisplay’smodewindow.

Display Text

NOTE Whentwoconditionsoccursimultaneously,theconditionwiththehigherprioritywillbedisplayed.

Meaning

Thebreathdetectionaudiblealarmhasbeenenabled.

Thebreathdetectionaudiblealarmisdisabled.Thisisthedefaultcondition.

Icon

! C

www.ActiveForever.com

23AudibleandVisibleSignalsChapter4

English

Informational MessagesThefollowinginformationdisplaysarenotaccompaniedbyanyaudiblefeedbackoranyvisualchangeintheindicatorlights.

Condition/Action/Explanation

TheInogenlogoisdisplayedatstartup.

Defaultdisplaywhenoperatingonbatterypower.“X”representstheselectedflowsetting(e.g.,Setting2).“HH:MM”representstheapproximatetimeremainingonthebatterycharge(e.g.,1:45).

Defaultdisplaywhenoperatingonanexternalpowersupplyandthebatteryischarging.“xx%”representsthepercentbatterycharge(e.g.,86%).

Defaultdisplaywhenthebatteryisnotchargingorwhenthetimeremainingisnotavailablefromthebattery.

Displayedwhentheconcentratorispluggedinandbeingusedtochargeabattery(notbeingusedforoxygenproduction).Itisnormaltoseeafullychargedbatteryreadbetween95%and100%whenexternalpowerisremoved.Thisfeaturemaximizestheusefullifeofthebattery.

Message Display & Text

SettingXBatteryHH:MM

SettingXChargingxx(or)BatteryFull

SettingXBatteryxx%

Chargingxx%(0r) BatteryFull

Notifications

! C

wArNING Audiblenotifications,rangingfrom55dbato65dbadependingontheusersposition,aretowarntheuserofproblems.Toinsurethataudiblenotificationsmaybeheard,themaximumdistancethattheusercanmoveawayfromitmustbedeterminedtosuitthesurroundingnoiselevel.

www.ActiveForever.com

24

Low Priority AlertsThefollowinglowpriorityalertmessagesareaccompaniedbyadouble beepandasolid red light.

Condition/Action/Explanation

Batterypowerislow,withlessthan10minutesremaining.Attachexternalpowersupplyorpowerdownandinsertafullychargedbattery.

Columnmaintenanceisrequiredwithin30days.Pressandholdthelightbuttonfor3secondstoclearmessagefromdisplay.Themessagewillbesilencedfor72hours.Contactyourequipmentprovidertoarrangeforservice.

Message Display & Text

BatteryLow AttachPlug

O2Service Soon

Notifications (continued)TheInogenOne®G3monitorsvariousparametersduringoperationandutilizesanintelligentalarmsystemtoindicateamalfunctionoftheconcentrator.Mathematicalalgorithmsandtimedelaysareusedtoreducetheprobabilityoffalsealarmswhilestillensuringpropernotificationofanalarmcondition.Ifmultiplealarmconditionsaredetected,thehighestpriorityalarmwillbedisplayed.Thefollowingnotificationmessagesareaccompaniedbyasingle, short beep.

Condition/Action/Explanation

On/Offbuttonhasbeenpressedfortwoseconds.Concentratorisperformingsystemshutdown.

AudibleAlarmbuttonhasbeenpressedforfiveseconds.

Message Display & Text

InogenOne® ShuttingDown

HH:MM Vx.x:SerialNumber

www.ActiveForever.com

25AudibleandVisibleSignalsChapter4

English

Low Priority Alerts (continued)Condition/Action/Explanation

Batteryerrorhasoccurred.Checktheconnectionofyourbatteryandensurethatitisproperlyattachedandlatchedonconcentrator.Ifbatteryerrorrecurswithsamebattery,stopusingthebatteryandswitchtoanewbatteryorremovebatteryandoperateconcentratorusingexternalpowersupply.

Concentratorisproducingoxygenataslightlylowlevel(<82%)foraperiodof10minutes.Ifconditionpersists,contactyourequipmentprovider.

Batteryhasexceededitschargingtemperatureandcharginghasstopped.Thebatterywillnotchargewhilethisalertispresentbutwillbegintochargewhenthebatterytemperaturereturnstothenormaloperatingrange.Ifbatterychargingisdesiredsooner,removethebatteryfromtheconcentratorandallowittocoolinanopenareaforapproximately10-15minutes.Then,re-insertthebatteryintotheInogenOne®G3.Iftheproblemstillpersists,contactyourequipmentprovider.

Concentratorisproducingoxygenbutcannotreportbatterystatus.Replacebattery.Ifconditionpersists,contactyourequipmentprovider.

Theconcentratorrequiresservicingattheearliestconvenience.Theconcentratorisoperatingtospecificationandmaycontinuetobeused.Contactyourequipmentprovidertoarrangeforservice.

Theconcentrator’soxygensensorhasmalfunctioned.Youmaycontinuetousetheconcentrator.Iftheconditionpersists,contactyourequipmentprovider.

Message Display & Text

CheckBattery

OxygenLow SeeManual

RemoveBatterytoCool

CommError SeeManual

ServiceNeeded

O2SensorFailSeeManual

www.ActiveForever.com

26

Condition/Action/Explanation

Concentratorhasnotdetectedabreathfor60seconds.Checkthatcannulaisconnectedtoconcentrator,therearenokinksintubingandcannulaispositionedproperlyinyournose.

Oxygenoutputconcentrationhasbeenbelow50%for10minutes.Ifconditionpersists,switchtoyourbackupoxygensourceandcontactyourequipmentprovidertoarrangeforservice.

Abreathhasbeenrecognized,butproperoxygendeliveryhasnotbeendetected.

Concentratorisexperiencinganerrorbutiscapableofcontinuedoperation.Removeandre-insertbatteryand/orcheckexternalpowersupplyconnections.Ifconditionpersists,contactyourequipmentprovider.

Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Ifpossible,moveconcentratortoacoolerlocationorpowerunitwithanexternalpowersupplyandremovebattery.Ifconditionpersists,contactyourequipmentprovider.

Message Display & Text

NoBreathDetect CheckCannula

OxygenErrorServiceNeeded

O2DeliveryError

SystemError SeeManual

BatteryHOT Warning

Medium Priority AlertsThefollowingmediumpriorityalertmessagesareaccompaniedbyatriple beep,repeatedevery25seconds,andaflashing red light.

www.ActiveForever.com

27

Condition/Action/Explanation

Concentratorhasinsufficientbatterypowertoproduceoxygen.Attachexternalpowersupplyorexchangebattery,thenrestartunitifnecessarybypressingOn/Offbutton.

Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Concentratorhasstoppedproducingoxygen.Ifpossible,moveconcentratortoacoolerlocation,thenturnpoweroffandbackon.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Concentratortemperatureistoohighandoxygenproductionisshuttingdown.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Message Display & Text

BatteryEmpty AttachPlug

BatteryHOT ShutDown

SystemHOT ShutDown

High Priority Alerts

CAUTION IfyouarenotneartheInogenOne®G3youmaynotbeabletohearorseethehighpriorityalerts.MakesuretheInogenOne®G3isinalocationwherethealertsandalarmswillberecognizediftheyoccur.

Thefollowinghighpriorityalertmessagesareaccompaniedbyafive beep pattern,repeatedevery10secondsandaflashing red light.

AudibleandVisibleSignalsChapter4

English

www.ActiveForever.com

28

Condition/Action/Explanation

Thismayresultfromtheconcentratorbeingstoredinacoldenvironment(below0ºC(32ºF)).Movetoawarmerenvironmenttoallowtheunittowarmupbeforestartingit.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Concentratorhasstoppedproducingoxygenandisshuttingdown.Youshould:

1. Noteerrornumber2.Switchtobackupoxygensource3.Contactyourequipmentprovider

Message Display & Text

SystemCOLD ShutDown

Error(###) ServiceNeeded

High Priority Alerts (continued)

www.ActiveForever.com

29

Troubleshooting

TroubleshootingChapter5

Useexternalpowersupplyorreplacebatterywithonethatisfullycharged

Checkpowersupplyconnectionandverifygreenlightissolid

CheckDCCableconnectionattheConcentratorandatcigarettelighterorauxiliaryDCpowersource

Contactyourequipmentprovider

Batteryisdischargedornobatteryispresent

ACPowersupplyisnotconnectedproperly

DCCableisnotconnectedproperly

Malfunction

ConcentratordoesnotpoweronwhenOn/Offbuttonispressed

recommended Solution

RefertoChapter4

Possible Cause

RefertoChapter4

Problem

Anyproblemaccompaniedbyinformationonconcentratordisplay,indicatorlightsand/oraudiblesignals

Solutionstosomepossibleissuesyoumayencounteraredescribedinthissection.

Inogen One® G3 Oxygen Concentrator

5English

www.ActiveForever.com

30

PressOn/Offbuttontopowerconcentrator

Checkcannulaanditsconnectiontoconcentratornozzle

Concentratorisnotpoweredon

Cannulaisnotconnectedproperlyoriskinkedorobstructed

Nooxygen

recommended SolutionPossible CauseProblem

Troubleshooting (continued)

www.ActiveForever.com

31

Cannula replacement

Yournasalcannulashouldbereplacedonaregularbasis.Consultwithyourphysicianand/orequipmentproviderand/orcannulamanufacturer’sinstructionsforreplacementinformation.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.

CAUTION Useofacannulaotherthanahighflowcannula(e.g.Salter1600Q) mayrestrictoxygendeliveryand/orattachmenttonozzlefittings.

Case CleaningYoumaycleantheoutsidecaseusingaclothdampenedwithamildliquiddetergent(suchasDawnTM)andwater.

wArNING DonotsubmersetheInogenOne®G3oritsaccessoriesinwaterorallowwatertoenterintothecase;thismayleadtoelectricalshockand/ordamage.

wArNING DonotusecleaningagentsotherthanthosespecifiedinthisUserManual.Donotusealcohol,isopropylalcohol,ethylenechlorideorpetroleumbasedcleanersonthecasesorontheparticlefilters.

Cleaning, Care and Maintenance

Cleaning,CareandMaintenanceChapter6

6

! C

! C

English

www.ActiveForever.com

32

Filter Cleaning and replacementBothparticlefiltersmustbecleanedweeklytoensuretheeaseofairflow.Removefiltersfromthefrontofthedevice.Cleantheparticlefilterswithamildliquiddetergent(suchasDawnTM)andwater;rinseinwateranddrybeforereuse.

NOTE Itmaybenecessarytocleantheparticlefiltersmoreoftenindustyenvironments.

TopurchaseadditionalparticlefilterscontactyourequipmentproviderorInogen.

Output FilterTheoutputfilterisintendedtoprotecttheuserfrominhalationofsmallparticlesintheproductgasflow.TheInogenOne®G3includesanoutputfilterconvenientlylocatedbehindtheremovablecannulanozzlefitting.Inogenrequiresthatthisfilterbereplacedbetweenpatients.

TheoutputfiltermaybereplacedbytheequipmentproviderorbytheownerusingtheOutputFilterReplacementKit(RP-107).

TheInogenOneG3Concentratormustbecleanedanddisinfectedaspertheaboveinstructionsforeachnewpatient.Nospecialmaintenanceneedstobecarriedoutbythepatient.YourequipmentproviderperformsmaintenanceoperationstoassurecontinuedreliableservicefromyourInogenOneG3.Themanufacturer’sinstructionsforthepreventativemaintenanceofthedevicesaredefinedintheservicemanual.Allworkmustbeperformedbytrainedtechnicianscertifiedbythemanufacturer.

DC Input Cable Fuse replacementTheCigaretteLighterDCpowerplugcontainsafuse.IftheDCinputcableisbeingusedwithaknowngoodpowersourceandtheunitisnotreceivingpower,thefusemayneedtobereplaced.

Toreplacethefuse,followtheseinstructionsandrefertothephotographbelow.1. Removethetipbyunscrewingtheretainer.Useatoolifnecessary.2. Removetheretainer,tipandfuse.3. ThespringshouldremaininsidetheCigaretteLighterAdapterhousing.Ifthespringis

removed,makesuretoreplacethespringfirstbeforeinsertingthereplacementfuse.

www.ActiveForever.com

33Cleaning,CareandMaintenanceChapter6

English

4. Installareplacementfuse,InogenRP#125(BUSSMDA-12)andreassemblethetip.Ensuretheretainerringisproperlyseatedandtightened.

Inogen One® G3 Column Change Procedure

NOTE Columnchangeinstructionsareonlytobeusedwhenmaintenanceis requiredandarenotintendedforpracticepurpose.

1. TurnofftheInogenOne®G3concentratorbypressingthepowerbuttontoshutdownthedevice.

2.RemovetheInogenOne®G3concentratorfromthecarrybag.

3.RemovethebatteryfromtheInogenOne®G3concentrator.

4.TurnInogenOne®G3oversothatisstandingintheupsidedownpositionwherethedisplaypanelis

facingdownandthebottomoftheInogenOne®G3isfacingup.

5.Therearetwocolumns(metaltubes)onthe InogenOne®G3concentratorandoneislocatedon eachsideofthedevice.

FuseCigarette Adapter Plug retainer Tip retainer

CAUTION Forcontinuedprotectionagainstriskoffire,onlyusefusespecified.

www.ActiveForever.com

34

Inogen One® G3 Column Change Procedure (continued) 6.Removecolumnbypressingthelatchbutton withthumborfinger.

7. Whileholdingthebuttonopen,slidethecolumn(metaltube)outbygrippingtheexposedaluminumsurfacewithotherhand.

8.Removethecolumn(metaltube)completelyfromtheInogenOne®G3.

9.Repeatsteps6-8toremovetheothercolumn(metaltube).

Column(metaltube)Installation10.Removedustcapsofnewcolumn(metaltube).

Makesurethereisnodustordebriswherethedustcapswerelocated.

Openandunlocked

www.ActiveForever.com

35Cleaning,CareandMaintenanceChapter6

English

Inogen One® G3 Column Change Procedure (continued)

11.Insertcolumn(metaltube)intotheInogenOne®G3concentrator.Donotleavethecolumnexposed;

itshouldbeinsertedintotheInogenOneG3as soonasthedustcapshavebeenremoved.

12.Pushthecolumn(metaltube)intothedevicesuchthatthecolumn(metaltube)isfullyseatedintotheInogenOne®G3concentrator.Thespringloadedlatchbuttonshouldfullyreturntotheclosedposition.

13.Repeatsteps10-12toinstalltheothercolumn(metaltube).

14.ConnecttheACpowersupplycordtotheInogen One®G3andplugthepowersupplyACcordintoan electricaloutlet.Do Not Power on the Inogen One® G3 concentrator.

15.Pressandholdthelightbuttonfor10secondsand

screenwilldisplay“sievereset”message.Releasebuttononcemessageisdisplayedonscreen.

16.Pressthebellbuttononceandscreenwilldisplay“sieveresetsuccessful”.

17.PressthepowerbuttontoturnontheInogenOne®G3,andusenormally.

wArNING Donotuseanycolumnsotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedcolumnsmaycreateasafetyhazardand/orimpairequipmentperformanceandwillvoidyourwarranty.! C

Closedandlocked

www.ActiveForever.com

36

Disposal of Equipment and AccessoriesFollowyourlocalgoverningordinancesfordisposalandrecyclingoftheInogenOne®G3andaccessories.IfWEEEregulationsapply,donotdisposeofinunsortedmunicipalwaste.WithinEurope,contacttheEUAuthorizedRepresentativefordisposalinstructions.Thebatterycontainslithiumioncellsandshouldberecycled.Thebatterymustnotbeincinerated.

Maintenance Items List• InogenOne®G3singlebattery(model#BA-300)• InogenOne®G3doublebattery(model#BA-316)• Replacementintakeparticlefilters(model#RP-300)• OutputFilterReplacementKit(model#RP-107)• InogenOne®G3columns(model#RP-320)

Forassistance,ifneeded,insettingup,using,maintaining,ortoreportunexpectedoperationorevents,contactyourequipmentprovider,ormanufacturer.

Other Service and Maintenance wArNING DonotdisassembletheInogenOne®G3oranyoftheaccessoriesor

attemptanymaintenanceotherthantasksdescribedinthisusermanual;disassemblycreatesahazardofelectricalshockandwillvoidyourwarranty.Donotremovethetamperevidentlabel.Foreventsotherthanthosedescribedinthismanual,contactyourequipmentproviderforservicingbyauthorizedpersonnel.

CAUTION DonotuselubricantsontheInogenOne®G3oritsaccessories.

! C

www.ActiveForever.com

37

Symbols Used On Concentrator and Accessories

SymbolsUsedonConcentratorandAccessoriesChapter7

7English

Meaning

Awarningindicatesthatthepersonalsafetyofthepatientmaybeinvolved.Disregardingawarningcouldresultinsignificantinjury.

Acautionindicatesthataprecautionorserviceproceduremustbefollowed.Disregardingacautioncouldleadtoaminorinjuryordamagetoequipment.

SeeUserManualforInstructions.

U.S.FederalRegulationRestrictsthisDevicetoSalebyOrderofPhysician.MayalsobeapplicableinotherCountries.

ACPower

DCPower

NoSmokingwhiledeviceisinuse.

NoOpenFlames(Concentrator);Donotincinerate(Battery).Refertoinstructionmanual/booklet.

Symbol

WARNING

CAUTION

ONLY

ONLY

ONLY

ONLY

ONLY

! C

www.ActiveForever.com

38

Meaning

KeepDry

IndoororDryLocationUseOnly,DoNotGetWet

UseNoOilorGrease

DoNotDisassemble(contactyourequipmentproviderforservicingbyauthorizedpersonnel)

DoNotDisposeofInUnsortedMunicipalWaste

TypeBFAppliedPart,NotIntendedforCardiacApplication

ClassIIDevice

ElectricalSafetyAgencyCertificationLogo

CompliesWithApplicableEUDirectivesIncludingMedicalDeviceDirective

Symbol

ONLYONLY

ONLY

ONLYONLY

0123

93-01213-00-00 rev1

0123

COLOR SUMMARY: 1. Black

User Interface Label

Meaning

ON/OFFButton

DisplayBacklightButton

IncreaseFlowSetting

DecreaseFlowSetting

AudibleAlarmButton

Symbol

www.ActiveForever.com

39InogenOne®G3SystemSpecificationsChapter8

English

Inogen One® G3 ConcentratorDimensions:Withsinglebattery

L/W/H:8.75in.(22.2cm.)/3.0in.(7.6cm.)/7.25in.(18.4cm.)L/W/H:8.75in.(22.2cm.)/3.0in.(7.6cm.)/8.25in.(21cm.)

Weight: 4.8pounds(2.2Kg.)(includessinglebattery)Noise: Lessthan42dBA(aspackaged)onsetting2Warm-UpTime: 2minutesOxygenConcentration: 90%-3%/+6%atallsettingsFlowControlSettings: 4settings:1to4Power:ACPowerSupply:

DCPowerCable:RechargeableBattery:

ACInput: 100to240VAC50to60Hz

Auto-Sensing: 1.0A

DCInput: 13.5-15.5VDC,10AMax.Voltage: 12.0to16.8VDC

BatteryDuration: Upto4hourswithsinglebatteryUpto8hourswithdoublebattery

BatteryChargingTime: Upto4hoursforasinglebatteryUpto8hoursforadoublebattery

EnvironmentalRangesIntendedforUse:

Temperature:41to104˚F(5to40˚C)Humidity:0%to95%,non-condensingAltitude:0to10,000ft(0to3048meters)

EnvironmentalRangesIntendedforStorage:

Temperature:-13to158˚F(-25to70˚C)Humidity:0%to95%,non-condensingStoreinadryenvironment

Transportation: KeepDry,HandleWithCare

Inogen One® G3 System Specifications8

www.ActiveForever.com

40

Inogen One® G3 Concentrator (continued)

Classifications

ModeofOperation: ContinuousDutyTypeofProtectionAgainstElectricalShock:

ClassII

DegreeofProtectiontoConcentratorComponentsAgainstElectricalShock:

TypeBFNotintendedforcardiacapplication

DegreeofProtectiontoConcentratorComponentsAgainstIngressofWaterWhileUsedOutsideofCarryBag:

IP20-Notprotectedfromdrippingwater.Protectedagainstingressofsolidobjects>12.5mm.

DegreeofProtectiontoConcentratorComponentsAgainstIngressofWaterWhileUsedInsideofCarryBag:

IP22-Verticallydrippingwatershallhavenoharmfuleffect&protectagainstingressofsolidobjects>12.5mm.diameterwhentheenclosureistiltedatanangleupto15˚fromitsnormalposition

DegreeofProtectiontoConcentratorExteriorProvidedbytheCarryBag:

IP02-Verticallydrippingwatershallhavenoharmfuleffectwhentheenclosureistiltedatanangleupto15˚fromitsnormalposition

DegreeofSafetyforApplicationinPresenceofAnestheticGases:

Notsuitableforsuchapplication

TestedbyIndependentLaboratory:

Safety:IEC60601-1CAN/CSAC22.2No.60601-1ElectromagneticCompatibility:IEC60601-1-2RTCADO160

www.ActiveForever.com

41InogenOne®G3SystemSpecificationsChapter8

English

ELECTrOMAGNETIC COMPATIBILITY

ThisCEMarkedequipmenthasbeentestedandfoundtocomplywiththeEMClimitsfortheMedicalDeviceDirective93/42/EEC[EN55011ClassBandEN60601-1-2].Theselimitsaredesignedtoprovidereasonableprotectionagainstharmfulinterferenceinatypicalmedicalinstallation.

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

TheInogenOne®G3OxygenConcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheInogenOne®G3OxygenConcentratorshouldassurethatitisusedinsuchanenvironment.

Emissions Test Compliance Electromagnetic Environment - Guidance

RFemissionsCISPR11

Group1 TheInogenOne®G3OxygenConcentratorusesRFenergyonlyforitsinternalfunction.ThereforeitsRFemissionsareverylowandnotlikelytocauseanyinterferenceinnearbyequipment.

RFemissionsCISPR11

ClassB TheInogenOne®G3OxygenConcentratorissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.

HarmonicEmissionsIEC61000-3-2

ClassA

Voltagefluctuations/flickeremissionsIEC61000-3-3

Complies

www.ActiveForever.com

42

recommended Separation Distances between Portable and Mobile rF Communications Equipment and This Device:

ThisdeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.ThecustomerortheuseroftheInogenOne®G3OxygenConcentratorcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andthisInogenOne®G3OxygenConcentratorasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.

150 kHz to 80 MHzd=1.2√P

0.12

0.38

1.2

3.8

12

0.12

0.38

1.2

3.8

12

0.23

0.73

2.3

7.3

23

0.01

0.1

1

10

100

80 MHz to 800 MHzd=1.2√P

800 MHz to 2.5 GHzd=2.3√P

Separation Distance According to Frequency of Transmitter (M)rated Maximum Power Output of Transmitter (w)

Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.

NOTE At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.

NOTE Theguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.