Edition V 1.6 EN Technical Changes reserved 09.09.2014

User Manualfor

SENTA PUR L liftPart No. 278700

Illustration showsSENTA PUR L liftPart No. 278700

Page 2: User Manual - Handicare Canada

Contents1. Preface 4

2. Introduction 5 2.1 Intended Use 5 2.2 Treatment Duration 5 2.3 Contraindications 5 2.4 Safety Risk and possible Device Damage 5

3. Safety instructions 6

4. Operation instructions 8 4.1 Before Use 8 4.2 Manual Emergency Lowering 8 4.3 SENTA PUR L lift with wings 9 4.4 SENTA PUR L lift 9 4.5 Function of the Safety Arm Rests 10 4.6 Battery Pack position 10 4.7 Detachable Lifter for Lying Patients Wing 11

5. SENTA Digital Scale SL 12 5.1 Safety Instructions Digital Scale SL 13 5.2 Cleaning the Scale 13 5.3 Start-up 13 5.4 Zeros and Tarring 13 5.5 Replacing batteries 13

6. Combo Bath Lifter with calibrated scale 14 6.1 Use of Calibration 14 6.2 Approval 14 6.3 Product Characteristics 14 6.4 Operating the Scale 15 6.5 “Stable Error” Error Message 16 6.6 LIVE WT 17 6.7 Error Message during Operation 18 6.8 Error Mode in Setup Mode 18 6.9 Diagnosis Error Messages 19

7. Battery device 24 Volt 21 7.1 Electricity Supply SENTA PUR L lift 21 7.2 Start-up and Function (Electric motor / Battery unit) 21 7.3 Hand control SENTA 23 7.4 External Charging Device / Wall-mounted Station CHJ2 24 7.5 Type Label 24

Prism Medical Ltd. V1-2014

2

Page 3: User Manual - Handicare Canada

8. Operating the SENTA PUR L lift 25 8.1 Patient Pickup 25 8.2 Manual Emergency Lowering 26 8.3 Contact Partner for Technical Information 26

9. Cleaning / Disinfection 27 9.1 Cleaning 27 9.2 Disinfection 27 9.3 Sterilisation 27

10. Disposal 28 10.1 Disposal of the Packing Material 28 10.2 Product Disposal 28

11. Safety Check 28

12. Troubleshooting and Customer Support 29 12.1 Troubleshooting 29 12.2 Customer Support 30

13. Appendix 31 13.1 Technical Data 31 13.2 Dimensions 33 13.3 Icons and Symbols used 34 13.4 Declaration of Conformity 35 13.5 Spare Parts and Accessories 35

Contents

Prism Medical Ltd. V1-2014

3

Page 4: User Manual - Handicare Canada

1. Preface

Dear Customer,Thank you for purchasing an Senta Pur L lift from BEKA Hospitec. Please consider the following points:1. Read this user manual carefully.2. Use this device for its intended purpose only.3. Operating this device is only admissible by trained personnel.4. Maintenance and safety checks are only permitted by BEKA Hospitec or by authorized

BEKA Hospitec distributors.5. Installations, upgrades, changes or repairs are only permitted by BEKA Hospitec or by

authorized BEKA Hospitec distributors.

In the case of repairs or modifications to BEKA Hospitec products that are not done by BEKA Hospitec or by authorized BEKA Hospitec distributors, all warranties for the repairs or modifications as well as for the device and device functionalities connected with these repairs or modifications, will expire immediately.

Liability and warrantyThe manufacturer of the lift shall only be responsible for the safety and reliability of the device if:a. the device is used for its intended purpose only, andb. functionality checks are carried out on a regular basis.

Prism Medical Ltd. V1-2014

4

Page 5: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

2. Introduction

2.1 Intended useThe Senta Pur L lift is intended to be used to lift and transport patients in hospitals and nursing homes as well as nursing facilities and its purpose is to support taking a bath. The picking up of patients can be done with the patient lying or standing. The SENTA PUR L lift is only intended to transport patients indoors on smooth and even surfaces. The expected operating life of the combo bath lift is approx. 8 years.

2.2 Duration of treatmentThe SENTA PUR L lift is intended to be used for a limited amount of time without getting in contact with injured skin.

Please NoteSide effects caused by this treatment are not known. Please take a look at the contraindications in section 2.3.

2.3 ContraindicationsPregnant women should not be treated with the Senta Pur L lift.

2.4 Safety Risk and possible Device DamagePregnant women should not be treated with the SENTA PUR L lift.

Patients with:• Osteoporosis• Spine damages• Ontogenesis imperfect• Mental aberrations• Epileptic seizuresmay be treated with the Senta Pur L lift after consulting the physician in charge.

5

WARNING:

Page 6: User Manual - Handicare Canada

3. Safety instructions

Read the following safety instructions carefully before using the Senta Pur L lift for the first time. All notes and warnings on the device and in this user manual must absolutely be observed.

The manufacturer BEKA Hospitec shall not be liable for any errors or damage caused by improper use.• Operating instructions - Read this manual before using the Senta Pur L lift for the first time. Please keep these operating

instructions near the device for later use.• Product shall only be operated by trained personnel.• After each treatment the product must be disinfected.• Before and during height adjustment, always make sure that your feet are not placed in the area

of the castors and the patient.• In case of unusual noises, damage or malfunctions the lift must not be operated any longer.• Do not stay between the lift and an obstacle during transport.• Avoid steep or slippery surfaces.• The lift must only be used for its intended purpose.• Make sure the power supply is always switched on during treatment.• Before each use check all functions are safe to use and the device and its accessories work as

intended (e.g. visual inspections, functionalities, etc.).• Make sure especially during adjustment that nobody reaches into the danger areas (hinges) –

crushing hazard.• Always make sure the patient / resident is correctly positioned on the lift and the safety strap is

fastened.• Protect the SENTA PUR L lift from direct exposure to sun and heat.• The device is not approved to be used in explosion-prone areas. • Treatment must only be carried out under supervision.• Never leave the patient / resident unattended on the lift.• Ensure there is no interference within the movement area of the arm / hand rest while setting up

the lift (risk of injury).• Make sure that hair, arms and legs are positioned tightly at the body and cannot get pinched.• Make sure that the legs and feet of the patient / resident as well as those of the care-giver are free

from obstacles (risk of injury).• Lock the castors of the nursing bed, the stretcher etc. to guarantee that the patient can be picked

up safely. Do not leave the brakes of the lift in a released state.• Try to keep transport time of the patient as short as possible and never leave the patient

unattended.• Lift the patient only as high as necessary.• Operating time as well as the maximum load capacity must not be exceeded.• Make sure picking up takes place at a safe place.

Prism Medical Ltd. V1-2014

6

Page 7: User Manual - Handicare Canada

• During pick up the SENTA PUR L lift must not be in operation.• Make sure the patient has a safe grip on the bow-shaped handle.

• Ventilation Never cover, place something over or alter slots and openings on the device.

• Cable routing Lay the power and the connection cords in such a way that they cannot be damaged. A damaged power cord can cause a fire or risk of electric shock and may not be used.

• Customer service Contact customer service if the SENTA PUR L is not working correctly and troubleshooting in section 10.1 did not fix the problem.

• Cleaning Do not use any aerosol cleaners. See detailed cleaning and disinfecting instructions in section 9.

• Repair Repairs are only permitted to be carried out by manufacturer authorized personnel. Contact your customer service representative to arrange a service. Opening the device or the accessories will void the warranty along with all liability claims.

• Duty of care Before use, make sure that the tub is in good condition and that all functions are safe to use. Please never insert any foreign objects into the device. This may create a safety hazard as well as a life-threatening electrical shock or damage to the device.

• Safety checks A “Safety Check” need to be carried out on the Senta Pur L lift every 12 months in conjunction with maintenance work. See section 11.

• Accessories Operate the SENTA PUR L lift with original accessories only.

BEKA Hospitec GmbH Am Rübenmorgen 3 D-35582 Wetzlar Phone: +49(0)641-9 22 20 - 0 Fax: +49(0)641-9 22 20 - 20 [email protected] www.beka-hospitec.de

Prism Medical Ltd.485 Millway Avenue, Unit 2 Concord Ont. L4K 3V4 T 1.800.265.0677 F 416.260.5580 www.prismmedical.ca

Prism Medical Ltd. V1-2014

7

Any unauthorized repair, rebuilding and/or changes to the SENTA PUR L lift is forbidden for safety reasons and will void all liability of the manufacturer for the damages resulting from it.For the damages which result from the use of spare parts not released by the manufacturer, will void all liability of the manufacturer.

WARNING:

Page 8: User Manual - Handicare Canada

4. Operation instructions

4.1 Before UseIn order to guarantee a failure-free operation the following inspections should be carried out once a day:• Visual inspection of the lift system (check for damage on the outside and wear).• Check for missing or loose screws at the handle system.• Check handles system for malfunctions.• Check manual operation of malfunctions (up/down).• Check emergency lowering.• Check charging function.• Check castors for un-smooth operation.

4.2 Manual Emergency LoweringTriggering the emergency lowering function:The SENTA PUR L lift is equipped with an emergency lowering function. The hole in the cover plate on the lifting pole allows, through using the 6 mm allen key that comes with the product, to lower the respective patient manually. Use the manual emergency lowering only in the case of an emergency. Before performing a manual emergency lowering unplug from power supply and remove the battery pack.

• Open the battery pack to access the allen key.• Put back the allen key after an emergency lowering has been carried out (key will snap in).

Prism Medical Ltd. V1-2014

8

Allen key for emergency lowering

Illustration 1

Hole

Illustration 2

The emergency lowering mechanism must not be treated with oil, grease or any other lubricant.

NOTE:

Page 9: User Manual - Handicare Canada

4.3 SENTA PUR L lift with wingsSENTA PUR L lift consists of three parts, to facilitate positioning at any side of the patient’s bed. When the lift is positioned at the patient’s bed, the two wings (left and right) are folded up or down to allow back and legs of the patient to rest comfortably (see illustration 3). The edges of the lying surface are shaped in such a way that the patient can be conveniently lifted into and out of the bed.

4.4 SENTA PUR L liftSENTA PUR L (sit-on lift) provides safe support for patients thanks to its adjustable safety arms. The safety arms can simply be folded to the front or to the back. The special pads provide perfect support for the patient and can be cleaned easily. A bedpan bracket is available as an optional accessory.

Prism Medical Ltd. V1-2014

9

Adjustable wings

Adjustable safety arms with belt holder

Handles

Illustration 3 Secure grid (optional) Foot rest

Back support, foldable

Adjustable safety arms with belt holder

Hygiene opening with cover

Illustration 4

Page 10: User Manual - Handicare Canada

4.5 Function of the Safety Arm RestsThe two safety arm rests can be folded or removed completely, independently from each other. In order to fold the safety arm rest turn them towards the back, as shown on illustration 5.

To remove a safety arm rest turn it until the two red marking on the safety arm is aligned with the red mark on the socket (see illustration 6). The safety arm rest can now be removed by pulling it out. For installation, follow the instructions in reverse order.

4.6 Battery Pack positionOpening up the battery pack door allows the lift through a doorway 90 cm wide.

Prism Medical Ltd. V1-2014

10

Illustration 5 Illustration 6Marking

Illustration 7 Standard position

Illustration 8 Folded upwards

The safety arm rest must not be used to support patients getting out of the lift.

The battery pack must be put back into its standard position after getting through the door.

NOTE:

NOTE:

Page 11: User Manual - Handicare Canada

4.7 Detachable Wings for the SENTA PUR L lift

Use the SENTA PUR L lift with wings for lying patients.

Detaching:• Position the wing to be detached from the lift into its end position and hold it.• Turn the lock (illustration 11) by ¼ twist to the left or right to unlock it.• The grid rod can only be removed from the grid (illustration 9) when the wing is in its end position,

remove the grid rod from the grid.• You can now push the wing out of its guides (illustration 10) towards the lifting pole and remove it.

Installation:• Push the wing to be installed into the guides.

• Attach the grid rod to the grid after bringing the wing into the top end position.

• Turn the lock by ¼ twists to the left of right (see illustration 11).

Prism Medical Ltd. V1-2014

11

Illustration 9 Wing for lying patients installed with grid rod

Illustration 10Wing for lying patients installed with guides

Illustration 11 Lock wing for lying patients

Make sure the grid rod snaps in correctly.

NOTE:

Page 12: User Manual - Handicare Canada

5. SENTA Digital Scale SL

The SL digital scale consists of the hand-held part and a pick-up. This SL scale is designed for measuring forces and weights. This is a non-calibrated scale.

Calibration can be done in the units N, kN, g, kg, t or lb. Depending on the design SENTA PUR L can weigh persons with a weight of up to 150 kg.

Prism Medical Ltd. V1-2014

12

ON/OFF

Illustration 12 Digital scale SL

Tareing

Page 13: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

13

5.1 Safety Instructions Digital Scale SLAll who are involved in configuring, calibrating and maintaining the SL scale need to:• Be qualified accordingly,• Read, understand and observe these instructions precisely,• Make sure that safety instructions are observed.

5.2 Cleaning the Scale• Wipe lightly only, do not use abrasive or caustic detergents.

5.3 Start-upFor a flawless operation of the hanging scale please observe the following:• Before weighing:

Switch on the scale with the On/Off-button. Wait until the scale automatically tare in at 0.0 kg.• Preparing the weighing process:

Position the resident on the SENTA PUR L lift. Make sure that the resident floats freely and the weighing process is not impaired through that the resident still rests partly on e.g. a chair / wheel chair / bed. Now the patient’s / resident’s correct weight is displayed (in steps of 0.1 kg).

• Attention: When calculating the weight take into account that the strap and the resident’s clothing need to be subtracted for determining the actual body weight.

The scale can now be switched off by pressing the On/Off-button again.

5.4 Zeros and TarringBy pressing the button 0/T the digital scale is set to zero before measuring and without any pre-load. By using this function also pre-loads (handle, side wing, and strap) can be tarred. These pre-loads are to be attached correctly before measuring and to be tarred by pressing the respective button.

5.5 Replacing batteriesThe batteries’ charging state is monitored by the digital SL scale. If the battery is almost empty the word “Batt” will blink in the display’s bottom row. Replace batteries with 4 x 1.5 V type AA batteries.

Page 14: User Manual - Handicare Canada

6. Combo Bath Lift with calibrated scale

6.1 Use of CalibrationOnly correctly sealed devices may be used for calibrated applications. It is the operator’s responsibility to guarantee this. It is possible to carry out device settings not covered by weights and measures regulations. Reviewing whether settings meet legal regulations should be carried out on a case-by-case basis.BEKA Hospitec EDI 03 weighing device is a highly precise digital weighing indicator based on the state-of-the-art Sigma-Delta A/D technology. This guarantees for a precise and fast weight indication.

6.2 ApprovalIn compliance with EN 45501 OIML up to 4000d for non-automatic scales of categories III and IIII CE label.

6.3 Product CharacteristicsThe EDI 03 weighing indicator is suitable for simple weighing applications with tareing and pressure functions.Built-in functions:• Battery operation and power supply operation via power plug.

• RS232 interface for printers, PC-communication or secondary display.

• 2 x digital threshold value outlet with status indicator.

• Function button <f> for counting scale, weighing persons, peak value indicator, holding or kg / lb switching.

• Real-time clock and NVRAM for saving current values such as Tara value, zero point etc.

• Environmental conditions: Environmental temperatures: Between -10 and 40°C Relative humidity: < 90% NiMh-battery operation: 9.6 V DC Uninterrupted power supply: 12 – 24 V DC, 2.5W max.

• Electrical safety:For your own safety make sure to observe environmental conditions. Always unplug the device for carrying out maintenance works or repairs.

Prism Medical Ltd. V1-2014

14

Illustration 13 Calibrated scale

Page 15: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

• Cleaning:Never use aggressive detergents or solvents. Clean with moist cloth and soap suds.

• Area of Application:The EDI 03 weighing indicator is not meant to be used in explosion-prone areas. The device has the IP X5 protection type.

• Disposal:Remove the batteries before dispose of the device. All batteries are to be disposed of in an environmentally compatible manner.

6.4 Operating the Scale

15

Button Description

The <POWER> switches the device on/off. To switch it off keep the button pressed for 3 seconds.

In battery operation the background illumination is switched off automatically. Press the <POWER> button to switch the light back on.

The <POWER> button has a memory function. After a power blackout the original state will be restored, i.e. if the device was switched on it will be switched on again. This allows the device to be – once switched on – always available as soon as power supply is provided.

If empty scale display a number other than zero this button is used to put it back to zero. The EDI 03 weight indicator saves the zero without the need for power and will thus be kept even in case of a power blackout.

This button tares the scale. The display will switch to 0, and in the status row you will read NET.

The <TARE> button covers the whole possible weighing range. This is about subtractive tareing, i.e. the weighing range is reduced by the

tared value. The EDI 03 weighing indicator saves the tared value also without the need for power.

Switches between gross and net. For this, a tared value needs to be available, of course.

Page 16: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

16

Button Description

Symbol Name Description

If a printer is connected and the printing function has been activated in the menu, <PRINT> will read a weight value, the display will read PRINT and the data will be sent to the printer. If no printer is connected, the device will wait for 10 sec. max. and abort.

Each printed weight value will be automatically added to the sum storage. Long press on <PRINT> will print the sum. The sum storage will be deleted

afterwards.

This button is automatically mapped to the HOLD function. It is possible to freeze the weight indicator with the HOLD button. The HOLD icon will be displayed. By pressing the HOLD button again the normal weight indicator will be re-activated.

Zero Magnifier Gross display within ± 1/4d.

NET Scale is being tared.

MOTION Scale is being moved.

Upper Threshold Value Weight higher than the set threshold value. Target high.

Lower Threshold Value Weight lower than the set threshold value. Target low.

Zero band Weight within zero range.

Holding Display is being held.

Battery Battery low, connect to power plug or replace batteries.

6.5 “Stable Error” Error MessageIf the buttons <ZERO>, <TARE> or <PRINT> are pressed the device will wait, before carrying out the function, for a stable weighing cell signal.When no stable signal is available within 10 seconds, STABLE ERROR will appear on the display and the function will be cancelled. Repeat this action, if necessary.

Page 17: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

17

6.6 LIVE WTThis function allows persons to be weighed who keep moving causing their weight to fluctuate heavily.

In LIVE WT mode (person weighing) do as follows:

• If <LIVE WT> is pressed briefly then this function is identical to <HOLD>. • Keep <LIVE WT> pressed for 2 seconds to switch between person weighing and normal weight display. After this the screen will read NORMAL or LIVE.WT for a brief period.

• If the device is within the zero range the weight will be indicated as usual,

i.e. the live WT function is deactivated.

• Press button <TARE> or <ZERO> to set the emptied scale back to zero.

• Let the person get on the scale.

• As soon as the weight is outside zero an averaging is carried out over a longer period of time. After that the device will display Hold the weight indication is “frozen” and the device will bleep.

• Press the <LIVE WT> button again to trigger another averaging.

• As soon as the gross weight is within the zero range, i.e. the scale has been emptied, the process can be repeated.

2 seconds

Page 18: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

18

Error Description Measure

Error Description

(U - - - - ) Scale is under loaded. Increase weight or switch scale off and on again.

(O - - -- -) Scale is overloaded. Reduce weight.

(TARE) Tarring attempt outside allowed range Change weight. (ERROR) (Calibrated scale may e.g. be only tarred with positive gross weight).

(ZERO) Attempt to set zero outside allowed Change weight or switch scale (ERROR) range. off and on again.

(STABLE) <ZERO>, <TARE> or <PRINT> Repeat when scale are stable. (ERROR) command cannot be executed due to missing standstill.

(ENTRY) A setting has been tried to carry out in SAFE-Setup that is only (DENIED) possible in FULL-Setup.

(LIN.PT Linearization below zero has been attempted. (LO)

(PT.TOO) Adjustment points are close to each other (<2%), an adjustment is not (CLOSE) possible. Use other (mostly higher) weights.

(RES) Too few calibrated values (<100) set. (LO)

(RES) Too few calibrated values set. (HIGH)

(SPAN) Weighing cell signal too low for this. (LO)

(SPAN) Weighing cell signal too high for this setting or will be too high under full load. (HIGH)

(ZERO) Zero point below -2mV/V. (LO)

(ZERO) Zero point above +2mV/V. (HIGH)

6.7 Error Message during Operation

6.8 Error Mode in Setup Mode

Page 19: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

19

Error Description Rectification

E0001 Line voltage too low. Check

E0002 Line voltage too high. Check

E0010 Temperature too high / low Check environment (-10 to +50°C ambient).

E0020 Scale division not correct (target Change value between 100 and 30,000 divisions).

E0100 Settings are lost. Repeat settings

E0200 Scale adjustment is lost. Re-calibration

E0300 All settings and scale adjustments Full re-adjustment and re-calibration are lost.

E0400 Factory parameters are lost (FATAL). Service

E0800 EEPROM defective (FATAL). Service

E2000 Weighing cell input is not correct. Check weighing cell

E4000 Battery-buffered storage is lost. Repeat settings

E8000 FLASH memory defective (FATAL). Service

6.9 Diagnosis Error Messages

E – error messages add up. Example:• E0012 (E0010+E0002) indicates that line voltage as well as temperature are too high.• E0C00 (E0800+E0400) indicates that factory parameters are lost and that the EEPROM is

defective.Hexadecimal numbers (basis 16) are as follows:1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - A - B - C - D - E - F(example: 2 + 4 = 6, or 4 + 8)

Battery charging:The line voltage supply of the patient scale is established via an 8-cell nickel / metal hydrite battery pack.

Page 20: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

20

Charging process:1. Plug battery charger into the power outlet.2. Plug the battery charger’s connector into the charging

socket, right hand side below the scale cover at the Senta lift.

3. Charging indicator shines red, battery pack is being charged. Charging time of the complete battery pack is 16 hours (battery pack capacity is 2700 mAh, sufficient for approx. 24 hours of weighing).

4. There is no interruption; this is why you should quit the charging process after the given period of time.

Illustration 14 Charging socket

Page 21: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

7. Battery device 24 Volt

7.1 Power Supply for the SENTA PUR L liftSENTA PUR L lift is equipped with a 24 V battery system. SENTA PUR L lift does not need to be connected to the power supply system in order to operate. When charging the battery alwaysensure power supply is in compliance with local requirements.

7.2 Start-up and Function (Electric motor / Battery unit)SENTA PUR L lift is equipped with a 24 V electric motor. Fully charge the SENTA PUR L lift battery before initial start up.

Unlock the emergency stop switch by turning in it clock-wise until it unlocks.

21

Illustration shows Control box CBJ Part No. 278849

Emer

genc

y st

op s

witc

h

Charging socket

Han

d co

ntro

l

Mot

or c

onne

ctio

n

Fully charge battery before initial start up (charging time approx. 4 hours). Unlock the emergency stop before starting the seat pickup process.

NOTE:

Page 22: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

22

Illustration shows: SENTA PUR L Battery Part No.: 280680

Handle with trigger

For

rem

ova

l the

rel

ease

tri

gg

er

need

s to

be

pul

led

.

Illustration shows Wall mount charger Part No. 278104

Illustration shows Coiled cord

Page 23: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

7.3 Hand control for SENTA PUR L lift

Up / Down Command Hand control:Adjust the height for seat pickup by using up or down buttons.

LED explanation:Green = Power supply operation / Yellow = Charging indicator / Yellow = EMERGENCY – Battery capacity low, remaining capacity approx. 10-15% / LED off = No power supply connection.

Control box CBJ 2:SENTA PUR L lift is equipped with a 24 V battery system. In order to remove the battery unlock the battery by using the trigger in the battery handle. To insert the battery place the battery in the control box, lift it slightly and push to the back until you hear the trigger snap in.

23

Illustration 15 SENTA PUR L Hand control

Part No. 278703

Up command

Down command

Battery charging situation Service

Max. weight reached

Page 24: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

7.4 External Charging Device / Wall-mounted Station CHJ2The external charging device is a switch-mode power supply and is shipped pre-installed (on installation rail). It can be attached to suitable walls. The required power cord is included in the delivery. The charging time for the battery unit is only approx. 4 hours.

Maintenance and Care of the 24 Volt batteries:The battery as well as the control box must not be opened by the customer.Repairs shall only be carried out by BEKA Hospitec or by BEKA Hospitec authorized companies. Re-charge empty batteries immediately, if possible, since this will extend their life-cycle. Re-charge stored batteries at least every 6 months (a possible deep discharge may destroy the battery). The battery’s life-cycle depends heavily on the factors load (number of stroke cycles) and the state of charge. It can last for up to 5 years. Replace defective or worn batteries and defective charging devices as soon as possible.

7.5 Type Label

24

Illustration 16

Indicator for power supply operation (green LED)

Cord holder (optional)

Charging indicator (yellow LED)

The type label can be found on the pole of the combo bath lift. The serial number can look as follows “Example 1311L60011”. If you have any questions, please always refer to the serial number of the combo bath lift on the type label.

Page 25: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

8. Operating the SENTA PUR L lift

The seat of the SENTA PUR L lift is moved up or down by using the provided hand control. Symbols mark the respective movement direction. The steering rollers at the back can be locked in position. When picking up a patient the steering rollers need to be locked in position to prevent any uncontrolled movement of the SENTA PUR L lift.Travel range: Before operating the SENTA PUR L lift look for any possible interferences and correct them (e.g. wall shelves etc.). Furthermore, keep in mind that the max. Height is 1310 mm.

8.1 Patient PickupPick up a patient from a bed:1. Move the SENTA PUR L lift close to the bed.2. Open both arm rests (towards the back).3. Fold the side grid off the patient’s bed.4. Now turn the patient to the side.5. Position the patient lift with the lying area or the seat on the mattress.6. Lock the castors of the lift into position.7. Now turn the patient back onto the lying area / seat of the lift.8. Fold the front arm rest down so the patient can hold fast.9. Help the patient taking a seated position, fold the back arm rest (with back padding) down as

soon as the patient is in a seated position.10. Fasten the patient’s safety belts.11. Release the steering rollers’ brakes.12. Move the lift up before moving it away from the bed.13. Lower the lift into transport height and moving the patient/resident approach the bath tub.

25

Page 26: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

26

8.2 Manual Emergency LoweringTriggering the emergency lowering:The SENTA PUR L lift is equipped with an emergency lowering function. Find the hole in the end cap of the lifting pole to lower the respective patient pickup manually by using the provided 6 mm allen key. The emergency lowering functions must only be used in cases of emergency. The battery back needs to be removed, and the power plug must not be connected to the power supply grid.

Note: The emergency lowering mechanism must not be treated with oil, grease or other lubricants.

8.3 Contact Partner for Technical InformationFor technical information and for queries regarding the application please always get in touch with the medical product advisor of your sales partner or the manufacturer.

BEKA Hospitec GmbHAm Rübenmorgen 3 D-35582 WetzlarFon: +49(0)641-9 22 20 - 0Fax: +49(0)641-9 22 20 - [email protected]

485 Millway Ave. Concord Ont. L4K 3V4 | T 1.800.265.0677 | F 416.260.5580 www.prismmedical.ca

Page 27: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

9. Cleaning / Disinfection

9.1 Cleaning the SENTA PUR L liftClean the SENTA PUR L lift with a soft, lint-free cloth. For thorough cleaning use a cloth moistened with soap suds. For disinfection you can moisten the cloth with isopropyl-alcohol solvent.In order to avoid any damage no aerosol detergents, sprays, abrasive or solvents must be used for the control unit.

SENTA PUR L lift can be moist-cleaned with a common agent for cleaning synthetic material with.

9.2 Disinfection of the SENTA PUR L liftRegarding the transfer risk you have to thoroughly disinfect your SENTA PUR L lift after each usage. For surface disinfection, a commercial isopropyl-alcohol solvent or a disinfection aerosol can be used.

9.3 SterilizationThe SENTA PUR L lift is not suitable for sterilization.

27

Do not operate the lift during the cleaning process

“Unplug power cord“CAUTION:

Disinfection must be carried out after every use.

WARNING:

Page 28: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

10. Disposal 10.1 Disposal of the Packing MaterialTreat the packing material provided with the SENTA PUR L lift in compliance with local applicable laws and regulations (recycling). Metal parts as well as plastic and electronic components need to be treated in compliance with the WEEE (recycling).

10.2 Product DisposalAfter the end of the product life cycle get in touch with your BEKA contractual partner, for proper disposal in compliance with local applicable laws and regulations (recycling).As the manufacturer, BEKA Hospitec GmbH will provide more information on environment compatible disposal.

11. Safety Check

A safety check needs to be carried out with the SENTA PUR L lift every 12 months along with maintenance.

Safety checks and maintenance must be documented.

Checks must be carried out by the manufacturer or professional staff authorised by the manufacturer only.

28

Page 29: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

12. Troubleshooting / Customer support

12.1 Troubleshooting Guide

Problem with lift Solution

Lift does not work up and down. a) Check that the emergency stop button is not pushed.

b) Check hand control and all cables are connected correctly.

c) Check that the battery is charged. d) Check that the battery contacts are not

defective.

Lift stops in the top position. a) Check that the emergency stop button is not pushed.

b) Check that the battery is charged. c) Use the mechanical emergency lowering

to lower the patient. (see engine) d) Use the electrical emergency lowering to

lower the patient. (see hand control)

Base motor does not work. a) Check that the emergency stop button is not pushed.

b) Check connections in the control box.

Actuator “stalls” during lift. a) Battery is low. Charge battery. b) The lifting capacity is exceeded (max.

patient weight).

The electrical positioning adjustment does not respond. a) Check if the hand control is connected. b) Check the plug-in connection on the

control box.

Audible “beep” is heard from the control box. Battery is low. Charge battery.

Hand control does not respond. a) Check the connector on hand control cable.

b) Check the battery capacity (replace with a fully charged battery).

29

Page 30: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

Problem with lift Solution

Prism Medical Ltd. V1-2014

Raise and lower buttons on hand control do not respond. a) Check that the emergency stop button is not pushed or pressed.

b) Check the battery capacity (replace with a fully charged battery).

c) Check that the cables are properly connected at the control box.

Castors make loud noises. Clean castors.

If you hear unusual sounds. Contact BEKA Service.

If the SENTA PUR L Lift is damaged. Contact BEKA Service.

Problem with charger Solution

Battery charging does not work. a) Check that the battery contacts are ok. b) Check the power plug.

“Power on” light on charger is not lit. a) Check if the charger is connected to the wall receptacle.

b) Check the fuse socket.

Charger is plugged in, but “Power on” light is not lit. a) Check that there is power to the wall outlet.

b) Remove the battery and check for damage.

Problem with battery Solution

Battery is properly seated but no lights are visible. Contact BEKA Service.

Indicator light does not got off after several hours Battery needs replacing. Contact BEKA of charging time. Service.

Battery indicates it is fully charged when in the charger, Battery needs replacing. Contact BEKA but when placed in the lift, will only do a few lifts. Service.

Upon actuation of the hand control, a warning signal Check the battery charge state. and the indicator lamp (red) light up.

12.2 Customer SupportContact your sales partner’s customer support or the manufacturer if the SENTA PUR L lift does not work as intended and it is not resolved with troubleshooting (section 12.1).

30

Page 31: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014 Prism Medical Ltd. V1-2014

13. Appendix

13.1 Technical data

Dimensions and Weights

Length arms open: 86.0 cm

Length arms closed: 94.0 cm

Width: 71.8 cm

Height: 137.5 cm

Height min: 47.5 cm

Height max: 117.5 cm

Width wing: 43.0 cm

Length wings open: 184.1 cm

Scale precision: 100 gram

Weight without packing: approx. 55 kg

Safe Working Load (SWL): max. 150 kg

Electrical data

Control unit / charger

Type of protection: Class II

Application part: Type B

Operating mode: Short-time operation

Input voltage: 100 V - 240 V~ (AC) / 50 / 60 Hz

Output voltage: 24 V (DC)

Power consumption: I in max. 1 Amp. / I out max. 10 Amp.

Treatment duration: 6 double strokes, 25 min. break

Hoist motor

Operating mode: Short-time operation

Input voltage: 24 V (DC)

Power consumption: I max. 7 Amp.

Degree of protection: IP X4

Treatment duration: 2 min. continuous operation followed by an 18 min. break

31

Page 32: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

Spindle motor

Operating mode: Short-time operation

Input voltage: 24 V (DC)

Power consumption: I max. 7 Amp.

Degree of protection: IP X4

Treatment duration: 2 min. continuous operation followed by an 18 min. break

ENVIRONMENTAL CONDITIONS Operation

Temperature range: Between 10° C and 40° C

Relative humidity: Non-condensing between 30% and 75%

Atmospheric pressure range: 800 – 1060 hPa

Storage and Transport

Temperature range (Lift): Between -40° C and 70° C

Temperature range (Battery): Between -15° C and 40° C

Relative humidity: Non-condensing between 10% and 80%

Atmospheric pressure range: 500 – 1100 hPa

32

BEKA Hospitec GmbHAm Rübenmorgen 3 D-35582 WetzlarFon: +49(0)641-9 22 20 - 0Fax: +49(0)641-9 22 20 - [email protected]

485 Millway Ave. Concord Ont. L4K 3V4 | T 1.800.265.0677 | F 416.260.5580 www.prismmedical.ca

Page 33: User Manual - Handicare Canada

13.2 Dimensions SENTA PUR L lift

33

Page 34: User Manual - Handicare Canada

Prism Medical Ltd. V1-2014

13.3 Icons and Symbols used

Accompanying documents / Observe operating manual.

Warning of a hazard area.

Application part “Type B” according to DIN EN 60601-1.

No domestic waste. Do not put the treatment system and packing materials into the domestic waste.

CE mark according to 93/42/EEC for medical products.

Indoor use only.

Type of protection: Class II.

34

Page 35: User Manual - Handicare Canada

Distributed by:

485 Millway Ave. Concord Ont. L4K 3V4 | T 1.800.265.0677 | F 416.260.5580 www.prismmedical.ca

Prism Medical Ltd. V1-2014

13.4 Declaration of conformityThe declaration of conformity of the lifting tub will be provided by the manufacturer BEKA Hospitec GmbH upon your request.

13.5 Spare Parts and Accessories for SENTA PUR L lift

Spare Part No. Description

278703 SENTA PUR L Hand control HB52 for LA34

278713 SENTA PUR L cover

278714 SENTA PUR L Seat cover, with hygeine opening

278715 SENTA PUR L Back Rest

278716 SENTA PUR L Lid for hyg-opening

278738 SENTA PUR L Wings cover

278729 SENTA PUR L swing away attachment for Control box

278723 SENTA PUR L Handle complete

278849 Control box CBJ

280680 Battery

280104 Wall charger

Accessory No. Description

278700 SENTA PUR L lift (dual controls and on board charging)

278702 SENTA PUR L Digital Scale

278705 SENTA PUR L Side Bars/Rails

35