Copyright © 2020 Netzer Precision Motion Sensors Ltd. All rights reserved. USER MANUAL ABSOLUTE POSITION ROTARY ELECTRIC ENCODER TM Electric Encoder Explorer Software Auto-calibration Feature Table of Contents 1. Overview ................................................................................................................ 2 2. Main screen .......................................................................................................... 2 3. Auto-calibration process stages ................................................................... 2 4. Abort auto-calibration process ...................................................................... 3 5. Restart auto-calibration process .................................................................. 3 6. Auto-calibration failures ................................................................................... 3 EES Software | Auto-calibration -UM-V01

USER MANUAL - Netzer

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

Page 1: USER MANUAL - Netzer

Copyright © 2020 Netzer Precision Motion Sensors Ltd. All rights reserved.

USER MANUAL

ABSOLUTE POSITIONROTARY ELECTRIC ENCODERTM

Electric Encoder Explorer Software Auto-calibration Feature

Table of Contents1. Overview ................................................................................................................ 22. Main screen .......................................................................................................... 23. Auto-calibration process stages ................................................................... 24. Abort auto-calibration process ...................................................................... 35. Restart auto-calibration process .................................................................. 36. Auto-calibration failures ................................................................................... 3

EES

Soft

war

e | A

uto-

calib

ratio

n -U

M-V

01

Page 2: USER MANUAL - Netzer

2

EES

Soft

war

e | A

uto-

calib

ratio

n -U

M-V

01

USER MANUALElectric Encoder Explorer Sotware | Auto-calibration ABSOLUTE POSITIONROTARY ELECTRIC ENCODERTM

1. OverviewAuto-calibration is a new feature for easy calibration at the customer’s site. It is supported by encoders with FW 4 version 4.1.3 or higher.

For encoders that support auto-calibration, under the “Calibration” button, additional “Auto-calibration” button will be displayed.

2.2 Main screen for encoders with auto-calibration support

2. Main screen

2.1 Main screen for encoders without auto-calibration support

3. Auto-calibration process stagesThe Auto-calibration process consists of three stages:

• Noise test - This stage runs a jitter test for the Fine, Medium, and Coarse encoder channels. During the Noise test, the shaft must be stationary.

• Accuracy calibration - During Accuracy calibration, the shaft must rotate continuously. The accuracy calibration of Fine, Medium, and Coarse encoder channels includes:

• Amplitude test

• Offsets calibration

• AP (Absolute Position) calibration

• At this stage, CAA and MAA are calculated

• After connection, a FW 4 encoder with auto-calibration support, an additional “Auto-calibration” button” will be added beneath the “Calibration” button.

• Press the “Auto-calibration” button.

• The main auto-calibration window opens.

(It is still possible to perform manual calibration using the regular “Calibration” button)

After pressing the “Auto-calibration” button on the Encoder Explorer main screen, the bellow shown Auto-calibration main window will open:

1. Make sure the shaft is standstill

2. Press the “Start”

Noise test - The encoder should stay standstill at this stage.

The Auto-calibration process starts the Noise test.

If the Noise test completes successfully (in the event of failure – see below):

• The “Noise test” label will be marked with a green check mark.

• The Auto-calibration process starts the Accuracy calibration.

Once the noise test is successful, start rotating the shaft continuously until the auto-calibration process finishes.

Page 3: USER MANUAL - Netzer

3

EES

Soft

war

e | A

uto-

calib

ratio

n -U

M-V

01

USER MANUALElectric Encoder Explorer Sotware | Auto-calibration ABSOLUTE POSITIONROTARY ELECTRIC ENCODERTM

3. Wait until encoder is reset.

4. The Auto-calibration process is successfully finished.

5. (Optionally) press the “View data” button to review the calibration results.

6. Close the Auto-calibration window

4. Abort auto-calibration processAt any stage, it is possible to abort the auto-calibration process by pressing the “Abort” button.

5. Restart auto-calibration processIt is possible to restart the auto-calibration process by pressing the “Restart” button.

6. Auto-calibration failuresIf a test fails (for example the Noise test) – the result will be marked with in red X.

Page 4: USER MANUAL - Netzer

4

EES

Soft

war

e | A

uto-

calib

ratio

n -U

M-V

01

USER MANUALElectric Encoder Explorer Sotware | Auto-calibration ABSOLUTE POSITIONROTARY ELECTRIC ENCODERTM

In the event of an auto-calibration failure, corrective recommendations will be displayed.

After pressing the “Detailed data” button, detailed information regarding the failure will be displayed