User’s manual

®

Page 2: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

2 ● NG39EN020E ● 05/2013 ● (13) ● V5 ● J05451 ●User’s Manual ScanWave

figure 1

1

3

4

5

8

9

76

10

11

2

Page 3: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 3

ENGLISH

I – DESCRIPTION ............................................ 4

II – ROUTINE OPERATION ................................. 6

III – CONFIGURATION ...................................... 7

IV – SECURITIES ............................................ 7

V - ACCESSORIES ........................................... 8

VI - TROUBLESHOOTING GUIDE ......................... 8

VII – SYMBOLS AND ABBREVIATIONS .................... 10

Page 4: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

4 ● NG39EN020E ● 05/2013 ● (13) ● V5 ● J05451 ●User’s Manual ScanWave

This is the user’s manual for your Scanwave device.This document must be used together with the general user’s manual for SATELEC curing lamps.

I – DESCRIPTION

1.1 PresentationAs symbolised in figure 1, ScanWave includes the following components:

• The multi-fiber light guide, 45° curve, Ø7.5mm, sterilizable (fig. 1-1).

• The handpiece (fig. 1-3) fitted with a protective cap (fig. 1-2).

• Lithium-ion battery (fig. 1-4), assembled when delivered.

• Charging base (fig. 1-5).• Power supply.• The main plugs (USA, AUS, EEC, UK).• The rigid protection shield.• Accompanying documentation.

On option:• Kit of 5 flexible protective shields, sterilizable, Ø7,5 mm.• Multi-fiber light guide, 45° curve, Ø 5,5mm,

sterilizable.• Kit of 5 flexible protective shields, sterilizable, Ø5,5 mm.

1.2 technical descriPtionScanWave is equipped with a Light-Emitting Diode (LED) that emits blue visible light of a wavelength between 390 and 505nm for the photopolymerization of dental materials.

Scanwave incorporates a laser aiming system with a wavelength of 655nm and a light intensity

of less than 1mV, allowing the light ray to be directed with precision.

ATTENTION - LASER LIGHT - DIRECT EXPOSURE DANGEROUS FOR EYES - CLASS 2 LASER UNIT

IEC –60825 standard-1 second edition 2007 - Safety of laser devices - Max. power: 1mW - Pulse duration: duration

of user pressing button - Wavelength: 655nm

1.3 lcd screenThe LCD screen is backlit and allows the display of the different pieces of information required by the user.

2

35

1

4

As shown above, it comprises different zones identified from top to bottom as follows:• A text box displaying the curing mode selected. (1) • A display in seconds of the duration of the selected

curing cycle. During operation, this countdown display indicates the remaining activation time until the current cycle is complete. (2)

• Three icons represent the type of light emission. (3) Continuous light

Flashing light

Gradual light

• A battery charge level indicator symbolised by a ladder with 0 to 5 levels. (4)

• A laser function activation indicator: LA. (5)

1.4 Menu activation and selectionThe handpiece comprises several control devices.Two buttons located under the LCD screen can be used to: - Configure the required mode by pressing the left-

hand «M» Mode button (fig. 1-6) - Select in this mode, the curing cycle to be used.

Right-hand «C» Cycle button (fig. 1-7)The curing cycle is initiated by pressing the control trigger located under the device (fig. 1-8) or the ON/OFF button (fig. 1-9) located on the aluminium tip.

1.5 Battery autonoMyThe charge level indicator shows the remaining battery power. Its charge time is about 3 hours.• When the battery is completely discharged, the LCD

Page 5: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 5

screen displays the message «Lb» (Low battery), the handpiece beeps 4 times, and the start of a curing cycle will be inhibited.

• You must then replace the ScanWave on its charger.

1.6 rechargingThe translucent «Neck» of the charger (fig. 1-10) shows the charger operation status and the level of lighting power emitted by the light.Place the charger on a hard, stable surface with a flatness of less than 5°, then plug the connector into the grid supply. The translucent part of the support successively emits three flashes: red, green and blue.Then insert the light guide in the charger and ensure charging terminals are in contact with battery connections.The translucent part then emits a blue light signalling that the battery is charging. The blue light goes out when charging is complete.

1.7 checking lighting Power The charger can be used to check the lighting power emitted by ScanWave. To do this, a control window is located in the rear section of the support (fig. 1-11).To check the lighting power emitted:• Select Scan mode; continuous; 9 seconds.• Clean the optical guide, make sure that it is not

damaged, and then insert it into the handpiece.• Also make sure that the control window is unbroken

and clean.• Place the tip of the optical guide flat on the control

window and activate ScanWave.• If the transluscent «Neck» of the charger lights up

green, the lighting power is sufficient. If the lighting power is not sufficient, this will be signalled by a fixed red or flashing red/green light. In this case, refer to chapter X.

1.8 technical sPecificationsModel name ScanWaveManufacturer SATELECMedical category IIa according to the

directive 93/42/CEEHandpieceWeight 185gDimensions Ø34 x 220mm Operating cycle See II - Routine operation -

Clinical careCategory Type BProtection 1,5 A T Fu1 fuse (non

accessible) 125VProtection index IPX0AC power adapterInput voltage 100V AC to 240V ACFrequency 50Hz to 60HzOutput voltage 12 V DC Output current 0.8 A Category IIProtection index IP 41Charging baseInput voltage 12V DCProtection 3 A T Fu1 fuse (non

accessible) 125VOperating cycle Continuous operationProtection index IPX0Battery Type Lithium-IonDimensions Ø34 x 115mmCapacity 2500mAhOptical specifications LED sources 4 Wavelength range 390nm-505nmCentral wavelength 405nm / 440nm / 460nm

and 480nmIntensity 1500 mW/cm² ± 10% (for a 7.5mm fiber : active diameter 6.8)2200 mW/cm² ± 10% (for a 5.5mm fiber : active diameter 4.8)Maximum exposure time 60 seconds (pulse mode at 1500 mW/cm² ± 10%)Laser aiming systemTypical wavelength 655nm

Page 6: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

6 ● NG39EN020E ● 05/2013 ● (13) ● V5 ● J05451 ●User’s Manual ScanWave

Minimum wavelength 650nm Maximum wavelength 660nm Intensity < 1mW Class 2Exposure time As long as user presses the

buttonTemperatureOperation +10°C to +30°C Stockage 0°C to +50°CHumidityOperation 30% to 75%Storage 10% to 70% condensation

includedAtmospheric pressure 500hPa to 1060hPa

II – ROUTINE OPERATION

2.1 oPerating for the first tiMeOn receipt of the unit, look for any damage that may have occurred during transportation. If necessary, contact your supplier.

Prior to any use, it is essential that the sterilizable accessories (light guide and protective shield) be sterilized and that the unit base be disinfected (see the maintenance chapter in the attached general user’s manual for SATELEC curing lamps).

Remove the protective caps from the handpiece (to be kept during maintenance to prevent the liquid products from damaging the LEDs), and insert the sterilized light guide into the handpiece. Ensure that the light guide is properly inserted, confirmed by a click.Ensure that all the segments of the LCD screen are present The battery supplied is only charged to 60% prior to shipment. Charge it fully (See 1.6 - Recharging) prior to using it each time.

2.1 selecting curing cyclesScanWave has several curing cycles classified into two families called «Modes». At start-up, the device runs a sequence of automatic tests and positions itself on the last cycle used.

• To switch from one mode to the other, quickly press the «Mode» button located on the left under the LCD screen.

• To scan and select one of the cycles contained in these modes, quickly press the «Cycle» button located on the right under the LCD screen.

C C

M

2.2 curing Modes and cyclesScan modeTo systematically scan a continuous spectrum from 390nm to 505nm, the Scan mode can be used to activate all photosensitive dental equipment.This mode has 2 pre-programmed cycles:• A continuous 9-second cycle.• A gradual 20-second cycle including a 5-second

increase in power, 10 seconds at stable full power, and a 5-second linear descent at the end of the cycle.

Bonding modeDesigned for dental adhesives and marked «Bond.», this mode has 2 cycles:• A 5-second continuous cycle.• A 12 x 5-second pulse cycle.

In dentistry, the pulse cycle is used to bond a complete arch passing from one bracket to another, with each space coinciding with a short extinguishing of the light.

2.3 oPerationSelect the curing mode and cycle that you want to use. The time corresponding to the selected cycle is displayed on the LCD screen in seconds. In addition,

Page 7: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 7

an icon indicates the type of light emission selected: continuous, flashing or gradual.

• Place the optical guide as close as possible to the surface of the equipment to be photocured but avoid contact as this may reduce the quality of the curing process. The use of a transparent strip is recommended.

• Press the ON/OFF button to start the polymerization cycle.

• A countdown counter on the LCD screen shows the time remaining.

• Press the button again for longer to stop the current cycle at any time.

III – CONFIGURATION

7.1 laser sightScanWave has a laser guiding system which positions the light beam in the direction of the material to be photocured. At the start of a cycle, press and hold down the ON/OFF button or the trigger to display a red sight circle. Release them to make the sight circle disappear and activate the light. This function allows you to view the light zone associated with optimum power. The laser radiation passes into the optical guide. Do not look in the optical guide.

3.2 cycle initiationBy default, in order to ensure that the device can be used easily by everyone, cycles are initiated by pressing the ON/OFF button or by pushing the control trigger.However, depending on the required ergonomics, the cycle start control can be configured:• One long press on the Cycle C button makes the

following screen appear:

• Activation is then possible either by pressing the button or the control trigger. Bo signifies «Both».

• A short press on the menu C button allows you to

select the pen-style hold. Initiation is then only possible by pressing the ON/OFF button and the screen displays PE: «Pen-Style».

• To validate the modification and exit configuration mode, press the Mode M button again.

• However, if you prefer the gun-type hold, quickly press the Cycle C button. Launch is then only possible by pressing the control trigger and the screen displays GU: «Gun-type».

• To validate the modification and exit configuration mode, press the Mode M button.

• Proceed in the same way to change the initiation mode again.

3.3 standBy ModeAfter 3 minutes without use, the unit goes into standby mode (low consumption), the LCD screen backlighting switches off.The unit can be taken out of standby mode by simply pressing one of the buttons.

IV – SECURITIES

ScanWave is equipped with a system for the detection of possible unit operation anomalies.

4.1 excessive teMPerature of the unitDuring intensive use, an excessive temperature may be detected and the screen then displays the letters «OH» (Over Heat) and the audible warning (beep) sounds 4 times. The light can now not be switched on.

Page 8: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

8 ● NG39EN020E ● 05/2013 ● (13) ● V5 ● J05451 ●User’s Manual ScanWave

It is then recommended that the user leaves the handpiece to cool for a few minutes, until the screen displays the selected time cycle again and then the light can be switched on again.

V - ACCESSORIES

The following accessories are available for ScanWave• ScanWave handpiece Ref F05460• ScanWave battery Ref F05451• ScanWave charging base Ref F05452• Power supply Ref X05494• Light guide Ø7.5mm Ref F05470• Rigid protection shield Ref F05407• Light guide Ø5.5mm Ref F05471 (option)• Kit of 5 flexible protective shields, sterilizable, Ø7.5mm Ref F61507 (option)• Kit of 5 flexible protective shields, sterilizable, Ø5.5mm Ref F61505 (option)

VI - TROUBLESHOOTING GUIDE

The unit is not repairable in-situ.All maintenance/repair procedures must be performed in a specialist workshop.

Symptom Possible causes Solutions

No operation (LCD screen off)

Battery completely discharged Recharge battery

Battery defective Return to SATELEC after-sales department

ScanWave defectivePut ScanWave back on the charging base

Return to SATELEC after-sales departmentNo operation (LCD screen on)

ScanWave defectivePut ScanWave back on the charging base

Return to SATELEC after-sales department

Light power defect or no light power

LED defective and / or dirty Check cleanliness of LEDs, clean with dry air jet and / or return to SATELEC after-sales department

ScanWave defective Return to SATELEC after-sales department

Defective buttons Return to SATELEC after-sales departmentInsufficiently charged battery Check charge indicator on the unit’s LCD screen, charge the

batteryLight guide defective and / or dirty

Clean light guide and / or return to SATELEC after-sales department

Page 9: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 9

Symptom Possible causes Solutions

Laser defect or no light power

Defective laser diode or defective buttons

Return to SATELEC after-sales department

Light guide defective and / or dirty

Clean light guide and / or return to SATELEC after-sales department

Charger does not operate

AC power wall outlet defective Contact your electrician

AC power adapter defective Return to SATELEC after-sales departmentLight and audible alarms defective Return to SATELEC after-sales department

Fuse defective Return to SATELEC after-sales departmentWrong battery detection Clean battery contacts and charger and / or return to

SATELEC after-sales department

Dysfunction of power level verification

Light guide defective and / or dirty

Clean light guide and / or return to SATELEC after-sales department

LED defective and / or dirty Check cleanliness of verification window and / or return to SATELEC after-sales department

Indicator light defective Check cleanliness of LEDs, clean with dry air jet and / or return to SATELEC after-sales department

Window defective and / or dirty Clean window

It is particularly recommended to continuously wear protective glasses during repair operations.

Page 10: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

10 ● NG39EN020E ● 05/2013 ● (13) ● V5 ● J05451 ●User’s Manual ScanWave

VII – SYMBOLS AND ABBREVIATIONS

SYMBOL DEFINITION

Alternating current

Direct current

Do not discard as household waste

Please refer to the accompanying documentation

«ON»/«OFF» (pushbutton)

Type B

Class 2 laser equipment

Protective glasses recommended

CE marking

No access for persons with active implant as pacemaker

Rx onlyUnited States Federal Law restricts on its territory the use of this unit exclusively to trained, capable and qualified dental healthcare professionals, or under their control.

Note: Technical personnel of the SATELEC authorized dealer network can obtain from ACTEON GROUP on request all the information they need for repair of the parts of the curing light that ACTEON has identified as repairable.

Page 11: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 11

Page 12: User’s manual · IV – SECURITIES ... • Charging base (fig. 1-5). ... 2.2 curing Modes and cycles Scan mode To systematically scan a continuous spectrum from

17 av. Gustave Eiffel • BP 30216 • 33708 MERIGNAC cedex • FRANCETel + 33 (0) 556 34 06 07 • Fax + 33 (0) 556 34 92 92

E-mail : [email protected] • www.acteongroup.com

J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E