i Content 1 Safety Information ...................................................................... 1-1 2 Notices ........................................................................................ 2-1 2.1 Warranty Notice .....................................................................................................2-1 2.2 Copyright Notice ....................................................................................................2-1 2.3 EMC Notice.............................................................................................................2-1 2.4 MRI Notice..............................................................................................................2-1 2.5 Intended Use Notice...............................................................................................2-2 2.6 IEC Classification.....................................................................................................2-2 2.7 Declaration of Conformity Notice ..........................................................................2-2 3 Equipment Symbols ..................................................................... 3-1 4 Introduction ................................................................................ 4-1 4.1 General description ................................................................................................4-1 4.2 Abbreviations and Definitions ................................................................................4-2 5 Theory of Operation .................................................................... 5-1 5.1 Pneumatic System ..................................................................................................5-1 5.1.1 Pneumatic Theory ..........................................................................................5-1 5.1.2 Main Pneumatic Assemblies Overview ..........................................................5-3 5.2 Electrical System.................................................................................................. 5-12 5.2.1 Electrical Box ............................................................................................... 5-12 5.2.2 Expiratory Valve Heating Module ................................................................ 5-13 5.2.3 Rear Panel View........................................................................................... 5-14 5.2.4 Display Inside View...................................................................................... 5-15 5.2.5 ARM platform circuit board......................................................................... 5-16 5.2.6 Display interface board ............................................................................... 5-17 5.2.7 Interface Board............................................................................................ 5-18 5.2.8 Mother Board .............................................................................................. 5-19 5.2.9 Core Board................................................................................................... 5-21 5.2.10 Main Control Board ..................................................................................... 5-22 5.2.11 Power Supply Board .................................................................................... 5-23 5.2.12 System Wiring Diagram ............................................................................... 5-24 5.3 Software System.................................................................................................. 5-30 5.3.1 Software Units and Their Relationship ........................................................ 5-30 5.3.2 Structure of the Software System ............................................................... 5-31 5.3.3 Hardware and Software Interface ............................................................... 5-32 6 Disassemble and Functional Testing ............................................ 6-1 6.1 General ...................................................................................................................6-1

V8600 Critical Care Ventilator Service Manual

  • Upload
    others

  • View
    12

  • Download
    0

Embed Size (px)

Citation preview

Page 1: V8600 Critical Care Ventilator Service Manual

i

Content

1 Safety Information ...................................................................... 1-1

2 Notices ........................................................................................ 2-1

2.1 Warranty Notice .....................................................................................................2-1 2.2 Copyright Notice ....................................................................................................2-1 2.3 EMC Notice .............................................................................................................2-1 2.4 MRI Notice..............................................................................................................2-1 2.5 Intended Use Notice...............................................................................................2-2 2.6 IEC Classification .....................................................................................................2-2 2.7 Declaration of Conformity Notice ..........................................................................2-2

3 Equipment Symbols ..................................................................... 3-1

4 Introduction ................................................................................ 4-1

4.1 General description ................................................................................................4-1 4.2 Abbreviations and Definitions ................................................................................4-2

5 Theory of Operation .................................................................... 5-1

5.1 Pneumatic System ..................................................................................................5-1 5.1.1 Pneumatic Theory ..........................................................................................5-1 5.1.2 Main Pneumatic Assemblies Overview ..........................................................5-3

5.2 Electrical System .................................................................................................. 5-12 5.2.1 Electrical Box ............................................................................................... 5-12 5.2.2 Expiratory Valve Heating Module ................................................................ 5-13 5.2.3 Rear Panel View ........................................................................................... 5-14 5.2.4 Display Inside View ...................................................................................... 5-15 5.2.5 ARM platform circuit board ......................................................................... 5-16 5.2.6 Display interface board ............................................................................... 5-17 5.2.7 Interface Board ............................................................................................ 5-18 5.2.8 Mother Board .............................................................................................. 5-19 5.2.9 Core Board ................................................................................................... 5-21 5.2.10 Main Control Board ..................................................................................... 5-22 5.2.11 Power Supply Board .................................................................................... 5-23 5.2.12 System Wiring Diagram ............................................................................... 5-24

5.3 Software System .................................................................................................. 5-30 5.3.1 Software Units and Their Relationship ........................................................ 5-30 5.3.2 Structure of the Software System ............................................................... 5-31 5.3.3 Hardware and Software Interface ............................................................... 5-32

6 Disassemble and Functional Testing ............................................ 6-1

6.1 General ...................................................................................................................6-1

Page 2: V8600 Critical Care Ventilator Service Manual

ii

6.2 Main Assemblies Disassemble ...............................................................................6-1 6.2.1 Disassembling of the Display from the Host ..................................................6-1 6.2.2 Disassembling of the Host from the Trolley ...................................................6-2 6.2.3 Disassembling of the Arm Assembly from the Host .......................................6-3 6.2.4 Disassembling of the Humidification Assembly from the Host ......................6-4

6.3 Disassembling of Maintenance Spare Parts ...........................................................6-5 6.3.1 Disassembling of Oxygen Sensor ....................................................................6-5 6.3.2 Disassembling of Pressure Sensor Board .......................................................6-6 6.3.3 Disassembling of Supply O2 Assembly ............................................................6-7 6.3.4 Disassembling of TSI Mass Flow Sensor .........................................................6-9 6.3.5 Disassembling of Differential Pressure Transmitter .................................... 6-10 6.3.6 Disassembling of Diaphragm and Scale Board ............................................ 6-12 6.3.7 Disassembling of Core Board and Main Control Board ............................... 6-13 6.3.8 Disassembling of Components in the Display ............................................. 6-14 6.3.9 Disassembling of Electric Panel Assembly ................................................... 6-16 6.3.10 Disassembling of One-way Diaphragm ........................................................ 6-17 6.3.11 Disassembling of Fan Filter Cotton .............................................................. 6-18 6.3.12 Disassembling of Filter Element of Gas Inlet ............................................... 6-19 6.3.13 Disassembling of Filter (Part No.:130003930)............................................. 6-20

6.4 Functional Testing ............................................................................................... 6-21

7 User Maintenance ....................................................................... 7-1

7.1 Cleaning and Disinfection .......................................................................................7-2 7.1.1 Cleaning and Disinfecting Agents/ Autoclaving ..............................................7-3 7.1.2 Cleaning and Disinfection Methods ...............................................................7-3 7.1.3 Cleaning and Disinfection of Components .....................................................7-4

7.2 Regular Maintenance .............................................................................................7-6 7.2.1 Maintenance Principles ..................................................................................7-6 7.2.2 User Maintenance ..........................................................................................7-6 7.2.3 Service Life of Product/Accessories ................................................................7-7

7.3 Maintenance in Operation and Transportation ......................................................7-8 7.3.1 Transportation ................................................................................................7-8 7.3.2 Storage ...........................................................................................................7-9

7.4 Fuse Replacement ..................................................................................................7-9 7.5 Battery Maintenance ........................................................................................... 7-10

7.5.1 Battery Specification ................................................................................... 7-10 7.5.2 Precautions .................................................................................................. 7-11 7.5.3 Battery Storage ............................................................................................ 7-11 7.5.4 Battery Replacement ................................................................................... 7-11 7.5.5 Battery Charging and Calibration ................................................................ 7-12

7.6 Oxygen Sensor ..................................................................................................... 7-12 7.6.1 Oxygen Sensor Replacement ....................................................................... 7-12 7.6.2 Oxygen Sensor Calibration .......................................................................... 7-13 7.6.3 Technical Specifications of Oxygen Sensor .................................................. 7-13

Page 3: V8600 Critical Care Ventilator Service Manual

iii

7.6.4 Oxygen Sensor Maintenance ....................................................................... 7-14 7.7 Paramagnetic Oxygen Sensor (optional) ............................................................. 7-14

7.7.1 Paramagnetic Oxygen Sensor Calibration ................................................... 7-15 7.7.2 Technical Specifications of Paramagnetic Oxygen Sensor ........................... 7-15

7.8 Disposal ............................................................................................................... 7-16 7.9 Manufacturing techniques and materials: .......................................................... 7-16 7.10 Free Obligations: ................................................................................................. 7-17 7.11 Security, reliability and operating condition: ...................................................... 7-17 7.12 Return .................................................................................................................. 7-17

8 Pre-Use Test ................................................................................ 8-1

8.1 Pre-use test procedure ...........................................................................................8-1 8.2 Pre-use test failure analysis ....................................................................................8-2

9 Network ...................................................................................... 9-1

9.1 Overview ................................................................................................................9-1 9.2 What is Exported? ..................................................................................................9-1 9.3 Establishing a Connection ......................................................................................9-1 9.4 Interface Instructions .............................................................................................9-2 9.5 Transmits and Receives Data ..................................................................................9-4 9.6 Troubleshooting .....................................................................................................9-5

10 Service Menu ....................................................................... 10-1

10.1 Calibration ........................................................................................................... 10-2 10.1.1 Calibration operation guidance ................................................................... 10-2 10.1.2 Calibration failure analysis .......................................................................... 10-8

10.2 Event/Alarm Log .................................................................................................. 10-9 10.3 Machine Information ........................................................................................ 10-10 10.4 Language ........................................................................................................... 10-10 10.5 Test Page ............................................................................................................ 10-10

10.5.1 Demo ......................................................................................................... 10-11 10.5.2 Vlt. Monitor ............................................................................................... 10-11 10.5.3 Schematic .................................................................................................. 10-12 10.5.4 Cali. Data ................................................................................................... 10-13 10.5.5 Service Timer ............................................................................................. 10-13 10.5.6 Error Code ................................................................................................. 10-14 10.5.7 PT100 Cali. ................................................................................................. 10-14

10.6 Update ............................................................................................................... 10-15 10.7 Optional ............................................................................................................. 10-16

11 Troubleshooting ................................................................... 11-1

11.1 Technical Error ..................................................................................................... 11-1 11.1.1 Technical Test Error Code ............................................................................ 11-1 11.1.2 Other Errors ................................................................................................. 11-4

Page 4: V8600 Critical Care Ventilator Service Manual

iv

11.1.3 Service Tools ................................................................................................ 11-4

12 Appendix A Contact & Ordering Information ....................... 12-1

13 Appendix B Diagrams and Schematics .................................. 13-1

14 Appendix C Specifications .................................................... 14-1

15 Appendix D Spare parts and configurations ......................... 15-1

Page 5: V8600 Critical Care Ventilator Service Manual

1 Safety Information

1 Safety Information

The operator of this ventilator must recognize their full responsibility for choosing appropriate ventilation settings to ensure proper ventilation and patient safety. The responsibility for the selection of the appropriate level of patient monitoring depends solely on the equipment operator.

All the monitoring information is for reference only; it should not be used as the sole basis for therapeutic or diagnostic decisions.

Whenever a patient is connected to the ventilator, constant attention by qualified medical personnel is required in order to provide immediate corrective action in case of a malfunction and alarm occurrence.

Terms

This manual uses three special indicators to convey information of a specific nature. They include:

WARNING: Identify conditions or practices that could result in serious adverse reactions or potential safety hazards.

CAUTION: Identify conditions or practices that could result in damage to the ventilator or other equipment

NOTE: Identify supplemental information to help you better understand how the ventilator works.

WARNING: Do not use the system until you have read and understood this manual

including:

All connections of the system All warnings and cautions Operation procedure of each and every component of the system Test procedure of each and every component of the system

WARNING: The VG70 Ventilator System is intended for use by authorized and

trained medical personnel only.

WARNING: The users must familiarize themselves with the operation and use of this

machine prior to first clinical use with a patient.

1-1

Page 6: V8600 Critical Care Ventilator Service Manual

1-2

WARNING: To ensure proper servicing and avoid the possibility of physical injury,

only qualified personnel should attempt to service or make authorized modifications to the ventilator.

WARNING: An authorized service engineer must first install the ventilator and run

Aeonmed’s installation procedure, which includes calibration of various system components, before you connect a patient to the ventilator.

WARNING: If a fault is detected in the VG70 ventilator so that its life support functions

are no longer assured: start ventilation using an independent ventilation device (resuscitation bag) without delay, if necessary with PEEP and/or increased inspiratory O2 concentration.

WARNING: Before activating any part of the ventilator, be sure to check the

equipment for proper operation and, if appropriate, run PUT (pre-use test) as described in this manual,

WARNING: The VG70 ventilator is not intended to be a comprehensive monitoring

device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment.

WARNING: Patients on life-support equipment must be appropriately monitored by

competent medical personnel and suitable monitoring devices at all times.

WARNING: An alternative source of ventilation, such as manual respiratory

equipment, should always be available when using the VG70 ventilator.

WARNING: Do not connect inspiratory or expiratory circuits to the exhaust port.

WARNING: Ensure that inspiratory and expiratory circuits are connected to the

correct port before operation of equipment.

WARNING: The expiratory gas pathway may become contaminated with body fluids

or expired gases during normal use, and the inspiratory gas pathway may become contaminated during fault condition, such as occlusion, breath hoses disconnection.

Page 7: V8600 Critical Care Ventilator Service Manual

1 Safety Information

WARNING: Disposable breathing hoses shall not be reused. Reuse of the single use

hoses can cause cross infection.

WARNING: Assure that hoses used have the appropriate resistance and compliance

to ensure proper therapy.

WARNING: Do not disconnect the cable between the Main Control Unit and the GUI

screen while Ventilator is operating.

WARNING: The ventilator must not be connected to any anti-static or electrically

conductive hoses, tubing or conduit

WARNING: Adding attachments or other components or sub-assemblies to the

ventilator breathing system can change the pressure gradient across the ventilator breathing system and that such changes to the ventilation breathing system can affect the ventilator performance.

WARNING: Expiratory module is heated; use caution to avoid burns.

WARNING: Use caution when handling flammable or fragile components.

WARNING: Do not place containers of liquids (such as humidifier water reservoirs) on

top of or above ventilator. Liquids getting into the ventilator can cause equipment malfunction with the risk of patient injury.

WARNING: To avoid risk of electric shock, this equipment must only be connected to

a supply main with protective earth.

CAUTION: The breathing circuit must not be installed whenever the VG70 powers up

and whenever a pre-use test is performed.

CAUTION: If the system test fails, do not use the system. Attempt to troubleshoot and

fix the failure. If you are unable to fix the device, ask an authorized service representative to repair the device.

1-3

Page 8: V8600 Critical Care Ventilator Service Manual

1-4

CAUTION: Check the ventilator periodically as outlined in this manual; do not use if

defective. Immediately replace parts that are broken, missing, obviously worn, distorted, or contaminated.

CAUTION: Do not put ventilator into service until the patient setup is complete.

CAUTION: Measurements can be affected by mobile and RF communications

equipment.

CAUTION: Do not use oxygen hoses that are worn, frayed, or contaminated by

combustible materials such as grease or oils. Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.

CAUTION: Follow your hospital infection control guidelines for handling infectious

material. Aeonmed recognizes that cleaning, sterilization, sanitation, and disinfection practices vary widely among health care institutions. It is not possible for Aeonmed to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of cleaning, sterilization, and other practices carried out in the patient care setting.

CAUTION: Equipment not suitable for use in the presence of a Flammable Anesthetic

mixture with Air or with Oxygen or Nitrous Oxide.

CAUTION: To avoid an electrical shock hazard while servicing the ventilator, be sure

to remove all power to the ventilator by disconnecting the power source and turning off all ventilator power switches.

CAUTION: To avoid a fire hazard, keep matches, lighted cigarettes, and all other

sources of ignition (e.g., flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses.

CAUTION: In case of fire or a burning odor, immediately disconnect the ventilator

from the oxygen supply, facility power and backup power source.

CAUTION: During operation, do not block: Speaker Holes, Exhaust Port, Air Inlet or

Cooling Fan.

Page 9: V8600 Critical Care Ventilator Service Manual

1 Safety Information

CAUTION: Do not use the VG70 Ventilator in a MRI environment.

CAUTION: The ventilator shall not be used in a hyperbaric chamber.

CAUTION: The ventilator shall not be used with helium or mixtures with helium.

CAUTION: Tip over hazard; use care when moving ventilator mounted to cart as

device could tip over leading to injury or damage of equipment.

CAUTION: Do not use sharp objects to make selections on the LCD touch screen or

panel.

CAUTION: Do not connect a VGA or USB interface while the system is in service

CAUTION: The Network interface connection is for authorized service only.

CAUTION: Batteries should be removed if equipment will not be in service for more

than 6 months.

CAUTION: Do not immerse the oxygen sensor or the connector in any type of liquid.

CAUTION: When ventilator is exposed to conditions outside the specified operating

environment, allow 24 hours in normal environment before using.

CAUTION: Storage environment: -20℃~+60℃ and ≤95%RH.

CAUTION: Operating environment: 5℃~40℃ and 5%RH~95%RH.

CAUTION: Do not connect items that are not specified as part of the system.

CAUTION: The auxiliary outlet is only for the recommended humidifier; do not

connect to any other equipment or an additional multiple socket outlets.

CAUTION: When using a humidifier, user should frequently check the water trap and

look for water in the hose. If water is found in the hose, this water should be removed.

1-5

Page 10: V8600 Critical Care Ventilator Service Manual

1-6

Also, it is important the water trap is positioned in a way such that it is lower than the patient tubes.

CAUTION: Connecting electrical equipment to auxiliary outlet effectively leads to

creating a medical equipment system, and can result in a reduced level of safety, make sure the ME SYSTEM comply with requirements of IEC 60601-1:2005. The user who connects is responsible for the standard for the requirements applicable to the medical equipment system.

CAUTION: USA Federal law restricts this device for sale by or on the order of a

physician.

NOTE: The user of this product shall have sole responsibility for any ventilator

malfunction due to operation or maintenance performed by anyone not trained by Aeonmed.

NOTE: Usage of a filter on the expiratory side will increase the resistance of the

patient circuit.

NOTE: In non-invasive (NIV) ventilation, the exhaled volume of the patient can differ

from the measured exhaled volume due to leaks around the mask

NOTE: Do not sterilize or immerse the Mainstream CO2 Adapter in any fluids.

NOTE: All parts of the ventilator system are suitable for use within the patient

environment.

NOTE: All gas volume, flow, and leakage specifications in this manual are expressed

at STPD (standard temperature and pressure dry), except when specified with another condition.

Page 11: V8600 Critical Care Ventilator Service Manual

2 Notices

2 Notices

2.1 Warranty Notice

Do not make any service repairs on this equipment during the states warranty period. Any unauthorized work immediately voids the warranty. Aeonmed will not be liable for any repairs attempted by the owner. Any such attempted repairs other than specified no warranty repairs void the warranty.

2.2 Copyright Notice

Copyright Notice: Aeonmed.

This work is protected the copyright law and is the sole property of the Company. No part of this document permitted to be copied or otherwise reproduced, translated into other languages, or stored in any electronic information retrieval system, without the prior written consent of the Company.

2.3 EMC Notice

This equipment can radiate radio frequency energy ,as well can be interfered by electromagnetic If not installed and used in accordance with the instructions mentioned in the manual, electromagnetic interference may result. The equipment has been tested and found to comply with the limits set forth in EN60601-1-2 for Medical Products. These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments described in this manual. The ventilator has been tested to conform to the following specifications:

MIL-STD-461D: 1993, MIL-STD-462D:1993, EN 55011:2007+A2:2007, IEC 1000-4-2: 1994, IEC 1000-4-3:1994

IEC 1000-4-4:1994, IEC 1000-4-5:1994, QUASI-STATIC:1993

This ventilator is also designed and manufactured to comply with the safety requirements of IEC 60601-1, IEC 60601-2-12, CAN/CSA C22.2 No. 601.1-M90, and UL 60601-1.

2.4 MRI Notice

This equipment contains electromagnetic components whose operation can be affected by intense electromagnetic fields. Do not operate the ventilator in a MRI environment or in

2-1

Page 12: V8600 Critical Care Ventilator Service Manual

2-2

the vicinity of high-frequency surgical diathermy equipment, defibrillators, or short-wave therapy equipment. Electromagnetic interference could disrupt the operation of the ventilator.

2.5 Intended Use Notice

The VG70 Critical Care Ventilator is intended for patients ranging from pediatric to adult, and for use in a wide variety of clinical conditions. Specifically, the VG70 Critical Care Ventilator is applicable for adult and pediatric patients weighing at least 5 kg for adult and 3 Kg for child (7lbs.), who require the following types of ventilatory support: Positive Pressure Ventilation, delivered invasively (by ET or Tracheotomy tube) or non-invasively (by mask) via Assist/Control, SIMV, CPAP and other modes of ventilation.

The VG70 Critical Care Ventilator is intended for use in hospital and hospital-type facilities. It may be used during intra-hospital transport provided that electrical power and compressed gas are supplied.

WARNING: The VG70 Ventilator is not designed for use in an MRI environment. Do

not use the VG70 Ventilator near an MRI machine or injury or equipment damage could result.

WARNING: The user of the Ventilator must be professional and trained.

2.6 IEC Classification

Type of Equipment: Medical Equipment, Class I, Type B.

Adult/Pediatric Lung Ventilator

2.7 Declaration of Conformity Notice

This medical equipment complies with the Medical Device Directive, 93/42/EEC, and the following Technical Standards, to which Conformity is declared:

IEC 60601-1: The device classification is: Class I, Type B applied part (ventilator breathing circuit and mask), type BF applied part (CO2 module), ordinary enclosed equipment without protection against ingress of liquids, continuous operation

ISO 80601-2-12

IEC 60601-1

Page 13: V8600 Critical Care Ventilator Service Manual

3 Equipment Symbols

3 Equipment Symbols

Warnings and Cautions indicate all the possible dangers in case of violation of the

stipulations in this manual. Refer to and follow them.

Warnings Indicate potential hazards to operators or patients.

Cautions Indicate potential damage to equipment.

Instead of illustrations, symbols may be utilized. Not all of these symbols may necessarily appear on the equipment or in this User manual. The symbols include:

On (Power)

Protection Class Type B

Off (Power)

Protection Class Type BF

Follow operating instructions

Warning & Caution

Protective earth ground

Dangerous voltage

EQUIPOTENTIAL connection

loudspeaker

Lock

Manufacturer

Unlock

Date of production

Inspiratory hold

Serial Number

3-1

Page 14: V8600 Critical Care Ventilator Service Manual

3-2

Nebulization

Expiratory hold

Intelligent increase of oxygen

Manual inspiration

Standby

Waveform freeze

AC power int. Internal Battery

USB device

Refer to documentation

Prompt message

Already online

Flow trigger

Pressure trigger

Adult

Manual trigger

NIV modes

Child

Main Menu

Invasive modes

neonate

Alarm Silence Key

Do not reuse

Disposal of Waste

The system, with this label under the stipulations in the operating manual, complies with the requirements related from 93/42/EEC. 0123 is the certificate number to certify authorizations

Page 15: V8600 Critical Care Ventilator Service Manual

4 Introduction

4 Introduction

4.1 General description

The VG70 Critical Care Ventilator System consists of two required components: a Main Control Unit and a Graphical User Interface (GUI). Optional Components available for the VG70 Critical Care Ventilator system are: Ventilator Cart, Gas Cylinder Transport Cart, Patient Circuit Positioning Arm, and Hose Assemblies for O2, CO2 monitoring subsystem, etc.

1 Main Control Unit 2 Humidifier 3 Patient Circuit

4 Water Trap 5 Nebulizer Tube 6 Y-piece

7 Test Lung 8 Mask 9 Connector

10 Nebulizer Connector 11 Patient Circuit Positioning Arm

4-1

Page 16: V8600 Critical Care Ventilator Service Manual

4-2

4.2 Abbreviations and Definitions

(S) Means Set Value

(M) Means Measured Value

CPAP Continuous Positive Airway Pressure (Set value, hereinafter “S”)

F Breath rate (frequency) in bpm, i.e. ventilation times per minute (S)

fspont Patient’s spontaneous respiratory frequency (Measured value,

hereinafter “M”)

ftotal Total breath rate, i.e. the sum of breath rate f and spontaneous breath

rate fspont(M)

O2 Inspiratory O2 concentration (S & M)

I : E The ratio of Inspiration to Expiration (M)

MV Expiratory minute volume (M)

MVspont Spontaneously breathed minute volume(M)

MVleak Leakage minute volume (M)

Paw Patient airway pressure (M)

PEEP Positive End-Expiratory Pressure, which can improve the patient’s

oxygenation (S & M)

PEEPi Intrinsic Positive End-Expiratory Pressure (M)

Pinsp Upper pressure level in PCV mode (S)

Pmean Mean airway pressure. This value is updated at the end of the last

respiratory cycle, hence, is a continuous average (M)

Ppeak Airway pressure peak value during one ventilatory cycle (M)

Pplat End-inspiratory airway pressure (M)

Pmin Minimum airway pressure (M)

Psens Pressure sensitivity (S)

Page 17: V8600 Critical Care Ventilator Service Manual

4 Introduction

Psupp Pressure support (S)

Phigh Upper pressure level in BIVENT and APRV (S)

Plow Lower pressure level in BIVENT and APRV (S)

T Imax Maximum inspiratory time (S)

Tinsp Inspiratory Time (S)

T pause Inspiratory Pause Time, to increase the inspiratory time to improve the

patient’s oxygenation (S)

Trigger by flow rate (S)

VT Tidal volume of mechanical ventilation (S)

VTE Expiratory tidal volume (M)

VTI Inspiratory tidal volume (M)

Esens Expiratory trigger sensitivity (S)

ETCO2 End-expiratory CO2 concentration (M)

WOB Work of breathing (M)

R*C Time constant (M)

Leak% Leakage percentage (M)

Cdyn Dynamic compliance (M)

Cstatic Static compliance (M)

Rinsp Inspiratoryresistance (M)

Rexp Expiratory resistance (M)

Elastic Elastic resistance (M)

IP21 Solid particle protection level 2; Liquid ingress protection level 1

4-3

Page 18: V8600 Critical Care Ventilator Service Manual
Page 19: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5 Theory of Operation

5.1 Pneumatic System

5.1.1 Pneumatic Theory

5-1

Page 20: V8600 Critical Care Ventilator Service Manual

5-2

Figure 5-1 Pneumatic Theory

Page 21: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.1.2 Main Pneumatic Assemblies Overview

Figure 5-2 Main Airway Assemblies sketch view

1 Turbine driver 2 Expiratory valve module 3 Gas circuit module 4 Electrical box

1

23

4

5-3

Page 22: V8600 Critical Care Ventilator Service Manual

5-4

Pneumatic Connecting Diagram

Figure 5-3 Pneumatic Connecting Diagram

PU tube φ6 Silica gel tube φ2x1.5

Silica gel tube φ1x1

1 Oxygen inlet module 2 Oxygen supply module

3 Inferoanterior shell components 4 Differential pressure transmitter

5 MMB 6 Safety inspiration module

7 Exhalation valve components

4

7

Page 23: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

Gas Circuit Module

Figure 5-4 Gas circuit module

1 Differential pressure transmitter

2 Interface board 3 Inspiratory flow sensor

4 Pressure sensor board 5 Fan

1

3

2

4 5

5-5

Page 24: V8600 Critical Care Ventilator Service Manual

5-6

Oxygen Inlet Module Overview

Figure 5-5 Supply Oxygen Module

item Description

Gas flow direction

1 Oxygen connector, connecting with oxygen supplies.

2 Low flow oxygen connector, connecting with low flow oxygen supplies.

3 Inlet filter, filtering impurities from the oxygen.

4 Pressure sensor, monitoring the pressure of high-pressure oxygen gas.

5 one-way valve

6 Regulator valve, regulating gas pressure to 0.28MPa

Page 25: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

7 Connector, connecting with oxygen mixer module, supply low flow oxygen

8 Connector, connecting with oxygen mixer module, supply high pressure oxygen

9 Inlet filter

Supply Oxygen Module Overview

Figure 5-6 Oxygen Mixer Module

Item Description

Gas flow direction

1 Connector, connecting with connector 7(low flow oxygen) in Figure 5-5 Supply

Oxygen Module

2 Connector, connecting with Turbine and Inspiratory Control Module

3 Solenoid valve V1-V4, Flow 30L/min

4 Connector, connecting with connector 8(high pressure oxygen) in Figure 5-5

Supply Oxygen Module

5 Connector, connecting with nebulizer pipe

6 Solenoid Valve V9, controlling whether affording the oxygen to nebulizer

5-7

Page 26: V8600 Critical Care Ventilator Service Manual

5-8

7 Solenoid valve V7-V8, Flow 4L/min

8 Solenoid valve V5-V6, Flow 12L/min

Turbine and Inspiratory Control Module Overview

Figure 5-7 Turbine and Inspiratory Control Module

Page 27: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

Item Description

Gas flow direction

1 Inspiratory Valve, control of inspiratory flow and pressure

2 Connector, connecting with the pressure sensor of mother board

3 Air intake, including air filter

4 Mixing chamber, mixing oxygen and air

5 Muffler

6 Turbine

7 Radiator

8 Flow sensor, monitoring the volume of the gas

9 Muffler

10 Connector, connecting with safe inspiratory module

Safety Inspiratory Module Overview

5-9

Page 28: V8600 Critical Care Ventilator Service Manual

5-10

Figure 5-8 Safety Inspiratory Module

Item Description

Gas flow direction

1 Inspiratory connector, connecting with silica gel tube for patient

2 Oxygen sensor, monitoring the oxygen density of mixed gas

3 Relief valve

4 Emergency air inlet

5 Connector, connecting with flow sensor

6 One-way valve

7 Connector, connecting with the pressure sensor of mother board

Page 29: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

Expiratory Module Overview

Figure 5-9

Figure 5-10

Item Description

Movement direction

1 Connector, connecting with pressure sensor of mother board.

2 Driving valve

3 Press the lock nut of the expiration valve, unlock, then pull out the component.

4 Water trap

12

3

4

5-11

Page 30: V8600 Critical Care Ventilator Service Manual

5-12

5.2 Electrical System

5.2.1 Electrical Box

1

2

1

3

4

5

6

7

Figure 5-11

1 Core board 2 Main control board 3 Battery box

4 Battery 5 Power supply board 6 Mother board

7 Switch power

Page 31: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.2.2 Expiratory Valve Heating Module

Figure 5-12

1 Temperature sensor 2 Heating resistor

2

1

5-13

Page 32: V8600 Critical Care Ventilator Service Manual

5-14

5.2.3 Rear Panel View

Figure 5-13

1 CO2 module connector 2 Nurse call connector 3 DC power fuse

4 Air inlet 5 Oxygen inlet 6 Fan

7 Equipotential terminal 8 DC power input port 9 Power switch

10 AC power input socket 11 Battery cover

Page 33: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.2.4 Display Inside View

Figure 5-14

1 Indicator light board 2 Speaker 3 Mute key board

4 Encoder board 5 Display cable 6 Display platform circuit board

7 Display connector 8 VGA connector 9 USB connector

10 NET connector 11 Touch screen controller

12 Alarm LED board

5-15

Page 34: V8600 Critical Care Ventilator Service Manual

5-16

5.2.5 ARM platform circuit board

Figure 5-15

1 Power supply indicate lamp board connector

2 Backlight cable connector

3 Mute button board connector

4 Encoder board connector

5 Display cable connector

6 Interface board connector

7 Touch screen control board connector

8 Alarm light board connector

9 LVDS cable connector

10 Speaker cable connector

Page 35: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.2.6 Display interface board

Figure 5-16

1 CN1 interface (Connect Display platform circuit board)

2 VGA interface

3 USB 2 interface

4 USB 1 interface

5 NET interface

5-17

Page 36: V8600 Critical Care Ventilator Service Manual

5-18

5.2.7 Interface Board

Figure 5-17

1 Signal

2 Motor controller

3 Insp. valve

4 Turbine

5 Fan

6 Turbo tempeture sensor

7 PEEP valve

8 Expe. valve heating

9 Expe. flow sensor board

10 O2 sensor

11 Expe. valve tempeture sensor

12 TSI sensor

13 Power in

Page 37: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.2.8 Mother Board

Figure 5-18 Front of mother board

Figure 5-19 Back of mother board

1 Core board slot

2 Reserved

3 Analogy slot

4 Lithium-ion battery 1 connector

5 Lithium-ion battery 2 connector

5-19

Page 38: V8600 Critical Care Ventilator Service Manual

5-20

6 PSB slot

7 Main switch connector

8 24VDC input connector

9 DC power input connector

10 Display connector

11 Reserved

12 Reserved

13 Motor controller connector

14 Reserved

15 O2 supply pressure sensor

16 Paramagnetic oxygen sensor connector (optional)

17 Interface board signal line connector

18, 19 Solenoids cable connector

20 Interface board power line connector

Page 39: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.2.9 Core Board

Figure 5-20

1 Mother board connecting connector

2 DSP_JTAG

5-21

Page 40: V8600 Critical Care Ventilator Service Manual

5-22

5.2.10 Main Control Board

Figure 5-21

1 Mother board connecting connector

2 CO2 module connector

3 Nurse call connector

Page 41: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.2.11 Power Supply Board

Figure 5-22

1 Mother board connecting connector

3

5-23

Page 42: V8600 Critical Care Ventilator Service Manual

5-24

5.2.12 System Wiring Diagram

BOM size and diagram:

No. BOM size Name Diagram

1 122007322 Fan cable

2 122007320 Interface board signal cable

3 122007321 PT100 signal cable

4 122008090 Expiration valve heating component

5 122008011 Magnetic base component

Page 43: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

No. BOM size Name Diagram

6 122007331 Oxygen supply pressure cable

7 122007327 Expiration flow sensor cable

8 122007329 TSI sensor cable (analog)

9 122007330 TSI sensor cable (digital)

10 122007328 Oxygen sensor signal cable

5-25

Page 44: V8600 Critical Care Ventilator Service Manual

5-26

No. BOM size Name Diagram

11 122007315 Valve terminal drive cable (30L)

12 122007316 Valve terminal drive cable (12L)

13 122007317 Valve terminal drive cable(4L)

14 122007319 Interface board power cable

15 122007478 Turbine driver component

Page 45: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

No. BOM size Name Diagram

16 122007336 Switch power output cable

17 122012312 W0056 cable

18 122007486 Power lamp component

19 122007484 Speaker component

20 122007483 Mute button component

21 122012310 W0054 cable

5-27

Page 46: V8600 Critical Care Ventilator Service Manual

5-28

No. BOM size Name Diagram

22 122007455 Switch mounting plate component

23 122007463 Inspiration control valve

24 122007482 Encoder component

25 122007485 Alarm lamp component

26 122007503 Copper woven mesh cable

Page 47: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

No. BOM size Name Diagram

27 122012313 W0057 cable

28 122012311 W0055 cable

29 122009449 W0001 cable (optional)

30 122001124 W1 cable

5-29

Page 48: V8600 Critical Care Ventilator Service Manual

5-30

5.3 Software System

5.3.1 Software Units and Their Relationship

User Interface Software Unit bears the responsibility of user interface of the whole software system. It includes operation interface for the users, graphic interface for showing information of the device, communication with other units, etc.

Breathing Delivery Software Unit bears the responsibility of ventilation control.

Power Supply Software Unit bears the responsibility of power management.

The relationship between these three units is as following:

Figure 5-23 Software Units of NBP System

In above figure, the orientation of arrow points to information is flowing to where. Furthermore, a bidirectional arrow means communication between SW units is bi- directional.

User Interface SW Unit

Breathing Delivery SW Unit Power Supply SW Unit

Page 49: V8600 Critical Care Ventilator Service Manual

5 Theory of Operation

5.3.2 Structure of the Software System

Figure 5-24 Structure of Software Units

5-31

Page 50: V8600 Critical Care Ventilator Service Manual

5-32

5.3.3 Hardware and Software Interface

Figure 5-25 Interface between Software and Hardware System

Page 51: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6 Disassemble and Functional Testing

6.1 General

The ventilator is packaged with two cartons. One carton filled with cart. Another carton filled with ventilator main units, display, and accessories crate.

Accessories crate including standard accessories and optional accessories. Standard accessories are divided into European standard and American standard, the optional accessories is changed based on order. Details please refer to Appendix D configuration List.

6.2 Main Assemblies Disassemble

6.2.1 Disassembling of the Display from the Host

Figure 6-1

To separate the display from the host: pull out the display (1) along the guide (2).

6-1

Page 52: V8600 Critical Care Ventilator Service Manual

6-2

6.2.2 Disassembling of the Host from the Trolley

Figure 6-2

To separate the host from the trolley

a. Remove the screw (2);

b. Remove the host (1) along “ ” direction;

c. Remove the host installing piece (3) from the tray connector (4).

d. Lift up the host (1);

NOTE: When installing the host to the trolley, please align the host installing piece

with the tray connector.

Page 53: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.2.3 Disassembling of the Arm Assembly from the Host

Figure 6-3

To separate the arm assembly from the Main Control Unit

a. Pull out the knob plunger(1);

b. Remove the arm assembly (3) from the handrail (2).

6-3

Page 54: V8600 Critical Care Ventilator Service Manual

6-4

6.2.4 Disassembling of the Humidification Assembly from

the Host

Figure 6-4

To separate the humidification assembly from the host: remove the humidification

assembly (1) from the guide-rail (2) along “ ” direction.

Page 55: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.3 Disassembling of Maintenance Spare Parts

6.3.1 Disassembling of Oxygen Sensor

Figure 6-5

To separate the oxygen sensor from the host:

a. Pull out oxygen sensor cover(1);

b. Remove the oxygen sensor(2).

6-5

Page 56: V8600 Critical Care Ventilator Service Manual

6-6

6.3.2 Disassembling of Pressure Sensor Board

Figure 6-6

To remove the pressure sensor board from the host:

a. Remove the six screws (2) (Cross recessed Pan head screw M4x8);

b. Move the side panel(1) to a side;

c. Disconnect the cable connector of the pressure sensor;

d. Remove the two screws (3) (Cross recessed Pan head screw M3x6);

e. Remove the pressure sensor board (4).

Test: When you replaced the oxygen pressure sensor, turn on the machine to check if the oxygen pressure value is back to normal.

Page 57: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.3.3 Disassembling of Supply O2 Assembly

1

2

Figure 6-7

a. Remove the screws (Hexagon socket head cap screws M4×16) of the upper cover, and remove the upper cover (1);

b. Remove the inlet link block of the oxygen inlet module, and remove the filter cover of the air inlet.

c. Remove the screws (Cross recessed Pan head screw M3x6) of the lower rear cover, and remove the lower rear cover (2);

1

Figure 6-8

d. Remove the screws of the electrical box, pull out the three connectors(2, 3 4) of supply O2 module from the mother board of the electrical box (5), and then remove the electrical box (1);

6-7

Page 58: V8600 Critical Care Ventilator Service Manual

6-8

1

2

1

Figure 6-9

e. Remove the four screws (1) (Cross recessed Pan head screw M3x25); f. Remove supply O2 module (2).

Page 59: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.3.4 Disassembling of TSI Mass Flow Sensor

Figure 6-10

To remove TSI mass flow sensor from the host:

a. Remove the six screws (2) (Cross recessed Pan head screw M4x8);

b. Move the right cover to a side(1);

c. Disconnect the two cables of TSI mass flow sensor;

d. Drag out the TSI mass flow sensor along “ ” direction (3).

Test: When the TSI mass flow sensor is replaced, the inspiratory valve must be calibrated. Please refer to Chapter 10.1.

6-9

Page 60: V8600 Critical Care Ventilator Service Manual

6-10

6.3.5 Disassembling of Differential Pressure Transmitter

Figure 6-11

To remove the differential pressure transmitter from the host:

a. Remove the handle (1), and then remove the host from Cart assembly;

b. Remove screws, and then remove the two side covers (2);

c. Remove Head cover components (3);

1

2

3

Page 61: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

Figure 6-12

d. Remove the Electrical box components backwards to the right place (1);

e. Remove the differential pressure transmitter (2).

Test: When the differential pressure transmitter is replaced, the inspiratory valve must be calibrated. Please refer to Chapter 8 and Chapter 10.1.

1

2

6-11

Page 62: V8600 Critical Care Ventilator Service Manual

6-12

6.3.6 Disassembling of Diaphragm and Scale Board

Figure 6-13 Figure 6-14

To remove the diaphragm and the scale board from expiratory module assembly:

a. Press the button(1 in Figure 6-13);

b. Pull out the expiratory valve core component (2 in Figure 6-13);

c. Remove the end cover (1 in Figure 6-14) and the gasket (2 in Figure 6-14);

d. Remove the scale board (3 in Figure 6-14) and the diaphragm (4 in Figure 6-14).

Test: When the diaphragm and scale board are replaced, the inspiratory valve must be calibrated. Please refer to the Chapter 10.1; Also perform a pre-use test and verify all test pass Refer to Chapter 8.

Page 63: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.3.7 Disassembling of Core Board and Main Control

Board

Figure 6-15

Remove of the core board and control board:

a. Remove the four screws(1) (Cross recessed Pan head screws M3×6);

b. Remove the electric board cover (2);

6-13

Page 64: V8600 Critical Care Ventilator Service Manual

6-14

c. Pull out the core board (3);

d. Pull out the main control board (4).

Test: when you replaced the core board and main control board. Cut off the battery, connect the alternating current, and check if the machine operates normally.

6.3.8 Disassembling of Components in the Display

a. Remove the four screws (1) (Cross recessed countersunk head screw M4×8), take off the rear shell hanger and hanging block interface board (Figure 6-16);

Figure 6-16

b. Pry the back-shell of display open with the tool. You can use the tool to pry the black circle signed places (Figure 6-17).

Figure 6-17

Page 65: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

c. Open the back-shell slowly and remove the back-shell.

d. Remove the six screws (2) (Cross recessed Pan head screws M3×6), and then remove the shielding mask (1) (Figure 6-18);

Figure 6-18

e. The components in the display can be easy to be taken away.

6-15

Page 66: V8600 Critical Care Ventilator Service Manual

6-16

6.3.9 Disassembling of Electric Panel Assembly

Figure 6-19

a. Remove the four screws (1) (Cross recessed Pan head screws M3×6);

b. Remove the electric panel cover (2);

c. Remove the power supply board component (3).

Test: When any circuit board has been removed and or replaced, you must perform ad calibration on all sensors and perform a pre-use check prior to use. Please refer to Chapter 8 and Chapter 10.1.

Page 67: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.3.10 Disassembling of One-way Diaphragm

Figure 6-20

To remove the one-way diaphragm:

a) Remove the expiratory valve core component;

b) Remove the one-way valve core component (1);

c) Remove one-way diaphragm(2);

Test: When the one-way diaphragm has been replaced, the inspiratory valve must be calibrated. Please refer to chapter 8 and chapter 10.1.

6-17

Page 68: V8600 Critical Care Ventilator Service Manual

6-18

6.3.11 Disassembling of Fan Filter Cotton

Figure 6-21

To remove the fan filter cotton:

a. Remove the fan filter cotton cover(2);

b. Remove the fan filter cotton(1).

Page 69: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.3.12 Disassembling of Filter Element of Gas Inlet

Figure 6-22

To remove filter element of gas inlet:

a. Remove the filter cover(1);

b. Remove the filter support sleeve(2);

c. Remove the filter element of gas inlet(3).

6-19

Page 70: V8600 Critical Care Ventilator Service Manual

6-20

6.3.13 Disassembling of Filter (Part No.:130003930)

Figure 6-23

To remove the filter:

a. Remove the two screws (1) (Hexagon socket head cap screws M4×10);

b. Remove the oxygen inlet connector(2);

c. Remove the filter (3).

Page 71: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

6.4 Functional Testing

Note: Before the functional testing you need to set ventilator in standard states.

Ventilator standard working states:

Breathing patterns: A / C (VCV);

Inspiratory time: 1 seconds;

Breath-hold time: 0 seconds;

Respiratory rate: 20 beats/min;

Tidal volume: 400 ml;

The upper pressure limit: 80 cmH2O;

Pressure lower limit: 0 cmH2O;

Oxygen concentration: 21%;

Trigger sensitivity: -3 cmH2O;

PEEP: 3 cmH2O;

Gas source rated working pressure: 0.4 mpa.

Test items Test point Inspection standards

Alarm sound and alarm lamp test

Respectively to trigger the high, middle, low-level alarm, observe warning lights and alarm sounds

Advanced alarm: flashing red light, 10-pulse cycle play an alarm sound. Intermediate warning: flashing yellow light, 3-pulse cycle play an alarm sound. Low-level alarm: yellow light, a single alarm sound of the two pulse.

The mute button

Unplug piping appear sound the alarm, press the mute button;

Press the mute button after the sound the alarm disappears, the Screen 120s countdown;

6-21

Page 72: V8600 Critical Care Ventilator Service Manual

6-22

Tidal volume

In standard work status, breathing test device is connected to ventilator inspiratory port in VCV mode, after the output tidal volume is stable, record Vti monitoring values of breathing test equipment and ventilator. Tidal volume settings as the following table: Volume

(mL) Inspiratory time(s)

Frequency(bpm)

Child 20 0.2 60 Adult 500 2 10

Set the value of 20mL, the detection value is 10 ~ 30mL; Set the value of 500mL, the detection value should be 450 ~ 550mL;

PEEP

The standard working condition, connecting pipes and the simulated lung, in the loop access breath testing equipment, ventilator work, set end-expiratory pressure for 20cmH2O, other parameters default. Observed ventilator monitoring values, values and ventilator settings.

Set PEEP 20cmH2O, as follows: 18cmH2O ~ 22cmH2O;

Peak velocity

Perform the calibration of the flow sensor, disconnect the breath-side piping, and the recorded value of the maximum flow rate of the ventilator.

Inspiratory peak flow rate of 160 to 200 the LPM

Pressure to control levels of PCV

PCV mode, respiratory rate 10 bpm, the upper pressure limit set maximum PEEP was the value of 0 cmH2O with three links breath test equipment and ventilator connection, set the value of the pressure level in the order of 5 cmH2O, 30 cmH2O, 60 cmH2O, breathing machine at the platform when the pressure of the test equipment in the record show the value and set value.

Set 30 cmH2O, as follows: 27 to 33 cmH2O;

FiO2

Set the oxygen concentration in the standard working condition, followed by 21%, 60%, 100%, the value of the oxygen concentration stabilized, record the test equipment and the ventilator oxygen concentration detection value, the ventilator settings value.

Set the oxygen concentration of 21%: 18% ~ 24%; Set the oxygen concentration of 60%: 55% ~ 65%; Set for 100% of the oxygen concentration: 95% ~ 100% (which may appear on the arrow, shall be deemed adopted);

Exhalation valve

Boot more than 40 minutes to check exhalation valve roots of the great circle the

Can be heated

Page 73: V8600 Critical Care Ventilator Service Manual

6 Disassemble and Functional Testing

heating lower half of the heating temperature;

AC,

battery,

display

check

NOTE: These items relate to the external battery, the application of optional external battery, or NA.

Switched by the AC with standard and optional batteries;

Power indicator light display standard and optional the battery indicator, external power supply; two batteries with a charging, the charge indicator lights are bright;

Turned on Unplug the AC standard and optional batteries connection at the same time;

Power indication, the charge lamp is off, the screen shows the standard and optional batteries, battery powered;

Boot after 10 minutes with alone the standard battery, connect the AC;

Boot, standard battery can be charged, the display shows the standard battery charging; charging lights;

Boot after 10 minutes with alone the optional battery, connect the AC;

Boot, optional battery can be charged, the display shows the optional battery charging; charging lights;

6-23

Page 74: V8600 Critical Care Ventilator Service Manual
Page 75: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

7 User Maintenance

WARNING: Aeonmed recognizes that cleaning, disinfection, and sterilization

practices vary widely among medical institutions. It is not possible to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of cleaning, sterilization, and other practices carried out in the patient care setting.

WARNING: Use a cleaning and disinfection schedule that conforms to your

institution’s disinfection and risk-management policies.

• Refer to the material safety data as applicable.

• Refer to the operation and maintenance manuals of all disinfection equipment.

• Do not inhale fumes that may result from any disinfection process.

WARNING: Movable and removable parts may clamp or even crush your hand. Use

caution when moving or replacing system components.

CAUTION: The disposal of environmentally harmful devices (such as batteries and

LCD display) must be in accordance with local regulations.

WARNING: Use a cleaning and disinfection schedule that conforms to your

institution’s disinfection and risk-management policies.

• Refer to the material safety data as applicable.

• Refer to the operation and maintenance manuals of all disinfection equipment.

• Do not inhale fumes that may result from any disinfection process.

WARNING: Do not use talc, zinc stearate, calcium carbonate, corn starch or similar

material to prevent sticking of the bellows, as these materials may enter the patient’s lungs or airway, causing irritation or injury.

CAUTION: To prevent system damage:

• Refer to the literature supplied by the manufacturer of the cleaning agent.

7-1

Page 76: V8600 Critical Care Ventilator Service Manual

7-2

• Never use organic, halogenated or petroleum-based solvents, anesthetic, glass cleaning agents, acetone or other irritant agents.

• Never use abrasive agents (i.e. steel wool or silver polish) to clean components.

• Keep all liquids away from electronic components.

• Prevent liquid from entering the equipment.

• All cleaning solutions used must have a pH between 7.0 and 10.5.

CAUTION: Never immerse the oxygen sensor or its connector in any type of liquid.

• Dispose of the oxygen sensor per the manufacturer’s specification.

CAUTION: Do not wash the inner surface of the oxygen sensor.

CAUTION: Prior to use after cleaning or disinfecting, power up the system as

described in section 6 and follow the on-screen Pre-Use test prompts to perform the Leak Test and Circuit Compliance Test.

Discard one time use components after using. Don’t use hard brushes or other sharp tools in cleaning to avoid damage to parts.

• Clean components with warm water and light detergent.

• Dry the components after rinsing with water.

• Check each component when cleaning and replace damaged ones.

• After a component is replaced, test the ventilator first prior to putting the ventilator

into service.

CAUTION: Follow the instructions of detergent manufacturer. Concentrated detergent

may hurt some components. Detergent residue may cause spots and cracks, especially in high temperature disinfection.

7.1 Cleaning and Disinfection

CAUTION: Before the first use clean, disinfect, and sterilize the ventilator.

Disposable components must be disposed in accordance with local regulations. Don’t use hard brushes or other sharp tools in cleaning to avoid damage to parts

Page 77: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

7.1.1 Cleaning and Disinfecting Agents/ Autoclaving

Agent Classification

• Mild dishwashing detergent

• Soapy water with detergent ph between 7.0 and 10.5

• Isopropyl alcohol (70% solution)

• Window cleaning solution (with isopropyl alcohol and ammonia)

• Sodium hypochloride- (bleach) in water (10% solution)

• Hydrogen peroxide (3% solution)

Detergent

Detergent

Intermediate level disinfectant

Intermediate level disinfectant

Intermediate level disinfectant

Intermediate level disinfectant

• Gluteraldehyde 2% solution

• Steam autoclaving up to a maximum temperature of 134°C (273°F).

High level disinfectant

High level disinfectant

7.1.2 Cleaning and Disinfection Methods

Different parts of the ventilator have their respective cleaning and disinfection methods. The following categories are defined for the parts noted in Table 7-1. The parts need to be cleaned, disinfected and thoroughly dried before reassembly.

A: Wipe: If there is a potentially infectious substance on the breathing system, such as blood or secretion, wipe away the substance with disposable cloth using proper disinfectant. Use a soft cloth with a water-soluble detergent or disinfectant wipes.

B: Machine washing: Automatic washing with washer and disinfecting with disinfection machine.

C: Immersion disinfection: Soak in glutaraldehyde-based formulations of 2%.

D: High temperature and pressure disinfection: At 121℃ for 20 minutes minimum, or at 134℃ for 8 minutes minimum. Follow the manufacturer’s instructions for high level disinfection.

High-temperature disinfection does not have any cleaning effect. It should only be used on components that have already been cleaned by hand or machine and then thoroughly dried.

7-3

Page 78: V8600 Critical Care Ventilator Service Manual

7-4

Table 7-1

Part name Cleaning and disinfection methods

A B C D

Ventilator exterior, including housing, gas supply hoses and power cord

All components in Figure 7-3 √ √ √ √

CO2 sensor, other breathing circuit parts or accessories

Follow the manufacture’s guidelines

7.1.3 Cleaning and Disinfection of Components

7.1.3.1 External Surfaces

Using a soft cloth with a water-soluble detergent or disinfectant wipes, clean the housing, gas supply hoses and power cord.

7.1.3.2 Expiratory Module

(1) Disassembly

Figure 7-1 Figure 7-2

Page 79: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

Figure 7-3

To remove the components from expiratory module:

e. Press the button(1 in Figure 7-1);

f. Pull out the expiratory valve core component (2 in Figure 7-1, Figure 7-2);

g. Remove the end cover(1 in Figure 7-3) and the gasket (2 in Figure 7-3);

h. Remove the scale board(3 in Figure 7-3)and the diaphragm(4 in Figure 7-3);

i. Rotate the expiratory connector (5 in Figure 7-3) clockwise;

j. Pull out one-way valve core(6 in Figure 7-3), rubber o-sealing ring(7 in Figure 7-3) and one-way diaphragm(8 in Figure 7-3);

k. Rotate the water trap (9 in Figure 7-3), then pull it out. (2) Cleaning

a. Wash each component using a mild detergent and water solution.

b. Rinse with clean, hot water and allow to thoroughly dry.

(3) Disinfection

NOTE: Ensure that all the components have been cleaned before disinfecting.

Using the Gluteraldehyde disinfection solution, follow the manufacturer’s instructions for high level disinfection and rinsing of all components while adhering to facility policies and procedures.

7-5

Page 80: V8600 Critical Care Ventilator Service Manual

7-6

All the components can also be high temperature and pressure disinfected. Using an autoclave, follow the manufacturer’s instruction for high level disinfection of all the components while adhering to facility procedures.

(4) Assembly

Reassemble the components in the reverse order.

After installation, please perform a pre-use test and verify all tests passed.

7.2 Regular Maintenance

7.2.1 Maintenance Principles

Do not use a faulty machine. Ask an authorized agent of our company to carry out all necessary maintenance tasks. Test the machine after maintenance for normal operation. Every parameter should meet specifications.

In order to ensure the reliability of the machine, all maintenance and repair work should be carried out by an authorized agent of our company. In case such person is unavailable, a qualified person with similar maintenance experience is acceptable for the work.

CAUTION: Person with no sufficient experience is strictly prohibited to perform

maintenance.

Only use products provided by Aeonmed to replace damaged items. Verify and test all items ensuring they meet specifications.

Contact local service agent of our company in case support is needed. In all cases, maintenance fee is the current component price plus reasonable labor cost, except for those within warranty period.

7.2.2 User Maintenance

Minimum maintenance interval

Task

Daily Check liquid in expiratory module water trap (collected liquid volume cannot be more than half of the bottle)

Weekly Calibrate oxygen sensor

Page 81: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

Minimum maintenance interval

Task

1-3 month(s) Replace air filter

Every 6 months Charge and discharge the batteries once (Charge time: at least 3.5h).

Every year Calibrate flow and pressure sensors; Calibrate inspiratory valve and expiratory valve (if necessary)

Every year or after calibration Replace the O2 sensor(actual life depends on temperature and O2 concentration)

When cleaning and installing Check components and replace or repair when necessary.

WARNING: If the Ventilator will not be in use for a period of more than 6 months, the

internal batteries must be disconnected or removed to prevent possible damage to the equipment or risk to users or service personnel.

WARNING: The ventilator must not use, nor be connected to, any anti-static or

electrically conductive hoses and tubing.

7.2.3 Service Life of Product/Accessories

Service Life is defined as the time that the manufactured device can be expected to be 'serviceable' or supported by our company and the maximum time the device can be used safely. Some items within the device may require maintenance, repair or replacement within this time. Such items will be available from our company for the service life of the product. The calculation of the service life begins at the installation of the product at the customer site. Our company recommends for safe use that each device is replaced after its service life is completed.

CAUTION:The service life of the following items is based on normal operating

conditions.

7-7

Page 82: V8600 Critical Care Ventilator Service Manual

7-8

Mask, patient pipeline Sterilize for 20 times at 121℃

Power supply cable, gas supply pipe 8 years

Machine 8 years

Battery 500 cycles of full charge(The battery from exhausted to full )

7.3 Maintenance in Operation and Transportation

The location of machine should be proper so that it cannot obstruct or be disturbed by medical care personnel. Fix power supply cable well to avoid failure. Use caution not to touch accidental keys on the panel, which may make tidal volume setting wrong.

During transportation of the ventilator with or without a patient connected, make sure that the following conditions are fulfilled:

Gas cylinders are connected with a sufficient amount of gas, the Battery module is functioning. Follow the hospital guidelines.

Use the handles on the Mobile Cart. Transport the bed and the ventilator slowly, and watch the patient connection carefully to see that no pulling or other movement occurs.

Be careful not to tip the Mobile Cart when crossing an obstacle such as a doorstep.

7.3.1 Transportation

Use care when moving machine within hospital or clinical environment.

WARNING: If Control Unit of Ventilator is dropped or damaged during transportation,

equipment failure could result in patient injury.

WARNING: Tip over hazard – use care when moving Ventilator mounted to Cart as

device could tip over leading to injury or damage to the equipment and possible subsequent patient injury.

User can carry packaged machine while riding vehicle, plane and train. Impact, severe shock and damp should be avoided during transportation (other conditions should accord with purchase contract), with ambient temperature -20°C~+60°C and relative humidity not more than 95%. In case transportation conditions don’t agree, put the machine in specified operating environment at least 24h before using.

Page 83: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

7.3.2 Storage

CAUTION: Do not put ventilator into the shock environment.

CAUTION: Do not lay anything heavy on the top.

The machine should be stored in a room with temperature -20°C~+60°C and relative humidity –not more than 95% non-condensing, with ventilation and no corrosive gas.

CAUTION: The device should be stored in a room that is drafty where no corrosion

gas exists.

CAUTION: When the storage conditions are beyond the requirements of operational

environment, and the storage state is transferred into operation state, the product only can be used after being stored in environment for over 24 hours.

7.4 Fuse Replacement

WARNING:Before replacing fuse, first disconnect AC power. Otherwise, it may cause

injury or even death.

WARNING:When replacing fuse, make sure the new fuse is the same type and size

as the old one; otherwise, the ventilator may be damaged.

CAUTION: The fuse is a damageable part, and care must be taking when replacing it

so no damage occurs. 1. AC Circuit Fuse

Fuse replacement steps:

• Insert screwdriver into the trench (2) of end of fuse box, see Figure 7-4. • Pull out fuse holder (1). • Remove the fuse (3). • Load new fuse. • Push the new fuse into the original position gently. • Connect AC power, and then start the ventilator to test.

7-9

Page 84: V8600 Critical Care Ventilator Service Manual

7-10

Figure 7-4

2. DC Power Fuse Fuse replacement steps:

• Insert screwdriver into the trench (1) of end of fuse box, see Figure 7-2. • Pull out fuse holder (2). • Remove the fuse (3). • Load new fuse. • Push the new fuse into the original position gently.

Figure 7-5

7.5 Battery Maintenance

7.5.1 Battery Specification

Battery module - DC12V, 6.6AH, 14.4V lithium-ion battery

- Typical charge time: 3.5h

- Typical discharge time: 2h

When the main power supply voltage is too low or the main power supply fails, two backup batteries (one is necessary, and one is optional) can protect the ventilator. When having a power failure, the ventilator can switch to battery supply automatically, and can normally work without pneumatic power supply failure. The two batteries are usually available for the ventilator working for 4 hours.

Page 85: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

7.5.2 Precautions

Charge: When operating with AC power supply on, the system will maintain the battery automatically. Charge time is less than 3.5h.

Discharge: The machine is operating on battery.

In case of low battery condition, an alarm message “Low battery” will appear, notifying the user to restore AC power supply to charge batteries, otherwise the batteries will be depleted and alarm “Limited Battery Capacity” will be displayed, and eventually the system will be shut down. (For safety reason, manual power-on is required to start the machines again after an automatic shutdown).

Before the machine is put into patient use, or disconnecting AC power for transport, or other purposes, check the battery power. If the battery is not fully charged, connect the ventilator to AC power for at least 3.5hours and recharge the battery until the power reaches 80%~100%.

7.5.3 Battery Storage

In case the battery is to be stored for a long time, charge it fully prior to storage.

To keep the battery power and prolong the battery life, please ensure that the ventilator is connected to the main power. Charge the battery every six months. Actual charging time depends on the storage environment.

High humidity and high temperature environments should be avoided for storage.

If battery is damaged due to improper maintenance, replace promptly, otherwise liquid leakage may corrupt the machine. Contact the manufacturer when replacing battery.

7.5.4 Battery Replacement

Same model battery with CE certification is required. Make sure AC power supply is disconnected before replacing.

CAUTION: An authorized Aeonmed service representative can replace battery. If the

machine will not be used for long-time, please contact Aeonmed service representatives to disconnect battery. A depleted battery should be disposed in accordance with the local policies.

7-11

Page 86: V8600 Critical Care Ventilator Service Manual

7-12

7.5.5 Battery Charging and Calibration

Use the battery charger supplied by our company to charge or calibrate the battery. After calibrating the battery, the ventilator can read the residual battery capacity accurately.

Please charge or calibrate the battery according to the instructions of the battery charger.

CAUTION: When ‘low battery’ alarm occurs, charging should be done immediately.

The VG70 Ventilator System will shut off in several minutes automatically.

7.6 Oxygen Sensor

The oxygen sensor can be used to measure the local oxygen concentration when it is connected to the ventilator or other equipment. The oxygen sensor is suitable for adult and child.

7.6.1 Oxygen Sensor Replacement

Disassembly

Step 1: Open outward (1), and then remove the cover.

Step 2: Pull up the crystal joint of the oxygen sensor, turn the oxygen sensor (3) anticlockwise, and then remove it.

Page 87: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

Assembly

Inspect the oxygen sensor for damage and replace as necessary. Then reassemble the oxygen sensor.

7.6.2 Oxygen Sensor Calibration

For oxygen sensor calibration, refer to section 9.1.

7.6.3 Technical Specifications of Oxygen Sensor

Measure range 0 ~ 100 Vol.%

Measure accuracy <1% Response time <10s Operating temperature 10℃~ 45 ℃ Pressure range 600hPa ~ 1750hPa Humidity effect -0.03× current humidity Storage temperature -20℃~ 50℃ Drift Air: <1%/month Reproducibility ±1% Type Chemical fuel cell Expect life time 12 months Total system response time

<60s

Working principle of O2 monitor

The O2 monitor surveys and displays the O2 concentration in the patient loop. The oxygen sensor component contains an oxygen sensor, which can produce the voltage proportional to the oxygen partial pressure (concentration) on its detection surface. The oxygen sensor is an electrochemical device (chemical battery). Oxygen expands in this device through a layer of film and oxidizes the base metal electrode. The oxidation process

7-13

Page 88: V8600 Critical Care Ventilator Service Manual

7-14

produces a current with an amplitude proportional to the oxygen partial pressure indicated by the electrode sensor. The base metal electrode is gradually exhausted during the oxidation process. The voltage of the sensor is influenced by the temperature of the monitoring gas mixture. The surgical thermosensitive resistor of the sensor automatically compensates temperature change in the sensor. The O2 monitor converts the sensor signal into the corresponding oxygen percentage value by using signal processing and circuit analysis. The system displays the value and compares it with the stored alarm limit value. If the value exceeds the threshold, the monitor will alarm.

NOTE: After being in a condensing atmosphere, the oxygen sensor shall be stored for

more than 24 hours in an environment equivalent to operating humidity.

7.6.4 Oxygen Sensor Maintenance

The oxygen sensor should be regularly calibrated. For the calibration interval, refer to section 7.2.2.

To improve the life time of the oxygen sensor, when the ventilator is not in use, the oxygen sensor should be avoided contact with the high-concentration oxygen.

The oxygen sensor is consumptive, and the period of valid is ordinarily 12 months. So the user should pay attention to the use of the oxygen sensor. When the oxygen sensor fails, replace it.

The recommended oxygen sensor is supplied by Aeonmed.

WARNING: Do not immerse oxygen sensor in liquid. Do not conduct autoclave or

high temperature fumigation on the oxygen sensor.

7.7 Paramagnetic Oxygen Sensor (optional)

The paramagnetic oxygen sensor can be used to measure the local oxygen concentration when it is connected to the ventilator or other equipment. The oxygen sensor is suitable for adult and child.

Page 89: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

7.7.1 Paramagnetic Oxygen Sensor Calibration

For the paramagnetic oxygen sensor calibration, refer to section 10.1.

7.7.2 Technical Specifications of Paramagnetic Oxygen

Sensor

Performance

Technology Paramagnetic

Range 0-100% O2 (with over range -15% O2 to 200% O2)

Accuracy (Intrinsic Error)

<± 0.2% O2

Linearity <± 0.2% O2

Repeatability <± 0.2% O2

Zero Drift <± 0.4% O2 in first 24 hours, then <± 0.2% O2 /week, then <± 0.2% O2 /month

Response Time (T10 –T90)

8 to 20 seconds dependent on application and filter selection (biological filter on request)

Outputs/Controls

Signal Output Digital UART or linear mV output (0.5mV or 10mV per % O2)

Physical

Weight 70g (2.47oz)

Dimensions Molex low profile connector: 33.5×30.0×46.1mm (1.32×1.18×1.81”)

Diffusion Port Aperture diameter: 15.5mm (0.61”)

Sample condition

Sample Gas Condition Clean dry gas, free of entrained oil, particulates <3μm, non-condensing

Gas Exchange Diffusion

Ambient conditions

Operation Temperature 5℃ to 50℃ (41℉ to 122℉)

Storage Temperature -30℃ to 70℃ (-22℉ to 158℉)

Temperature Coefficient Within a range of 0℃ to 50℃: Zero <± 0.5% O2 /10℃, Span: <±0.5% O2 /10℃

Operating Pressure Range

±33kPag (±5psig)

7-15

Page 90: V8600 Critical Care Ventilator Service Manual

7-16

Ambient Humidity 0 to 95% non-condensing

External Power Supply 5V dc, 70mA nominal

Power Consumption 350mW

7.8 Disposal

This product must not be disposed of with your other waste. Instead, it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment, or by returning it to our company for reprocessing. The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. For more information about where you can drop off your waste equipment for recycling, please contact your local city office, your waste disposal service, or your product distributor or retailer.

Correct Disposal of Batteries and O2 Sensors

WARNING: Treatment of batteries and O2 sensor:

Follow all local regulations with respect to environmental protection when disposing of batteries and O2 sensor. These products contain toxic compounds irrespective of physical condition. They should be disposed of according to local waste management requirements and environmental legislation. They should not be burned since they may give off toxic fumes. Do not throw into fire! Risk of explosion. Do not force open! Danger of bodily injury.

7.9 Manufacturing techniques and materials:

For a period of one year from the date of original delivery, the components and assemblies of this product are warranted to be free from defects manufacturing techniques and materials, provided that the same is properly operated under the conditions of normal use and regular maintenance. The warranty period for other parts is three months. Expendable parts are not included. Aeonmed’s obligation under the above warranties is limited to repairing free of charge.

Page 91: V8600 Critical Care Ventilator Service Manual

7 User Maintenance

7.10 Free Obligations:

• Aeonmed’s obligation under the above warranties does not include the freight and other fees;

• Aeonmed is not responsible for any direct, indirect or failed product or delay which result from improper use, alteration by using unapproved assemblies and maintenance by anyone other than Aeonmed;

• This warranty does not apply to the followings: Improper use; Machines without maintenance or machines broken; The label of Aeonmed original serial number or mark is removed or replaced; Other manufacturers’ products.

7.11 Security, reliability and operating condition:

Aeonmed is not responsible for the security; reliability and operating condition of this product in case that:

• The assemblies are disassembled, extended, altered and/or readjusted • This product is not operated correctly in accordance with the manual instruction. The power supply used or operating environment does not follow the requirements in this manual.

7.12 Return

Follow the steps in case the product needs to be returned to Aeonmed:

1、Obtain the rights of return

Contact customer service of Aeonmed by informing them of the serial number and type of the product. The number is marked on the surface of the product. Return is unacceptable if the number cannot be identified. Enclose a statement of the number, type and the reason of return.

2、Transportation charges

Transportation and insurance charges must be prepaid by the user for transporting the product to Aeonmed for repairing.

7-17

Page 92: V8600 Critical Care Ventilator Service Manual
Page 93: V8600 Critical Care Ventilator Service Manual

8 Pre-Use Test

8 Pre-Use Test

8.1 Pre-use test procedure

After power on, the machine enter Pre-use test. The user can also do all the pre-use test items on Standby screen.

Test items on Standby screen are as follows.

Test Items Remarks

Technical Test

After each system has completed its initialization, technical tests will be performed, including: voltage checks at critical points in the circuitry; data collection necessary for system operation; test of communication between sub-systems; tests of measurement circuits and valve control circuits.

AC/Battery test

This test will verify whether the batteries can supply enough power to operate the ventilator normally. Please follow instructions as displayed.

Gas supply test Test will proceed when the hyperbaric oxygen is functional.

Oxygen sensor test.

This test requires that oxygen supply is available. If oxygen source is not available then a message “oxygen source is inadequate” and the test cannot be carried out.

Leak test Internal leakage test.

Flow sensor test Flow sensors function and accuracy test.

Pressure sensor test Pressure sensors function and accuracy test.

Safety valve test Safety valve function and accuracy test.

Circuit test Circuit compliance value measurement.

CO2 Sensor test Performed if a CO2 module is detected. An Alert shall be posted if the test fails.

Methods for testing the function of the alarm system for conditions are specified by IEC60601-2-12. These are to be performed at the user’s discretion.

Perform the following procedure to prove operation of the Low MVe and High Airway pressure alarms:

1. Turn the Power switch On.

2. Connect a breathing circuit and test lung to VG70.

8-1

Page 94: V8600 Critical Care Ventilator Service Manual

8-2

3. Press Start Ventilation and ventilate with default settings, set O2 to 21%.

4. After 5 breaths, observe the MVe reading on the display.

5. Set the MVe low alarm limit to a value greater than the observed MVe reading.

6. Verify that a low level MVe low alarm is present on the 3rd breath.

7. Return the MVe low alarm limit to original setting.

8. After 5 breaths, observe the Ppeak reading on the display.

9. Set the PAW upper alarm limit to a value lower than the observed Ppeak reading.

10. Verify that a low level High Airway Pressure alarm is present after 1 breath and that a high level High Airway Pressure alarm is present at the start of the 4th breath.

11. Return the PAW upper alarm limit to original setting.

12. Turn the Power switch OFF.

8.2 Pre-use test failure analysis

If the pre-use test failed, you need to follow the way to troubleshoot.

Pre-use test item Possible reason and recommended action if failed:

Technical Test See the Technical Test Error Code below.

AC/Battery Test Make sure at least one battery is installed and charged. Make sure the AC supply is OK.

Gas Supply Test Make sure the oxygen supply is connected, and the pressure is in the spec

Oxygen Sensor Test

Perform 21% and 100% oxygen sensor test at first, and try again. Make sure the oxygen supply is connected, and the pressure is in the spec. Make sure the oxygen sensor is valid and connected to the main box. Make sure the oxygen auto calibration valve is connected right and no leak.

Leak Test Calibrate Inspiratory valve at first, and try again; Make sure machine and connected tube have no leak.

Flow Sensor Test

Calibrate Inspiratory valve at first, and try again; Make sure machine and connected tube have no leak; Make sure inspiratory flow sensor and expiratory flow sensor work fine.

Page 95: V8600 Critical Care Ventilator Service Manual

8 Pre-Use Test

Pre-use test item Possible reason and recommended action if failed:

Pressure Sensor Test

Calibrate Inspiratory valve at first, and try again; Make sure machine and connected tube have no leak; Make sure inspiratory pressure sensor and expiratory pressure sensor work fine.

Safety Valve Test

Calibrate Inspiratory valve at first, and try again; Make sure safety valve works fine; Make sure machine and connected tube have no leak; Make sure expiratory pressure sensor works fine.

Circuit Test

Calibrate Inspiratory valve and Expiratory valve at first, and try again. Make sure machine and connected tube have no leak; Make sure expiratory pressure sensor works fine.

CO2 Sensor Test

Take off the CO2 sensor from the breath circuit, and put it in the air. Make sure the CO2 sensor has warmed up. The LED indicator in the sensor turns green. Make sure the adapter is installed into the sensor.

8-3

Page 96: V8600 Critical Care Ventilator Service Manual
Page 97: V8600 Critical Care Ventilator Service Manual

9 Network

9 Network

9.1 Overview

The ventilator can export patient monitoring data in HL7 format over a local area network (LAN) using the TCP/IP transport protocol. The data can be sent to a server and then forwarded to an Electronic Medical Records (EMR) system or sent directly to the EMR system.

9.2 What is Exported?

Network completes ventilator local IP and port settings, specifies IP (common PC) of node and port number that connect and send data. After doing the above settings can establish a connection to the ventilator. This machine data is sent to the remote destination node. Currently send data by two ways. One is sent cyclically, the period is 1second, and sending waveform data and monitoring values. Another is sent immediately, send alarm information, ventilation parameters, alarm limit settings and so on at the beginning of connection. When the information is changed that send synchronal information. Detailed data communication protocols refer to "HL7 agreement."

9.3 Establishing a Connection

Using data transmission functions require checking whether connection is open. Check by following steps:

1) Connect the ventilator and network of purposes PC with network cable.

Connection can be achieved by two ways: First, the ventilator access network equipment, such as Hub, switches and so on. Second, connect the ventilator and the destination PC's network card with a network cable. Two connections use ordinary network cable.

2) Ensure that the ventilator set the IP address manually, not automatically obtain IP.

USB key board is connected to the ventilator while click two keys Win and Esynchronously, then appear “My Device” interface, press“ control Panel”“Network and Dial-up Connections”“PCI-E100E1”, enter the connection settings interface. Obtaining the IP address is set to: Specify an IP address, and the ventilator is specified an IP address.

Set IP address of ventilator and destination PC in the same network segment. Such as

9-1

Page 98: V8600 Critical Care Ventilator Service Manual

9-2

PC's subnet mask is 255.255.255.0, subnet mask of ventilator also performs the same setting, first three digits setting of ventilator IP address is the same as PC's IP address and ensure that the final one setting is different other hosts on the network segment. Then click “win” button, and select “reboot”, hot restart ventilator.

After this operation, ventilator enter the local IP and remote IP and click the Connect button to complete the local IP settings by the ventilator settings →UI interface network interface. You can enter the ventilator control panel to set without keyboard each time

3) Use “Ping” command to verify network connectivity.

Use “Ping” command to verify network connectivity on the PC. “Ping” open indicates the network is available, you can click connection button to start data transmission. Click the Connect button, if the ventilator displays IP settings are different the current IP, change interface IP settings for ventilator IP, or retain the current setting. Then start data transmission.

4) The PC starts data receiving test software to check whether the data is sent correctly. The PC starts data receiving test software, input ventilator IP, click “connection”.

Ventilator system setting web interface, complete the above 1), 2), 3) operation, click the “connection” that starts data transmission. Received data can be seen by PC-data receiving software.

9.4 Interface Instructions

Click “Configurations” to enter the configurations interface. On this page, the user needs to enter the password, as shown in Figure 9-1.

Figure 9-1

Page 99: V8600 Critical Care Ventilator Service Manual

9 Network

Input the correct password to enter, Click “Site Configuration” to enter the site configuration interface. As shown in Figure 9-2.

Figure 9-2

Various parts of the interface are described as follows:

Figure 9-3

1) Local IP Address. Each number is from 1 to 255. If the input is outside this range, a message is displayed: “* is invalidated. Please insure the value is between 1 and 255.” If it is not set then display 0.0.0.0.

2) Local subnet mask. If it is not set then display 0.0.0.0. This is generally filled 255.255.255.0 or 255.255.0.0. This set is the same as subnet mask of destination PC.

3) Local port. Ventilator sends data through the port. Receiver gets data from the ventilator through the port. Port 9101 is used by default. It can use from 65535 to 5000 as

9-3

Page 100: V8600 Critical Care Ventilator Service Manual

9-4

the port number, If the input is outside this range, a message is displayed:” * is invalidate. Please insure the value is between 5000 and 65535.”

4) Remote IP address. It is the IP address of the data receiving end. Once connected, the ventilator only sends data to this address of the host. Usually the IP and ventilator IP is in the same segment.

5) Remote port. It is data receiving end port. Ventilator establishes a connection with the receiver through the port, receiving end receives data through the port. This machine is generally associated with Ventilator port. Port 9101 is used by default. It can use from 65535 to 5000 as the port number, If the input is outside this range, a message is displayed:” * is invalidate. Please insure the value is between 5000 and 65535.”

6) “Connect”. Click this button to start data transmission. A message is displayed: message is displayed on (8) area, while a small icon is displayed in the upper right corner of interface within three seconds. Then all the data will be saved in the ventilator file, when open this interface again, it will display the last settings.

7) “Disconnect”. Click this button to stop the data transfer.

8) Data sharing status. Connected, a message is displayed: “Data Share On”; disconnected, a message is displayed: “Data Share Off”.

9) Digital input keypad. It is used to enter the digital.

9.5 Transmits and Receives Data

Data is packaged in HL7 format and transmitted by UDP packets. You can set up UDP listening service to receive data, unpack the data according to the rules of the relevant agreements. Detailed rules refer to "HL7 Protocol."

1) Waveform data. The period of Waveform is 1second. Receiver receives the waveform data and saves into the buffer, and then it is drew on the interface according to the sampling frequency (50ms). Drawing frequency and sampling frequency may be different. It is adjusted according to the number of data in the buffer. Since the network delay and buffer, there will be a delay drawing, the maximum delay is about 7 seconds.

2) Monitoring data. The transmission period is 1 second. Only the network latency is affect monitoring data updates, it is generally not delay in the open network.

3) Alarm information. It sends all current alarm information when connection is established and ventilation is started. It will also send the appropriate information when the new alarm generates or alarm disappears.

4) Mode and ventilation parameters. It sends all current mode and ventilation parameters when connection is established and ventilation is started. It will send the change mode and ventilation parameters settings in the mode transitions.

5) Alarm limit. It sends all current alarm limits when connection is established and

Page 101: V8600 Critical Care Ventilator Service Manual

9 Network

ventilation is started. It will send the change value when alarm limit is changed.

9.6 Troubleshooting

Right transmission data requires that network settings of ventilator is correct, receiver test software can start normally, and network port of ventilator is installed correctly, and so on.

1) Local network settings of ventilator can refer to 9.3.

2) Receiver test software can move in windows XP and windows 7 system, if it can’t receive data and ventilator and network cable are normally, network configuration information may be incorrect. In this case, you can use the command:”netsh winsock reset” that reset the network configuration and reboot the computer.

3) Ventilator network card. The above case is correctly, the network is still blocked, namely, ping is failure. You should check that ventilator network card is properly installed and connected. Use correct network port which is connected to the ventilator, if you can Unicom network, show the network adapter or interface have problems. Can replace card or connect afresh.

4) The PC's network connection is disconnected if you connect the ventilator and computer with network cable, you should check whether the line is connected properly. Make sure the cable is normal, then should open the ventilator screen, check the card is properly connected.

5) If uld op is fail, make sure “remote IP Address” is the same as PC IP. If the PC uses a wireless connection that can be disconnected, then verify properly connected.

9-5

Page 102: V8600 Critical Care Ventilator Service Manual
Page 103: V8600 Critical Care Ventilator Service Manual

10 Service Menu

10 Service Menu

CAUTION: On “System” of “Main Menu”, “Service” tab is reserved for service only.

Press “Service” to enter the “Service” page. On this page, the user needs to enter the password, as shown in Figure 10-1.

Figure 10-1

Input the correct password to enter, There are six choices on this page: Calibration, Event/Alarm Log, Machine Information, Language, Test Page and Update. The default page is Calibration. See Figure 10-2.

Figure 10-2

10-1

Page 104: V8600 Critical Care Ventilator Service Manual

10-2

10.1 Calibration

10.1.1 Calibration operation guidance

The calibration choices include: 1)Pressure Sensor Calibration, 2) Flow Sensor Calibration, 3) O2 Sensor Calibration, 4)CO2 Sensor Calibration, 5) Inspiratory valve Calibration, 6) Expiratory Valve Calibration, 7) Atmospheric Sensor Calibration, 8) Touch Screen Calibration, 9) Leak Test and 10) Breath Circuit test, as shown in Figure 10-2.

1) Pressure Sensor Calibration: Press “Pressure Sensor” to enter the calibration interface. A message is displayed: “This step is to zero the pressure sensor. Please remove the breathing circuit from the ventilator before calibration.” A legend is displayed as well as shown in Figure 10-3.

Figure 10-3

Press “Start” to start pressure sensor calibration. A progress bar and a message “Calibration in progress. Please Wait” will be displayed as shown in Figure 10-4.

NOTE: During this period no other operation can be performed. Pressing other areas

will have no response.

Page 105: V8600 Critical Care Ventilator Service Manual

10 Service Menu

Figure 10-4

2) Flow Sensor Calibration: Press “flow sensor” to enter the interface. A message is displayed: “This step is to calibrate the flow sensor. Please connect the insp. Port and Exp. Port directly with a tube” as shown in Figure 10-5.

Figure 10-5

Press the “Start” button to start flow sensor calibration, the remaining procedure is the same as the pressure sensor calibration.

3) O2 Sensor Calibration: Press “O2 sensor” to enter the interface. A message displayed: “Please verify that the oxygen source are connected correctly. Verify that the gas inlet pressure is within specification.” A legend will also be shown. There are two keys below the legend: “Start 21%” and “Start 100%”. Choose the needed one and Press, as shown in Figure 10-6. The remaining procedure is the same as described above.

10-3

Page 106: V8600 Critical Care Ventilator Service Manual

10-4

Figure 10-6

4) CO2 Sensor Calibration: Press “CO2 Sensor” to enter the interface, as shown in Figure 10-7. A message displayed: “Disconnect the CO2 sensor with the adapter from breathing circuit and ensure it is in ambient air. Wait 1 minute for warm up after the unit is powered on or after installing an airway adapter. Press “Zero” when the State-Area turns green”. Please follow the prompt message to calibrate.

Figure 10-7

5) Inspiratory Valve Calibration: Press “inspiratory valve” to enter the calibration interface. A message is displayed: “This step is to calibrate the inspiratory valve. Please connect the Insp. Port and Exp. Port directly with a tube.” as shown in Figure 10-8. To get a precise calibration, please connect patient tubes with a humidifier before inspiratory valve calibration.

Click the “Start” button to start inspiratory valve calibration, the remaining procedure is the same as the pressure sensor calibration.

Page 107: V8600 Critical Care Ventilator Service Manual

10 Service Menu

Figure 10-8

6) Expiratory Valve Calibration: Press “Expiratory valve” to enter the calibration interface. A message is displayed: “This step is to calibrate the expiratory valve. Please connect patient circuit and test lung before calibration.” as shown in Figure 10-9.

Click the “Start” button to start expiratory valve calibration, the remaining procedure is the same as the pressure sensor calibration.

Figure 10-9

7) Touch Screen Calibration: Press “Touch screen” to enter the calibration interface. A message is displayed: “This step is to calibrate the touch screen. The ventilator’s screen will disappear during the calibration. Please follow the instruction in the calibration program.” as shown in Figure 10-10.

CAUTION: Please calibrate the touch screen periodically or when it works

abnormally.

10-5

Page 108: V8600 Critical Care Ventilator Service Manual

10-6

Figure 10-10

Press the “Start” key to start, the remaining procedure is as described in steps above.

8) Leakage Test Calibration: Press “Leakage Test” to enter the test interface. A message is displayed: “This step is to test the internal leakage of ventilator. Please connect the Insp. Port and Exp. Port directly with a tube”, as shown in Figure 10-11.

Figure 10-11

Press “Start” to start the leakage test. A progress bar and a message “Test in progress, please wait” will be displayed as shown in Figure 10-12.

Page 109: V8600 Critical Care Ventilator Service Manual

10 Service Menu

Figure 10-12

9) Breath Circuit Test Calibration: Press “Breath Circuit Test” to enter the test interface. A message is displayed: “This step is to test the compliance and leakage of breathing circuit. Please connect the patient circuit to the T-piece, and plug up the patient end of the T-piece”, as shown in Figure 10-13.

Before starting the test, ensure the patient circuit has been connected to the T-piece and the patient end of the T-piece has been plugged up. Press the “Start” button to start the breath circuit test.

Figure 10-13

10-7

Page 110: V8600 Critical Care Ventilator Service Manual

10-8

10.1.2 Calibration failure analysis

Calibration items Possible reason and recommended action if failed:

Pressure Sensor

Make sure inspiratory pressure sensor and expiratory pressure sensor work fine; Make sure no gas flow pass inspiratory pressure sensor and expiratory pressure sensor.

Flow Sensor

Make sure inspiratory flow sensor and expiratory flow sensor work fine; Make sure no gas flow pass inspiratory flow sensor and expiratory flow sensor.

O2 Sensor

Calibrate Inspiratory valve and Expiratory valve at first, and try again; Make sure the oxygen supply is connected, and the pressure is in the spec; Make sure the oxygen sensor is valid and connected to the main box.

CO2 Sensor See Section 8.2

Inspiratory Valve

Make sure the oxygen supply is connected and work normal; Make sure machine and connected tube have no leak; Make sure inspiratory flow sensor and expiratory flow sensor work fine. Make sure inspiratory valve works fine.

Expiratory Valve

Make sure the oxygen supply is connected and work normal; Make sure machine and connected tube and lung have no leak; Make sure inspiratory pressure sensor and expiratory pressure sensor work fine; Make sure inspiratory valve and expiratory valve work fine.

Touch Screen

Trying the fewest test points calibration (4 points). Make sure press the center of each test point (the red region). Make sure the plastic frame of screen doesn’t constrict the touching area of touch screen.

Page 111: V8600 Critical Care Ventilator Service Manual

10 Service Menu

10.2 Event/Alarm Log

Press “Event/Alarm Log” to enter “Event/Alarm Log” page, as shown in Figure 10-14.

Figure 10-14

The middle area of this page is the message area. It can store up to 1000 messages, including event messages and alarm messages. All messages will be listed in time sequence. The top is the latest event or alarm message, and the bottom is the oldest. Use the scroll bar to check all the messages. An asterisk (*) in front of an alarm message indicates that alarm message was not displayed in the alarm message area. As shown in Figure 10-15. The event/alarm log records all alarms and most actions.

Figure 10-15

Below is the message area of the Settings area resulting from highlighting an alarm. All settings will be given here, as shown in Figure 10-16.

Figure 10-16

NOTE: Alarm/Event Log data is retained during a power interruption and can be viewed when power returns.

10-9

Page 112: V8600 Critical Care Ventilator Service Manual

10-10

10.3 Machine Information

Press the “Machine information” key to enter the Machine information page, this area includes the following information, as shown in Figure 10-17.

1. Software Version: a. UI b. BDU c. Power Supply

2. Runtime Hours 3. O2 Source Pressure 4. Atmospheric Pressure

Figure 10-17

10.4 Language

Press “Language” to enter the language screen. English and other languages are available for the user to choose.

10.5 Test Page

Press “Test Page” to enter the test page. There are five choices on this page: Demo, Vlt. Monitor, Schematic, Cali. Data, Service Timer, Error Code and PT100 Cali. as shown in Figure 10-18.

Page 113: V8600 Critical Care Ventilator Service Manual

10 Service Menu

Figure 10-18

10.5.1 Demo

Press “Demo” to enter the demo setting page, as shown in Figure 10-19. In this page, the operator can turn on or turn off the demo mode, and test the broken pixel of the screen.

Figure 10-19

10.5.2 Vlt. Monitor

Press “Vlt. Monitor” to enter the vlt. monitor page, as shown in Figure 10-20. This page displays all the BDU and PS circuit monitoring value.

10-11

Page 114: V8600 Critical Care Ventilator Service Manual

10-12

Figure 10-20

10.5.3 Schematic

Press “Schematic” to enter the schematic page. The schematic is as shown in Figure 10-21. In this page, the operator can control all valves and read all sensors state.

The valves (#1 to #9) and the Exp. Valve Heater shall have two states: on and OFF; The Turbine voltage range: 0-5, the units: V, the resolution: 1; The Insp. Valve range: 0-180, the units: LPM, the resolution: 0.5; The IP Valve range: 0-100, the units: cmH2O, the resolution: 1.

Figure 10-21

Page 115: V8600 Critical Care Ventilator Service Manual

10 Service Menu

10.5.4 Cali. Data

Press “Cali. Data” to enter the cali. data page, as shown in Figure 10-22. This page displays the calibration data of Inspiration /Expiration Sensor.

Press “Refresh” button to update the calibration data and the zero AD value of the sensor. Press “Manual Cali.” button to display the manual calibration page; in the page, the operator can manually calibration the Insp/Exp sensor.�

Figure 10-22

10.5.5 Service Timer

Press “Service Timer” to enter the service timer page, as shown in Figure 10-23. This page displays the service time about the ventilator, the turbine runtime, and the int./opt. battery charging/discharging circle count.

Figure 10-23

10-13

Page 116: V8600 Critical Care Ventilator Service Manual

10-14

10.5.6 Error Code

Press “Error Code” to enter the Error Code interface, as shown in Figure 10-24.

Figure 10-24

10.5.7 PT100 Cali.

Press “PT100 Cali.” to enter the PT100 calibration interface, as shown in Figure 10-25.

Figure 10-25

Page 117: V8600 Critical Care Ventilator Service Manual

10 Service Menu

The calibration steps are as follows.

Step 1: Press “Stop Heater” and wait for 40min to make the expiration valve to cool to ambient temperature. Then measure and record the temperature of the expiration valve.

Step 2: Press “ ” turning the key background color to yellow, adjust the value equal to the temperature recorded at Step 1, then read the value “ ” and record the value in Textbox 1.

Step 3: Press “ ” turning the key background color to yellow, adjust the value equal to the setting temperature, then read the value “ ” and record the value in Textbox 3.(The setting temperature: the highest temperature of the expiration value

heating, default: 35℃. )

Step 4: In turn, press “Write”, “Read”, “Set Point”, “Read Point”, then determine that the

value of Textbox 2 is equal to the value of Textbox 1 and the value of Textbox 4 is equal to the value of Textbox 3. If same, the calibration is finished; if not, carry out Step 5. Step 5: Repeat the above steps.

10.6 Update

Press “Update” to enter the update page, as shown in Figure 10-26. In this page, the operator can update the software.

Figure 10-26

10-15

Page 118: V8600 Critical Care Ventilator Service Manual

10-16

10.7 Optional

Press “Optional” to enter the optional page, as shown in Figure 10-27. In this page, the operator can choose configurable functions.

Figure 10-27

Page 119: V8600 Critical Care Ventilator Service Manual

11 Troubleshooting

11 Troubleshooting

11.1 Technical Error

11.1.1 Technical Test Error Code

The error code is made up by 8 digits hexadecimal numbers, which are four byte using binary system. The highest two bytes are internal AD test result flag, the lowest byte are external AD test result flag. The meaning of each bit is shown in the below table.

External AD##(0xFFFF00xx)

Test Point Possible Reason and Recommended Action

Bit 0 Zero point of expiratory flow rate

1. Remove the test lung and breath circuits, perform the test again.

2. Don’s start PUT too quickly after quit from ventilation.

3. Calibrate the pressure, flow rate sensor, Insp. valve and Exp valve.

4. Proportional solenoid valve can’t close completely. Replace proportional solenoid valve.

5. Replace expiratory flow rate sensor.

Bit 1 Zero point of inspiratory flow rate

1. Remove the test lung, connect Insp. port and Exp. port directly with a tube.

2. Don’s start PUT too quickly after quit from ventilation.

3. Calibrate the pressure, flow rate sensor, Insp. valve and Exp valve.

4. Proportional solenoid valve can’t close completely. Replace proportional solenoid valve.

5. Replace mother board.

Bit 2 Inspiratory temperature

1. Check the cable between the mother board to the temperature sensor

2. The interface IC on the mother board is broken. Replace mother board.

3. The input port of external AD chip on the mother board is broken. Replace mother board.

4. Replace temperature sensor.

11-1

Page 120: V8600 Critical Care Ventilator Service Manual

11-2

Bit 3 -- --

Bit 4 Zero point of expiratory pressure

1. Remove the test lung, connect Insp. port and Exp. port directly with a tube.

2. Don’s start PUT too quickly after quit from ventilation.

3. Calibrate the pressure, flow rate sensor, Insp. valve and Exp valve.

4. Proportional solenoid valve can’t close completely. Replace proportional solenoid valve.

5. The amplifier on the mother board is broken. Replace mother board.

6. The input port of external AD chip on the mother board is broken. Replace mother board.

7. The sensor is broken. Replace mother board.

Bit 5 Zero point of inspiratory pressure

1. Remove the test lung, connect Insp. port and Exp. port directly with a tube.

2. Don’s start PUT too quickly after quit from ventilation.

3. Calibrate the pressure, flow rate sensor, Insp. valve and Exp valve.

4. Proportional solenoid valve can’t close completely. Replace proportional solenoid valve.

5. The amplifier on the mother board is broken. Replace mother board.

6. The input port of external AD chip on the mother board is broken. Replace mother board.

7. The sensor is broken. Replace mother board.

Bit 6 Barometric pressure

1. Proportional solenoid valve can’t close completely. Replace proportional solenoid valve.

2. The amplifier on the mother board is broken. Replace mother board.

3. The input port of external AD chip on the mother board is broken. Replace mother board.

4. The sensor is broken. Replace mother board.

Bit 7 -- -- Internal AD(0xxxxx00FF)

-- --

Bit 15 -- -- Bit 16 -- -- Bit 17 -- -- Bit 18 -- --

Page 121: V8600 Critical Care Ventilator Service Manual

11 Troubleshooting

Bit 19 Expiratory valve AD

1. Poor connection between the mother board and daughter board. Re-plug the daughter board. 2. The driver chip of expiratory valve on the mother board is broken. Replace mother board. 3. The daughter board is broken. Replace daughter board.

Bit 20

Inspiratory valve AD

1. Poor connection between the mother board and daughter board. Re-plug the daughter board.

2. The daughter board is broken. Replace daughter board.

Bit 21 -- -- Bit 22 -- -- Bit 23 -- -- Bit 24 -- -- Bit 25 -- -- Bit 26 -- --

Bit 27 2.5V reference voltage

--

Bit 28 -- -- Bit 29 -- -- Bit 30 -- -- Bit 31 -- --

1. Error Codes in the Log

Error Code

Test Point

Possible Reason and Recommended Action

1000 T12 DC The A5V out is out of range. Replace PS board. 1001 T1 DC The W24V output is out of range. Replace switch PS. 1002 T13 DC The P12V output is out of range. Replace PS board. 1004 T19 DC The N24V output is out of range. Replace PS board. 1008 T24 DC PS board failure. Replace PS board. 1009 T14 DC The A12V output is out of range. Replace PS board. 1013 -- During expiratory valve heating the temperature is out of range.

Replace PS board. 2000 T31 DC During expiratory valve heating the output voltage is out of range.

Replace PS board. 2001 T30 DC During expiratory valve heating the output current is out of range.

Replace PS board. 3006 -- PS board communication failure. Replace PS board. 3005 -- BDU communication failure. Replace core board. 3002 -- Software mismatch failure. Update software.

11-3

Page 122: V8600 Critical Care Ventilator Service Manual

11-4

11.1.2 Other Errors

Alarm sound from buzzer in the PS board

1) Communication failure between UI to PS board

2) No AC and battery connected

Check if the UI are working. Make sure the cable linking to the PS board is OK. Replace power supply board.

11.1.3 Service Tools

Service tools are listed underside:

Sequence number

Name Quantity Remark

1 Metric hex driver set One set ----

2 #2cross screwdriver One entries Staff width is about

5mm

3 Multi-meter One block ----

4

VG70 checkout frock and its attachments for ESD

(Electro-Static Discharge) protection)

One desk

If anyone of the electric machine, electric

machine driver and the main board is replaced

5 Breath pipeline cover bag One set ----

6 Simulated lung One entries ----

Page 123: V8600 Critical Care Ventilator Service Manual

12 Appendix A

12 Appendix A Contact & Ordering Information

How to Call for Service

This service manual provides procedures for testing and maintaining the Ventilator VG70. It is not intended to be a complete maintenance document; therefore, it contains no detailed disassembly or reassembly instructions.

Refer any repairs or adjustments that exceed the scope of this manual to the Service Center of Aeonmed by calling 800-810-8333.

This manual contains proprietary information. It is intended for use only by individuals qualified in the installation and maintenance of the Aeonmed ventilator. Receipt, purchase, or possession of this document in no way confers or transfers any other rights for the use of this information. Disclosure or reproduction of the enclosed, without the written permission of Aeonmed is prohibited.

This manual is intended for use only by technicians who have successfully completed Aeonmed training on this product.

Aeonmed believes the information herein is accurate but accepts no responsibility for errors, omissions, or misrepresentation.

12-1

Page 124: V8600 Critical Care Ventilator Service Manual
Page 125: V8600 Critical Care Ventilator Service Manual

13 Appendix B

13 Appendix B Diagrams and Schematics

Pneumatic diagram

13-1

Page 126: V8600 Critical Care Ventilator Service Manual
Page 127: V8600 Critical Care Ventilator Service Manual

14 Appendix C

14 Appendix C Specifications

System

Operating Conditions

Operating Temperature Range: +5 to +40 ℃

Relative Humidity: 5 to 95% non-condensing

Atmospheric Pressure: 700 to 1060 hPa

Non-operating Conditions

Storage Temperature Range: -20 to +60 ℃

Storage Relative Humidity: ≤95% non-condensing

Storage Atmospheric Pressure: 700 to 1060 hPa

Power Supply

Power Supply, Automatic Range Selection

Battery Backup (Standard and Optional)

Maximum Power Consumption

100-240 +/- 10% VAC 50/60 Hz

Two rechargeable battery modules, 12 V, 6.6 Ah each. Recharge time approximately 3.5hours. Battery backup time of 120minutes minimum with only standard internal battery.

CAUTION: The Power Supply should meet the above specifications.

Ventilator

General Dimensions:

User Interface: 350 wide x 55 deep x 244 high (mm)

Ventilation Delivery Unit: 322 wide x 375 deep x 366 high (mm)

System with Cart (optional): 547 wide x 675 deep x 950 high (mm)

Weight: Total: 40kg

(Approximate) User Interface: 2.5kg

Ventilation Delivery Unit: 12.5kg

14-1

Page 128: V8600 Critical Care Ventilator Service Manual

14-2

Cart: 25kg

Trigger Method: Flow and Pressure

Maximum Limited pressure: 80 cmH2O

Maximum Working pressure: 80cmH2O

Gas Supply

Supplied gases must be free of water, oil and particles

Inlet Gas Pressure

O2: 280kPa to 600kPa (41– 87PSI)

Connection Standards Available

DISS, NIST

Patient System Connectors

Male 22 mm Conical Fittings in accordance with ISO 5356-1

User Interface

Attaches to the cart, a rail, or other mounting system.

Standard Conditions Specifications

Error ranges in this document assume the following standard conditions:

Ambient pressure: 101.3 kPa (1 atmosphere)

Room temperature: 20 ℃

Dry gases in patient system

Inlet pressure: 345 kPa (50 PSI)

Inspiratory Channel

Pressure Drop

Maximum 5 cmH2O at a flow of 60 liters/min without CO2 airway adapter.

Compliance

Maximum 2 ml/cmH2O (with Fisher &Paykel MR 810 humidifier and reusable silicone adult patient circuit)

Gas Delivery System

Microprocessor controlled valves

Gas Delivery Device

Flow Range:

Page 129: V8600 Critical Care Ventilator Service Manual

14 Appendix C

1 to 180 liters/min. (Adult)

0 to 60 liters/min. (Child)

Maximum Pressure Setting: 70 cmH2O

NIV Max leakage compensation level

Adult: 60liters/min

Child: 30liters/min

Expiratory Channel

Pressure Drop (Resistance)

Maximum 5 cmH2O at a flow of 60 liters/min without CO2 airway adapter.

Compliance

Maximum 2 mL/cmH2O (with Fisher & Paykel MR 810 humidifier and reusable silicone adult patient circuit)

PEEP Regulation

Microprocessor controlled valve

PEEP Setting Range: 0 – 35 cmH2O

Expiratory Flow Measurements

Range: 0 – 180 liters/min.

Monitoring

Expiratory Minute Volume

Range: 0 – 60 liters/min.

Accuracy: +/- 1 LPM or +/- 15% of measured value (whichever is greater)

Resolution: 0.1 liters > 1 liter/min, 0.001 liters < 1 liter/min

Expiratory Tidal Volume

Range: 0 – 4000 mL

Adult Accuracy: +/- 25 mL or ±15% of the measured value (whichever is greater)

Child Accuracy: +/- 10 mL or ±10% of the measured value (whichever is greater)

Resolution: 1 mL

O2 Concentration

Range: 18 – 100%

Accuracy: +/- 3 vol. %

14-3

Page 130: V8600 Critical Care Ventilator Service Manual

14-4

Resolution: 1%

Airway Pressure

Range: -20 to 80 cmH2O

Accuracy: ± (2 cmH2O + 4% of reading)

Resolution: 1 cmH2O

Page 131: V8600 Critical Care Ventilator Service Manual

15 Appendix D

15 Appendix D Spare parts and configurations

Spare parts list: No. Code Description

Replacement period

Remarks

1 122007310 NBP mother board As needed ——

2 122007306 NBP power supply board As needed ——

3 122007308 NBP main control board As needed ——

4 122007309 NBP core board As needed ——

5 122007311 NBP interface board As needed ——

6 122005527 Gas supply pressure board

As needed ——

7 122007482 Encoder knob component As needed ——

8 122007483 Mute button component As needed ——

9 122007304 NBP display interface board

As needed ——

10 122007303 Common display platform core board

As needed ——

11 122007305 Common display platform mother board

As needed ——

12 122007485 Alarm lamp component As needed ——

13 122007486 NBP indicate lamp component

As needed ——

14 122010401 Differential pressure transmitter board As needed ——

15 210003677 Turbine As needed ——

16 210002403 Flower sensor TSI 840201

As needed ——

17 122007463 Inspiration control valve As needed ——

18 240000089 Check valve As needed ——

19 240000252 Electromagnetic valve SY114-6G

As needed ——

20 240001020 Mini-electromagnetic valve N332/2B As needed ——

21 210003856 High rupture fuse T250V/3.15A As needed ——

15-1

Page 132: V8600 Critical Care Ventilator Service Manual

15-2

No. Code Description Replacement

period Remarks

22 210003676 Brushless motor driver As needed ——

23 210003693 Touch panel AMT9542 As needed ——

24 210003675 LCD As needed ——

25 122007471 Exhalation valve components

As needed ——

26 122007890 Exhalation valve core assemble

As needed ——

27 210003491 Switch power supply SNP-G169-M

As needed ——

28 210003678 Switch power supply As needed

29 122007322 Fan cable As needed —— Consumables list:

No. Code Description Replacement Period Remarks

1 130000231 Valve 1 years or as needed ——

2 130001347 One-way diaphragm 1 years or as needed ——

3 130009651 Water trap 2 years or as needed ——

4 130010423 Fan filter cotton 2 years or as needed ——

5 210001975 O2 sensor 1 years or as needed ——

6 230000135 Filter element 2 years or as needed ——

7 210003734 Lithium-ion battery RC2024 2 years or as needed ——

8 130004358 Filter element 3 months or as needed ——

9 130003930 Filter 2 years or as needed ——

Page 133: V8600 Critical Care Ventilator Service Manual

This manual No.: 130015826

Directive 93/42/EEC concerning Medical Devices

CE mark in this manual apply only to product with CE mark.

Edition 01.04 May. 2018