13
VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng and Brett Pugh Boeing Satellite Systems El Segundo, California Paul Wilkin, Chetan Sayal and Don Chu Swales Aerospace Beltsville, Maryland ABSTRACT Comparison of GOES instrument thermal model predictions with on-orbit data shows that the models capture the observed temperature and misalignment trends. Lack of precise knowledge as to spacecraft pointing precludes such comparison with instrument pointing predictions. Based on the models, thermally induced instrument attitude variation will dominate GOES N-Q Image Motion Compensation (IMC). Errors due to day-to-day changes in the attitude profiles are predicted to be under 10 microradians except for rapid scans where disturbances may reach 30 microradians. THE VALIDATION PROBLEM The ITT Industries Imager and Sounder are scanning multi-channel imaging instruments flown on the Geostationary Operational Environmental Satellites (GOES). To ensure that the line-of-sight points in the desired direction, nominal scan mirror orientation is adjusted to compensate for predicted instrument pointing and gimbal misalignment errors. This is called Image Motion Compensation (IMC) and ideally avoids the need for image adjustments on the ground. When there is a problem, it usually comes about because the pointing and misalignment predictions used to compute IMC are not correct. On the current GOES I-M momentum-bias spacecraft built by Space Systems Loral (SSL), pointing errors come not only from instrument thermal deformations but also from the Earth sensors used to control spacecraft attitude. Upcoming GOES N-Q three-axis-stabilized spacecraft being built by Boeing Satellite Systems (BSS) control spacecraft attitude using star trackers and gyros. This approach is expected to leave instrument thermal deformation as the primary source of pointing error. The archive of GOES IMC sets provides instrument pointing and misalignment profiles for every day of the year. Although instrument misalignment is spacecraft-independent, instrument pointing includes spacecraft attitude. Obtaining pure instrument pointing from the IMC set requires precise knowledge of spacecraft attitude. Unfortunately, it is the instruments themselves that provide the most accurate observations of spacecraft attitude, and no way has been found to distinguish between instrument pointing and spacecraft attitude effects. 1 This work was supported under NASA contract NAS5-01090. 203 https://ntrs.nasa.gov/search.jsp?R=20010084974 2020-05-02T04:59:48+00:00Z

VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

  • Upload
    others

  • View
    10

  • Download
    0

Embed Size (px)

Citation preview

Page 1: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

VALIDATING GOESINSTRUMENTTHERMAL DEFORMATIONS1

Peter Harter and Benny GhaffarianITT Industries

Fort Wayne, Indiana

Ray Ng and Brett Pugh

Boeing Satellite Systems

El Segundo, California

Paul Wilkin, Chetan Sayal and Don Chu

Swales Aerospace

Beltsville, Maryland

ABSTRACT

Comparison of GOES instrument thermal model predictions with on-orbit data shows that the models

capture the observed temperature and misalignment trends. Lack of precise knowledge as to spacecraft

pointing precludes such comparison with instrument pointing predictions. Based on the models, thermallyinduced instrument attitude variation will dominate GOES N-Q Image Motion Compensation (IMC).

Errors due to day-to-day changes in the attitude profiles are predicted to be under 10 microradians exceptfor rapid scans where disturbances may reach 30 microradians.

THE VALIDATION PROBLEM

The ITT Industries Imager and Sounder are scanning multi-channel imaging instruments flown on the

Geostationary Operational Environmental Satellites (GOES). To ensure that the line-of-sight points in the

desired direction, nominal scan mirror orientation is adjusted to compensate for predicted instrument

pointing and gimbal misalignment errors. This is called Image Motion Compensation (IMC) and ideally

avoids the need for image adjustments on the ground. When there is a problem, it usually comes about

because the pointing and misalignment predictions used to compute IMC are not correct.

On the current GOES I-M momentum-bias spacecraft built by Space Systems Loral (SSL), pointing

errors come not only from instrument thermal deformations but also from the Earth sensors used to control

spacecraft attitude. Upcoming GOES N-Q three-axis-stabilized spacecraft being built by Boeing Satellite

Systems (BSS) control spacecraft attitude using star trackers and gyros. This approach is expected to leave

instrument thermal deformation as the primary source of pointing error.

The archive of GOES IMC sets provides instrument pointing and misalignment profiles for every day

of the year. Although instrument misalignment is spacecraft-independent, instrument pointing includes

spacecraft attitude. Obtaining pure instrument pointing from the IMC set requires precise knowledge of

spacecraft attitude. Unfortunately, it is the instruments themselves that provide the most accurate

observations of spacecraft attitude, and no way has been found to distinguish between instrument pointingand spacecraft attitude effects.

1 This work was supported under NASA contract NAS5-01090.

203

https://ntrs.nasa.gov/search.jsp?R=20010084974 2020-05-02T04:59:48+00:00Z

Page 2: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

If therewereawaytopropagatespacecraftpointingeitherkinematicallyordynamically,it wouldbepossibletoremovespacecraftpointingfromtheIMCpointing.Propagationaccuracy,however,wouldhavetobeontheorderofone-tenththemaximumIMCorabout100microradiansover24hours.UnfortunatelyGOESI-M gyropropagationisnotthisaccurate.GOESI-Mdoesproviderelativelyaccurateactuatortelemetry,but100microradiansoutofthedaily360°pitchrotationisonlyonepartinfiftythousand,andknowledgeofthespacecraftinertiaisnotthataccurate.

Todetermineinstrumentpointingandmisalignment,ITT,SSLandBSSdevelopeddetailedthermal,structuralandopticalmodelsfortheinstruments.Thethermalmodelsimulateselectronicandsolarheatingplusreradiationandthencomputestemperaturesathundredsofinstrumentpoints.Fromthesetemperatures,theexpansionofinstrumentpartsiscomputed.Thenewdimensionsarefedintoastructuralmodelthatcomputesthemovementofthousandsofinstrumentpoints.Deflectionsandrotationsof fourcriticalopticalpointsarethenextractedandtransformedtopointingandmisalignmenterrors.

INSTRUMENT MODELS

The Imager and Sounder are similar in construction. The optical components consist of a flat scanning

mirror, a Cassegrain telescope made up of parabolic primary and secondary mirrors plus a detector array.

These are housed in an aluminum box having an optical port for incoming light, cooling louvers above the

scan mirror and radiant coolers above the detector array. Light from the Earth enters the optical port, is

reflected at the scan mirror and enters the telescope and detector. The baseplate at the back of the

instruments is attached to the spacecraft and is heated to keep its temperature above 12 °C. Figure 1 showsthe structure of the two instruments.

The instruments are insulated everywhere except over the optical port, radiant cooler and louver. At

midnight (local solar time), the optical points toward the Sun. This is the time of greatest thermal loadingand deformation. For much of the year, the Sun crosses the instrument field-of-view and around the

equinoxes is eclipsed by the Earth. This intense heating interrupted by sudden cooling causes the most

rapid thermal pointing and misalignment disturbances of all. Because Sun shining into the coolers reduces

their effectiveness, 180 ° yaw maneuvers are planned for equinox to keep the coolers pointing away fromthe Sun.

Figure 1. Imager and Sounder Structure

Fadia_t

++_tcrWhee_%4++":'".+_%>Li+a+ia.," ?_

%£_ti+m " "+"

C_I_

Pmc'h

For purposes of Image Navigation and Registration (INR), the instruments are characterized by threepointing and two intemal misalignment angles which are collectively called the instrument attitude. The

roll (qb),pitch (0) and yaw (_g) pointing angles are defined with respect to the orbital coordinate system x-,y- and z-axes respectively. In the upright orientation, orbital coordinates coincide with nominal instrument

204

Page 3: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

coordinates.Intheinvertedorientation, the instrument x- and y-axes coincide with the orbital minus x-

and minus y-axes. This means that the same deformations in instrument coordinates imply opposite roll

and pitch pointing in the upright and inverted orientations.

Misalignments are those of the outer scan mirror gimbal axis with respect to the Cassegrain telescope

axis. Roll misalignment (d_m. )and pitch misalignment (0ma) are not tied to body coordinates as might be

expected but are yaw-dependent. This is done to make the misalignments correspond to roll and pitch in

either yaw orientation. Neither are the polarities the same for Imager and Sounder. This is done to make

the upright Imager look like an inverted Sounder. The directions of positive misalignment rotation areindicated in Table 1.

Table 1. Misalignment Sign Conventions

Upright Inverted

Imager Cma -zaxis z_axis

0ma -y_axis y_axis

Sounder Cma z_axis -z_axis

0ma -y_axis y_axis

Instrument observations are east-west (E) and north-south (N) scan angles derived from the scan

mirror inner and outer gimbal angles c_ and q. The zero position for cxis such that the mirror normal

makes a 45 °. angle with the outer gimbal axis. The zero position for r 1 is such that the mirror normal lies in

the x-z plane. Unlike the misalignment angles, these angles are fixed in instrument coordinates along the

positive body y- and x-axes respectively. This is Figure 2 shows this schematic INR model for the twoinstruments.

Figure 2. INR Instrument Model

z

Upright Orientation l Y

I /_'_'_ Sounder

telescope 1"1 /_/ rl _ _ X, ,telescope

Inverted Orientation

x

telescope_._)c rI _ Imager

Sounder _ ytelescope

205

Page 4: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

Themagnitudeofthe1]anglecorrespondsapproximatelytothenorth-southscanangle(N),butthemagnitudeofthee_angleisapproximatelyhalftheeast-westscanangle(E).Becausethegimbalanglesaredefinedwithrespecttoinstrumentcoordinates,however,thesignschangewithyaworientation.Ifyfequalto+1indicatestheuprightorientationand-1 indicatestheinvertedorientationandis equal to +1 indicates

the Imager while is equal to -1 indicates the Sounder, the dependence of scan angle errors on pointing and

misalignment for all four cases can be represented by the following pair of equations

AE_(O-C N -S N -is.yf.S u 0).(_b, 0 n g/ _b,,a 0,,a) r (1)f

AN_(-1 -SNT E CNT _ 1 CN SN (SE+is.yf)].(f/) n O. _ _m. 0._) r (2)

\

\ CE CE J

Here, C and S denote the cosine and sine of the subscript angles, and d_nand 0, are the modified roll and

pitch used in the GOES Orbit and Attitude Tracking System (OATS)

(/)_= q_+ q_ma (3)

On =-0 + Om_ (4)

Although the thermal and structural models predict deformations for thousands of instrument points,

the optical model requires the three displacements and three rotations of only four points [ 1, 2, 3, 4, 5]. As

shown in Figure 3, these four points are the centers of the scan mirror, primary mirror, secondary mirror

and detector. Their twenty-four coordinates are multiplied by an Optical Sensitivity Matrix (OSM) that is

specific for each instrument in each of the two possible yaw orientations. No deformations of the opticalcomponents are considered.

Figure 3. Four-Point Optical Model

scan mirror

image plane

se:°;do7

primarymirror

GOES I-M MODEL VALIDATION

Over the years since 1987 when the ITT instrument model was first developed, it has been repeatedly

compared to ground test or on-orbit data and against general purpose modeling software. Temperature

predictions have been checked to test the thermal model. Natural vibration frequency predictions have

206

Page 5: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

beencheckedtotestthestructuralmodel.Pointingandmisalignmentpredictionshavebeencheckedtotestthe optical model. When necessary, the model has been corrected or enhanced.

Thermal modeling is the first step in the simulation process, and predicted temperatures have been

compared to both ground test and on-orbit telemetry data. In 1994, Harter showed that temperature

predictions matched test data and predicted 1NR performance [6]. In 1995, Zurnaehly showed that

predicted temperatures matched GOES-8 on-orbit temperatures [7]. In 1996, Ghaffarian and Sprunger

predicted that secondary mirror temperatures would exceed operating limits and verified their predictionswith GOES-8 thermistor data [8].

Structural models relate temperatures to deformations. In 1997, Harter validated the structural model

by successfully predicting instrument natural vibration frequencies [9]. He also showed that uniform

temperature gives minute pointing errors as expected. In 1998, he identified the contributions of various

instrument sub-assemblies by setting coefficients of thermal expansion to zero for all but the components

under consideration [10]. In this way, Harter showed that roughly three-fifths oflNR errors came from

deformations of the instrument housing and one third came from deformations of the scan assembly.

The optical model transforms the three translations and three rotations for each of the four optical

points into three pointing and two misalignment angles. Predictions have been compared with on-orbit

IMC sets by Harter in 1991, Walker in 1996, Hampton in 1997, and Harter in 1997, but agreement was

weak due to the overriding effect of Earth sensor errors [11, 12, 13, 14]. From the 1997 study, Harter

found that structural translations and rotations corresponding to an instrument rigid body rotation wrongly

produced optical internal misalignments. He also discovered that the OSM predicted results of the wrongsign for the Sounder. In 1999, Harter revised the Imager OSM and created a distinct OSM for the Sounder

[15]. In 2000, Harter and Wickholm showed that the corrected OSM matched results obtained with the

Code V optics modeling program [ 16].

GOES instrument thermal models are important for day-to-day operations as well as 1NR prediction.The instrument is susceptible to overheating, and operators avoid scanning close to the Sun rather than risk

damage. The cost of this caution is lost images. So, there is motivation to predict temperatures as

accurately as possible. Current models capture trends and actual temperatures within several degrees.

Figure 5 shows typical agreement between predicted and observed Imager secondary mirror temperaturesfor GOES-10 at summer solstice.

Figure 5. Predicted and Observed Mirror Temperatures

Imagez Secondary ikrttrror Te_a'atme Stmm'ler Solstice EOL

25

_p._ On-°rbita_a Secona_ I

_¢t_ _ 7001

Io !

0 3 6 9 12 15 18 21 24

Time l-k

207

Page 6: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

Unfortunately,therearenostraingaugesor other devices to measure deformations on-board. So,

structural predictions cannot be checked. The next level of on-orbit instrument validation possible is thatof instrument attitude. As mentioned earlier, attitude is available from the GOES I-M IMC sets but is not

ideal. Pointing includes Earth sensor errors, and misalignment observability is often poor. To minimize

the effect of day-to-day variability, the IMC values in the following plots were averaged over fifteen days.

Figure 5 shows predicted and observed misalignments for winter solstice. The pointing curves show

little agreement. When the shapes of the predicted and observed misalignment curves are compared, they

are almost identical. Roll misalignment values are within the uncertainty of the IMC profiles. The larger

pitch misalignment bias may be due to unmodeled effects either in the thermal or INR models. Lower

bounds on the pitch misalignment estimate standard deviation are l0 microradians, and the misalignments

are not among the most highly correlated of the solved-for state variables.

50

0

-50

_, -lOOE

-150

-200

Figure 5. Predicted and Observed Misalignment

-250

roll ma pitch ma

.................. .....; "_..\...., \\'\. ///" ...........................

15 20 25 30 35

200

0

-200

.400

-600

-800

-1000

•.////"". \.

/....

15 20 25 30 35

5O

0

-50

o_ -100

-150

-200

GOES-8 1999 days 348 to 362

7".- ... '"-.. / "/'/ ....................t

15 20 25 30 35

local solar time (hours)

10-Apr-2001400

200

0

-200

-40O

-600

.600

-1000

//. .................

\ . ./

15 20 25 30 35

local solar time (hours)

GOES N-Q MODEL VALIDATION

The preceding comparisons with on-orbit data were made against the GOES I-M model, but it is

primarily the GOES N-Q models that are of interest now. Without GOES N-Q temperatures or IMC sets

for comparison, the N-Q models can still be checked for reasonableness. The N-Q models are qualitatively

similar to those for GOES I-M. The primary differences are the attitude stability of the spacecraft and the

instrument mounting. Thin metal strips called flexures attach the instrument to the bench. This holds the

instrument in place but allows it more freedom to expand and contract. As shown in Figure 7, theinstrument is mounted using six flexures whose normal vectors intersect at the center of the instrument

footprint.

2O8

Page 7: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

Figure 7. Flexure Mounting

I -.., ,,.,.r.'..--':"I'x°r'7" ---W --\'lax°reI

I .....ii: :_i..............-.\ ,,o_ore /".........Ir i _'-,I

Harter identified flexures in 1998 as a way to reduce pointing and misalignment errors without

redesigning the instrument [11]. As shown in Figure 8, the GOES N-Q predictions are generally smallerthan those for GOES I-M. Pointing improvement may be due in part to the GOES N-Q bus attitude

stability, but misalignment improvement is due to the flexures. Pitch misalignment is greatly reduced

while roll misalignment variation is only slightly reduced. The results for summer and equinox are similar.

Figure 8. GOES I-M / N-Q Model Comparison

roll

200 200

_ /_ -100 ..400

-200 -600

-300 -80015 20 25 30 35

pitch yaw

400,

:!

i"" i

/ i

/ i .::::

15 20 25 30 35

3OO

200 ,'" i

100 !

i ::"" ""

-100

-20015 20 25 30 35

-5O

-10C

-15C

-200

roll ma

.._...,.\.. . . .

\?,

15 20 25 30 35

yaw

200,

,oot

-400) • . . _ -'-15 20 25 30 35

upright

0 600

-5o i 4oo./ i ......

-150 -2001 '_/:

-400

-250 -60015 20 25 30 35 15 20 25 30 35

Ist (hours) Ist (hours)

winter

300 20 rI

I 0 r,..250

/", ,. :/....I

'°°t /: _ -4° 'i /, [: ':i _o J

so_ -8o

0 -100

.;:°o :?15 20 25 30 35 15 20 25 30 35

Ist (hours) Ist (hours)

10-Apr-23ol

400

100

0 _

.,co !./

-20015 20 25 30 35

Ist (hours)

The strange shape of some of the GOES N-Q predicted curves, particularly the double peaks, raised

concem that the GOES N-Q model might not be consistent with real instrument behavior. To resolve that

question, GOES I-M IMC sets were checked to see if double peaks had been seen in operations and

temperature predictions were checked for anything that could cause the double peaks.. In addition, specialsimulations were run holding the spacecraft at a uniform temperature and holding the instrument at a

uniform temperature in order to isolate the contributions of the spacecraft alone and the instrument alone.

209

Page 8: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

Checking IMC sets showed some days with and some without double peaks. That day-to-day

variability may have been due to Earth sensor and estimation errors. So, other verification was still

necessary. Given the success of the thermal model in capturing thermal variations, baseplate heater power

and instrument temperatures were examined for features coincident with instrument attitude variations.

The instrument temperatures checked were those of the baseplate, optical port sunshade, scan mirror

gussets, north panel, primary and scan mirror. What was found were that the jogs in the 1NR profiles did

correspond to instrument thermal events and that reradiation or backloading from the optical port sunshade

was a significant source of heating. Figure 9 shows one such plot of roll overplotted on top of baseplate

heater power [17].

Figure 9. Roll and Baseplate Heater Power

50 20L L L____

Ill IMAGER BASEPLATE (TOTAL)

+IMAGER BASEPLATE TEMP #3 TELE N (10023245 --

-- -- -- Rx (ROLL) Pointing Enor 2

_'35 %.

30 "

m

25 / • II

= I. ;2O

= ]=

E 15

/ [ TemperatureBaseplatem 10 /

/I J

5 m

-20

/',j-jj A

'\ , L/ .oo

/ = ¢u

r -100I U

a= _ -120

. _, /-14 0

m

/ • -160

• \,...,

• i

-180

0 4 8 12 16 20 24 28 32 36

Time ( Hours )

The special simulations to separate instrument and spacecraft effects also suggested that the double

peaks were due to the instrument. Figure 10 shows the original combined attitude profiles plus those for

the instrument alone and spacecraft alone. The prediction for the instrument alone closely follows that for

the spacecraft and instrument combination. Due to the spacecraft stability, the spacecraft-only profile is a

small fraction of the combination. Also as expected, spacecraft thermal variation contributes very little toinstrument misalignment.

Underlying the interpretation of these simulations as the effect due to spacecraft alone and that due to

the instrument alone is the assumption that the combined profile equals the sum of the individual profiles,

i.e. that the spacecraft and instrument effects are independent of each other. Conceivably, there could be

interactions between components of the spacecraft and instrument attitudes that would cause the individual

profiles not to add up. When summed together, however, the spacecraft-only plus instrument-only profiles

do match the combined profiles within one microradian. This agreement lends credence to the

interpretation of the results as being spacecraft-only and instrument-only effects.

210

Page 9: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

o

-50

-lOO

-150

-200

Figure 10. Spacecraft versus Instrument Effects

roll

-50

-100

-150

-200

-25015 20 25 30 35

pitch50

1 _/ -20

-40

_0

-80

-100

-12015 20 25 30 35

yaw4O 0

20_0 ', -50

"J'_ { ....

il -100-150

-20015 20 25 30 35

roll ma pitch ma50r

15 20 25 3O 35

-5oL

-lOOL

-1sol

-200115 20 25 30 35

_nter upright

-o_E O.Otlu NIIII I I 0.50 0.50_).5 4).5 _].5

-1 -1 -115 20 25 30 35 15 20 25 30 35 15 20 25 30 35

Ist (hours) Ist (hours) Ist (hours)

12-Apr-20011 . 1

:IHI O' i it0.5

0

4) _).5

-115 20 25 30 35 15 20 25 30 35

Ist (hoUrs) Ist (hours)

The variation of pointing and misalignment with season also provides insight into the thermal behavior

of the instruments. Figure 11 shows predicted pointing for winter, spring and summer in the upright yaw

orientation. At midnight, the summer Sun illuminates the north face of the instruments, shines into the

louvers and heats the scan cavity. This causes the large long-lasting excursions that dominate the roll and

yaw profiles. In contrast, the winter Sun illuminates the south face of the instruments and does not shine

into the louvers. It does not cause the same large deformations at midnight. As expected, the equinox case

is intermediate between the winter and summer cases for most of the day. At midnight, however, the Earth

blocks the Sun, and the instruments cool down rapidly pushing the equinox profile in the winter direction.

Figure 11. Predicted Pointing for Different Seasons

roll5O

_0............. '.\ '7 _"......

-15O _i X,/

-_._ _ /

_ eq_nox 1

. summer _ !

_- N 25 3o 35

q

-100 <-150

-2_

-250 L

-300

-3_15

pitch yaw5O

-10O _

-150 _ f

-2OO\ /

-25O

5O

0

_0

-1oo

-15o

-20O

-250

-3OO

..........1,....,L¢I...//

/

15 2O 25 30 35

local solartime (h)

withs/c4_

3O

-1C_15 2O 25 5O 35

I_al solartime (h)

30-Mar-2o01

;zoo,

/,'-

5o ....._--,/

13 ',

_oIt5 2O 25 3O 3S

local solartime (h)

211

Page 10: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

The dominant effect of Sun coming into the louvers suggests that summer in the upright orientation

should look more like winter in the inverted orientation than summer in the upright orientation. This is

borne out in Figure 12 where the curves in the second row of plots are more similar than those in the first

row. The differences between the summer upright and winter inverted curves may be due to the fact that

the Sun travels in different directions with respect to the instrument and also to the greater distance from

the Sun during summer.

50

0

-5O

-100

,_= -150

-200

-250

-300

-35C

5O

0

-50

_o) -100.=

E-150

E -200

-250

-300

-35C

Figure 12.

roll

?

_nterupnght

[_ _nter in_aedJ

TS

15 20 25 30 35

N_ model

-- . !. {

Summer Upright _ Winter Inverted

pitch150.

.',.

lOO_ :50| /

-5o

-100 /"

-150

-200

-250

5C

C

-5C

-10C

-15C

-200

-25(]

yaw

i /

15 20 25 30 35

with sic

150.

IOOL !

5oL / :u f

-50 "

-100 _ "'

-150 ,/-200

-2501 ""

11-Apr-200150

ol

-50

-10(1

-150

-200

-25015 20 25 30 35 15 20 25 30 35 15 20 25 30 35

Ist (hours) Ist (hours) Ist (hours)

INR PREDICTIONS

To ensure that GOES N-Q meets its 1NR requirements, day-to-day variations in the thermal profiles

are specified to be under 10 microradians. The thermal model provides a means of predicting whether or

not this requirement will be met. By fitting the day 1 profile to a Fourier series and comparing that fit to

the raw profile for day 2, one can predict the day-to-day 1NR error. This error depends on season because

the thermal profiles depend on season. Vernal and autumnal equinox are expected to be the worst cases,

but special short span IMC sets will be used over the eclipse periods. The only case considered here is that

of winter solstice which is in the normal season for upright yaw.

Rather than fit points at the same thne as the next day's "observations", day 1 points were first

interpolated to uniform 15 minute intervals staggered 7.5 minutes from the original points. Then the

interpolated values were fit to Fourier series with the recommended [12 15 8 8 8] fit orders for the roll,pitch, yaw, roll misalignment and pitch misalignment. This was considered a better simulation of the

random landmark observation spacing encountered in operations. As shown in Figure 13, fit errors range

from 3 microradians for roll to 10 microradians for more jagged roll misalignment.

In addition to solar heating, another source of nonrepeatability is the rapid scan mode of operation used to

image severe storms. Rapid scanning may heat and deform the instrument on one day but not necessarily

the next day. To assess the impact on repeatability, rapid scans were simulated for the Imager on day 2 butnot for the Sounder.

212

Page 11: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

.4C

_C

_C

-10C

-12C

-14C

-166

-18C

Figure 13. Curve Fit Error

roll (12) pitch (15) yaw (8) roll ma (8)

50 r .40 . . O

Imager

_50

lOO/

150l200

-L__250L15 20 25 5

40,

2ol

Ot

-20

.40L

_o _ _

_60!i/ _6°-100

_0_

-I00[-120

_/_ -1OO

-120

-160-140 I[. -150

-180 - - V . -20015 20 25 30 35 15 20 25 30 35

no rapid scan winter

-5 5 _ _ '52'0 3'035

pitch ma (8)

[ " .

/\//

/

upright10 8

5 4

2

, -2

-5 -4

-1015 20 25 30 35

Ist (hours)

15 20 25 30 35

09-Apr-2001

15 20 25 30 35

Ist (hours) Ist (hours) Ist (hours) Ist (hours)

During rapid scanning, the servo motor generates more heat than usual. This alters the thermaldeformation profiles and causes errors in the profiles predicted from the previous day. Figure 14 shows theerror for the Imager which performs two rapid scans on day 2 and for the Sounder which does none. Theerror is greater at noon when the scan cavity is otherwise cool than at midnight when sunlight enters theoptical port. Without rapid scans, repeatability differences are under one microradian (Sounder). Withrapid scans (Imager), differences reach 20 microradians in pitch and 8 microradians in pitch misalignment.This exceeds the specification for day-to-day variation and requires special attention.

roll

1

!!i0.60,4

02

0 20 2 35

_hnter

1 0

O.E -0.2e4

0.4 4).6

to 0.2 -0.8

C -1lfi 20 25 30 35

Ist (hours)

Figure 14. Repeatability

pitch yaw

20 i 1 5

15 ! 1

10 05

i' o

-5 5 20 25 30 35 5

roll ma

2 _

1

!! !!!

° 110 •

15 20 25 3(] 35

pitch ma

10

't

-2 15 20'2S go 65

upnght

15 20 25 30 35 15 20 25 30 35

Ist(hours) Ist(houm)

08. <12

0.6. 4) I

0.4. 4) 3

0.2, 4) 3

0 15 20 25 30 35 .I

Ist (hours)

11-Apt-2001

15 i0 25 30 35

Ist (hours)

213

Page 12: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

Although instrument pointing and misalignment have theft own requirements, navigation error is the

bottom line for INR. Scan angle errors can be computed using the equations (1-4) given earlier from

instrument attitude and the scan angles themselves. Roll, pitch, roll misalignment and pitch misalignment

effects depend weakly on scan angle, but yaw effects increase from zero at nadir to a maximum at the edge

of the field of view. For a point on the Earth limb 8.3 ° to the north and east of nadir, the east-west, north-

south angle and rss errors are as shown in Figure 15. Overall navigation error may be computed as the root

sum square of the east-west and north-south scan angle errors. With rapid scans, navigation error reaches

30 microradians. Without rapid scans, navigation error is only 10 microradians.

Figure 15. Navigation Errors

30

o_20

xJ

lO

E- 0

-10

east-west error north-south error root sum square error40 10, 35,

-lO[

3ol

251

201

15lI

lO[

5t

o[15 20 25 30 35 15 20 25 30 35 15 20 25 30 35

upright8 10r 12

_ 8

o o4

_4 -5t

co 2-6

-8 -101 015 20 25 30 35 15 20

local solar time (hours)

winter 11-Apt-2001

15 20 25 30 35 25 30 35

local solar time (hours) local solar time (hours)

CONCLUSIONS

The GOES I-M Imager and Sounder thermal, structural and optical models have been shown to agree

with on-orbit data. The GOES N-Q instrument models are derived from those for GOES I-M but predict

smaller pointing and misalignment errors due to improved spacecraft attitude stability and stress-relieving

instrument flexure mounts. In the absence of on-orbit GOES N-Q data for comparison, the instrumentmodels have been shown to be reasonable and self-consistent.

The GOES N-Q models predict that the instrument itself will be the primary source of pointing and

misalignment errors. In the absence of rapid scans, day-to-day pointing and misalignment repeatability are

predicted to be 1 microradian at winter solstice. Curve fitting these profiles with the planned [12 15 8 8

8] order Fourier series introduces additional error on the order of 8 microradians. Rapid scanning heats the

instruments and causes deformations that add 20 microradians to the nonrepeatable error. The resultingroot sum square of the east-west and north-south errors is 30 microradians which is of concern.

214

Page 13: VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1€¦ · VALIDATING GOESINSTRUMENT THERMAL DEFORMATIONS1 Peter Harter and Benny Ghaffarian ITT Industries Fort Wayne, Indiana Ray Ng

REFERENCES

I. Roy, N. and W. Haile, "Optical Ray Tracing in Finite Element Models", Advanced TechnologyOptical Telescopes II, SPIE Proc. Vol. 444, 1983

2. "Thermal/Structural Analysis for GOES Image Navigation/Registration", ITT Rept MD-1053,September 14, 1988, by J. Monirian and P. Harter

3. "Thermal/Structural Analysis for GOES Image Navigation/Registration Instrument-Instrument Cross-

over Effect", ITT Rept MD-1127, February 20, 1989, by P. Harter and J.Monirian

4. Memo, "GOES One-Model Report", September 19, 1994, E. Zurmehly to C. Young

5. "GOES 1NR Prediction Evaluation", S-GOES-NSH-99-0140, September 28, 1999, by N. Hodgman

6. "Analysis of GOES Image Navigation/Registration Performance with Test-Correlated Temperatures",

ITT Rept MD-1546, February 21, 1994, by P. Harter

7. Memo, "Correlation of the GOES Instrument Thermal Model to GOES-8 Flight Data", 26 April 1995,E. Zurmehly to C. Young

8. "Solar Intrusion Thermal Analysis", SPIE Vol. 2812, pp. 251-259, 1996, by Benny Ghaffarian and

Kent Sprunger

9. "GOES Imager Model Correlation with Overall Damping", S-GOES-PKH-97-0044, May 9, 1997, byP. Harter

10. "Effect of Composite Structural Components on GOES-NQ INR Performance", February 6, 1998, S-

NQ-PKH-98-0011, by P. Harter

11. "Preliminary Analysis of GOES Image Navigation/Registration Performance in Response to CCN

#116, ITT Rept MD-1355, August 16, 1991, by P. Harter

12. "NAS5-96090, Systems Engineering Analysis #4, Development and Correlation of Thermal Distortion

Model", GSFC #214.2-G03670, December 10, 1996, Clelia Walker (GSFC) Correspondence to C.Cierniak (ITT)

13. "Sounder_18 Correlation", June 12, 1997, J. Hampton (Swales) to M. Clark (GSFC)

14. "Analysis of GOES INR Performance w/Second Generation Instrument Models", S-GOES-PKH-97-

0082, August 15, 1997, by P. Harter

15. "Revised Optical Sensitivity Matrix for GOES Sotmder 1NR Error Analysis", S-GOES-PKH-99-0146,

October 4, 1999, by P. Harter

16. "Verification of GOES Optical Sensitivity Matrix using Code V Program", S-GOES-PKH-00-0031,March 1, 2000, by P. Harter and D. Wickhoim

17. "Investigation of Integrated Thermal Model Effect on INR pointing Error", ITT MEA report01-0010,NOPQDCN-274, 20 Feb 2001, by B. Ghaffarian

215