BCM350N Operator’s Manual 1 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13) Table Contents Prior to beginning ………………………………………………………… 2 BCM350N Minimum system requirements ……………………………………… 3 Precautions on install and use BCM350N ………………………………………… 3 Chpater 1 General ……………………………………………………… 4 1.Introduction of BCM350N ……………………………………………… 4 1.BCM350N installation …………………………………………………… 6 Chpater 2 BCM350N Description ………………………………………… 10 2.Main window …………………………………………………………… 10 2.Patient basic registration window ……………………………… 13 2.Doctor registration widow ……………………………………………… 15 2.NST Review window …………………………………………… 15 2.Bed view setting window …………………………………………… 16 2.Language setup window …………………………………………… 16 2.Setting(Configuration) window …………………………………………… 17 2.Close ……………………………………………………………… 17 Chpater 3 How to use BCM350N ………………………………………… 18 3.How to setup language ………………………………………… 18 3.How to setup communication port ………………………………… 19 3.How to register patient basic information ………………………… 20 3.How to assign a patient to the relevant bed ………………… 22 3.How to start diagnosis …………………………………………… 24 3.How to stop diagnosis ………………………………………… 25 3.How to reset bed information …………………………………… 27 3.How to setup bed view ……………………………………………… 27 3.How to listen fetal monitor sound ………………………………………… 29 3.10 How to control fetal monitor ……………………………………………… 29 3.11 How to review existing diagnosis ………………………………………… 30 3.12 How to save to image and print ………………………………………… 33 Chpater 4 BCM350N Uninstallation ……………………………………… 38 Chpater 5 Specification ………………………………………………… 39 Appendix 1. Wired communication system ………………………………………… 40 Appendix 2. Wireless communication system ……………………………………… 42

Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 1

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Table Contents Pr ior to beginn ing ………………………………………………………… 2

BCM350N Minimum system requirements ……………………………………… 3

Precautions on install and use BCM350N ………………………………………… 3

Chpater 1 General ……………………………………………………… 4

1.1 Introduction of BCM350N ……………………………………………… 4

1.2 BCM350N installation …………………………………………………… 6

Chpater 2 BCM350N Description ………………………………………… 10

2.1 Main window …………………………………………………………… 10

2.2 Patient basic registration window ……………………………… 13

2.3 Doctor registration widow ……………………………………………… 15

2.4 NST Review window …………………………………………… 15

2.5 Bed view setting window …………………………………………… 16

2.6 Language setup window …………………………………………… 16

2.7 Setting(Configuration) window …………………………………………… 17

2.8 C lose ……………………………………………………………… 17

Chpater 3 How to use BCM350N ………………………………………… 18

3.1 How to setup language ………………………………………… 18

3.2 How to setup communication port ………………………………… 19

3.3 How to register patient basic information ………………………… 20

3.4 How to assign a patient to the relevant bed ………………… 22

3.5 How to start diagnosis …………………………………………… 24

3.6 How to stop diagnosis ………………………………………… 25

3.7 How to reset bed information …………………………………… 27

3.8 How to setup bed view ……………………………………………… 27

3.9 How to listen fetal monitor sound ………………………………………… 29

3.10 How to control fetal monitor ……………………………………………… 29

3.11 How to review existing diagnosis ………………………………………… 30

3.12 How to save to image and print ………………………………………… 33

Chpater 4 BCM350N Uninstallation ……………………………………… 38

Chpater 5 Specification ………………………………………………… 39

Appendix 1. Wired communication system ………………………………………… 40

Appendix 2. Wireless communication system ……………………………………… 42

Page 2: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 2

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Prior to beginning

In this document, all information for correct using and operation are

included.

If you need a help for instruction or questions, please contact Bistos Co.,

Ltd.

Bistos Co.,Ltd. and the authorized agency do not take responsibility for

the harmful cases that are happened by user’s obvious careless or

incorrect uses.

In this document, the program indicates BCM350N. And the company and

we indicate Bistos Co.,Ltd.

May 2013

Rev 01

Copyright © 2013 Bistos Co., Ltd. All rights reserved.

Information and descriptions contained in this manual are the property of

Bistos Co.,Ltd. and may not be copied, reproduced, disseminated, or

distributed without express written permission from Bistos Co.,Ltd. The

materials in this manual can contain typographic errors and be subject to

change without notice.

Printed in Korea

Page 3: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 3

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Minimum system requirements

To install BCM350 to PC, minimum system requirements are as follow.

1. CPU clock speed : 2.3 GHz

2. Memory (RAM): 4 Giga Bytes

3. Display resolution : 1024 * 768 Graphic LCD

4. Operating system : Windows XP or Window 7

5. Sound card and armored speaker

6. Ethernet adaptor or RS-232C port

Precautions on install and use BCM350N

1. When install BCM350N, a user must follow the regular install procedure. If

not, it is possible that BCM350N does not work properly.

2. Before use BCM350N, please read this manual closely. To operate some

functions what user wants, please read and follow this manual. If not, it is

possible that BCM350N does not work properly.

3. If you need a help for instruction or questions during installing or using

BCM350N, please contact Bistos Co., Ltd. or the authorized agency. Bistos

Co.,Ltd. and the authorized agency do not take responsibility for the harmful

cases that are happened by using BCM350N with problem.

Page 4: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 4

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Chapter 1 General

1.1 Instruction of BCM350N

BCM350N is an integrated program of NST(non-stress test) central monitoring

system and networking system. That is, the networking function is added to

existing NST central monitoring system. So it is possible to access to the Server

PC through the Local Area Network (LAN) and so multiple monitoring also is

possible at the same time. The concept diagram is shown as below.

<BCM350N concept diagram>

The maximum number of accessible clients at the same time is limited 8 to

reduce the server overflow.

Server mode and Client mode is selectable in [Configuration] window in the

program. In the client mode operation, the server IP address should be

registered and the relevant directory (DB and Date) of the server must be

shared to enable to remote access.

Page 5: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 5

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

The useable devices in this system are BT-300, BT-330, BT-350 and Biocare

FM-1 which are fetal monitors manufactured by Bistos Co.,Ltd.

Using PC serial port, it is possible to connect from 4 to 32 fetal monitors to this

system. Default is 8 devices in upper of [Configuration] window.

Using System Base corporation’s multi-port product named ‘Multi-8’, it is

possible to comprise wire communication (refer to Appendix 1).

Using Sena Technology corporation’s Bluetooth module named ‘MSP-1000’

and ‘Parani-SD1000’, it is possible to comprise wireless communication (refer

to Appendix 2). To comprise wireless communication, it needs to create a

virtual port in PC. To create a virtual port, it needs to install extra program. The

install method of this extra program is described in Appendix 2.

Appendix 1. Wire communication system

Appendix 2. Wireless communication system

Please run BCM350N after check out there is no problem in wire or wireless

communication system.

To run the program, select as below;

‘C:\programfiles\bcm350n’ or ‘C:\programfiles(x86)\bcm350n’

When the program installation is finished, the ‘Sort cut icon’ is created in

wallpaper. If the ‘Short cut icon’ is not created, we recommend that create

‘Short cut icon’ manually and use that icon.

Page 6: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 6

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

1.2 BCM350N installation

1. Run BCM-350N [Setup.exe].

2. Select the language for installation. (Korean / English)

Page 7: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 7

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3. Installation Wizard will come out. Click [Next] button.

4. Specify the path of the installed directory for BCM350N by clicking

[Browse…] button. Select a destination folder and click [Install] button. We

recommend that using the default path.

Page 8: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 8

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5. Please wait while the software is being installed.

If ‘Microsoft. Net Framework 3.5’ is not installed in your desktop, it is installed

automatically during the installation of BCM350N. In this case, it may take few

minutes.

6. Click [Finish] button to complete the installation. (Check the small box if

you want to start the BCM350N immediately.)

Page 9: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 9

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

7. ‘BCM350N Short cut icon’ is created in desktop.

Run BCM350N and the initial window will be shown up.

8. The initial window is set with English. If you want to change language,

please click [Language] button on the toolbar and change language what

you want.

9. Initial network of BCM350N is ‘Server’. Change network mode to ‘Client’ as

follow.

[Setting] button -> [Configuration] window -> Check ‘Run as client’ of

Communication region -> Application restart required. -> Input Server IP

And then run BCM350 again. You can find the client mode main window.

Page 10: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 10

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Chapter 2 BCM350N Description

2.1 Main window

[Figure 1] BCM350N main window

Such as [Figure 1], the main window has a toolbar. The toolbar is located in

upper region of the main window. Also, there are 8 control panels for remote

control 8 fetal monitors in the left side of the main window. And there are Patient

Information region and NST chart region on the right of the control panels.

1. Toolbar button

There are several buttons for dealing with program on the upper region of the

main window.

These buttons are listed in the below table.

Patient Registration

New patient can be registered by using the

button. If you click the button, the [Patient

Registration] window will come out.

Doctor Registration

New doctor can be registered by using the button.

If you click the button, the [Doctor Registration]

small window will come out.

Page 11: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 11

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Review

If you click the button, the [Review] window will

come out. You can review and print the saved

NST chart. and convert the NST chart to jpg type

of image file.

1,2,3,4,6,8 Bed If you click each of the buttons, you can see the

NST chart according to number of bed view.

Bed View setting

If you click the button, the [Multi Bed View

Setting] window will come out. You can select the

bed that you want to see at the relevant bed view

respectively.

Language

If you click this button, available languages are

lists on pop-up menu. You can change the

language by choosing one of them.

Setting

If you click the button, the [Configuration]

window will come out. You can change the

necessary setting (or configuration).

Close To terminate NST Program.

(1) Control panel

Every bed has its own control panel which is located on the left of the main

window. The below figure is an example control panel of “Bed 1”. The functions

are starting or stopping diagnosis, turning off fetal monitor, starting or stopping

print and disabling alarm.

The patient’s name will be displayed in bed button region(①). If the user clicks

this button, the relevant graph region will be shown up and the graph will be

displayed. Also, The FHR1, FHR2 and UC will be displayed in value button

region(②) in real time.

In the case of the program is connected to fetal monitor, the connection

indicator on the left of the bed button is changed to green from red. When the

program starts to receive diagnosis data, the connection indicator is blinking.

The follows are functional description of each button in control panel.

① ②

Page 12: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 12

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Button Name Functional description

( ) Start / Stop To start or stop NST diagnosis

( ) Sound Listen On or Off the real time fetal heart sound.

Device Printer To control the device printer

Device Power Fetal monitor power off function

Device Alarm

Sound

You can control the alarm enable and disable

function, start or stop printing function.

Assign Patient To assign the patient

Reset When clicked, the bed configuration is reset.

Connected If connected, the connection indicator is changed

to green from red. Disconnected

Printer During print, it blinks.

Saving While the patient is assigned, if diagnosis is

ongoing, It means real time saving and blinks.

(2) Patient basic profile

In this region, patient basic information is displayed. Displaying parameters are

shown as below.

※ GA : Gestational Age

(3) NST chart region

In this region, FHR1, FHR2 and UC are displayed.

Use the scroll bar If you want to see other bed’s NST chart.

Page 13: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 13

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

2.2 Patient registration window

Click [Patient Registration] button. Then ‘Patient Registration’ window will be

shown as below.

This window is composed of patient search region, patient information region,

guardian information region and medical information. Also the number of

registered patient is displayed.

[Figure 2] Patient Registration window

Page 14: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 14

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

(1) Searching existing patient.

To use this function, insert the patient’s ID or name and press ‘Enter’ key. Also

click [+] region of tree structure and search the patient directly. The tree

structure is classified between two methods, registration date and name.

(2) New patient’s basic information registration region: [New]->[OK]

To register a new patient, you can insert ID, name, SSN(Social Security number),

zip code, address, home phone number, cell phone, birthday, attending doctor

and registration date in the ‘Patient Information’ region.

Also you can insert guardian’s name, contact and relation in the ‘Guardian

Information’ region.

To register a new patient, you must insert ID and name. The other items are

optional. Then, click [OK] button to complete and save it.

(3) Medical information region :

In this region, a doctor’s comment and delivery history such as LMP, EDC, GA

and Parity information is recorded. Then, press [OK] button to complete and

save it.

※ LMP : Last Menstrual Period EDC : Estimated Day of Confinement

GA : Gestational Age Parity : The number of pregnancies

If you want to delete it, select the item of patient and click [Delete] button.

Note : Same name can be registered, but Same ID cannot be registered. (When

you try to register the patient, if there is same ID, message comes out.)

Page 15: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 15

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

2.3 Doctor registration window

Click [Doctor Registration] button. Then ‘Doctor Registration’ window will be

shown as below.

[Figure 3] Doctor Registration window

To register a new doctor, you can insert name and telephone number and press

[Save] button. If you want to delete it, select the item and use [Delete] button.

2.4 NST Review window

To review saved data, click [Review] button. Then [Review] window will be

shown as below.

[Figure 4] Review window

Page 16: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 16

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

The NST Review window is composed of Direct / Tree search region, patient

information region, history region, and graph display region. Using direct or tree

search, you can search and select the patient. If the patient is selected, the

patient’s information is displayed in the patient information region. Also saved

diagnosis list is displayed in history region. When you click one item of the

diagnosis list, the relevant graph is displayed in the graph display region. In the

Print option region, you can select the range of printing. If you want to print the

test result as soon as the test is finished, you can select the relevant function

in After test print region. The printing paper size is fixed to A4.

2.5 Bed view setting window

[Figure 5] Bed view setting window

You can select beds what you want to view at the same time in this window.

Select beds to want to see when the relevant bed view is clicked. And then click

[Save] button.

2.6 Language setup window

Click [Language] button. Then the popup menu comes out. The default

language is English. BCM350N supports English, Korean, Japanese and

Spanish. You can change another language after registering.

Page 17: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 17

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

2.7 Setting(Configuration) window

To change the configuration, click [Setting] button. Then [Configuration]

window will be shown as below.

[Figure 6] Configuration window

In this window, you can select the port to use and wave grid of graph mode.

Also you can modify the path of data base and image.

Note : You can change the path of database(DB), data, and image, but we

recommend that you hardly change them.

2.8 Close

To finish BCM350N, click [Close] button in the main window.

Page 18: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 18

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Chapter 3. How to use BCM350

After installing communication module and setting communication program, run

BCM 350N as follow path. <Start Programs BCM350N BCM350N > Then

the main window comes out as Figure 1. Operate BCM350N according to the

following procedure.

1. Language setting: Select the language what you want.

2. Port setting: Select the ports relevant to each bed and the monitor type.

(The baud rate is set automatically.) (See 6. 2.)

3. Register a new patient.

4. Select the bed what you want to connect in the control panel.

5. Move the mouse pointer to the graph display region and click on the right

mouse button. And then the popup menu comes out. Select [Patient Select]

item in the popup menu.

6. [Patient Basic Registration] window comes out. Then select the patient

using [Direct Search] or [Tree Search].

7. The selected patient’s name will be presented in the bed button region of

the control panel which is located in the left of the main window.

8. Click [Start] button in the control panel and start the diagnosis.

9. Click [Stop] button in the control panel and stop the diagnosis.

3.1 How to setup language

1 [Language] Click the Language setup button in the main

window.

2

The popup menu comes out such as left. The

present language is checked. Select the

language what you want. Default language is

English. Supporting language are Korean,

English, Japanese and Spanish. If you want to

add another language, please contact us.

Page 19: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 19

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.2 How to setup communication port

1 [Setting] Click [Setting] button in the

main window.

2 Refer to [Figure 6].

The [Configuration] window

will be shown such as

[Figure 6].

3

Click [Serial port Setup]

button.

4

The [Port Setup] popup

window comes out. Select a

bed number and select

connect serial port number.

(If the connected device to

the bed uses 1.6k protocol

such as BT-300, check

“1.6k”)

☞ When you select serial port, if the port is not connected, selectable port

may do not appears.

5 [Save] When the setting is finished,

click [Save] button.

☞ [Cancel]

If you want to exit without

change, click [Cancel]

button.

Page 20: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 20

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.3 How to register patient basic information

1 [Patient Registration] Click [Patient Registration] button in the main

window.

2 Refer to [Figure 2]. The [Patient Registration] window will be

shown such as [Figure 2].

3 [New]

If you want to register the new patient, click

[New] button. (Button to make all of the

blanks)

3

(1) Input the patient’s information. To register a new patient, you must

input the patient’s ‘ID’ and ‘name’. If not, it is impossible to register a

new patient. ‘Reg. Date’ in the Patient Information region means the date

of the first registration date.

(2) When you input all information, click [OK] button.

4

The dialogue box comes out. Click [OK]

button. Then, new patient is added.

Page 21: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 21

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

or

When the registration is

completed, the directory named

by the patient name(Name) and

registered date(Reg. Date) is

created. You can confirm the

patient’s information.

3.3.1 How to edit existing patient registration information

1 [Patient Registration]

Click [Patient Registration]

button in the main window.

2 Refer to [Figure 2].

The [Patient Registration]

window will be shown such as

[Figure 2].

4

Input the patient’s name or ID

what you want to find in the

Direct Search region and press

enter. For example, input ‘Lee’

and press enter.

☞ If there is no patient to find, any information is not appeared.

4-1

or

Also using the directory named

by the patient name(Name) and

registered date(Reg. Date), you

can search the patient what you

want to find. Refer to the

registration date and click [+]

button or patient’s name directly

and select the patient.

5

Page 22: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 22

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

In this window, edit the item what you want, and then click [OK]

button. (In this window, the Home Phone number is modified.)

6

The [Update] popup window

comes out. If you want to edit

and save, click [Yes] button.

7

When the update or edit is

completed, the dialogue box

comes out one more time.

Click [OK] button.

If you want to delete the existing patient registration information, click

[Delete] button in the [Patient Registration] window after selecting the

patient what you want.

3.4 How to assign a patient to the relevant bed.

There are three ways to assign a patient to a bed. First, click the [Assign] icon

in the control panel of left menu.

Second, move the mouse pointer to FHR grid of the graph region and click on

the right button. Then the popup menu comes out. Select <Assign Patient>.

Finally, Select a patient from [Patient Registration] window, and click [Assign

to Bed] button directly.

OK

Page 23: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 23

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

1-1

Click the [Assign] icon in the control

panel of left menu.

1-2

Click the bed button and move the

mouse pointer to the graph region and

click on the right button.

The popup menu comes out. Select

<Assign Patient>.

※ <Start>/<Stop> has same function with [Start/Stop] icon in the

control panel.

※ <Reset> has same function with [Reset] icon in the control panel.

1-3 [Patient Registration] Click [Patient Registration] button in the

main window.

2 Refer to [Figure 2]. The [Patient Registration] window will

come out such as [Figure 2].

3

Input the patient’s name or ID what you

want to find and press enter. For

example, input ‘Lee’ and press enter.

☞ If there is no patient to find, any information is not appeared.

4

Selected patient’s information is presented.

Then click [Assign to Bed] button. And this window disappears.

Page 24: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 24

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5

The [Assign Patient] popup window

comes out. Select the bed number.(For

example, select <Bed 1>.)

And click [OK] button.

6

The selected patient’s name is

presented in the control panel.

Example) ‘Bed 1’ is connected to the

patient ‘Lee’.

3.5 How to start diagnosis

1-1

Click [Start] button in the control panel.

Then the diagnosis starts.

1-2

Click the bed button and move the mouse

pointer to the graph region and click on

the right button.

The popup menu comes out. Select

<Start>.

2

If the communication port is normal, the

signal indicator changes to green from

red. The FHR and UC are displayed.

Also FHR and UC are displayed in the

graph region.

The graph of FHR and UC are displayed.

(FHR1 and FHR2: ①, UC: ②)

Page 25: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 25

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

If the communication port is abnormal, the

dialogue box such like the left comes out.

To clear this box, click [OK] button.

In this case, exit program and check the

communication port setting program.

3.5.1 Display of information during diagnosis

During diagnosis, various information is displayed in the graph region window.

The meaning are as follows.

Patient’s Information

HR1 HR2 UC Diagnosis

time Functions

HR1 FHR value from Doppler 1 probe.

HR2 FHR value from Doppler 1 probe.

UC UC value from UC probe.

Note : [Save to JPG] and [Print] buttons for function appears after diagnosis

completion.

3.6 How to stop diagnosis

1-1

Click [Stop] button in the control panel.

OK

OK

Page 26: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 26

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

1-2

Click the bed button and move the mouse

pointer to the graph region and click on

the right button.

The popup menu comes out. Select

<Stop>.

2

The [Stop] popup window comes out. If

you like to stop, click [Yes] button

3

If the diagnosis is stopped, the signal

indicator is change to red.

☞ If a patient is not assigned to the bed, the following dialogue

box comes out.

1

The [Save] popup window comes out. If

you like to save data, click [Yes] button

2

Yes

The [Patient Registration] window comes out.

If you want to save diagnosis data, assign the patient

to the bed.

[Exit]

The [Discard] popup window comes out. . If you sure

to discard the data , click [Yes] button

No Stop monitoring without saving.

☞ Even if you do not save the patient information, diagnosis data still remain.

Page 27: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 27

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.7 How to reset bed information

If you want to disconnect a bed and patient, click this button.

1-1

In the status of the diagnosis is

stopped, click the [Reset] icon in the

control panel of left menu.

1-2

Click the bed button and move the

mouse pointer to the graph region and

click on the right button.

The popup menu comes out. Select

<Reset>.

This function does not work during

diagnosis. In this case, the dialogue

box such like the left comes out. To

clear this box, click [OK] button.

3.8 How to setup bed view

When BCM350N starts, there is one bed in the graph region of the main window.

Using this button, you can setup bed view. The number of bed you can view at

the same time is 2, 3, 6 and 8. The default view is one bed view. If you want to

view several beds, you have to setup bed view first. Follow next procedure to

setup bed view.

1 [Bed View Setting] Click [Bed View Setting] button in the

main window.

2

Select the beds to display and click [Save] button.

OK

Page 28: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 28

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

If there is a checking error, If there is a

checking error, the error message window

comes out. Click [OK]. Then check the

error and re-check beds. And click [OK]

button.

3

[1 Bed], [2 Bed], [3 Bed],

[4 Bed], [6 Bed], [8 Bed]

If you click each of the buttons, you can

see the NST chart according to number of

bed view.

☞ If you want to see only one, double click the necessary one bed on the

multi view screen in the main display.

Follow next procedure to change the bed number in the control panel.

1 [Setting]

Click [Setting] button in the main window.

Then, The [Configuration] window will

come out.

2

Click [Bed Number Setting] button. And this window disappears.

OK

Page 29: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 29

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3

The [Configuration] window will come out and you can modify the bed

number you want in the control panel. To apply the setting, click [Save]

button.

3.9 How to listen fetal heart sound

BCM350N support the hearing the real time fetal heart sound function.

1

Choose the bed what you want to hear

fetal heart sound and click [Sound Listen]

button in the control panel. The other

panel’s [Sound Listen] buttons are

deactivated.

2

To deactivate this button, click again this

button. In this case, this button is

activated like a procedure 1.

☞ In the case of the [Sound Listen] button is activated, you can hear fetal

heart sound from the device or the PC speaker.

3.10 How to control fetal monitor

3.10.1 Device Printer

1

Click [Device Printer] button in the control

panel.

Page 30: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 30

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.10.2 Device Alarm Sound

3.10.3 Device Power

3.11 How to review existing diagnosis

1 [Review] Click [Review] button in the main

window.

2 Refer to [Figure 4]. The [Review] window comes out.

4

Input the patient’s name or ID

what you want to find and press

enter. For example, input ‘Lee’

and press enter.

If there is no patient to find, the

error message comes out. To

clear this message, click [OK]

button.

☞ During printing, [Device Printer] icon blinks.

1

Click [Device Alarm Sound] button in the

control panel.

☞ You can hear the alarm sound from the device or the PC speaker. Alarm

on and off can be checked on the device only.

1

If you want to turn off the device power,

Click [Device Power] button in the control

panel.

☞ This function is only available during diagnosis. It can be only set the

device off.

OK

Page 31: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 31

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

4-1

Or using Tree Search, you can

search the patient what you want

to find. Refer to the registration

date and select the patient.

5

The existing diagnosis list is

displayed in the upper right

region. Click the list what you

want to review.

6

Then the saved chart and relevant information are displayed.

You can save to JPG or print 20 minutes of the stored data you want. In

the [Review] window, there is a track bar on the displayed chart. The

position of the track bar is point to save to JPG or print.

7

Page 32: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 32

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

When you place the mouse pointer on the NST chart, time appears. At

that time, if you double click, it is possible to enter a record of the text.

After completing the input, click the [Save] button. Then text is saved.

Double click to delete the text

entered.

Click the [Yes] button.

☞ If you want to delete the existing diagnosis in upper right, press [Delete]

key after selecting diagnosis information of the patient what you want.

Note: Not only you can read and save the USB data that is stored in the device

to PC(desktop or notebook), but you can save to JPG or print it.

After connecting the USB memory stick to coputer, enter the [Review]

window and USB drive is shown on the left top.

Page 33: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 33

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.12 How to save to image and print

3.12.1 Save to JPG

1-1

In the graph region window, click the [Save to JPG] button.

1-2 Refer to [Figure 4]. Click the [Save to JPG] button in the lower

left of [Review] window.

2.

According to the path designated in the path item of [Configuration]

window, the image is stored by ‘.jpg’ format.

3.12.2 Print

1-1

In the graph region window, click the [Print] button.

Page 34: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 34

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

1-2 Refer to [Figure 4]. Click the [Print] button in the lower left of

[Review] window.

2

The [Print Preview] popup window comes out.

3 Click [Print] button.

3.12.3 Setting to JPG and Print

When saving to JPG or printing the stored data, information of each page and

graph line of 10 units is shown.

Page 35: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 35

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

<Detailed Grid>

According to the printing of [Configuration] window, you can change

the printing mode as ‘Landscape’ or ‘Portrait’. Also, if you check

‘Detailed Grid’, you can print graph line of 10 units

Page 36: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 36

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Note : When the connection is lost with the device, the reconnection is tried

automatically and it is marked with dotted line.

Note: If you have two DOP1, DOP2 graph, ‘offset 20’ applied like a device when

they save to JPG print only.

Page 37: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 37

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Chapter 4. BCM350N Uninstallation

Two ways to uninstall the BCM350N

- 1st

1. Run [uninst.exe].

2. The dialogue box comes out. Click [Yes] button.

- 2nd

1. Use the ‘Control panel’ of your desktop.

Page 38: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 38

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Chapter 5. Specification

BT-350 fetal monitor specification

Physical Chracteristics

Size 9.6 cm H x 32.6cm W x 27.6cm D

Weight About 5.5kg(except battery)

Safety

EN60601-1, EN60601-1-1, EN60601-1-2 Standard applied

Class I equipment & Internal powered equipment

Type BF applied parts

Continuous Operation

Power

Internal Battery

Ni-MH rechargeable battery

14 hours to full recharge during monitoring

14 hours to full recharge when monitor is

off

Using time : about 30 minutes(Dop 1/2

on[vol4], Print on[3cm/min, FM print on])

Power consumption : max. 10watt

External AC power AC 100~240V, 50/60Hz, 80VA maximum

Environmental

Operating temperature 10°C ~ 40°C (50°F ~ 104°F)

Storage temperature –20°C ~ 60°C (–4°F ~ 140°F)

Relative humidity 20% to 90%, non-condensing

Altitude 0 -3048m (0 -10,000 ft)

Page 39: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 39

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

FHR measuring

Item Value

Probe surface IR(SATA) ≤ 10 mW/cmP

2P

Incidence beam area 19.95 cmP

2P, circle type

Ultrasound frequency 0.985 MHz ±0.5%

Pulse lasting time 97.65625 microseconds

Pulse repeat frequency 3.2 kHz

Range of FHR measuring 30-240 BPM

Accuracy ±2% of range

Leakage current <10 µA @ 264 VAC applied to transducer

Voltage withstand >4 kV RMS, Type BF applied part

UC measurement

Item Value

UC Range 0-99, relative unit

Resolution 1 Count

Accuracy ±1% relative unit

Leakage current <10 µA @ 264 VAC applied to transducer

Voltage withstand >4 kV RMS, Type BF applied part

Paper

Pack style Z-Fold.

Pack size 150 mm x 90 mm x 15 mm

End of pack Mark along paper edge

Loading Side-door, slide-in

Paper Detector Paper out

Door open

Paper speed

Normal 1, 2, and 3 cm/min ±1%

High speed 10 cm/min (only in Trend

mode)

Paper tracking accuracy : ±1% (exclusive of paper accuracy)

Page 40: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 40

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Appendix 1. Wired communication system

1. How to making connection cable

1.1 In case of connecting a fetal monitor (BT-300, BT-350 etc.) to PC

directly

You can get RS-232C cross cable in the market. Both side of

connector is D-SUB9 (Female). Or you can make a cable with the

connection information as below.

(Please secure the enough length of cable considering setting place

and distance between PC and a fetal monitor.)

1.2 In case of connecting monitors to the central monitoring PC through

multi-port. The cable information is as follow;

RXD

TXD

GND

Page 41: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 41

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

2. Specification of Multi-port

Use the SystemBase Co.,Ltd.’s product - Multi-8/ Multi-8/USB Series.

SystemBase Co.,Ltd.

www.sysbas.com

Page 42: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 42

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Appendix 2. Wireless communication system

Table contents

1. Introduction of wireless communication system …………… 48

2. Checking IP address of installing PC ………………………… 48

3. How to set MSP-1000 IP address ………………………………… 51

4. How to set MSP-1000 ……………………………………………… 57

5. How to connect SD-1000 …………………………………………… 64

6. How to set Serial IP address ……………………………………… 70

Page 43: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 43

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

1. Introduction of wireless communication system

MSP-1000 is communication device. Using this device, it is possible to

use bluetooth connection through Ethernet. Also this device supports

simultaneous access of bluetooth device. Parani-SD1000 is a Bluetooth

serial adaptor. Using MSP-100 and SD-1000, it is possible to comprise

wireless communication system.

2. Checking IP address of installing PC

2.1. Pre-installation before checking IP address

2.1. Connect a DC 5V power adapter to MSP-1000. And connect

internet LAN cable to MSP-1000 “EHT0” port. And connect MSP-1000

“ETH1” port and PC LAN port using LAN cable. Using serial cable,

connect MSP-1000 and PC.

2.2. Checking IP address

2.2.1. Run [Run] as below.

Input [cmd] and click [OK].

Page 44: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 44

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

2.2.2. MS-DOS window comes out as below.

2.2.3. Input [ipconfig/all] and press [Enter]. Then below window comes out.

2.1.4. The red box in the upper window is IP address.

Make a note IP addresses in the blue box.

※ For example, make a note as follow.

1) IP Adress : 192. 168. 131. 65

2) Subent Mask : 255. 255. 255. 0

3) Default Gateway : 192. 168. 131. 254

4) DHCP Server : 192. 168. 131. 254

5) DNS Servers : 61. 97. 112. 155

61. 97. 113. 250

Page 45: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 45

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

2.3. Check whether IP address is occupied or not

To install MSP-1000, you should be allowed to use IP address. If you use IP

address without checking whether that is occupied or not, it may occurs the

communication problem because of IP collision. Therefore you must check

whether IP address is occupied or not as follows.

2.3.1. In the case of IP address from a hospital is [192. 168. 131. 200].

1) Input [ping 192.168.131.200] in the MS-DOS window.

2) And then press [Enter].

3) The results are as below.

4) In the case of IP address is occupied, you can check the message like

as “Reply from 192.168.10.65: byte=32 time<1ms TTL=128”. In the case

of IP address is not occupied, you can find the message like as

“ Request timed out.”. In the case of IP is not occupied, you can use IP

from a hospital.

IP is occupied.

IP is not occupied.

Page 46: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 46

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3. How to set MSP-1000 IP address

3.1. How to connect MSP-1000 and PC

3.1.1. Connect MSP-1000 and PC using a serial cable.

3.1.2. Run [Hyperterminal]. The path is [Start →All Programs→Accessories→

Communications→Hyper Terminal].

3.1.3. After run [Hyper Terminal], the window as below comes out.

Input name (all value is available) and click [OK].

Page 47: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 47

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

※ If [New connection] window does not come out and [Location Information]

window comes out, follow the next procedure.

Click 1 →2 →3.

3.1.4. Return to 3.1.3 and select COM port(in this manual, COM3 is used) for

[Connect using] and click [OK].

2

1

1

1

1

2

3

Page 48: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 48

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.1.5. [COM3 Properties] window comes out. Then follow the below procedure.

Click [Restore Defaults] and [OK]. Then the basic values are changed.

And click [OK] and complete this step.

<Before run [Resotre Defaults]> <After run [Resotre Defaults]>

3.1.6. When there is no sign in the hypertermial window, press enter. Then

logging in message comes out. Input “root” for ID and password. You must

input in small characters. After logging in, input “editconf” and press enter.

1

1 2

1 3

1

Page 49: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 49

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.1.7. If all steps are completed normally, the below messages come out.

3.1.8. Input ‘1’ and press enter. Repeat this step until the below messages

come out.

Page 50: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 50

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.1.9. After complete this step like as upper window, check ‘2. IP address’. If IP

address differs from IP address what you received from the hospital, input ‘2’

and press enter. Then input IP address what you received from the hospital and

press enter. You can change IP address through this step. And check items

from 3 to 6. If IP addresses are different, change IP address using the same

changing method of the item ‘2. IP address’.

3.1.10. After change MSP-1000 IP address, press [ESC] 3 times and return to

the initial window.

Input received IP address

= 192.168.10.200

Page 51: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 51

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

3.1.12. In this window, input ‘8’ and press [ENTER]. Then two questions come

out step by step. Input ‘Y’ twice.

3.1.13. Input ‘Editconf’ and press [ENTER]. Then check IP address.

3.1.14. If there is no problem in IP address, input ‘8’ and press [ENTER]. Then

two questions come out step by step. Input ‘Y’ twice. And disconnect serial

cable from MSP-1000 and PC.

Page 52: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 52

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

4. How to set MSP-1000

4.1. Set MSP-1000

4.1.1. Run Internet Explore.

4.1.2. Input IP address in the web browser and press [ENTER].

For example) Input 192.168.131.200 then the login box comes out.

4.1.3. Input ‘root’ for User ID and Password. Then the window such as below

comes out. In this window, you can find IP address and change IP

address.

Click [Save & apply] button.

Page 53: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 53

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

4.1.4. How to connect MSP-1000 and SD-1000.

Click 1 → 2 → 3.

4.1.5. Then the dialogue box comes out as below.

4.1.6. Click [Ok] and complete the configuration.

4.1.7. In this window, you can set the number of SD-1000. Click [Add] button.

1

2

3

Check after click 1.

Ok Cancel

Page 54: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 54

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

If you want to add more SD-1000, click [Add] button again.

4.1.8. The below is the case of the number of SD-1000 is 3. In this case, you

should click [Add] button three times.

4.1.9. In the upper window, you should change each MAC address. Follow

next steps.

4.1.10. When you select a SD-1000 what you want to change MAC address, the

Page 55: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 55

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

below window comes out. In this window, you should change MAC

address.(MAC address is printed in the SD-1000 label.)

4.1.11. Input SD-1000’s MAC address.

The values for the first 6 are default values. They are fixed to [00:01:95].

You should input the rest values.

00:01:95:08:ab:f4

Each SD-1000 has different MAC address.

Please check and input the right address.

Page 56: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 56

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

4.1.12. Input MAC address and click [Save & apply]. Repeat these steps

according to the number of SD-1000.

4.1.13. When you complete these steps, the dialogue box comes out. Then

click [OK] and save configurations.

4.2. How to set Port

4.2.1. Click [Network Service] and input the port number.

4.2.2. Click [Network service] and add port by clicking [Add] button.

OK

Page 57: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 57

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

4.2.3. In the upper window, click “Listening on 7000” and change port number.

4.2.4. MSP-1000’s “Local port number” is start from 7001.

Repeat these steps to change port number according to the number of

SD-1000. For example, the port numbers are 7002, 7003 and so on.

After change the port numbers, click [Save & apply] button and save

configurations.

4.2.5. After you click [Save & apply] button, click [Port configuration]. (Serial

Page 58: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 58

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Port Profile Port configuration) Then the below window comes out.

Check the MAC addresses are correct.

4.2.6. When you complete configurations, click [Apply changes] and [Logout].

Then exit program.

Page 59: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 59

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5. How to set SD-1000

5.1. Install SD-1000 program and configuration

5.1.1. Setup SD-1000 program.

Run Setup_sd1000_conf-v2.01.exe.

5.1.2. The below window comes out and click [Finish] button.

Page 60: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 60

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5.1.3. Run SD-1000 program.

5.1.4. After running SD-1000 program, check the communication status.

Connect ‘USB power cable’ to SD-1000. Then connect this cable to

PC‘USB-Port’. And connect SD-1000 to PC’s serial port. Check that

Power led is on and Connect led is blinking with green.

5.1.5. The default Baud Rate value of SD-1000 is 9600. Change the Baud Rate

to 115200 and click [OK].

Page 61: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 61

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5.2. How to initialize SD-1000

5.2.1. Click [Device Setting] in the main window.

5.1.2. To initialize SD-1000, click [Hard Reset] and [Apply].

Page 62: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 62

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5.1.3. After complete SD-1000 initialization, click [Connection].

And click [Search].

5.1.4. After searching process, click [MSP-1000-0] and [Connect].

1

2

Page 63: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 63

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5.1.5. If MSP-1000 is connected successfully, click [OK].

5.1.6. After complete previous steps, click [Disconnect] and [OK].

2

1

Connection completed

Connection dropped

Page 64: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 64

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

5.2. How to set SD-1000 connection

5.2.1. Click [Device Setting]. And follow the next step. Click 1 → 2 → 3 → 4.

5.2.2. Disconnect SD-1000 from PC. Then connect another SD-1000 to PC.

Follow clause from 5.1.4 to 5.2.1.

1

3

2

4

H/W flow control

Page 65: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 65

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

6. How set Serial IP

6.1. Install program

6.1.1. Run Setup_Sena_serialip-v4.8.4 and install this program.

6.1.2. Follow the below steps.

Page 66: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 66

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

6.1.3. Click [Finish] and complete installing.

6.2. How to set SD-1000 Port

6.2.1. Run the Serial/IP Control Panel software (Start → Programs → Sena →

Serial IP → Control Panel). And click [Select Ports].

Page 67: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 67

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

6.2.2. Select COM ports according to the number of SD-1000.

We recommend that you should select COM number under 20.

6.2.3. The below window shows that the method of setting IP address and port

number when two COM ports are used. Note that you must input IP

address which is received from a hospital and Port number. (Ex. IP

address from a hospital is 192.168.10.200 and SD-1000 port numbers

are 7001~700X.)

Page 68: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 68

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

6.3. Checking the connection status

6.3.1. Click [Configuration Wizard...] and check the connection status.

6.3.2. The below windows show that the normal connection status and

abnormal connection status.

To test the communication

status, click [Start] button.

SD-1000 must be connected

to the monitor.

Normal connection

Abnormal connection / Disconnected

Page 69: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 69

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Warranty Product name Central monitoring & Networking system

Model name BCM350N

Permission No.P

1) N/A

Serial No.

Warranty Period 2 years from purchasing date

Date of purchase

C

U

S

T

O

M

E

R

Hospital :

Address :

Name :

Contact :

Agent / Dealer

Manufacturer Bistos Co.,Ltd.

※ Thanks for your purchasing our product.

※ This product is manufactured and passed through strict quality control system and

inspection.

※ Compensation standard concerning repair, replacement, refund of the product

complies with “Consumer’s protection law” noticed by Economic Planning Dept..

※ If an abnormal condition is happened with normal use in the term of guarantee,

the service center will offer free repair service.

※ If there is problem in the warranty period, please inform us model name, serial

number, purchasing date, problem and so on.

Note1) ‘Permission No.’ is a identify number by Korean Food and Drug Administration.

Manufacturer Information

Manufacturer

Head office /

Manufacturing

site

7P

thP Fl. A Bldg., Woolim Lions Valley 5cha,

144-3, Sangdaewon-dong, Jungwon-gu,

Seongnam-si, Gyeonggi-do, Korea

Telephone 82-31-750-0340

Tele fax 82-31-750-0344

Web site www.bistos.co.kr

e-mail [email protected]

Page 70: Vente de materiel sage femme, gynécologue et materiel medical … LOGIVIEL BCM 350... · 2020. 4. 2. · BCM350N Operator’s Manual 4 P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

BCM350N Operator’s Manual 70

P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)

Contact to Technical support

Telephone: +82 31 750 0340

Fax: +82 31 750 0344

Bistos Co., Ltd.

7P

thP Fl. A Bldg., Woolim Lions Valley 5cha, 302,

Galmachi-ro, Jungwon-gu, Seongnam-si,

Gyeonggi-do, Korea

2TUwww.bistos.co.krU2T

[email protected]

Model Name: BCM350N

EC Representative: Medical Econet Gmbh

High-t'Park Mainstr. 6c-6d D-45768 Marl/Germany

Telephone: +49 2365 92 437 - 0

Fax: +49 2365 92 437 – 55